← Product Code [ITI](/productcode/ITI) · K220747

# Power Wheelchair (N5515B) (K220747)

_Zhejiang Innuovo Rehabilitation Devices Co.,Ltd · ITI · May 12, 2022 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K220747

## Device Facts

- **Applicant:** Zhejiang Innuovo Rehabilitation Devices Co.,Ltd
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** May 12, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The Power Wheelchair (N5515B) is a motor driven, indoor wansportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

## Device Story

Motor-driven power wheelchair for mobility-impaired or elderly users. Input: user manipulation of joystick controller. Transformation: controller signals drive motors; automatic electromagnetic brake system engages upon joystick release. Output: controlled movement (speed/direction) of wheelchair. Used in indoor/outdoor environments; operated by patient. Healthcare provider uses output (mobility) to assist patient independence. Benefits: provides seated mobility for individuals with limited physical function.

## Clinical Evidence

Bench testing only. No clinical or animal studies performed. Performance validated against ISO 7176 series (stability, brakes, energy, dimensions, strength, climatic, obstacle climbing, power/control systems, EMC) and ISO 10993 (biocompatibility).

## Technological Characteristics

Carbon fiber frame; 24V 12Ah Li-ion battery; brushless DC motors (250W x 2); PU solid tires; joystick-based electronic controller. Electromagnetic braking. Standards: ISO 7176 series (performance/safety), ISO 10993 (biocompatibility), IEC 60601-1-2 (EMC).

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Power wheelchair, PL00I ([K113463](/device/K113463.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 12, 2022

Zhejiang Innuovo Rehabilitation Devices Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China

Re: K220747

Trade/Device Name: Power Wheelchair (N5515B) Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: March 14, 2022 Received: March 14, 2022

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K220747

Device Name

Power Wheelchair (N5515B)

Indications for Use (Describe)

The Power Wheelchair (N5515B) is a motor driven, indoor wansportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officen Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### l. SUBMITTER

Name: Zhejiang Innuovo Rehabilitation Devices Co., Ltd.

Address: No.196 Industry Road, Hengdian Movie Zone, Dongyang, Zhejiang, China

Name of contact person: Leo Zheng

Telephone: +86 18358936043

Fax: +86-579-89327232

Date prepared: 2022-05-05

#### II. Device

Device trade name: Power Wheelchair (N5515B)

Classification name: Powered wheelchair

Regulation class: 2

Regulation number: 21CFR 890.3860

Panel: Physical Medicine

Product code: ITI

### III. Predicate device

K113463

Power wheelchair, PL00I

SUZHOU KD Medical Appliance Co. Ltd.

### IV. Device description

This Power wheelchair is a motor driven, indoor and outdoor transportation vehicle, which a device for assisting action handicapped people and disabled people to move. It is suitable for disabled people with mobility difficulties and elderly people.

The device consists of front wheel, drive wheel, frame, controller, motor, armrest, backrest, seat cushion, pedal, battery box and charger.

The device is powered by Li-ion Battery pack (24V 12Ah) with 15 Km (9.32 miles) range, which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet (100-240V, 50/60Hz) when the device is not in use.

{4}------------------------------------------------

The patient can activate the controller handle (joystick) to control the speed and direction of the wheelchair movement. In addition, when the patient releases the joystick, the joystick will return back to the central position and the wheelchair will be automatically stopped soon due to automatic electromagnetic brake system. Once the joystick is activated again move to other position, the wheelchair will be re-energized.

# V. Indication for use

The Power Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

# VI. Comparison of technological characteristics with the predicate device

| Attribute                                               | Subject device                                                                                                                                                                            | Predicate device                                                                                                                                                         | Discussion/<br>Conclusion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                            | Zhejiang Innuovo Rehabilitation Devices Co., Ltd.                                                                                                                                         | SUZHOU KD Medical Appliance Co. Ltd.                                                                                                                                     | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Proprietary name,<br>model                              | Power Wheelchair, N5515B                                                                                                                                                                  | power wheelchair, PL00I                                                                                                                                                  | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| 510(k) number                                           | K220747                                                                                                                                                                                   | K113463                                                                                                                                                                  | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device classification<br>name                           | Class II                                                                                                                                                                                  | Class II                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Classification<br>regulations                           | 21 CFR 890.3860                                                                                                                                                                           | 21 CFR 890.3860                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product code                                            | ITI                                                                                                                                                                                       | ITI                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Similarities                                            |                                                                                                                                                                                           |                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indication for use                                      | The Power Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | They are motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Intended user                                           | disabled people with mobility difficulties and elderly people                                                                                                                             | disabled or elderly person limited to a seated position                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Use condition                                           | indoor and outdoor use                                                                                                                                                                    | indoor and outdoor use                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Number of wheels                                        | 4, including two front wheels and two rear wheels                                                                                                                                         | 4, including two pivoting casters and two rear drive wheels                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Function of wheels                                      | Front wheels: driven                                                                                                                                                                      | two pivoting casters: driven                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Attribute                                               | Subject device                                                                                                                                                                            | Predicate device                                                                                                                                                         | Discussion/ Conclusion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                         | wheels suitable for rotation, acceleration, retrograde<br>Rear wheels: driving wheels to control the speed and direction                                                                  | wheels suitable for rotation, acceleration, retrograde<br>two rear drive wheels: driving wheels to control the speed and direction                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Movement control method                                 | By Joystick control                                                                                                                                                                       | By Joystick control                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Driving system                                          | Direct drive on the rear wheels                                                                                                                                                           | Direct drive on the rear wheels                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Brake system                                            | Automatic electromagnetic brake system                                                                                                                                                    | Intelligent regenerative Electromagnetic brake                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Braking distance                                        | $\leq1.5$ m                                                                                                                                                                               | Forward: 1.5m (59") at max speed                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Maximum safe operational incline degree                 | 9°                                                                                                                                                                                        | 9°                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Armrest                                                 | PU                                                                                                                                                                                        | PU                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Battery charger                                         | Off-board charger<br>Input: 100-240V, 50/60Hz, 1.5A,<br>Output: 24 Vdc, 2A;                                                                                                               | Off-board, Automatic Type<br>Input: 110-220 V / 50-60 Hz,<br>Output: 24 Vdc, 2A;                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Differences                                             |                                                                                                                                                                                           |                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Main frame material                                     | Carbon fiber material                                                                                                                                                                     | aluminum alloy                                                                                                                                                           | Different material used for frame, that such difference will not impact the safety and effectiveness of the subject device as the performance tests are conducted according to ISO 7176 series. The carbon fiber material is lighter and easy to carry.                                                                                                                                                                                                                                                                             |
| back cushion                                            | Polyester fabric                                                                                                                                                                          | PU foam covered by nylon fabric cloth                                                                                                                                    | different material used for parts in contact with user, which such differences will not impact the safety and effectiveness of the subject device as                                                                                                                                                                                                                                                                                                                                                                                |
| Attribute                                               | Subject device                                                                                                                                                                            | Predicate device                                                                                                                                                         | Discussion/Conclusion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| seat cushion                                            | rubber patch cloth and Oxford fabric                                                                                                                                                      | PU foam covered by nylon fabric cloth                                                                                                                                    | carried out according to ISO 10993 series. different material used for parts in contact with user, which such differences will not impact the safety and effectiveness of the subject device as biocompatibility tests are carried out according to ISO 10993 series.                                                                                                                                                                                                                                                               |
| Overall Dimension (length*width*height)                 | 940*610*960mm                                                                                                                                                                             | 880*570*890mm                                                                                                                                                            | larger size is designed for bearing more loading weight.                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Stowage Dimension (length*width*height)                 | 720*310*610mm                                                                                                                                                                             | 720*570*400mm                                                                                                                                                            | All safety and performance have been validated with the maximum rated weight dummy.                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Front wheel size/type                                   | 7" x 1.75"/PU Solid tire                                                                                                                                                                  | 6" x 2" /PU Solid tire                                                                                                                                                   | Minor difference on dimension of driven wheel will not cause different performance.                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Rear wheel size/type                                    | 8.5"x 2"/ PU Solid tire                                                                                                                                                                   | 8" x 2.4"/PU Solid tire                                                                                                                                                  | Minor difference on dimension of driving wheel will not cause different performance.                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Max speed forward                                       | Up to 6 km/h (1.6 m/s), adjustable                                                                                                                                                        | Up to 6 km/h (3.75 mph), variable                                                                                                                                        | minor difference on max. forwarding speed will not cause different performance. lower speed will be more safety.                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Max backward Speed                                      | Less than 3 km/h (0.5 m/s)                                                                                                                                                                | 2.4 mph (3.84 km/h)                                                                                                                                                      | lower speed on max. backward speed will be more safety.                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Max loading weight                                      | 136kg (≈300 lbs)                                                                                                                                                                          | 114 kg (251 lbs)                                                                                                                                                         | Difference on loading weight will not cause different performance. more loading weight provide more convenient and stable performance                                                                                                                                                                                                                                                                                                                                                                                               |
| Attribute                                               | Subject device                                                                                                                                                                            | Predicate device                                                                                                                                                         | Discussion/Conclusion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Battery                                                 | li-ion battery pack; rechargeable, 24 VDC 12Ah                                                                                                                                            | Li-ion, Rechargeable; 24 VDC 20Ah                                                                                                                                        | for the transportation. the battery capacity will impact the travel distance, which will not cause new safety and effectiveness concerns raised.                                                                                                                                                                                                                                                                                                                                                                                    |
| Maximum distance of travel on the fully charged battery | 15 km                                                                                                                                                                                     | 20 km                                                                                                                                                                    | Shorter driving distance will not cause new safety and effectiveness concerns raised.                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Motor                                                   | Brushless DC motor; 24VDC; 250W; 2pcs                                                                                                                                                     | Brushless DC motor; 24 VDC; 180 W; 2 pcs                                                                                                                                 | minor difference on motor power will not cause different performance. larger power will provide more driving force, no safety and effectiveness concerns raised.                                                                                                                                                                                                                                                                                                                                                                    |
| Electronic controller                                   | Brushless dual-drive rocker controller                                                                                                                                                    | Brushless dual-drive rocker controller                                                                                                                                   | Similar controller is used, both the control system, including the joystick controller, the electromagnetic brakes and the user interface are similar. The joystick controls the directions and speed of movement, and when the joystick is released, the powered wheelchair will slow down to stop and the brakes will automatically re-engage. The controller also provides the battery status displaying and abnormal condition displaying. Both of the control systems are evaluated according to standard ISO 7176-14:2008 and |
| Attribute                                               | Subject device                                                                                                                                                                            | Predicate device                                                                                                                                                         | Discussion/<br>Conclusion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                         |                                                                                                                                                                                           |                                                                                                                                                                          | software validation<br>requirement and there<br>are no new safety and<br>effectiveness concerns<br>due to the difference.                                                                                                                                                                                                                                                                                                                                                                                                           |
| Turning Radius                                          | 900 mm                                                                                                                                                                                    | 31.5" (800 mm)                                                                                                                                                           | The minor difference in<br>the turning radius is<br>caused by different size<br>of wheelchair and may<br>cause a little bit<br>inconvenience when it<br>turns in a narrow space<br>while the minor<br>difference will not raise<br>any new safety and<br>effectiveness concerns.                                                                                                                                                                                                                                                    |
| Maximum obstacle<br>climbing                            | 40 mm                                                                                                                                                                                     | 1.2" (30 mm)                                                                                                                                                             | Longer distance in the<br>obstacle climbing will not<br>impact the safety and<br>effectiveness of the<br>subject device.                                                                                                                                                                                                                                                                                                                                                                                                            |

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

## VII. Summary of substantial equivalence discussion

The power wheelchair complied with the requirements of ISO 7176-1:2014, ISO 7176-2:2001, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2018, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2008, ISO 7176-13:1989, ISO 7176-14:2008, ISO 7176-15:1996, ISO 7176-16:2012, ISO 7176-21:2009, ISO 7176-22:2014, ISO 7176-25:2013, IEC 60601-1-2: 2014, ISO 10993-1:2018, ISO10993-5:2009, ISO 10993-10:2010.

The intended uses for both devices are the same. Mainframes of two devices are folded by way of front and rear close, and frame materials all meet the Tensile Strength, Yield Load, and Elongation tests. The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14:2008. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3:2012. Maximum obstacle climbing and Maximum safe operational incline are slightly different while such differences will not impact the safety and 

{9}------------------------------------------------

effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2:2001, ISO 7176-10:2008. The biocompatibility of the Predicate device and Subject device meet the requirements of the ISO 10993-5:2009 & ISO 10993-10:2010.

The flame retardant test of the seat cushion/backrest of both subject device and predicate device is carried out according to the ISO 7176-16 test. Therefore, both devices are assured to be under the same safety level.

In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

### VIII. Summary of non-clinical testing

#### A Performance testing-bench

The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.

- . Software validation
- . ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
- . ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs
- . ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
- . ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
- . ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space
- . ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
- . ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
- . ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength

{10}------------------------------------------------

- . ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
- . ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
- . ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies
- . ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces.
- . ISO 7176-14:2008 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods
- . ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling.
- . ISO 7176-16:2012 Wheelchairs -- Part 16: Resistance to ignition of postural support devices
- . ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters
- ISO 7176-22: 2014 Wheelchairs Part 22: Set-up procedures .
- . ISO 7176-25:2013 Wheelchairs - Batteries and chargers for powered wheelchairs
- . Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2014.

#### A Biocompatibility of patient-contacting material

Biocompatibility Tests are carried out in accordance with ISO 10993-1: 2018, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2010) and irritation (ISO 10993-10:2010).

# IX. Summary of clinical testing

No animal study and clinical studies are available for our device. Clinical testing was not required

to demonstrate the substantial equivalence of the electric wheelchair to its predicate device.

#### X. Conclusions

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K113463.

---

**Source:** [https://fda.innolitics.com/device/K220747](https://fda.innolitics.com/device/K220747)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
