Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001)
K220657 · Zhejiang Rongsheng Industrial Co., Ltd. · FLL · Feb 10, 2023 · General Hospital
Device Facts
Record ID
K220657
Device Name
Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001)
Applicant
Zhejiang Rongsheng Industrial Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Feb 10, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Medical Infrared Thermometer is a non-sterile. reusable clinical thermometer intended for the intermittent determination of human body temperature in no touch mode on the center of the forehead as the measurement site on people of one month and above.
Device Story
Non-contact, hand-held, battery-powered infrared thermometer; measures human body temperature via infrared energy emitted from forehead. Optical system gathers infrared radiation, focuses it on a photoelectric detector, and converts it to an electrical signal. Internal algorithm, including target emissivity correction, processes signal to calculate temperature. Device features digital LCD, buzzer, and three-color backlight (green/yellow/red) based on temperature range. Used in home or clinical settings by patients or caregivers. Output provides temperature reading to assist in clinical decision-making. Includes memory function for 20 records. Powered by rechargeable Li-ion battery via USB.
Clinical Evidence
Clinical accuracy validation performed on 165 subjects across age groups (1 month to >5 years) per ISO 80601-2-56:2017+A1:2018. Comparator was a mercury axillary thermometer. Results confirmed compliance with ISO 80601-2-56 requirements. Bench testing included risk analysis (ISO 14971), electrical safety (IEC 60601-1, -1-2, -1-11), battery safety (IEC 62133-2), and biocompatibility (ISO 10993-5, -10).
Technological Characteristics
Infrared thermopile sensor; optical system with parabolic mirror; MCU-based signal processing. Materials: ABS housing, PMMA/silicone buttons. Rechargeable 3.7V Li-ion battery (300mAh). Connectivity: USB charging. Standards: ISO 80601-2-56, ASTM E1965-98, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62133-2, ISO 10993.
Indications for Use
Indicated for intermittent determination of human body temperature in no-touch mode on the center of the forehead for patients one month of age and older.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Braun No Touch + Forehead NTF3000 Thermometer (K163516)
K202111 — Medical infrared forehead thermometer · Hunan Honggao Electronic Technolofy Co., Ltd. · Feb 26, 2021
K193213 — Vicks VNT200 No Touch Forehead Thermometer · Kaz USA, Inc., A Helen of Troy Company · Apr 28, 2020
K201536 — Infrared Forehead Thermometer · Hangzhou Qingyuan Medical Equipment Technology Co., Ltd. · Feb 10, 2021
K221800 — Infrared Thermometer (Model: Y20001, Y20002) · Zhangzhou Easepal Medical Science and Technology Co., Ltd. · Dec 14, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
February 10, 2023
Zhejiang Rongsheng Industrial Co., Ltd. Wu Bentuan Chief Engineer 2 Jinheng Road Xiaoshun Town, Jinhua, Zhejiang 321035 China
Re: K220657
Trade/Device Name: Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: January 5, 2023 Received: January 12, 2023
## Dear Wu Bentuan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220657
Device Name
Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001)
Indications for Use (Describe)
The Medical Infrared Thermometer is a non-sterile. reusable clinical thermometer intended for the intermittent determination of human body temperature in no touch mode on the center of the forehead as the measurement site on people of one month and above.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K220657
# 510(k) Summary
### SUBMITTER
Name: Zhejiang Rongsheng Industrial Co., Ltd.
Address: NO. 2 Jinheng Road, Xiaoshun Town, Jindong District Jinhua City Zhejiang, CN 321035
Name of contact person: Wu Bentuan
Telephone: +86-579-82953477
Preparation date: 2/1/2023
## Device
Device trade name: Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001) Classification name: Clinical Electronic Thermometer Regulation class: II Regulation number: 21CFR 880.2910 Panel: General Hospital Product code: FLL
#### Predicative device
Predicate Submission Number: K163516
Predicate Device Trade Name: Braun No Touch + Forehead NTF3000 Thermometer
Classification name: Clinical Electronic Thermometer
Regulation class: II
Regulation number: 21CFR 880.2910
Panel: General Hospital
Product code: FLL
### Device description
The Medical Infrared Thermometer is a non-contact, hand-held, battery powered device designed to measure human body temperature, which has the feature of simple and easy use, rapid measurement and accurate temperature feedback. The Medical Infrared Thermometer is an infrared thermometer to measure human body temperature using the infrared energy emitted in the area around the subject's forehead. The reference body site of the temperature measurement is axillary. The temperature
{4}------------------------------------------------
measurement distance is 1-5 cm.
The product is composed of enclosure, control panel, LCD screen, infrared sensor, PCB board, battery pack and USB charging inlet.
The thermometer has three functions of digital LCD display, buzzer indication for measurement results and three color back light indication for temperature (32.0℃ to 37.3℃ with green back light; 37.4℃ to 38.0℃ with yellow back light; 38.1℃ to 43.0℃ with red back light).
The device is powered by Li-ion Battery (3.7VDC; 30mAh), which can be recharged through a computer with USB port or by switching power supply (100-240V, 50/60Hz) connected with a USB charging cable. The switching power supply is not provided with the subject device and the USB charging cable will be provided with the subject device as an accessory.
There are two models RST-TC59-1001 and RST-TC60-1001 included in this document. Both models share the same construction except differences on product structure including appearance of enclosure, PCB design, sensor, MCU and working condition (ambient temperature). There are three buttons on control panel of both models, the first button is for power on/off and temperature measurement function, the second button is for temperature unit shift (°C and °F) function, the third button is for memory function.
#### Indication for use
The Medical Infrared Thermometer is a non-sterile, reusable clinical thermometer intended for the intermittent determination of human body temperature in no touch mode on the center of the forehead as the measurement site on people of one month and above.
| Attribute | Subject device | Predicative device | Verdict |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Device trade name | Medical Infrared<br>Thermometer (Model: RST-TC59-1001, RST-TC60-1001) | Braun No Touch + Forehead<br>NTF3000 Thermometer | / |
| Product model | RST-TC59-1001, RST-TC60-1001 | NTF3000 | / |
| 510(k) number | K220657 | K163516 | / |
| Device classification name | Class II | Class II | Same |
| Classification | 880.2910 | 880.2910 | Same |
| Attribute | Subject device | Predicative device | Verdict |
| regulations | | | |
| Product code | FLL | FLL | Same |
| thermometer type | skin IR thermometers | skin IR thermometers | Same |
| Intended<br>use(s)/<br>Indication for use | The Medical Infrared<br>Thermometer is a non-sterile,<br>reusable clinical thermometer<br>intended for the intermittent<br>determination of human body<br>temperature in no touch mode<br>on the center of the forehead<br>as the measurement site on<br>people of one month and<br>above. | The Braun No Touch +<br>Forehead NTF3000<br>Thermometer is a non-sterile,<br>reusable clinical thermometer<br>intended for the intermittent<br>determination of human body<br>temperature in a touch and no<br>touch mode on the center of the<br>forehead as the measurement<br>site on people of all ages. | Similar |
| Operational principle | The optical system gathers<br>the infrared radiation energy<br>of the target in its field of<br>view, size of which depends<br>on the optical parts and their<br>positions. The gathered<br>infrared energy is focused on<br>the photoelectricity detector<br>and converted to be the<br>electrical signal accordingly.<br>After passing through the<br>amplifier and signal<br>processing circuit, the<br>electrical signal will be<br>converted to be the<br>temperature of the measured<br>target according to the<br>algorithm internally<br>contained in the device and<br>target emissivity correction. | The thermometer uses a<br>thermopile sensor with<br>integrated thermistor for the<br>target reading, a thermistor<br>mounted in the head of the<br>thermometer for ambient<br>temperature readings, a<br>parabolic mirror to help focus<br>the infrared energy emitted<br>from the forehead, and an<br>infrared proximity sensor for<br>detection of contact or non-<br>contact use and compensation<br>of the temperature reading. | Same |
| Sensor type | thermopile | thermopile | Same |
| signal processing and<br>display | MCU and LCD | MCU and LCD | Same |
| Low<br>power<br>indication | yes | yes | Same |
| Measurement site | forehead | forehead | Same |
| Operation mode | Adjusted mode | Adjusted mode | Same |
| °C/°F switchable | yes | yes | Same |
| Temperature<br>measurement range | 32 °C~43 °C (89.6 °F~109.4 | 34.4°C (93.9°F) - 42.2°C<br>(108°F) | Similar |
| Attribute | Subject device | Predicative device | Verdict |
| | °F) | | |
| Measuring accuracy | 35.0 °C~42.0 °C (95.0 °F<br>~107.6°F): ±0.2 °C(±0.4 °F); | 35-42 °C (95-107.6 °F):± 0.2 °C (± 0.4 °F);<br>Outside the range 35-42 °C<br>(95-107.6 °F): | Same |
| | 32.0 °C ~ 34.9 °C (89.6 °F<br>~94.8 °F): ±0.3 °C (±0.5°F); | ± 0.3 °C (± 0.5 °F) | |
| | 42.1 °C~43.0 °C (107.8 °F<br>~109.4 °F): ±0.3 °C (±0.5 °F) | | |
| | | | |
| Measurement<br>precision | 0.1°C (0.1°F) | 0.1°C (0.1°F) | Same |
| Electric Safety and<br>EMC | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 62133-2 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 | Different |
| Biocompatibility | Meets ISO 10993 and FDA<br>guidance "Use of<br>International Standard ISO<br>10993-1"<br>ISO 10993-5<br>ISO 10993-10 | Meets ISO 10993 and FDA<br>Bluebook memo G95-1<br>ISO 10993-5<br>ISO 10993-10 | Same |
| Performance | ASTM E1965-98<br>ISO 80601-2-56 | ASTM E1965-98<br>ISO 80601-2-56 | Same |
| Power requirements | rechargeable li-ion battery<br>pack; 3.7 V; 300 mAh;<br>recharged through USB port<br>by external power source, e.g.<br>compatible charger or<br>computer with USB port etc. | 2 AA batteries | Different |
| | | | |
| Patient<br>contact<br>materials | RST-TC59-1001: ABS with<br>colorants (white and green)<br>for device housing including<br>handle, PMMA for buttons | Patient contacting materials<br>include ABS (device housing /<br>handle and power button) and<br>TPR (temperature button and<br>nose / forehead touch bumper). | Similar |
| | RST-TC60-1001: ABS with<br>colorants (white and blue) for<br>device housing including<br>handle, PMMA for<br>Transparent LCD cover,<br>silicone rubber for buttons | | |
| Memory | 20 sets | Not available | Different |
| Attribute | Subject device | Predicative device | Verdict |
| Ambient temperature<br>environment<br>Operating condition | RST-TC59-1001: 15°C ~ 40°C (59 °F~104 °F); Relative humidity: ≤85%; non-condensing | 15 °C – 40 °C (59 °F –104 °F), 15–95% non-condensing | Similar |
| | RST-TC60-1001: 0°C ~ +40°C (32°F~104°F); Relative humidity: ≤85%; non-condensing | | |
| Ambient temperature<br>environment<br>Storage condition | RST-TC59-1001: temperature: -25°C ~ +50°C (-13 °F~+122 °F); Relative humidity: ≤90% non-condensing; Atmospheric pressure: 70kPa~106 kPa | temperature: -25°C to 60°C (-13°F to 140°F)<br>Humidity: 15–95% non-condensing<br>Pressure: 700-1060hPA | Similar |
| | RST-TC60-1001: -25°C ~ +70°C (-13°F~+158°F); Relative humidity: ≤90% non-condensing; Atmospheric pressure: 70kPa~106 kPa | | |
| Measurement<br>distance | 1cm ~ 5cm | 0-5cm | Similar |
| Response time | 1 second | 2 seconds | Different |
Comparison of technological characteristics with the predicate device
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Discussion for similarities and differences:
- 1. Intended use(s)/Indication for use: indication for use of the subject device has the similar meaning with different words as the predicate device. In addition, the subject device has more restricted people age range than the predicate device. Such difference will not affect the safety and effectiveness of the subject devices.
- 2. Temperature measurement range: the subject devices measurement range from 32 ℃~43 ℃ (89.6 °F~109.4 °F), which is different from the predicate device's measurement range, 34.4℃ ( 93.9F ) – 42.2℃ (108°F ). While the subject devices have been tested and validated according to standard ISO 80601-2-56 including clinical accuracy and the measuring range of the subject devices meet the minimum rated output range of clinical thermometer requirement, from 35°C to 42°C, therefore, such difference will not affect the safety and effectiveness of the subject devices.
{8}------------------------------------------------
- Electric safety and EMC: the batteries used in the subject device are rechargeable batteries, different 3. from the predicate device, and tested according to IEC 62133-2. Such difference between the predicate device and the subject device will not affect the safety and effectiveness of the subject device.
- Power requirements: the subject devices are powered with rechargeable li-ion battery pack and 4. different from the predicate device, which is powered with 2 AA batteries. The rechargeable battery pack has been tested according to IEC 62133-2: 2017, the differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
- Patient contact materials: there are some material differences between the subject devices and the 5. predicate device, but biocompatibility of both the subject devices and the predicate device has been evaluated and tested according to ISO 10993 series and FDA guidance. The patient contact materials have passed the biocompatibility testing in accordance with ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010. Such difference between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
- Memory function: The subject devices have the memory function to store 20 sets history 6. measurement records while the predicate device has not such memory function. The validation test of the software used in the subject device was conducted in accordance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The difference will not raise safety or effectiveness issues on the subject device.
- Ambient temperature environment operating condition: there are differences between ambient 7. temperature of the subject devices and the predicate device, but performance and safety tests of the subjective device have been conducted and passed under the specific operating condition according to standards IEC 60601-1, ISO 80601-2-56 and ASTM E1965-98, and the user manuals of the subjective devices provide the operating and storage condition as well. Therefore, such difference will not affect the safety and effectiveness of the subject devices.
- Ambient temperature environment Storage condition: Performance and safety tests of the 8. subjective device have been passed under the specific storage condition, and the user manual of the subjective device provides the storage condition as well, therefore, the difference of the storage conditions between subject device and predicate device will not affect the safety and effectiveness of subject device.
{9}------------------------------------------------
- Measurement distance: there is slight difference between the subject devices and the predicate device 9. on measurement distance and the subject devices have more restricted measurement range (1cm-5cm) than the predicate device (0cm-1cm). The performance tests of the subject device are conducted in accordance with standard ISO 80601-2-56 and ASTM E1965-98, and the test results showed that the measurement accuracy of the subject device is in conformity with the standard requirement. In addition, clinical accuracy testing of the subject devices was validated according to standard ISO 80601-2-56, therefore, such difference will not affect the safety and effectiveness of the subject devices.
- 10. Response time: there is slight difference between the subject devices and the predicate device on response time and the subject devices are faster in response than the predicate device. The performance tests of the subject device are conducted in accordance with standard ISO 80601-2-56 and such difference will not affect the safety and effectiveness of the subject devices.
#### Summary of non-clinical testing (Performance testing-bench)
The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.
- . Risk Analysis developed in accordance with ISO 14971:2019.
- . Software Evaluation according to Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software validation and verification tests are conducted at unit level, integration level and system level. The software and hardware system is verified that no defects are found.
- IEC 60601-1: 2005+A1:2012 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2: 2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances -Requirements and tests
- IEC 60601-1-11: 2015 Medical electrical equipment. Part 1-11: General requirements for basic ● safety and essential performance. Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 80601-2-56: 2017+A1: 2018 medical electrical equipment part 2-56: particular ●
{10}------------------------------------------------
requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- . ASTM E1965-98(2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
- . IEC 62133-2: 2017 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
- . ISO 10993-1: 2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
Biocompatibility Tests are carried out in accordance with ISO 10993-1: 2018, summary level biocompatibility information sees below table:
| Name of test | Endpoint | Result |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| In vitro<br>cytotoxicity test | Quantitative evaluation according to ISO 10993-5: there shall be no reduction of cell viability by more than 30 %, which is considered a cytotoxic effect. | Pass |
| Skin sensitization<br>test | According to ISO 10993-10: Magnusson and Kligman grades shall not be of 1 or greater in the test group, provided grades of less than 1 are seen in control animals. | Pass |
| Skin irritation test | According to ISO 10993-10: the primary irritation index (PII) determined on the test group shall be 0-0.4 | Pass |
#### Summary of clinical Accuracy Validation Test
Clinical accuracy validation for the Medical Infrared Thermometer is designed and performed according to standard ISO 80601-2-56: 2017+A1: 2018 and ISO 14155: 2011 under clinical setting.
Both models RST-TC59-1001 and RST-TC60-1001 are selected to implement clinical accuracy validation respectively. A blackbody radiator and a Platinum resistance thermometer are selected to validate the laboratory accuracy of the infrared thermometer under testing before and after the clinical trial. And total of 165 subjects are selected to temperature measurement with the infrared thermometer under testing (DUT) and the reference clinical thermometer (RCT) respectively. RCT used
{11}------------------------------------------------
in this clinical trial is mercury axillary thermometer (Geratherm Medical AG). The whole trial procedure and the clinical accuracy validation is in accordance with ISO 80601-2-56: 2017+A1: 2018.
The clinical tests were evaluated on 165 subjects and the thermometer was evaluated in three age groups including subgroup A1 and A2: A1 – one month up to three months; A2 - three months to one year; B older than one years and younger than five years; and C - older than five years old.
Subjects selected for each model in this clinical trial are: A1) group 29 b) group: 55 c) group: 55. Total number of subjects is 165.
The clinical performance test protocol and data analysis were conducted in accordance with the requirement of ISO 80601-2-56: 2017+A1: 2018. The test report showed the clinical performance of the subject devices complied with the requirement of ISO 80601-2-56: 2017+A1: 2018.
#### Conclusions
Based on comparison of the intended use, technological characteristics, applicable safety standards, verification and validation testing, the differences between Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001) and the predicate device do not raise new issues of safety and effectiveness. it can be concluded that Medical Infrared Thermometer (Model: RST-TC59-1001, RST-TC60-1001) is substantially equivalent to the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.