← Product Code [LOX](/productcode/LOX) · K220629

# Emerge Monorail PTCA Dilatation Catheter (K220629)

_Boston Scientific · LOX · Mar 24, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K220629

## Device Facts

- **Applicant:** Boston Scientific
- **Product Code:** [LOX](/productcode/LOX.md)
- **Decision Date:** Mar 24, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.5100
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (1.20 mm balloon models) are indicated as pre-dilatation catheters in the stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis). The Emerge Over-The-Wire (balloon models 1.50-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 1.50-5.00 mm) are indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Emerge Over-The-Wire (balloon models 2.00-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 2.00-5.00 mm) are also indicated for the postdelivery expansion of balloon expandable stents (bare metal and drug-eluting).

## Device Story

Emerge Monorail PTCA Dilatation Catheter is a sterile, single-use intravascular device; used by physicians in clinical settings to widen coronary artery blockages or dilate stents. Device consists of a shaft with a Pebax balloon at the distal tip. Operation involves inserting the catheter into the coronary artery/bypass graft; inflating the balloon to dilate the stenosis or stent; and deflating for removal. Output is mechanical dilation of the vessel or stent, improving myocardial perfusion. Benefits include restored blood flow and optimized stent apposition. This submission extends the existing balloon size matrix to include 4.50mm and 5.00mm diameters and 40mm lengths.

## Clinical Evidence

Bench testing only. Performance testing included effective length, crossing profile, balloon preparation/deployment/retraction, withdrawal, fatigue, inflation/deflation, coating integrity, and particulate evaluation. Additional data leveraged from predicate/reference devices included shaft dimensions, rated burst pressure, compliance, bond strength, flexibility, torque, radiopacity, shelf life, and biocompatibility.

## Technological Characteristics

Intravascular PTCA dilatation catheter; Pebax balloon; shaft construction. Sterilization method and materials are identical to predicate. Dimensions: 1.20-5.00 mm balloon diameters; 8-40 mm lengths. Connectivity: None (mechanical device).

## Regulatory Identification

Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.

## Predicate Devices

- Emerge™ Monorail™ PTCA Dilatation Catheter ([K163174](/device/K163174.md))

## Reference Devices

- Apex™ PTCA Dilatation Catheter ([P860019](/device/P860019.md)/S208)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 24, 2022

Boston Scientific Corporation Ms. Aoife O'Flaherty Senior Regulatory Affairs Specialist Ballybrit Business Park Galway, Ireland

### Re: K220629

Trade/Device Name: Emerge Monorail PTCA Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: March 3, 2022 Received: March 4, 2022

Dear Ms. O'Flaherty:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220629

Device Name Emerge Monorail PTCA Dilatation Catheter

#### Indications for Use (Describe)

The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (1.20 mm balloon models) are indicated as pre-dilatation catheters in the stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis).

The Emerge Over-The-Wire (balloon models 1.50-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 1.50-5.00 mm) are indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.

The Emerge Over-The-Wire (balloon models 2.00-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 2.00-5.00 mm) are also indicated for the postdelivery expandable stents (bare metal and drug-eluting)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary per 21 CFR §807.92

| Sponsor                         | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA                       |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Contact Name and<br>Information | Aoife O'Flaherty<br>Ballybrit Business Park,<br>Galway, Ireland.<br>Phone: 353-91-517299<br>Email: Aoife.OFlaherty@bsci.com |
| Date Prepared                   | March 3, 2022                                                                                                               |
| Proprietary Name                | Emerge™ Monorail™ PTCA Dilatation Catheter                                                                                  |
| Common Name                     | PTCA Dilatation Catheter                                                                                                    |
| Product Code                    | LOX                                                                                                                         |
| Classification                  | Class II, 21 CFR Part 870.5100                                                                                              |
| Predicate Device                | Emerge™ Monorail™ PTCA Dilatation Catheter, K163174,<br>cleared December 14, 2016                                           |
| Reference Device                | Apex™ PTCA Dilatation Catheter, P860019/S208, approved<br>November 7, 2008                                                  |

#### Device Description

The Emerge™ Monorail™ (MR) PTCA Dilatation Catheter is a sterile, single-use, intravascular medical devices used to widen coronary artery blockages and/or dilate coronary artery stents upon stent placement. The catheter consists of a shaft with a Pebax balloon near the distal tip and is available in a range of balloon sizes.

#### Indications for Use / Intended Use

The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (1.20 mm balloon models) are indicated as pre-dilatation catheters in the stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis).

The Emerge Over-The-Wire (balloon models 1.50-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 1.50-5.00 mm) are indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.

The Emerge Over-The-Wire (balloon models 2.00-4.00 mm) and Emerge Monorail PTCA Dilatation Catheters (balloon models 2.00-5.00 mm) are also indicated for the postdelivery expansion of balloon expandable stents (bare metal and drug-eluting).

{4}------------------------------------------------

## Comparison of Technological Characteristics

The proposed Emerge™ Monorail™ PTCA Dilatation Catheter sizes are substantially equivalent to the predicate Emerge™ Monorail™ PTCA Dilatation Catheter, cleared by FDA under premarket notification K163174, December 14, 2016. The proposed 4.50mm and 5.00mm diameter (for 8 - 20mm lengths) and 40mm length (for 2.00, 2.50, 3.00, 3.50, and 4.00mm diameter) Emerge™ Monorail™ balloon sizes serve as an extension to the cleared Emerge™ Monorail™ matrix of balloons 1.20 - 4.00mm in diameter and 8 - 30mm in length.

The proposed and predicate devices share the following technological characteristics:

- . Fundamental device design and intended use
- . Device materials
- Packaging materials and design ●
- Sterilization method ●
- Fundamental manufacturing processes .

The proposed and predicate devices differ in the following technological characteristic:

- . Balloon lengths and diameters

{5}------------------------------------------------

### Non-clinical Performance Data

Bench testing was performed to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed Emerge MR device has been designed and tested to assure conformance to the requirements for its intended use.

The following performance tests were completed on the proposed Emerge MR devices:

- Effective Length .
- Crossing Profile ●
- Balloon Preparation, Deployment and Retraction ●
- Withdrawal into a Guide Catheter ●
- Balloon Fatigue ●
- Balloon Inflation and Deflation ●
- Balloon Coating Integrity
- Particulate Evaluation .

The following tests were leveraged from the predicate/reference devices:

- Shaft Inner and Outer Diameter
- Balloon Rated Burst Pressure ●
- Balloon Rated Burst Pressure (in Stent) ●
- Balloon Fatique (select balloon sizes only) ●
- Balloon Fatigue (Repeat Balloon Inflations; in Stent) ●
- Balloon Compliance ●
- Catheter Bond Strength ●
- Tip Pull Test
- Flexibility and Kink Test ●
- Torque Strength
- Radiopacity .
- Shaft Coating Integrity ●
- Shelf Life
- Biocompatibility ●
- Sterilization

### Conclusion

Based on the indications for use, technological characteristics, and performance testing, the proposed Emerge™ Monorail™ PTCA Dilatation Catheter has been shown to be appropriate for its intended use and is considered to be substantially equivalent to Emerge™ Monorail™ PTCA Dilatation Catheter (K161374).

---

**Source:** [https://fda.innolitics.com/device/K220629](https://fda.innolitics.com/device/K220629)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
