K220628 · Waldemar Link GmbH & Co. KG · KRO · May 3, 2022 · Orthopedic
Device Facts
Record ID
K220628
Device Name
LINK Endo-Model EVO Knee System
Applicant
Waldemar Link GmbH & Co. KG
Product Code
KRO · Orthopedic
Decision Date
May 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3510
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LINK Endo-Model EVO Knee System is intended for mobility limiting diseases, fractures or defects of the knee joint, distal femur or proximal tibia which cannot be treated by conservative or osteosynthetic procedures. This device is intended for cemented use only unless a cementless modular stem is indicated for use. The Link OptiStems are for use with the following LINK knee femoral components, both intracondylar and distal femoral replacement versions: LINK Endo-Model SL, LINK Endo-Model EVO-M, and LINK Endo-Model EVO-W and are indicated for mobility limiting diseases, fractures or defects of the knee joint, distal femur or proximal tibia which cannot be treated by conservative or osteosynthetic procedures. The Link OptiStems cemented are intended for cemented use only. The Link OptiStems cementless are intended for cementless use only.
Device Story
Constrained total knee prosthesis; evolution of previously cleared LINK Endo-Model system. Available in Rotating Hinge and Non-rotating (Pure) Hinge designs; Standard (fixed stems) and Modular (taper connection) versions. Components include femoral/tibial implants, modular stems, adapters, fixation screws, and 3-peg patella. Used in orthopedic surgery for knee joint replacement; implanted by surgeons. Provides mechanical stability for mobility-limiting conditions; restores joint function; enables revision of complex bone defects. Compatible with existing LINK femoral/tibial cones and MEGASYSTEM-C components.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical performance testing, including range of motion analysis, wear analysis (ISO 14243-1/2), particle analysis (ISO 17853, ASTM F1877), tibial baseplate fatigue testing (ISO 14879, ASTM F1800), contact area/stress analysis, and fretting/corrosion testing (ASTM F1875).
Technological Characteristics
Materials: CoCrMo, Ti6Al4V (Tilastan), cPE, and optional TiNbN (PorEx) coating. Design: Constrained total knee prosthesis, rotating or pure hinge, fixed or modular stems with taper connections. Standards: ISO 14243-1/2 (wear), ISO 17853/ASTM F1875/F1877 (particle/corrosion), ISO 14879/ASTM F1800 (fatigue).
Indications for Use
Indicated for patients with primary/secondary osteoarthritis, rheumatoid arthritis, revision after total knee replacement, bone necroses, or varus/valgus deformity with stabilizer laxity. Pure/Rotating Hinge versions indicated for extreme deformities (20-30°), muscular deficiency, genu laxum, and oncological/revision surgery. OptiStems indicated for revision arthroplasty with bone defects, loosened femoral components, peri-/subprosthetic fractures, and tumor-related bone deficiencies.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Waldemar Link LINK Endo-Model Knee System (K143179)
Waldemar Link LINK Endo-Model Knee System with PorEx (TiNbN) coating (K152431)
Waldemar Link LINK Endo-Model Knee System with PorEx (TiNbN) coating (K212742)
Tibial and Femoral Cones of Waldemar Link GmbH & Co. KG (K200113)
Tibial and Femoral Cones of Waldemar Link GmbH & Co. KG (K201364)
Submission Summary (Full Text)
{0}------------------------------------------------
May 3, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that.
Waldemar Link GmbH & Co. KG % Terry Powell Regulatory Affairs LinkBio Corp 69 King Street Dover, New Jersey 07801
Re: K220628
Trade/Device Name: LINK Endo-Model EVO Knee System Regulation Number: 21 CFR 888.3510 Regulation Name: Knee Joint Femorotibial Metal/Polymer Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: KRO Dated: March 4, 2022 Received: March 4, 2022
Dear Terry Powell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220628
Device Name
Link Endo-Model EVO Knee System, and Link OptiStems
Indications for Use (Describe)
General Indications LINK Endo-Model EVO Knee System:
The LINK Endo-Model EVO Knee System is intended for mobility limiting diseases, fractures or defects of the knee joint, distal femur or proximal tibia which cannot be treated by conservative or os- teosynthetic procedures.
This device is intended for cemented use only unless a cementless modular stem is indicated for use.
Indications:
All LINK Endo-Model EVO components:
- · Primary and secondary osteoarthritis.
- Rheumatoid arthritis.
- · Revision after primary or revision total knee replacement.
- · Bone necroses which won't compromise the successful implantation of a hinged total knee endoprosthesis.
- · Varus and valgus deformity with contracture or laxity of the medial or lateral stabilizers.
- The LINK Endo-Model EVO Pure Hinge is additionally indicated for:
· Extreme cases of varus/valgus deformities (20-30°), theumatoid arthritis, muscular deficiency and any kind of genu laxum.
· Oncological and revision surgery in lower limb (in conjunction with the Endo- Model EVO -W and the Megasystem-C) The LINK Endo-Model EVO Rotating Hinge is additionally indicated for:
· Oncological and revision surgery in lower limb (in conjunction with the Endo- Model EVO - W and the Megasystem-C)
General Indications Link OptiStems:
The Link OptiStems are for use with the following LINK knee femoral components, both intracondylar and distal femoral replacement versions: LINK Endo-Model SL. LINK Endo-Model EVO-M. and LINK Endo-Model EVO-W and are indicated for mobility limiting diseases, fractures or defects of the knee joint, distal femur or proximal tibia which cannot be treated by conservative or osteosynthetic procedures.
The Link OptiStems cemented are intended for cemented use only. The Link OptiStems cementless are intended for cementless use only.
Indications:
· Revision arthroplasty due to juxta-articular bone defects.
- · Revision of loosened femoral prosthesis components by peri-/subprosthetic fracture.
- · Bone deficiencies, e.g. due to tumors, as well as in large post-revision and posttrauma segmental bone defects.
· Oncological and revision surgery in the area of the distal femur (in conjunction with Endo-Model SL Rotational and
Hinge Knee Prostheses or LINK Endo-Model EVO - M/ - W).
# Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| 510(k) Submitter: | Waldemar Link GmbH & Co. KG<br>Barkhausenweg 10<br>22339 Hamburg, Germany<br>Phone: +49-40-539950<br>Facility Registration #:3003386935 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Waldemar Link GmbH & Co. KG<br>Stefanie Fuchs (Regulatory Affairs)<br>Oststraße 4-10<br>Norderstedt, GERMANY 22844<br>Phone: +49-40 53995-530<br>Fax: +49-40 53995-174<br>E-Mail: st.fuchs@linkhh.de |
| Date Prepared: | March 3, 2022 |
| Trade Name: | LINK Endo-Model EVO Knee System |
| Common Name: | Knee Joint Replacement |
| Classification Name: | Knee joint femorotibial metal/polymer constrained cemented<br>prosthesis, 21 CFR §888.3510, product code KRO |
| Classification and Panel: | Class II, Orthopedic / 87 |
| Predicate Devices: | Primary Predicate<br>Waldemar Link LINK Endo-Model Knee System: K143179<br>Additional Predicates<br>Waldemar Link LINK Endo-Model Knee System with PorEx (TiNbN)<br>coating: K152431, K212742<br>Waldemar Link LinkSymphoKnee: K202924<br>Waldemar Link LINK MEGASYSTEM-C: K151008 |
| Device Description: | The LINK Endo-Model EVO Knee System is available in multiple<br>versions with different applications, characteristics, and materials.<br>The LINK Endo-Model EVO is a constrained total knee prosthesis,<br>which is based on the previously cleared LINK Endo-Model Knee<br>System (K143179, K152431, K212742), and is considered an<br>evolution ('EVO') of the previous designs. The Knee System is<br>available in two different hinge designs - Rotating Hinge and Non-<br>rotating (Pure) Hinge.<br>Also like the previously cleared system, the subject LINK Endo-<br>Model EVO is available in two versions - LINK Endo-Model EVO<br>Standard and LINK Endo-Model EVO - Modular. The femoral and |
{5}------------------------------------------------
### K220628 Page 2 of 4
tibial components of the standard version have fixed stems while the femoral and tibial components of the modular system are available with stems having a taper connection. The modular system includes femoral components with either a male taper or a female taper, and are designated as LINK Endo-Model EVO-M (male taper) and Endo-Model EVO-W ("weiblich" = female taper).
The LINK Endo-Model EVO - Standard consists of a cemented monoblock femoral component made from CoCrMo and cPE and a cemented monoblock tibial component made from CoCrMo and an articulating surface made from cPE. The Standard System is available in Rotating Hinge or Pure Hinge.
The LINK Endo-Model EVO – M consists of a cemented modular femoral component made from CoCrMo, Ti6A14V (Tilastan) and cPE and a cemented modular tibial component made from CoCrMo, Ti6Al4V (Tilastan) and an articulating surface made from cPE and with modular stems, cemented made from CoCrMo or cementless made from Ti6Al4V (Tilastan). The Modular System is available in Rotating Hinge or Pure Hinge.
The LINK Endo-Model EVO – W intracondylar and condylar replacements consists of a cemented modular femoral component made from CoCrMo, Ti6Al4V (Tilastan) and cPE and a cemented modular tibial component made from CoCrMo, Ti6Al4V (Tilastan) and an articulating surface made from cPE and with modular stems, cemented made from CoCrMo or cementless made from Ti6Al4V (Tilastan). The Modular System is available in Rotating Hinge or Pure Hinge.
The femoral components can be used with distal and femoral (straight and L-shaped) segments made from Ti6A14V (Tilastan).
The modular tibial components can be used with tibial spacers made from Ti6Al4V (Tilastan).
The femoral and tibial components Rotating Hinge versions are also available in LINK PorEx (TiNbN) coated version.
Additionally a 3-peg patella can be used with the LINK Endo-Model EVO Knee System. It is made from cPE.
The LINK Endo-Model EVO Knee System femoral and tibial components are compatible with the previously cleared LINK Endo-Model Knee System (K143179, K152431 and K212742).
The femoral components of the LINK Endo-Model EVO – W are compatible with the previously cleared LINK MEGASYSTEM-C (K151008).
{6}------------------------------------------------
The LINK Endo-Model EVO Knee System is compatible with previously cleared Tibial and Femoral Cones of Waldemar Link GmbH & Co. KG (K200113 and K201364).
The Link OptiStems are an expansion of the modular stem portfolio. The Link OptiStems consist of a modular stem, adapter and fixation screw. The Link OptiStems come in cemented and cementless version. The Link OptiStems have to be joined and implanted in combination with the Femoral Components of LINK Endo-Model SL (K151008) or LINK Endo-Model EVO – M / -W.
#### Indications for Use: General Indications LINK Endo-Model EVO Knee System: The LINK Endo-Model EVO Knee System is intended for mobility limiting diseases, fractures or defects of the knee joint, distal femur or proximal tibia which cannot be treated by conservative or osteosynthetic procedures.
This device is intended for cemented use only unless a cementless modular stem is indicated for use.
### Indications:
All LINK Endo-Model EVO components:
- . Primary and secondary osteoarthritis.
- Rheumatoid arthritis. ●
- Revision after primary or revision total knee replacement.
- Bone necroses which won't compromise the successful implantation of a hinged total knee endoprosthesis.
- Varus and valgus deformity with contracture or laxity of the medial or lateral stabilizers.
The LINK Endo-Model EVO Pure Hinge is additionally indicated for:
- Extreme cases of deformities (20-30°). rheumatoid arthritis, muscular deficiency and any kind of genu laxum.
- Oncological and revision surgery in lower limb (in conjunction with the Endo- Model EVO -W and the Megasystem-C)
The LINK Endo-Model EVO Rotating Hinge is additionally indicated for:
- . Oncological and revision surgery in lower limb (in conjunction with the Endo- Model EVO -W and the Megasystem-C)
General Indications Link OptiStems:
The Link OptiStems are for use with the following LINK knee femoral components, both intracondylar and distal femoral
{7}------------------------------------------------
| replacement versions: LINK Endo-Model SL, LINK Endo-Model | | | |
|----------------------------------------------------------------------------|--|--|--|
| EVO-M, and LINK Endo-Model EVO-W and are indicated for | | | |
| mobility limiting diseases, fractures or defects of the knee joint, distal | | | |
| femur or proximal tibia which cannot be treated by conservative or | | | |
| osteosynthetic procedures. | | | |
The Link OptiStems cemented are intended for cemented use only. The Link OptiStems cementless are intended for cementless use only.
## Indications:
- Revision arthroplasty due to juxta-articular bone defects.
- Revision of loosened femoral prosthesis components by peri-● /subprosthetic fracture.
- Bone deficiencies, e.g. due to tumors, as well as in large postrevision and posttrauma segmental bone defects.
- . Oncological and revision surgery in the area of the distal femur (in conjunction with Endo-Model SL Rotational and Hinge Knee Prostheses or LINK Endo-Model EVO - M/ -W).
Comparison to the predicate: The LINK Endo-Model EVO Knee System is substantially equivalent to the commercially available device LINK Endo-Model Knee System in that all have similar indications, design, materials and mechanicals safety. All devices are intended for cemented use only.
| Performance Testing: | Non-clinical performance testing and analysis were provided, including:<br>Range of Motion analysis Wear analysis (ISO 14243-1 and -2) Particle analysis (ISO 17853 and ASTM F1877) Tibial baseplate component fatigue testing (ISO 14879 and ASTM F1800) Contact area/stress analysis Fretting / Corrosion (ASTM F1875) The results of non-clinical performance testing demonstrate that the device is suitable for its intended purpose and substantially equivalent to the predicates. |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Testing: | Clinical performance testing was not required to demonstrate the substantial equivalence of this device. |
| Conclusion: | The subject LINK Endo-Model EVO Knee System is substantially equivalent to the predicate devices identified in this premarket notification |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.