K220614 · Chirana T. Injecta · FMI · Jun 15, 2022 · General Hospital
Device Facts
Record ID
K220614
Device Name
MEDOJECT fine Pen Needles
Applicant
Chirana T. Injecta
Product Code
FMI · General Hospital
Decision Date
Jun 15, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MEDOJECT™ fine Pen Needles for single use are intended for subcutaneous delivery of insulin in conjunction with injection pens.
Device Story
MEDOJECT™ fine Pen Needles are sterile, single-use, hypodermic needles designed for subcutaneous insulin delivery. The device consists of a stainless steel cannula, polypropylene hub, protective caps, and a paper seal. It is intended for use by patients or clinicians in home or clinical settings. The user attaches the needle to a compatible injection pen (ISO 11608-1 compliant) and manually inserts the needle tip into the skin to deliver the insulin dose. The device facilitates drug administration by providing a sterile, sharp, and compatible interface between the pen injector and the patient's subcutaneous tissue. The needle caps ensure sterility and protection until the point of use. The device is available in various gauges (29G-32G) and lengths (4mm-12.7mm) to accommodate different patient needs.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including conformance to ISO 7864 (sterile needles), ISO 11608-2 (needle-based injection systems), and ISO 9626 (stainless steel tubing). Biocompatibility testing was performed per ISO 10993-1, including cytotoxicity, sensitization, irritation, systemic toxicity, material-mediated pyrogenicity, and hemocompatibility. Sterility, shipping, and shelf-life were validated per ISO 11135, ASTM D4169, and ASTM F1980.
Technological Characteristics
Cannula: 304 stainless steel; Hub/Outer Cap: polypropylene; Inner Cap: polyethylene; Joint: medical grade glue; Lubricant: medical grade silicone; Seal: medical paper. Sterilization: Ethylene Oxide (SAL 10⁻⁶). Connectivity: Mechanical screw-thread attachment per ISO 11608-1. Dimensions: 29G-32G gauges; 4mm-12.7mm lengths.
Indications for Use
Indicated for subcutaneous delivery of insulin in patients requiring insulin therapy using injection pens.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Chirana T. Injecta
% Nathan Wright
Engineer and Regulatory Specialist
Empirical Testing Corp.
4628 Northpark Drive
Colorado Springs, Colorado 80918
Re: K220614
Trade/Device Name: MEDOJECT fine Pen Needles
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle
Regulatory Class: Class II
Product Code: FMI
Dated: May 11, 2022
Received: May 13, 2022
Dear Nathan Wright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K220614 - Nathan Wright
Page 2
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Courtney Evans -S
Digitally signed by Courtney Evans -S
Date: 2022.06.15 15:32:31 -04'00'
For CAPT Alan Stevens
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{2}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K220614 | |
| Device Name MEDOJECT™ fine Pen Needles | |
| Indications for Use (Describe) MEDOJECT™ fine Pen Needles for single use are intended for subcutaneous delivery of insulin in conjunction with injection pens. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection including suggestions for reducing this burden to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{3}
K220614 - 510(K) SUMMARY
| Preparation Date: | March 2, 2022 |
| --- | --- |
| Submitter’s Name: | CHIRANA T. Injecta, a.s. |
| Submitter’s Address: | Nám. Dr. Schweitzera 194
Stará Turá 916 01
Slovak Republic |
| Submitter’s Telephone: | 421-32-770-9972 |
| Contact Person: | Nathan Wright MS
Empirical Testing Corp.
1-719-351-0248
nwright@empiricaltech.com |
| Trade or Proprietary Name: | MEDOJECT™ fine Pen Needles |
| Classification Name: | Hypodermic single lumen needles |
| Classification: | Class II per 21 CFR §880.5570 |
| Product Code: | FMI |
| Classification Panel: | Drug Delivery and General Hospital Devices, and Human Factors (DHT3C) |
Predicate Device: K213407, UltiCare Disposable Pen Needles, UltiMed Incorporated
## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The MEDOJECT™ fine Pen Needles consist of a cap, needle cap, needle, hub, and paper. The cap and paper function to maintain the sterility of the needle. The device is single use and supplied sterile (ethylene oxide sterilized). The hub is connected to the injection pen for manual subcutaneous injection of appropriate does to the user. The needle cap protects the needle. The available size options are shown below.
| Needle Size | Color Code | | Item Code |
| --- | --- | --- | --- |
| 29G x 10mm | | Yellow | CHPN2910 |
| 29G x 12.7mm | | Pink | CHPN29127 |
| 30G x 6mm | | Dark Blue | CHPN3006 |
| 30G x 8mm | | Light Blue | CHPN3008 |
| 31G x 4mm | | Green | CHPN3104 |
| 31G x 5mm | | Dark Violet | CHPN3105 |
| Needle Size | Color Code | | Item Code |
| --- | --- | --- | --- |
| 31G x 6mm | | Dark Blue | CHPN3106 |
| 31G x 8mm | | Light Blue | CHPN3108 |
| 32G x 4mm | | Green | CHPN3204 |
| 32G x 5mm | | Dark Violet | CHPN3205 |
| 32G x 6mm | | Dark Blue | CHPN3206 |
| 32G x 8mm | | Light Blue | CHPN3208 |
## INDICATIONS FOR USE
The MEDOJECT™ fine Pen Needles are single use and are intended for subcutaneous delivery of insulin in conjunction with injection pens.
| Characteristic | Subject Device
MEDOJECT™ fine Pen Needles | Predicate Device
UltiCare Disposable Pen Needles, UltiMed Incorporated (K213407) |
| --- | --- | --- |
| Indications for Use | MEDOJECT™ fine Pen Needles for single use are intended for subcutaneous delivery of insulin in conjunction with injection pens. | The UltiCare™ Disposable Pen Needles are used with pen injector devices for the subcutaneous injection of drugs. |
| Rx or OTC | Over the Counter | Over the Counter |
## Discussion of differences in Indications for Use:
There are editorial differences to the indications for use statement between the subject and predicate device which do not change the indications. The predicate generically indicates the pen needles for drug delivery while the subject limits the delivery to only insulin, which is within the predicate indication.
CHIRANA T. Injecta MEDOJECT™ fine Pen Needles
{4}
K220614 - 510(k) Summary
# TECHNOLOGICAL CHARACTERISTICS
The table below includes a comparison of the technological characteristics between the subject and the predicate devices:
| | MEDOJECT™ fine Pen Needles | UltiCare Disposable Pen Needles | Comparison |
| --- | --- | --- | --- |
| 510(k) Number | Subject | K213407 | |
| Manufacturer | CHIRANA T. Injecta, a.s. | UltiMed Incorporated | |
| Regulation Number | 21 CFR § 880.5570 | 21 CFR § 880.5570 | Same |
| Classification Name | Hypodermic Single Lumen Needle | Hypodermic Single Lumen Needle | Same |
| Regulatory Class | Class II | Class II | Same |
| Product Code | FMI | FMI | Same |
| Operating Principle | The user manually inserts the needle tip into the skin to deliver insulin to the patient through a pen injector to which the pen needle is attached. | The user manually inserts the needle tip into the skin to deliver drug to the patient through a pen injector to which the pen needle is attached. | Same |
| Needle Gauge | 29G, 30G, 31G, 32G | 29G, 30G, 31G, 32G | Same |
| Needle Length¹ | 4mm, 5mm, 6mm, 8mm, 10mm, 12.7mm | 4mm, 5mm, 6mm, 8mm, 12.7mm | Similar |
| Material² | Cannula: 304 stainless steel
Needle hub: polypropylene
Needle Outer Cap: polypropylene
Needle Inner Cap: polyethylene
Joint medium: Medical grade glue
Lubricant: medical grade silicone
Seal: medical paper | Cannula: 304 stainless steel
Needle hub: polyethylene
Needle Outer Cap: Unknown
Needle Inner Cap: Unknown
Joint medium: Unknown
Lubricant: Unknown
Seal: Unknown | See Note |
| Needle Connector Type | Compatible with pen injectors which comply to ISO 11608-1 with screw threading attachment. | Compatible with pen injectors which comply to ISO 11608-1 with screw threading attachment. | Same |
| Design³ | Outer Cap; Inner Cap; Cannula;
Hub; Paper Seal | Protective Outer Container; Shield; Cannula; Hub | Similar |
| Biocompatibility | Per ISO 10993 | Per ISO 10993 | Same |
| Sterility | Ethylene Oxide to SAL of 10⁻⁶ | Ethylene Oxide to SAL of 10⁻⁶ | Same |
| Shelf Life⁴ | 5 years | Unknown | See Note |
¹The subject offers an additional needle length (10mm) not offered by the predicate to accommodate a wider range of patient needs. This difference does not affect the safety and effectiveness of the subject because all subject needle lengths are according to the size specifications in ISO 11608-2.
²Subject and predicate materials are similar but not identical. Differences in hub materials or in joint medium, lubricant, and seal do not introduce concerns for safety and efficacy of the subject device. Biocompatibility testing and performance testing confirmed that all materials of subject are safe and effective for use.
³The subject and predicate have the same needle, hub, and outer and inner cap design. The use of a paper seal and its material in the predicate is unknown. Differences in paper seal do not affect safety and effectiveness of the subject device. Performance testing confirmed that the subject design was safe and effective for use.
⁴Differences in shelf life would have no effect on the safety and effectiveness of the subject device. The shelf life of the sterile subject device was established through shelf life testing.
# PERFORMANCE DATA
The MEDOJECT™ fine Pen Needles described in this summary were tested and demonstrated to be in conformance with the following FDA recognized standards:
- ISO 7864 4th Ed-2016, Sterile hypodermic needles for single use – Requirements and test methods
- Limits for extractable metals
CHIRANA T. Injecta MEDOJECT™ fine Pen Needles
{5}
K220614 - 510(k) Summary
- ISO 11608-2 2nd Ed-2012, Needle-based injection systems for medical use – Requirements and test methods – Part 2: Needles
- Flow rate through rate
- Bond between hub and needle tube
- Functional compatibility with insulin injectors
- Ease of assembly and disassembly
- ISO 9626 2nd Ed-2016, Stainless steel needle tubing for the manufacture of medical devices – Requirements and test methods
- Needle sharpness – penetration force
- Limits of acidity and alkalinity
- Stiffness
- Resistance to breakage
- Resistance to corrosion
## Biocompatibility
In accordance with ISO 10993-1, the pen needle is classified as Externally Communicating Device, Blood Contact Indirect, Prolonged Contact (24 hours to 30 days). The following testing was conducted:
- Cytotoxicity (ISO 10993-5)
- Sensitization (ISO 10993-10)
- Irritation/Intracutaneous reactivity (SIO 10993-10)
- Subacute/Sub-chronic Toxicity (ISO 10993-11)
- Acute Systemic Toxicity (ISO 10993-11)
- Material-Mediated Pyrogenicity (ISO 10993-11/USP <151>)
- Hemocompatibility (ASTM F756)
## Sterility, Shipping, and Shelf-Life
- EO sterilization validation (ISO 11135)
- Simulated Packaging (ASTM D4169)
- Accelerated aging for shelf-life (ASTM F1980)
## CONCLUSION
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The MEDOJECT™ fine Pen Needles are substantially equivalent to the UltiCare Disposable Pen Needle (K213407) with respect to the indications for use, target populations, treatment method, and technological characteristics.
CHIRANA T. Injecta MEDOJECT™ fine Pen Needles
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.