CoolTone

K220601 · Zimmer Medizinsysteme GmbH · NGX · Apr 1, 2022 · Physical Medicine

Device Facts

Record IDK220601
Device NameCoolTone
ApplicantZimmer Medizinsysteme GmbH
Product CodeNGX · Physical Medicine
Decision DateApr 1, 2022
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 890.5850
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K220601 · Apr 1, 2022CoolToneZimmer Medizinsysteme GmbHPost-market surveillance activityInformation obtained from post-market surveillance was used to update the user's manual for the device.post-market surveillance; labeling update

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Post-market surveillance activity; Post-market surveillance—Not applicable for this studyDevice performance and safety information

Indications for Use

CoolTone is indicated to be used for: - Improvement of abdominal tone, strengthening of the abdominal muscles, development for firmer abdomen. · Strengthening, toning and firming of buttocks and thighs.

Device Story

CoolTone is a non-invasive, mobile, powered muscle stimulator used in clinical settings by healthcare providers. It generates a magnetic field via two large applicators to induce muscle contractions. The device features a 12-inch color touch screen that guides the user through treatment parameters (frequency, time, intensity) for abdomen, buttocks, or thigh protocols. The system uses a securement system to maintain applicator position. By initiating nerve action potentials, the device causes muscle contraction, which helps strengthen, tone, and firm targeted muscle groups. The device includes a compressor cooler and optimized pulsing sequences to manage heat and noise. The healthcare provider operates the device to deliver therapy, with the output directly affecting muscle conditioning. Benefits include improved muscle tone and firmness in the treated areas.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (ANSI AAMI ES 60601-1), electromagnetic compatibility (IEC 60601-1-2), software verification/validation, and performance testing confirming magnetic field intensity (0.5–1.35T) and thermal safety.

Technological Characteristics

Powered muscle stimulator; magnetic field energy source (0.5-1.35T); symmetrical biphasic sine wave; 1-150 Hz frequency; 370 µs pulse duration. Materials: steel and injection-molded plastics. Features: 12-inch touch screen, two applicators, compressor cooling system. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-10, ISO 62304, ISO 14971, and ISO 10993 standards.

Indications for Use

Indicated for individuals seeking improvement of abdominal tone, strengthening of abdominal muscles, development of a firmer abdomen, and strengthening, toning, and firming of buttocks and thighs.

Regulatory Classification

Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. April 1, 2022 Zimmer MedizinSysteme GmbH % Scott Blood Principal Consultant Quality and Regulatory Services 151 Glensondale Road Stow, Massachusetts 01775 Re: K220601 Trade/Device Name: CoolTone Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: March 1, 2022 Received: March 2, 2022 Dear Scott Blood: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Heather Dean, PhD Assistant Director THT5B3: Acute Injury Devices DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K220601 Device Name CoolTone Indications for Use (Describe) CoolTone is indicated to be used for: - Improvement of abdominal tone, strengthening of the abdominal muscles, development for firmer abdomen. · Strengthening, toning and firming of buttocks and thighs. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------| | <span style="text-decoration: underline;"><b> Prescription Use (Part 21 CFR 801 Subpart D) </b></span> | <span style="text-decoration: underline;"><b> Over-The-Counter Use (21 CFR 801 Subpart C) </b></span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z" and "i". Below the word "zimmer" is the phrase "MedizinSysteme" in a smaller, sans-serif font. The logo is simple and modern, and the use of a bold font makes it easily recognizable. # 6. 510(K) SUMMARY This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | I. SUBMITTER: | Zimmer MedizinSysteme GmbH | |------------------------|------------------------------------------------------| | | Junkersstrasse 9 | | | 89231 Neu-Ulm, Germany | | | Establishment Registration: 8010720 | | | Ms. Julia Kredler | | | Manager, Regulatory Affairs | | | Phone: +49-731-9761-625 | | | Fax: +49-731-9761-118 | | | E-mail: j.kredler@zimmer.de | | CONTACT: | Scott Blood | | | Principal Consultant | | | Phone: 978-729-5978 | | | Fax: +49-731-9761-118 | | | Email: scottqara@gmail.com | | DATE PREPARED: | March 31, 2022 | | II. DEVICE: | | | TRADE NAME: | CoolTone | | COMMON NAME: | Powered Muscle Stimulator | | CLASSIFICATION NAME: | Stimulator, Muscle, Powered, For Muscle Conditioning | | DEVICE CLASSIFICATION: | Class II, 21 CFR §890.5850 | | PRODUCT CODE: | NGX | | III. PREDICATE DEVICE: | CoolTone (K192940) | | | | ## IV. DEVICE DESCRIPTION: The CoolTone system is a non-invasive, therapeutic device. The device produces an electromagnetic field that interacts with the tissues of the human body. By muscle stimulation, the CoolTone helps to strengthen, tone and firm the abdomen, buttocks and thighs. {4}------------------------------------------------ The device is a mobile, standalone unit with four wheels and the control unit housing protects the patient from electrical shock and mechanical injury.. These features are unchanged. Two large applicators are connected to the control unit and can be used simultaneously depending on the treatment. CoolTone is a medical device that generates a magnetic field by applying a strong current to an applicator. It is equipped with the securement system, designed to maintain the position of the applicator throughout treatment. A large color touch screen facilitates the use of the device. The on-screen information guides the user, step-by-step, through the entire treatment process. The treatment is operated through variable parameters such as frequency, time and intensity. Three, pre-set treatment options are available for users to choose from: Abdomen, buttocks and thighs. These features are unchanged. #### V. INDICATION FOR USE: CoolTone is indicated to be used for: - Improvement of abdominal tone, strengthening of the abdominal muscles, development ● for firmer abdomen. - . Strengthening, toning and firming of buttocks and thighs. ## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE: The technological characteristics and operating principles associated with the treatment remain unchanged from the predicate device. The subject device is identical to the primary predicate device, the CoolTone system, as cleared under K192940. In comparison with the predicate device, the following changes have been made on the subject device: Hardware: Control unit: - A compressor cooler was added to the cooling circuit to enhance cooling ● performance; - . The single diode was replaced by a double diode on the capacitor stage to reduce the potential for treatment interruption; - . Foam tape was added to the compressor module to reduce noise; Applicator: - . Design and manufacture process improvement was made to the applicator to prevent cracking and the leaking of coolant fluid; #### Software: - . Modified to allow customers limited access to a device check and an advanced customer screen to facilitate troubleshooting; {5}------------------------------------------------ - . Implemented a treatment protocol that was within the previously cleared range of software specification; - Adjusted the coolant flow, implemented relax-sequences and extended cooldown to reduce the potential interruption of treatment incurred by applicator warmth; - Optimized pulsing sequence to reduce noise; Labeling: - The user's manual was updated to capture information obtained in post-market . surveillance activity. The technological characteristics of the subject device, as outlined in the comparison table below, remain the same as those of the predicate device. | Device Name | Predicate Device<br>CoolTone K192940 | Subject<br>Device<br>CoolTone | |-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------| | Company Name | Zimmer MedizinSysteme GmbH | Same | | Product Code and<br>Regulation | Physical Medicine 21 CFR 890.5850<br>NGX-Stimulator, Muscle, Powered, Muscle<br>Conditioning | Same | | Intended Use | Improvement of abdominal tone,<br>•<br>strengthening of the abdominal muscles,<br>development of firmer abdomen.<br>Strengthening, Toning and Firming of<br>•<br>buttocks and thighs. | Same | | Principle of Action | Initiating action potential of nerves results in<br>muscle contraction | Same | | Clinical Use | Prescription use | Same | | Electrical Protection | Class I, BF | Same | | User Interface | Touch screen, 12" | Same | | Firmware Controlled | Yes | Same | | Type of Energy | Magnetic field | Same | | Number of outputs | 2 | Same | | Number of Applicators | Up to two applicators can be operational at<br>same time | Same | | Applicator Connection | Detachable from the control unit | Same | | Total Induced Current in<br>Tissue (mA) | 327 | Same | | Type of Operation | Continuous | Same | | Number of Magnetic Coils<br>in the Applicator | 1 | Same | | Positioning of Applicator | Securement system | Same | | Magnetic Field Intensity | 0.5-1.35T ± 20% | Same | | Pulse Repetition Rate | 1 - 150 Hz | Same | | Pulse Duration | 370 µs ± 20% | Same | | Pulse Amplitude | 0 - 100% | Same | #### Table 6.1: Technological Characteristics between Subject and Predicate Device {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo is in black and white. The word "zimmer" is in a bold, sans-serif font. Below the word "zimmer" is the phrase "MedizinSysteme" in a smaller, sans-serif font. | Device Name | Predicate Device<br>CoolTone K192940 | Subject<br>Device<br>CoolTone | |----------------------------------------------|--------------------------------------|-------------------------------| | Selection of parameters<br>(Intensity, Time) | Yes | Same | | Therapy Time | Up to 30 min | Same | | Shape of Stimulation<br>Pulse | Symmetrical Biphasic Sine<br>Wave | Same | | Energy Source | 220-240V, 50/60 Hz | Same | | System Dimensions<br>(WxHxD) | 600x1100x600 mm | Same | | Operating Ambient<br>Temperature | 10°C – 28 °C | Same | | Environmental<br>Specifications | For indoor use only | Same | # Table 6.2: Technological Similarities between Proposed and Predicate Device per FDA Guidance for Industry for Powered Muscle Stimulators for 510(k)s (June 9, 1999) | Technological<br>Characteristics | Predicate Device<br>CoolTone K192940 | Subject Device<br>CoolTone | Comments | |-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|----------------------------|-------------------------------------------------------------------| | Power Source(s)<br>- Method of Line<br>Current Isolation<br>- Patient Leakage<br>Current<br>Normal condition<br>Single fault condition | 60601 compliant<br>< 1 μΑ<br>3.9 μΑ | Same | 60601 Compliant | | Number of Output<br>Modes | 1 | Same | | | Number of Output<br>Channels<br>- Synchronous or<br>Alternating?<br>- Method of Channel<br>Isolation | Synchronous<br>N/A | Same | No electrodes -<br>applicators are<br>not connected to<br>patient | | Regulated Current or<br>Regulated Voltage? | Voltage | Same | Controlled<br>voltage to the coil | | Software/Firmware/<br>Microprocessor Control? | Yes | Same | | | Automatic Overload<br>Trip? | N/A | N/A | No electrodes -<br>applicators are<br>not connected to<br>patient | | Automatic No-Load<br>Trip? | N/A | N/A | No electrodes -<br>applicators are<br>not connected to<br>patient | | Technological<br>Characteristics | Predicate Device<br>CoolTone K192940 | Subject Device<br>CoolTone | Comments | | Automatic Shut Off? | Yes | Same | Unit shuts off<br>with specified<br>timer | | Patient Override<br>Control? | No | Same | Treatment is<br>delivered by<br>healthcare<br>provider | | Indicator Display:<br>- On/Off Status?<br>- Low Battery?<br>- Voltage/Current Level? | Yes<br>N/A<br>No | Same | | | Timer Range<br>(minutes) | Up to 30 min | Same | | | Compliance with<br>Voluntary Standards? | Yes | Same | Refer to<br>performance data<br>for details | | Compliance* with<br>21 CFR 898? (*Becomes<br>mandatory beginning<br>May 9, 2000) | Yes | Same | | | Weight | 80 Kg | Same | | | Housing Materials<br>and Construction | Steel and Injection<br>Molded Plastics | Same | | | Waveform (e.g., pulsed<br>monophasic, biphasic) | Symmetrical Biphasic<br>Sine<br>Wave | Same | | | Shape (e.g., rectangular,<br>spike, rectified<br>sinusoidal) | Sinusoidal | Same | | | Maximum Output<br>Voltage (specify units) | N/A | N/A | No electrodes -<br>applicators are<br>not connected to<br>patient | | Maximum Output<br>Current (specify units) | N/A | N/A | No electrodes -<br>applicators are<br>not connected to<br>patient | | Frequency (Hz) | 1-150 Hz | Same | | | For interferential<br>modes only: - Beat<br>Frequency (Hz) | N/A | N/A | | | For multiphasic<br>waveforms only: -<br>Symmetrical phases? | Yes | Same | Biphasic | | Phase Duration<br>(include units) (state<br>range, if applicable) | 370 $\mu$ s +/- 20% | Same | | | Technological<br>Characteristics | Predicate Device<br>CoolTone K192940 | Subject Device<br>CoolTone | Comments | | (both phases, if<br>asymmetrical) | N/A | N/A | | | Net Charge (mC per<br>pulse) (If zero, state<br>method of achieving<br>zero net charge.) | N/A | N/A | No electrodes -<br>applicators are<br>not connected to<br>patient | | Maximum Phase<br>Charge, (mC) | N/A | N/A | No electrodes -<br>applicators are<br>not connected to<br>patient | | Maximum Current<br>Density, (mA/cm²) | N/A | N/A | No electrodes -<br>applicators are<br>not connected to<br>patient | | Maximum Power<br>Density, (W/cm²)<br>(using smallest<br>electrode conductive<br>surface area) | N/A | N/A | No electrodes -<br>applicators are<br>not connected to<br>patient | | Burst Mode7 (i.e., pulse<br>trains) a. Pulses per<br>burst b. Bursts per<br>second c. Burst<br>duration (seconds) d.<br>Duty Cycle<br>[Line (b) x Line (c)] | N/A | N/A | | | ON Time (seconds) | N/A | N/A | | | OFF Time (seconds) | N/A | N/A | | | Additional Features (if<br>applicable) | N/A | N/A | | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a line above the "z" and "i". Below the word "zimmer" is the text "MedizinSysteme" in a smaller, sans-serif font. The logo is in grayscale. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The word "zimmer" is in bold, lowercase letters on the top line. Below that, the words "MedizinSysteme" are in a smaller font size. The logo is in grayscale. # VII. PERFORMANCE DATA: The CoolTone System is the same as the previously cleared predicate device (K192940). It has been tested/assessed against and fully complies with the following voluntary standards: | Table 6.3: Compliance with Voluntary Standards | | |------------------------------------------------|--| |------------------------------------------------|--| | Standards | Standards<br>Organization | Standards Title | |-------------------------------------------------------------------------------------------------|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ANSI AAMI ES<br>60601-1:2005/(R)2012<br>and A1:2012,<br>C1:2009/(R) 2012 and<br>A2:2010/(R)2012 | ANSI<br>AAMI | Medical electrical equipment – Part 1: General<br>requirements for basic safety and essential<br>performance | | Standards | Standards<br>Organization | Standards Title | | 60601-1-2:2014<br>(Edition 4.0) | IEC | Medical electrical equipment – Part 1-2: General<br>requirements for basic safety and essential<br>performance – Collateral standard:<br>Electromagnetic disturbances - Requirements<br>and tests | | 60601-1-6:2013<br>(Edition 3.1) | IEC | Medical electrical equipment – Part 1-6: General<br>requirements for basic safety and essential<br>performance - Collateral standard: Usability | | 60601-2-10:2016<br>(Edition 2.1) | IEC | Medical electrical equipment - Part 2-10:<br>Particular requirements for the Basic Safety and<br>Essential Performance of Nerve and Muscle<br>Stimulators | | 62366-1:2015<br>(Edition 1.0) | IEC | Medical devices - Application of usability<br>engineering to medical devices | | 62304:2015<br>(Edition 1.1) | ISO | Medical devices software –software life cycle<br>processes | | 14971:2012 | ISO | Medical devices - Application of risk management<br>to medical devices | | 10993-1: 2018 | ISO | Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management<br>process | | 10993-5: 2009 | ISO | Biological evaluation of medical devices - Part 5:<br>Tests for in vitro cytotoxicity | | 10993-10: 2010 | ISO | Biological evaluation of medical devices - Part 10:<br>Tests for irritation and skin sensitization | {9}------------------------------------------------ # Software Verification and Validation Testing Software verification and validation testing were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Software verification was performed in three steps: verification of the software requirements, verification of the design specifications and verification of the software architecture. All tests, including Integration Tests (Tests of the System Architecture and the {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The word "zimmer" is in bold, sans-serif font. Below the word "zimmer" is the word "MedizinSysteme" in a smaller, sans-serif font. The logo is in grayscale. Design Specification) and the System Level Test (Test of functional requirements that were evaluated in the software requirements verification) were performed successfully and met their acceptance criteria. ## Cybersecurity Cybersecurity risk management for the device was performed as part of the overall risk management process for the medical device and follows the guidance in the FDA document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". The changes made to the predicate's (K192940) system software have no impact on cybersecurity. ## Biocompatibility Testing There were no changes to patient-contacting materials in comparison with the previously cleared predicate device. ## Electrical safety and electromagnetic compatibility (EMC) The CoolTone system (control unit and applicator) has undergone electrical and mechanical safety performance testing as a result of the changes referenced. Additionally, the control unit underwent electromagnetic compatibility testing. The applicator did not require additional electromagnetic compatibility testing because the changes did not affect its electromagnetic compatibility. The system complies with ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R)2012), and IEC 60601-1-2:2014 (Fourth Edition). ## Performance Testing The proposed CoolTone system performance is the same as the previously cleared predicate device. The System Level Testing confirmed that the two applicators, either operated separately or simultaneously, performed within the magnetic field intensity of 0.5 – 1.35T +/- 20% and that the tissue being treated by the device is not subjected to an appreciable rise in temperature at maximum intensity to cause a risk to the patient. # Clinical Study No clinical testing was required for these changes. ## VIII. CONCLUSION: The Indication for Use for the subject CoolTone system is the same as the predicate device cleared in K192940. The changes that have been made to the system's hardware and software do not affect the intended use, performance or risk profile of the device. The subject CoolTone system is therefore substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...