K220538 · E.M.S Electro Medical Systems S.A · ISA · Jul 11, 2022 · Physical Medicine
Device Facts
Record ID
K220538
Device Name
DolorClast Radial
Applicant
E.M.S Electro Medical Systems S.A
Product Code
ISA · Physical Medicine
Decision Date
Jul 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5660
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The DolorClast® Radial is indicated for: Relief of minor muscle aches and pains Temporary increase in local blood circulation Activation of connective tissue
Device Story
DolorClast® Radial is a portable radial pressure pulse device for therapeutic massage. System includes a console with integrated air compressor, handpiece, interchangeable stainless-steel applicator heads, and coupling gel. Practitioner operates device via handpiece trigger button. Console generates pneumatic pressure pulses; air drives internal projectile against applicator head. Applicator transmits radial pressure pulses to patient's skin via coupling gel. Device used in clinical settings to provide mechanical vibration. Output affects local blood circulation and connective tissue, providing relief for minor muscle aches and pains. Benefits include non-invasive therapeutic massage.
Clinical Evidence
Bench testing only. No clinical data. Testing included system functional performance (pressure output, console/handpiece operation), acoustic output characterization (displacement, penetration depth, energy flow density, peak pressure, pulse duration), lifetime testing (console, handpiece, applicators), biocompatibility (leveraging P050004), transportation testing (ISTA 2A/3A), and electrical safety/EMC (IEC 60601-1, IEC 60601-1-2).
Technological Characteristics
Pneumatic/ballistic radial pressure pulse device. Materials: stainless-steel applicators. Energy source: mains-powered integrated air compressor. Dimensions: 450 x 340 x 180 mm. Weight: 14.5 kg. Operating frequency: 1-25 Hz. Pressure range: 1-4 bar. Software: moderate level of concern, compliant with IEC 62304. Standards: IEC 60601-1, IEC 60601-1-2, ISTA 2A/3A.
Indications for Use
Indicated for relief of minor muscle aches and pains, temporary increase in local blood circulation, and activation of connective tissue in patients requiring therapeutic massage.
Regulatory Classification
Identification
A therapeutic massager is an electrically powered device intended for medical purposes, such as to relieve minor muscle aches and pains.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font size.
7/11/22
EMS Electro Medical Systems SA % Sheila Hemeon-Heyer President Heyer Regulatory Solutions LLC 125 Cherry Lane Amherst, Massachusetts 01002
Re: K220538
Trade/Device Name: DolorClast® Radial Regulation Number: 21 CFR 890.5660 Regulation Name: Therapeutic massager Regulatory Class: Class I Product Code: ISA Dated: February 25, 2022 Received: February 25, 2022
Dear Sheila Hemeon-Heyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K220538
Device Name
DolorClast® Radial
Indications for Use (Describe)
The DolorClast® Radial is indicated for:
- · Relief of minor muscle aches and pains
- · Temporary increase in local blood circulation
- · Activation of connective tissue
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
| A. | Submitter: | EMS Electro Medical SA<br>Ch. de la Vuarpillière 31<br>1260 Nyon, Switzerland<br>Contact: Vivian Yee<br>Phone: +44 7305 232869<br>Email: vyee@ems-ch.com |
|----|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
|----|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
- B. Date Prepared: July 7, 2022
#### Device Name and Classification Information: D.
| Trade name: | DolorClast® Radial |
|----------------------------|-------------------------------|
| Common/Usual Name: | Therapeutic Electric Massager |
| Classification name: | Therapeutic Massager |
| Review Panel | Physical Medicine |
| Classification regulation: | 21 CFR 890.5660 |
| Product code: | ISA |
| Class: | Class I |
#### E. Predicate Device(s): K173692, D-Actor® 200 Vibration Massage System
#### F. Device Description:
The DolorClast® Radial is a portable radial pressure pulse device with an integrated air compressor. The DolorClast® Radial system is composed of the DolorClast® Radial console, handpiece, interchangeable applicator heads, and optional cart. Swiss DolorClast® Gel, previously approved under P050004, is also provided for coupling the treatment applicator to the patient's skin.
The DolorClast® Radial console generates and controls the pressure pulse frequency and intensity. Air pressure from the console drives a projectile within the handpiece against the applicator head to generate the pressure pulses at the set frequency and intensity. Device activation is done by the practitioner via a trigger button on the handpiece. The applicator at the distal end of the handpiece transmits the radial pressure pulses to the targeted tissues via the coupling gel applied to the treated skin.
#### G. Indications for Use Statement
The DolorClast® Radial is indicated for:
- . Relief of minor muscle aches and pains
- Temporary increase in local blood circulation
- Activation of connective tissue ●
{4}------------------------------------------------
#### Comparison with Predicate Device H.
| Parameter | Proposed Device<br>(K220538) | Predicate Device<br>(K173692) | Comparison |
|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Device Trade Name | DolorClast® Radial | D-Actor® 200 Vibration Massage<br>System | N/A |
| Device Manufacturer | EMS Electro Medical System<br>SA | Storz Medical AG | N/A |
| Indications for Use<br>Statement | The DolorClast® Radial<br>is indicated for:<br>Relief of minor muscle<br>●<br>aches and pains<br>Temporary increase in local<br>●<br>blood circulation<br>Activation of connective<br>●<br>tissue | The D-Actor® 200<br>Vibration Massage<br>System is intended for:<br>Relief of minor muscle<br>aches and pains<br>Temporary increase in local<br>●<br>blood circulation<br>Activation of connective<br>tissue | Same |
| Mode of Action | Radial pressure waves, or<br>extracorporeal pulse activation<br>respectively | Radial pressure waves, or<br>extracorporeal pulse activation<br>respectively | Same |
| Mechanism of action | Pneumatically generated<br>vibrations | Pneumatically generated<br>vibrations | Same |
| System components | Control console with integrated<br>air compressors to control<br>treatment parameters<br>Mains cable<br>Handpiece containing projectile<br>within guiding tube<br>Interchangeable applicators<br>Coupling gel<br>Cart (optional) | Control unit<br>Mains cable<br>Handpiece containing projectile<br>within guiding tube<br>Interchangeable applicators<br>Coupling gel<br>Footswitch (optional)<br>Tablet PC (optional) | Same except for<br>optional<br>accessories |
| Method of triggering<br>vibration pulses | Handpiece | Handpiece or optional<br>footswitch | Same except for<br>predicate<br>optionally can be<br>controlled using a<br>footswitch |
| Type of acoustic wave<br>generation | Pneumatic/ballistic | Pneumatic/ballistic | Same |
| Pressure setting range | 1-4 bar | 1-5 bar | Similar |
| Frequency setting range | 1-25 Hz | 1-21 Hz | Similar |
| Number of treatment<br>applicators | 7 | 4 | Similar |
| Treatment applicator<br>sizes | OD in mm: 5, 10, 15, 25, 40 | OD in mm: 6, 15, 20, 35 | Similar |
| Projectile mass (g) | 3.2 | 3 | Similar |
| Displacement of<br>applicator heads | 0.18 - 0.28 mm | 0.6 – 2.0 mm | Similar |
| Operating mode | Continuous vibration at a fixed<br>frequency | Continuous vibration at a fixed<br>frequency | Same |
| Parameter | Proposed Device<br>(K220538) | Predicate Device<br>(K173692) | Comparison |
| Maximum penetration<br>depth | 40 mm | 32.3 mm | Similar |
| Number of pulses per<br>treatment | Variable as set by operator<br>Max 5 000 pulses/treatment | Variable<br>Max not publicly available | Similar |
| Maximum energy flow<br>density | 0.29 mJ/mm² at 4 bar<br>0.14 mJ/mm² at 2.4 bar | 0.284 mJ/mm² at 5 bar<br>0.176 mJ/mm² at 3 bar | Similar |
| Maximum positive peak<br>pressure amplitude | 17 MPa at 4 bar<br>11.24 MPa at 2.4 bar | 18.5 MPa at 5 bar<br>13.4 MPa at 3 bar | Similar |
| Maximum negative peak<br>pressure amplitude | 10 MPa at 4 bar<br>7.2 MPa at 2.4 bar | 6.8 MPa at 5 bar<br>5.0 MPa at 3 bar | Similar |
| Derived focal acoustic<br>pulse energy | 5.9 mJ at 4 bar*<br>2.2 mJ at 2.4 bar* | 6.5 mJ at 5 bar<br>2.4 mJ at 3 bar | Similar |
| Derived pulse-intensity<br>integral (max), integrated<br>over total temporal<br>integration limits<br>(mJ/mm²) | 0.29 mJ/mm² at 4 bar<br>0.14 mJ/mm² at 2.4 bar | 0.284 mJ/mm² at 5 bar<br>0.176 mJ/mm² at 3 bar | Similar |
| Rise time | 3.2 µs | 2.5 µs | Similar |
| Compressional pulse<br>duration | 2.6 µs (1st peak)<br>62.7 µs (1st phase) | Ultrasonic pulse: 5.0 µs<br>Sonic pulse: 50 µs - 5.0 ms | Similar |
| Power supply | Mains power | Mains power | Same |
| Operating temperature<br>range | 10 °C - 30 °C | 10 °C - 40 °C | Similar |
| Control console<br>dimensions | 450 x 340 x 180 mm<br>17.7 x 13.4 x 7.1 inches | Not publicly available | - |
| Control console weight | 14.5kg | Not publicly available | - |
{5}------------------------------------------------
*Averaged across all applicator sizes.
#### -Substantial Equivalence Discussion
The DolorClast® Radial has the same indications for use and mechanism of action as the predicate D- Actor® 200. Both use compressed air to drive a projectile against a stainless steel applicator in contact with the skin, transferring low-amplitude pressure pulses to the treatment target via standard coupling gel. Both devices are operated using a handpiece; however, the predicate device can also optionally be operated using a footswitch. There are minor differences in the technical characteristics of the pressure pulses. The maximum pulse pressure setting of the DolorClast® Radial is slightly lower than the D-Actor® 200 Vibration Massage System (4 bar vs 5 bar, respectively), while the maximum pulse frequency setting of the DolorClast® Radial is slightly higher than the D-Actor® 200 Vibration Massage System (25Hz vs 21Hz, respectively). The DolorClast® Radial has more options for applicators than the D-Actor® 200 Vibration Massage System (7 vs 4, respectively); however, the applicator size ranges for the two devices are very similar. The technical characteristics of the pressure pulses are very similar between the two devices. In particular, the maximum energy flux density is nearly identical at 0.29 mJ/mm² for the DolorClast® Radial and 0.28mJ/mm² for the D-Actor® 200 Vibration Massage System. The DolorClast® Radial applicator heads demonstrate less movement than the D-Actor® 200
{6}------------------------------------------------
Vibration Massage System, ranging from 0.18 to 0.28 mm depending on applicator size for the DolorClast® Radial versus 0.6 to 2.0 mm for the D-Actor® 200 Vibration Massage System. However, the role of the applicator is to transfer the pressure pulse energy from the handpiece piston to the treatment site, which is independent of applicator movement. The technical differences do not present any new risks or raise any new questions of safety or effectiveness.
#### J. Summary of Data Submitted to Support Substantial Equivalence
The following testing was included in this 510(k) to support substantial equivalence:
### System functional performance testing
- The operation of the DolorClast® Radial air compressors was tested under an internal EMS protocol to confirm that the compressors were only activated when the handpiece trigger was pressed and to verify that the pressure output at each pressure setting met the pre - defined pressure setting tolerance. All tests were passed.
- . The operation of the DolorClast® Radial console was tested under an internal EMS protocol to confirm correct operation of the console at all treatment settings, including for the Smart Protocols and Special Modes. Testing included error conditions to verify correct device response. All tests were passed.
- . The operation of the DolorClast® Radial handpiece was tested under an internal EMS protocol to confirm correct operation of the handpiece including adjustment of the pressure, starting and stopping treatment, and recording piston use information. All tests were passed.
- Acoustic output testing was conducted to characterize the output pressure wave for each . of the system applicators on the following parameters:
- Displacement of applicator heads o
- O Maximum penetration depth
- Energy flow density O
- Positive and negative peak pressure amplitudes O
- Derived focal acoustic pulse energy o
- Derived pulse-intensity integral O
- Rise time O
- Compressional pulse duration o
### Lifetime Testing
Testing was conducted under internal EMS protocols to confirm that the DolorClast® Radial console, handpiece, and applicators continued to function and meet their performance specifications for the claimed useful lifetimes of:
- . Console: 7 years or 50,000,000 pulses
- . Handpiece: 2 years or 5,000,000 pulses
- . Applicators: 2 years or 5,000,000 pulses
{7}------------------------------------------------
# Biocompatibility
The only components of the DolorClast® Radial system that come into contact with the patient are the applicator tips and the Swiss DolorClast® gel, both of which are in contact with intact skin for less than 24 hours. Cumulative use may exceed 24 hours for heavy users. The Swiss DolorClast® gel was previously approved under P050004 and has not changed since that time. The applicators are made of the same stainless-steel material as the applicators for the Swiss DolorClast® applicators, previously approved under P050004, and there have been no changes to the applicator material, manufacturing processes, or patient contact category since that time. Therefore, the prior biocompatibility testing conducted to support P050004 remains applicable for both the applicators and the gel, and no further biocompatibility testing was required to support the DolorClast® Radial 510(k) application.
# Transportation Testing
Transportation testing was conducted for the DolorClast® Radial system in accordance with the International Safe Transit Authority standards ISTA 2A and ISTA 3A. Packages were visually inspected after being subjected to environmental and mechanical conditioning. The device and device accessories were tested for functionality following the package conditioning. All tests were passed.
# Software Validation
The DolorClast® Radial software was evaluated as having a moderate level of concern. The 510(k) included software documentation commensurate with a moderate level of concern and compliant with IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cycle processes. The software was validated in both software testing and system level testing with the software integrated into the final device.
# Electrical Safety and EMC
The DolorClast was tested and demonstrated to comply with the following standards:
- . IEC 60601-1:2005 + A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
#### K. Conclusion
The information presented in this 510(k) demonstrates that the DolorClast® Radial is substantially equivalent to the predicate device as a Class I therapeutic massager under 21 CFR 890.5660, Product Code ISA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.