Versana Balance

K220446 · GE Medical Systems Ultrasound and Primary Care Diagnostics · IYN · May 11, 2022 · Radiology

Device Facts

Record IDK220446
Device NameVersana Balance
ApplicantGE Medical Systems Ultrasound and Primary Care Diagnostics
Product CodeIYN · Radiology
Decision DateMay 11, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesAI/ML, Pediatric

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Anatomical structure identification———————
Color flow optimization———————

Indications for Use

The Versana Balance is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Versana Balance clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Urology, Pediatric, Small Parts (includes breast, testes, thyroid), Cardiac Pediatric, Vascular/Peripheral Vascular, Musculoskeletal Conventional, Musculoskeletal Superficial, Thoracic/Pleural, Transcranial, Transvaginal, Interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD, B Flow/B Flow Color. Versana Balance is intended to be used in a hospital or medical clinic.

Device Story

Versana Balance is a mobile diagnostic ultrasound console for use by trained clinicians in hospitals or clinics. It acquires, processes, and displays ultrasound data via various transducers (convex, linear, sector, mechanical 4D). Inputs include acoustic signals from transducers and optional external ECG gating. The system features AI-based anatomy labeling (Whizz Label), automated color flow optimization (Whizz Color Flow), and volume calculation tools (VOCAL). It provides real-time imaging, measurement, and analysis to assist clinicians in diagnostic and interventional procedures. Data management includes export via USB, LAN, WiFi, or Bluetooth. The device includes an internal battery for mobile operation and supports peripheral connectivity. Clinical benefits include enhanced visualization, standardized reporting (LI-RADS), and improved workflow efficiency through automated features, aiding in accurate diagnosis and procedural guidance.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including acoustic output, biocompatibility, electrical/electromagnetic safety, and software verification/validation.

Technological Characteristics

Mobile ultrasound console; transducers (convex, linear, sector, 4D); modes: B, M, Doppler, Harmonic, Elastography, 3D/4D. Connectivity: USB, LAN, WiFi, Bluetooth, HDMI, VGA. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, IEC 62359, ISO 14971, NEMA PS 3.1-3.20 (DICOM). Software includes AI-based anatomy labeling and automated parameter optimization.

Indications for Use

Indicated for qualified healthcare professionals to perform ultrasound imaging, measurement, and analysis of human body and fluid. Clinical applications: Fetal/Obstetrics, Abdominal, Gynecology, Urology, Pediatric, Small Parts (breast, testes, thyroid), Cardiac (Adult/Pediatric), Vascular/Peripheral Vascular, Musculoskeletal (Conventional/Superficial), Thoracic/Pleural, Transcranial, Transvaginal, and Interventional guidance (biopsy, drainage, access).

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. GE Medical Systems Ultrasound and Primary Care Diagnostics % Lee Bush Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226 ### Re: K220446 Trade/Device Name: Versana Balance Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: February 15, 2022 Received: February 16, 2022 May 11, 2022 ## Dear Lee Bush: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Michael D. O'Hara, Ph.D. Deputy Director DHT 8C: Division of Radiological Imaging and Radiation Therapy Devices OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K220446 Device Name Versana Balance #### Indications for Use (Describe) The Versana Balance is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Versana Balance clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Urology, Pediatric, Small Parts (includes breast, testes, thyroid), Cardiac Adult, Cardiac Pediatric, Vascular, Musculoskeletal Conventional, Musculoskeletal Superficial, Transcranial, Transrectal, Transvaginal, Interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, and Combined modes: B/M, B/Color, B/PWD, B/ Color/PWD, B/Power/PWD, B Flow/B Flow Color. Versana Balance is intended to be used in a hospital or medical clinic. > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame is adorned with three droplet-like shapes evenly spaced around the circle. The logo is presented in a blue color scheme. # 510(k) Summary K220446 In accordance with 21 CFR 807.92 the following summary of information is provided: | Date: | May 5, 2022 | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | GE Medical Systems Ultrasound and Primary Care Diagnostics<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 | | Primary Contact Person: | Lee Bush<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262) 309-9429 | | Secondary Contact Person: | Andrew Turner<br>Regulatory Affairs<br>GE Healthcare | | Device Trade Name: | Versana Balance | | Common/Usual Name: | Diagnostic Ultrasound System | | Classification Names: | Class II | | Product Code(s): | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX | | Primary Predicate Device: | Versana Balance (K191792), Diagnostic Ultrasound System | | Classification Names: | Class II | | Product Code(s): | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle is surrounded by three water droplet-like shapes, also in blue, positioned around the top and sides of the circle. The logo is simple, clean, and recognizable. Reference Device: Versana Premier (K210438), Diagnostic Ultrasound System Classification Names: Class II Product Code(s): Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX Reference Device: Vivid i (K121062), Diagnostic Ultrasound System Classification Names: Class II Product Code(s): Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX; Picture Archiving and Communication System 21 CFR 892.2050, 90-LLZ Reference Device: LOGIQ e (K113690), Diagnostic Ultrasound System Classification Names: Class II Product Code(s): Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. There are three water droplet-like shapes evenly spaced around the outside of the circle. The logo is a light blue color. #### Device Description: The Versana Balance is a general purpose, Track 3, diagnostic ultrasound system for use by qualified and trained healthcare professionals. The system is a mobile console that includes an operator control panel, display monitor and transducers. The console provides digital acquisition, processing and display capability. The system has an internal battery to allow for acquisition while the system is not plugged into a power source. Acquisition can also be done while the system is connected to an AC power source. The operator control panel includes function keys, trackball, an alfa-numeric keyboard and a touch panel as input sources of the device. The variety of transducers include convex, linear, sector, and mechanical 4D transducers. The access types include trans- body surface, transrectal, transvaginal and transcranial. Data can be imported or exported by DVD, USB, LAN or WiFi if the USB wireless adapter is connected to the system. An external ECG module has been verified to use as input for gating/triggering during scanning. The system has a HDMI port, VGA connection port, Audio out port, S-Video port, and a Composite Out port connection. The system has an external AC outlet to allow connection of a printer box of the console and an option for external Printer USB Isolator for other printers to connect. The system supports one way, Bluetooth communication capability from the system to a personal device to allow for sharing of the patient's data/images when the external Bluetooth USB adapter is connected to the system. ### Intended Use/Indication for Use: The Versana Balance is a general-purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. Versana Balance clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Urology, Pediatric, Small Parts (includes breast, testes, thyroid), Cardiac Pediatric, Vascular/Peripheral Vascular, Musculoskeletal Conventional, Musculoskeletal Superficial, Thoracic/Pleural, Transcranial, Transvaginal, Interventional guidance (includes tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode, Elastography, and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD, B Flow/B Flow Color. Versana Balance is intended to be used in a hospital or medical clinic. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. There are three water droplet-like shapes evenly spaced around the outside of the circle. The logo is a solid blue color and is set against a white background. ## Technology: The Versana Balance employs the same fundamental scientific technology as its predicate device and reference devices. ## Determination of Substantial Equivalence: The proposed Versana Balance system is substantially equivalent to the predicate Versana Balance device with regard to intended use, imaging capabilities, technological characteristics, imaging modes, hardware, and safety effectiveness. ### Indications for use - o Vascular has been added with peripheral vascular, this clarifies vessels and is not a change in imaging or product. No impact to safe or effective use. This is the same as on reference device Versana Premier (K210438). - . Removal of the wording "Cephalic" as this is covered by "Transcranial" and to align indications with the reference device Versana Premier (K210438). No impact to safe or effective use. - Gynecology and Urology are listed independently instead of within Abdominal as it was in the predicate Versana Balance (K191792). There is no change in imaging or the product. No impact to safe or effective use. - Small Organ is changed to Small Parts, it is a wording use preference change only. There is no change in imaging or the product. No impact to safe or effective use. This is the same as on reference device Versana Premier (K210438). - Cardiac (includes Adult and Pediate Versana Balance (K191792) is changed to Cardiac Adult and Cardiac Pediatric; it is a wording use preference change only. There is no change in imaging or the product. No impact to safe or effective use. This is the same as on reference device Versana Premier (K210438). ## Transducers and Modes: - The proposed Versana Balance and predicate Versana Balance (K191792) systems transducers are similar, except for: - o Addition of L8-18i-RS, which was first cleared on LOGIQ e (K113690). The clinical indications and modes of L8-18i-RS are the same as on the proposed Versana Balance as they are on the reference device Versana Premier (K210438). - Addition of 12S-RS, which was first cleared on Vivid i, (K121062). The clinical indications and modes o of 12S-RS are the same as on proposed Versana Balance as they are on the reference device Versana Premier (K210438). - Elastography mode has been added to the following existing transducers for Versana Balance for o same clinical indications as the reference device Versana Premier (K210438): - 트 4C-RS - . E8Cs-RS - L6-12-RS - 12L-RS {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'GE' in a stylized font inside. The letters are connected and have a flowing, almost cursive appearance. The circle is surrounded by a thin white border, and there are four curved lines emanating from the circle, resembling water droplets or stylized flames. The logo is simple, clean, and recognizable, representing the brand identity of General Electric. # Software: - Addition of Whizz Label: an Artificial Intelligence (AI) feature that identifies key anatomies/organs ● during ultrasound imaging. Whizz Label was first cleared with Versana Premier (K210438). - . Addition of LI-RADS: Liver Imagining Reporting and Data System, based on the ACR (American College of Radiology) published literature. LI-Rads enables the user to enter categorization, standardized reporting, and data collection for the liver. - Addition of Probe Check (transducer element check): a tool that evaluates the probe performance and provides the user with an indication of the potential impact to the diagnostic image if it is compromised due to a transducer malfunction. - . Addition of Whizz Color Flow: Whizz Color Flow is used to optimize color flow by adjusting Color Flow parameters (gain and frequency) automatically. Whizz Color Flow is the same feature as on reference device Versana Premier (K210438). - Addition of Lateral Gain Compensation (LGC): Lateral Gain Compensation (LGC) amplifies returning signals to correct for the attenuation due the beam penetrating through tissue. This is a similar feature as on reference device Versana Premier (K210438). - . Addition of VOCAL: a function to visualize and calculate the volume of anatomical structures running under 3D/4D mode for volume calculation. - Addition of Standby mode (Fast boot up): a power saving functionality and is the same feature as on Versana Premier (K210438). - Addition of V-Live: a feature that allows the user to rotate a single light source by track ball to easily highlight details in otherwise darkened areas within a region of interest. This is the same feature as on reference device Versana Premier (K210438). ## Mechanical and Industrial Design Change: Optimization of the industrial design, no change in the materials themselves Hardware: - Addition of an optional 13.3' multi-touch panel ● - Addition of an optional cable tray - Updated high level internal computer components (Hard drive, CPU, etc) - Addition of optional barcode reader {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame is adorned with small, teardrop-shaped elements that resemble water droplets or stylized leaves. The color of the logo is a bright, solid blue. ### Summary of Non-Clinical Tests: Versana Balance has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to comply with applicable medical device safety standards. The Versana Balance complies with voluntary standards: - AAMI/ANSI ES60601-1, Medical Electrical Equipment – Part 1: General requirements for basic safety and essential performance - 2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 - . IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests - Edition 4.0, 2014 - IEC 60601-2-37, Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment -Edition 2.1, 2015 - ISO 10993-1, Biological evaluation of medical devices Part 1: Evaluation and testing within a ● risk management process - Fourth edition, 2009 - IEC 62359, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields - Edition 2.1, 2017 - ISO 14971, Medical devices - Application of risk management to medical devices, 2019 - NEMA PS 3.1 - 3.20, Digital Imaging and Communications in Medicine (DICOM) Set, 2016 The following quality assurance measures are applied to the development of the system: - Risk Analysis - Requirements Reviews ● - Design Reviews ● - Testing on unit level (Module verification) - Integration testing (System verification) - Performance testing (Verification) - . Safety testing (Verification) Transducer material and other patient contact materials are biocompatible. ### Summary of Clinical Tests: The subject of this premarket submission, Versana Balance, did not require clinical studies to support substantial equivalence. ### Conclusion: GE Healthcare considers the Versana Balance to be as safe, effective, and performs in a substantially equivalent manner as the predicate and reference devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...