The Saffron™ Fixation System is indicated for attaching suture to ligaments of the pelvic floor.
Device Story
Saffron Fixation System is a manual, single-use surgical instrument for transvaginal pelvic floor ligament fixation. System comprises a delivery tool and implantable polysulfone anchors. Physician inserts distal end of tool through dissected tissue to target ligament; manually presses plunger to advance cannula and deploy anchor. Anchor provides permanent fixation point for suture (size 0 or smaller). Used in clinical settings by surgeons. Device facilitates anchor placement with or without direct visualization. Benefits include permanent suture attachment to pelvic floor ligaments for surgical repair.
Clinical Evidence
No clinical or animal studies were provided. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), mechanical performance (penetration, detachment, holding, and actuation forces), and sterile barrier integrity testing.
Technological Characteristics
Materials: Polysulfone anchor with titanium oxide colorant; delivery tool uses polypropylene, polycarbonate, stainless steel 304, and ABS. Dimensions: Anchor 9.29mm x 3.44mm x 1.65mm; tool shaft 4mm diameter, 200mm length. Energy: Manual mechanical actuation. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 10993 (biocompatibility), ISO 11135 (sterilization), ISO 11607 (packaging), ASTM D4169 (transportation).
Indications for Use
Indicated for attaching suture to ligaments of the pelvic floor.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 10, 2022
Coloplast A/S Gayatri Ghadge Principal Regulatory Affairs Specialist 1601 West River Road North Minneapolis, MN 55411
Re: K220420
> Trade/Device Name: Saffron™ Fixation System Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: PBQ Dated: May 9, 2022 Received: May 10, 2022
Dear Gayatri Ghadge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220420
Device Name Saffron™ Fixation System
### Indications for Use (Describe)
The Saffron™ Fixation System is indicated for attaching suture to ligaments of the pelvic floor.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-family: sans-serif; font-size: 10pt;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span style="font-family: sans-serif; font-size: 10pt;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| I.<br>SUBMITTER<br>510(K) Owner's Name: | Coloplast A/S |
|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Holtedam 1<br>3050 Humlebaek, Denmark |
| Phone/Fax/Email: | Office: 612-422-3206<br>Email: usggh@coloplast.com |
| Name of Contact Person: | Gayatri Ghadge<br>Principal Regulatory Affairs Specialist |
| Address/Contact: | 1601 West River Road North<br>Minneapolis, MN 55411 |
| Date Prepared: | June 7, 2022 |
| II.<br>DEVICE<br>Trade or Proprietary Name: | Saffron™ Fixation System |
| Common or Usual Name:<br>Regulation Number:<br>Regulation Name:<br>Product Code: | Pelvic ligament fixation system<br>21 CFR 884.4530<br>Obstetric-gynecologic specialized manual instrument<br>PBQ (Fixation, Non-Absorbable Or Absorbable, For Pelvic<br>Use) |
| Regulatory Class: | II |
#### III. PREDICATE DEVICE
Anchorsure, 510(k)-cleared on October 12, 2012 under K120831.
The predicate device has not been subject to a design-related recall.
#### DEVICE DESCRIPTION IV.
The Saffron Fixation System consists of a single-use disposable Saffron Fixation Tool (delivery tool) and implantable tissue fixation Saffron Anchors (packaged and sold separately) made of polysulfone that are intended to be compatible with commercially available sutures up to size 0 (not provided as a part of the system). The delivery tool facilitates anchor placement with or without direct visualization and may be used to implant multiple anchors in a single patient as needed by the surgeon. The fixation anchors are multi-barbed hooks that provide permanent fixation points for suture attachment to the sacrospinous ligament..
{4}------------------------------------------------
The physician transvaginally inserts the distal end of the delivery tool with a loaded anchor through the dissected tissue to the desired location. The physician manually positions the tip of the distal end at the location where the anchor is to be placed and manually presses the plunger advancing the cannula in the fixation tool and releasing the anchor into the tissue. The anchor penetrates the chosen landmark to provide a permanent fixation point for the loaded suture.
The Saffron Fixation device is a single use device which is supplied sterile. The Saffron Anchor is an implantable device which is supplied sterile. The Saffron Fixation System has a shelf life of 6 months.
#### V. INDICATIONS FOR USE
The Saffron™ Fixation System is indicated for attaching suture to ligaments of the pelvic floor.
| VI. | COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE |
|-----|------------------------------------------------------|
| | PREDICATE DEVICE |
| | Subject Device | Predicate Device |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| | Saffron Fixation System | Anchorsure |
| | K220420 | K120831 |
| Regulation<br>Number | 21 CFR 884.4530 | 21 CFR 884.4530 |
| Product Code | PBQ | PBQ |
| Indications for<br>Use | The Saffron™ Fixation System is<br>indicated for attaching suture to<br>ligaments of the pelvic floor. | Anchorsure is indicated for<br>attaching suture to ligaments of the<br>pelvic floor. |
| Target Structures | Pelvic floor ligaments | Pelvic floor ligaments |
| Anchor<br>Placement<br>Duration | Permanent Implant | Permanent Implant |
| Anchor Materials | Polysulfone, titanium oxide<br>colorant | Polyether ether ketone (PEEK) |
| Anchor<br>Dimensions | Length: 9.29 mm<br>Height: 3.44 mm<br>Width: 1.65 mm | Diameter: 3 mm<br>Length: 7 mm |
| Delivery System<br>Materials | Polypropylene (plunger),<br>polycarbonate (handle), stainless<br>steel 304 (shaft), polycarbonate /<br>acrylonitrile butadiene styrene<br>(ABS) (distal tip) | Stainless steel 303,<br>polyoxymethylene |
| Delivery System<br>Dimensions | Fixation Tool shaft diameter: 4 mm<br>Fixation Tool shaft length: 200 mm | Anchoring Handle tube diameter: 6<br>mm<br>Anchoring Handle tube length: 204<br>mm |
| Accessories<br>Provided | No, compatible with any suture<br>size 0 or smaller | Yes, USP size 0 polypropylene<br>suture and curved stainless steel |
{5}------------------------------------------------
| | | surgical needle |
|---------|---------------------|---------------------|
| Sterile | Yes, ethylene oxide | Yes, ethylene oxide |
Saffron Fixation System and the Anchorsure are both permanent anchoring devices that can be used to secure the suture to pelvic floor ligaments. Saffron has the same intended use, target population, sterilization technique, and duration of use as the predicate device. The differences between the two devices are the dimensions, provided accessories, anchor and delivery system design, and component materials. These differences do not raise different questions of safety or effectiveness. These differences were evaluated with performance testing to demonstrate substantial equivalence to the predicate device.
#### PERFORMANCE DATA VII.
The following performance data was provided in support of the substantial equivalence determination.
## Biocompatibility Testing
Biocompatibility testing was conducted based upon ISO 10993-1 (2020): Biological evaluation of medical devices – Part 1: "Evaluation and testing within a risk management process" and FDA Guidance for Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" - Guidance for Industry and Food and Drug Administration Staff - September 4, 2020. The delivery tool was evaluated for the following biocompatibility endpoints, in accordance with an external communicating device contacting tissue/bone/dentin for a limited duration (≤ 24 hours):
- Cytotoxicity per ISO 10993-5:2009 ●
- Guinea Pig Maximization Sensitization per ISO 10993-10:2010 ●
- Intracutaneous Reactivity Irritation per ISO 10993-10:2010 ●
- Acute Systemic Toxicity per ISO 10993-11:2017 ●
- . Material-Mediated Pyrogenicity per USP<151>
The anchor was evaluated for the following biocompatibility endpoints, in accordance with an implant device contacting tissue/bone for a permanent duration:
- Cytotoxicity per ISO 10993-5:2009
- Guinea Pig Maximization Sensitization per ISO 10993-10:2010
- Intracutaneous Reactivity Irritation per ISO 10993-10:2010 .
- Acute Systemic Toxicity per ISO 10993-11:2017 ●
- Material-Mediated Pyrogenicity per USP<151> ●
- . Subacute/subchronic toxicity with intramuscular implantation per ISO 10993-11:2017
- Genotoxicity per ISO 10993-3:2014
- Implantation per ISO 10993-6:2016 ●
{6}------------------------------------------------
Chemical characterization and toxicological risk assessment in accordance with ISO 10993-18:2020 was completed on the anchor to address the biocompatibility endpoints of:
- Chronic toxicity
- . Carcinogenicity
### Mechanical/ Performance Testing
The below testing was performed:
- Anchor penetration force ●
- Anchor detachment force ●
- Cannula compressive strength
- Anchor eyelet strength ●
- Anchor holding force
- Anchor insertion force
- Anchor tab torque strength
- Anchor tab shear strength
- Anchor tab fracture strength
- Anchor tab removal force
- Anchor eyelet/suture compatibility ●
- Anchor eyelet capacity
- Anchor MR compatibility assessment per 2014 FDA guidance document Establishing ● Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment
- Shaft rotation
- Delivery device maximum actuation force ●
- Delivery device feedback force
- Delivery device minimum actuation force ●
- Delivery device shaft stiffness
- Delivery device torque strength
- Delivery device tensile strength
- Delivery device plunger strength
- Visual and dimensional testing of the anchor and delivery device ●
- Label and barcode evaluation
- Sterile barrier seal strength
- Sterile barrier integrity after aging
- Full package integrity after transportation simulation per ASTM D4169-16
- Shelf Life/Expiration date
- Usability testing per the 2016 FDA guidance document Applying Human Factors and ● Usability Engineering to Medical Devices
- Non-clinical Cadaver testing
{7}------------------------------------------------
## Sterilization
The Saffron Fixation System is sterilized using ethylene oxide in a validated cycle, demonstrating a sterility assurance level (SAL) of 106. Sterilization was completed per ISO 11135:2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices and summarized in accordance with the 2016 FDA guidance document Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile.
### Packaging and Distribution
The Saffron Fixation System was subjected to packaging performance and stability testing to establish compliance with ISO 11607-1:2019 "Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems, and packaging systems". Results of testing demonstrate that the sterile barrier packaging systems can withstand the hazards of climatic stressing, transportation, and storage while adequately maintaining the functional integrity of the product and sterile barrier integrity throughout the product's shelf life of 6 months.
### Animal/Clinical Testing
No animal studies or clinical testing were provided to support substantial equivalence between the subject and predicate devices.
## VIII. CONCLUSIONS
The Saffron™ Tissue Fixation System has been demonstrated to be substantially equivalent to the predicate, Anchorsure, cleared under premarket notification number K120831, based on the non-clinical performance data provided. The Saffron™ Fixation System is as safe and effective as the predicate device to support a substantial equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.