K220407 · Visby Medical · QEP · Mar 7, 2023 · Microbiology
Device Facts
Record ID
K220407
Device Name
Visby Medical Sexual Health Test
Applicant
Visby Medical
Product Code
QEP · Microbiology
Decision Date
Mar 7, 2023
Decision
SESE
Submission Type
Dual Track
Regulation
21 CFR 866.3393
Device Class
Class 2
Indications for Use
The Visby Medical Sexual Health Test is a single-use (disposable), fully integrated, automated Polymerase Chain Reaction (PCR) in vitro diagnostic test intended for use in point-of-care or clinical laboratory settings for the rapid detection and differentiation of DNA from Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis in self-collected female vaginal swab specimens using Visby Medical Sexual Health Vaginal Specimen Collection Kit in a health care setting. The test results are to aid in the diagnosis of symptomatic or asymptomatic infections with Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis.
Device Story
Visby Medical Sexual Health Test is a single-use, fully integrated, automated PCR diagnostic device for point-of-care or clinical laboratory use. Input: self-collected female vaginal swab specimen in collection media. Operation: operator transfers sample to device port, slides switch to initiate automated lysis, PCR thermocycling, and amplicon detection. Detection: enzyme-linked colorimetric assay using streptavidin-HRP and substrate forming purple precipitate on flow cell. Output: visual interpretation of status LEDs (test progress/validity) and colorimetric results (purple spot indicates positive). Benefit: rapid (approx. 30 min) detection of CT, NG, and TV to aid clinical diagnosis. Healthcare providers use visual results to guide patient management.
Clinical Evidence
Performance evaluated via three studies: 1) External CLIA-waived site testing (n=102) using archived specimens; 2) Internal testing with trained operators (n=351) using archived/banked specimens; 3) In-house testing with contrived samples (n=80). Results showed high agreement with the predicate: CT PPA 100%, NG PPA 100%, TV PPA 93.5-100%. Initial invalid rate was 3.1%, final invalid rate 0.3%. Bench testing confirmed LoD comparability to the predicate.
Technological Characteristics
Single-use, fully integrated, automated PCR device. Principle: multiplex PCR with colorimetric detection. Reagents: lyophilized PCR reagents, target-specific primers/probes, streptavidin-HRP, colorimetric substrate. Form factor: compact, disposable device with external power supply. Connectivity: standalone. Sterilization: not specified. Software: embedded firmware for automated thermal cycling and fluidics control.
Indications for Use
Indicated for symptomatic or asymptomatic females aged 14 and older for the detection and differentiation of Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis DNA from self-collected vaginal swabs.
Regulatory Classification
Identification
A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) is an in vitro diagnostic device intended for the detection and identification of nucleic acids from non-viral microorganism(s) and their associated resistance markers in clinical specimens collected from patients suspected of sexually transmitted infections. The device is intended to aid in the diagnosis of non-viral sexually transmitted infections in conjunction with other clinical and laboratory data. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist that are not detected by the device.
Special Controls
A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) must comply with the following special controls: (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended. (2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device: alternatively, the sample collection device must be cleared in a premarket submission as a part of this device. (3) The 21 CFR 809.10(b) labeling must include: (i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens; (ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including, but not limited to. Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, with-in lab precision, and reproducibility, as appropriate; (iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing. (iv) Limiting statements indicating that: (A)a negative test result does not preclude the possibility of infection; (B) the test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician; (C) reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe procedures in any one of these steps can lead to incorrect results; and (D)if appropriate (e.g., recommended by CDC, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type. (v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that: (A)negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present; (B) detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora; (C) detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and (D) therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy. (4) Design verification and validation must include: (i) Detailed device description documentation, including, but not limited to, methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported result). (ii) Detailed documentation of analytical studies including but not limited to, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, with-in lab precision, and reproducibility, as appropriate. (iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses. (iv) A detailed description of the impact of any software, including, but not limited to, software applications and hardware-based devices that incorporate software, on the device's functions.
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate;
(iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) Limiting statements indicating that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) Reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(D) If appropriate (
*e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type; and(v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that:
(A) Negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present;
(B) Detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora;
(C) Detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and
(D) Therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy.
(4) Design verification and validation must include:
(i) Detailed device description documentation, including methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (
*e.g.,* how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies, including, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate.
(iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses.
(iv) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.
Predicate Devices
Visby Medical Sexual Health Click Test (K200748/CW200003)
Submission Summary (Full Text)
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K220407
B Applicant
Visby Medical
C Proprietary and Established Names
Visby Medical Sexual Health Test
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QEP | Class II | 21 CFR 866.3393 - Device To Detect Nucleic Acids From Non-Viral Microorganism(S) Causing Sexually Transmitted Infections And Associated Resistance Marker(S) | MI - Microbiology |
| MKZ | Class I, reserved | 21 CFR 866.3120 - Chlamydia serological reagents | MI - Microbiology |
| LSL | Class II | 21 CFR 866.3390 - Neisseria spp. direct serological test reagents | MI - Microbiology |
| OUY | Class II | 21 CFR 866.3860 - Trichomonas vaginalis nucleic acid assay | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To demonstrate that the performance of the Visby Medical Sexual Health Test (Visby Test, K220407/CW220001) is equivalent to the performance of the FDA cleared Visby Medical
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
Sexual Health Click Test (Click Test, K200748/CW200003). The Visby Test is a modified version of the Click Test. The Visby Test was developed to further simplify the user interface and to enable the automated manufacture of the device. The Visby Test has no changes to the PCR primer and probe sequences, reagent formulations, detection method, or result interpretations.
## B Measurand:
- Chlamydia trachomatis (CT) DNA
- Neisseria gonorrhoeae (NG) DNA
- Trichomonas vaginalis (TV) DNA
## C Type of Test:
Qualitative PCR
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The Visby Medical Sexual Health Test is a single-use (disposable), fully integrated, automated Polymerase Chain Reaction (PCR) in vitro diagnostic test intended for use in point-of-care or clinical laboratory settings for the rapid detection and differentiation of DNA from Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis in self-collected female vaginal swab specimens collected using Visby Medical Sexual Health Vaginal Specimen Collection Kit in a health care setting. The test results are to aid in the diagnosis of symptomatic or asymptomatic infections with Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis.
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
### D Special Instrument Requirements:
Single-use test without special instrument requirement.
## IV Device/System Characteristics:
### A Device Description:
Similar to the previously FDA cleared Visby Medical Sexual Health Click Test (referred as "Click Test" hereafter, K200748/CW200003), the Visby Medical Sexual Health Test (referred as "Visby Test" hereafter) is a single-use (disposable), fully integrated and compact device containing a PCR-based assay for direct qualitative detection and differentiation of DNA from CT, NG, and TV from self-collected female vaginal swab samples. The Visby Test was developed to further simplify the user interface and to enable the automated manufacture of the device. The Visby Test has no changes to the PCR primer and probe sequences, reagent formulations, detection method, or result interpretations.
K220407 - Page 2 of 13
{3}
The test system includes the Visby Medical Sexual Health device (see figure below), the reusable Visby Medical power adaptor, the Visby Medical Vaginal Specimen Collection kit, and fixed-volume transfer pipette. The Visby Medical Vaginal Collection kit includes a flocked swab and a tube containing the Visby Collection Media. Patients are instructed to self-collect the sample and elute it into the Visby Collection Media.

Visby Medical Sexual Health Device (Visby Test)
A test operator transfers the collection media (containing patient specimen) into the sample port of the device using the provided fixed-volume pipette and then slides a purple switch on the front of the device to both close the sample port and initiate the fully automated testing process. The device processes self-collected vaginal swab samples by automatically performing all steps required to complete lysis, polymerase chain reaction, and amplicon detection.
There are four LED lights on the lower left corner of the device that are used to communicate the test status, see below. Stable white light in the shape of a power symbol will appear indicating the test is plugged in. In addition, three progress lights to the right of the power light will each blink for $\sim 10$ minutes in sequence and then provide a stable white light to indicate the progression of the test. At the conclusion of the test, a green check mark next to the white lights signals that the test run completed successfully. A 'Red X' during or at the end of the run indicates a run failure.

When the run has completed, the operator visually interprets the status lights on the front of the device and the colorimetric output in the detection window. As shown in the figure below, a valid test will have both an illuminated green check mark and a purple spot adjacent to "RESULTS VALID" in the detection window as shown below.

K220407 - Page 3 of 13
{4}
For valid tests, a purple spot next to the name of a pathogen indicates a positive result for that target. The following figures illustrate some examples of valid test results, including a negative result for all pathogens (upper left figure) and a positive result for NG and TV (upper right figure). In addition, any shade of purple with distinct edges next to the target is a positive result (figure on the bottom).



In the case of a "Red X" error or lack of a purple "RESULTS VALID" spot (see figure below), the test is invalid, and the operator is instructed to repeat the test. If the subsequent test is also invalid, the operator is instructed to collect a new sample and repeat the test with a new device.

The device has built-in procedural controls. These include an internal process control and built-in electronic control. The internal process control monitors effective sample preparation, PCR amplification, and detection. The electronic control monitors if device is being operated outside of its temperature range or if there are any other device errors. The result of the internal process control is displayed in the detection window while the results of the electronic controls are displayed using the status lights.
Third-party external positive and negative controls are recommended for use with this test. These controls are not provided with the device and must be purchased separately by the customer directly from the manufacturer. The labeling states that the external controls "must be tested once with each new shipment received and once for each untrained operator."
K220407 - Page 4 of 13
{5}
K220407 - Page 5 of 13
## B Principle of Operation:
The device houses all the reagents to perform the testing process. The patient sample enters the lysis module where it rehydrates a lyophilized internal process control organism. DNA from the sample and the internal process control are released using a combination of chemical lysis and high temperature. The mixture then enters a mixing chamber where it rehydrates lyophilized PCR reagents, followed by 40 cycles of thermocycling to amplify target DNA. If present, the amplified pathogen target (CT, NG, and/or TV) and the amplified internal process control hybridize to specific probes located on a flow cell. Detection of the target-specific PCR product is accomplished via an enzyme-linked colorimetric assay using streptavidin bound horseradish peroxidase (HRP) and a colorimetric substrate that forms a purple precipitate. Test results can be expected in approximately 30 minutes and will be visually interpreted by the operator.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
Visby Medical Sexual Click Health Test
### B Predicate 510(k) Number(s):
K200748
### C Comparison with Predicate(s):
| Device & Predicate Device(s): | K220407 | K200748 |
| --- | --- | --- |
| Device Trade Name | Visby Medical Sexual Health Test | Visby Medical Sexual Health Click Test |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Same, only change is to the name of the device.
The Visby Medical Sexual Health Test is a single-use (disposable), fully integrated, automated Polymerase Chain Reaction (PCR) in vitro diagnostic test intended for use in point-of-care or clinical laboratory settings for the rapid detection and differentiation of DNA from Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis in self-collected female vaginal swab specimens | The Visby Medical Sexual Health Click Test is a single-use (disposable), fully integrated, automated Polymerase Chain Reaction (PCR) in vitro diagnostic test intended for use in point-of-care or clinical laboratory settings for the rapid detection and differentiation of DNA from Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis in self-collected female vaginal swab specimens collected using Visby |
{6}
| | collected using Visby Medical Sexual Health Vaginal Specimen Collection Kit in a health care setting. The test results are to aid in the diagnosis of symptomatic or asymptomatic infections with Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis. | Medical Sexual Health Vaginal Specimen Collection Kit in a health care setting. The test results are to aid in the diagnosis of symptomatic or asymptomatic infections with Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis. |
| --- | --- | --- |
| Collection/Transport Device | Same | Vaginal Specimen Collection Kit consisting of a flocked swab and a tube containing collection media (3.2 mL). |
| Specimen | Same | Self-collected female vaginal swab |
| Assay Results | Same | Qualitative |
| Organisms Detected | Same | CT, NG and TV |
| Technology | Same | PCR |
| Assay Controls | Same | Internal process control |
| Target Detection | Same | Visual interpretation |
| General Device Characteristic Differences | | |
| Device Assembly | Automated | Manual |
VI Standards/Guidance Documents Referenced:
1. Recommendations for Clinical Laboratory Improvement Amendments of 1988 (CLIA Waiver Applications for Manufacturers of In Vitro Diagnostic Devices, Guidance for Industry and Food and Drug Administration Staff. February 26, 2020
2. Guidance for the Contents of Premarket Submissions for Software Contained in Medical Devices. May 11, 2005
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Reproducibility:
K220407 - Page 6 of 13
{7}
A reproducibility study was conducted with the Visby Test at three CLIA waived study sites by a total of six untrained operators (two operators at each site). Each operator tested a panel (see table below) three times each testing day over six non-consecutive days. The panel consisted of seven members prepared by spiking cultured organisms in negative pooled clinical vaginal swab matrix. Three reagent lots were used in this study.
Panel Members Tested for Reproducibility Study
| Panel Member | Target Concentration |
| --- | --- |
| CT Moderate Positive | 4xLoD |
| CT Low Positive | 1xLoD |
| NG Moderate Positive | 4xLoD |
| NG Low Positive | 1xLoD |
| TV ModeratePositive | 4xLoD |
| TV Low Positive | 1xLoD |
| Negative | No target |
A summary of percent agreement with expected results for each panel member by site and overall is presented in the following table. The results of this study are comparable to the results observed for the Click Test (refer to K200748). This study demonstrates that untrained users could perform the Visby Test accurately including testing samples with organism concentrations near the assay LoD.
| Panel Member | Site 1 | Site 2 | Site 3 | Overall Agreement | |
| --- | --- | --- | --- | --- | --- |
| | % Agreement with Expected Results (No. Correct/ Total Tested) | | | % Agreement | 95% CI |
| CT Moderate Positive | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(108/108) | 96.6%-100.0% |
| CT Low Positive | 97.2%
(35/36) | 100.0%
(36/36) a | 100.0%
(36/36) | 99.1%
(107/108) | 94.9%-99.8% |
| NG Moderate Positive | 100.0%
(36/36) | 100.0%
(36/36) | 97.2%
(35/36) | 99.1%
(107/108) | 94.9%-99.8% |
| NG Low Positive | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(108/108) | 96.6%-100.0% |
| TV Moderate Positive | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(36/36) | 100.0%
(108/108) | 96.6%-100.0% |
| TV Low Positive | 97.2%
(35/36) | 97.2%
(35/36) b | 94.4%
(34/36) | 96.3%
(104/108) | 90.9%-98.6% |
| Negative | 97.2%
(35/36) c | 100.0%
(36/36) | 97.2%
(35/36) c | 98.1%
(106/108) | 93.5%-99.5% |
a One CT Low Positive sample was unexpectedly positive for TV
b One TV Low Positive sample was unexpected positive for CT
c One Negative sample was unexpectedly positive for TV
K220407 - Page 7 of 13
{8}
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
a. Cross-reactivity
Please refer to K200748. The cross-reactivity study was performed on the Click Test as demonstrated in K200748, and the data is applicable to the Visby Test.
b. Microbial Interference
Please refer to K200748. The Microbial Interference study was performed on the Click Test as demonstrated in K200748, and the data is applicable to the Visby Test.
c. Competitive Interference
Please refer to K200748. The Competitive Interference study was performed on the Click Test as demonstrated in K200748, and the data is applicable to the Visby Test.
d. Interfering Substances
Please refer to K200748. The Interfering Substances study was performed on the Click Test as demonstrated in K200748, and the data is applicable to the Visby Test.
4. Assay Reportable Range:
Not applicable.
This is a qualitative colorimetric test with binary output (positive/negative). The intensity of the color is not proportional to the concentration of the target organisms.
5. Specimen Stability:
Please refer to K200748. The Specimen Stability study was performed on the Click Test as demonstrated in K200748, and the data is applicable to the Visby Test.
6. Detection Limit:
To demonstrate that the LoD of the Visby Test is comparable to that of the Click Test, pooled negative clinical vaginal swab matrix spiked at the LoD established with the Click Test were prepared and tested with 20 Visby Tests and 20 Click Tests by trained operators in house. If at least 19/20 devices returned a positive test result, the organism was diluted three-fold and the testing was repeated until the detection rate was <19/20 for both devices. The LoD values were determined to be equivalent if the lowest concentrations of organism with at least 19/20 detection
K220407 - Page 8 of 13
{9}
rate from the two devices were within three-fold of each other. The results presented in the following table confirm that the LoD of the Visby Test is comparable to that of the Click Test.
LoD Comparison between Visby Test and Click Test
| Organism | LoD Multiple | Target Concentration | Detection Rate (Positive Valid Devices / Total Valid Devices Tested) | |
| --- | --- | --- | --- | --- |
| | | | Click Test | Visby Test |
| CT Serovar H (VR-879) | 1x | 16.00 EB/mL | 20/20 | 20/20 |
| | 1/3x | 5.33 EB/mL | 13/20 | 18/20 |
| CT Serovar D (VR-885) | 1x | 5.90 EB/mL | 20/20 | 20/20 |
| | 1/3x | 1.97 EB/mL | 11/20 | 19/20 |
| | 1/9x | 0.66 EB/mL | 7/20 | 10/20 |
| NG (ATCC 49226) | 1x | 6.20 cfu/mL | 19/20 | 20/20 |
| | 1/3x | 2.06 cfu/mL | 10/20 | 19/20 |
| | 1/9x | 0.68 cfu/mL | 5/20 | 9/20 |
| NG (ATCC 19424) | 3x | 17.1 cfu/mL | 20/20 | 20/20 |
| | 1x | 5.7 cfu/mL | 17/20 | 20/20 |
| | 1/3x | 1.9 cfu/mL | 17/20 | 19/20 |
| | 1/9x | 0.63 cfu/mL | 8/20 | 9/20 |
| TV (ATCC 30001)a | 1x | 1.2 troph/ mL | 19/20 | 20/20 |
| | 1/3x | 0.4 troph/mL | 20/20 | 19/20 |
| | 1/9x | 0.13 troph/mL | 16/20 | 16/20 |
| TV (ATCC 30238)b | 1x | 0.24 troph/mL | 20/20 | 20/20 |
| | 1/3x | 0.08 troph/mL | 17/20 | 17/20 |
a. Metronidazole susceptible
b. Metronidazole resistant
7. Assay Cut-Off:
Not applicable. This is a visually interpreted test where any shade of purple indicates a positive reaction.
8. Carry-Over:
Not applicable. This is a single-use test and is not subject to carryover from previously tested specimens.
B Comparison Studies:
1. Method Comparison between the Visby Test and the Click Test:
To demonstrate that the performance of the Visby Test is comparable to that of the Click Test, the following three studies were conducted:
K220407 - Page 9 of 13
{10}
- Testing performed at external CLIA waived sites. A total of 102 remnant vaginal swab specimens, a subset from the previous clinical study for Click Test, were tested with the Visby Test at external CLIA waived sites by untrained operators.
- In-house testing with trained operators. A second clinical comparison study was performed to supplement the number of positive specimens tested, with a focus on testing a sample set representing a natural distribution of the pathogen loads in the intended use population, including those with low target levels. Therefore, for this study, all remnant specimens, with sufficient volume and patient consent, from the previous clinical study of the Click Test were included in this study. The above specimens were further supplemented with archived and banked frozen self-collected samples from two other previous studies conducted by Visby Medical. These specimens were tested in house by trained operators.
- In-house study with contrived samples at near LoD target concentrations. Each individual sample was prepared in an individual negative clinical matrix, with the intention to mimic the patient specimen as much as possible.
1) Testing at external CLIA waived sites
Archived frozen vaginal swab patient samples that had been previously characterized in the clinical study for the Click Test were tested. The frozen samples were de-identified, randomized, and blinded to the study staff and test operators. The panel included 30 CT positive, 20 NG positive, 30 TV positive and 33 specimens that were negative for all three pathogens. The testing was performed with Visby Tests at three external study sites in the United States representing CLIA waived settings by a total of six untrained operators (two operators at each site). Operators thawed the samples and tested them on-site by following the instructions in the QRG. The performance of the Visby Test was evaluated by comparing the Visby Test results to the results from Click Test recorded during the previous clinical study. Of the 102 specimens (seven of these samples are double positive and two are triple positive) included in the study, two (2.4%) had an initial invalid test result. One sample was excluded from the performance evaluation because it was invalid upon retesting. There were no additional exclusions. Please refer to the following table for performance evaluation.
Visby Test vs. Click Test Performance Comparison (External CLIA Waived Sites, Testing Archived Specimens)
| Target | N | TP | FP | TN | FN | PPA (95% CI) | NPA (95% CI) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| CT | 101 | 30 | 0 | 71 | 0 | 100.0%
(88.6%-100.0%) | 100%
(94.9%-100.0%) |
| NG | 101 | 20 | 0 | 81 | 0 | 100.0%
(83.9%-100.0%) | 100.0%
(95.5%-100.0%) |
| TV | 101 | 30 | 3 | 68 | 0 | 100.0%
(88.6%-100.0%) | 95.8%
(88.3%-98.6%) |
PPA=Positive Percent agreement; NPA=Negative Percent Agreement.
TP=true positive; FP=false positive; TN=true negative; FN=false negative.
2) In-house testing with trained operators
K220407 - Page 10 of 13
{11}
All specimens meeting the inclusion criteria (see below) from the previous clinical study of Click Test were included in this study. To obtain a sufficient number of positive specimens (by the comparator assays) for each analyte, these specimens were further supplemented with archived and banked frozen self-collected samples meeting the same inclusion criteria from two other previous studies conducted by Visby Medical. Target negative patient specimens were also included in this study, and all positive and negative specimens were de-identified, randomized and blinded to the operators.
The inclusion criteria for this study:
- Self-collected vaginal swab in Visby Medical Collection Media from a female subject ≥14 years of age at the time of specimen collection.
- Age and symptomatic status of the subject at the time of specimen collection is available.
- Swab specimen was obtained from a collection wherein the subject consented/assented to future unspecified research use.
- Minimum sample volume of 1.0 mL.
A total of 359 de-identified archived frozen self-collected vaginal swab specimens were provided to the operators for testing. One specimen had less than 1 mL of volume and thus could not be tested. Of the remaining 358 specimens, seven were excluded from performance evaluation due to the following listed reasons.
- One specimen had invalid result for both the initial and retest on the Visby Test and thus was excluded.
- Five specimens yielded invalid results on the Visby Test and there was insufficient volume to run a retest, thus these specimens were excluded.
- One specimen had invalid result for both the initial and retest on the Click Test and thus was excluded.
Therefore, a total of 351 specimens were included in the performance evaluation.
Testing with Visby devices was conducted by seven trained operators internally at Visby Medical over seven days under variable lighting conditions throughout the day. When sufficient sample volume was present, specimens were tested on both the Click Test and the Visby Test to allow for performance comparison. Specimens without sufficient volume (27/351) to run on both devices were tested only on the Visby Test, and the original Click Test results from the previous Click Test clinical studies were used for the comparison. Please refer to the following table for performance estimates calculations.
K220407 - Page 11 of 13
{12}
Visby Test vs. Click Test Performance Comparison (Internal Testing of Archived Specimens)
| Visby Test vs. Click Test | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Target | N | TP | FP | TN | FN | PPA | NPA |
| CT | 351 | 116 | 3 | 232 | 0 | 100.0% (95% CI: 96.8% - 100.0%) | 98.7% (95% CI: 96.3% - 99.6%) |
| NG | 351 | 34 | 0 | 317 | 0 | 100.0% (95% CI: 89.9% - 100.0%) | 100.0% (95% CI: 98.8% - 100.0%) |
| TV | 351 | 100 | 4 | 240 | 7^{a} | 93.5% (95% CI: 87.1% - 96.8%) | 98.4% (95% CI: 95.9% - 99.4%) |
a. TV PIS (based on the data from the clinical study for Click Test) for all seven specimens were negative.
Note that 4 CT, 3 NG, and 6 TV positive specimens tested at the external sites in the first study had sufficient volume and were also tested in the internal study. Therefore, a total of 142 CT, 51 NG and 124 TV positive patient specimens (by the comparator assays) were tested in the above two Clinical Comparison studies.
The initial invalid rate for Visby Test is 3.1% (11/358) and the final invalid rate is 0.3% (1/353, five out the 11 initial invalids did not have sufficient volume for retest).
3) In-house testing with contrived samples
20 contrived samples with target concentrations at 1.5x LoD (N=6), 2x LoD (N=10) and 3x LoD (N=4) for each analyte were prepared in individual negative clinical matrices. An additional 20 negatives, prepared with pooled negative clinical matrix collected from consented subjects under IRB-approved protocols, were included in the study. All samples were randomized, blinded and tested on both the Click Test as well as the Visby Test in-house. All 80 contrived samples yielded valid results (no initial invalids) and were included in the final data analysis. The results of this study are presented in the table below.
Visby Test vs. Click Test Performance Comparison (Internal Testing of Contrived Samples)
| Target | Concentration | Correct Results/Total Tested | |
| --- | --- | --- | --- |
| | | Click Test | Visby Test |
| CT
Serovar H
(VR-879) | 1.5x LoD | 6/6 | 5/6 |
| | 2x LoD | 10/10 | 9/10 |
| | 3x LoD | 4/4 | 4/4 |
| NG
(ATCC
49226) | 1.5x LoD | 5/6 | 5/6 |
| | 2x LoD | 9/10 | 10/10 |
K220407 - Page 12 of 13
{13}
| | 3x | 4/4 | 4/4 |
| --- | --- | --- | --- |
| TV (ATCC 30001) | 1.5x | 6/6 | 6/6 |
| | 2x | 10/10 | 10/10 |
| | 3x | 4/4 | 4/4 |
| Negative | N/A | 20/20 | 19/20^{a} |
a. One device was unexpectedly positive for TV.
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
For the complete clinical data set in support of the Click Test, refer to K200748.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Please refer to K200748 for Expected Values for the Click Test.
F Other Supportive Instrument Performance Characteristics Data:
Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K220407 - Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.