The EasyStat 300 is designed for clinical laboratory use, making quantitative measurements of potassium (K+), ionized calcium (Ca++), and chloride (Cl-) in whole blood (arterial/venous) samples from Li-Heparinized Syringes. This Analyzer should only be used by trained technicians in clinical laboratories to aid in the diagnosis and treatment of patients with electrolyte and/or acid-base disturbances. Potassium (K+) measurements are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels. Calcium (Ca++) (ionized) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). Chloride (Cl-) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Device Story
EasyStat 300 is a clinical laboratory analyzer for rapid measurement of K+, Ca++, and Cl- in whole blood (175µL sample). It utilizes Ion Selective Electrode (ISE) sensors; automated calibration via Reagent Module (solutions A2, B2, C2) establishes sensor slopes. Operators use the device in clinical labs; results are used to monitor electrolyte balance and diagnose metabolic disorders. The system stores QC data (Levey-Jennings plots) and supports optional barcode scanning for sample entry. It provides calculated parameters based on measured electrolytes. Clinical benefit includes rapid, accurate electrolyte assessment to guide treatment of critical conditions like tetany or acidosis.
Clinical Evidence
Bench testing only. Studies included 5-day repeatability and 20-day reproducibility precision, linearity (CLSI EP06-A), method comparison (CLSI EP9-A2) against EasyStat (k063376) using 198 samples, sensitivity (CLSI EP17-A), and selectivity/interference testing. Results showed high correlation (R2 > 0.98) and precision within specifications for all analytes. No adverse effects reported.
Technological Characteristics
Ion Selective Electrode (ISE) sensing principle for potentiometric measurement of K+, Ca++, and Cl-. Aqueous calibrant base. Dimensions/form factor: clinical benchtop analyzer. Connectivity: USB (5 ports), Ethernet, RS232, SD card. Software: embedded system for automated calibration, QC tracking, and result calculation. Sensors have one-year shelf life.
Indications for Use
Indicated for patients requiring electrolyte monitoring for acid-base disturbances, parathyroid disease, bone disease, chronic renal disease, tetany, cystic fibrosis, or diabetic acidosis. For use with whole blood (arterial/venous) samples from Li-Heparinized Syringes. Contraindicated for patients taking perchlorate.
Regulatory Classification
Identification
A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K220396
B Applicant
Medica corporation
C Proprietary and Established Names
EasyStat 300
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JFP | Class II | 21 CFR 862.1145 - Calcium test system | CH - Clinical Chemistry |
| CGZ | Class II | 21 CFR 862.1170 - Chloride test system | CH - Clinical Chemistry |
| CEM | Class II | 21 CFR 862.1600 - Potassium test system | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
Potassium (K+), Chloride (Cl-), ionized Calcium (Ca++)
C Type of Test:
Quantitative, ion specific electrodes
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The EasyStat 300 is designed for clinical laboratory use, making quantitative measurements of potassium (K+), ionized calcium (Ca++), and chloride (Cl-) in whole blood (arterial/venous) samples from Li-Heparinized Syringes. This Analyzer should only be used by trained technicians in clinical laboratories to aid in the diagnosis and treatment of patients with electrolyte and/or acid-base disturbances.
Potassium (K+) measurements are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
Calcium (Ca++) (ionized) measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Chloride (Cl-) measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For in vitro diagnostic use only
D Special Instrument Requirements:
EasyStat 300 Analyzer
IV Device/System Characteristics:
A Device Description:
The candidate device is a small bench-top analyzer for use by health care professionals. The analyzer incorporates a replaceable EasyStat 300 ISE cartridge that is comprised of sensors for measurement of potassium, chloride, and calcium. The analyzer also incorporates a replaceable reagent module (i.e., EASYSTAT 300 REAGENT MODULE) containing calibrating solutions for the sensors. The reagent module contains encoded reagent information (calibration values and expiration date), which is read by the analyzer upon installation of the reagent module. Calibrations are performed automatically or on-demand by the user. The analyzer draws 175μL of whole blood (lithium heparinized venous whole blood and arterial blood) when operated in the syringe mode. Quality control materials (i.e., EasyQC BGME Level 1, 2, and 3), cleaning solution, replacement reference electrode, bubble detection, and troubleshoot kit are provided for the system.
K220396 - Page 2 of 11
{2}
The EasyStat 300 Analyzer was also cleared for making quantitative measurements of pO2 (partial pressure of oxygen), pCO2 (partial pressure of carbon dioxide), and pH (hydrogen ion activity) in K211559.
## B Principle of Operation:
The potassium, chloride, and ionized calcium parameters are measured by ion-selective electrodes (ISE). When the ISE is in fluid contact with a sample, an electrochemical potential is developed versus a reference electrode, and is proportional to the logarithm of the ionic activity, as described by the Nernst equation.
## V Substantial Equivalence Information:
A Predicate Device Name(s): EasyLyte Calcium/Chloride Analyzer
B Predicate 510(k) Number(s): K963694
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K220396 | K963694 |
| --- | --- | --- |
| Device Trade Name | EasyStat 300 | EasyLyte Calcium/Chloride Analyzer |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | For the measurement of ionized calcium and chloride. | Same |
| General Device Characteristic Differences | | |
| Analytes Measured | Potassium, Ionized Calcium, and Chloride | Ionized Calcium and Chloride |
The sponsor referenced K063376 to support the substantial equivalence of potassium measurements.
## VI Standards/Guidance Documents Referenced:
Clinical and Laboratory Standards Institute (CLSI) EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition.
CLSI EP07 Interference Testing in Clinical Chemistry. 3rd Edition.
K220396 - Page 3 of 11
{3}
CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures. 2nd Edition.
CLSI EP37 Supplemental Tables for Interference Testing in Clinical Chemistry. 1st Edition.
## VII Performance Characteristics (if/when applicable):
## A Analytical Performance:
### 1. Precision/Reproducibility:
The precision performance of the EasyStat 300 was evaluated in two studies.
**Study #1 - Within-Device using controls**
In the study, three samples of aqueous control material were assayed on each of three analyzers in duplicates per run, two runs per day for 20 days, for a total of 80 measurements per sample per analyzer. The samples were tested on the analyzer operated in syringe mode. For each of the three analyzers, the data was analyzed for repeatability (i.e., within-run) and within-device precision. The results from one representative analyzer are summarized below.
| Analyte | Level | Mean, mmol/L | N | Repeatability | | Within-Device | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | CV | SD | CV |
| Potassium | 1 | 2.6 | 80 | 0.00 | 0.1% | 0.01 | 0.3% |
| | 2 | 4.2 | 80 | 0.01 | 0.2% | 0.01 | 0.2% |
| | 3 | 6.0 | 80 | 0.01 | 0.2% | 0.01 | 0.2% |
| Chloride | 1 | 79 | 80 | 0.0 | 0.1% | 0.2 | 0.2% |
| | 2 | 101 | 80 | 0.1 | 0.1% | 0.1 | 0.1% |
| | 3 | 130 | 80 | 0.1 | 0.1% | 0.1 | 0.1% |
| Calcium | 1 | 1.73 | 80 | 0.00 | 0.2% | 0.01 | 0.3% |
| | 2 | 1.12 | 80 | 0.00 | 0.3% | 0.00 | 0.4% |
| | 3 | 0.53 | 80 | 0.00 | 0.3% | 0.00 | 0.3% |
**Study #2 - Venous whole blood**
In the study, with each analyte (K+, Cl-, and Ca++) and for each of five days, a lithium heparin venous whole blood sample was collected from a donor, and each day a different donor. Each sample per day was prepared into three analyte levels with normal analyte (n=5), spiked (n=5), and diluted (n=5) for a total of 15 samples of each level. On each of five days, the samples were each assayed in replicates of five using each of three analyzers (n=25 for each sample tested). The analyzer was operated in the syringe mode. The results were analyzed for each sample using a fully-nested random effects variance components model to compute the mean result, and the within-sensor (repeatability), between-analyzer and total (reproducibility or within-laboratory) standard deviations and associated %CVs were estimated, using the methodology as recommended in CLSI guidance document EP05-A3.
K220396 - Page 4 of 11
{4}
The results summarized as follows:
| K+ (Syringe Mode) | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| level | Mean | Repeatability | | Between-Analyzer | | Within Laboratory | |
| | | SD | %CV | SD | %CV | SD | %CV |
| Low 1 | 2.46 | 0.004 | 0.2% | 0.017 | 0.7% | 0.018 | 0.7% |
| Low 2 | 1.86 | 0.011 | 0.6% | 0.016 | 0.8% | 0.019 | 1.0% |
| Low 3 | 2.18 | 0.004 | 0.2% | 0.013 | 0.6% | 0.014 | 0.6% |
| Low 4 | 2.60 | 0.009 | 0.3% | 0.020 | 0.8% | 0.022 | 0.8% |
| Low 5 | 1.75 | 0.010 | 0.6% | 0.016 | 0.9% | 0.019 | 1.1% |
| Normal 1 | 4.67 | 0.022 | 0.5% | 0.022 | 0.5% | 0.041 | 0.9% |
| Normal 2 | 4.05 | 0.015 | 0.4% | 0.004 | 0.1% | 0.022 | 0.5% |
| Normal 3 | 4.53 | 0.044 | 1.0% | 0.002 | 0.1% | 0.045 | 1.0% |
| Normal 4 | 5.14 | 0.018 | 0.4% | 0.000 | 0.0% | 0.041 | 0.8% |
| Normal 5 | 4.59 | 0.022 | 0.5% | 0.014 | 0.3% | 0.034 | 0.7% |
| High 1 | 8.13 | 0.025 | 0.3% | 0.094 | 1.2% | 0.131 | 1.6% |
| High 2 | 6.88 | 0.035 | 0.5% | 0.000 | 0.0% | 0.064 | 0.9% |
| High 3 | 8.53 | 0.189 | 2.2% | 0.076 | 0.9% | 0.203 | 2.4% |
| High 4 | 6.92 | 0.027 | 0.4% | 0.000 | 0.0% | 0.066 | 0.9% |
| High 5 | 8.38 | 0.069 | 0.8% | 0.080 | 1.0% | 0.119 | 1.4% |
| Cl- (Syringe Mode) | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| level | Mean | Repeatability | | Between-Analyzer | | Within Laboratory | |
| | | SD | %CV | SD | %CV | SD | %CV |
| Low 1 | 72.46 | 0.18 | 0.2% | 0.00 | 0.0% | 0.25 | 0.3% |
| Low 2 | 85.79 | 0.21 | 0.2% | 0.00 | 0.0% | 0.21 | 0.2% |
| Low 3 | 71.43 | 0.52 | 0.7% | 0.09 | 0.1% | 0.52 | 0.7% |
| Low 4 | 70.96 | 0.37 | 0.5% | 0.37 | 0.5% | 0.52 | 0.7% |
| Low 5 | 68.49 | 0.30 | 0.4% | 0.18 | 0.3% | 0.35 | 0.5% |
| Normal 1 | 103.13 | 0.20 | 0.2% | 0.29 | 0.3% | 0.35 | 0.3% |
| Normal 2 | 100.18 | 0.26 | 0.3% | 0.36 | 0.4% | 0.45 | 0.4% |
| Normal 3 | 103.17 | 0.21 | 0.2% | 0.57 | 0.6% | 0.63 | 0.6% |
| Normal 4 | 99.97 | 0.30 | 0.3% | 0.58 | 0.6% | 0.66 | 0.7% |
| Normal 5 | 99.23 | 0.31 | 0.3% | 0.22 | 0.2% | 0.41 | 0.4% |
| High 1 | 123.74 | 0.25 | 0.2% | 0.55 | 0.4% | 0.61 | 0.5% |
| High 2 | 121.81 | 0.20 | 0.2% | 0.66 | 0.5% | 0.73 | 0.6% |
| High 3 | 125.66 | 0.21 | 0.2% | 0.87 | 0.7% | 0.94 | 0.7% |
| High 4 | 123.84 | 0.24 | 0.2% | 0.81 | 0.7% | 0.88 | 0.7% |
| High 5 | 117.09 | 0.21 | 0.2% | 0.65 | 0.6% | 0.73 | 0.6% |
K220396 - Page 5 of 11
{5}
| Ca++ (Syringe Mode) | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| level | Mean | Repeatability | | Between-Analyzer | | Within Laboratory | |
| | | SD | %CV | SD | %CV | SD | %CV |
| Low 1 | 0.48 | 0.000 | 0.0% | 0.006 | 1.2% | 0.006 | 1.2% |
| Low 2 | 0.45 | 0.003 | 0.6% | 0.002 | 0.4% | 0.003 | 0.7% |
| Low 3 | 0.43 | 0.000 | 0.0% | 0.000 | 0.0% | 0.006 | 1.3% |
| Low 4 | 0.42 | 0.002 | 0.4% | 0.000 | 0.0% | 0.005 | 1.3% |
| Low 5 | 0.42 | 0.002 | 0.4% | 0.003 | 0.8% | 0.006 | 1.3% |
| Normal 1 | 1.27 | 0.004 | 0.4% | 0.011 | 0.9% | 0.013 | 1.0% |
| Normal 2 | 1.25 | 0.004 | 0.4% | 0.009 | 0.7% | 0.010 | 0.8% |
| Normal 3 | 1.31 | 0.009 | 0.7% | 0.005 | 0.4% | 0.010 | 0.8% |
| Normal 4 | 1.29 | 0.003 | 0.2% | 0.004 | 0.3% | 0.006 | 0.4% |
| Normal 5 | 1.31 | 0.008 | 0.6% | 0.003 | 0.3% | 0.009 | 0.7% |
| High 1 | 2.60 | 0.008 | 0.3% | 0.038 | 1.4% | 0.042 | 1.6% |
| High 2 | 2.20 | 0.008 | 0.4% | 0.016 | 0.7% | 0.020 | 0.9% |
| High 3 | 2.61 | 0.045 | 1.7% | 0.034 | 1.3% | 0.056 | 2.2% |
| High 4 | 2.50 | 0.006 | 0.2% | 0.012 | 0.5% | 0.018 | 0.7% |
| High 5 | 2.23 | 0.012 | 0.5% | 0.009 | 0.4% | 0.015 | 0.7% |
# 2. Linearity:
A linearity study was conducted using a linearity set of 9 samples prepared from lithium heparin venous whole blood samples. Each of the 9 samples was assayed in triplicate on each of three analyzers. The study was conducted with the analyzers operated in syringe mode. For potassium, chloride, and calcium, regression analysis demonstrated first order linearity. The results of the linear regression analysis from one representative analyzer are summarized below:
| Analyte | Claimed Measuring Range (mmol/L) | Sample Range Tested (mmol/L) | Slope | Intercept | R2 |
| --- | --- | --- | --- | --- | --- |
| Potassium | 1.0 - 20 | 0.51 - 21.12 | 0.99 | 0.12 | 0.999 |
| Chloride | 50 - 150 | 46.6 - 165.8 | 0.97 | -0.36 | 1.000 |
| Calcium | 0.25 - 5.0 | 0.22 - 5.28 | 0.96 | 0.01 | 1.000 |
# 3. Analytical Specificity/Interference:
The analytical specificity performance of the EasyStat 300 Analyzer for potassium, chloride, and calcium was evaluated by testing for interference from endogenous and exogenous substances. In the study, lithium heparin venous whole blood samples were prepared at two analyte concentrations. Each sample was further divided into two aliquots; i.e. test sample (with added interferent) and control sample (with no added interferent). Each test and control sample was assayed in replicates of five on each of three analyzers operated in syringe mode. A substance was identified as an interferent if the difference in the mean between the test and control sample was outside of the predefined allowable difference, as shown below:
K220396 - Page 6 of 11
{6}
The following tables list the concentrations of each substance at which no significant interference was found.
Potassium
| Substance | Highest concentration tested at which no significant interference is observed |
| --- | --- |
| Acetaminophen | 16 mg/dL |
| Ammonium (Chloride) | 0.151 mmol/L |
| Benzalkonium (Chloride) | 5 mg/dL |
| (Sodium) Bromide | 37.5 mmol/L |
| Bilirubin, conjugated | 20 mg/dL |
| Calcium (Chloride) | 5.0 mmol/L |
| (Sodium) Citrate | 12 mmol/L |
| Ethanol | 130 mmol/L |
| Heparin-Na | 330 U/dL |
| Hydroxyurea | 3.08 mg/dL |
| Intralipid® | 1% |
| Ipratropium Bromide | 0.08 mg/L |
| Lithium (Chloride) | 3.2 mmol/L |
| Magnesium (Chloride) | 4.1 mmol/L |
| pH | 8.0 pH units |
| (Sodium) Perchlorate | 20 mg/dL |
| (Sodium) Salicylate | 4.3 mmol/L |
| Na (Chloride) | 170 mmol/L |
Chloride
| Substance | Highest concentration tested at which no significant interference is observed |
| --- | --- |
| Acetaminophen | 16 mg/dL |
| (Sodium) Bromide | 1.5 mmol/L |
| Bilirubin, conjugated | 20 mg/dL |
| Ethanol | 130 mmol/L |
| (Sodium) Fluoride | 63.2 μmol/L |
| Heparin-Na | 330 U/dL |
| Hydroxyurea | 3.08 mg/dL |
| Ibuprofen | 1.06 mM |
| Intralipid® | 1% |
| (Potassium) Iodide | 0.55 mmol/L |
| Ipratropium Bromide | 0.08 mg/L |
K220396 - Page 7 of 11
{7}
| Substance | Highest concentration tested at which no significant interference is observed |
| --- | --- |
| (Potassium) Oxalate | 12.4 mg/L |
| pH | 8.0 units |
| (Sodium) Perchlorate | 20 mg/dL |
| (Sodium) Salicylate | 4.3 mmol/L |
Calcium
| Substance | Highest concentration tested at which no significant interference is observed |
| --- | --- |
| Acetaminophen | 16 mg/dL |
| Ammonium (Chloride) | 0.151 mmol/L |
| Benzalkonium (Chloride) | 0.5 mg/dL |
| (Sodium) Bromide | 37.5 mmol/L |
| Bilirubin, conjugated | 20 mg/dL |
| Ethanol | 130 mmol/L |
| Heparin-Na | 330 U/dL |
| Hydroxyurea | 3.08 mg/dL |
| Ibuprofen | 1.06 mM |
| Intralipid® | 1% |
| (Potassium) Iodide | 3 mmol/L |
| Ipratropium Bromide | 0.08 mg/L |
| Lithium (Chloride) | 3.2 mmol/L |
| Magnesium (Chloride) | 12.35 mg/dL |
| (Potassium) Oxalate | 12.4 mg/L |
| pH | 8.0 units |
| (Sodium) Salicylate | 4.3 mmol/L |
| Na (Chloride) | 170 mmol/L |
| (Potassium) Thiocyanate | 898 μmol/L |
For those substances that on initial screening were found to interfere, dose response testing was conducted to establish the concentration limit below which no significant interference is expected. The results are summarized in the table below:
| Substance | Affected Analyte | Interferant concentration | Interference |
| --- | --- | --- | --- |
| Fluoride | Potassium | ≥ 1.0 mmol/L | Decreased results |
| Ibuprofen | Potassium | ≥ 0.8 mM | Increased results |
| Perchlorate | Calcium | ≥ 6.0 mg/dL | Decreased results |
| Perchlorate | Chloride | ≥ 6.0 mg/dL | Decreased results |
| Thiocyanate | Chloride | ≥ 225 μmol/L | Increased results |
| Thiosulfate | Potassium | ≥ 5.0 mmol/L | Decreased results |
| Thiosulfate | Chloride | ≥ 5.0 mmol/L | Decreased results |
| Thiosulfate | Calcium | ≥ 5.0 mmol/L | Decreased results |
| Bromide | Chloride | ≥ 1.5 mmol/L | Increased results |
K220396 - Page 8 of 11
{8}
| Substance | Affected Analyte | Interferant concentration | Interference |
| --- | --- | --- | --- |
| Citrate | Chloride | ≥ 3.0 mmol/L | Decreased results |
| Intralipid | Potassium | ≥ 1.0 g/dL | Decreased results |
| Iodide | Chloride | ≥ 0.65 mmol/L | Increased results |
| pH | Calcium | ≥ 7.55 | Decreased results |
The sponsor added the following limitations to their labeling:
- Potassium levels should not be relied upon in patients suspected of ibuprofen toxicity, as very high concentrations of ibuprofen may cause positive bias for potassium.
- Bromide and iodide from therapeutic drugs and ointments may cause a positive bias for chloride. Normal physiological levels of bromide and iodide do not interfere.
- This device should not be used with patients taking perchlorate.
4. **Assay Reportable Range:**
See linearity.
5. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
**Traceability**
The metrological traceability of the EasyStat 300 for potassium, chloride, and calcium was reviewed and found acceptable.
6. **Detection Limit:**
The performance at the lower end of the measuring range for potassium and chloride is supported by the linearity studies (see section VII.A.2 above). The sponsor performed a study to evaluate the limit of blank (LoB), limit of detection (LoD) and limit of quantification (LoQ) for ionized calcium following the recommendations in CLSI EP17-A2.
**LoB**
In the study, five pooled plasma samples ultracentrifuged to lower the calcium concentration were assayed in replicates of two with each of two reagent lots on one analyzer over three days, for a total of 60 measurements. The LoB was determined using parametric analysis as described in CLSI EP17-A2.
**LoD**
In the study, five pooled plasma samples ultrafiltered to lower the calcium concentration near the LoD were assayed in replicates of two with each of two reagent lots on one analyzer over three days, for a total of 60 measurements. The LoD was analyzed using parametric data analysis as described in CLSI EP17-A2.
**LoQ**
In the study, four pooled samples with concentrations near the expected LoQ were prepared, and then analyzed with each of two reagent lot on one EasyStat 300 analyzer with three replicates per day over three consecutive days total of 18 replicates per sample. The LoQ was
K220396 - Page 9 of 11
{9}
determined as the measured sample concentration meeting the total allowable error goals of $\pm 0.10\mathrm{mmol / L}$.
The detection limit studies support the claimed measuring range of $0.25 - 5.0\mathrm{mmol / L}$ for ionized calcium.
## 7. Assay Cut-Off:
Not applicable.
## B Comparison Studies:
### 1. Method Comparison with Predicate Device:
The agreement accuracy of the candidate device was evaluated in a method comparison study versus a comparator device. In the study, lithium heparin venous and arterial whole blood samples were collected from 198 subjects across two sites (one hospital and one internal), and each tested in duplicate on the EasyStat 300 Analyzer in syringe mode, and in duplicate on the comparator. The average of the two readings using the comparator and the first of the two EasyStat 300 Analyzer reading was used in the data analysis by linear regression. To cover the measurement range, no more than $10\%$ of the samples were diluted and/or spiked. The sponsor provided acceptable evidence to support pooling the results from venous and arterial samples in the same analysis. The results of the method comparison study are summarized as follows.
Linear regression
| Analyte | Slope | Intercept | R² | Range Tested |
| --- | --- | --- | --- | --- |
| Potassium | 0.962 | 0.13 | 0.997 | 1.45 - 18.24 mmol/L |
| Chloride | 1.007 | -0.38 | 0.981 | 64.3 - 145.4 mmol/L |
| Calcium | 0.987 | -0.02 | 0.994 | 0.44 - 4.5 mmol/L |
Bias at medical decision levels (MDL)
| Analyte | N | MDL (mmol/L) | Bias at MDL |
| --- | --- | --- | --- |
| Potassium | 198 | 3.0 | 0.02 |
| | | 5.8 | -0.09 |
| | | 7.5 | -2.1% |
| Chloride | 198 | 90 | 0.3% |
| | | 112 | 0.4% |
| Calcium | 198 | 0.37 | -0.02 |
| | | 0.82 | -0.03 |
| | | 1.58 | -2.6% |
### 2. Matrix Comparison:
Not applicable, the device is only intended for use with lithium heparin whole blood.
K220396 - Page 10 of 11
{10}
K220396 - Page 11 of 11
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data:
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
The EasyStat 300 Analyzer operator manual lists the following expected values from scientific literature for the normal ranges in arterial, mixed venous, or venous samples:
Potassium: 3.5 – 5.1 mmol/L
Chloride: 98.0 – 106.0 mmol/L
Calcium: 1.05 – 1.32 mmol/L
Sources:
Tietz: Fundamentals of Clinical Chemistry, 4th edition, (1996)
B. Statland: Clinical Decision Levels for Lab Test, 2nd edition
Fink: Clinical Practice in Respiratory Care
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.