The i-STAT PTplus cartridge is intended for use in the in vitro quantitative measurement of the clot time of the extrinsic coagulation pathway when activated by thromboplastin in non-anticoagulated whole blood (venous or capillary), using the i-STAT 1 System. Measurements of prothrombin time are used to aid in the monitoring of patients receiving anticoagulant therapy with coumarin derivatives. The i-STAT PTplus Prothrombin Time test result is reported in seconds and as an International Normalized Ratio (INR). The test is intended for point of care use and is for prescription use only.
Device Story
i-STAT PTplus cartridge is a single-use, self-contained disposable unit for the i-STAT 1 System. It accepts ~20 µl of non-anticoagulated whole blood (venous or capillary). Inside the cartridge, tissue thromboplastin activates the extrinsic coagulation cascade. Electrochemical sensors detect thrombin activity via amperometric measurement of thrombin substrate conversion to fibrin. The i-STAT 1 analyzer processes these signals to calculate PT (seconds) and INR. Used by trained medical professionals at point-of-care or clinical labs. Output assists clinicians in monitoring coumarin derivative anticoagulant therapy. System includes internal/external electronic simulators for quality control.
Clinical Evidence
Clinical method comparison study conducted at three sites using venous and capillary whole blood from patients on coumarin therapy and healthy subjects. Compared against the CoaguChek XS predicate and Sysmex CS-2500 reference method. Results showed strong correlation (R=0.97-0.98 vs. CoaguChek; R=0.91-0.92 vs. Sysmex). Precision evaluated over 20 days and via whole blood repeatability; within-laboratory CVs for INR ranged from 1.7% to 6.8%. Sensitivity to factors, hematocrit, and interfering substances was assessed.
Technological Characteristics
Handheld, point-of-care system. Cartridge contains electrochemical sensors and human recombinant thromboplastin. Amperometric detection principle. Sample volume ~20 µl. Factory calibrated. Connectivity via i-STAT 1 analyzer (wireless or wired). Shelf-life 193 days at 2–8°C. Liquid quality controls (lyophilized citrated human plasma) available.
Indications for Use
Indicated for patients receiving anticoagulant therapy with coumarin derivatives requiring monitoring of the extrinsic coagulation pathway. Intended for use with non-anticoagulated venous or capillary whole blood.
Regulatory Classification
Identification
A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K220282
B Applicant
Abbott Laboratories
C Proprietary and Established Names
i-STAT PT*plus* Cartridge with i-STAT 1 System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| GJS | Class II | 21 CFR 864.7750 - Prothrombin Time Test | HE - Hematology |
## II Submission/Device Overview:
A Purpose for Submission:
Clearance of a new device
B Measurand:
Prothrombin Time (PT)
International Normalized Ratio (INR) (seconds)
C Type of Test:
Quantitative
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
B Indication(s) for Use:
The i-STAT PT$^{plus}$ cartridge is intended for use in the in vitro quantitative measurement of the clot time of the extrinsic coagulation pathway when activated by thromboplastin in non-anticoagulated whole blood (venous or capillary), using the i-STAT 1 System. Measurements of prothrombin time are used to aid in the monitoring of patients receiving anticoagulant therapy with coumarin derivatives. The i-STAT PT$^{plus}$ Prothrombin Time test result is reported in seconds and as an International Normalized Ratio (INR). The test is intended for point of care use and is for prescription use only.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
i-STAT 1 analyzer (K001387)
i-STAT 1 Wireless Analyzer (K103195)
IV Device/System Characteristics:
A Device Description:
The i-STAT 1 System consists of the i-STAT 1 analyzer and the i-STAT cartridges. The i-STAT PT$^{plus}$ cartridge is a coagulation cartridge for determining the time required for complete activation of the extrinsic coagulation cascade. The cartridge contains electrochemical sensors and test reagents that must be mixed with the sample. The reagents include the reactive ingredient to activate the coagulation cascade as well as electrochemical markers that generate a sensor signal when the cascade is fully activated.
The analysis time of the i-STAT PT$^{plus}$ cartridge is up to 300 seconds or 5 minutes. The sample volume required for the i-STAT PT$^{plus}$ cartridge is approximately 20 µl of whole blood (venous or capillary) without added anticoagulant. The i-STAT PT$^{plus}$ cartridge is a single-use disposable unit that is self-contained. The test reagents and sample fluids do not contact the instrument or user. All the test steps and fluid movements occur within the cartridge.
Other components of the i-STAT 1 System are the Electronic Simulator, the i-STAT 1 Downloader/Recharger and the i-STAT Printer. The i-STAT 1 analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The system is designed for use by trained medical professionals at the patient point of care or in the clinical laboratory and is for prescription use only. The i-STAT 1 System has an internal quality control (internal simulator) and an external quality control (Electronic Simulator). The internal and external simulators are used to check the instrument signal-reading function. In addition to the quality controls within the i-STAT 1 System, liquid quality controls are available as an optional quality control methodology to meet the regulatory compliance requirements applicable to the facility where they are to be used.
The liquid quality controls are the i-STAT PT$^{plus}$ Control Levels 1 and 2 and can be used for the quality control of the i-STAT PT$^{plus}$ cartridge. The coagulation controls consist of lyophilized citrated human plasma and calcium chloride fluid for reconstitution. i-STAT PT$^{plus}$ Control Level 1 has been formulated to produce a normal prothrombin time. Level 2 has been formulated to produce an extended prothrombin time. Each level of control is packaged as a box of 5 vials
K220282 - Page 2 of 13
{3}
containing 1 mL of lyophilized citrated human plasma and 5 vials of 1.5 mL of calcium chloride diluent.
The i-STAT PTplus controls are intended for use with the i-STAT PTplus cartridge on the i-STAT System, and values assigned to these controls may not be commutable with other commercial methods.
## B Principle of Operation:
The i-STAT PTplus cartridge with the i-STAT 1 Analyzer uses electrochemical technology with amperometric (electric current) detection of thrombin activity to generate a clot time. The prothrombin time test result is reported as an International Normalized Ratio (INR) and/or in seconds.
In a prothrombin time test, coagulation is initiated by exposing the sample to tissue thromboplastin. In the i-STAT Prothrombin time test, complete activation is indicated when activated thrombin converts a thrombin substrate to fibrin and extensive or localized clots are detected by an electrochemical sensor.
## V Substantial Equivalence Information:
## A Predicate Device Name(s):
Coaguchek XS Plus System
## B Predicate 510(k) Number(s):
K071041
## C Comparison with Predicate(s):
| Device & Predicate Device(s): | K220282 | K071041 |
| --- | --- | --- |
| Device Trade Name | i-STAT PTplus cartridge with the i-STAT 1 System | CoaguChek® XS Plus System |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The i-STAT PTplus cartridge is intended for use in the in vitro quantitative measurement of the clot time of the extrinsic coagulation pathway when activated by thromboplastin in non-anticoagulated whole blood (venous or capillary), using the i-STAT PTplus | Intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. The system uses fresh capillary or non-anticoagulated venous whole blood. |
K220282 - Page 3 of 13
{4}
| | STAT 1 System. Measurements of prothrombin time are used to aid in the monitoring of patients receiving anticoagulant therapy with coumarin derivatives. The i-STAT PT^{plus} Prothrombin Time test result is reported in seconds and as an International Normalized Ratio (INR). The test is intended for point of care use and is for prescription use only. | |
| --- | --- | --- |
| Principle of Measurement | Electrochemical technology with amperometric (electric current) detection of thrombin activity | Same |
| Reportable Range | 0.8 – 8.0 INR | 0.8 – 8.0 INR |
| Sample Type | Fresh whole blood from venous or capillary samples | Same |
| Sample Preparation | Ready to use. No sample preparation required | Same |
| Traceability | Traceable to the WHO international reference method; rTF-16 | Same |
| Reagent | Human recombinant thromboplastin | Same |
| Electronic Quality Controls | On-board quality control built into the analyzer and external Electronic Simulator | On-board quality control which uses electrochemical signal to detect test strip integrity |
| Liquid Quality Controls | Two levels of liquid controls | Same |
| Analyzer Type | Handheld | Same |
| Output | PT in seconds and INR | Same |
K220282 - Page 4 of 13
{5}
K220282 - Page 5 of 13
| General Device Characteristic Differences | | |
| --- | --- | --- |
| Measurement software | Results calculated from analyzer-measured parameters | Converts raw signal into PT result |
| Sample Volume | Approximately 20 μL | Minimum of 10 μL |
| Quality Control Lockout | Not available for external liquid controls | If QC result fails, the operator is locked from performing testing |
| Control Target Values | Lot-specific target values provided in QC package insert | Lot-specific target values provided on QC code chip |
VI Standards/Guidance Documents Referenced:
- CLSI EP05: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
- CLSI EP28: Defining, Establishing, and Verifying, Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition
- CLSI EP07: Interference Testing in Clinical Chemistry; Second Edition
- CLSI POCT14-ED2 Point-of-Care Coagulation Testing and Anticoagulation Monitoring; Second Edition
- CLSI H54; Procedures for Validation of INR and Local Calibration of PT/INR Systems; Approved Guideline
- CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
VII Performance Characteristics:
A Analytical Performance:
1. Precision/Reproducibility:
a. 20-day Liquid Control Precision Study
A single-site precision study was performed over the course of 20 non-consecutive days, two runs per day, by two POC healthcare professionals using three lots of i-STAT PTplus cartridges and i-STAT PTplus Controls (L1 and L2) as well as an internal control fluid (L3) on 18 i-STAT 1 analyzers. The 20-day precision of the prothrombin time test in the i-STAT PTplus cartridge with the i-STAT 1 System has been demonstrated to be within the acceptance criteria at each control level and for each cartridge lot. The results of this study demonstrated acceptable within-laboratory precision for all three control levels.
| 20 Day Precision Result Summary (seconds) | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Fluid Level | N | Mean | Repeatability | | Between-lot | | Between-run | | Between-day | | Within-lab | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| L1 | 359 | 10.44 | 0.274 | 2.6 | 0.113 | 1.1 | 0.084 | 0.8 | 0.133 | 1.3 | 0.335 | 3.1 |
| L2 | 355 | 23.64 | 0.764 | 3.2 | 0.841 | 3.6 | 0.183 | 0.8 | 0.226 | 1.0 | 1.173 | 4.5 |
| L3 | 339 | 51.77 | 2.329 | 4.5 | 1.530 | 3.0 | 0.644 | 1.2 | 0.96 | 0.8 | 2.887 | 5.9 |
{6}
| 20 Day Precision Result Summary (INR) | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Fluid Level | N | Mean | Repeatability | | Between-lot | | Between-run | | Between-day | | Within-lab | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| L1 | 359 | 1.03 | 0.027 | 2.6 | 0.011 | 1.1 | 0.008 | 0.8 | 0.013 | 1.3 | 0.033 | 3.1 |
| L2 | 355 | 2.34 | 0.076 | 3.2 | 0.083 | 3.6 | 0.018 | 0.8 | 0.022 | 1.0 | 0.106 | 4.5 |
| L3 | 339 | 5.13 | 0.231 | 4.5 | 0.151 | 3.0 | 0.064 | 1.2 | 0.038 | 0.8 | 0.286 | 5.9 |
## b. Lot-to-Lot Precision Study
A lot-to-lot precision study was performed over five days by three POC healthcare professionals using three lots of i-STAT PT$^{plus}$ cartridges and one lot of i-STAT PT$^{plus}$ Controls (L1 and L2) as well as an internal control fluid (L3). A total of 32 i-STAT 1 analyzers were used. The lot-to-lot precision of the prothrombin time test in the i-STAT PT$^{plus}$ cartridge on the i-STAT 1 System demonstrated acceptable precision, as the results are shown to be within the acceptance criteria for all levels.
| Lot-to-Lot Precision Summary for Prothrombin Time (seconds) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Control Fluid | N | Mean (sec) | Within-run | | Between-day | | Between-lot | | Within-lab | |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | %CV |
| L1 | 88 | 9.15 | 0.212 | 2.3 | 0.414 | 4.5 | 0.150 | 1.6 | 0.489 | 5.3 |
| L2 | 86 | 23.72 | 0.743 | 3.1 | 1.010 | 4.3 | 1.022 | 4.3 | 1.617 | 6.8 |
| L3 | 88 | 58.27 | 1.351 | 2.3 | 2.269 | 3.9 | 2.846 | 4.9 | 3.882 | 6.7 |
| Lot-to-Lot Precision Summary for Prothrombin Time (INR) | | | | | | | | | | |
| Control Fluid | N | Mean (INR) | Within-run | | Between-day | | Between-lot | | Within-lab | |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | %CV |
| L1 | 88 | 0.91 | 0.021 | 2.3 | 0.041 | 4.5 | 0.015 | 1.6 | 0.048 | 5.3 |
| L2 | 86 | 2.35 | 0.074 | 3.1 | 0.100 | 4.3 | 0.101 | 4.3 | 0.160 | 6.8 |
| L3 | 88 | 5.77 | 0.134 | 2.3 | 0.225 | 3.9 | 0.282 | 4.9 | 0.384 | 6.7 |
## c. Instrument-to-Instrument Precision Study
An instrument-to-instrument precision study was performed over five days by two healthcare professionals using one lot of i-STAT PT$^{plus}$ cartridges and one lot of i-STAT PT$^{plus}$ Controls (L1 and L2) as well as an internal control fluid (L3). Each fluid level was tested on three i-STAT 1 analyzers with six replicates per analyzer, once per day for five days. The instrument-to-instrument precision of the prothrombin time test in the i-STAT PT$^{plus}$ cartridge on the i-STAT 1 System demonstrated acceptable precision, as the results are shown to be within the acceptance criteria for all levels.
| Instrument-to-Instrument Precision Summary for Prothrombin Time (seconds) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Control Fluid | N | Mean (sec) | Within-run | | Between-day | | Between-instrument | | Within-lab | |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | %CV |
| L1 | 90 | 10.26 | 0.209 | 2.0 | 0.034 | 0.3 | 0.031 | 0.3 | 0.031 | 0.3 |
| L2 | 90 | 23.58 | 0.536 | 2.3 | 0.086 | 0.4 | 0.105 | 0.4 | 0.105 | 0.4 |
| L3 | 90 | 53.34 | 0.811 | 1.5 | 0.129 | 0.2 | 0.381 | 0.7 | 0.38 | 0.7 |
K220282 - Page 6 of 13
{7}
| Instrument-to-Instrument Precision Summary for Prothrombin Time (INR) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Control Fluid | N | Mean (INR) | Within-run | | Between-day | | Between-instrument | | Within-lab | |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | %CV |
| L1 | 90 | 1.02 | 0.021 | 2.0 | 0.003 | 0.3 | 0.003 | 0.3 | 0.003 | 0.3 |
| L2 | 90 | 2.33 | 0.053 | 2.3 | 0.008 | 0.4 | 0.010 | 0.4 | 0.010 | 0.4 |
| L3 | 90 | 5.28 | 0.080 | 1.5 | 0.013 | 0.2 | 0.038 | 0.7 | 0.038 | 0.7 |
## d. Operator-to-Operator Precision Study
An operator-to-operator precision was performed over five days by three POC healthcare professionals using one lot of i-STAT PTplus cartridges and one lot of i-STAT PTplus Controls (L1 and L2) as well as an internal control fluid (L3). Each control fluid level was tested by three operators with six replicates per operator, once per day for five days. The operator-to-operator precision of the prothrombin time test in the i-STAT PTplus cartridge on the i-STAT 1 System demonstrated acceptable precision, as the results are shown to be within the acceptance criteria for all levels.
| Operator-to-Operator Precision Summary for Prothrombin Time (seconds) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Control Fluid | N | Mean (sec) | Within-Run | | Between-Day | | Between-Operator | | Within-Lab | |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | %CV |
| L1 | 90 | 10.49 | 0.267 | 2.5 | 0.052 | 0.5 | 0.095 | 0.9 | 0.288 | 2.7 |
| L2 | 90 | 24.31 | 0.456 | 1.9 | 0.124 | 0.5 | 0.016 | 0.1 | 0.473 | 1.9 |
| L3 | 89 | 53.04 | 0.811 | 1.5 | 0.360 | 0.7 | 0.369 | 0.7 | 0.961 | 1.8 |
| Operator-to-Operator Precision Summary for Prothrombin Time (INR) | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Control Fluid | N | Mean (INR) | Within-Run | | Between-Day | | Between-Operator | | Within-Lab | |
| | | | SD | % CV | SD | % CV | SD | % CV | SD | %CV |
| L1 | 90 | 1.04 | 0.026 | 2.5 | 0.005 | 0.5 | 0.009 | 0.9 | 0.029 | 2.7 |
| L2 | 90 | 2.41 | 0.045 | 1.9 | 0.012 | 0.5 | 0.002 | 0.1 | 0.047 | 1.9 |
| L3 | 89 | 5.25 | 0.080 | 1.5 | 0.036 | 0.7 | 0.037 | 0.7 | 0.095 | 1.8 |
## e. Multi-Day Precision Study
A multi-day precision study was performed at three POC sites by one POC healthcare professional per site. The study was performed using one lot of i-STAT PTplus Controls (L1 and L2) as well as an internal control fluid (L3) tested once per day for five days using six i-STAT PTplus cartridges on six i-STAT 1 Wireless Analyzers. The results of the multi-day precision study for the i-STAT PTplus cartridge with the i-STAT 1 System met the acceptance criterion for total precision by site and by level, with a range of 2.5–3.7 %CV for PT(seconds) and a range of 3.2–5.0 %CV for INR.
| Multi-Day Precision Summary for Prothrombin Time (seconds) | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Fluid Level | N | Mean | Within-Day | | Between-Day | | Within-Site | | Between-Site | | Reproducibility |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD |
| L1 | 90 | 1.03 | 0.039 | 3.8 | 0.024 | 2.3 | 0.045 | 4.4 | 0.005 | 0.4 | 0.045 |
| L2 | 91 | 2.46 | 0.089 | 3.6 | 0.000 | 0.0 | 0.089 | 3.6 | 0.000 | 0.0 | 0.089 |
| L3 | 90 | 5.36 | 0.170 | 3.2 | 0.071 | 1.3 | 0.184 | 3.4 | 0.000 | 0.0 | 0.184 |
K220282 - Page 7 of 13
{8}
| Multi-Day Precision Summary for Prothrombin Time (INR) | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Fluid Level | N | Mean | Within-Day | | Between-Day | | Within-Site | | Between-Site | | Reproducibility |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD |
| L1 | 90 | 10.49 | 0.210 | 2.0 | 0.164 | 1.6 | 0.266 | 2.2 | 0.023 | 0.2 | 0.267 |
| L2 | 91 | 24.82 | 0.771 | 3.1 | 0.221 | 0.9 | 0.802 | 3.2 | 0.000 | 0.0 | 0.802 |
| L3 | 90 | 54.12 | 1.708 | 3.2 | 0.712 | 1.3 | 1.851 | 3.4 | 0.000 | 0.0 | 1.851 |
## f. Whole Blood Precision Study
Whole blood precision (repeatability) was evaluated using venous and capillary specimens at three ranges: non-therapeutic (INR 0.8 – 1.9), therapeutic (INR 2.0 – 4.5), and very high therapeutic (INR 4.6 – 8.0). The repeatability analysis was conducted using the data collected across three point of care sites from duplicate testing in both the capillary and the venous method comparison studies. The mean, pooled standard deviation, %CV and their respective 95% confidence intervals (CI) for PT in seconds and INR were calculated for each subject level. The table below summarizes this data for all sites combined. All results met the acceptance criteria.
| Capillary Whole Blood Precision Summary Prothrombin Time (seconds) | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample Range | N | Mean | | SD | | %CV | |
| | | Estimate | 95% CI | Estimate | 95% CI | Estimate | 95% CI |
| Non-therapeutic | 58* | 14.89 | (14.06, 15.73) | 1.414 | (1.197, 1.727) | 9.5 | (8.0, 11.6) |
| Therapeutic | 119* | 28.51 | (27.73, 29.30) | 1.495 | (1.327, 1.713) | 5.2 | (4.7, 6.0) |
| Very High Therapeutic | 9 | 50.71 | (42.88, 58.54) | 2.109 | (1.451, 3.851) | 4.2 | (2.9, 7.6) |
| Capillary Whole Blood Precision Summary Prothrombin Time (INR) | | | | | | | |
| Non-therapeutic | 58* | 1.48 | (1.39, 1.56) | 0.143 | (0.121, 0.174) | 9.7 | (8.2, 11.8) |
| Therapeutic | 119* | 2.82 | (2.74, 2.90) | 0.148 | (0.131, 0.169) | 5.2 | (4.6, 6.0) |
| Very High Therapeutic | 9 | 5.02 | (4.24, 5.80) | 0.201 | (0.139, 0.368) | 4.0 | (2.8, 7.3) |
| Venous Whole Blood Precision Summary Prothrombin Time (seconds) | | | | | | | |
| Non-therapeutic | 65* | 14.69 | (13.95, 15.43) | 1.047 | (0.894, 1.263) | 7.1 | (6.1, 8.6) |
| Therapeutic | 131 | 28.78 | (27.97, 29.59) | 0.660 | (0.589, 0.751) | 2.3 | (2.0, 2.6) |
| Very High Therapeutic | 13* | 54.37 | (48.44, 60.31) | 0.785 | (0.569, 1.264) | 1.4 | (1.0, 2.3) |
| Venous Whole Blood Precision Summary Prothrombin Time (INR) | | | | | | | |
| Non-therapeutic | 65* | 1.45 | (1.38, 1.53) | 0.109 | (0.093, 0.131) | 7.5 | (6.4, 9.0) |
| Therapeutic | 131 | 2.85 | (2.77, 2.93) | 0.069 | (0.061, 0.078) | 2.4 | (2.1, 2.7) |
| Very High Therapeutic | 13* | 5.38 | (3.80, 5.97) | 0.088 | (0.064, 0.141) | 1.6 | (1.2, 2.6) |
*Results with outliers included
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
a. Interference
K220282 - Page 8 of 13
{9}
The interference performance of the prothrombin time test in the i-STAT PTplus cartridge with the i-STAT 1 System was evaluated in the presence of potentially interfering endogenous or exogenous substances. The effect of each substance at each prothrombin time level (normal and therapeutic) was evaluated by comparing the performance of a test sample spiked to a high concentration of the substance and a control sample spiked with an equal volume of solvent. For an identified interferent, a dose-response was performed to determine the degree of interference as a function of the substance concentration. The common exogenous and endogenous interfering substances and their interference results are listed below and showed no significant interference up to the indicated concentration for all samples tested.
| Substance | Maximum Test Concentration with no interference |
| --- | --- |
| Acetaminophen | 1324 μmol/L |
| Acetylsalicylic Acid | 3.62 mmol/L |
| Ascorbic Acid | 342 μmol/L |
| Atorvastatin | 750 μg/L |
| Unconjugated Bilirubin | 684 μmol/L |
| Conjugated Bilirubin | 475 μmol/L |
| Enoxaparin | 2.0 IU/mL |
| Epsilon-aminocaproic acid | 0.39 mg/mL |
| Fondaparinux | 3.78 mg/L |
| Hemoglobin | 10 g/L |
| Ibuprofen | 2425 μmol/L |
| Prasugrel | 265.5 ng/mL |
| Tranexamic Acid | 45 μg/mL |
| Triglycerides | 37 mmol/L |
| Uric Acid | 1.4 mmol/L |
For the identified PT test interferents chlorhexidine digluconate, daptomycin, and oritavancin, a dose-response experiment was performed.
| Substance | Interference Result |
| --- | --- |
| Chlorhexidine digluconate | Prolonged results ≥ 9.58×10⁻⁴ % |
| Daptomycin | Prolonged results ≥ 0.22 mg/mL |
| Oritavancin | Prolonged results ≥ 104 mg/L |
b. Factor Sensitivity
The factor sensitivity for factors FII, FV, FVII and FX was determined using samples prepared by proportionately combining pooled normal plasma, red blood cells and factor-deficient plasma with various percent (%) factor activity ranging from 20%–100%. Each factor was tested at eight factor activity levels with three cartridge lots and three replicates per lot on nine i-STAT 1 analyzers. The factor sensitivity for the i-STAT PTplus test was estimated to be FII (39.5%); FV (42.0%); FVII (21.5%); FX (22.0%).
c. Fibrinogen Sensitivity
Fibrinogen sensitivity was evaluated using one cartridge lot across five fibrinogen levels, 63, 80, 336, 599, and 702 mg/dL. The study demonstrated that the prothrombin time test
K220282 - Page 9 of 13
{10}
in the i-STAT PTplus cartridge with the i-STAT 1 System is insensitive to fibrinogen across a concentration range of 63–702 mg/dL.
d. Platelet Sensitivity
Platelet sensitivity was evaluated at three platelet levels (low, normal, and high). The study demonstrated that the prothrombin time test in the i-STAT PTplus cartridge with the i-STAT 1 System is insensitive to platelets in the range of 70,000–670,000/mm³.
e. Hematocrit Sensitivity
The hematocrit sensitivity was evaluated by comparing the low hematocrit level (~23% packed-cell volume (PCV)) and the high hematocrit level (~55% PCV) to the nominal hematocrit level (~39% PCV) using three lots of i-STAT PTplus cartridges. The results of the internal study demonstrate that the prothrombin test time in the i-STAT PTplus cartridge on the i-STAT 1 System is insensitive to hematocrit between 24 to 54% PCV.
f. Heparin Sensitivity
Sensitivity to unfractionated heparin was evaluated at two heparin levels. The study demonstrated that the prothrombin time test in the i-STAT PTplus cartridge with the i-STAT 1 System is insensitive to unfractionated heparin concentrations up to 1.0 IU/mL.
g. Sensitivity for Lupus Anticoagulant
Sensitivity to lupus anticoagulant antibodies (LA) was evaluated using two levels (low and high) of positive LA and one level of negative LA (control). The study demonstrated that the prothrombin time test in the i-STAT PTplus cartridge with the i-STAT 1 System is sensitive to both low and high levels of lupus anticoagulant antibodies.
4. Assay Reportable Range:
The reportable range of the prothrombin time test in the i-STAT PTplus cartridge with the i-STAT 1 System was determined through a method comparison study by using venous and capillary whole blood specimens collected from subjects undergoing anticoagulant therapy with coumarin derivatives and from subjects who were not on anticoagulant therapy. The reportable range for prothrombin time test in the i-STAT PTplus cartridge with the i-STAT 1 System is 0.8–8.0 INR and 8.1–80.8 seconds.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability
The traceability of the prothrombin time test in the i-STAT PTplus has been established against the World Health Organization (WHO) International Reference Preparation (IRP) of human recombinant tissue factor (RTF) using the tilt tube technique. The i-STAT PTplus cartridge is traceable to rTF/16.
Calibration
The i-STAT PTplus thromboplastin is specific for the i-STAT instrument. Calibration and ISI assignment is performed using multiple i-STAT instruments running i-STAT cartridges containing the i-STAT PTplus thromboplastin resulting in a factory calibrated PT test.
K220282 - Page 10 of 13
{11}
K220282 - Page 11 of 13
# Controls
Liquid quality controls (i-STAT PT$^{plus}$ Control Levels 1 and 2) are available as an optional quality control methodology to meet facility regulatory compliance requirements. The controls are citrated pooled lyophilized plasma with a calcium chloride (CaCl$_2$) reconstitution solution. i-STAT PT$^{plus}$ Control Level 1 has been formulated to produce a normal/non-therapeutic prothrombin time and is designed to provide results within a range of 0.8–1.6 INR, while i-STAT PT$^{plus}$ Control Level 2 has been formulated to produce a therapeutic prothrombin time and is designed to provide results within a range of 1.7–3.5 INR. The controls are manufactured for Abbott Point of Care by Cliniqa Corporation.
# In-Use (Reconstituted) Stability Study for Controls
The instructions for use for the i-STAT PT$^{plus}$ Controls specify that the control fluid components be equilibrated at room temperature for a minimum of 45 minutes to a maximum of 4 hours prior to reconstitution. The user is instructed to test the fluids immediately after reconstitution. The in-use stability of the controls was assessed by testing of i-STAT PT$^{plus}$ cartridges with of i-STAT PT$^{plus}$ Controls stored at room temperature over the range of the equilibration period specified above. An in-use stability study was conducted using nine vials from one lot of i-STAT PT$^{plus}$ Control Levels 1 and 2 on i-STAT PT$^{plus}$ cartridge lot using 60 i-STAT 1 analyzers. At each time interval, three vials of each fluid level were tested with 20 i-STAT PT$^{plus}$/aPTT cartridges per vial on i-STAT 1 analyzers for a total sample size of 60 cartridges per test condition per fluid level. The results for the in-use stability study for the i-STAT PT$^{plus}$ Controls support that the controls are stable for four hours when stored at room temperature.
# Shelf-Life Stability Study for Controls
A real-time shelf-life stability study was conducted using three lots for each level of i-STAT PT$^{plus}$ Control Levels 1 and 2 with seven i-STAT PT$^{plus}$ cartridge lots using six i-STAT 1 analyzers. Testing was performed at the beginning of the study (T= 0, corresponding to date of control lot manufacture), at three intervals throughout the duration of the intended shelf life and two intervals beyond the intended shelf life for a total of six test time points. The results for the real time stability study for the i-STAT PT$^{plus}$ Controls support a shelf-life of 12 months when stored refrigerated at 2–8°C.
# i-STAT PT$^{plus}$ Cartridge Shelf-Life
Shelf-life stability study was conducted using three lots of i-STAT PT$^{plus}$ cartridges at 2–8°C and 18–30°C storage conditions. Three levels of the i-STAT PT$^{plus}$ control fluids, unaltered whole blood at normal prothrombin time and factor depleted samples to reflect therapeutic prothrombin times were tested at seven different test events: 0, 19, 60, 90, 120, 180, 210, and 240 days.
The stability results for the prothrombin time test in the i-STAT PT$^{plus}$ cartridge on the i-STAT 1 System support a shelf-life claim of 193 days (6 months) at 2–8°C and 14 days at 18–30°C.
6. Detection Limit:
Not applicable.
7. Assay Cut-Off:
Not applicable.
{12}
# B Comparison Studies:
# 1. Method Comparison with Predicate Device:
A method comparison study was conducted. Venous and capillary whole blood specimens were prospectively collected from subjects undergoing anticoagulant therapy with coumarin derivates and from subjects who were not on anticoagulant therapy at three clinical sites. Both venous and capillary specimens were tested in duplicate on the i-STAT 1 Analyzer and were tested in singlicate on the Roche CoaguChek XS Plus System (K071041). Passing-Bablok regression analysis was performed for the first replicate of the i-STAT prothrombin time result versus the singlicate result from the CoaguChek.
| Method Comparison Results for i-STAT PTplus Cartridge with i-STAT 1 System vs. CoaguChek XS System (Seconds) | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Matrix | N | i-STAT 1 PT (sec) Range | CoaguChek PT (sec) Range | Correlation coefficient (r) | | Slope | | Intercept | |
| | | | | Estimate | 95% CI | Estimate | 95% CI | Estimate | 95% CI |
| Venous | 191 | 8.5 – 67.9 | 10.8 – 84.6 | 0.98 | 0.97, 0.98 | 0.736 | 0.715, 0.757 | 1.352 | 0.708, 1.920 |
| Capillary | 153 | 8.6 – 61.9 | 11.1 – 83.7 | 0.97 | 0.96, 0.98 | 0.743 | 0.711, 0.776 | 1.051 | 0.309, 1.892 |
| Method Comparison Results for i-STAT PTplus Cartridge with i-STAT 1 System vs. CoaguChek XS System (INR) | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Matrix | N | i-STAT 1 INR Range | CoaguChek INR Range | Correlation coefficient (r) | | Slope | | Intercept | |
| | | | | Estimate | 95% CI | Estimate | 95% CI | Estimate | 95% CI |
| Venous | 191 | 0.8 – 6.7 | 0.9 – 7.0 | 0.98 | 0.97, 0.98 | 0.875 | 0.857, 0.906 | 0.113 | 0.045, 0.157 |
| Capillary | 153 | 0.8 – 6.1 | 0.8 – 7.0 | 0.97 | 0.96, 0.98 | 0.885 | 0.846, 0.923 | 0.104 | 0.009, 0.171 |
# 2. Method Comparison with Reference Device:
Venous and capillary whole blood specimens were prospectively collected from subjects undergoing anticoagulant therapy with coumarin derivates and from subjects who were not on anticoagulant therapy at three POC clinical sites. Both venous and capillary specimens were tested in duplicate on the i-STAT 1 Analyzer and citrated plasma from the venous whole blood specimens was tested in duplicate on the Sysmex CS-2500 (K172286) reference instrument using Dade Innovin reagent. Passing-Bablok regression analysis was performed for the first replicate of the i-STAT prothrombin time result versus the first replicate result from the Sysmex CS-2500.
| Method Comparison Results for i-STAT PTplus Cartridge with i-STAT 1 System vs. Sysmex CS-2500 (seconds) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Matrix | N | i-STAT 1 PT (sec) Range | Sysmex PT (sec) Range | Correlation coefficient (r) (95% CI) | Slope (95% CI) | Intercept |
| Venous | 211 | 8.5 – 80.5 | 9.7 – 83.1 | 0.92 (0.90, 0.94) | 1.037 (0.994, 1.082) | -0.591 (-1.500, 0.151) |
| Capillary | 203 | 8.6 – 80.5 | 9.7 – 83.1 | 0.91 (0.88, 0.93) | 1.023 (0.979, 1.070) | -0.189 (-1.052, 0.641) |
K220282 - Page 12 of 13
{13}
| Method Comparison Results for i-STAT PT^{plus} Cartridge with i-STAT 1 System vs. Sysmex CS-2500 (INR) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Matrix | N | i-STAT 1 INR Range | Sysmex INR Range | Correlation coefficient (r) (95% CI) | Slope (95% CI) | Intercept (95% CI) |
| Venous | 211 | 0.8 – 8.0 | 0.9 – 8.1 | 0.92 (0.90, 0.94) | 1.037 (1.000, 1.078) | 0.004 (-0.079, 0.075) |
| Capillary | 203 | 0.8 – 8.0 | 0.9 – 8.1 | 0.91 (0.89, 0.93) | 1.022 (0.984, 1.061) | 0.047 (-0.029, 0.127) |
## C Clinical Studies:
1. Clinical Sensitivity: Not applicable.
2. Clinical Specificity: Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable.
## D Clinical Cut-Off:
Not applicable.
## E Expected Values/Reference Range:
i-STAT PT<sup>plus</sup> cartridge reference intervals were established with 154 venous and 146 capillary specimens on the i-STAT 1 analyzer. Testing was performed with three cartridge lots on the i-STAT 1 System at three clinical sites. The reference range for venous and capillary specimens combined is INR: 0.9–1.3.
| Reference Range | Venous | | Capillary | |
| --- | --- | --- | --- | --- |
| | PT (sec) | INR | PT (sec) | INR |
| i-STAT PT^{plus} Cartridge with i-STAT 1 Wireless | 9.2–13.0 | 0.9–1.3 | 9.0–13.8 | 0.9–1.3 |
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K220282 - Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.