K220257 · 3M Deutschland GmbH · ELW · Feb 1, 2022 · Dental
Device Facts
Record ID
K220257
Device Name
Permadyne
Applicant
3M Deutschland GmbH
Product Code
ELW · Dental
Decision Date
Feb 1, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3660
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Impression of inlay, onlay, crown, bridge, and veneer preparations Functional impressions
Device Story
Permadyne is a polyether impression material used by dental professionals to reproduce the structure of a patient's teeth and gums. It consists of two components: a light-bodied wash material and a heavy-bodied tray material. The clinician manually mixes the base and catalyst pastes (7:1 volume ratio) and applies the material to a preformed impression tray. The material sets in the patient's mouth to create a physical mold of the dental anatomy. This mold is used by dental laboratories to fabricate restorations such as crowns, bridges, inlays, onlays, and veneers. The device benefits the patient by enabling the creation of accurate, custom-fitted dental prosthetics.
Clinical Evidence
Bench testing only. The device was evaluated against ISO 4823:2015 standards for elastomeric impression materials. Comparative testing against the predicate (K800031) included mixing time, consistency, working time, detail reproduction, linear dimensional change, compatibility with gypsum, elastic recovery, and strain in compression. Biocompatibility was assessed per ISO 10993 series and ISO 7405 by a board-certified toxicologist.
Technological Characteristics
Polyether-based elastomeric impression material. Components: base paste and catalyst paste. Manual mixing. Form factor: tubes. Standards: ISO 4823:2015 (performance), ISO 10993 series/ISO 7405 (biocompatibility).
Indications for Use
Indicated for dental patients requiring impressions for inlay, onlay, crown, bridge, and veneer preparations, or functional impressions for edentulous/partially dentate situations.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
February 1, 2022
3M Deutschland GmbH % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K220257
Trade/Device Name: Permadyne Regulation Number: 21 CFR 872.3660 Regulation Name: Impression material Regulatory Class: Class II Product Code: ELW Dated: January 28, 2022 Received: January 31, 2022
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K220257
Device Name
Permadyne
Indications for Use (Describe)
- · Impression of inlay, onlay, crown, bridge, and veneer preparations
- · Functional impressions
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Carl-Schurz-Straße 1 41453 Neuss Germany
Postal Address: ESPE Platz 82229 Seefeld Germany
☎+49 (0)8152 700-0 鳥 +49 (0)8152 700-1366 Internet: www.3MESPE.de E-Mail: info3MESPE@3M.com WEEE-Req.-Nr. DE 36963167 VAT-ID: DE 120679179
# 510(k) Summary
| Submitter: | 3M Deutschland GmbH<br>ESPE Platz<br>82229 Seefeld<br>Germany<br>Establishment Registration Number: 9611385 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ruediger Franke<br>Regulatory Affairs Specialist<br>Phone: +49-8152-700 1802<br>Fax: +49-8152-700 1869<br>e-mail: ruediger.franke@3M.com |
| Date: | November 25, 2021 |
| Trade Name: | Permadyne |
| Common Name: | Polyether impression material |
| Classification Name: | Impression material<br>(21 CFR 872.3660, product code ELW) |
| Device Class: | Class II |
| Predicate Device: | Permadyne Polyether Rubber Impression (K800031) |
#### Description of Device
The polyether impression material Permadyne consists of the two material components lightbodied consistency and heavy-bodied consistency. It's classified as impression materials (21 CFR 872.3660) because it's intended to reproduce the structure of a patient's teeth and qums. Permadyne is a hand mixed (base paste/catalyst paste) impression material used in one step technique. Thereby Permadyne – Light Bodied Consistency is used as wash material in combination with Permadyne – Heavy Bodied Consistency as tray material. The mixing ratio is 7 volumes base paste :1 volume catalyst paste.
Permadyne – Light Bodied Consistency is a light-bodied polyether impression material. The base paste is blueish colored and the catalyst paste reddish colored.
Permadyne - Heavy Bodied Consistency is a heavy-bodied polyether impression material. The base paste is salmon colored and the catalyst paste reddish colored.
Image /page/3/Picture/11 description: The image shows a logo for the German Dental Industry. The logo consists of a stylized letter 'G' formed by concentric circles, with a triangular shape at the bottom. To the right of the symbol, the text 'Deutsche Dental Industrie' is written in a simple, sans-serif font, indicating the organization or industry the logo represents.
{4}------------------------------------------------
### Applicable Standards for Product Tests
- . ISO 4823:2015 Dentistry — Dentistry — Elastomeric Impression Materials
### Indications for Use for Permadyne reformulated
- Impressions of inlay, onlay, crown, bridge, and veneer preparations
- . Functional impressions
### Comparison
The reformulated Permadyne was compared to Permadyne (K800031) regarding indications for use, intended use, composition technology and physical and mechanical properties.
The tables below summarize the indications and technology of the reformulated Permadyne and the predicate device:
| Permadyne (K800031) | Permadyne reformulated |
|-----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Polyether elastomeric impression material for<br>crowns, bridges and edentulous impression<br>procedures. | • Impression of inlay, onlay, crown, bridge, and<br>veneer preparations |
| | • Functional impressions |
Table Comparison of indications
| Technology | Permadyne (K800031) | Permadyne reformulated |
|------------------------------------------------------------|---------------------|------------------------|
| Heavy-bodied and light-bodied material<br>components | × | × |
| Material family – Polyether impression<br>material | × | × |
| Material offered in tubes with base and<br>catalyst pastes | × | × |
| Manual mixing | × | × |
Table Comparison to predicate technology
Permadyne (K800031) and Permadyne reformulated, are both intend to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. Therefore, the intended use of the product is not affected by the change. The indications for use of Permadyne reformulated and the predicate device Permadyne (K800031) are clinically very similar. Permadyne is a precision impression material and therefore suitable for all kinds of impressions. The requirements of an impression material for inlay, onlay, crown, bridge and veneer preparations are comparable as all examples describe the impression of prepared teeth. The impression of an inlay, onlay or veneer don't require additional or other performance characteristics of the impression material than the impression of a crown or bridge. The procedure "functional impression" describes the impression technique that is used for
{5}------------------------------------------------
page 3
edentulous and almost edentulous /partially dentate situations and is therefore don't describe a new condition compared to the edentulous impression procedures description of the predicate. Therefore, 3M Deutschland concludes that Permadyne reformulated is substantially equivalent to the predicate device Permadyne (K800031).
The reformulated Permadyne (K800031) are polyether impression materials containing reactive components, fillers, softeners, setting regulators, flavors, pigments and catalysts. 3M Deutschland GmbH is providing information to the Agency regarding information to FDA about the composition of the reformulated Permadyne and compared this to the predicate device. Considering the results of the biocompatibility evaluation of the reformulated Permadyne material components, it was shown that the new raw material does not have a negative influence on the biocompatibility of the materials. Also, the slightly different quantitative compositions of the reformulated material components in comparison to the predicate's composition submitted to the FDA in 1982 are covered by these favorable biocompatibility evaluations. In addition, the reformulated Permadyne material components fulfil the material performance requirements of ISO 4823:2015. No negative influence of the composition change on the material properties was observed. Therefore, 3M Deutschland GmbH concludes that Permadyne reformulated is substantially equivalent to the predicate device Permadyne (K800031) in technology.
In vitro testing was conducted to show that the reformulated Permadyne fulfils the requirements of FDA recognized standard ISO 4823. Additionally, the reformulated Permadyne was compared to the predicate device Permadyne (K800031) regarding colors, mixing-time, consistency, working time, detail reproduction, linear dimensional change, compatibility with gypsum (type 3 / type 4 or 5), elastic recovery and strain in compression. The results of the reformulated Permadyne fulfill the requirements of ISO 4823 and are similar to Permadyne (K800031). In summary, 3M Deutschland GmbH concludes that the reformulated Permadyne is substantially equivalent to the predicate device regarding performance and physical and mechanical properties.
{6}------------------------------------------------
### Biocompatibility
The biocompatibility assessment for the products was conducted in accordance with the following guidance:
| Guidance | Edition | Title |
|------------------------------------------------------------------------------|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| US FDA Docket Number<br>FDA-2013-D-0350. CDRH<br>Document Number 1811-<br>R1 | September 4, 2020 | Use of International Standard ISO 10993-1, Biological<br>evaluation of medical devices - Part 1: Evaluation and<br>testing within a risk management process - Guidance<br>for Industry and Food and Drug Administration Staff |
| ISO 10993-1 | 20018 | Evaluation and testing within a risk management<br>process |
| ISO 10993-3 | 2014 | Tests for genotoxicity, carcinogenicity and reproductive<br>toxicity |
| ISO 10993-5 | 2009 | Tests for in vitro cytotoxicity |
| ISO 10993-10 | 2010 | Tests for irritation and skin sensitization |
| ISO 10993-11 | 2017 | Tests for systemic toxicity |
| ISO 10993-12 | 2021 | Sample preparation and reference materials |
| ISO 10993-18 | 2020 | Chemical characterization of materials |
| ISO/TR 10993-22 | 2017 | Guidance on nanomaterials |
| ISO 10993-23 | 2021 | Tests for irritation |
| ISO 7405 | 2018 | Evaluation of biocompatibility of medical devices used<br>in dentistry |
The biocompatibility of Permadyne has been assessed by a board-certified toxicologist according to recommendations in FDA guidance and internationally recognized standards for medical and dental devices. The conclusion of the assessment is that Permadyne is safe for its intended use.
## Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe, as effective, and performs as well as the legally marketed predicate device."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.