K220227 · Heartway Medical Products Co., Ltd. · INI · Oct 12, 2022 · Physical Medicine
Device Facts
Record ID
K220227
Device Name
Auto Folding Scooter, S21F
Applicant
Heartway Medical Products Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
Oct 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Indications for Use
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Device Story
Auto Folding Scooter S21F is a battery-powered, motorized 3-wheeled vehicle designed for indoor mobility for adults with physical restrictions. The device features a seat, handlebar steering, and four solid wheels (two rear drive, two front castors). It is operated by the user via steering handles that control forward/backward movement and speed. The device is compact and foldable for transport. It does not include external wired or wireless communication interfaces. The healthcare provider or user evaluates the device's suitability based on the patient's physical and mental capacity. The device provides mobility assistance, potentially improving independence for users with restricted mobility. It is not intended for use as a seat in motor vehicles.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including EMC testing (ISO 7176-21, RESNA WC-2 Section 21), electrical safety (ANSI/AAMI ES60601-1), and performance testing (ISO 7176-1, -2, -3, -4, -5, -6, -7, -8, -9, -10, -11, -13, -14, -15, -16, -22; RESNA WC-1 and WC-2 series). Biocompatibility testing (ISO 10993-5, -10) was performed on patient-contacting materials.
Technological Characteristics
Motorized 3-wheeled scooter; aluminum alloy frame; solid PU tires. Powered by 25.2 Vdc Lithium-ion battery. Dynamic R-series DR50 controller. Max speed 3.75 mph; max load 115 kg. Biocompatible materials: TPU (handlebar), PU foam (seat cushion), Vinyl (seat/backrest). Standards: ISO 7176 series, RESNA WC-1/WC-2, ANSI/AAMI ES60601-1, ISO 10993. No wireless/wired communication interfaces.
Indications for Use
Indicated for adults with mobility restrictions who are restricted to a sitting position. Not intended for visually impaired individuals. Users must be mentally and physically capable of operating the scooter and possess functional hand/finger dexterity.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
Predicate Devices
Heartway Medical Products Co., Ltd. Power Mobility Scooter, Brio S19 (K150987)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 12, 2022
Heartway Medical Products Co., Ltd. % Ke-Min Jen Contact Person Roc Chinese-European Industrial Research Society No. 58. Fu Chiun Street Hsin Chu City, Taiwan 30067 Taiwan
Re: K220227
Trade/Device Name: Auto Folding Scooter, S21F Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: September 6, 2022 Received: September 14, 2022
Dear Ke-Min Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K220227
Device Name Auto Folding Scooter, S21F
Indications for Use (Describe)
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Heartway. The logo features a blue square at the top, with a stylized pink butterfly-like design inside. Below the square, the word "HEARTWAY" is written in a bold, sans-serif font.
HEARTWAY MEDICAL PRODUCTS CO., LTD. No. 18, Jingke Central 1st Rd., Nantun Dist., Taichung City 408, Taiwan TEL: +886-4-23583232 FAX: +886-4-23590786 E-mail: sales(@heartway.com.tw
### 510(k) number: K220227 510(k) Summary
| Submitter's Name: | <b>HEARTWAY Medical Products Co., Ltd.</b><br>No.18, Jingke Central 1st Rd., Nantun Dist., Taichung City,<br>Taiwan, R.O.C. 40852 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------|
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Date summary prepared: | October 12, 2022 |
|------------------------|-----------------------------------------------------------|
| Proprietary Name: | HEARTWAY Auto Folding Scooter, S21F |
| Common or Usual Name: | POWERED SCOOTER |
| Classification Name: | Vehicle, Motorized 3- Wheeled (Class II, 21 CFR 890.3800) |
| Product Code: | INI |
| Official Correspondent: | Dr. KE-MIN JEN (email: ceirs.jen@msa.hinet.net) |
|-------------------------|-------------------------------------------------|
| TEL: | 886-3-5208829 |
| Predicate Device: | Heartway Medical Products Co., Ltd. |
| | Power Mobility Scooter, Brio S19 |
| | K150987 |
#### ● Indications for Use:
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
#### . Description of the device:
The Auto Folding Scooter S21F is designed for indoor use for adults with mobility restrictions. It is classified as Class II medical device per US FDA 21 CFR 890.3800. It is compact, maneuverable and not necessarily able to overcome obstacles. The maximum load is 115 kg.
The product is not intended for visually impaired people. The drivers need to be mentally and physically suitable to drive the scooters. The fingers need to work functionally. The driving distance will be reduced if the scooter is used frequently on slopes, rough ground or climbing curb. The scooter is not for use as a seat in motor vehicle.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for HEARTWAY. The logo features a blue square in the background with a pink butterfly-like shape in the foreground. Below the graphic is the word "HEARTWAY" in a bold, sans-serif font.
HEARTWAY MEDICAL PRODUCTS CO., LTD. No. 18, Jingke Central 1st Rd., Nantun Dist., Taichung City 408, Taiwan TEL: +886-4-23583232 FAX: +886-4-23590786 E-mail: sales(@heartway.com.tw http://www.heartway.com.tw
Auto Folding Scooter S21F is battery powered and configured with four solid wheels (03) (08), a seat (04), a handle bar to control the driving function (01), a cross bar (05), a main frame (06), a foot-rest (02), a back-rest (04), a set of 2 anti-tippers (09), referred to the components and parts diagrams of S21F below.
S21F is driven by two solid rear wheels (08) as the drive wheel, two solid front castors (03) as the steering wheel, using the steering handles (01) to control the front frame assembly (02) to control the front wheels (03) as the power scooter steering direction mechanism. Steering handles (01) is able to control driving forward, driving backward, speed control. The main frame (06) is equipped with a rear bumper to allow the scooter to sustain an impact without damage to the power scooter safety system. The maximum loading weight of S21F is 253 lbs. (115 kg), and its maximum speed is 3.75 mph (6 km/h). The dimensions of (Length * Width *Height) of the unfolded device are 33.0" x 18.1" x 30.3" (840 mm x 460 mm x 770 mm) and 28.7" x 17.3" x 16.5" (730 mm * 420 mm * 420 mm) for the folded device. The weight of the device is 51.8 lbs. (23.5 kg) with battery and 46.9 lbs. (21.3 kg) without battery. Seat belt is a standard accessory to S21F power scooter, and is not a 510(k)-clearance accessory. It can be installed onto the seat (as seen in the diagram below) by consulting with the local authorized dealer for seat belt installation.
Image /page/4/Picture/4 description: The image shows a line drawing of what appears to be a car frame. The frame is mostly rectangular with curved edges. There are also some wires and tubes attached to the frame, possibly indicating the car's electrical or plumbing system. The drawing is simple and lacks detail, focusing on the basic structure of the car frame.
Seat belt of S21F
The static stability results of tilt over tests performed laterally, posteriorly and anteriorly with a user of maximum allowable weight 253 lbs. and a fixed seat height 14.8" (376 mm) are 3 degrees for three cases.
There are no any external wired and /or wireless communication interfaces which may impact the cybersecurity information of our subject device. The following surfaces are recommended NOT to operate on:
- Sand surface A
- A Wet or icy surface
- A Road maintenance hole metal cover
- A Avoid going up multiple steps.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Heartway. The logo features a blue square with a pink butterfly-like design inside. Below the square, the word "HEARTWAY" is written in a bold, sans-serif font.
HEARTWAY MEDICAL PRODUCTS CO., LTD. No. 18, Jingke Central 1st Rd., Nantun Dist., Taichung City 408, Taiwan TEL: +886-4-23583232 FAX: +886-4-23590786 E-mail: sales(@heartway.com.tw http://www.heartway.com.tw
- A Avoid using escalators. Use the elevator.
- A Too steep incline over 3 degrees.
- A Ground clearance to battery: 1.18" (30 mm)
- A Curb climbing ability: 0.59" (15 mm)
EMC & Electrical Safety Testing and Performance Bench Testing are conducted per the following standards:
- EMC standards (applied for S21F & Charger)
- ISO 7176-21:2009 Wheelchairs Part 21: Requirements and test method . for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers
- RESNA WC-2:2019 Section 21: 2009 Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and motorized scooters
- Electrical Safety standards (applied for battery chargers: METCO, NL07C-25HT)
- . ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, Medical electrical equipment-Part 1: General requirements for basic safety and essential performance
- Scooter safety performance standards applied for S21F
- . ISO 7176-1, -3, -5, -7, -8, 15, -22 series & RESNA WC-1 series
- . ISO 7176-2, -4, -6, -9, -10 series & RESNA WC-2 series
- ISO 7176-11, ISO 7176-13
- I ISO 7176-14:2008 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods
- . RESNA WC-2 Section 14:2009 - Power and control systems for electrically powered wheelchair and scooters - Requirements and test methods
- . ISO 7176-16:2012 Resistance to Ignition of Postural Support Devices Test
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Heartway. The logo features a stylized pink butterfly shape against a blue background. Below the graphic is the word "HEARTWAY" in a bold, black sans-serif font.
## HEARTWAY MEDICAL PRODUCTS CO., LTD.
No. 18, Jingke Central 1st Rd., Nantun Dist., Taichung City 408, Taiwan TEL: +886-4-23583232 FAX: +886-4-23590786 E-mail: sales@heartway.com.tw
#### ● Biocompatibility information
| Patient-contacting parts | Materials | Testing standards | Testing laboratory | Verdict | Attached files |
|---------------------------|--------------------------------------------|-------------------------------------------------------|-------------------------------------------------------------------|---------|----------------|
| Dial set &<br>Handle bar | DIOSHY® TPU<br>(Thermoplastic<br>Urethane) | *ISO 10993-5:2009<br>in vitro Cytotoxicity test | *Super Laboratory Co., Ltd. | Pass | VOL_015_003 |
| | | *ISO 10993-10:2010 Maximization<br>Sensitization test | *Super Laboratory Co., Ltd. | Pass | VOL_015_004 |
| | | *ISO 10993-10:2010 Skin Irritation<br>test | *Super Laboratory Co., Ltd. | Pass | VOL_015_005 |
| Seat cushion | PU Foam | *ISO 10993-5:2009<br>in vitro Cytotoxicity test | *SGS Taiwan Ltd.<br>Ultra Trace &<br>Industrial Safety<br>Hygiene | Pass | VOL_015_006 |
| | | *ISO 10993-10:2010<br>Maximization Sensitization test | *SGS Mechanical<br>& Hardgoods Lab | Pass | VOL_015_007 |
| | | *ISO 10993-10:2010<br>Skin Irritation test | *SGS Mechanical<br>& Hardgoods Lab | Pass | VOL_015_008 |
| Seat leather<br>Seat belt | Vinyl Fabric | *ISO 10993-5:2009<br>in vitro Cytotoxicity test | *SGS Mechanical<br>& Hardgoods Lab | Pass | VOL_015_009 |
| | | *ISO 10993-10:2010<br>Maximization Sensitization test | *SGS Mechanical<br>& Hardgoods Lab | Pass | VOL_015_010 |
| | | *ISO 10993-10:2010<br>Skin Irritation test | *SGS Mechanical<br>& Hardgoods Lab | Pass | VOL_015_011 |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for HEARTWAY. The logo features a blue square with a pink butterfly-like design inside. Below the square, the word "HEARTWAY" is written in a bold, sans-serif font.
## HEARTWAY MEDICAL PRODUCTS CO., LTD.
No. 18, Jingke Central 1st Rd., Nantun Dist., Taichung City 408, Taiwan TEL: +886-4-23583232 FAX: +886-4-23590786 E-mail: sales@heartway.com.tw
| Comparison Item | Predicate device | Subject (New) Device | Verdict |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------|
| Manufacturer | Heartway medical products Co.,<br>Ltd. | Heartway Medical Products Co.,<br>Ltd. | -- |
| Trade Name | Power mobility scooter | Auto Folding Scooter | -- |
| Model Name | Brio S19 | S21F | -- |
| 510(K) Number | K150987 | To Be Assigned | -- |
| Indications for Use | The device is intended for<br>medical purposes to provide<br>mobility to persons restricted to<br>a sitting position. | The device is intended for<br>medical purposes to provide<br>mobility to persons restricted to<br>a sitting position. | Same |
| Number of Wheels | 4 | 4 | Same |
| Anti-Tipper | One set of 2 anti-tippers | One set of 2 anti-tippers | Same |
| Frame Type | Fixed / Aluminum alloy | Fixed / Aluminum alloy | Same |
| Biocompatibility testing | Iso 10993-1:2009<br>Iso 10993-5:2009<br>Iso 10993-10:2010 | ISO 10993-1:2009<br>ISO 10993-5:2009<br>ISO 10993-10:2010 | Same |
| Rear Wheels | Size 8" x 2"<br>Type Pu Solid Tire x 2 | Size 8" x 2"<br>PU solid tire x 2 | Same |
| Front Castor | Size 7" x 1.6"<br>Type Pu Solid Tire x 2 | Size 7" x 1.6"<br>PU solid tire x 2 | Same |
| Armrest type | NA | NA | Same |
| Wheel Lock type | Push-to- lock | Push-to- Lock | Same |
| Upholstery<br>inflammability testing | Pass | Pass | Same |
| Overall Dimensions | Length 36.6" (930 mm)<br>Width 19.0" (485 mm)<br>Height 37.2" (945 mm) | 33.0" (840 mm)<br>18.1" (460 mm)<br>30.3" (770 mm) | Different |
| Folded Dimensions<br>(L x W x H) | 36.6" x 19" x 13.3"<br>(930 x 485 x 340) mm | 28.7" x 17.3" x 16.5"<br>(730 x 440 x 420) mm | Different |
| Weight Capacity | 220 lbs. (100 kg) | 253 lbs. (115 kg) | Different |
| Max Speed | 5.0 mph (8.0 km/h) | 3.75 mph (6 km/h ) | Different |
| Battery Rating | Two 12 Vdc, 12 Ah<br>Sealed lead acid Batteries | One 25.2 Vdc, 11.5 Ah<br>Lithium-ion Battery | Different |
| | Off-board external type<br>CTE 4C24050A<br>Input: 110 -240 Vac, 50 Hz, 2 A<br>Output: 24 Vdc, 5 A | Off -board external type<br>METCO, NL07C-25HT<br>Input: 110-240 Vac, 50/60 Hz, 84 W<br>Output: 29.05 Vdc, 2.5 A | Different |
| Charger Type | | | |
| Cruise Range | 11.25 miles (18 km) | 9.3 miles (15 km ) | Different |
| Patient- contacting parts and materials | Seat -PVC material<br>Hand grip - PVC material<br>Safety belt - PVC material | Dial set / Handle bar: TPU<br>Seat cushion: PU Foam<br>Seat & backrest leather: Vinyl Fabric | Different |
| Controller Type | Penny & Giles S-drive | Dynamic, R series, DR50 | Different |
| Motor Type | 3A, 24 Vdc / 270 W | 2.5 A max, 24 Vdc, 180 W | Different |
| Scooter Weight: | with Battery 69 lbs. (31.3 Kg)<br>without Battery 53 lbs. (24.0 Kg ) | 51.8 lbs. (23.5 kg)<br>46.9 lbs. (21.3 kg) | Different |
| Suspension | Cross brace | None | Different |
| Turning Radius | 32.2" (820 mm) | 43.7" / 1110 mm | Different |
| Static / Dynamic<br>Stabilities (Degrees) | 10 /6 | 6 / 3 | Different |
| Ground Clearance | 2.3" (60 mm) | 1.18" (30 mm) | Different |
| Kerb Climbing Ability | 1.7" (45 mm) | 0.59" (15 mm) | Different |
| Warranty | 3 years: main frame<br>1 year: controller /gear motor /batteries<br>Without exhaustive and wear parts | 2 years: Main frame<br>1.5 years: Controller<br>1 year: Charger<br>Without exhaustive and wear parts | Similar |
#### ● Comparison table
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Heartway. The logo features a blue square with a pink butterfly-like design inside. Below the square, the word "HEARTWAY" is written in bold, black letters.
# HEARTWAY MEDICAL PRODUCTS CO., LTD.
No. 18, Jingke Central 1st Rd., Nantun Dist., Taichung City 408, Taiwan TEL: +886-4-23583232 FAX: +886-4-23590786 E-mail: sales@heartway.com.tw
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo for HEARTWAY. The logo features a blue square with a pink butterfly-like design inside. Below the square, the word "HEARTWAY" is written in a bold, black font.
HEARTWAY MEDICAL PRODUCTS CO., LTD. No. 18, Jingke Central 1st Rd., Nantun Dist., Taichung City 408, Taiwan TEL: +886-4-23583232 FAX: +886-4-23590786 E-mail: sales@heartway.com.tw http://www.heartway.com.tw
#### ● COMPARISON DISCUSSION
Both devices are classified as Vehicle, Motorized 3-Wheeled per 21 CFR 890.3800 under product code INI. The indications for use for both devices are the same. Both devices, Brio S19 and S21F, use the same scooter technologies.
The overall dimensions of the subject device are smaller than those of the predicate device. Both devices, Brio S19 and S21F, are designed to function as small-scale motorized 3-wheeled vehicles, so their overall dimensions and weight capacities are small and similar, and their differences are not large. The subject device S21F meets the performance requirements of ISO 7176 series, ISO 7176-14-2008: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test method. and RESNA WC-2 Sec 14: Power and control systems for electrically powered wheelchairs and scooters- Requirements and test methods. There are no any new safety and effectiveness concerns raised by the S21F due to the minor differences of the overall dimensions and weight capacity.
The folded dimensions of both devices are different, and the length and width of S21F while folded are smaller than Brio S 19 and height of the S21F while folded is larger than the Brio S19. We fold the scooter when transported, and smaller length and width need less space of the motor vehicle to store the scooter, and this brings out more convenience for the patient. There are no any new safety and effectiveness concerns raised by S21F due to the smaller volume or space of S21F.
The maximum speed for the subject device is 3.75 mph and the predicate device has maximum speed of 5.0 mph. Regarding the subject device has a slower maximum speed, it has higher dynamic stability performance than the predicate device. So. there are no any new safety and effectiveness concerns raised by the subject device S21F.
The battery ratings for both devices are different. The predicate device used two sealed lead acid batteries rating at 12 Vdc, 12 Ah, and the subject device used one Lithium -ion battery rating at 25.2 Vdc, 11.5 Ah. The Lithium-ion battery for the subject device has passed the requirements of ISO 7176-14:2008 and RESNA WC-2 Sec 14, so there are no any new safety and effectiveness concerns raised by S21F.
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for HEARTWAY. The logo features a stylized pink butterfly shape above the word "HEARTWAY" in black, block letters. The butterfly is composed of four heart shapes, and the background behind the butterfly is a blue square.
HEARTWAY MEDICAL PRODUCTS CO., LTD. No. 18, Jingke Central 1st Rd., Nantun Dist., Taichung City 408, Taiwan TEL: +886-4-23583232 FAX: +886-4-23590786 E-mail: sales@heartway.com.tw
Because the subject device uses a new type battery: Lithium-ion battery charger is also adapted to use the METCO Lithium-ion battery charger NL07-25HT. The battery chargers for the predicate and subject devices are different, but the battery charger for the subject device passes the EMC standard: ISO 7176-21:2009, and Electrical Safety standard: ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012. There are no any new safety and effectiveness concerns raised by using different battery charger for the subject device S21F.
The cruise ranges for the predicate and the subject devices are 11.25 miles and 9.3 miles. The cruise range of S21F is less than the predicate device, and this difference relates with the frequency of battery charging and how far the patient will get to. The difference of the cruise ranges will not raise any new safety and effectiveness concerns for the subject device.
The patient-contacting parts and materials are different for both devices. But, the patient-contacting parts of the subject device passed biocompatibility testing per ISO 10993-5:2009 and ISO 10993-10:2010, the differences between them are not to raise any new safety and effectiveness concerns for the subject device.
The electronic controllers used by two devices are different. The Dynamic R-series DR50 controller used with the subject device S21F passed the ISO 7176-14:2008 and RESNA WC-2 Sec 14:2009. There are no any new safety and effectiveness concerns raised by the S21F due to using different electronic controllers.
The motor types for both devices are different. The predicate device uses motor of 270 W / 24 Vdc / 3A, and the subject device uses motor of 180 W /24 Vdc / 2.5 A. But the motor used in the subject device passed the requirements of the EMC standards: ISO 7176-21:2009. RESNA WC-2 Sec 21:2009. and performance standards: ISO 7176-14:2008, and RESNA WC-2 Sec 14:.2009. So, there are no any new safety and effectiveness concerns raised due to different motor used by the subject devices.
With the advancement of battery technology, S21F now uses one Lithium-ion battery, instead of using two Sealed Lead Acid batteries. The weight of one Lithium-ion battery is only 4.9 lbs. and the total weight of two Sealed Lead Acid batteries is 16 lbs. So, the scooter weight with battery for the S21F is 51.8 lbs. and the predicate device with battery weighs 69 lbs. There are no any new safety and effectiveness concerns raised by the different scooter weights.
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image shows a logo for "HEARTWAY". The logo features a stylized pink butterfly-like shape against a blue square background. Below the graphic is the word "HEARTWAY" in a bold, sans-serif font.
HEARTWAY MEDICAL PRODUCTS CO., LTD. No. 18, Jingke Central 1st Rd., Nantun Dist., Taichung City 408, Taiwan TEL: +886-4-23583232 FAX: +886-4-23590786 E-mail: sales(@heartway.com.tw
There is a suspension system for the predicate device and no suspension for the subject device. The suspension used by the predicate device is installed to decrease the impact from ground to the patient, since the maximum speed of the subject device is 3.75 mph, which is less than the predicate device of 5 mph; there is no need to install suspension system on the S21F to reduce the impact to the scooter. Without suspension system for the subject device will let the patient feel more impact from the ground, but this is not related with the safety and effectiveness.
The turning radius for the predicate device is 32.2" and it is 43.7" for the subject device. A smaller turning radius will let the scooter move in a smaller space. But this is not related to the safety and effectiveness.
The static/dynamic stabilities for the subject device are 6/3 degrees, which are lower than the predicate device 10/6 degrees. We place the safety notice in the user manual to inform the users of these limitations - not to use the device on the slope angle greater than 3 degrees or not to use it outdoors. We install a set of anti-tippers on the back of the S21F to prevent the tipping on the slope. By doing these actions, there are no any new safety and effectiveness concerns raised by S21F due to the smaller stabilities of the subject device.
The ground clearance for the predicate device is 2.3" and it is 1.18" for the subject device. So, there are more limitations for the subject device than the predicate device when moving across an obstruct. But the purpose of the limitations is to let the patient know the capability of the subject device. A smaller clearance will not lead to the dangerous situations, instead it will prevent the patient from the potential harms.
The kerb climbing ability for the predicate device is 1.7" and it is 0.59" for the subject device. A smaller kerb climbing ability will not lead to dangerous situations, but it will bring about more cautions and more safety.
The warranty for the subject device is similar to the predicate device. The different warranty contents will result to the different costs paid by the customers, which is not related to the safety and effectiveness concerns. There are no any new safety and effectiveness concerns raised by the subject device due to the different warranty contents.
{13}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Heartway. The logo features a blue square with a pink butterfly-like design inside. Below the square, the word "HEARTWAY" is written in a bold, sans-serif font.
HEARTWAY MEDICAL PRODUCTS CO., LTD. No. 18, Jingke Central 1st Rd., Nantun Dist., Taichung City 408, Taiwan TEL: +886-4-23583232 FAX: +886-4-23590786 E-mail: sales@heartway.com.tw
At last, the subject device passes the EMC testing per ANSI/RESNA WC-2:2009 Section 21 & ISO 7176-21:2009, and the performance testing per ISO 7176-14:2008 and RESNA WC-2 Section 14. The overall performances of the subject device are ensured.
#### ● CLINICAL INFORMATION
Not Applicable
#### ● CONCLUSIONS
The subject device, HEARTWAY Auto Folding Scooter, S21F, is as safe and effective as, and functions in a manner equivalent to the predicate device, HEARTWAY Power Mobility Scooter, Brio S19 (K150987). The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device identified in the submission. Thus, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.