← Product Code [ODC](/productcode/ODC) · K220210

# Disposable Endoscopy Adapter Set (K220210)

_Yangzhou Fartley Medical Instrument Technology Co., Ltd. · ODC · Jun 17, 2022 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K220210

## Device Facts

- **Applicant:** Yangzhou Fartley Medical Instrument Technology Co., Ltd.
- **Product Code:** [ODC](/productcode/ODC.md)
- **Decision Date:** Jun 17, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Disposable Endoscopy Adapter Set is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials. - Disposable Air/Water Valve: This unit is intended to be fitted to an endoscope air/water channel to control the inflow of medical gases and water, whilst preventing back-flow. - Disposable Suction Valve: This unit is intended to be fitted to an endoscope suction channel to control the operations of suction, whilst preventing inflow of air. - Disposable Biopsy Valve: This unit is intended to be fitted to an endoscope biopsy port to prevent leakage of gases and body fluids during an endoscopic procedure. - Disposable Water Jet Adapter: This unit is intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

## Device Story

Disposable Endoscopy Adapter Set comprises sterile units (Air/Water Valve, Suction Valve, Biopsy Valve, Water Jet Adapter) for attachment to endoscope working channels/ports. Device controls fluid/gas flow, prevents back-flow/leakage, and facilitates irrigation during GI endoscopy. Used in hospitals/clinics by trained endoscopists. Air/Water valve regulates gas/water inflow; Suction valve manages suction; Biopsy valve prevents leakage; Water Jet adapter connects to irrigation pumps. Output is mechanical control of endoscope channel function. Benefits include maintaining endoscope integrity, preventing fluid/gas leakage, and enabling irrigation, thereby supporting procedural safety and efficacy.

## Clinical Evidence

No clinical data included. Bench testing only: appearance and physical properties testing performed. Biocompatibility testing conducted per ISO 10993. Sterilization validated per ISO 11135-1 and ISO 10993-7.

## Technological Characteristics

Materials: Silicone rubber, ABS, SUS304, SUS303, PC. Components: Air/Water valve, Suction valve, Biopsy valve, Water Jet adapter. Energy: None (mechanical). Connectivity: None. Sterilization: Ethylene Oxide (EO). Form factor: Disposable, single-use, up to 4 valves per set. Compatible with Olympus, PENTAX, and Fujifilm endoscope series.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Disposable Endoscope Valves Set, Disposable Endoscope Valves Set B ([K200769](/device/K200769.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.

June 17, 2022

Yangzhou Fartley Medical Instrument Technology Co., Ltd. % Ethan Liu RA Specialist Shanghai Thinkwell Consulting Co., Ltd Room 211/6F, Xinling Road, Minhang District Shanghai, Shanghai 201100 China

Re: K220210

> Trade/Device Name: Disposable Endoscopy Adapter Set Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: ODC Dated: May 9, 2022 Received: May 16, 2022

Dear Ethan Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

#### Device Name

Disposable Endoscopy Adapter Set

The Disposable Endoscopy Adapter Set is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials.

- Disposable Air/Water Valve: This unit is intended to be fitted to an endoscope air/water channel to control the inflow of medical gases and water, whilst preventing back-flow.

- Disposable Suction Valve: This unit is intended to be fitted to an endoscope suction channel to control the operations of suction, whilst preventing inflow of air.

- Disposable Biopsy Valve: This unit is intended to be fitted to an endoscope biopsy port to prevent leakage of gases and body fluids during an endoscopic procedure.

- Disposable Water Jet Adapter: This unit is intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

Type of Use (*Select one or both, as applicable*)

| <div> <b> <span style="font-size:10pt">       ☑ Prescription Use (Part 21 CFR 801 Subpart D)      </span> </b> </div> | <div> <span style="font-size:10pt">      ☐ Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "ARTLEY" in a stylized, blue font. The letters are interconnected and have a three-dimensional appearance. A yellow line extends from the top right of the letter "Y", adding a contrasting color element to the design.

## 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

### 5.1 Submitter

| Submitted by:   | Yangzhou Fartley Medical Instrument Technology Co., Ltd.<br>Address:Beizhou Road, Lidian Town, Guangling District, Yangzhou<br>225106 Jiangsu, China                                                       |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ethan Liu<br>RA Specialist<br>Shanghai Thinkwell Consulting Co., Ltd<br>Address: Room 211/6F, Xinling Road, Minhang District, Shanghai,<br>China.<br>Phone: 0086-15216699240<br>Email: xtdeepwater@126.com |
| Date Prepared:  | Jan 27, 2022                                                                                                                                                                                               |

### 5.2 Device

| Device Name:         | Disposable Endoscopy Adapter Set |
|----------------------|----------------------------------|
| Classification Name: | Endoscope Channel Accessory      |
| Regulatory Class:    | II                               |
| Regulation Number:   | 21 CFR 876.1500                  |
| Regulation Name:     | Endoscope and Accessories        |
| Product Code:        | ODC                              |

## 5.3 Predicate Device

| Device Name:         | Disposable Endoscope Valves Set, Disposable Endoscope<br>Valves Set B, K200769 |
|----------------------|--------------------------------------------------------------------------------|
| Manufacturer:        | Wilson Instruments (SHA) Co., LTD                                              |
| Classification Name: | Endoscope Channel Accessory                                                    |
| Regulatory Class:    | II                                                                             |
| Regulation Number:   | 21 CFR 876.1500                                                                |
| Regulation Name:     | Endoscope and Accessories                                                      |
| Product Code:        | ODC                                                                            |

## 5.4 Device Description

The Disposable Endoscopy Adapter Set is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials. The sterile units may consist of Disposable Air/Water Valve, Disposable Suction Valve, Disposable Biopsy Valve and Disposable Water Jet Adapter. Disposable Endoscopy Adapter Set

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "ARTLEY" in a stylized, blue font. The letters are bold and interconnected, with a slight slant to the right. A yellow line extends from the top right of the "Y", adding a touch of color to the otherwise monochromatic design. The overall impression is of a modern and dynamic logo.

may be configured as single valve or multiple valves in any combination up to a maximum of 4 valves.

#### 5.5 Indication for Use:

The Disposable Endoscopy Adapter Set is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials.

- Disposable Air/Water Valve: This unit is intended to be fitted to an endoscope air/water channel to control the inflow of medical gases and water, whilst preventing back-flow.

- Disposable Suction Valve: This unit is intended to be fitted to an endoscope suction channel to control the operations of suction, whilst preventing inflow of air.

- Disposable Biopsy Valve: This unit is intended to be fitted to an endoscope biopsy port to prevent leakage of gases and body fluids during an endoscopic procedure.

- Disposable Water Jet Adapter: This unit is intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

#### 5.6 Comparison of Technological Characteristics

The Disposable Endoscopy Adapter Set has substantially equivalent device design, configuration, packaging fundamental technology, sterilization process and intended use as those featured in the predicate device Wilson Instruments (SHA) Co., LTD's Disposable Endoscope Valves Set, Disposable Endoscope Valves Set B, K200769. The differences between the proposed device and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below.

| Item                     | Disposable Endoscopy<br>Adapter Set(Proposed<br>Device)                                                                                                                                                                                | Disposable Endoscope Valves<br>Set, Disposable Endoscope<br>Valves Set B, K200769                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Discussion                                                                  |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Indication<br>for Use    | The Disposable Endoscopy<br>Adapter Set is a collection<br>of several sterile units. It is<br>intended to be fitted to<br>multiple endoscope<br>working channels/ports to<br>control the flow of fluids,<br>gases and other materials. | The Disposable Endoscope<br>Valves Set is a collection of<br>several sterile units. It is<br>intended to be fitted to<br>multiple endoscope working<br>channels/ports to control the<br>flow of fluids, gases and other<br>materials. It includes an<br>air/water valve, a suction<br>valve, a biopsy valve and a<br>water connector.                                                                                                                                                                                                                                                                                                                 | Substantially<br>equivalent                                                 |
|                          | - Disposable Air/Water<br>Valve: This unit is intended<br>to be fitted to an endoscope<br>air/water channel to control<br>the inflow of medical gases<br>and water, whilst<br>preventing back-flow.                                    | - Disposable Air/Water Valves:<br>This unit is intended to be<br>fitted to an endoscope<br>air/water channel to control the<br>inflow of medical gases and<br>water, whilst preventing                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                             |
|                          | - Disposable Suction Valve:                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                             |
| Item                     | Disposable Endoscopy<br>Adapter Set(Proposed<br>Device)                                                                                                                                                                                | Disposable Endoscope Valves<br>Set, Disposable Endoscope<br>Valves Set B, K200769                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Discussion                                                                  |
|                          | This unit is intended to be<br>fitted to an endoscope<br>suction channel to control<br>the operations of suction,<br>whilst preventing inflow of<br>air.                                                                               | back-flow.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                             |
|                          | - Disposable Biopsy Valve:<br>This unit is intended to be<br>fitted to an endoscope<br>biopsy port to prevent<br>leakage of gases and body<br>fluids during an endoscopic<br>procedure.                                                | - Disposable Suction Valves:<br>This unit is intended to be<br>fitted to an endoscope suction<br>channel to control the<br>operations of suction, whilst<br>preventing inflow of air.                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                             |
|                          | - Disposable Water Jet<br>Adapter: This unit is<br>intended to provide<br>irrigation via sterile water<br>supply during GI<br>endoscopic procedures<br>when used in conjunction<br>with an irrigation pump.                            | - Disposable Biopsy Valves:<br>This unit is intended to be<br>fitted to an endoscope biopsy<br>port to prevent leakage of<br>gases and body fluids during<br>an endoscopic procedure.                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                             |
|                          |                                                                                                                                                                                                                                        | - Disposable Endoscope Water<br>Connectors: This unit is<br>intended to provide irrigation<br>via sterile water supply during<br>GI endoscopic procedures<br>when used in conjunction with<br>an irrigation pump.                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                             |
|                          |                                                                                                                                                                                                                                        | The Disposable Endoscope<br>Valves Set B is a collection of<br>sterile device intended to be<br>fitted to multiple endoscope<br>working channels/ports to<br>enable an endoscope operator<br>control the function of the<br>working channels/ports and<br>prevent retrograde flow of<br>fluids, gases, and other<br>materials. It includes a suction<br>valve, an air/water valve. This<br>is a single-use device. -<br>Air/Water Valves: This device<br>is intended to be fitted to an<br>endoscope air/water channel to<br>enable the operator to control<br>inflow of medical gases and<br>water, whilst preventing<br>backflow. - Suction Valves: |                                                                             |
| Item                     | Disposable Endoscopy<br>Adapter Set(Proposed<br>Device)                                                                                                                                                                                | Disposable Endoscope Valves<br>Set, Disposable Endoscope<br>Valves Set B, K200769                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Discussion                                                                  |
|                          |                                                                                                                                                                                                                                        | The device is intended to be<br>fitted to an endoscope suction<br>channel to enable the operator<br>to control suction whilst<br>preventing inflow of air.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                             |
| Product<br>Code          | ODC                                                                                                                                                                                                                                    | ODC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                        |
| Regulation<br>Number     | 21 CFR 876.1500                                                                                                                                                                                                                        | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                        |
| Classification           | II                                                                                                                                                                                                                                     | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                        |
| Air/Water Valve          |                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                             |
| Compatible<br>Endoscopes | Olympus<br>140/160/180/190/240/260/2<br>90 series endoscopes;<br>PENTAX GI Video<br>Endoscope 90K/90i Series,<br>K10/i10 Series;<br>Fujifilm® 500/600/<br>700 series endoscopes                                                        | WS-VO-01:Olympus<br>140/160/180/190/240/260/2 90<br>series endoscopes<br>WS-VP-01: PENTAX GI<br>Video Endoscope 90K/90i<br>Series, K10/i10 Series                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Substantially<br>equivalent<br>supported by<br>bench testing.               |
| Sterile                  | EO Sterilization                                                                                                                                                                                                                       | EO Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                        |
| Material                 | Silicone Rubber, ABS,<br>SUS304                                                                                                                                                                                                        | WS-VO-01: SIR, ABS, TPE,<br>Stainless steel 304<br>WS-VP-01: SIR, ABS, TPE,<br>Stainless steel 304                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Substantially<br>equivalent<br>supported by<br>biocompatibility<br>testing. |
| Environment<br>Use       | Hospital/clinics                                                                                                                                                                                                                       | Hospital/clinics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                        |
| Suction Valve            |                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                             |
| Compatible<br>Endoscopes | Olympus<br>140/160/180/190/240/260/2<br>90 series endoscopes;<br>PENTAX GI Video<br>Endoscope 90K/90i Series;<br>K10/i10 Series;<br>Fujifilm® 500/600/<br>700 series endoscopes                                                        | WS-VO-02: Olympus<br>140/160/180/190/240/260/2 90<br>series endoscopes<br>WS-VP-02: PENTAX GI<br>Video Endoscope 90K/90i<br>Series; K10/i10 Series                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Substantially<br>equivalent<br>supported by<br>bench testing.               |
| Sterile                  | EO Sterilization                                                                                                                                                                                                                       | EO Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                        |
| Material                 | ABS, Silicone Rubber,<br>SUS304                                                                                                                                                                                                        | WS-VO-02: SIR, ABS, TPE,<br>Stainless steel 304                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Substantially<br>equivalent                                                 |
| Item                     | Disposable Endoscopy<br>Adapter Set(Proposed<br>Device)                                                                                                                                                                                | Disposable Endoscope Valves<br>Set, Disposable Endoscope<br>Valves Set B, K200769                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Discussion                                                                  |
|                          |                                                                                                                                                                                                                                        | WS-VP-02: SIR, ABS, TPE,<br>Stainless steel 304                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | biocompatibility<br>testing.                                                |
| Environment<br>Use       | Hospital/clinics                                                                                                                                                                                                                       | Hospital/clinics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                        |
| Biopsy Valve             |                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                             |
| Compatible<br>Endoscopes | Olympus<br>140/160/180/190/240/260/2<br>90 series endoscopes;<br>PENTAX GI Video<br>Endoscope 90K/90i Series;<br>K10/i10 Series;<br>Fujifilm® 500/600/<br>700 series endoscopes                                                        | Olympus<br>140/160/180/190/240/260/290<br>series endoscopes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Substantially<br>equivalent<br>supported by<br>bench testing.               |
| Sterile                  | EO Sterilization                                                                                                                                                                                                                       | EO Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                        |
| Material                 | Silicone Rubber                                                                                                                                                                                                                        | Silicone Rubber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                        |
| Environment<br>Use       | Hospital/clinics                                                                                                                                                                                                                       | Hospital/clinics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                        |
| Water Jet Adapter        |                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                             |
| Compatible<br>Endoscopes | Olympus<br>140/160/180/190/240/260/2<br>90 series endoscopes;<br>PENTAX GI Video<br>Endoscope 90K/90i Series;<br>K10/i10 Series;<br>Fujifilm® 500/600/<br>700 series endoscopes                                                        | Olympus<br>140/160/180/190/240/260/290<br>series endoscopes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Substantially<br>equivalent<br>supported by<br>bench testing.               |
| Sterile                  | EO Sterilization                                                                                                                                                                                                                       | EO Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                        |
| Material                 | Metal Type:PC,<br>SUS303,Silicone Rubber<br>Plastic Type: PC, Silicone<br>Rubber                                                                                                                                                       | NBR, SIR, ABS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Substantially<br>equivalent<br>supported by<br>biocompatibility<br>testing. |
| Environment<br>Use       | Hospitals/clinics                                                                                                                                                                                                                      | Hospitals/clinics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                        |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "FRITIEY" in a stylized, sans-serif font. The word is in blue, except for a yellow extension on the right side of the letter "Y". The letters are slanted to the right, giving the word a dynamic appearance.

K220210
Page 3 of 6

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "ARTLEY" in a stylized, blue font. The letters are connected and slanted to the right, giving the word a dynamic appearance. A thin, yellow line extends from the top of the "Y", adding a subtle accent to the design. The overall impression is modern and professional.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "ARTLEY" in a stylized, italicized font. The letters are primarily blue, with a yellow accent on the right side of the "Y". The font appears to be bold and slanted, giving the word a dynamic and modern look. The overall design is clean and simple, with a focus on the word itself.

## 5.7 Non-clinical Performance Data

The proposed device meets the requirements of ISO 10993 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing", ISO 11135-1 "Sterilization of Health Care products Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and ISO 10993-7 "Biological evaluation of medical

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "ARTLEY" in a stylized, blue font. The letters are bold and slightly slanted to the right, giving the word a dynamic appearance. A thin, yellow line extends from the top of the "Y", adding a subtle accent to the overall design.

devices - Part 7: ethylene oxide sterilization residuals" .

The following bench tests were performed on Disposable Endoscopy Adapter Set: Appearance, Physical properties. The results of all testing were passing.

## 5.8 Clinical Test Data

No Clinical Study is included in this submission.

# 5.9 Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Yangzhou Fartley Medical Instrument Technology Co., Ltd. has demonstrated that proposed device Disposable Endoscopy Adapter Set is substantially equivalent to Wilson Instruments (SHA) Co., LTD's currently marketed Disposable Endoscope Valves Set, Disposable Endoscope Valves Set B, K200769.

---

**Source:** [https://fda.innolitics.com/device/K220210](https://fda.innolitics.com/device/K220210)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
