K220202 · The O R Company Pty, Ltd. · LKF · Mar 31, 2022 · Obstetrics/Gynecology
Device Facts
Record ID
K220202
Device Name
Uterine ElevatOR PRO with OccludOR Balloon
Applicant
The O R Company Pty, Ltd.
Product Code
LKF · Obstetrics/Gynecology
Decision Date
Mar 31, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Uterine ElevatOR PRO with OccludOR Balloon™ is indicated for manipulation of the uterus, and injection of fluids or gases during laparoscopic procedures including total laparoscopic hysterectomy, laparoscopic assisted vaginal hysterectomy, laparoscopic tubal occlusion, and diagnostic laparoscopy. The Uterine ElevatOR PRO with OccludOR Balloon™ maintains pneumoperioneum during laparoscopic procedures by sealing the vagina once colpotomy is performed.
Device Story
Sterile, single-use uterine manipulator/injector for laparoscopic surgery. Device features stainless steel shaft, intrauterine balloon, cervical cup, and vaginal cup. Surgeon inserts device into uterus; inflates intrauterine balloon via syringe to secure position; adjusts vaginal cup along shaft to seal vagina. OccludOR balloon provides additional seal to maintain pneumoperitoneum during colpotomy. Port at handle allows injection of fluids/gases. Used in OR by gynecological surgeons. Output is mechanical manipulation of uterus and maintenance of surgical field pressure. Benefits include improved uterine positioning and reduced gas leakage during procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Biocompatibility testing performed per ISO 10993-1, -5, -10, -11 and USP <151>. Mechanical testing included balloon integrity, burst, fatigue, and adhesion tests; vaginal/cervical cup leakage and strength tests; and handle/joint strength tests. All tests met pre-specified acceptance criteria.
Indicated for patients undergoing laparoscopic procedures including total laparoscopic hysterectomy, laparoscopic assisted vaginal hysterectomy, laparoscopic tubal occlusion, and diagnostic laparoscopy requiring uterine manipulation, fluid/gas injection, and maintenance of pneumoperitoneum.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 31, 2022
The O R Company Pty Ltd Nicole Conway Quality Assurance & Regulatory Affairs Manager 1/32 Silkwood Rise Carrum Downs, Victoria 3201 Australia
Re: K220202
> Trade/Device Name: Uterine ElevatOR PRO with OccludOR Balloon™ Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: LKF Dated: January 28, 2022 Received: February 3, 2022
Dear Nicole Conway:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220202
Device Name
Uterine ElevatOR PRO with OccludOR Balloon TM
Indications for Use (Describe)
The Uterine ElevatOR PRO with OccludOR Balloon™ is indicated for manipulation of the uterus, and injection of fluids or gases during laparoscopic procedures including total laparoscopic hysterectomy, laparoscopic assisted vaginal hysterectorny, laparoscopic tubal occlusion, and diagnostic laparoscopy. The Uterine ElevatOR PRO with OccludOR Balloon™ maintains pneumoperioneum during laparoscopic procedures by sealing the vagina once colpotomy is performed.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for "THE O.R. COMPANY". The words "THE" and "COMPANY" are in a blue sans-serif font, with "O.R." in a blue circle. Below the company name is the tagline "The smart way to operate".
Image /page/3/Picture/2 description: The image shows the logo for Symmetry Surgical Company. The logo features a stylized red and gray shape to the left of the text. The text reads "symmetry surgical" in a gray sans-serif font, with the word "company" in a smaller font size to the right.
Image /page/3/Picture/3 description: The image shows the word "Bovie" in a sans-serif font. The word is in a dark blue color. Above the word is a light blue curved line with a dot at the end. The registered trademark symbol is to the right of the word.
# Uterine ElevatOR PRO with OccludOR Balloon™ - 510(k) Summary
#### l. SUBMITTER
The O R Company Pty Ltd 1/32 Silkwood Rise, Carrum Downs, VIC 3201, Australia Phone: +61 3 9413 5555 Fax: + 61 3 9413 5556
Contact Person: Nicole Conway Email: nicole@theorcompany.com Date Prepared: January 19, 2022
#### II. DEVICE
Trade Name: Uterine ElevatOR PRO with OccludOR Balloon™ Model numbers: UE-OBPRO-32, UE-OBPRO-35, UE-OBPRO-37 and UE-OBPRO-40 Common Name: Uterine Manipulator / Injector Regulation Number: 21 CFR 884.4530 Classification Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: LKF
#### III. PREDICATE DEVICE
Conmed Corporation, VCARE (Vaginal-Cervical Ahluwalia's Retractor-Elevator), K142716
This predicate has not been subject to a design-related recall.
## Reference devices:
Coopersurgical, Inc., Koh Colpotomizer System, K954311
#### IV. DEVICE DESCRIPTION
The Uterine ElevatOR PRO with OccludOR Balloon™ is a sterile, single-use medical device used to manipulate the uterus and cervix in surgical and diagnostic procedures. The device includes an insulated and anatomically designed stainless steel shaft with an intrauterine balloon and reference graduations (centimetres) at the proximal [to patient] end and an external handle at the distal end. The intrauterine balloon is inflated with air using a standard syringe (not included). The device has a vaginal cup that is pushed along the shaft to support a green cervical cup. The cervical cup provides a guide for colpotomy. There are four sizes of cervical cups available, 32mm, 37mm, and 40mm, which correspond to the different model numbers. A locking assembly secures the position of the vaginal cup and therefore the cervical cup. Sites for suturing are located on the cervical cup. An Occluder Balloon further reduces
### Asia Pacific
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Symmetry Surgical Company. The logo features the words "symmetry surgical" in a sans-serif font, with the word "company" in a smaller font size to the right. To the left of the text is a red and gray graphic element.
Image /page/4/Picture/2 description: The image shows the logo for "THE O.R. COMPANY". The letters "O.R." are inside of a blue circle. Below the company name is the tagline "The smart way to operate".
leakage of pneumoperitoneum during laparoscopic surgery. A port at the distal end of handle provides a channel for injection of fluids or gases.
#### V. INDICATIONS FOR USE
The Uterine ElevatOR PRO with OccludOR Balloon™ is indicated for manipulation of the uterus, and injection of fluids or gases during laparoscopic procedures including total laparoscopic hysterectomy, laparoscopic assisted vaginal hysterectomy, laparoscopic tubal occlusion, and diagnostic laparoscopy. The Uterine ElevatOR PRO with OccludOR Balloon™ maintains pneumoperitoneum during laparoscopic procedures by sealing the vagina once colpotomy is performed.
| Attributes | Uterine ElevatOR<br>PRO with<br>OccludOR Balloon<br>K220202 | VCARE® Vaginal-<br>Cervical-<br>Ahluwalia's-<br>Retractor-<br>Elevator K142716 | Discussion |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | The O R Company Pty<br>Ltd | CONMED<br>Corporation | --- |
| Class | II | II | Same |
| Regulation Number | 884.4530 | 884.4530 | Same |
| Regulation Generic Name | Cannula,<br>Manipulator/Injector,<br>Uterine | Cannula,<br>Manipulator/Injector,<br>Uterine | Same |
| Indications for Use | The Uterine ElevatOR<br>PRO with OccludOR<br>Balloon™is indicated<br>for manipulation of<br>the uterus, and<br>injection of fluids or<br>gasses during<br>laparoscopic<br>procedures including<br>total laparoscopic<br>hysterectomy,<br>laparoscopic assisted<br>vaginal<br>hysterectomy,<br>laparoscopic tubal<br>occlusion, and<br>diagnostic<br>laparoscopy. The<br>Uterine ElevatOR<br>PRO with OccludOR<br>Balloon™ maintains<br>pneumoperitoneum<br>during laparoscopic | The ConMed VCARE®<br>Retractor/Elevator is<br>indicated for<br>manipulation of the<br>uterus and injection<br>of fluids or gases<br>during laparoscopic<br>procedures such as<br>Laparoscopic<br>Assisted Vaginal<br>Hysterectomy<br>(LAVH), Total<br>Laparoscopic<br>Hysterectomy (TLH),<br>minilap, laparoscopic<br>tubal occlusion, or<br>diagnostic<br>laparoscopy and also<br>maintains<br>pneumoperitoneum<br>by sealing the vagina<br>once a colpotomy is<br>performed. | Same |
| | procedures by<br>sealing the vagina<br>once colpotomy is<br>performed. | | |
| Sterile | Yes, Ethylene Oxide | Yes, Ethylene Oxide | Same |
| Prescription Use | Yes | Yes | Same |
| Single Use | Yes | Yes | Same |
| Intrauterine Balloon inflation volume | 7 – 10 сс | 7 – 10 сс | Same |
| Intrauterine Balloon inflation method | The intrauterine<br>balloon is inflated by<br>passing air from a<br>syringe through a<br>pilot balloon located<br>at the distal end of<br>the device. Inflation<br>is maintained by a<br>one-way valve<br>positioned in the<br>pilot balloon. The<br>pilot balloon also<br>serves as an indicator<br>of intrauterine<br>balloon inflation. | The intrauterine<br>balloon is inflated by<br>passing air from a<br>syringe through a<br>pilot balloon located<br>at the distal end of<br>the device. Inflation<br>is maintained by a<br>one-way valve<br>positioned in the<br>pilot balloon. The<br>pilot balloon also<br>serves as an indicator<br>of intrauterine<br>balloon inflation. | Same |
| Occluder Balloon | Yes | No | Different |
| Mode of action | Mechanical device<br>positioning prior to<br>performing indicated<br>laparoscopic<br>procedures | Mechanical device<br>positioning prior to<br>performing indicated<br>laparoscopic<br>procedures | Same |
| Rigid, insulated and anatomically<br>curved shaft | Yes | Yes | Same |
| Handle at the distal [to patient] end | Yes | Yes | Same |
| Intrauterine balloon at proximal end<br>[to patient] end | Yes | Yes | Same |
| Intrauterine balloon material | PVC | PVC | Same |
| Device Components | Direct patient<br>contacting:<br>Plug cap<br>Intrauterine Balloon<br>Intrauterine Balloon<br>Colorant<br>Heat-shrink on the<br>shaft<br>Stopper on the shaft<br>Graduation marking<br>Cervical cup | Direct patient<br>contacting:<br>Intrauterine Balloon<br>Intrauterine Balloon<br>Colorant<br>Heat-shrink on the<br>shaft<br>Stopper on the shaft<br>Graduation marking<br>Cervical cup<br>Vaginal Cup | Similar |
| | OccludOR balloon<br>OccludOR balloon<br>inflation line<br>Non-contacting:<br>Shaft<br>Thumbscrew &<br>locking assembly<br>Thumbscrew &<br>locking assembly<br>colorant<br>Handle<br>Pilot balloon<br>Air inflation valve<br>Air Inflation line<br>Dye Injection port<br>Dye Injection port<br>cap<br>Robert Clamp<br>Inflation port for<br>OccludOR balloon | Non-contacting:<br>Shaft<br>Thumbscrew &<br>locking assembly<br>Thumbscrew &<br>locking assembly<br>colorant<br>Handle<br>Pilot balloon<br>Air inflation valve<br>Air Inflation line<br>Dye Injection port<br>Dye Injection port<br>cap | |
| Anatomical site | Mucosal<br>Membranes- Uterus<br>and Vagina | Mucosal<br>Membranes- Uterus<br>and Vagina | Same |
| Contact Period | Short-term contact<br>(≤24 hours) | Short-term contact<br>(≤24 hours) | Same |
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VI.
### Asia Pacific
T +61 3 9413 5555 F +61 3 9413 5556
North America
USA
3034 Owen Drive T +1 800-251-3000 Antioch TN, 37013 F +1 800-342-3272
Contact
info@theorcompany.com theorcompany.com
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "THE O.R. COMPANY". The letters "O.R." are inside of a blue circle. Below the company name is the tagline "The smart way to operate".
Image /page/5/Picture/1 description: The image shows the logo for Asymmetry. The logo consists of the letter A followed by a stylized letter S. The letter S is split into two parts, with the top part being gray and the bottom part being red. The word "symmetry" is written in gray to the right of the letter S.
symmetry surgical" company
vie.
### Asia Pacific
T +61 3 9413 5555
F +61 3 9413 5556
North America
3034 Owen Drive
Antioch TN, 37013
USA
T +1 800-251-3000
F+1 800-342-3272
Contact
info@theorcompany.com theorcompany.com
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for "THE O.R. COMPANY". The company name is in blue font, with the "O.R." in a blue circle. Below the company name is the tagline "The smart way to operate".
Image /page/6/Picture/1 description: The image shows the logo for Symmetry Surgical. The logo consists of the letter "A" followed by a stylized "S" made up of two curved shapes, one red and one gray. To the right of the logo is the text "symmetry surgical" in a gray sans-serif font.
Both the subject and predicate devices are sterile, single-use medical devices used to manipulate the uterus and cervix in surgical and diagnostic procedures. At a high level, the subject and predicate devices are based on the following same technological elements:
- An anatomically contoured shaft. ●
- An intrauterine balloon at the proximal [to patient] end, a handle at the distal end. ●
- A cervical cup and a vaginal cup.
- . A locking mechanism to secure the vaginal cup in position.
- An inflation line for the intrauterine balloon.
- A dye injection port at the distal end of the handle. ●
The following technological differences exist between the subject and predicate devices:
- The subject device includes the OccludOR Balloon, which is similar to a feature of ● the Colpo-Pneumo Occluder device cleared under K954311.
These differences do not raise different questions of safety and effectiveness as compared to the predicate.
### Asia Pacific
3201, Australia
1/32 Silkwood Rise T +61 3 9413 5555
USA
3034 Owen Drive T +1 800-251-3000 Antioch TN, 37013 F +1 800-342-3272 Contact
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "THE O.R. COMPANY". The "O.R." is inside of a blue circle. Below the company name is the tagline "The smart way to operate".
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA guidance "Use of International Standard ISO 10993-1: 2018, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The subject device underwent the following tests:
- Cytotoxicity per ISO 10993-5:2009
- Irritation per ISO 10993-10:2010
- Sensitization per ISO 10993-5:2010
- Acute Systemic toxicity per ISO 10993-11:2017
- Pyrogenicity per ISO 10993-11 and USP <151>
The subject device is considered externally communicating device & surface contact device that may come in contact with tissue and breached or compromised surface respectively for a duration of less than 24 hours.
## Performance & mechanical testing
The following performance and mechanical tests demonstrated that the subject device performs substantially equivalent to the predicate device:
- Intrauterine balloon integrity & concentric inspection ●
- Intrauterine balloon inflation stability test for a defined time period.
- Intrauterine balloon burst test
- Intrauterine balloon fatigue test ●
- Intrauterine balloon adhesion strength test ●
- OccludOR balloon outer diameter inspection & leakage free test
- OccludOR balloon deflation time test
- OccludOR balloon capacity test- reliability/safety
- OccludOR balloon fatigue durability test
- OccludOR balloon burst test
- Vaginal cup air leakage test
- Cervical cup temperature test
- Air inflation valve test
- Dye-injection port compliance & leakage test
- Thumb screw strength test
- Locking force test of vaginal cup
- Ease of movement of vaginal cup along the shaft ●
- Cervical cup detachment force test
- Cervical cup stopper adhesion strength test ●
### Asia Pacific
North A
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for "The O.R. Company". The letters "O.R." are inside of a blue circle. Below the company name is the tagline "The smart way to operate".
symmetry surgical" company
- . Device tensile strength
- . Handle joint strength
- . Tip deflection
- Ultrasonic welding strength
In all testing, the subject device met the pre-specified acceptance criteria.
## Conclusion:
The intended use of the subject device is identical to the predicate device. The subject device has similar technological characteristics to the predicate device. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. All testing undertaken to demonstrate substantial equivalence of the subject device meet the predetermined acceptance criteria. The test results demonstrate that the subject device is as safe and effective as the predicate device.
Asia Pacific
North America
3034 Owen Drive Antioch TN, 37013 USA
T +1 800-251-3000 F +1 800-342-3272 Contact
info@theorcompany.com theorcompany.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.