K220194 · Winner Medical Co., Ltd. · FXX · May 17, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220194
Device Name
Procedure mask/Surgical mask/Face mask
Applicant
Winner Medical Co., Ltd.
Product Code
FXX · General, Plastic Surgery
Decision Date
May 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Procedure mask/Surgical mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.
Device Story
Three-layer, flat-pleated, single-use surgical face mask; inner/outer layers polypropylene nonwoven; middle layer polypropylene melt-blown nonwoven. Available in ear-loop (polyester/spandex) or tie-on (polypropylene) configurations. Includes iron/polypropylene nose clip for fit adjustment. Used in clinical settings by healthcare personnel and patients to reduce exposure to blood, body fluids, and particulates. Device provides physical barrier; performance validated against ASTM F2100-19 standards for fluid resistance, particulate filtration, and bacterial filtration. Non-sterile, disposable.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ASTM F1862 (synthetic blood penetration), ASTM F2299 (PFE), ASTM F2101 (BFE), EN 14683 (differential pressure), 16 CFR 1610 (flammability), and ISO 10993-5/10 (cytotoxicity, sensitization, irritation). All results met ASTM F2100-19 Level 1 and Level 2 acceptance criteria.
Technological Characteristics
Three-layer polypropylene nonwoven construction. Sensing/actuation: None (passive barrier). Dimensions: 17.5x9.5 cm or 14.5x9 cm. Connectivity: None. Sterilization: Non-sterile. Standards: ASTM F2100-19, ASTM F1862, ASTM F2299, ASTM F2101, EN 14683, 16 CFR 1610, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for use by patients and healthcare personnel to protect against transfer of microorganisms, body fluids, and particulate material in infection control practices. Single-use, disposable, non-sterile.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Disposable Ear-loop Medical Face Mask Disposable Tie-On Medical Face Mask (K211462)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 17, 2022
Winner Medical Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O. Box 120-119 Shanghai, 200120 China
Re: K220194
Trade/Device Name: Procedure mask/Surgical mask/Face mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: April 9, 2022 Received: April 20, 2022
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K220194
Device Name
Procedure mask/Surgical mask/Face mask
## Indications for Use (Describe)
The Procedure mask/Surgical mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section
807.92. The assigned 510(k) Number: K220194
- 1. Date of Preparation: 01/18/2022
- 2. Sponsor Identification
## Winner Medical Co., Ltd.
Winner Industrial Park, No.660 Bulong Road, Longhua District, Shenzhen Guangdong, China 518109
Establishment Registration Number: 9616433
Contact Person: Mingni Liu Position: Regulatory Affairs Tel: +86-755-28138888-8822 Email: 2346@winnemedical.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jinlei Tang (Alternative Contact Person)
## Mid-Link Consulting Co., Ltd.
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Procedure mask/Surgical mask/Face mask Common Name: Surgical Face Mask
Regulatory Information Classification Name: Mask, Surgical; Classification: II; Product Code: FXX; Regulation Number: 21CFR 878.4040; Review Panel: General Hospital;
### Indications for use:
The Procedure mask/Surgical mask/Face mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.
### Device Description:
The proposed device is a three-layer, single-use, flat- pleated mask. The inner and outer layers of the mask are made of polypropylene nonwoven, and the middle layer is made of polypropylene melt-blown nonwoven. The proposed devices are available in two types, ear loop and tie-on. The ear loops are made of polyester and spandex, and the ties are made of polypropylene nonwoven. The ear loops/ties are used to secure the mask over the users' mouth and face. The nose clip is made of Iron and polypropylene. Users can adjust the nose piece according to the shape of the nose, and fix the mask on the bridge of the nose to prevent the mask from falling off. The ear loop masks are available in two size, 17.5×9.5 cm and 14.5×9 cm; the tie-on mask is available in one size, 17.5×9.5 cm. And the colors for the ear loop mask are blue and black; the color for the tie-on mask is blue. Both the ear loop and tie-on masks are available in level 1 and level 2 masks based on ASTM F2100-19 due to the difference in gram weight of the mask body. The proposed device is provided in non-sterile. All specifications of the proposed device are provided in table 1.
| Product name | ASTM F2100-19<br>level | Ear strap type | Size (cm) | Color | Layers | Sterilization |
|-------------------------------------------------|------------------------|----------------|-----------|-------|--------|---------------|
| Procedure<br>mask/Surgical<br>mask/Face<br>mask | Level 1 | Ear loop | 17.5x9.5 | Blue | 3 | Non-sterile |
| | | | 17.5x9.5 | Black | | |
| | | Ear loop | 14.5x9 | Blue | | |
| | | | 14.5x9 | Black | | |
| | | Tie-on | 17.5x9.5 | Blue | | |
| Table 1 Procedure mask/Surgical mask Description | | |
|--------------------------------------------------|--|--|
| | | |
{5}------------------------------------------------
| | Level 2 | Ear loop | 17.5x9.5 | Blue | |
|--|---------|----------|----------|-------|--|
| | | | | Black | |
| | | | 14.5x9 | Blue | |
| | | | | Black | |
| | | Tie-on | 17.5x9.5 | Blue | |
#### న్. Identification of Predicate Device
510(k) Number: K211462 Product Name: Disposable Ear-loop Medical Face Mask Disposable Tie-On Medical Face Mask (Select Level 1 and Level 2 masks as predicate device)
#### Summary of Non-Clinical Test 6.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was same/similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ASTM F1862/F1862M: 2017 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity);
- > ASTM F2299/F2299M-03 (2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres;
- > ASTM F2101: 2019 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus;
- > ASTM F2100: 2019 Standard Specification for Performance of Materials Used in Medical Face Masks
- > EN 14683:2019+AC: 2019 Annex C Medical face masks - Requirements and test methods
- > 16 CFR Part 1610 Standard for the Flammability of Clothing Textiles;
- > ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritaion and skin sensitization
| Test Methodology | Purpose | Acceptance Criteria | Result |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Resistance to<br>Penetration by<br>Synthetic blood | The test was performed in accordance with ASTM<br>F1862/F1862M: 2017<br>Standard Test Method for<br>Resistance of Medical Face | Level 1: No penetration<br>at 80 mmHg | Level 1: Pass at<br>80mmHg |
| | | Level 2: No penetration<br>at 120 mmHg | Level 2: Pass at<br>120mmHg |
| | Masks to Penetration by<br>Synthetic Blood (Horizontal<br>Projection of Fixed Volume at<br>a Known Velocity) to evaluate<br>the effectiveness of the test<br>article from possible exposure<br>to blood and other body fluids. | | |
| | The test was performed in<br>accordance with ASTM<br>F2299/F2299M-03 (2017)<br>Standard Test Method for<br>Determining the Initial<br>Efficiency of Material Used in<br>medical Face Masks to<br>Penetration by Particulates<br>using Latex Spheres to<br>determine the particle<br>filtration efficiency (PFE) of<br>the test article. | Level 1: ≥95% | Blue mask:<br>Pass at 96.05%<br>Black mask:<br>Pass at 96.03% |
| Particulate<br>Filtration Efficiency | | Level 2: ≥98% | Blue mask:<br>Pass at 98.78%<br>Black mask:<br>Pass at 98.75% |
| Bacterial Filtration<br>Efficiency | The test was performed in<br>accordance with ASTM<br>F2101: 2019 Standard Test<br>Method for Evaluating the<br>Bacterial Filtration Efficiency<br>(BFE) of Medical Face Mask<br>Materials, Using a Biological<br>Aerosol of Staphylococcus<br>aureus to determine the<br>bacterial filtration efficiency<br>(BFE) of the test article. | Level 1: ≥95% | Blue mask:<br>Pass at 98.25%<br>Black mask:<br>Pass at 98.25% |
| | | Level 2: ≥98% | Blue mask:<br>Pass at 98.72%<br>Black mask:<br>Pass at 98.73% |
| Differential<br>Pressure | The test was performed in<br>accordance with EN<br>14683:2019+AC: 2019 Annex<br>C Medical face masks -<br>Requirements and test<br>methods to determine the<br>differential pressure of the test<br>article. | Level 1:<br><5.0 mmH2O/cm² | Blue mask:<br>Pass at 3.5 mmH2O/cm²<br>Black mask:<br>Pass at 3.5 mmH2O/cm² |
| | | Level 2:<br><6.0 mmH2O/cm² | Blue mask:<br>Pass at 3.5 mmH2O/cm² |
| | | | |
| | | | Black mask:<br>Pass at 3.6 mmH2O/cm2 |
| Flammability | The test was performed in accordance with 16 CFR Part 1610 Standard for the Flammability of Clothing Textiles to evaluate the flammability of the test article. | Class 1 | Class 1 |
| Cytotoxicity | The test was performed in accordance with ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity to evaluate the cytotoxicity of the test article. | The viability should be ≥70% of the blank. And the 50% extract of the test sample should have at least the same or a higher viability than the 100% extract. | The viability was ≥70% of the blank. And the 50% extract of the test sample had a higher viability than the 100% extract.<br>Under the conditions of the study, the proposed device was non-cytotoxic. |
| Sensitization | The test was performed in accordance with ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization to evaluate the sensitization of the test article. | Non-sensitizing | Under the conditions of the study, the proposed device was non-sensitizing. |
| Irritation | The test was performed in accordance with ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization to evaluate the irritation of the test article. | Non-irritating | Under the conditions of the study, the proposed device was non-irritating. |
Table 2 Comparison of Procedure mask/Surgical mask
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### 7. Clinical Test Conclusion
No clinical study is included in this submission.
#### 8. Technological Characteristics Comparison
| ITEM | Proposed Device | Predicate Device<br>K211462 | Remark | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|------|
| Product Code | FXX | FXX | Same | |
| Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same | |
| Class | II | II | Same | |
| Indications<br>for<br>Use | The Procedure mask/Surgical<br>mask is intended to be worn to<br>protect both the patient and<br>healthcare personnel from transfer<br>of microorganisms, body fluids,<br>and particulate material. These<br>face masks are intended for use in<br>infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This is a<br>single-use, disposable device,<br>provided non-sterile. | The Disposable Ear-loop Medical Face<br>Mask/Disposable Tie-On Medical Face<br>Mask is intended to be worn to protect<br>both the patient and healthcare personnel<br>from transfer of microorganisms, body<br>fluids, and particulate material. These<br>face masks are intended for use in<br>infection control practices to reduce the<br>potential exposure to blood and body<br>fluids. This is a single-use, disposable<br>device, provided nonsterile. | Same | |
| Mask style | Flat pleated | Flat pleated | Same | |
| Design feature | Ear loop / Tie-on | Ear loop / Tie-on | Same | |
| Dimension | Ear loops:<br>Body: 175 mm×95 mm, nose clip:<br>110mm, Ear-loop: 178mm<br>Body: 145 mm×90 mm, nose clip:<br>85mm, Ear-loop: 145mm<br>Tie-on:<br>Body: 175 mm×95 mm, nose clip:<br>110mm, Ear-loop: 900mm | Ear loops: Body: 175 mm×95 mm, nose<br>clip: 125mm, Ear-loop: 175mm<br>Tie-on: Body: 175 mm×95 mm, nose<br>clip: 125mm, Tie strings: 910mm | Different | |
| ASTM<br>F2100<br>Level | Level 1 | Level 1 | Level 2 | Same |
| Fluid<br>Resistance | Pass<br>at<br>80mmHg | Pass at 80mmHg | Pass<br>at<br>120<br>mmHg | Same |
## Table 3 Comparison of Procedure mask/Surgical mask
{9}------------------------------------------------
| Particulate<br>filtration<br>efficiency | | Blue mask:<br>Pass at 96.05% | Blue mask:<br>Pass at 98.78% | Pass at 98.2% | | Pass at 99.3% | Different |
|-----------------------------------------|--------------------|-----------------------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------------------------------------|---------------------------------|--------------------------------|-----------|
| | | Black mask:<br>Pass at 96.03% | Black mask:<br>Pass at 98.75% | | | | |
| Bacterial<br>filtration<br>efficiency | | Blue mask:<br>Pass at 98.25% | Blue mask:<br>Pass at 98.72% | Pass at 98.9% | | Pass at 99.4% | Different |
| | | Black mask:<br>Pass at 98.25% | Black mask:<br>Pass at 98.73% | | | | |
| Differential<br>pressure | | Blue mask:<br>Pass at 3.5 mmH2O/cm² | Blue mask:<br>Pass at 3.5 mmH2O/cm² | Passed at 3.6mmH2O/cm² | | Pass at 4.0mmH2O/cm2 | Different |
| | | Black mask:<br>Pass at 3.5 mmH2O/cm² | Black mask:<br>Pass at 3.6 mmH2O/cm² | | | | |
| | | | | | | | |
| Flammability | | Class 1 | | Class 1 | | | Same |
| Label/Labeling | | Complied with 21 CFR part 801 | | Complied with 21 CFR part 801 | | | Same |
| Materials | | | | | | | |
| Ear loop | | Polyester and spandex | | Nylon and Spandex | | | |
| Tie strings | | 38 /m² polypropylene nonwoven | | 35g/ m² PP non-woven cloth | | | |
| Nose clip | | Iron and polypropylene | | Aluminum wire | | | |
| Mask<br>body | Outer<br>material | 25g/m²<br>polypropylene<br>nonwoven | 30g/m²<br>polypropylene<br>nonwoven | 23g/ m² PP non-woven cloth | | | Different |
| | Middle<br>material | 25g/m²<br>polypropylene<br>melt-blown<br>nonwoven | 30g/m²<br>polypropylene<br>melt-blown<br>nonwoven | 22g/ m² PP<br>non-woven<br>cloth | 25g/m² PP<br>non-woven<br>cloth | 33g/m²PP<br>non-woven<br>cloth | |
| | Inner<br>material | 25g/m²<br>polypropylene<br>nonwoven | 30g/m²<br>polypropylene<br>nonwoven | 20g/ m² PP non-woven cloth | | | |
| Colors | | Blue; Black | | Blue | | | Different |
| Biocompatibility | | | | | | | |
| Cytotoxicity | | Under the conditions of the study, the proposed device was non-cytotoxic. | | Under the conditions of the study, the proposed device was non-cytotoxic. | | | Same |
| Sensitization | | Under the conditions of the study, the proposed device was non-sensitizing. | | Under the conditions of the study, the proposed device was non-sensitizing. | | | Same |
| Irritation | | Under the conditions of the study, the proposed device was non-irritating. | | Under the conditions of the study, the proposed device was non-irritating. | | | Same |
{10}------------------------------------------------
### 510(k) Summary
| | non-irritating. | | |
|---------------|-----------------|-------------|------|
| Sterilization | | | |
| Method | Non-sterile | Non-sterile | Same |
## Different - Dimension (mm)
The dimension for the proposed device is different from the predicate device. The mask body of the proposed device is provided in two dimensions: ear loop masks - 17.5×9.5 cm and 14.5×9 cm; tie-on mask - 17.5×9.5 cm. While the predicate device is only provided in one dimension, 17.5×9.5 cm. However, the difference in dimension of mask body does not affect indications for use and will not raise any safety issues. The dimensions of the nose clip, ear loop and ties for the proposed device are different from the predicate device. However, these difference does not affect the protective performance of masks. The nose clip is just to fix the mask on the bridge of the nose and keep it from falling off. And the ear loop and ties are also designed to attach masks to the user's mouth and nose. Thus, this difference will not affect the safety and effectiveness of the proposed device.
## Different - Particulate efficiency
The test result for particulate filtration efficiency for the proposed device is different from the predicate device. However, the test result for the proposed device can meet the requirements of level 1 / level 2 mask based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of proposed device.
### Different - Bacterial filtration efficiency
The test result for bacterial filtration efficiency for the proposed device is different from the predicate device. However, the test result for the proposed device can meet the requirements of level 1 / level 2 mask based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of proposed device.
## Different - Differential pressure
The test result for differential pressure for the proposed device is different from the predicate device. However, the test result for the proposed device can meet the requirements of level 1 / level 2 mask based on ASTM F2100-19. Thus, this difference will not affectiveness of proposed device.
### Different - Materials
The material for the proposed device is different from predicate device. However, biocompatibility test has been performed on the proposed device and the results does not show any adverse effect. Thus, this difference in materials will not affect the safety and effectiveness of the proposed device.
## Different - Colors
The color of proposed device is provided in blue and black, while the predicate device is blue. However,
{11}------------------------------------------------
the biocompatibility test has been conducted and test results did not show any adverse effects. Thus, this difference will not affect the safety and effectiveness of the proposed device.
- Conclusion 9.
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device K211462.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.