K220189 · Omni-Guide Holdings, Inc. · GEX · Apr 29, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220189
Device Name
LISA Laser Surgical Fibers
Applicant
Omni-Guide Holdings, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 29, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Omni-Guide Holdings, Inc. single-use LISA Laser Surgical Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500mm have received regulatory clearance. Omni-Guide Holdings, Inc. single-use LISA Laser Surgical Fibers devices are intended for use with any cleared surgical laser with an SMA 905 connector, SMA 906 connector, or manufacturer specific connectors and adapters.
Device Story
LISA Laser Surgical Fibers are sterile, single-use fiber optic delivery systems for surgical lasers. Input: laser energy from compatible surgical laser systems. Transformation: propagation of light energy through glass fiber optic cable to the surgical site. Output: directed laser energy for tissue ablation, coagulation, incision, excision, or vaporization. Used in open, laparoscopic, or endoscopic procedures by physicians. Construction includes fused silica core, fused silica or fluoropolymer hard cladding, and various jackets (nylon, polyimide, Teflon). Connectors (SMA-905, SMA-906, or proprietary) attach via crimp or adhesive. Benefits include precise energy delivery to target tissue, facilitating various surgical applications.
Clinical Evidence
Bench testing only. No clinical or animal data required. Performance testing included power/energy transmission, beam profile/homogeneity, high-power testing, bend radius, pull testing, and feature-specific testing (e.g., navigation tip deflection). Packaging and sterilization validated per ISO 11135-1, ISO 11607-1, and ASTM standards.
Technological Characteristics
Single-use fiber optic delivery system. Materials: Fused silica core, fused silica or fluoropolymer hard cladding, silicone acrylate buffer, nylon/polyimide/Teflon jacket. Numerical Aperture: 0.22-0.48. Wavelengths: 500nm-2200nm. Power: 1-300W. Connectors: SMA 905, SMA 906, or manufacturer-specific. Sterilization: Ethylene Oxide (EtO). Standards: ISO 10993 (biocompatibility), ISO 11135-1 (sterilization), ISO 11607-1 (packaging), ASTM F88/F2096/F1980 (seal/leak/aging).
Indications for Use
Indicated for use in all surgical specialties requiring laser energy delivery (500nm–2200nm) for ablation, coagulation, incision, excision, or vaporization of soft/hard tissue. Compatible with cleared surgical lasers using SMA 905, SMA 906, or manufacturer-specific connectors.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Laser Peripherals' Family of Surgical Fiber Optic Laser Delivery Device (K170366)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
April 29, 2022
Omni-Guide Holdings, Inc. Carlos Acosta Global Director Regulatory Affairs & Quality Assurance 46 Manning Road Billerica, Massachusetts 01821
Re: K220189
Trade/Device Name: LISA Laser Surgical Fibers Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 12, 2022 Received: January 24, 2022
Dear Carlos Acosta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K220189
Device Name Omni-Guide LISA Laser Surgical Fibers
#### Indications for Use (Describe)
Omni-Guide Holdings, Inc. single-use LISA Laser Surgical Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500mm have received regulatory clearance. Omni-Guide Holdings, Inc. single-use LISA Laser Surgical Fibers devices are intended for use with any cleared surgical laser with an SMA 905 connector, SMA 906 connector, or manufacturer specific connectors and adapters.
| Type of Use (Select one or both, as applicable) | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <table border="0"><tr><td><span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </td></tr><tr><td> Over-The-Counter Use (21 CFR 801 Subpart C) </td></tr></table> | <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| | | | | 008_510(k)_Summary |
|--|--|--|--|--------------------|
|--|--|--|--|--------------------|
| Classification Name | Common Name | Trade Name/Proprietary Name |
|----------------------------------------------------------|----------------------------|-----------------------------|
| Part 878 General and Plastic Surgery 21 CFR<br>§874.4680 | LISA Laser Surgical Fibers | Surgical Fibers |
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of Omni Guide knowledge.
| Submitter: | Omni-Guide Holdings, Inc.<br>46 Manning Drive<br>Billerica, MA 01821 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Carlos O. Acosta<br>Phone: 617-551-4400 Ext. 563<br>Fax: 888-490-6020 |
| 510(k) Co-<br>Author | Irina Fedorov<br>Phone: 617-551-4400 Ext.<br>Fax: 888-490-6020 |
| Date of<br>Preparation: | December 16, 2021 |
| Type of 510(k)<br>Submission: | Traditional |
| Device<br>Identification: | Trade Name: LISA Laser Surgical Fibers<br>Classification Name: Laser surgical instrument for use in general and plastic surgery and in<br>dermatology |
| Regulatory<br>Class: | II |
| Product Code: | GEX |
| Panel<br>Classification | General and Plastic Surgery (21 CFR §878.4810) |
{4}------------------------------------------------
# K220189
| Predicate Device: | Primary Predicate Device:<br>Laser Peripherals' Family of Surgical Fiber Optic Laser Delivery<br>Device 510(k) = K170366 | | | |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|---------------------|
| Device<br>Description: | Omni-Guide Holding, Inc. LISA Laser Surgical Fibers are individually packed sterile devices to provide high<br>quality surgical laser fiber optic delivery systems for laser surgery. The Fibers are for use in laser surgical<br>procedures, including open, laparoscopic, or endoscopic ablation, coagulation, incision, and excision<br>or vaporizing, in soft/hard-tissue application for which compatible surgical lasers are applicable. The key<br>components are the fiber optic, and either an SMA-905, SMA-906, or manufacturer's specific<br>connector. The fibers allow energy to be delivered to the surgical site through the use of glass fiber optic cable.<br>Propagation of the light energy through the glass provides users the ability to direct the energy to the appropriate<br>areas in the body.<br>The Omni Guide Surgical Holding, Inc. single-use LISA Laser Surgical Fibers family of fibers are marketed to<br>physicians as a sterile device within a typical Tyvek/poly pouch. Internally, the fiber is either secured in a wound<br>circle with a Tyvek band, backer-card, or placed in a polyethylene tubing hoop to allow ease of removal. The<br>construction of the fibers is dependent of the application to which it's marketed. There are two main<br>methods of attaching the connector : crimp and adhesive. The crimp offers a reliable connection that will not<br>be affected by heat whereas the epoxy allows a lower cost alternative where high temperatures are not an<br>issue. All fibers offer various connector nuts, heat shrinks, abrasion jackets, strain reliefs or caps that best serve<br>the environment where they are utilized. | | | |
| Indications For<br>Use: | Omni-Guide Holdings, Inc. single-use LISA Laser Surgical Fibers are indicated for use in all surgical<br>specialties in which compatible laser systems with operational wavelengths between 500nm – 2200nm have<br>received regulatory clearance. Omni-Guide Holdings, Inc. single-use LISA Laser Surgical Fibers devices are<br>intended for use with any cleared surgical laser with an SMA 905 connector, SMA 906 connector, or<br>manufacturer specific connectors and adapters. | | | |
| Substantial<br>Equivalence<br>Comparison<br>Table | Characteristic | Subject Device<br>LISA Laser Surgical Fiber | Primary Predicate, Laser<br>Peripherals surgical fiber<br>optic laser delivery devices<br>K170366 | Equivalence |
| | 510(k) Number | Pending | K170366 | N/A |
| | Product Code | GEX | GEX | Same/<br>Equivalent |
| | Regulation<br>Number | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| | Regulation<br>Name | Laser surgical instrument for<br>use in general and plastic<br>surgery and in dermatology | Laser surgical instrument for<br>use in general and plastic<br>surgery and in dermatology | Same/equivalent |
| | Components | Connector configurations<br>offered suitable to<br>multiple laser platforms | Connector configurations<br>offered suitable to<br>multiple laser platforms | Same/Equivalent |
| | Technological<br>characteristics | High OH or Low OH silica<br>material w/low attenuation of<br>light wavelengths between<br>532nm - 2100nm | High OH or Low OH silica<br>material w/low attenuation of<br>light wavelengths between<br>532nm - 2100nm | Same/Equivalent |
| | Silica/Silica<br>fibers/hard clad<br>fibers | Fiber having a Numerical<br>Aperture (NA) between<br>0.22 - 0.48 | Fiber having a Numerical<br>Aperture (NA) between 0.22<br>- 0.48 | Same/Equivalent |
| | Fiber distal tip | Multiple configurations of<br>distal tips offered to provide<br>the most suitable performance<br>for the application | Multiple configurations of<br>distal tips offered to provide<br>the most suitable performance<br>for the application | Same/Equivalent |
| | Used with<br>attachments | Core diameters are offered<br>in a range of sizes suitable to<br>user needs | Core diameters are offered<br>in a range of sizes suitable to<br>user needs | Same/Equivalent |
| Connectors | SMA 905 connector, SMA 906 connector, or manufacturer specific connectors and adaptors | SMA 905 connector, SMA 906 connector, or manufacturer specific connectors and adaptors | Same/Equivalent | |
| Fiber<br>Construction | Core - Fused Silica<br>Clad – Fused Silica or Fluoropolymer Hard Cladding<br>Buffer – Fluoropolymer Hard Cladding or Silicone Acrylate<br>Jacket - Nylon, Polyimide, or Teflon | Core - Fused Silica<br>Clad – Fused Silica or Fluoropolymer Hard Cladding<br>Buffer – Fluoropolymer Hard Cladding or Silicone Acrylate<br>Jacket - Nylon, Polyimide, or Teflon | Same/Equivalent | |
| Fiber<br>Numerical<br>Aperture | Fiber having a numerical<br>Aperture (NA) between 0.22-<br>048 | Fiber having a Numerical<br>Aperture (NA) between 0.22-<br>048 | Same/Equivalent | |
| Peak and<br>Continuous<br>Wave<br>lengths | 500nm-2200nm | 500nm-2200nm | Same/Equivalent | |
| Power Ranges | 1-300 Watts | 1-300 Watts | Same/Equivalent | |
| Diameter | Core diameters are<br>offered in a range of sizes<br>suitable to user needs | Core diameters are<br>offered in a range of sizes<br>suitable to user needs | Same/Equivalent | |
| Fiber Distal<br>Tip | Multiple configurations of<br>distal tips offered to provide<br>the most suitable performance<br>for the application | Multiple configurations of<br>distal tips offered to provide<br>the most suitable performance<br>for the application | Same/Equivalent | |
| Compatibility<br>with surgical<br>laser systems | Fibers are compatible<br>with any cleared laser<br>system with an<br>appropriate connection<br>system | Fibers are compatible<br>with any cleared laser<br>system with an<br>appropriate connection<br>system | Same/Equivalent | |
| | Sterilization | EtO | EtO | Same/Equivalent |
| Non-Clinical<br>Performance<br>Data: | Bench testing on the subject device has shown the device to perform as intended with the same technological<br>principle, fit, form, function and method of operating, as the single use fibers of K170366. Fibers included<br>in this premarket notification were tested for conformance to the following performance standards: | | | |
| | Document Number | Title | | |
| | ISO 11135-1 | Sterilization of Health Care Products - Ethylene<br>Oxide - Requirements for the development, validation and routine<br>control sterilization process for medical devices | | |
| | ISO 11607-1 | Packaging for terminally sterilized medical devices - Part 1:<br>Requirements for materials, sterile barrier systems and packaging<br>systems | | |
| | ISO 10993-1 | Biological Evaluation of Medical Devices - Part 1:<br>Evaluation and testing | | |
| | ISO 10993-5 | Biological Evaluation of Medical Devices - Part 5:<br>Tests for in vitro cytotoxicity | | |
| | ISO 10993-7 | Biological Evaluation of Medical Devices - Part 7:<br>Tests for Ethylene Oxide Sterilization Residuals | | |
| | ISO 10993-10 | Biological Evaluation of Medical Devices - Part 5:<br>Tests for Irritation and Skin Sensitization | | |
| | ISO 10993-11 | Biological Evaluation of Medical Devices - Part 5: Tests for Systemic<br>Toxicity | | |
| | ISO 14971 | Medical devices - Application of Risk Management<br>to Medical Devices | | |
| | ASTM D4169 | Standard Practice for Performance Testing of<br>Shipping Containers and Systems | | |
| | ASTM F88-09…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.