Magnus Neuromodulation System (MNS) with SAINT Technology, Model Number 1001K
K220177 · Magnus Medical, Inc. · OBP · Sep 1, 2022 · Neurology
Device Facts
Record ID
K220177
Device Name
Magnus Neuromodulation System (MNS) with SAINT Technology, Model Number 1001K
Applicant
Magnus Medical, Inc.
Product Code
OBP · Neurology
Decision Date
Sep 1, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5805
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Magnus Neuromodulation System (MNS) with SAINT Technology is intended for the delivery of SAINT neuromodulation therapy to treat major depressive disorder (MDD) in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Device Story
Non-invasive repetitive transcranial magnetic stimulation (rTMS) system; delivers individualized, navigationally directed intermittent theta burst stimulation (iTBS) to left dorsolateral prefrontal cortex (L-DLPFC). System inputs: structural and functional MRI data for target identification; motor threshold determination via coil and evoked response monitor. Proprietary software identifies individualized stimulation targets; hardware includes stimulator, treatment coil, neuronavigation system. Used in inpatient or outpatient settings by trained medical professionals. Output: magnetic pulses to brain. Clinical benefit: rapid-acting antidepressant treatment for refractory MDD. Accelerated protocol condenses treatment to five days. Healthcare providers use neuronavigation for real-time visualization of target relative to coil position; output informs clinical decision-making for depression management.
Clinical Evidence
Four clinical studies (n=70 total) evaluated SAINT technology. Study 3 (n=29, randomized sham-controlled) showed 85.7% response and 78.6% remission in active group vs 26.7% response and 13.3% remission in sham. Three open-label studies showed response rates 83.3%-92.8% and remission rates 78.6%-90.5%. Primary endpoint: MADRS/HDRS reduction. No serious adverse events reported. Data compared favorably to THREE-D trial (iTBS predicate).
Technological Characteristics
Non-sterile rTMS system. Components: stimulator, treatment coil, motor threshold coil, neuronavigation system, cooling unit, isolation transformer. Materials compliant with ISO 10993-1. Sensing: EMG for motor threshold. Energy: magnetic pulses. Connectivity: cloud software integration. Software developed per IEC 62304 and ISO 14971. Neuronavigation uses structural/functional MRI data.
Indications for Use
Indicated for treatment of Major Depressive Disorder (MDD) in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
Predicate Devices
Nexstim Navigated Brain Therapy (NBT) System 2 (K182700)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 1, 2022
Magnus Medical, Inc. Susan Noriega VP Regulatory Affairs 1350 Old Bayshore Highway, Suite 600 Burlingame, CA 94010
## Re: K220177
Trade/Device Name: Magnus Neuromodulation System (MNS) with SAINT Technology, Model Number 1001K Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP Dated: August 2, 2022 Received: August 2, 2022
Dear Susan Noriega:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) [not yet assigned] K220177
Device Name
Magnus Neuromodulation System (MNS) with SAINT® Technology
Indications for Use (Describe)
The Magnus Neuromodulation System with SAINT Technology is indicated for the treatment of Maior Depressive Disorder (MDD) in adult patients who have failed to achieve satisfactory improvement from prior antidenressant medication in the current episode.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg height="16" width="16"> <path d="M2 2 L14 2 L14 14 L2 14 Z M2 2" fill="none" stroke="currentColor" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpa </div> |
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Provided in Accordance with 21 CFR §807.92(c)
# magnusmedical
| Date Summary Prepared: | August 30, 2022 |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number: | K220177 |
| 510(k) Owner: | Magnus Medical, Inc.<br>1350 Old Bayshore Highway, Suite 600<br>Burlingame CA 94010 USA<br>(415) 690-7358 |
| Submitter and Official Contact: | Susan Noriega<br>Magnus Medical, Inc.<br>1350 Old Bayshore Highway, Suite 600<br>Burlingame CA 94010 USA<br>susan@magnusmed.com<br>(650) 793-1966 |
| Trade Name: | Magnus Neuromodulation System (MNS) with SAINT<br>Technology |
| Common Name: | Transcranial Magnetic Stimulation Device |
| Classification Name: | Repetitive Transcranial Magnetic Stimulator for Treatment<br>of Major Depressive Disorder |
| Primary Classification Regulation: | 21 CFR §882.5805 |
| Primary Product Code: | OBP |
| Secondary Classification and Code: | 21 CFR §882.1870/GWF<br>21 CFR §882.4560/HAW |
| Substantially Equivalent Device: | Nexstim Navigated Brain Therapy (NBT) System 2<br>Nexstim Plc<br>Helsinki, Finland<br>Premarket Notification K182700<br>Cleared on March 22, 2019 |
| Device Description: | The Magnus Neuromodulation System (MNS) with SAINT<br>Technology is a non-invasive repetitive transcranial<br>magnetic stimulation (rTMS) system that delivers<br>individualized and navigationally directed repetitive<br>magnetic pulses to the left dorsolateral prefrontal cortex (L- |
{4}------------------------------------------------
DLPFC) to treat Major Depressive Disorder (MDD) in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
The MNS with SAINT Technology is available for prescription use only and is intended for use by trained medical professionals in either an inpatient or outpatient setting.
The Magnus Neuromodulation System consists of hardware devices (stimulator with treatment coil and neuronavigation system) intended to deliver SAINT Technology; that is, rTMS (as intermittent theta burst stimulation (iTBS)) to a target area within the L-DLPFC along with proprietary software informed by structural and functional MRI and designed to identify the individualized target within the L-DLPFC for stimulation. Also included in the system are a coil and monitor for motor threshold determination, which is used to inform patient-specific stimulation settings. SAINT Technology is the combination of using the specific individualized target for treatment along with a proprietary accelerated treatment protocol that condenses treatment to five days. The Magnus Neuromodulation System is designed to support successful delivery of SAINT Technology.
The Magnus Neuromodulation System with SAINT Intended Use: Technology is intended for the delivery of SAINT neuromodulation therapy to treat major depressive disorder (MDD) in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Technology Comparison:
The Magnus Neuromodulation System with SAINT Technology has the same intended use and technological characteristics as the predicate device.
Compatible Hardware: The Magnus Neuromodulation System is a complete set of mutually compatible hardware components that have been specifically selected and evaluated to safely and effectively deliver SAINT Technology in conjunction with the Magnus Cloud Software. The following includes the set of hardware components included with the Magnus Neuromodulation System:
{5}------------------------------------------------
| Item | Qty | Magnus Part # | Supplier Name/Model | Supplier Part # | 510(k) Clearance |
|----------------------------------------------------------------|-----|---------------|---------------------------------------------------|----------------------------------------------------------|------------------|
| Stimulator | 1 | 1002 | MagVenture/X100 | 9016E0711 | K173620 |
| Treatment Coil | 1 | 1003 | MagVenture/Cool-B65 | 9016E0491 | K171967 |
| Motor Threshold Coil | 1 | 1004 | MagVenture/C-B60 | 9016E0673 | K171967 |
| Coil Arm | 1 | 1005 | MagVenture/Super Flexible<br>Arm | 9016B0181 | K173620 |
| Stimulator Cart | 1 | 1006 | MagVenture/Trolley with<br>Holding Arrangements | 9016B0383 | K173620 |
| Isolation Transformer | 1 | 1007 | MagVenture/Isolation<br>Transformer | 9016D0031 | K173620 |
| Cooling System | 1 | 1008 | MagVenture/Coil Cooler Unit | 9016B0151 | K173620 |
| Evoked Response<br>Monitor | 1 | 1009 | MagVenture/MEP Monitor | 9016C0711 | K162873 |
| Neuronavigation<br>System with<br>accessories (cart and<br>PC) | 1 | 1010 | Brain Science Tools<br>(Soterix)/Neural Navigator | HD-SWN<br>(nav)<br>HD-SWTA<br>(cart)<br>HD-SWAPC<br>(pc) | K191422 |
MagVenture devices and accessories are supplied by Tonica Elektronik / MagVenture Brain Science Tools Neural Navigator is supplied by Soterix Medical
All hardware components were utilized in non-clinical bench performance testing as included in their respective 510(k) clearances. Stimulator, Treatment Coil, Motor Threshold Coil, Coil Arm, Stimulator Cart, Isolation Transformer, Cooling System, and Evoked Response Monitor hardware components were additionally utilized in clinical performance testing.
| Table 1: Technology Comparison for the Magnus Neuromodulation System vs. the Predicate | | | |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | | | |
| | Predicate Device | Proposed Device | Notes |
| Characteristic | Nexstim NBT System 2<br>K182700 | Magnus Neuromodulation<br>System with SAINT<br>Technology | |
| Intended Use | The Nexstim Navigated<br>Brain Therapy (NBT)<br>System 2 is intended for the<br>treatment of major<br>depressive disorder in adult<br>patients who have failed to<br>achieve satisfactory<br>improvement from prior | The Magnus<br>Neuromodulation System<br>(MNS) with SAINT<br>Technology is intended for<br>the delivery of SAINT<br>neuromodulation therapy to<br>treat major depressive<br>disorder (MDD) in adult | No difference |
| | Predicate Device | Proposed Device | Notes |
| | antidepressant medication in<br>the current episode. | patients who have failed to<br>achieve satisfactory<br>improvement from prior<br>antidepressant medication in<br>the current episode. | |
| Main System<br>Hardware | • TMS Stimulator<br>• Motor Threshold and<br>Treatment Coil<br>• Tracking System for coil<br>positioning<br>• Evoked response monitor<br>for motor threshold<br>determination<br>• Cooling Unit | • TMS Stimulator<br>• Treatment Coil<br>• Tracking System for coil<br>positioning<br>• Evoked response monitor<br>for motor threshold<br>determination<br>• Cooling Unit<br>• Motor Threshold Coil | The predicate device<br>combines motor threshold<br>measurement and stimulation<br>treatment in one coil while<br>the Magnus System utilizes<br>two separate dedicated<br>purpose coils. |
| Treatment Target<br>Identification | Proprietary software utilizing<br>structural MRI to assist in<br>locating the target area<br>within the L-DLPFC for<br>depression therapy. | Proprietary software utilizing<br>structural and functional<br>MRI data to locate the target<br>area within the L-DLPFC for<br>depression therapy. | The Magnus System uses<br>functional MRI connectivity<br>data in addition to structural<br>MRI data to inform the<br>individual treatment target<br>localization software. Both<br>Magnus and predicate<br>methods aid in localization<br>of appropriate target areas<br>within the L-DLPFC for<br>stimulation and both<br>methods provide more<br>individual specificity than<br>the historically used external<br>anatomical landmarks. |
| Treatment Coil<br>Positioning | Real time visualization of the<br>treatment target relative to<br>the treatment coil position<br>via a tracking system<br>(neuronavigation system) | Real time visualization of the<br>treatment target relative to<br>the treatment coil position<br>via a tracking system<br>(neuronavigation system) | No difference |
| Motor Threshold<br>Determination | Visual inspection of finger<br>movement + EMG | Visual inspection of finger<br>movement + EMG | No difference |
| Treatment Intensity<br>Determination | Motor threshold and<br>electrical field model is used<br>to adjust stimulation<br>intensity | Motor threshold and depth-<br>correction is used to adjust<br>stimulation intensity | Nexstim and Magnus both<br>primarily rely on MT to<br>determine treatment<br>intensity. Nexstim also uses<br>a proprietary electrical field<br>model to orient the rotation<br>of the coil, while Magnus<br>uses a depth correction to<br>adjust for differences in<br>distance between motor<br>cortex and frontal cortex at<br>the target. |
| Stimulation<br>Protocol | 10 Hz and iTBS | iTBS | The Magnus<br>Neuromodulation System<br>with SAINT Technology<br>uses only iTBS; the 10 Hz<br>protocol is not needed for the<br>Magnus System |
| | Predicate Device | Proposed Device | Notes |
| Area of Brain<br>Stimulated | L-DLPFC | L-DLPFC | No difference |
{6}------------------------------------------------
{7}------------------------------------------------
Treatment Protocol Comparison: The Magnus Neuromodulation System with SAINT Technology utilizes the same type of treatment (intermittent theta burst stimulation/iTBS) as the predicate device but with more pulses (90,000 vs. 18,000) delivered over a shorter period of time (five days vs. six weeks) thus allowing for a more rapid response and more flexibility in terms of where treatment can be delivered; the more compressed schedule makes acute inpatient treatment more feasible.
| Table 2: Treatment Protocol Comparison of SAINT Technology iTBS vs. Predicate iTBS | | | |
|------------------------------------------------------------------------------------|---------------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | |
| Treatment<br>Parameter | Nexstim NBT System 2<br>K182700<br>(predicate device) | Magnus<br>Neuromodulation<br>System with SAINT<br>Technology (proposed<br>device) | Notes |
| Stimulation<br>Dose | 600 pulses per<br>session/day, 12,000-<br>18,000 pulses total | 1,800 pulses per<br>session, 18,000 pulses<br>per day, 90,000 pulses<br>total | The number of pulses and timing of<br>sessions are optimized for effectiveness<br>and compressed treatment schedule with<br>no impact on safety. |
| Magnetic Field<br>Intensity | 120% Motor Threshold | 90% Motor Threshold | iTBS administered at amplitudes less than<br>100% of motor threshold may be safer<br>and more focal, and may be more<br>effective than iTBS at amplitudes greater<br>than 100% motor threshold. |
| Pulse<br>Frequency | 50 Hz | 50 Hz | No difference |
| Pulses per<br>burst | 3 | 3 | No difference |
| Burst<br>Frequency | 5 Hz | 5 Hz | No difference |
| Stimulus Train<br>Duration | 2 seconds | 2 seconds | No difference |
| Inter-train<br>Interval | 8 seconds | 8 seconds | No difference |
| Magnetic<br>Pulses per<br>Session | 600 | 1800 | |
| Treatment<br>Session<br>Duration | 3.3 minutes | 10 minutes | The number of pulses and timing of<br>sessions are optimized for effectiveness |
| Treatment<br>Sessions per<br>Day | 1 | 10 | and compressed treatment schedule with<br>no impact on safety. |
| Treatment<br>Sessions per<br>Week | 5 | 50 | |
{8}------------------------------------------------
| Treatment<br>Parameter | Nexstim NBT System 2<br>K182700<br>(predicate device) | Magnus<br>Neuromodulation<br>System with SAINT<br>Technology (proposed<br>device) | Notes |
|--------------------------------|-------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Total<br>Treatment<br>Minutes | 99 | 500 | |
| Total<br>Treatment<br>Sessions | 30 | 50 | |
| Total<br>Treatment<br>Pulses | 18,000 | 90,000 | |
| Treatment<br>Duration | 6 weeks | 5 days | Shorter total treatment duration than the<br>predicate is possible because of an<br>optimized and compressed treatment<br>schedule. |
Summary of Performance Data:
Sterilization and Shelf Life: The Magnus Neuromodulation System with SAINT Technology is not provided as a sterile device and is not intended for sterilization by the user. The shelf life and/or useful life of the hardware components are as indicated in their respective labeling and instructions for use. Software components will have updates managed as appropriate by Magnus Medical or its suppliers and therefore do not have a defined useful life at this time. Biocompatibility: Patient contact materials that are part of the Magnus Neuromodulation System were tested and are compliant with ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. Software for the Magnus Neuromodulation System with SAINT Software: Technology was designed and developed in accordance with current FDA guidance and industry standards including IEC 62304, Medical Device Software – Software lifecycle processes and ISO 14971, Application of risk management to medical devices. Hardware: The hardware components of the Magnus Neuromodulation System have undergone all applicable electrical safety, electromagnetic compatibility, performance, and usability testing required.
{9}------------------------------------------------
Clinical: Clinical testing was performed to provide assurance of safety and effectiveness of the Magnus Neuromodulation System with SAINT Technology. The system was evaluated in four clinical studies, three open-label and one randomized and sham controlled. Data from these studies are summarized in the following tables:
| Title of Study | Number of<br>Participants | Summary of Results |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| High-dose Spaced theta-burst TMS<br>as a Rapid-acting Antidepressant in<br>Highly Refractory Depression1 | 6 | 5/6 (83.3%) responded (≥50% decrease in Hamilton<br>Depression Rating Scale (HDRS-17) and 4/6 remitted<br>(HDRS<7) with no serious adverse events. |
| Stanford Accelerated Intelligent<br>Neuromodulation Therapy for<br>Treatment Resistant Depression2 | 21 | 19/21 (90.5%) of participants met remission criteria (score<br><11 on the Montgomery-Asberg Depression Rating Scale<br>[MADRS]) with no serious adverse events or negative<br>cognitive side effects. |
| Stanford Neuromodulation Therapy<br>(SNT): A Double-Blind<br>Randomized Controlled Trial3 | 29 (14 active,<br>15 sham) | 12/14 (85.7%) in the active treatment group met response<br>criteria (≥50% reduction in MADRS) and 11/14 (78.6%) met<br>remission criteria (MADRS ≤10) vs 4/15 (27%) in the sham<br>group met response criteria and 2/15 (13.3%) remitted. The<br>study was terminated early based on clear superiority of active<br>treatment vs. sham at the planned interim analysis. No serious<br>adverse events were reported. |
| An Open Label Pilot Trial to Assess<br>the Feasibility of Using the Magnus<br>Neuromodulation System (MNS)<br>with Magnus Intelligent<br>Neuromodulation Therapy (MINT4)<br>as a Maintenance Treatment for<br>Depression | 14 | 13/14 (92.8%) of subjects responded (≥50% reduction in<br>MADRS) and 11/14 (78.6%) met remission criteria (MADRS<br>≤10) after the initial 5 days of treatment. No serious adverse<br>events have been reported. |
Table 3: Summary of Clinical Studies Performed with SAINT Technology
Data from four clinical trials using SAINT Technology have demonstrated similar clinical outcomes with equivalent results as compared to the predicate device.
Demographic and effectiveness data are presented separately for the each of the four SAINT studies and compared to the published iTBS data5 in Table 4. Patient demographics in terms of severity of depression and resistance to depression are comparable for the SAINT and THREE-D
3 Cole EJ, Phillips AL, Bentzley BS, et al. Am J Psychiatry. 2021 Oct 29:appiajp202120101429. doi: 10.1176/appi.ajp.2021.20101429. Epub ahead of print. PMID: 34711062.
<sup>4</sup> Williams NR, Sudheimer KD, Bentzley BS, et al., High-dose spaced theta-burst TMS as a rapid-acting antidepressant in highly refractory depression. Brain 2018;141:1-5.
<sup>2</sup> Cole EJ, Stimpson KH, Bentzley BS, at al., Stanford Accelerated Intelligent Neuromodulation Therapy for Treatment Resistant Depression. Am J Psychiatry 2020;177:716-726.
<sup>4</sup> Note that MINT, SNT, and SAINT all refer to the same combination of target identification and treatment delivery that is SAINT Technology.
<sup>5</sup> Blumberger DM et al. Effectiveness of theta burst vs. high-frequency repeitive transcranial magnetic stimulation in patients with depression (THREE-D): a randomized non-inferiority trial. Lancet 2018;391:1683-92.
{10}------------------------------------------------
populations and the outcomes for SAINT trials, both open-label and randomized controlled trials, demonstrate equivalent effectiveness as compared to the results from the THREE-D trial without introduction of any new safety concerns.
| SAINT Technology Compared to the Predicate iTBS (THREE-D) | Study #1 | Study #2 | Study #3 | Study #4 | THREE-D* |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|-------------------------------------------------------|
| # of Subjects | 6 | 22 (21 per protocol) | 29<br>(15 active/14 sham) | 14 | 209 active |
| Baseline Characteristics | | | | | |
| Age (mean) | 56 | 45 | 49/52 | 52 | 42 |
| Age (range) | 38-69 | 19-78 | 27-73 | 23-82 | 18-65 |
| Gender (% female) | 67% | 57% | 34% | 50% | 59% |
| Maudsley Staging<br>Method Score (mean) | 14 | 10 | 9/9 | 9.5 | 6.3 |
| # having prior TMS | 6 | 7 (only 1 previously<br>remitted) | 0 | - | 0 |
| # having prior ECT | 6 | 0 | 0 | - | 16 |
| # having prior VNS | 1 | 0 | 0 | - | 0 |
| MADRS at baseline<br>(mean) | 40.3 | 34.86 | 31/35 | 32 | - |
| HDRS at baseline<br>(mean) | 16 (6-item)<br>28.8 (17-item) | 13.9 (6-item)<br>25.9 (17-item) | 14/15 (6-item)<br>24/26 (17-item) | 12.5 (6-item) | 23.7 (17-item) |
| Type of Study | Single arm,<br>Open label | Single arm<br>Open label | Randomized,<br>blinded | Single arm<br>Open label | Randomized,<br>Open label |
| Treatment Arms | Single, active | Single, active | 2 arms - active<br>(n=14) vs.<br>sham (n=15) | Single, active | 2 arms - 10 Hz<br>rTMS (n=205)<br>vs. iTBS<br>(n=209) |
| Treatment | 5 days SAINT<br>iTBS | 5 days SAINT<br>iTBS | 5 days SAINT<br>iTBS vs. 5 days<br>sham SAINT | 5 days SAINT<br>iTBS | 4-6 weeks 10<br>Hz rTMS vs. 6<br>weeks iTBS |
| Outcome | | | | | |
| Response<br>(MADRS reduction<br>≥50%) | 83.3%<br>(assessed at end<br>of 5 days of<br>treatment) | 90.5%<br>(assessed at end<br>of 5 days of<br>treatment) | 85.7% active<br>vs. 26.7% sham<br>(assessed<br>during the<br>month after<br>treatment) | 92.8%<br>(assessed<br>during the<br>week after<br>treatment) | - |
| Remission (MADRS<br>≤10) | 83.3%<br>(assessed at end<br>of 5 days of<br>treatment) | 90.5%<br>(assessed at end<br>of 5 days of<br>treatment for<br>the 21 per<br>protocol<br>subjects) | 78.6% active<br>vs. 13.3% sham<br>(assessed<br>during the<br>month after<br>treatment) | 78.6%<br>(assessed<br>during the<br>week after<br>treatment) | - |
| Response (HDRS<br>reduction ≥50%) | 83.3% | 86.4 %<br>(HDRS-17) | - | 71.4% (HDRS-<br>6) | 49% (HDRS-<br>17) |
| | Study #1 | Study #2 | Study #3 | Study #4 | THREE-D* |
| | of 5 days of<br>treatment) | 81.8% (HDRS-<br>6) | | | |
| | 66.7% | 77.3% (HDRS-<br>17) | | | |
| Remission (HDRS-17<br>≤7 or HDRS-6 ≤4) | (assessed at end<br>of 5 days of<br>treatment) | 81.8% (HDRS-<br>6) | | 57.1% (HDRS-<br>6) | 32% (HDRS-<br>17) |
| Serious Adverse<br>Events | 0 | 0 | 0 | 0 | 3 (1%) |
| Comments | The single non-<br>responding<br>subject was<br>later<br>determined to<br>have a primary<br>diagnosis of<br>OCD | 23 subjects<br>recruited, 1<br>ineligible, 1<br>withdrew on<br>treatment day 1<br>because of<br>anxiety | The trial was<br>terminated at<br>the planned<br>interim analysis<br>due to the<br>superiority of<br>the active<br>treatment | This trial is<br>ongoing | |
Table 4: Summary of Demographic and Effectiveness Data for Clinical Studies Performed with SAINT Technology Compared to the Predicate iTBS (THREE-D)
{11}------------------------------------------------
*The predicate device, Nexstim Navigated Brain Therapy (NBT) System 2, relied on the THREE-D Clinical Study data for clearance, via its own predicate (K173620).
Note: The clinical performance data for the Magnus Neuromodulation System (MNS) with SAINT Technology were obtained from a total of 70 patients enrolled in four clinical trials (one randomized double-blind, sham-controlled study and three open-label studies) conducted in close geographical proximity. As a result of studies being performed at a single site, generalizability to the broader United States population has not been evaluated.
Note: The Magnus Neuromodulation System (MNS) with SAINT Technology has been evaluated at four weeks post treatment in all four clinical trials. Effectiveness has not been established beyond the timepoints evaluated in the four clinical studies.
## Note: Blinding Assessment Study #3
The adequacy of blinding utilized in the investigation was assessed by asking participants to guess their treatment allocation and to report their confidence in their guess (on a scale of 1 to 5) on the last day of treatment. This guess and confidence level were translated to a guess metric that ranged from 0 (participant had full confidence that he or she received the sham treatment) to 1 (participant had full confidence he or she received the active treatment). Twenty-three (23) participants provided guesses as to which treatment they received, and 19 indicated their confidence in their guess. One-way t tests indicated no significant differences from chance (chance guess metric=0.50) in the sham (mean guess metric=0.39, p=0.56) and active (mean guess metric=0.43, p=0.52) treatment groups. Because not all participants indicated their confidence in their guess, binomial tests were also used to determine whether the number of correct guesses exceeded chance. Binomial tests indicated no significant differences from chance in proportion of correct guesses in the sham (6 of 10 correct, p=0.38) and active (7 of 13 correct, p=0.50) treatment groups. Finally, linear regression analysis detected no relationship between the guess metric and the change in depression severity as indicated by magnitude of proportional change in MADRS scores (=0.11. p=0.66). Since only 19 of 29 participants completed both parts of the blinding assessment. uncertainty remains as to the adequacy of the blinding for the remainder of the study population.
Adverse events from the Study #3 active treatment arm and the open label study (Study #4) of SAINT are listed in Table 5 below along with the adverse events reported in the THREE-D iTBS Study. (Note that Study #1 and Study #2 are not shown in the table, because these universitybased studies did not formally tabulate non-serious adverse events; zero serious adverse events were observed in either study.) All the adverse events reported for SAINT are similar to those
{12}------------------------------------------------
reported for THREE-D iTBS in type and incidence with the exception that fatigue is reported more frequently with SAINT, likely because of the increased time spent in the clinic for the five days of SAINT treatment. Notably, rates of fatigue were the same for active and sham SAINT groups.
| Event | Number participants reporting event (%)6 | | |
|--------------------------------------|------------------------------------------|-----------------------|----------------------|
| | SAINT n=14 (active arm<br>from Study #3) | SAINT n=14 (Study #4) | iTBS (THREE-D) n=209 |
| Headache | 8 (57%) | 6 (43%) | 136 (65%) |
| Nausea | - | - | 14 (7%) |
| Dizziness | - | 1 (7%) | 18 (9%) |
| Unrelated medical<br>problem | - | 5 (36%) | 46 (22%) |
| Fatigue | 8 (57%) | 5 (36%) | 16 (8%) |
| Insomnia | - | 5 (36%) | 10 (5%) |
| Anxiety or agitation | 4 (29%) | 2 (14%) | 9 (4%) |
| Back or neck<br>pain/discomfort | 7 (50%) | 3 (21%) | 6 (3%) |
| Unrelated accidents | - | 1 (7%) | 3 (1%) |
| Vomiting | - | - | 1 (<1%) |
| Tinnitus | - | 2 (14%) | 3 (1%) |
| Migraine aura | - | - | 4 (2%) |
| Abnormal sensations | - | - | 4 (2%) |
| Pain/discomfort at<br>treatment site | 5 (36%) | 4 (29%) | - |
| Other head/neck area<br>pain | 3 (21%) | 3 (21%) | - |
| Any serious adverse<br>events | 0 | 0 | 3 (1%) |
Table 5: Adverse Events: SAINT compared to THREE-D iTBS
In summary, the clinical study results indicate that iTBS delivered using the Magnus Neuromodulation System with SAINT Technology is safe, effective, and substantially equivalent to iTBS delivered using the predicate device, the Nexstim NBT System 2 (K182700), and without introduction of any new risks or safety concerns.
Conclusion: The Magnus Neuromodulation System (MNS) with SAINT Technology has equivalent hardware components, the same general principle for target identification, the same intended use, and the same technological features as the predicate device. The Magnus Neuromodulation System (MNS) with SAINT Technology does not raise any new issues of safety and effectiveness and is substantially equivalent to the predicate device.
<sup>6</sup> Shown here are outcomes from the active SAINT arm (Study #3), active SAINT (Study #4), and the iTBS arm from the THREE-D study.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.