← Product Code [FGB](/productcode/FGB) · K220159

# Flexible Ureterorenoscope (Model: US31D-12-EU), Flexible Ureterorenoscope (Model: US31D-12-US) (K220159)

_Shanghai AnQing Medical Instrument Co., Ltd. · FGB · Aug 26, 2022 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K220159

## Device Facts

- **Applicant:** Shanghai AnQing Medical Instrument Co., Ltd.
- **Product Code:** [FGB](/productcode/FGB.md)
- **Decision Date:** Aug 26, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Flexible Ureterorenoscope is intended to be used to visualize organs, cavities, and canals in the urinary tract (urethra, bladder, ureter, calyces, and renal papillae) via transurethral access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.

## Device Story

Single-use flexible ureterorenoscope; consists of handle, insertion section, and connector. Handle features deflection lever, lever lock, and Luer port for irrigation/accessories. Distal tip contains CMOS sensor, LED illumination, and working channel. Device connects to external Video Processor (Model EOS-H-01) for power and signal processing. User operates deflection lever to maneuver tip. Light from distal LEDs reflects off tissue; CMOS sensor captures image; RGB signals transmitted to processor for display on monitor. Used in healthcare facilities by clinicians for visualization and minimally invasive procedures in the urinary tract. Benefits include real-time visualization for diagnosis and therapeutic intervention.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing: electrical safety (IEC 60601-1, 60601-2-18, 60601-1-2), photobiological safety (IEC 62471), mechanical/optical performance (ISO 8600), biocompatibility (ISO 10993), sterilization validation (ISO 11135), and package integrity (ISO 11607, ASTM standards).

## Technological Characteristics

Single-use flexible endoscope. Materials: Pebax, polyamide, stainless steel, epoxy, glass, polycarbonate, PTFE, PET. Sensing: CMOS sensor. Illumination: Internal LED. Energy: Electricity. Deflection: ≥275° up/down. Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: Wired to external video processor.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Ureterorenoscope System ([K201293](/device/K201293.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 26, 2022

Shanghai AnQing Medical Instrument Co., Ltd. Shuwen Fan RA Manager 3 & 4 Floor, No.2 Building, 366 Huiging Rd, East Zhangjiang High-Tech Park Shanghai, 201201 China

Re: K220159

Trade/Device Name: Flexible Ureterorenoscope Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FGB Dated: July 27, 2022 Received: July 27, 2022

Dear Shuwen Fan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Reginald K. Avery, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220159

Device Name Flexible Ureterorenoscope

#### Indications for Use (Describe)

The Flexible Ureterorenoscope is intended to be used to visualize organs, cavities, and canals in the urinary tract (urethra, bladder, ureter, calyces, and renal papillae) via transurethral access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.

Type of Use (Select one or both, as applicable)

| <span style="font-size:12pt"> <span style="font-family:Symbol">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-----------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:12pt">     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span>                                      |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains a logo for a company called "Innovex Anqing Medical". The logo features the word "Innovex" in a bold, sans-serif font, with the word "Anqing Medical" appearing below it in a slightly smaller font. To the left of the text is a purple square with a white, stylized leaf design inside. A curved line extends from the top of the square over the word "Innovex".

#### 510(k) Summary

| Date Prepared: | Aug. 18th, 2022                                                                                                                                          |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:  | Shanghai AnQing Medical Instrument Co., Ltd.<br>3 & 4 Floor, No.2 Building, 366 Huiqing Rd, East<br>Zhangjiang High-Tech Park, 201201<br>Shanghai, China |

Contact Person:

Shuwen Fan RA Manager Shanghai AnQing Medical Instrument Co., Ltd. Tel: +86-21-61117375 Fax: +86-21-61117374 ra_dept@anqing-sh.com

#### Identification of the Device:

| Proprietary/Trade Name: | Flexible Ureterorenoscope                    |
|-------------------------|----------------------------------------------|
| Model:                  | US31D-12-EU, US31D-12-US                     |
| Common or Usual Name:   | Ureteroscope and accessories, flexible/rigid |
| Classification Name:    | Endoscope and accessories                    |
| Regulatory Number:      | 21 CFR Part 876.1500                         |
| Product Code:           | FGB                                          |
| Device Class:           | Class II                                     |
| Review Panel:           | Gastroenterology/Urology                     |

# Identification of the Legally Marketed Predicate Device:

| Trade Name:              | Ureterorenoscope System                                                                            |
|--------------------------|----------------------------------------------------------------------------------------------------|
| Classification Name:     | Endoscope and accessories                                                                          |
| Regulatory Number:       | 21 CFR Part 876.1500                                                                               |
| Product Code:            | FGB                                                                                                |
| Device Class:            | Class II                                                                                           |
| Review Panel:            | Gastroenterology/Urology                                                                           |
| Submitter/510(k) Holder: | Shanghai AnQing Medical Instrument Co., Ltd                                                        |
| Clearance:               | K201293 (cleared July 24, 2020)<br>This predicate has not been subject to a design-related recall. |

# Device Description:

The Flexible Ureterorenoscope (Model: US31D-12-EU, US31D-12-US) is an endoscope which is used with the Video Processor (Model: EOS-H-01) (FDA cleared

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Innovex Anqing Medical. The logo is purple and white. On the left side of the logo is a purple square with a white leaf design. To the right of the square is the word "INNOVEX" in a sans-serif font, with the words "Anqing Medical" below it.

#K211169) produced by AnQing for providing endoscopic imaging of the ureter and the renal pelvis for the purpose of diagnosis and treatment. The 2 proposed models are identical except the deflection versions, which is opposite from each other (EU version or US version).

The Flexible Ureterorenoscope is a single-use endoscope, which consists of a Handle, an Insertion Section, and an Endoscope Connector. The handle includes a deflection lever, a lever lock, a push button for picture taking/video recording and a Luer port for insertion of accessory devices and irrigation to the working channel. The insertion section contains one working channel and wiring to transmit the image signals to the Video Processor. The distal bending section of the insertion is controlled by the user via the deflection lever on the handle. The distal end of the insertion section contains a CMOS sensor for capturing image and transmitting it to the Video Processor, LEDs for illumination, and the distal opening of the working channel. The endoscope connector connects the endoscope handle to the video processor, which provides power and processes video signals from the endoscope.

Mechanism of action:

The light emitted by the LED cold light source at the distal tip of the disposable Flexible Ureterorenoscope is irradiated into the body cavity, and the light reflected from the cavity enters the optical system and is captured by the CMOS image sensor. The CMOS acquisition image is controlled by the CMOS drive circuit, and the RGB video signal is output to the Video Processor via the VI circuit. The Video Processor receives video signals from the endoscope, processes the video signals, and outputs the processed video signal to the attached monitor. The video processor also controls the brightness of the LEDs on the endoscope.

Flexible Ureterorenoscope has the following physical and performance characteristics:

- Maneuverable tip controlled by the user
- Flexible insertion cord
- Camera and LED light source at the distal tip
- Sterilized by Ethylene Oxide
- For single use

# Indications for Use:

The Flexible Ureterorenoscope is intended to be used to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.

# Comparison with Predicate Device:

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a logo for a company called "Innovex Anqing Medical". The logo has a purple graphic on the left side, which looks like a stylized leaf or wing. To the right of the graphic, the word "Innovex" is written in a bold, sans-serif font, with "Anqing Medical" written in a smaller font below it. The overall design is clean and modern.

The Flexible Ureterorenoscope and its predicate device, the Shanghai AnQing Medical Ureterorenoscope System (K201293), have the same intended use, sterilization Methods and similar physical characteristics, optical characteristics.

# Substantial Equivalence:

The Flexible Ureterorenoscope employs the same fundamental scientific technology as its predicate devices, as below table:

|                                                     | Subject Device                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device,<br>K201293                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison            |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Indications For Use                                 |                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                       |
| Indications<br>For Use:                             | The Flexible<br>Ureterorenoscope is<br>intended to be used to<br>visualize organs, cavities<br>and canals in the urinary<br>tract (urethra, bladder,<br>ureter, calyces and renal<br>papillae) via transurethral<br>access routes. It can also be<br>used in conjunction with<br>endoscopic accessories to<br>perform various diagnostic<br>and therapeutic procedures<br>in the urinary tract. | The Ureterorenoscope<br>System consists of a<br>sterile single-use<br>Flexible Ureteroscope<br>to be introduced within<br>the urinary tract and<br>video processor for<br>clinical image<br>processing. The device<br>is indicated for<br>endoscopic<br>examination in the<br>urinary tract and can be<br>used to examine the<br>interior of the kidney,<br>and using additional<br>accessories, to perform<br>various diagnostic and<br>therapeutic<br>procedures. | Same.<br>See Note2.   |
|                                                     | Physical Characteristics                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                       |
| Type<br>of<br>Scope                                 | Flexible                                                                                                                                                                                                                                                                                                                                                                                        | Flexible                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                  |
| Distal<br>end<br>outer<br>diameter                  | ≤3.2mm (9.6Fr)                                                                                                                                                                                                                                                                                                                                                                                  | ≤3.1mm (9.3 Fr)                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Similar               |
| Insertion<br>Section<br>length                      | 670mm                                                                                                                                                                                                                                                                                                                                                                                           | 650mm                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Similar               |
|                                                     | Subject Device                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device, K201293                                                                                                                                                                                                                                                                                                                                                                                                                                           | Comparison            |
| Deflection                                          | ≥275°up/275°down                                                                                                                                                                                                                                                                                                                                                                                | ≥275°up/275°down                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                  |
|                                                     | Optical Characteristics                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                       |
| Type of Imager                                      | CMOS                                                                                                                                                                                                                                                                                                                                                                                            | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                  |
| Direction of View                                   | forward                                                                                                                                                                                                                                                                                                                                                                                         | forward                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                  |
| Field of View                                       | 110°                                                                                                                                                                                                                                                                                                                                                                                            | 110°                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                  |
| Depth of Field                                      | 3mm~100mm                                                                                                                                                                                                                                                                                                                                                                                       | 3mm~100mm                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                  |
| Light Source                                        | Internal LED                                                                                                                                                                                                                                                                                                                                                                                    | Internal LED                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                  |
|                                                     | Patient Contacting Materials                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                       |
| General material type of main patient- contact part | Pebax, polyamide, Stainless Steel, Epoxy glue, glass, Polycarbonate, PTFE, PET                                                                                                                                                                                                                                                                                                                  | Pebax, PTFE, LCP, TPU, Fluoro elastomers, Epoxy glue, glass                                                                                                                                                                                                                                                                                                                                                                                                         | Different See Note 1. |
|                                                     | Sterilization Methods                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                       |
| Number of Users                                     | Single-Use                                                                                                                                                                                                                                                                                                                                                                                      | Single-Use                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                  |
| Sterilization                                       | EO Sterilized, SAL 10-6                                                                                                                                                                                                                                                                                                                                                                         | EO Sterilized, SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                  |
|                                                     | Technological Characteristics                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                       |
| Environmen t of use                                 | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                    | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                  |
| Energy source                                       | Electricity                                                                                                                                                                                                                                                                                                                                                                                     | Electricity                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                  |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Innovex Anqing Medical. The logo is purple and white. On the left side of the logo is a purple graphic that looks like a leaf. To the right of the graphic is the word "INNOVEX" in a sans-serif font. Below "INNOVEX" is the text "Anqing Medical" in a larger, bolder font.

#### Note :

1. The subject and predicate devices have the same fundamental technology, direction of view, depth of field, light source, type of imager, number of uses and sterilization method. The subject ureterorenoscope differs from the predicate in insertion section length, distal end outer diameter and patient-contacting materials. These differences do not raise different questions of safety and effectiveness as compared to the predicate, and can be evaluated through performance testing.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Innovex Anqing Medical. The logo features the word "INNOVEX" in a bold, sans-serif font, with the word "Anqing MEDICAL" in a smaller, bolder font below it. To the left of the text is a stylized graphic of a leaf or wing in a purple color. The logo has a clean and professional design.

| Subject Device                                                                                                                                                                                                                                                                                                                                                                                                          | Predicate Device, K201293 | Comparison |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|------------|
| 2. The indications for use statement for the subject ureterorenoscope is very similar to<br>that of the predicate device. A slightly different wording is chosen and specific<br>anatomical locations within the urinary tract are listed. The differences do not alter the<br>intended use of the device nor do they raise different questions of safety and<br>effectiveness of the device relative to the predicate. |                           |            |

# Summary of Testing:

#### Summary of Non-Clinical Tests:

#### Electrical Safety and Electromagnetic Compatibility Summary

The electrical safety and EMC data included in the submission is in compliance with the following FDA recognized standards:

•ANSI/AAMI ES: 60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012

- ・IEC 60601-2-18:2009
·IEC 60601-1-2:2014

# Bench Testing Summary

# Photobiological safety

The LEDs in submitted Ureterorenoscope were tested according to the following FDA recognized standards:

· IEC 62471:2006 Medical electrical equipment, Photobiological safety of lamps and lamp systems

# Mechanical and Optical Performance

The Flexible Ureterorenoscope was designed to comply with applicable parts of ISO 8600. Optical measurements were performed according to applicable part of ISO 8600 standard.

Mechanical characteristics were tested and include leakage tightness, bending. deflection endurance, and tensile strength testing.

In addition, comparative testing related to image quality parameters including Color performance (color reproduction), optical performance (resolution, depth of field and image intensity uniformity, distortion), and dynamic range test was performed for submitted Flexible Ureterorenoscope and the predicate device to support substantial equivalence.

# Biocompatibility Summarv

The biocompatibility evaluation for the patient contacting components of the Flexible Ureterorenoscope was performed according to ISO 10993-1 and FDA Guidance. The

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Innovex Anqing Medical. The logo features the text "INNOVEX Anqing Medical" in a stylized font. The word "INNOVEX" is in a larger, bolder font than "Anqing Medical". Above the text is a graphic element that resembles a stylized leaf or wing shape. The color scheme of the logo is primarily purple.

following tests were conducted based contact category of "Surface-Breached or compromised surface" with a contact duration of "Limited (< 24 hours):

- · Cytotoxicity: ISO 10993-5:2009/(R) 2014
- · Sensitization, Intracutaneous reactivity/irritation: ISO 10993-10:2010
- · Material-mediated pyrogenicity: ISO 10993-11:2017
- · Acute systemic toxicity: ISO 10993-11:2017

# Sterilization and shelf life testing

The sterilization method has been validated to ISO 11135:2014, which has thereby determined the routine control and monitoring parameters.

EO/ECH residual test was performed according to ISO 10993-7:2008.

The shelf life (3 years) of the Flexible Ureterorenoscope is determined based on stability study which includes ageing test according to ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier.

# Package Validation

Package validation was conducted according to ISO 11607-1:2019 and ISO 11607-2:2019, and visual inspection (ASTM F1886), seal strength (F88/F88M-15), dye penetration (ASTM F1929-15), and bubble emission (ASTM D3078). Transport and shipping testing as per ASTM D4169-16.

# Summary of Clinical Tests:

The subject of this premarket submission did not require clinical studies to support substantial equivalence.

# Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the subject device, the Flexible Ureterorenoscope is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K220159](https://fda.innolitics.com/device/K220159)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
