← Product Code [NCX](/productcode/NCX) · K220008

# Solex 7 Intravascular Heat Exchange Catheter, Cool Line Intravascular Heat Exchange Catheter, ICY Intravascular Heat Exchange Catheter, Quattro Intravascular Heat Exchange Catheter, Thermogard HQ Start-Up Kit, Thermogard HQ Start-Up Kit EX, Thermogard HQ Console (K220008)

_Zoll Circulation, Inc. · NCX · Jun 2, 2022 · Cardiovascular · SESU_

**Canonical URL:** https://fda.innolitics.com/device/K220008

## Device Facts

- **Applicant:** Zoll Circulation, Inc.
- **Product Code:** [NCX](/productcode/NCX.md)
- **Decision Date:** Jun 2, 2022
- **Decision:** SESU
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5900
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K220008 · Jun 2, 2022 | Solex 7 Intravascular Heat Exchange Catheter, Cool Line Intravascular Heat Exchange Catheter, ICY Intravascular Heat Exchange Catheter, Quattro Intravascular Heat Exchange Catheter, Thermogard HQ Start-Up Kit, Thermogard HQ Start-Up Kit EX, Thermogard HQ Console | Zoll Circulation, Inc. | Randomized controlled trial (clinical trial data) | The document references a randomized controlled trial of 296 patients to establish safety and effectiveness profiles and to support mandatory labeling warnings regarding fever reduction in patients with subarachnoid hemorrhage or primary traumatic brain injury. | Fever reduction; Clinical trial; Mortality results; Safety warning |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Randomized controlled trial of endovascular cooling; Randomized controlled trial | Patients with fever associated with subarachnoid hemorrhage (SAH), primary traumatic brain injury (PTBI), cerebral infarction (CI), and intracerebral hemorrhage (ICH); Sample Size: 296 patients | Standard of care | Mortality |

## Indications for Use

The Solex 7 Intravascular Heat Exchange Catheter connected to the Coolgard/Thermogard Thermal Regulation System is indicated for use: · In cardiac surgery patients to achieve and/or maintain normothermia during surgery and recovery/intensive care. (Maximum use period: 4 days) · To induce, maintain and reverse mild hypothermia in neurosurgery and recovery/intensive care. (Maximum use period: 4 days) · In fever reduction, as an adjunct to other antipyretic therapy, in adult patients with cerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated. (Maximum use period: 7 days) The Cool Line Catheter Model CL-2295A, when used with the ZOLL Thermal Regulation System, is indicated for use in fever reduction, as an adjunct to antipyretic therapy. in adult patients with cerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated. The ZOLL ICY Intravascular Heat Exchange Catheter Model IC-3893A, connected to the ZOLL Coolgard/Thermogard Thermal Regulation System, is indicated for use: • In cardiac surgery adult patients to achieve and/or maintain normothermia during surgery and recovery/intensive care, and - · To induce, maintain and reverse mild hypothermia in neurosurgery adult patients in surgery and recovery/intensive care. The ZOLL Quattro Catheter Model IC-4593, connected to a ZOLL Thermal Regulation System, is indicated for use: • In cardiac surgery adult patients to achieve and/or maintain normothermia during surgery and recovery/intensive care, and - · To induce, maintain and reverse mild hypothermia in neurosurgery adult patients in surgery and recovery/intensive care.

## Device Story

IVTM system comprises Thermogard HQ console, heat exchange catheter, and sterile tubing kit. Console circulates saline through catheter balloons placed in central venous circulation to heat/cool patient via feedback control. Physician sets target temperature; system monitors patient temperature via YSI 400 series thermistor and adjusts saline temperature in glycol bath. Used in OR/ICU by clinicians. Provides automated thermal regulation to achieve/maintain normothermia or induce/reverse hypothermia. Benefits include precise temperature management for neurocritical/cardiac patients. Subject submission updates console and tubing kit components; catheters remain unchanged from predicate.

## Clinical Evidence

No clinical data provided. Substantial equivalence demonstrated via bench testing, electrical safety, electromagnetic compatibility, and software verification.

## Technological Characteristics

System: Hypothermia/Thermal Regulating System (21 CFR 870.5900). Console: 115-230V, glycol/water bath, 0-42°C range. Catheters: Polyurethane/PET, heparin-coated, 9.3 Fr, EO sterilized. Connectivity: USB, Wi-Fi, EMR output. Software: IEC 62304 compliant. Standards: ISO 14971, IEC 60601-1, ISO 10993.

## Regulatory Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

## Predicate Devices

- Solex 7 Intravascular Heat Exchange Catheter, Cool Line Intravascular Heat Exchange Catheter, ICY Intravascular Heat Exchange Catheter, Quattro Intravascular Heat Exchange Catheter, Start-Up Kit, Coolgard 3000 Console, Thermogard XP Console ([K213031](/device/K213031.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

June 2, 2022

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left and features a stylized human figure. The U.S. Food & Drug Administration logo is on the right and features the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

ZOLL Circulation, Inc. Elizabeth Haines Senior Director, Regulatory Affairs 2000 Ringwood Avenue San Jose, California 95131

# Re: K220008

Trade/Device Name: Solex 7 Intravascular Heat Exchange Catheter, Cool Line Intravascular Heat Exchange Catheter, ICY Intravascular Heat Exchange Catheter, Quattro Intravascular Heat Exchange Catheter, Thermogard HQ Start-Up Kit, Thermogard HQ Start-Up Kit EX, Thermogard HQ Console, Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II

Product Code: NCX Dated: April 29, 2022 Received: May 2, 2022

Dear Elizabeth Haines:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

The OHTS: Office of Neurological and Physical Medicine Devices has determined that there is a reasonable likelihood that the Solex 7 Intravascular Heat Exchange Catheter will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the box warning immediately following the indications for use statement of the device's labeling:

{1}------------------------------------------------

# Warning - Fever Reduction

The safety of this device has not been demonstrated for fever reduction in patients presenting with subarachnoid hemorrhage or primary traumatic brain injury. A randomized controlled trial of endovascular cooling in patients with fever associated with subarachnoid hemorrhage and primary traumatic brain injury has shown increased mortality as compared to patients receiving standard of care.

The OHT5: Office of Neurological and Physical Medicine Devices has determined that there is a reasonable likelihood that the Cool Line Intravascular Heat Exchange Catheter will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the box warning immediately following the indications for use statement of the device's labeling:

# Warning - Fever Reduction

The safety of this device has not been demonstrated for fever reduction in patients presenting with subarachnoid hemorrhage or primary traumatic brain injury. The safety and effectiveness of this device was examined in a randomized controlled trial of 296 patients. The mortality results reported in this trial, for the four patient cohorts enrolled, are presented in the table below (CI - cerebral infarction, ICH - intracerebral hemorrhage, PTBI - primary traumatic brain injury, SAH subarachnoid hemorrhage).

|      |    | Cool Line |      |   |    | Control |      |  |  |
|------|----|-----------|------|---|----|---------|------|--|--|
|      | n  | N         | %    | n | N  | %       | p*   |  |  |
| CI   | 3  | 16        | 18.8 | 3 | 14 | 21.4    | 0.74 |  |  |
| ICH  | 8  | 33        | 24.2 | 7 | 27 | 25.9    | 1.00 |  |  |
| PTBI | 10 | 44        | 22.7 | 4 | 38 | 10.5    | 0.24 |  |  |
| SAH  | 13 | 61        | 21.3 | 7 | 63 | 11.1    | 0.15 |  |  |

*Fischer's exact test

For more details on the clinical trial results, refer to the Physician's Manual - "Normothermia for the Neurocritically Ill stroke patient."

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Christopher M. Loftus M.D. Acting Director OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

## 510(k) Number (if known) K220008

#### Device Name

Solex 7 Intravascular Heat Exchange Catheter, Cool Line Intravascular Heat Exchange Catheter, ICY Intravascular Heat Exchange Catheter, Quattro Intravascular Heat Exchange Catheter, Thermogard HQ Start-Up Kit, Thermogard HQ Start-Up Kit EX, Thermogard HO Console

#### Indications for Use (Describe)

The Solex 7 Intravascular Heat Exchange Catheter connected to the Coolgard/Thermogard Thermal Regulation System is indicated for use:

· In cardiac surgery patients to achieve and/or maintain normothermia during surgery and recovery/intensive care. (Maximum use period: 4 days)

· To induce, maintain and reverse mild hypothermia in neurosurgery and recovery/intensive care.

(Maximum use period: 4 days)

· In fever reduction, as an adjunct to other antipyretic therapy, in adult patients with cerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated. (Maximum use period: 7 days)

#### Warning - Fever Reduction

The safety of this device has not been demonstrated for fever reduction in patients presenting with subarachnoid hemorrhage or primary traumatic brain injury. A randomized controlled trial of endovascular cooling in patients with fever associated with subarachnoid hemorrhage and primary traumatic brain injury has shown increased mortality as compared to patients receiving standard of care.

The Cool Line Catheter Model CL-2295A, when used with the ZOLL Thermal Regulation System, is indicated for use in fever reduction, as an adjunct to antipyretic therapy. in adult patients with cerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.

|      |    |           | WARNING - Fever Reduction<br>The safety of this device has not been demonstrated for fever<br>reduction in patients presenting with subarachnoid hemorrhage<br>or primary traumatic brain injury. The safety and effectiveness of<br>this device was examined in a randomized controlled trial of 296<br>patients. The mortality results reported in this trial, for the four<br>patient cohorts enrolled, are presented in the table below (CI -<br>cerebral infarction, ICH - intracerebral hemorrhage, PTBI - pri-<br>mary traumatic brain injury, SAH - subarachnoid hemorrhage). |   |         |      |      |
|------|----|-----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---------|------|------|
|      |    |           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |   |         |      |      |
|      |    | Cool Line |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |   | Control |      |      |
|      | n  | N         | %                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | n | N       | %    | D*   |
| Cl   | ਤੋ | 16        | 18.8                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | ਤ | 14      | 21.4 | 0.74 |
| ICH  | 8  | 33        | 24.2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 7 | 27      | 25.9 | 1.00 |
| РТВІ | 10 | 44        | 22.7                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 4 | 38      | 10.5 | 0.24 |

For more details on the clinical trial results, refer to the Physician's Manual - "Normothermia for the Neuro- critically III stroke patient"

{4}------------------------------------------------

The ZOLL ICY Intravascular Heat Exchange Catheter Model IC-3893A, connected to the ZOLL Coolgard/Thermogard Thermal Regulation System, is indicated for use:

• In cardiac surgery adult patients to achieve and/or maintain normothermia during surgery and recovery/intensive care, and

- · To induce, maintain and reverse mild hypothermia in neurosurgery adult patients in surgery and recovery/intensive care.
The ZOLL Quattro Catheter Model IC-4593, connected to a ZOLL Thermal Regulation System, is indicated for use:

• In cardiac surgery adult patients to achieve and/or maintain normothermia during surgery and recovery/intensive care, and

- · To induce, maintain and reverse mild hypothermia in neurosurgery adult patients in surgery and recovery/intensive care.
Type of Use (Select one or both, as applicable)

| <span></span> | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------------------------------------------|
| <span></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)             |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

#### K220008 510(k) SUMMARY

| Date Prepared:       | May 31, 2022                                                                                                                                                                                                                                                                             |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:           | ZOLL Circulation, Inc.                                                                                                                                                                                                                                                                   |
| Address:             | 2000 Ringwood Avenue<br>San Jose, CA 95131                                                                                                                                                                                                                                               |
| Phone:               | 978-421-9291                                                                                                                                                                                                                                                                             |
| Fax:                 | 408-541-1030                                                                                                                                                                                                                                                                             |
| Contact Person:      | Elizabeth Haines, Senior Director Regulatory Affairs                                                                                                                                                                                                                                     |
| Submission Author:   | Melissa Paffenroth, Manager Regulatory Affairs                                                                                                                                                                                                                                           |
| Trade Name:          | Solex 7 Intravascular Heat Exchange Catheter<br>Cool Line Intravascular Heat Exchange Catheter<br>ICY Intravascular Heat Exchange Catheter<br>Quattro Intravascular Heat Exchange Catheter<br>Thermogard HQTM Start-Up Kit<br>Thermogard HQTM Start-Up Kit EX<br>Thermogard HQTM Console |
| Common Name:         | Central Venous Catheter (short term) and Thermal Regulating System                                                                                                                                                                                                                       |
| Classification/Name: | Class II; System, Hypothermia, Intravenous, Cooling                                                                                                                                                                                                                                      |
| Regulation:          | 21 CFR 870.5900, Thermal Regulating System                                                                                                                                                                                                                                               |
| Product Code:        | NCX                                                                                                                                                                                                                                                                                      |
| Predicate Device:    | K213031, Solex 7® Intravascular Heat Exchange Catheter, Cool<br>Line® Intravascular Heat Exchange Catheter, ICY® Intravascular<br>Heat Exchange Catheter, Quattro® Intravascular Heat Exchange<br>Catheter, Start-Up Kit, Coolgard 3000® Console, Thermogard XP®<br>Console              |

#### I. DEVICE DESCRIPTION

The ZOLL® Intravascular Temperature Management (IVTM™) System is comprised of an external heat exchange console (Thermogard HQ™ console) and intravascular heat exchange catheter connected via a sterile heat exchanger and tubing circuit (Thermogard HQTM Start-Up Kit or Thermogard HQ™ Start-Up Kit EX). These components together comprise a patient temperature regulation apparatus employing feedback control. The subject devices of this submission are the Thermogard HQ™ Console, Thermogard HQ™ Start-Up Kit and Thermogard HQTM Start-Up Kit EX. The Solex 7 Intravascular Heat Exchange Catheter, Cool Line Intravascular Heat Exchange Catheter, ICY Intravascular Heat Exchange Catheter, and Quattro Intravascular Heat Exchange Catheter are also included as part of the system but are unchanged compared to the predicate device.

#### II. INDICATIONS FOR USE

The Thermogard HQ™ Console and Thermogard HQ™ Start-Up Kit (EX) are not intended to be used separately from the heat exchange catheters and do not have specific indications for use. The indications for use of the Solex 7® Intravascular Heat Exchange Catheter, Cool Line® Intravascular Heat Exchange Catheter, ICY® Intravascular Heat Exchange Catheter, Quattro® Intravascular Heat Exchange Catheter are identical to the indications for use of the predicate device.

{6}------------------------------------------------

Indications for Use

Solex 7® Intravascular Heat Exchange Catheter

The Solex 7® Intravascular Heat Exchange Catheter connected to the CoolgardThermal Regulation System is indicated for use:

- . In cardiac surgery patients to achieve and/or maintain normothermia during surgery and recovery/intensive care. (Maximum use period: 4 days)
- To induce, maintain and reverse mild hypothermia in neurosurgery patients in surgery and . recovery/intensive care. (Maximum use period: 4 days)
- . In fever reduction, as an adjunct to other antipyretic therapy, in adult patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated. (Maximum use period: 7 days)

## Warning - Fever Reduction

The safety of this device has not been demonstrated for fever reduction in patients presenting with subarachnoid hemorrhage or primary traumatic brain injury. A randomized controlled trial of endovascular cooling in patients with fever associated with subarachnoid hemorrhage and primary traumatic brain injury has shown increased mortality as compared to patients receiving standard of care.

Cool Line® Intravascular Heat Exchange Catheter

The Cool Line® Catheter Model CL-2295A, when used with the ZOLL® Thermal Regulation System, is indicated for use in fever reduction, as an adjunct to antipyretic therapy, in adult patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.

# Warning - Fever Reduction

The safety of this device has not been demonstrated for fever reduction in patients presenting with subarachnoid hemorrhage or primary traumatic brain injury. The safety and effectiveness of this device was examined in a randomized controlled trial of 296 patients. The mortality results reported in this trial, for the four patient cohorts enrolled, are presented in the table below (CI - cerebral

infarction, ICH - intracerebral hemorrhage, PTBI - primary traumatic brain injury, SAH subarachnoid hemorrhage).

|      | Cool Line® |    |      | Control |    |      |      |
|------|------------|----|------|---------|----|------|------|
|      | n          | N  | %    | n       | N  | %    | p*   |
| CI   | 3          | 16 | 18.8 | 3       | 14 | 21.4 | 0.74 |
| ICH  | 8          | 33 | 24.2 | 7       | 27 | 25.9 | 1.00 |
| PTBI | 10         | 44 | 22.7 | 4       | 38 | 10.5 | 0.24 |
| SAH  | 13         | 61 | 21.3 | 7       | 63 | 11.1 | 0.15 |

*Fischer's exact test

For more details on the clinical trial results, refer to the Physician's Manual – "Normothermia for the Neuro-critically Ill stroke patient."

{7}------------------------------------------------

## ICY® Intravascular Heat Exchange Catheter

The ZOLL® ICY® Intravascular Heat Exchange Catheter Model IC-3893A, connected to the ZOLL® Coolgard/Thermogard Thermal Regulation System, is indicated for use:

- . In cardiac surgery adult patients to achieve and/or maintain normothermia during surgery and recovery/intensive care, and
- . To induce, maintain and reverse mild hypothermia in neurosurgery adult patients in surgery and recovery/intensive care.

# Quattro® Intravascular Heat Exchange Catheter

The ZOLL® Quattro® Catheter Model IC-4593, connected to a ZOLL Thermal Regulation System, is indicated for use:

- In cardiac surgery adult patients to achieve and/or maintain normothermia during surgery . and recovery/intensive care, and
- . To induce, maintain and reverse mild hypothermia in neurosurgery adult patients in surgery and recovery/intensive care.

# III. TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT DEVICES COMPARED TO THE PREDICATE DEVICES

There were no changes to the intravascular heat exchange catheters as shown through comparison of the technological characteristics in Table 1 and 2. The comparison with the predicate Start-Up Kit and Start-Up Kit EX shows the technological characteristics of the proposed HOM Start-Up Kit and Thermogard HQTM Start-Up Kit EX to be substantially equivalent in Table 3. The comparison with the predicate Thermogard XP® Console shows the technological characteristics of the proposed Thermogard HOTM Console to be substantially equivalent in Table 4.

| Feature                                                        | PREDICATE DEVICES                                                                                                                                                                                                                                         | SUBJECT DEVICES                                                                                                                                       |                                                                           |      |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|------|
| Device Name                                                    | Quattro® Intravascular Heat<br>Exchange Catheter, Cool Line®<br>Intravascular Heat Exchange<br>Catheter, ICY® Intravascular<br>Heat Exchange Catheter                                                                                                     | Quattro® Intravascular Heat<br>Exchange Catheter, Cool Line®<br>Intravascular Heat Exchange<br>Catheter, ICY® Intravascular<br>Heat Exchange Catheter |                                                                           |      |
| 510(k) Number                                                  | K213031                                                                                                                                                                                                                                                   | K220008                                                                                                                                               |                                                                           |      |
| Class                                                          | II                                                                                                                                                                                                                                                        | Same                                                                                                                                                  |                                                                           |      |
| Classification/Regulation<br>Name                              | System, Hypothermia, Intravenous,<br>Cooling/Thermal Regulating<br>System                                                                                                                                                                                 | Same                                                                                                                                                  |                                                                           |      |
| Regulation Number                                              | 21 CFR 870.5900                                                                                                                                                                                                                                           | Same                                                                                                                                                  |                                                                           |      |
| Product Code                                                   | NCX                                                                                                                                                                                                                                                       | Same                                                                                                                                                  |                                                                           |      |
| Feature                                                        | PREDICATE DEVICES                                                                                                                                                                                                                                         | SUBJECT DEVICES                                                                                                                                       |                                                                           |      |
| Insertion Site                                                 | Cool Line® - Femoral vein,<br>jugular vein, subclavian vein<br>ICY® – Femoral vein<br>Quattro® - Femoral vein                                                                                                                                             | Same                                                                                                                                                  |                                                                           |      |
| Heparin Coating                                                | SurModics Applause<br>Heparin Coating                                                                                                                                                                                                                     | Same                                                                                                                                                  |                                                                           |      |
| Luer Designs                                                   | Inflow and outflow Luers:<br>ZOLL® Custom Luers<br>Infusion Luers: ZOLL®<br>Standard Luers<br>Vent Caps: ZOLL® Custom<br>Vent Caps                                                                                                                        | Same                                                                                                                                                  |                                                                           |      |
| Luer Materials                                                 | Base material is polyurethane<br>(for all Luers except for vent<br>caps)                                                                                                                                                                                  | Same                                                                                                                                                  |                                                                           |      |
|                                                                | Inflow and outflow Luers:<br>Polyurethane: Tecoplast<br>OP-770-164 Orange                                                                                                                                                                                 |                                                                                                                                                       |                                                                           |      |
|                                                                | Distal infusion Luer:<br>Polyurethane:<br>Tecoplast OP-770-477Brown<br>(pad printing white ink)                                                                                                                                                           |                                                                                                                                                       |                                                                           |      |
|                                                                | Medial Infusion Luer:<br>Polyurethane:<br>Tecoplast OP-770-White                                                                                                                                                                                          |                                                                                                                                                       |                                                                           |      |
|                                                                | Proximal infusion Luer:<br>Polyurethane: Tecoplast<br>OP-770-541 Dark Blue<br>(pad printing white ink)                                                                                                                                                    |                                                                                                                                                       |                                                                           |      |
|                                                                | Male vent cap: ABS: Ineos<br>Lustran (P/N 348-012002)                                                                                                                                                                                                     |                                                                                                                                                       |                                                                           |      |
|                                                                | Female vent cap: ABS: Ineos<br>Lustran (P/N 348-012002)                                                                                                                                                                                                   |                                                                                                                                                       |                                                                           |      |
|                                                                | Catheter Working Length<br>(tip to manifold)                                                                                                                                                                                                              |                                                                                                                                                       | Cool Line® – 22 cm<br>ICY® - 38 cm<br>Quattro® - 45 cm                    | Same |
|                                                                | Shaft Diameter                                                                                                                                                                                                                                            |                                                                                                                                                       | 9.3 Fr                                                                    | Same |
|                                                                | Number of Lumens                                                                                                                                                                                                                                          |                                                                                                                                                       | 5 lumens: 2 infusion, 1 guidewire<br>(plus infusion), 1 inflow, 1 outflow | Same |
| Guidewire Compatibility                                        | 0.032 in.                                                                                                                                                                                                                                                 | Same                                                                                                                                                  |                                                                           |      |
| Feature                                                        | PREDICATE DEVICES                                                                                                                                                                                                                                         | SUBJECT DEVICES                                                                                                                                       |                                                                           |      |
| Heat Exchange Balloons                                         | Cool Line® - 2 (straight/coaxial)<br>ICY®- 3 (straight/coaxial)<br>Quattro® - 4 (straight/coaxial)                                                                                                                                                        | Same                                                                                                                                                  |                                                                           |      |
| Flow Rate (by lumen)                                           | Cool Line®<br>Distal - 2100 mL/hr<br>Medial - 1200 mL/hr<br>Proximal – 1400 mL/hr<br><br>ICY®<br>Distal – 1700 mL/hr<br>Medial – 900 mL/hr<br>Proximal - 1200 mL/hr<br><br>Quattro®<br>Distal – 1300 mL/hr<br>Medial – 800 mL/hr<br>Proximal - 1100 mL/hr | Same                                                                                                                                                  |                                                                           |      |
| Approx. Inflated Balloon OD<br>(Cross-sectional Area)          | Cool Line®: ~5 mm (20 mm2)<br>ICY® and Quattro®: ~8 mm (50 mm2)                                                                                                                                                                                           | Same                                                                                                                                                  |                                                                           |      |
| Heat Exchange Power                                            | Cool Line® – 65 Watts nominal<br>ICY®- 140 Watts nominal<br>Quattro® – 190 Watts nominal                                                                                                                                                                  | Same                                                                                                                                                  |                                                                           |      |
| Materials:<br>Shaft                                            | Polyurethane<br>PET and Polyurethane (ICY® and Quattro®)                                                                                                                                                                                                  | Same                                                                                                                                                  |                                                                           |      |
| Balloon                                                        | Polyurethane (Cool Line®)                                                                                                                                                                                                                                 | Same                                                                                                                                                  |                                                                           |      |
| Sterilization Method and<br>Sterility Assurance Level<br>(SAL) | Ethylene Oxide (EO) and SAL 10-6                                                                                                                                                                                                                          | Same                                                                                                                                                  |                                                                           |      |

## Table 1: Comparison of Subject and Predicate Catheters

{8}------------------------------------------------

{9}------------------------------------------------

# Table 2: Comparison of Subject Solex 7® Intravascular Heat Exchange Catheter with
Predicate

| Feature                                                   | PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                 | SUBJECT DEVICE                                   |
|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Device Name                                               | Solex 7® Intravascular<br>Heat Exchange Catheter                                                                                                                                                                                                                                                                                                                                                                                                 | Solex 7® Intravascular<br>Heat Exchange Catheter |
| 510(k) Number                                             | K213031                                                                                                                                                                                                                                                                                                                                                                                                                                          | K220008                                          |
| Class                                                     | II                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                             |
| Classification/Regulation<br>Name                         | System, Hypothermia,<br>Intravenous, Cooling/Thermal<br>Regulating System                                                                                                                                                                                                                                                                                                                                                                        | Same                                             |
| Regulation Number                                         | 21 CFR 870.5900                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                             |
| Product Code                                              | NCX                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                             |
| Heparin Coating                                           | SurModics Applause Heparin<br>Coating                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                             |
| Feature                                                   | PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                 | SUBJECT DEVICE                                   |
| Luer Designs                                              | Inflow and Outflow Luers:<br>ZOLL® Custom Luers<br>Infusion Luers: Standard Luers<br>Vent Caps: ZOLL® Custom Vent<br>Caps                                                                                                                                                                                                                                                                                                                        | Same                                             |
| Luer Materials                                            | Luer Base: Polyurethane (for all<br>Luers except for vent caps)<br><br>Inflow and outflow Luers:<br>Polyurethane: Tecoplast Orange<br><br>Distal infusion Luer:<br>Polyurethane: Tecoplast Brown<br>(pad printing white ink)<br><br>Medial infusion Luer:<br>Polyurethane: Tecoplast White<br><br>Proximal infusion Luer:<br>Polyurethane: Tecoplast Dark<br>Blue (pad printing white ink)<br><br>Male vent cap: ABS<br><br>Female vent cap: ABS | Same                                             |
| Catheter Working Length (tip<br>to manifold)              | 20 cm                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                             |
| Shaft Diameter                                            | 9.3 Fr                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                             |
| Number of Lumens                                          | 5 lumens:<br>2 infusion<br>1 guidewire (also infusion)<br>1 inflow<br>1 outflow                                                                                                                                                                                                                                                                                                                                                                  | Same                                             |
| Guidewire Compatibility                                   | 0.032 in.                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                             |
| Flow Rate (by lumen)                                      | Distal - 1900 mL/hr<br>Medial – 1300 L/hr<br>Proximal - 1300 mL/hr                                                                                                                                                                                                                                                                                                                                                                               | Same                                             |
| Heat Exchange Balloons                                    | 1 (serpentine)                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                             |
| Inflated Balloon OD<br>(Cross- Sectional Area)            | Balloon OD: N/A<br>Cross-sectional Area: 54 mm2                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                             |
| Cross Sectional Area (approx.<br>inflated outer diameter) | 54 mm2 (12.2 mm OD)                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                             |
| Insertion Site                                            | Jugular and Subclavian Veins                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                             |
| Max. Use Period                                           | 7 Days                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                             |
| Materials                                                 | Shaft: Polyurethane<br>Heat Exchange Balloon: PET                                                                                                                                                                                                                                                                                                                                                                                                | Same                                             |
| Feature                                                   | PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                 | SUBJECT DEVICE                                   |
| Sterilization Method and SAL                              | EO and SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                             |

{10}------------------------------------------------

{11}------------------------------------------------

| Table 3: Comparison of the subject Thermogard HQ™ Start-Up Kit with predicate Start-Up Kit |  |  |  |  |
|--------------------------------------------------------------------------------------------|--|--|--|--|
|--------------------------------------------------------------------------------------------|--|--|--|--|

| Feature                                                   | PREDICATE DEVICE                                                                                                                                   | SUBJECT DEVICE                                                                                                                                     |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                                               | Start-Up Kit<br>Start-Up Kit EX                                                                                                                    | Thermogard HQTM Start-Up Kit<br>Thermogard HQTM Start-Up Kit EX                                                                                    |
| 510(k) number                                             | K213031                                                                                                                                            | K220008                                                                                                                                            |
| Model                                                     | CG-500D<br>CG-500D EX                                                                                                                              | TGHQ-500D<br>TGHQ-500D EX                                                                                                                          |
| Class                                                     | II                                                                                                                                                 | Same                                                                                                                                               |
| Classification/Regulation Name                            | System, Hypothermia,<br>Intravenous, Cooling/Thermal<br>Regulating System                                                                          | Same                                                                                                                                               |
| Regulation Number                                         | 21 CFR 870.5900                                                                                                                                    | Same                                                                                                                                               |
| Product Code                                              | NCX                                                                                                                                                | Same                                                                                                                                               |
| Patient Contact                                           | Indirect Patient Contact                                                                                                                           | Same                                                                                                                                               |
| Luer Function                                             | Join the SUK to the<br>InFlow/OutFlow Lumens of the<br>catheters, and allow saline to<br>circulate through the<br>catheter/start-up kit fluid path | Same                                                                                                                                               |
| Supplied 20 ml Sterile<br>Deflation (Slip-Fit)<br>Syringe | Syringe provided with SUK for<br>optional removal of saline from<br>catheter heat exchange balloons<br>prior to catheter removal                   | Same                                                                                                                                               |
| Sterilization Method and<br>SAL                           | Gamma sterilization SAL 10-6                                                                                                                       | Same                                                                                                                                               |
| Shelf Life                                                | 2 years                                                                                                                                            | Same                                                                                                                                               |
| IV Spike design                                           | Bonded into a short section of<br>larger tubing at the end of the<br>tubing set.                                                                   | New component with integrated<br>pockets into which the inlet and<br>outlet tubing are bonded.<br>Revised spike geometry for ease of<br>insertion. |
| Materials:                                                |                                                                                                                                                    |                                                                                                                                                    |
| Start-up Kit tubing                                       | Polyvinyl chloride                                                                                                                                 | Same                                                                                                                                               |
| Luer                                                      | Polyvinyl chloride                                                                                                                                 | Same                                                                                                                                               |
| Spike                                                     | Acrylonitrile butadiene styrene                                                                                                                    | Same                                                                                                                                               |

# Table 4: Comparison of the subject Thermogard HQ™ Console and predicate Thermogard XP® Console

| Feature                                    | PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                             | SUBJECT DEVICE                                                                       |                                                                           |      |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|---------------------------------------------------------------------------|------|
| Device Name                                | Thermogard XP® Console                                                                                                                                                                                                                                                                                                                                                                                                       | Thermogard HQ TM Console                                                             |                                                                           |      |
| 510(k) number                              | K213031                                                                                                                                                                                                                                                                                                                                                                                                                      | K220008                                                                              |                                                                           |      |
| Principle of Operation                     | It automatically adjusts the temperature of a<br>heater/chiller glycol bath to achieve the<br>patient target temperature that has previously                                                                                                                                                                                                                                                                                 | Same                                                                                 |                                                                           |      |
| Feature                                    | PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                             | SUBJECT DEVICE                                                                       |                                                                           |      |
|                                            | been set by the attending physician. A<br>temperature probe is used to monitor the<br>patient temperature, a catheter is used to<br>exchange heat to and from the patient, and<br>the delta between the target temperature and<br>patient temperature defines the temperature<br>of the saline pumped through the catheter. A<br>heat exchanger placed within the glycol bath<br>heats and cools the saline in the catheter. |                                                                                      |                                                                           |      |
|                                            | Class                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                      | Class II                                                                  | Same |
|                                            | Classification/Regulation<br>Name                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                      | System, Hypothermia, Intravenous, Cooling                                 | Same |
|                                            | Regulation Number                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                      | 21 CFR 870.5900                                                           | Same |
|                                            | Product Code                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                      | NCX                                                                       | Same |
|                                            | Patient Contact                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                      | No direct patient contact                                                 | Same |
|                                            | Physical:                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                      |                                                                           |      |
|                                            | 1. Dimensions                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                      | Height: 45 in. (114 cm)<br>Width: 17 in. (43 cm)<br>Depth: 30 in. (76 cm) | Same |
|                                            | 2. Weight                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                      | 107 lb. (49 kg)                                                           |      |
| Electrical:                                |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                      |                                                                           |      |
| 1. Configuration                           | 100-120 VAC, 50/60 Hz, 5 A<br>220-240 VAC, 50/60 Hz, 2.25 A                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                 |                                                                           |      |
| 2. Voltage                                 | 115 V - 230 V                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                      |                                                                           |      |
| 3. Fuse protection                         | T6.3A (slow blow) 5 x 20mm<br>Lag: 60ms@60A Breaking<br>Capacity: 63A @ 250VAC                                                                                                                                                                                                                                                                                                                                               |                                                                                      |                                                                           |      |
| Environmental:                             |                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                 |                                                                           |      |
| 1. Operating<br>temperatures               | 10° C – 27° C (50° F – 81° F)                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                      |                                                                           |      |
| 2. Operating humidity                      | 30% to 75% noncondensing<br>70 kPa to 106 kPa                                                                                                                                                                                                                                                                                                                                                                                |                                                                                      |                                                                           |      |
| 3. Atmospheric pressure:                   |…

---

**Source:** [https://fda.innolitics.com/device/K220008](https://fda.innolitics.com/device/K220008)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
