K214091 · Hefei Dentafilm Medical Equipment Co., Ltd. · MUH · Mar 2, 2022 · Dental
Device Facts
Record ID
K214091
Device Name
Dental Image Plate Scanner, Model DFC-4T-SMART
Applicant
Hefei Dentafilm Medical Equipment Co., Ltd.
Product Code
MUH · Dental
Decision Date
Mar 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
The Dental Image Plate Scanner, Model DFC-4T-SMART is intended to scan dental X-ray latent images contained in the intraoral imaging plate and then generate, browse, process and review the intraoral dental digital X-ray image.
Device Story
Device scans intraoral phosphor imaging plates to digitize latent dental X-ray images. Operation: plate inserted into scanner; laser beam scans plate progressively; blue fluorescence excited proportional to image density; high-efficiency photoconductor collects light; photomultiplier tube converts light to electrical signal; A/D converter digitizes signal. Output: digital image for display, storage, or transmission. Used in hospitals/clinics by trained dental/medical staff. Software (DFC V1.0) manages patient data, examinations, and image processing. Benefits: enables digital review of dental X-rays.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence was demonstrated through non-clinical bench testing, including risk analysis (ISO 14971:2019), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), and photobiological safety (IEC 62471).
Indicated for scanning dental X-ray latent images from intraoral imaging plates to generate, browse, process, and review digital X-ray images in dental/medical clinical settings.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
Predicate Devices
Digital Intraoral Imaging Plate System, Model F200, F210 (K192766)
Submission Summary (Full Text)
{0}------------------------------------------------
March 2, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
Hefei DentaFilm Medical Equipment Co., Ltd. % Ming Chang Qiu Management Representative No. 98 Tangkou Road, Economic and Technological Development Zone Hefei, Anhui 230601 CHINA
Re: K214091
Trade/Device Name: Dental Image Plate Scanner, Model DFC-4T-SMART Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: January 5, 2022 Received: January 5, 2022
Dear Ming Qiu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K214091
Device Name
Dental Image Plate Scanner, Model DFC-4T-SMART
Indications for Use (Describe)
The Dental Image Plate Scanner, Model DFC-4T-SMART is intended to scan dental X-ray latent images contained in the intraoral imaging plate and then generate, browse, process and review the intraoral dental digital X-ray image.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> X Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K214091
# Hefei DentaFilm Medical Equipment Co., Ltd
# Dental Image Plate Scanner, Model DFC-4T-SMART
Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-ray System Regulatory Class: II Product Code: MUH Traditional 510(K) Submission Files
## SECTION 5
510(k) Summary or 510(k) Statement
{4}------------------------------------------------
# 510(k) Summary
#### Submission Number: K214091
#### Submitter
Name: Hefei DentaFilm Medical Equipment Co., Ltd
Address: No. 98 Tangkou Road, Economic and Technological Development Zone,
230601 Hefei City, Anhui Province, People's Republic of China
Name of contact person: Ming Chang Qiu
Telephone: +86-551-63528008
Submission date: 2022-01-26
#### Device
Device trade name: Dental Image Plate Scanner, Model DFC-4T-SMART
Regulation Name: Extraoral Source X-ray System
Regulation class: 2
Regulation number: 21CFR 872.1800
Panel: Radiology
Product code: MUH
### Predicative device
Predicate Submission Number: K192766 Predicate Device Trade Name: Digital Intraoral Imaging Plate System, Model F200, F210 Regulation Name: Extraoral Source X-ray System Regulation number: 21CFR 872.1800 Panel: Radiology Product code: MUH
### Device description
The Dental Image Plate Scanner, Model DFC-4T-SMART is intended to scan and process image data contained in the intraoral imaging plate. Once the intraoral imaging
Traditional 510(K) Submission Files
{5}------------------------------------------------
plate is inserted into the entrance of the scanner, the scanning program will be started to progressively scan the imaging plate, and move forward in the meantime to scan the imaging plate completely with the laser beam in a specified time. In the place where the laser is irradiated, blue fluorescence will be excited and the intensity of the fluorescence brightness is linearly related to the image data density at that point. The fluorescence will be collected by a high-efficiency photoconductor, which is positioned along the laser scanning line and then will be introduced into a photomultiplier tube, where it will be converted into a corresponding electrical signal. After the electrical signal is converted by A/D (analog/digital) converter in the electrical circuit, it can be used for digital image processing and then output to the image display, storage, or transmission communication system.
This Dental Image Plate Scanner, Model DFC-4T-SMART mainly consists of image plate scanner, power adapter, intraoral imaging plate and Dental image plate scanner control system software (Version: V1.0).
The image plate scanner is composed of lasers, optical scanners, photomultiplier tubes, amplifiers, A/D converters, image processing unit and output interfaces etc.
Dental image plate scanner control system software (DFC software) is mainly used to acquire and display the image data of the intraoral imaging plate as well as used for patient management, examination management, image storage, image printing etc. Dental image plate scanner control system software (Version: V1.0) is of Moderate level of concern.
### Indication for use
The Dental Image Plate Scanner, Model DFC-4T-SMART is intended to be used to scan dental X-ray latent images contained in the intraoral imaging plate and then generate, browse, process and review the intraoral dental digital X-ray image.
| Attribute | Subject device | Predicative device | Discussion/<br>Conclusion |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product name | Dental Image Plate<br>Scanner, Model<br>DFC-4T-SMART | Digital Intraoral<br>Imaging Plate System,<br>Model F200, F210 | / |
| Manufacturer | Hefei DentaFilm | Fussen Technology Co., | / |
| DentaFilm<br>Section 5 510(k) Summary | | | |
| Attribute | Subject device | Predicative device | Discussion/<br>Conclusion |
| | Medical Equipment<br>Co., Ltd | Ltd. | |
| 510(k) number | K214091 | K192766 | / |
| Device<br>classification<br>name | Class II | Class II | Same |
| Classification<br>regulations | 21 CFR 872.1800 | 21 CFR 872.1800 | Same |
| Product code | MUH | MUH | Same |
| Product picture | Image: DentaFilm | Image: FUSSEN | / |
| Similarities | | | |
| Indication for<br>Use | The Dental Image<br>Plate Scanner is<br>intended to be used to<br>scan dental X-ray<br>latent images<br>contained in the<br>intraoral imaging plate<br>and then generate,<br>browse, process and<br>review the intraoral<br>dental digital X-ray<br>image. | Digital Intraoral Imaging<br>Plate System is indicated<br>for capturing, digitization<br>and processing of<br>intraoral x-ray images<br>stored in imaging plate<br>recording media. | Same<br>meaning |
| Intended user | The Dental Image<br>Plate Scanner is<br>expected to be used<br>only by professionally<br>qualified<br>dental/medical staffs.<br>Typical user is a dental<br>assistant with specific<br>training for using<br>dental diagnostic | It is intended for uses in<br>hospitals and clinics, and<br>shall be operated and used<br>by trained professionals<br>with physician's<br>guidance. | Same<br>meaning |
| DentaFilm | | | Section 5 510(k) Summary |
| Attribute | Subject device | Predicative device | Discussion/<br>Conclusion |
| Working<br>principle | The Dental Image<br>Plate Scanner is<br>intended to scan and<br>process image data<br>contained in the<br>intraoral imaging<br>plate. Once the<br>intraoral imaging plate<br>is inserted into the<br>entrance of the<br>scanner, the scanning<br>program will be started<br>to progressively scan<br>the imaging plate, and<br>move forward in the<br>meantime to scan the<br>imaging plate<br>completely with the<br>laser beam in a<br>specified time. In the<br>place where the laser is<br>irradiated, blue<br>fluorescence will be<br>excited and the<br>intensity of the<br>fluorescence<br>brightness is linearly<br>related to the image<br>data density at that<br>point. The<br>fluorescence will be<br>collected by a high-<br>efficiency<br>photoconductor, which<br>is positioned along the<br>laser scanning line and<br>then will be introduced<br>into a photomultiplier<br>tube, where it will be<br>converted into a<br>corresponding<br>electrical signal. After | The imaging plate<br>scanner functions as<br>interpreting the latent<br>image information stored<br>in the imaging plate, and<br>it consists of laser, optical<br>scanner, photomultiplier,<br>amplifier, A/D converter,<br>image processing unit,<br>output interface, etc.<br>After a phosphor imaging<br>plate is loaded into the<br>imaging plate scanner<br>inlet, start the scanning<br>program to scan the<br>imaging plate with latent<br>image information to be<br>interpreted row by row,<br>move forwards at the<br>same time, and it can be<br>completely scanned by<br>the laser beams once<br>within set time. Blue<br>florescence can be<br>excited from parts<br>radiated by the laser, and<br>the intensity of the<br>florescence brightness is<br>in linear relationship with<br>the density of the latent<br>image information at the<br>point. The florescence is<br>collected by efficient<br>photoconductor arranged<br>along laser scanning line,<br>is introduced into the<br>photomultiplier, and is<br>further converted into<br>corresponding electric<br>signals. After being<br>transmitted into circuit | Same<br>meaning |
| Attribute | Subject device | Predicative device | Discussion<br>Conclusion |
| Attribute | the electrical signal is<br>converted by A/D<br>(analog/digital)<br>converter in the<br>electrical circuit, it can<br>be used for digital<br>image processing and<br>then output to the<br>image display, storage,<br>or<br>transmission<br>communication<br>system. | A/D (analog/digital)<br>signal conversion, the<br>electric signals can be<br>used in digital image<br>processing, and can be<br>output to image display,<br>storage or transmission<br>communication system. | |
| Product<br>structure | The image plate<br>scanner is composed<br>of lasers, optical<br>scanners,<br>photomultiplier tubes,<br>amplifiers, A/D<br>converters, image<br>processing unit and<br>output interfaces etc. | The imaging plate<br>scanner consists of laser,<br>optical<br>scanner,<br>photomultiplier,<br>amplifier, A/D converter,<br>image processing unit,<br>output interface, etc. | Same |
| Software<br>function design | Dental image plate<br>scanner control system<br>software (DFC<br>software) is mainly<br>used to acquire and<br>display the image data<br>of the intraoral | Digital intraoral imaging<br>plate control system<br>software (CRFC) is<br>mainly used to read<br>images from oral imaging<br>plate and for patient and<br>examination management | Same |
| DentaFilm Section 5 510(k) Summary | | | |
| Attribute | Subject device | Predicative device | Discussion/<br>Conclusion |
| scanning<br>plate reuse. | | | |
| Transport/feed<br>mechanism of<br>imaging plate | Imaging plate tray | Imaging plate holding<br>groove | Same |
| Imaging plate | Intra Oral Dental<br>Phosphor Plates<br>Size 0: 22 x 35 mm<br>Size 1: 24 x 40 mm<br>Size 2: 31 x 41 mm<br>Size 3: 27 x 54 mm | Intra Oral Dental<br>Phosphor Plates<br>Size 0: 22 x 35 mm<br>Size 1: 24 x 40 mm<br>Size 2: 31 x 41 mm<br>Size 3: 27 x 54 mm | Same |
| Image data bit<br>depth | 14 bits/pixel | 14 bits/pixel | Same |
| Data<br>transmission<br>interface | USB 2.0 port | USB 2.0 port | Same |
| Communication | DICOM3.0…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.