The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate.
Device Story
Robotized image-guided system for neurosurgery; assists surgeons in spatial positioning/orientation of instrument holders. Inputs: preoperative 3D MRI/CT images; patient registration data (fiducial markers, optical registration, or stereotactic frame). Operation: surgeon plans instrument trajectories (entry/target points) on images; robot arm moves to calculated position; provides stable, accurate guidance for instruments like biopsy needles, electrodes, or endoscopes. Used in neurosurgical operating rooms by neurosurgeons. Output: physical guidance of surgical tools; real-time display of instrument position relative to anatomy. Benefits: provides reproducible, accurate instrument placement; assists in complex stereotactic procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including system applicative accuracy (robot arm positioning <0.75 mm RMS; device accuracy <2mm), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-1), and software verification/validation.
Technological Characteristics
Robotized electromechanical arm with absolute encoders; touch screen interface. Connectivity: DICOM compliant. Laser: Class 2 (658 nm, 1 mW). Sterilization: Reusable instruments (ISO 17665-1, ISO 17664, ANSI/AAMI ST79). Software: Major level of concern, IEC 62304 compliant.
Indications for Use
Indicated for neurosurgical procedures requiring stereotactic guidance, including biopsy, stimulation/recording electrode placement, and endoscopy. For use by trained neurosurgeons.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
May 4, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Medtech S.A.S Paul Hardy Regulatory Affairs Associate Director 432 rue du Rajol Mauguio, Languedoc-Roussillon 34130 France
## Re: K214065
Trade/Device Name: ROSA ONE Brain Application Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: April 28, 2022 Received: May 4, 2022
## Dear Paul Hardy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D. Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
## 510(k) Number (if known)
K214065
Device Name ROSA ONE Brain Application
Indications for Use (Describe)
The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate.
| <div> <span> <span style="text-decoration: overline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete.
the complete instructions, search existing data sources, gather and maintain the data needed and and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
{3}------------------------------------------------
## 510(k) Summary
This 510(k) summary is submitted in accordance with the requirements of 21 C.F.R. Part 8807.92.
#### SUBMITTER l
Medtech S.A.S 432 rue du Rajol 34130 Mauguio, France Tel +33 (0)4 67 10 77 40
## Contact Person :
Paul HARDY Regulatory Affairs Associate Director paul.hardy@zimmerbiomet.com
Dated prepared: May 4, 2022
### ll DEVICE
| Name of Device: | ROSA ONE Brain application |
|-----------------------|------------------------------------------|
| Common Name : | Neurological Stereotaxic Instrument |
| Classification Name: | Stereotaxic Instrument (21 CFR 882.4560) |
| Classification Panel: | Neurology |
| Regulatory Class: | II |
| Product Code: | HAW |
| 510k #: | K214065 |
### PREDICATE DEVICES ���
ROSA ONE Brain application, manufactured by Medtech S.A., K200511, cleared on May 29, 2020
{4}------------------------------------------------
## IV DEVICE DESCRIPTION
The ROSA One Brain application device is a robotized image-guided device that assists the surgeon during brain surgeries.
lt provides guidance of any surgical instruments compatible with the diameter of the adaptors supplied by Medtech. It allows the user to plan the position of instruments or implants on medical images and provides stable, accurate and reproducible guidance in accordance with the planning.
The device is composed of a robot stand with a compact robotic arm and a touch screen.
Different types of instruments may be attached to the robot arm and changed according to the intended surgical procedure. For Brain applications, these neurosurgical instruments (e.g. biopsy needle, stimulation or recording electrode, endoscope) remain applicable for a variety of procedures as shown below in Figure 1 for the placement of recording electrodes.
Image /page/4/Picture/6 description: The image shows four different views of a brain scan with various colored lines and markers overlaid on them. The top left panel shows a 3D view of a head with lines extending from the scalp. The top right panel displays an axial view of the brain, while the bottom left panel shows a sagittal view. The bottom right panel presents a coronal view, all with the same colored lines and markers indicating specific points or regions of interest within the brain.
Figure 1: Example planning for recording electrodes in a stereo electroencephalography (SEEG) procedure with the ROSA ONE Brain application
The touchscreen ensures the communication between the device and its user by indicating the actions to be performed with respect to the procedure.
Adequate guidance of instruments is obtained from three-dimensional calculations performed from desired surgical planning parameters and registration of spatial position of the patient.
{5}------------------------------------------------
### INDICATIONS FOR USE V
The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate.
{6}------------------------------------------------
# VI COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| | ROSA ONE Brain application | ROSA ONE Brain application | |
|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Device | v.3.1.3.2<br>(K200511) | v.3.1.6.0<br>(submission subject) | Comparison Analysis |
| | Device description and indications for use | | |
| General device description | Computer controlled electromechanical arm providing guidance of neurosurgical instruments | Computer controlled electromechanical arm providing guidance of neurosurgical instruments | Identical |
| Indications for use | The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate. | The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate. | Identical |
| Where used | Neurosurgical operating room | Neurosurgical operating room | Identical |
| User | Neurosurgeon | Neurosurgeon | Identical |
| Anatomical site | Head | Head | Identical |
| Principle of operation | Pre & intraoperative images Surgical planning Patient registration Guidance of instruments | Pre & intraoperative images Surgical planning Patient registration Guidance of instruments | Identical |
| Preoperative images & surgical planning | | | |
| Images type | 3D MRI / CT | 3D MRI / CT | Identical |
| DICOM compliance | Yes | Yes | Identical |
| Device | ROSA ONE Brain application<br>v.3.1.3.2<br>(K200511) | ROSA ONE Brain application<br>v.3.1.6.0<br>(submission subject) | Comparison Analysis |
| Merge images (multimodality<br>image fusion capability) | Yes | Yes | Identical |
| Integrated planning software | ROSANNA BRAIN<br>(Medtech) | ROSANNA BRAIN<br>(Medtech) | Similar<br>Software modified as part of this<br>submission |
| Define regions of interest<br>(ROI) | Yes | Yes | Identical |
| Trajectory planning<br>parameters | Entry point, target point, length of the<br>instrument, diameter, name, color | Entry point, target point, length of the<br>instrument, diameter, name, color | Identical |
| Trajectory definition<br>(endoscopy module) | Parameters for planning trajectories: entry<br>point, target point, length of the instrument,<br>diameter, name, security radius (10mm by<br>default), security aperture (10° by default) | Parameters for planning trajectories: entry<br>point, target point, length of the instrument,<br>diameter, name, security radius (10mm by<br>default), security aperture (10° by default) | Identical |
| Save/load planning | Yes | Yes | Identical |
| Patient Registration | | | |
| Localization means | Robot arm absolute encoders | Robot arm absolute encoders | Identical |
| Controller | Axis controller for each joint<br>Kinematic transformation between the<br>Cartesian space and joint space<br>Supervisor module | Axis controller for each joint<br>Kinematic transformation between the<br>Cartesian space and joint space<br>Supervisor module | Identical |
| Patient registration methods | • Fiducial markers<br>• Optical registration device<br>• Stereotactic frame (fiducials mounted<br>on the frame) | • Fiducial markers (skin, bone)<br>• Optical registration device<br>• Stereotactic frame (fiducials mounted<br>on the frame) | Identical |
| Fiducial markers registration<br>with pointer probe | Yes | Yes | Identical |
| Surface matching registration<br>with optical distance sensor | Yes | Yes | Identical |
| Laser class for optical<br>registration | Class 2 laser<br>Wavelength – 658 nm, Maximum output – 1<br>mW (complies with 21 CFR 1040.10) | Class 2 laser<br>Wavelength – 658 nm, Maximum output – 1<br>mW (complies with 21 CFR 1040.10) | Identical |
| Device | ROSA ONE Brain application<br>v.3.1.3.2<br>(K200511) | ROSA ONE Brain application<br>v.3.1.6.0<br>(submission subject) | Comparison Analysis |
| Cooperative movement | Yes | Yes | Identical |
| Accuracy verification on<br>anatomical landmarks | Yes | Yes | Identical |
| | Instruments guidance | | |
| Image-guided | Yes | Yes | Identical |
| Real time display of the<br>instrument position | Yes | Yes | Identical |
| Provide guidance for surgical<br>instruments | Yes | Yes | Identical |
| Instrument guide position<br>adjustment | Automatic (robotized) | Automatic (robotized) | Identical |
| Surgeon carries out final<br>gesture through the<br>instrument guide with<br>traditional surgical<br>instrument | Yes - through the instrument guide | Yes - through the instrument guide | Identical |
| Instrument fixation | Instruments are mounted onto robot arm's<br>flange | Instruments are mounted onto robot arm's<br>flange | Identical |
| Instruments | Instrument holder, endoscope holder and<br>adaptors, optical sensor | Instrument holder, endoscope holder and<br>adaptors, optical sensor | Identical |
| Instrument calibration method | Factory calibration | Factory calibration | Identical |
| Associated equipment | Pointer probe<br>•<br>Standard tool holder<br>•<br>Endoscope holder<br>•<br>Microdrive holder<br>•<br>Optical sensor<br>•<br>Fiducial markers<br>•<br>Head holder<br>•<br>Leksell frame registration plates<br>•<br>CPW Frame | Pointer probe<br>•<br>Standard tool holder<br>•<br>Endoscope holder<br>•<br>Microdrive holder<br>•<br>Optical sensor<br>•<br>Fiducial markers<br>•<br>Head holder<br>•<br>Leksell frame registration plates<br>•<br>CPW Frame | Identical |
| Device | ROSA ONE Brain application<br>v.3.1.3.2<br>(K200511) | ROSA ONE Brain application<br>v.3.1.6.0<br>(submission subject) | Comparison Analysis |
| Patient immobilization | Yes - The device is attached to the head<br>holder or the frame via an adaptor. | Yes - The device is attached to the head<br>holder or the frame via an adaptor | Identical |
| Device mobility | Yes - Mobile stand with wheels,<br>immobilized with 4 stabilization feet | Yes - Mobile stands with wheels; Robot<br>stand immobilized with stabilization<br>feet | Identical |
| Vigilance system | Yes - foot pedal | Yes - foot pedal | Identical |
| Sterility | Non-sterile and sterile instruments<br>Disposable sterile drapes for the robot arm<br>and touch screen | Non-sterile and sterile instruments<br>Disposable sterile drapes for the robot arm<br>and touch screen | Identical |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
ર
{10}------------------------------------------------
## VII PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The biocompatibility evaluation for the ROSA ONE Brain application device has been conducted in accordance with FDA quidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". The evaluation reveals that biocompatibility requirements are met by the ROSA ONE device.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on ROSA ONE Brain application. The device complies with recognized electrical safety standards: IEC 60601 -1 standard for electrical safety and IEC 60601-1-2 standard for electromagnetic compatibility. The EMC testing was performed according to the FDA EMC guidance document "Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices".
## Software Verification and Validation Testing
Software tests were conducted to satisfy the requirements of the FDA guidance document for the Content of Premarket Submissions for Software Contained in Medical Devices and IEC 62304 Standard (Medical Device Software - Life Cycle Process). The software was considered as a "major" level of concern, since a failure of the software could result in serious injury or death to the patient.
Software verification activities were performed during the "Design, coding & testing" and "Verification" phases of software lifecycle. Outputs generated during these phases include:
- · Code guidelines
- Unit test reports
- · Integration test reports
- · Overall software test report
- · Verification test reports
- · Overall software verification report
Code inspections and software tests at the unit and integration levels were performed according to the Software Test Plan. Verification tests were performed for each software requirement according to the Software Verification Plan.
Conformity of software with the user needs and intended use of the device were performed through the "Validation" phase of the ROSA ONE Brain application.
{11}------------------------------------------------
## Cleaning- and Sterilization Validation
MEDTECH has performed an automated cleaning validation according to FDA guidance document Reprocessing of Reusable Medical Devices: Information for Manufacturers and AAMI TIR 30 Technical report. Additionally, the sterilization was performed according to ISO 17665-1, ISO 17664, ANSI/AAMI ST79, and AAMI TIR 12 Technical report using two cycles.
## Animal studies
Data from animal studies were not required to support the safety and effectiveness of ROSA ONE Brain application.
## Clinical Studies
Clinical data were not required to support the safety and effectiveness of ROSA ONE Brain application. All validation was performed based on non-clinical performance tests.
## VIII SUMMARY OF NON CLINICAL PERFORM ANCE TESTING
| Test | Test Method Summary | Results |
|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System applicative accuracy<br>In vitro testing | Performance bench Testing in<br>compliance with internal Biomet<br>Medtech/Zimmer<br>robotics procedures | Testing were performed on the predicate<br>device.<br><br>The subject devices were evaluated<br>against the predicate testing and<br>determined to be substantially equivalent:<br>• Robot arm positioning accuracy<br><0.75 mm RMS<br>• Device applicative accuracy<br><2mm |
| Electrical safety and<br>electromagnetic compatibility<br>(EMC) | Testing in compliance with the<br>IEC 60601-1:2005/A1:2012 and<br>IEC 60601-1-2:2014 | Evaluation and testing were performed on<br>the predicate device.<br><br>The subject device was evaluated against<br>the predicate testing and determined to be<br>substantially equivalent. |
| Biocompatibility testing | Testing in compliance with FDA<br>Guidance "Use of International<br>Standard IS10993, Biological<br>evaluation of medical Devices<br>Part 1". | The following non clinical tests were<br>performed on the predicate device:<br>Cytotoxicity, Sensitization, Irritation and<br>Acute systemic toxicity.<br><br>The subject devices were evaluated<br>against the predicate testing and<br>determined to be substantially equivalent. |
| Software Verification and<br>Validation Testing | Software verification testing in<br>compliance with FDA guidance<br>"General Principles of Software<br>Validation" and IEC 62304: 2015 | Evaluation and testing were performed on<br>the subject device and demonstrated<br>substantially equivalent performance to<br>identified predicate device. |
| Cleaning-and Sterilization<br>Validation | Testing in compliance with FDA<br>Guidance "Reprocessing<br>Medical Devices in Health Care<br>Settings: Validation Methods | Evaluation and testing were performed on<br>the predicate device. |
{12}------------------------------------------------
| | and Labeling" and the following<br>standards: ISO 17665-1<br>Sterilization of health care<br>products -Moist heat - Part 1:<br>Requirements for<br>the Development, Validation and<br>Routine Control of a Sterilization<br>Process for Medical Devices and<br>ISO 17664 Sterilization of<br>medical devices --Information to<br>be provided by the manufacturer<br>for the processing of re-<br>sterilizable medical devices | The subject device was evaluated against<br>the predicate testing and determined to be<br>substantially equivalent. |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Animal studies | Not applicable | Not applicable |
| Clinical Studies | Not applicable | Not applicable |
## IX CONCLUSIONS
ROSA ONE Brain application (v.3.1.6.0) is substantially equivalent in design and intended use to the predicate device - ROSA ONE Brain application (v.3.1.3.2) (K200511).
Any differences between the subject and the predicate device have no significant influence on safety or effectiveness as established through performance testing. Therefore, ROSA ONE Brain application (v.3.1.6.0). raises no new issues of safety or effectiveness when compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.