Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder System
Applicant
Biomet Manufacturing Corp
Product Code
PHX · Orthopedic
Decision Date
Mar 15, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Biomet Comprehensive Reverse Shoulder products are in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically suited to receive the implants and a functional deltoid muscle is necessary. The Comprehensive Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency. Glenoid components with Hydroxyapatite (HA) coating applied over the porous coating are indicated only for uncemented biological fixation applications. The Glenoid Baseplate components are intended for cementless application with the addition of screw fixation. Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented applications. Humeral components with porous coating are indicated for either cemented or uncemented biological fixation applications.
Device Story
Comprehensive Shoulder System is a modular shoulder joint replacement prosthesis consisting of humeral stems, modular heads, glenoid components, and reverse configuration components (humeral trays and glenospheres). Used in orthopedic surgery for primary, fracture, or revision shoulder arthroplasty. The device replaces damaged joint surfaces to relieve pain and restore function. This submission specifically updates the device labeling to include MR Conditional status. The system is implanted by orthopedic surgeons in a clinical/hospital setting. The device provides mechanical structural support for the shoulder joint.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, specifically MR compatibility evaluation.
Technological Characteristics
Modular shoulder prosthesis components (humeral stems, heads, glenoid baseplates, glenospheres). Materials include metal and polymer. Fixation methods include cemented, press-fit, and screw fixation. Features include Hydroxyapatite (HA) and porous coatings. MR Conditional labeling supported by ASTM F2503-20, F2119-07R13, F2213-17, F2052-21, and F2182-19E02.
Indications for Use
Indicated for patients with grossly deficient rotator cuff, severe arthropathy, or failed shoulder replacement requiring primary, fracture, or revision total shoulder arthroplasty. Requires functional deltoid muscle and suitable anatomy. Specific components indicated for cemented or cementless/press-fit fixation.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
March 15, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two distinct elements: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized depiction of a human figure, while the FDA portion includes the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Biomet Manufacturing Corp. Aishwarya Pandey Regulatory Affairs Specialist 56 East Bell Drive Warsaw, Indiana 46582
# Re: K214001
Trade/Device Name: Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, PAO, KWT, MBF, HSD Dated: December 14, 2021 Received: December 21, 2021
Dear Aishwarya Pandey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Farzana Farzana Sharmin -S Sharmin -S Date: 2023.03.15
For Jiping Chen, MD, Ph.D., M.P.H. Division Director DHT6A:Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K214001
#### Device Name
Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder System (Biomet® Comprehensive® Reverse Shoulder Screws)
#### Indications for Use (Describe)
Biomet Comprehensive Reverse Shoulder products are in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically suited to receive the implants and a functional deltoid muscle is necessary.
The Comprehensive Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
Glenoid components with Hydroxyapatite (HA) coating applied over the porous coating are indicated only for uncemented biological fixation applications. The Glenoid Baseplate components are intended for cementless application with the addition of screw fixation.
Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented applications. Humeral components with porous coating are indicated for either cemented or uncemented biological fixation applications.
Type of Use (Select one or both, as applicable)
| <span> <b> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> | <span> <b> Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known) K214001
#### Device Name
Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder System (Biomet® Comprehensive® Reverse Shoulder Products)
#### Indications for Use (Describe)
Biomet Comprehensive Reverse Shoulder products are indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically suited to receive the implants and a functional deltoid muscle is necessary.
The Comprehensive Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
Titanium glenospheres are intents with Cobalt Alloy material sensitivity. The wear of these devices has not been tested but, based on pin on disk testing, the wear rate is inferior to that of cobalt alloy glenospheres. A Cobalt Alloy glenosphere is the recommended component for reverse shoulder arthroplasty patients without material sensitivity to cobalt alloy.
Glenoid components with Hydroxyapatite (HA) coating applied over the porous coating are indicated only for uncemented biological fixation applications. The Glenoid Baseplate components are intended for cementless application with the addition of screw fixation.
Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented applications. Humeral components with porous coating are indicated for either cemented or uncemented biological fixation applications.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known) K214001
#### Device Name
Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder System (Biomet® Shoulder Joint Replacement Prostheses)
Indications for Use (Describe)
- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis.
- 3. Correction of functional deformity.
- 4. Fractures of the proximal humerus, where other methods of treatment are deemed inadequate.
5. Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate.
Optional use in revision: in some medical conditions (e.g. revision when healthy and good bone stock exists), the surgeon may opt to use primary implants in a revision procedure.
Humeral components with a porous coated surface coating are indicated for either cemented biological fixation applications.
The Comprehensive® Modular Hybrid® Glenoid is intended to be implanted with bone cement. The optional porous titanium peg may be inserted without bone cement. The optional polyethylene peg should be inserted with bone cement.
The Comprehensive Humeral Positioning Sleeves are for cemented use only and are intended for use with the Comprehensive Fracture Stem.
The Versa-Dial Humeral Head Prosthesis is intended for use only with the Comprehensive Shoulder Stems (Fracture, Primary and Revision), and the glenoid components of the Comprehensive Shoulder System.
The Titanium Versa-Dial Humeral Head Prosthesis is indicated for patients with suspected cobalt alloy sensitivity. The wear properties of Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head is not recommended for patients who lack suspected material sensitivity to cobalt alloy*
*The Titanium Versa-Dial Humeral Head Prosthesis is not for sale in Canada
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
510(k) Number (if known) K214001
#### Device Name
Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder System (Biomet® Comprehensive® Convertible Glenoid Prostheses)
Indications for Use (Describe)
Anatomic Applications
1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis.
3. Revision where other devices or treatments have failed.
- 4. Correction of functional deformity.
5. Fractures of the proximal humerus, where other methods of treatment are deemed inadequate.
6. Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may be suitable or may be inadequate.
Reverse Applications
The Comprehensive Reverse Shoulder is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.
Comprehensive Convertible Glenoid Baseplate components are intended for cementless applications with the addition of screw fixation.
Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented application. Huments with porous coated surface coating are indicated for either cemented or uncemented biological fixation applications.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{8}------------------------------------------------
510(k) Number (if known) K214001
#### Device Name
Comprehensive® Shoulder System, Comprehensive® Reverse Shoulder System (Comprehensive Augmented Glenoid Components, Comprehensive Standard Baseplate, Comprehensive Mini Baseplate, Comprehensive® RSA Baseplates)
Indications for Use (Describe)
Biomet Comprehensive Reverse Shoulder products are indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically suited to receive the implants and a functional deltoid muscle is necessary.
The Comprehensive Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
Titanium glenospheres are intents with Cobalt Alloy material sensitivity. The wear of these devices has not been tested but, based on pin on disk testing, the wear rate is inferior to that of cobalt alloy glenospheres. A Cobalt Alloy glenosphere is the recommended component for reverse shoulder arthroplasty patients without material sensitivity to cobalt alloy.
Glenoid components with Hydroxyapatite (HA) coating applied over the porous coating are indicated only for uncemented biological fixation applications. The Glenoid Baseplate components are intended for cementless application with the addition of screw fixation.
Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented applications. Humeral components with porous coating are indicated for ether cemented or uncemented biological fixation applications.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{9}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{10}------------------------------------------------
## 510(k) Summary
| Sponsor: | Biomet Inc.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581<br>Establishment Registration Number: 1825034 |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Aishwarya Pandey<br>Regulatory Affairs Specialist<br>Telephone: 574.373.3740<br>Fax: +1(574)-373-3740 |
| Date: | March 15, 2023 |
| Subject Device: | Trade Name: Comprehensive® Shoulder System, Comprehensive®<br>Reverse Shoulder System<br>Common Name: Shoulder replacement prosthesis |
| | Classification Name:<br>PHX - Shoulder Prosthesis, Reverse Configuration (21 CFR 888.3660) KWS - Shoulder joint, metal/polymer, semi-constrained, cemented prosthesis (21 CFR 888.3660) PAO - Shoulder joint metal/polymer (+additive) semi-constrained cemented prosthesis (21 CFR 888.3660) KWT - Shoulder joint metal/polymer non-constrained cemented prosthesis (21 CFR 888.3650) MBF - Shoulder joint metal/polymer/metal Nonconstrained or semi-constrained porous-coated uncemented prosthesis (21 CFR 888.3670) HSD - Shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis (21 CFR 888.3690) |
## Predicate Device(s):
| Primary Predicate | 510(k) Number |
|---------------------------------------|---------------|
| Comprehensive Primary Shoulder | K193038 |
| Comprehensive Reverse Shoulder System | K193373 |
| Comprehensive Augmented Baseplates | K172502 |
| Comprehensive Convertible Glenoid | K211729 |
## Device Description
Comprehensive Shoulder System consists of partial and total shoulder replacement components use in cemented and uncemented applications. The devices are modular components consisting of humeral stems, modular heads and glenoid components for anatomic and humeral tray and glenosphere components for reverse configuration. The purpose of the current submission is to add MR Conditional labeling.
## Indications for Use:
## Biomet Comprehensive Reverse Shoulder Screws- IFU 01-50-0890
Biomet Comprehensive Reverse Shoulder products are indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy failed shoulder joint
{11}------------------------------------------------
replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.
The Comprehensive Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
Glenoid components with Hydroxyapatite (HA) coating applied over the porous coating are indicated only for uncemented biological fixation applications. The Glenoid Baseplate components are intended for cementless application with the addition of screw fixation.
Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented applications. Humeral components with porous coated surface coating are indicated for either cemented or uncemented biological fixation applications.
## Biomet® Comprehensive® Reverse Shoulder Products- IFU 01-50-0903
Biomet Comprehensive Reverse Shoulder products are indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically suited to receive the implants and a functional deltoid muscle is necessary.
The Comprehensive Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
Titanium glenospheres are intended for patients with Cobalt Alloy material sensitivity. The wear of these devices has not been tested but based on pin on disk testing, the wear rate is inferior to that of cobalt alloy glenospheres. A Cobalt Alloy glenosphere is the recommended component for reverse shoulder arthroplasty patients without material sensitivity to cobalt alloy.
Glenoid components with Hydroxyapatite (HA) coating applied over the porous coating are indicated only for uncemented biological fixation applications. The Glenoid Baseplate components are intended for cementless application with the addition of screw fixation.
Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented applications. Humeral components with porous coated surface coating are indicated for either cemented or uncemented biological fixation applications.
## Biomet® Shoulder Joint Replacement Prostheses- IFU 01-50-0944
1.Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2.Rheumatoid arthritis.
- 3.Correction of functional deformity.
4.Fractures of the proximal humerus, where other methods of treatment are deemed inadequate.
5.Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate.
Optional use in revision: in some medical conditions (e.g.,revision when healthy and good bone stock exists), the surgeon may opt to use primary implants in a revision procedure.
Humeral components with a porous coated surface coating are indicated for either cemented biological fixation applications.
{12}------------------------------------------------
The Comprehensive® Modular Hybrid® Glenoid is intended to be implanted with bone cement. The optional porous titanium peg may be inserted without bone cement. The optional polyethylene peg should be inserted with bone cement.
The Comprehensive Humeral Positioning Sleeves are for cemented use only and are intended for use with the Comprehensive Fracture Stem.
The Versa-Dial Humeral Head Prosthesis is intended for use only with the Comprehensive Shoulder Stems (Fracture, Primary and Revision), and the glenoid components of the Comprehensive Shoulder System.
The Titanium Versa-Dial Humeral Head Prosthesis is indicated for patients with suspected cobalt alloy sensitivity. The wear properties of Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head is not recommended for patients who lack suspected material sensitivity to cobalt alloy*
*The Titanium Versa-Dial Humeral Head Prosthesis is not for sale in Canada.
## Biomet® Comprehensive® Convertible Glenoid Prostheses- IFU 01-50-1256
Anatomic Applications
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 1.
- 2. Rheumatoid arthritis.
- 3. Revision where other devices or treatments have failed.
- 4. Correction of functional deformity.
- Fractures of the proximal humerus, where other methods of treatment are deemed inadequate. 5.
- Difficult clinical management problems, including cuff arthropathy, where other methods of 6. treatment may not be suitable or may be inadequate.
## Reverse Applications
The Comprehensive Reverse Shoulder is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.
Comprehensive Convertible Glenoid Baseplate components are intended for cementless applications with the addition of screw fixation.
Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented application. Humeral components with porous coated surface coating are indicated for either cemented or uncemented biological fixation applications.
### Comprehensive Augmented Glenoid Components. Comprehensive Standard Baseplate, Comprehensive Mini Baseplate (aka Comprehensive® RSA Baseplates)- IFU 01-50-1284
Biomet Comprehensive Reverse Shoulder products are in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.
The Comprehensive Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.
{13}------------------------------------------------
Titanium glenospheres are intended for patients with Cobalt Alloy material sensitivity. The wear of these devices has not been tested but, based on pin on disk testing, the wear rate is inferior to that of cobalt alloy glenospheres. A Cobalt Alloy glenosphere is the recommended component for reverse shoulder arthroplasty patients without material sensitivity to cobalt alloy.
Glenoid components with Hydroxyapatite (HA) coating applied over the porous coating are indicated only for uncemented biological fixations. The Glenoid Baseplate components are intended for cementless application with the addition of screw fixation.
Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented applications. Humeral components with porous coated surface coating are indicated for either cemented or uncemented biological fixation applications.
## Summary of Technological
Characteristics:
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use: Identical to predicate ●
- Indications for Use: Identical to predicate ●
- Materials: Identical to predicate ●
- Design Features: Identical to predicate
- Sterilization: Identical to predicate ●
# Summary of Performance Data
(Nonclinical and/or Clinical)
- Non-Clinical Tests:
- o Evaluation of MR compatibility to support MR Conditional labeling
- ASTM F2503-20 (Labeling) o
- ASTM F2119-07R13 (Artifact) O
- O ASTM F2213-17 (Torque)
- ASTM F2052-21 (Displacement Force) O
- ASTM F2182-19E02 (RF-heating) Preliminary Phantom O Evaluation
- Clinical Tests: ●
- o None provided
## Substantial Equivalence
Conclusion
Based on the information contained within this submission, it is concluded that the Comprehensive Shoulder System is substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.