← Product Code [HAW](/productcode/HAW) · K213989

# Cranial EM System (K213989)

_Brainlab AG · HAW · Sep 8, 2022 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K213989

## Device Facts

- **Applicant:** Brainlab AG
- **Product Code:** [HAW](/productcode/HAW.md)
- **Decision Date:** Sep 8, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

Cranial EM is intended as an image-guided planning and navigation system to enable neurosurgery procedures. The device is indicated for any medical condition in which a reference to a rigid anatomical structure can be identified relative to images (CT, CTA, X-Ray, MR, MRA, and ultrasound) of the anatomy, such as: - Cranial Resection - o Resection of tumors and other lesions - o Resection of skull-base tumor or other lesions - Intracranial catheter placement

## Device Story

Cranial EM System is an intraoperative image-guided navigation system for neurosurgery. It uses electromagnetic (EM) tracking to link patient anatomy (via patient reference) to preoperative image data (CT, MR, etc.). Hardware includes a navigation station (Kick 2 or Curve) and various EM-tracked instruments (pointer, stylet, references). The system emits a varying EM field via a field generator (NDI Aurora V3), inducing currents in instrument sensors to calculate position and orientation. This data is overlaid on 3D anatomical representations on a touchscreen, supporting surgeons during resection and catheter placement. The system is used in clinical settings by neurosurgeons to visualize instrument position relative to patient anatomy, aiding surgical precision and potentially improving patient outcomes by facilitating minimally invasive approaches. It does not perform automated diagnosis; it provides real-time spatial guidance.

## Clinical Evidence

No clinical testing was performed. Substantial equivalence is supported by bench testing, including system accuracy verification (mean location error ≤ 2 mm, trajectory angle error ≤ 2 degrees), platform service life, biocompatibility, reprocessing, sterility, software verification, and usability evaluation. Existing literature was leveraged to validate indications.

## Technological Characteristics

Electromagnetic tracking system using NDI Aurora V3 field generator (500x500x400 mm volume). Hardware platforms: Kick 2 (21.5" LCD) and Curve (31.5" LCD) navigation stations. Instruments include EM-tracked pointers, stylets, and references. Connectivity via USB and network. Software modules: EM Setup, Head Registration, EM Instruments, Navigation. Sterilization for applicable instruments. Cybersecurity measures implemented.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- StealthStation Cranial Software v1.3.0 ([K201175](/device/K201175.md))
- Kolibri Cranial Magnetic ([K042391](/device/K042391.md))

## Reference Devices

- Cranial IGS System ([K192703](/device/K192703.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 8, 2022

Brainlab AG Chiara Cunico Manager, Regulatory Affairs Olof-Palme-Str. 9 Munchen, 81829 Germany

Re: K213989

Trade/Device Name: Cranial EM System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: August 9, 2022 Received: August 9, 2022

## Dear Chiara Cunico:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213989

Device Name Cranial EM System

Indications for Use (Describe)

Cranial EM is intended as an image-guided planning and navigation system to enable neurosurgery procedures. The device is indicated for any medical condition in which a reference to a rigid anatomical structure can be identified relative to images (CT, CTA, X-Ray, MR, MRA, and ultrasound) of the anatomy, such as:

- · Cranial Resection
	- o Resection of tumors and other lesions
- o Resection of skull-base tumor or other lesions
- Intracranial catheter placement

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

#### September 8, 2022

| General Information                                                                          |                                                                                                                                                                                                         |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                                                                 | Brainlab AG; Olof-Palme Str.9; 81829, Munich, Germany                                                                                                                                                   |
| Establishment<br>Registration                                                                | 8043933                                                                                                                                                                                                 |
| Device Name                                                                                  | Neurological Stereotaxic Instrument                                                                                                                                                                     |
| Trade Name                                                                                   | • Cranial EM system                                                                                                                                                                                     |
| Product Code                                                                                 | HAW                                                                                                                                                                                                     |
| Regulation Number                                                                            | 882.4560                                                                                                                                                                                                |
| Regulatory Class                                                                             | II                                                                                                                                                                                                      |
| Panel                                                                                        | Neurology                                                                                                                                                                                               |
| Predicate Devices                                                                            | Primary: StealthStation Cranial Software v1.3.0 (K201175) by Medtronic<br>Secondary: Kolibri Cranial Magnetic (K042391) by Brainlab AG<br>Reference Device: Cranial IGS System (K192703) by Brainlab AG |
| Contact Information                                                                          |                                                                                                                                                                                                         |
| Primary Contact                                                                              | Alternate Contact                                                                                                                                                                                       |
| Chiara Cunico<br>Manager RA<br>Phone: +49 89 99 15 68 0<br>Email: chiara.cunico@brainlab.com | Regulatory Affairs Brainlab<br>Phone: +49 89 99 15 68 0<br>Fax: +49 89 99 15 68 5033<br>Email: regulatory.affairs@brainlab.com                                                                          |

#### 1. Indications for Use

Cranial EM is intended as an image-guided planning and navigation system to enable neurosurgery procedures.

The device is indicated for any medical condition in which a reference to a rigid anatomical structure can be identified relative to images (CT, CTA, X-Ray, MR, MRA and ultrasound) of the anatomy, such as:

- Cranial Resection
	- o Resection of tumors and other lesions
	- o Resection of skull-base tumor or other lesions
- Intracranial catheter placement

### 2. Device Description

The Subject Device, Cranial EM System, consists of software and hardware components. It links patient anatomy (using a patient reference) and instruments in the real world or "patient space" to patient scan data or "image space". This allows for the continuous localization of medical instruments and patient anatomy for medical interventions in cranial procedures. The tracking data are acquired via electromagnetic tracking.

{4}------------------------------------------------

# K213989 Page 2 of 9

Cranial EM is a touchscreen-based intraoperative navigation software. The placement of surgical instruments in a three-dimensional representation overlaid on anatomical image sets, such as MR and/or CT, can support the surgeon during various surgical interventions. Cranial EM uses scanned images of the patient that are acquired before surgery is performed.

The following software make up the main module of the device:

- 1. EM Setup 2.1 2. Head Registration 2.1 3. EM Instruments 2.1
- 4. Navigation 2.1

The Subject Device Consists of the following hardware components:

Platforms:

- 1. Kick 2 Navigation Station (Article Number: 18202)
- 2. Curve Navigation 17700 (Article Number: 17700)

Instruments:

- 1. EM Patient Reference 2.0 (18099-24)
- 2. EM Pointer (18099-02C)
- 3. EM Instrument Reference (18099-05A)
- 4. EM Registration Pointer (18099-23)
- 5. EM Stylet 2.0 (18097-01)
- 6. EM Short Pointer (18099-27)
- 7. EM Skull Reference Base (18099-06)

### 3. Substantial Equivalence

{5}------------------------------------------------

| Topic                                                                     | Subject Device: Cranial EM<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Primary Predicate:<br>StealthStation Cranial<br>Software v1.3.0 (K201175)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Secondary Predicate:<br>Kolibri Cranial Magnetic<br>(K042391)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Reference Device: Cranial<br>Image Guided Surgery<br>System (K192703)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Comment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|---------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use                                                    | Cranial EM is intended as an image-<br>guided planning and navigation system<br>to enable neurosurgery procedures.<br>The device is indicated for any medical<br>condition in which a reference to a rigid<br>anatomical structure can be identified<br>relative to images (CT, CTA, X-Ray, MR,<br>MRA and ultrasound) of the anatomy,<br>such as:<br>• Cranial Resection<br>o Resection of tumors and other<br>lesions<br>o Resection of skull-base tumor or<br>other lesions<br>• Intracranial catheter placement | The StealthStation™ System, with<br>StealthStation™ Cranial Software, is<br>intended as an aid for locating<br>anatomical structures in either open<br>or percutaneous neurosurgical<br>procedures. Their use is indicated<br>for any medical condition in which<br>the use of stereotactic surgery may<br>be appropriate, and where reference<br>to a rigid anatomical structure, such<br>as the skull, can be identified relative<br>to images of the anatomy.<br>This can include, but is not limited<br>to, the following cranial procedures<br>(including stereotactic frame-based<br>and stereotactic frame alternatives-<br>based procedures):<br>• Tumor resections<br>• General ventricular catheter<br>placement<br>• Pediatric ventricular catheter<br>placement<br>• Depth electrode, lead, and probe<br>placement<br>• Cranial biopsies | The BrainLAB Kolibri IGS<br>System is intended to be an<br>intraoperative image guided<br>localization system to enable<br>minimally invasive surgery. It<br>links a freehand probe, tracked<br>by a magnetic sensor system<br>or a passive marker sensor<br>system to a virtual computer<br>image space on a patient's<br>preoperative diagnostic image<br>data set being processed by<br>the Kolibri computer<br>workstation. The system is<br>indicated for any medical<br>condition in which the use of<br>stereotactic surgery may be<br>appropriate and where a<br>reference to a rigid anatomical<br>structure, such as the skull, a<br>long bone, or vertebra, can be<br>identified relative to a CT, X-ray<br>or MR based model of the<br>anatomy.<br>Example image guided cranial<br>surgery procedures include but<br>are not limited to:<br>1. Tumor resections<br>2. Skull base surgery<br>3. Cranial biopsies<br>4. Craniotomies/Craniect<br>omies<br>5. Pediatric Catheter<br>Shunt Placement<br>6. General Catheter<br>Shunt Placement | The Cranial IGS System, when used<br>with a compatible navigation<br>platform and compatible instrument<br>accessories, is intended as an<br>image-guided planning and<br>navigation system to enable<br>navigated surgery. It links<br>instruments to a virtual computer<br>image space on patient image data<br>that is being processed by the<br>navigation platform.<br>The system is indicated for any<br>medical condition in which a<br>reference to a rigid anatomical<br>structure can be identified relative to<br>images (CT, CTA, X-Ray, MR, MRA<br>and ultrasound) of the anatomy,<br>including:<br>•Cranial Resection<br>o Resection of tumors and<br>other lesions<br>o Resection of skull-base<br>tumor or other lesions<br>•AVM Resection<br>•Cranial biopsies<br>•Intracranial catheter placement<br>•Intranasal structures and Paranasal<br>Sinus Surgery<br>•Functional endoscopic sinus<br>surgery (FESS)<br>•Revision & distorted anatomy<br>surgery all intranasal structures and<br>paranasal sinuses | The wording has been<br>rephrased with the<br>purpose to improve<br>readability and clarity<br>without affecting the<br>content.<br>The included procedures<br>are grouped differently:<br>resections are bundled<br>and approaches as<br>craniotomy not listed<br>separately. Catheter<br>placement are bundle<br>without patient population<br>specifications. Biopsies,<br>depth electrode<br>placement, ENT<br>indications or AVM are not<br>part of the indications of<br>the subject device.<br>The Subject Devices'<br>indications are within the<br>subset of the primary<br>predicate and reference<br>devices' indications.                                                                                           |
| Topic                                                                     | Subject Device: Cranial EM<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Primary Predicate:<br>StealthStation Cranial<br>Software v1.3.0 (K201175)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Secondary Predicate:<br>Kolibri Cranial Magnetic<br>(K042391)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Reference Device: Cranial<br>Image Guided Surgery<br>System (K192703)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Comment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Localization<br>technique<br>based on:                                    | Electromagnetic tracking<br>The Field generator emits low<br>intensity and varying<br>electromagnetic field which induce<br>small currents in the sensors<br>embedded in the EM instruments.<br>The position and spatial orientation<br>of the sensors integrated in the EM<br>instruments are calculated in the<br>Base station.                                                                                                                                                                                   | Electromagnetic tracking<br>The Side Emitter emits low<br>intensity and varying<br>electromagnetic field which<br>induce small currents in the<br>sensors embedded EM<br>instruments. The position and<br>spatial orientation of the sensors<br>integrated in the EM instruments<br>are calculated in the Instrument<br>Interface Box.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Electromagnetic tracking<br>The Field generator emits<br>low intensity and varying<br>electromagnetic field which<br>induce small currents in the<br>sensors of the EM<br>instruments. The position<br>and spatial orientation of the<br>sensors integrated in the EM<br>instruments are calculated in<br>the Base station.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Optical Tracking- N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | General localization<br>techniques<br>remain the same w.r.t.<br>the predicate devices:<br>K201175 and K042391.<br>K19270 uses an optical<br>tracking system.<br>Therefore it´s not<br>comparable with the<br>other systems                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Tracking<br>Accuracy                                                      | Under representative worst-case<br>configuration, the Cranial EM<br>System is tested to ensure that its<br>mean location error is ≤ 2 mm and<br>its mean trajectory angle error is ≤ 2<br>degrees.                                                                                                                                                                                                                                                                                                                  | Under representative worst-case<br>configuration, the StealthStation<br>S8 System with StealthStation<br>Cranial Software v1.3.0, has<br>demonstrated performance in<br>3D positional accuracy with a<br>mean error ≤ 2.0 mm and in<br>trajectory angle accuracy with a<br>mean error ≤ 2 degrees.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | ≤ 2.5 mm distance RMS<br>(Root Mean Square)<br>≤ 2.5° angle RMS (Root<br>Mean Square)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Under representative worst-case<br>configuration, the Cranial EM<br>System is tested to ensure that<br>its mean location error is ≤ 2 mm<br>and its mean trajectory angle<br>error is ≤ 2 degrees.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | The subject device and<br>primary predicate<br>device are both<br>meeting the same<br>accuracy requirements.<br>Refer to Ground truth<br>accuracy test report<br>[0000279513]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Topic                                                                     | Subject Device: Cranial EM System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Primary Predicate:<br>StealthStation Cranial<br>Software v1.3.0 (K201175)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Secondary Predicate:<br>Kolibri Cranial Magnetic<br>(K042391)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Reference Device: Cranial<br>Image Guided Surgery<br>System (K192703)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Comment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Field<br>Generator                                                        | Manufacturer: NDI<br>Localizer: Aurora V3<br>Emitter Type: side<br>Tracking Volume: 500x500x400 mm<br>Image: Drawing of tracking volume<br>Image: Drawing of tracking volume<br>Field strength:<br>Image: Graph of field strength                                                                                                                                                                                                                                                                                   | Manufacturer: Medtronic<br>Localizer: AXIEM III<br>Emitter Type: side, flat<br>Tracking Volume: 460x460x310 mm (side), 400x400x375 mm (flat)<br>Image: Drawing of tracking volume<br>Image: Drawing of tracking volume<br>Field strength:<br>$< 2.10 x (f / 3350) V/m$ where f is<br>the frequency in Hz from 3kHz<br>to 5MHz as set by IEEE<br>Standard C95.1 - 2005                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Manufacturer: NDI<br>Localizer: Aurora V1<br>Emitter Type: side<br>Tracking Volume:<br>500x500x500 mm<br>Field strength: same as<br>subject device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Optical Tracking- N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | The tracking volume of<br>the subject device and<br>predicate devices<br>covers the region of<br>interest. The subject<br>device has a slightly<br>bigger volume in x, y<br>and z direction<br>compared to the<br>primary predicate.<br>Compared to the<br>secondary predicate,<br>the volume of the field<br>generator was reduced<br>in the new version to<br>400 mm in z direction to<br>ensure accurate<br>positioning of the head<br>in the tracking volume.<br>During Usability testing<br>it was verified that<br>representative clinical<br>users are able to<br>position, register and<br>navigate within the<br>reduced/new tracking<br>volume of the subject<br>device [Usability<br>summary 0000269739,<br>rev. 1]. |
| Topic                                                                     | Subject Device: Cranial EM<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Primary Predicate:<br>StealthStation Cranial<br>Software v1.3.0 (K201175)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Secondary Predicate:<br>Kolibri Cranial Magnetic<br>(K042391)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Reference Device: Cranial<br>Image Guided Surgery<br>System (K192703)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Comment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Touch screen                                                              | Curve Navigation 17700:<br>Type: LCD TFT display<br>Size: 31.5"<br>Color depth: 1.073B colors<br>Resolution: 3840 x 2160 pixels<br>Touch technology: Projected<br>capacitive<br><br>Kick 2 Navigation Station 18202:<br>Type: LCD TFT display<br>Size: 21.5"<br>Color depth: 16.7M colors<br>Resolution: 1920 x 1080 pixels<br>Touch technology: Projected<br>capacitive                                                                                                                                            | Monitor dimension: 15.3" x 25.5"<br>x 2.5"<br>Resolution: 2560x1440 pixels                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Type: LCD TFT display<br>Size: 15"<br>Color depth: 16M colors<br>Resolution: 1024 x 768<br>pixels<br>Touch technology:<br>AccuTouch resistive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | -----                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Two new platforms as<br>hardware components<br>have been included<br>with the Subject<br>Device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Electrical<br>specifications                                              | Curve Navigation 17700:<br>• Frequency: 50 / 60 Hz<br>• Power consumption: 7.6 A @<br>100 VAC, 3.3 A @ 240 VAC<br><br>Kick 2 Navigation Station 18202:<br>• Frequency: 50 / 60 Hz<br>• Power consumption: 3 A @ 100<br>VAC, 1.5 A @ 240 VAC                                                                                                                                                                                                                                                                         | Frequency: 50 / 60 Hz<br>Input voltage: 100 - 230 V<br>Maximum power: 900 VA<br>Typical power dissipation: 400 –<br>600 VA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Frequency: 50 / 60 Hz<br>Power consumption: 3 A @<br>100-115 V, 15 A @ 230 V                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | -----                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Two new platforms as<br>hardware components<br>have been included<br>with the Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Data loading<br>possibilities                                             | • USB<br>• Network<br><br>Wireless transfer                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | • USB<br>• CD / DVD<br><br>Wireless transfer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | • USB<br>• CD/DVD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | • USB<br>• Network<br><br>Wireless transfer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Only state of the art<br>data loading<br>possibilities are<br>supported.<br>Cybersecurity summary<br>[0000271450] contains<br>the details on risks<br>related to cybersecurity<br>and the measures<br>implemented included<br>additional<br>recommended<br>considerations.                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Topic                                                                     | Subject Device: Cranial EM<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Primary Predicate:<br>StealthStation Cranial<br>Software v1.3.0 (K201175)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Secondary Predicate:<br>Kolibri Cranial Magnetic<br>(K042391)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Reference Device: Cranial<br>Image Guided Surgery<br>System (K192703)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Comment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Imaging<br>modality<br>supported<br>[Head<br>Registration<br>application] | CT, MR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | X-Ray based, MR based,<br>Nuclear Medicine based                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | CT, MR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | CT, MR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | CT and MR are the<br>main supported<br>modalities where<br>patient registration is<br>also supported. Nuclear<br>medicine based<br>modalities are<br>supported for display in<br>the navigation.<br>Main supported<br>modalities are same.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Planning<br>features<br>[Navigation<br>App]                               | Plan Entry and Target<br>Selection 3D Model Building                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Plan Entry and Target<br>Selection 3D Model Building Advanced Visualization Create Patient Based<br>Anatomical Coordinate<br>Space Stereotactic Frame Settings Brain<br>Atlas: Schaltenbrand-<br>Wahren Atlas with<br>Talairach Grid<br>STarFixTM Designer Annotations                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Plan Entry and Target<br>Selection 3D Model Building Advanced Visualization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Planning features<br>relevant for the covered<br>indications for use are<br>same.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{6}------------------------------------------------

{7}------------------------------------------------

# K213989 Page 5 of 9

{8}------------------------------------------------

# K213989 Page 6 of 9

{9}------------------------------------------------

# K213989 Page 7 of 9

{10}------------------------------------------------

# K213989 Page 8 of 9

| Topic                                      | Subject Device: Cranial EM<br>System                                                                        | Primary Predicate:<br>StealthStation Cranial<br>Software v1.3.0 (K201175)                                                                                                                                     | Secondary Predicate:<br>Kolibri Cranial Magnetic<br>(K042391) | Reference Device: Cranial<br>Image Guided Surgery<br>System (K192703)                                                                                                                                                                                                                                                                                                  | Comment                                                                                           |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Navigation<br>Views<br>[Navigation<br>App] | Axial Coronal Sagittal Inline 1 Inline 2 Probe's Eye 3D Overview Video View (e.g: Endoscope) Autopilot-View | Axial Coronal Sagittal Trajectory 1 Trajectory 2 Probes Eye Target Guidance,<br>Trajectory Guidance Endoscopic Look Ahead Ultrasound Video In Ultrasound Overlay,<br>3D, Microscope Injection,<br>Video Input | Axial Coronal Sagittal Probes Eye                             | Axial Coronal Sagittal Inline 1 Inline 2 Probe's Eye Split Probe's Eye Skin Overview Skin Probe's Eye Bone Overview 3D Overview Video View Ultrasound video Acquired 2-D ultrasound<br>images Acquired 3-D ultrasound<br>data sets Ultrasound reconstruction:<br>superimposition of<br>ultrasound video with the<br>correlated diagnostic image<br>data Autopilot-View | Main navigation views<br>are identical to those in<br>primary and secondary<br>predicate devices. |

{11}------------------------------------------------

## 4. Performance Data

The following tests were conducted:

- System accuracy tests (Software + platforms + instruments)
- Platform expected service life verification
- Biocompatibility and reprocessing and shelf life sterility tests for applicable instruments
- User contact biocompatibility assessment
- Routine Software verification
- Platform Verification
- Usability Evaluation

No clinical testing submitted for the Subject Device. Data from existing literature was leveraged for validating the Subject Device's indications.

## 5. Conclusion

Based on the battery of non-clinical testing conducted for the Subject Device as listed above, and based on the Substantial Equivalence discussion, the Subject Device was demonstrated to be as safe and effective as the predicate devices for the intended use and indications stated.

---

**Source:** [https://fda.innolitics.com/device/K213989](https://fda.innolitics.com/device/K213989)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
