← Product Code [MAX](/productcode/MAX) · K213951

# FORZA XP Expandable Spacer System (K213951)

_Orthofix US, LLC · MAX · Jan 5, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K213951

## Device Facts

- **Applicant:** Orthofix US, LLC
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Jan 5, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

FORZA XP Expandable Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved level(s). FORZA XP Expandable Spacer System is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation. Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with FORZA XP Expandable Spacer System.

## Device Story

Expandable intervertebral body fusion device; titanium alloy and PEEK polymer construction; height expansion capability; bulleted nose for insertion; anti-migration surface ripples. Used in TLIF and PLIF procedures; implanted by surgeons; requires supplemental fixation; not for stand-alone use. Restores spinal curvature; facilitates fusion in DDD patients. Sterile packaging configuration added as product line extension.

## Clinical Evidence

No clinical data provided. Bench testing not required due to lack of design changes; substantial equivalence supported by comparison to predicate device.

## Technological Characteristics

Materials: Titanium alloy and PEEK polymer. Design: Expandable height, bulleted nose, anti-migration surface ripples. Configurations: Non-lordotic, lordotic, hyperlordotic. Sterilization: Sterile packaging configuration.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- FORZA XP Spacer System ([K211704](/device/K211704.md))

## Reference Devices

- FORZA Ti Spacer System ([K211704](/device/K211704.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 5, 2022

Orthofix US LLC Natalia Volosen Regulatory Affairs Program Manager 3451 Plano Parkway Lewisville, Texas 75056

Re: K213951

Trade/Device Name: FORZA XP Expandable Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: December 16, 2021 Received: December 17, 2021

Dear Ms. Volosen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213951

Device Name FORZA® XP Expandable Spacer System

### Indications for Use (Describe)

FORZA XP Expandable Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved level(s).

FORZA XP Expandable Spacer System is intended for use with autograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation.

Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with FORZA XP Expandable Spacer System.

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

Device Trade Name: FORZA XP Expandable Spacer System

| Manufacturer:        | Orthofix US LLC<br>3451 Plano Parkway<br>Lewisville, TX 75056, USA<br>Phone: (214) 937-2145<br>Fax: (214) 937-3322 |
|----------------------|--------------------------------------------------------------------------------------------------------------------|
| Contact Person:      | Ms. Natalia Volosen<br>Regulatory Affairs Program Manager<br>nataliavolosen@orthofix.com                           |
| Date Prepared:       | December 16, 2021                                                                                                  |
| Registration Number: | 2183449                                                                                                            |
| Product Code:        | MAX                                                                                                                |
| Classifications:     | Class II - 21 CFR §888.3080, Intervertebral body fusion device                                                     |
| Primary Predicate:   | K211704 – FORZA XP Spacer System, SE 8/31/2021                                                                     |

Reason for the 510(k) Submission: Product Line Extension

# Device Description:

The FORZA XP Expandable Spacer System is comprised of an assortment of single use spacers made from titanium alloy and PEEK Polymer with height expansion capability. The implants feature a bulleted nose for ease of insertion and anti-migration ripples on both the inferior and superior surfaces to provide increased stability and help prevent anterior/posterior movement of the device.

The implants are offered in non-lordotic, lordotic, and hyperlordotic configurations to help restore the natural curvature of the spine. The implants can be used in single placement or pairs with typical approaches being transforaminal lumbar interbody fusion (TLIF) and posterior lumbar interbody fusion (PLIF).

The FORZA XP Expandable Spacer System is not intended to be used as a stand-alone device and must be used with a supplemental fixation system.

# Indications for Use and Intended Use:

FORZA XP Expandable Spacer System is indicated for spinal fusion procedures in skeletally mature patients with deqenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surqery at the involved level(s).

{4}------------------------------------------------

FORZA XP Expandable Spacer System is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation.

Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with FORZA XP Expandable Spacer System.

### Performance Testing Summary:

No new mechanical testing was performed for the FORZA XP Expandable Spacer System because there were no design changes to the device.

The proposed sterile package offering does not change the design, materials, function, performance for the FORZA XP implants most recently cleared thru 510(k) K211704. FORZA XP sterile packaging confiquration use Orthofix standard sterile packaging confiquration recently cleared thru FORZA Ti Spacer System (K211704).

# Substantial Equivalence:

The proposed sterile package offering does not change the design, materials, function, performance, indications for use of the device; does not alter the intended use of the implants therefore the subject FORZA XP Expandable sterile packaged devices is substantially equivalent to the FORZA XP Expandable Spacer System (K211704) predicate device

# Conclusion:

FORZA XP Expandable Spacer System has the same intended use, indications for use, technological characteristics, materials, the same principles of operation and same design as the predicate device FORZA XP Expandable Spacer System (K211704).

---

**Source:** [https://fda.innolitics.com/device/K213951](https://fda.innolitics.com/device/K213951)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
