K213890 · Imicryl Dis Malzemeleri San. VE Tic. A.S. · ELW · Feb 11, 2022 · Dental
Device Facts
Record ID
K213890
Device Name
IMICRYL Impression Materials
Applicant
Imicryl Dis Malzemeleri San. VE Tic. A.S.
Product Code
ELW · Dental
Decision Date
Feb 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3660
Device Class
Class 2
Indications for Use
PUTTY SOFT is to be used as preliminary materials for: -Two-step Putty-wash impression technique -One-step Putty-wash impression technique HEAVY BODY materials for: -One-step impression technique (simultaneous technique) using single or dual viscosities -Two-step impression technique using dual viscosities -Functional impression MEDIUM/MONOPHASE/REGULAR BODY tray or syringeable impression material for: -Taking impressions over fixed/removable restorations and implants (i.e., transferring impression posts and bridge components) -Functional impressions -Fabricating crown and bridgework or inlays -Fabricating full or partial dentures -Reline impressions -Use in the simultaneous mixing technique as well as the putty-wash and triple tray techniques -Transferring root posts when fabricating posts and cores indirectly LIGHT BODY is to be used as syringeable impression materials for: -Two-step putty-wash impression technique -One-step putty-wash impression technique -Two-step impression technique using dual viscosities -Reline impressions -Fabricating full or partial dentures EXTRA LIGHT BODY is to be used as syringeable impression materials for: -Two-step putty-wash impression technique -One-step putty-wash impression technique -Two-step impression technique using dual viscosities -Reline impressions -Fabricating full or partial dentures BITE REGISTRATION is used for impression as below. -Taking occlusal surfaces -Confirming occlusal surfaces -Recording after putting the articulator
Device Story
IMICRYL Impression Materials are elastomeric, hydrophilic silicone-based materials (vinylpolysiloxane) used by dentists to capture anatomical impressions of the oral cavity. The family includes six viscosities: Putty Soft, Heavy Body, Medium/Monophase/Regular Body, Light Body, Extra Light Body, and Bite Registration. The material is applied via tray or syringe to the patient's mouth to create a negative impression, which is then used to produce a plaster cast for diagnosis, treatment planning, or fabrication of dental prosthetics (crowns, bridges, dentures, implants). The device aids clinicians in identifying necessary interventions and verifying the effectiveness of dental work. It is intended for professional use in a clinical dental setting.
Clinical Evidence
No clinical data. Substantial equivalence is supported by in-vitro bench testing performed in accordance with ISO 4823:2021, including consistency, working/mixing time, detail reproduction, compatibility with gypsum, linear dimensional change, elastic recovery, strain-in-compression, time in mouth, compression set, and hardness. Biocompatibility testing (cytotoxicity, irritation, sensitization) confirmed the device is biocompatible.
Technological Characteristics
Elastomeric vinylpolysiloxane impression material. Hydrophilic properties. Six viscosities (Putty Soft, Heavy, Medium/Monophase/Regular, Light, Extra Light, Bite Registration). Complies with ISO 4823:2021. Cured by addition reaction. Supplied in 1:1 250/300 mL jars (Putty) or 1:1 50 mL syringes.
Indications for Use
Indicated for dental patients requiring impressions for fixed/removable restorations, implants, crown/bridgework, inlays, full/partial dentures, relines, or occlusal registration. Used by dentists for anatomical data collection and plaster cast fabrication.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 11, 2022
IMICRYL Dis Malzemeleri San. Ve Tic. A.S. Husamettin Sonmez General Manager Fetih Mahallesi Mahir Sokak No:5/201 Konya, Karatay 42030 TURKEY
Re: K213890
Trade/Device Name: IMICRYL Impression Materials Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: Class II Product Code: ELW Dated: December 6, 2021 Received: December 13, 2021
Dear Husamettin Sonmez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE., Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213890
Device Name IMICRYL Impression Materials
| Indications for Use (Describe) | |
|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PUTTY SOFT is to be used as preliminary materials for: | |
| | -Two-step Putty-wash impression technique |
| | -One-step Putty-wash impression technique |
| HEAVY BODY materials for: | |
| | -One-step impression technique (simultaneous technique) using single or dual viscosities |
| | -Two-step impression technique using dual viscosities |
| | -Functional impression |
| MEDIUM/MONOPHASE/REGULAR BODY tray or syringeable impression material for: | |
| | -Taking impressions over fixed/removable restorations and implants (i.e., transferring impression posts and bridge components) |
| | -Functional impressions |
| | -Fabricating crown and bridgework or inlays |
| | -Fabricating full or partial dentures |
| | -Reline impressions |
| | -Use in the simultaneous mixing technique as well as the putty-wash and triple tray techniques |
| | -Transferring root posts when fabricating posts and cores indirectly |
| LIGHT BODY is to be used as syringeable impression materials for: | |
| | -Two-step putty-wash impression technique |
| | -One-step putty-wash impression technique |
| | -Two-step impression technique using dual viscosities |
| | -Reline impressions |
| | -Fabricating full or partial dentures |
| EXTRA LIGHT BODY is to be used as syringeable impression materials for: | |
| | -Two-step putty-wash impression technique |
| | -One-step putty-wash impression technique |
| | -Two-step impression technique using dual viscosities |
| | -Reline impressions |
| | -Fabricating full or partial dentures |
| BITE REGISTRATION is used for impression as below. | |
| | -Taking occlusal surfaces |
| | -Confirming occlusal surfaces |
| | -Recording after putting the articulator |
| Type of Use (Select one or both, as applicable) | |
| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for IMICRYL. The logo consists of a blue circle with the letters "IMI" inside of a white shape that resembles a tooth. To the right of the circle is the word "IMICRYL" in blue, and below the circle is the text "DIS MALZEMELERI SAN. VE TIC A.S." in a smaller blue font.
# 510(k) Summary - K213890
# IMICRYL Impression Materials
The IMICRYL family of silicone impression materials: includes the brand names SPIRIAS, NICETY, REFLECT and PE Sil.
Date of Summary Preparation: November 30, 2021
Type of Submission: Traditional 510(k)
#### SUBMITTER INFORMATION:
Company Name: IMICRYL DIS MALZEMELERI SANAYI VE TICARET A.S.
Company Address: Fetih Mahallesi Mahir Sokak No: 5/201 Zip Code:42030 Karatay/Konya/TURKEY
Phone Number: 0090 444 71 30
Fax Number: 0090 332 355 76 19
Contact Person: Husamettin SONMEZ
Phone: 0090 444 71 30 ext: 120
E-mail: kaliteguvence@imicryl.com
#### Secondary Contact Person: Muhammed Hulusi SONMEZ
Phone: 0090 542 372 27 23
E-mail: hsonmez@imicryl.com.tr
#### DEVICE INFORMATION:
Trade Name: IMICRYL Impression Materials Common Name: Impression Material Product Code: ELW Classification: Class II Classification Name: Material, Impression Regulation Number: 21 CFR 872.3660
Review Panel: Dental
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for IMICRYL. The logo consists of a circular emblem with the letters "IMI" inside, followed by the word "IMICRYL" in bold, blue letters. Below the logo is the text "DIŞ MALZEMELERI SAN. VE TIC A.Ş." in a smaller font size, also in blue.
# PREDICATE DEVICES:
IMICRYL Impression Materials is substantially equivalent to the following marketed product:
| COMPANY | DEVICE | 510(k) NUMBER | PRODUCT CODE |
|-----------------------------|----------------------------|---------------|--------------|
| Osstem Implant Co.,<br>Ltd. | HySil Impression Materials | K170736 | ELW |
# INDICATION FOR USE:
SPIRIAS, NICETY, REFLECT and PE Sil Putty Soft is to be used as preliminary materials for: Two-step Putty-wash impression technique One-step Putty-wash impression technique
SPIRIAS, NICETY, REFLECT and PE Sil Heavy Bodied materials for: One-step impression technique (simultaneous technique) using single or dual viscosities Two-step impression technique using dual viscosities Functional impression
SPIRIAS, NICETY, REFLECT and PE Sil Medium/Monophase/Reqular Bodies tray or syringeable impression material for:
- · Taking impressions over fixed/removable restorations and implants (i.e., transferring impression posts and bridge components)
- · Functional impressions
- · Fabricating crown and bridgework or inlays
- Fabricating full or partial dentures
- · Reline impressions
- Use in the simultaneous mixing technique as well as the putty-wash and triple tray techniques
- · Transferring root posts when fabricating posts and cores indirectly
SPIRIAS, NICETY, REFLECT and PE Sil Light Body is to be used as syringeable impression materials for: Two-step putty-wash impression technique
One-step putty-wash impression technique
Two-step impression technique using dual viscosities
Reline impressions
Fabricating full or partial dentures
SPIRIAS, NICETY, REFLECT and PE Sil Extra Light Body is to be used as syringeable impression
materials for: Two-step putty-wash impression technique One-step putty-wash impression technique Two-step impression technique using dual viscosities Reline impressions Fabricating full or partial dentures
SPIRIAS, NICETY, REFLECT and PE Sil Bite Registration is used for impression as below.
- Taking occlusal surfaces
- Confirming occlusal surfaces
- Recording after putting the articulator
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo for IMICRYL. The logo consists of a circular emblem on the left, containing the letters "IMI" inside a tooth-shaped figure. To the right of the emblem is the word "IMICRYL" in bold, blue letters, with a registered trademark symbol. Below the logo is the text "DIŞ MALZEMELERI SAN. VE TIC A.Ş."
# DEVICE DESCRIPTION:
The IMICRYL family of silicone impression materials: includes the brand names SPIRIAS, NICETY, REFLECT and PE Sil. It is used by dentists to obtain anatomical data of the patient's mouth and then to obtain a useful plaster cast to diagnose problems, identify necessary interventions and/or check their effectiveness. They are elastomeric materials with hydrophilic properties, high tear strength, dimensional accuracy and resistance to permanent deformation, cured by addition reaction.
The IMICRYL family of silicone impression materials consists of six different viscosities (putty soft, heavy body, medium/monophase/regular body, light body, extra light body, bite registration) for various application systems. Putty Soft consists of 1:1 250 and 1:1 300 mL jars, other consistencies 1:1 50 mL syringes. Although the consistency and composition of medium, monophase and regular body products, which have type 2 consistency, are exactly the same, they are offered for sale under three different names.
IMICRYL impression materials meet the requirements of ISO 4823:2021 Dentistry- Elastomeric impression and bite registration materials standard.
## SUBSTANTIAL EQUIVALENCE:
The applicant device has the same intended use as the 510(k) cleared predicates listed above. Table 1 below shows a comparison of IMICRYL Impression Materials and the predicates.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for IMICRYL. The logo consists of the letters "IMI" inside of an oval shape on the left, followed by the word "IMICRYL" in large, bold, blue letters. Below the logo is the text "DİŞ MALZEMELERİ SAN. VE TİC A.Ş." in smaller, black letters.
## Table 1: Comparison with Predicate Device
| DESCRIPTIVE | NEW DEVICE | PREDICATE DEVICE | REMARK |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| INFORMATION | IMICRYL Impression Materials (K213890) | HySil Impression Materials (K170736) | |
| INDICATIONS FOR USE | | | |
| | Putty Soft<br>SPIRIAS, NICETY, REFLECT and PE Sil Putty Soft is to be used as<br>preliminary materials for:<br>Two-step Putty-wash impression technique<br>One-step Putty-wash impression technique | Putty<br>HySil Putty is to be used as preliminary materials for:<br>Two-step Putty-wash impression technique<br>One-step Putty-wash impression technique | Same Indications for Use |
| | Heavy Body<br>SPIRIAS, NICETY, REFLECT and PE Sil Heavy Bodied materials<br>for:<br>One-step impression technique (simultaneous technique)<br>using single or dual viscosities<br>Two-step impression technique using dual viscosities<br>Functional impressions | Heavy Body<br>HySil Heavy is to be used as heavy-bodied materials for:<br>One-step impression technique (simultaneous technique)<br>using single or dual viscosities<br>Two-step impression technique using dual viscosities<br>Functional impressions | |
| | Medium/Monophase/Regular Body<br>SPIRIAS, NICETY, REFLECT and PE Sil<br>Medium/Monophase/Regular Bodies tray or syringeable<br>impression material for:<br>_ Taking impressions over fixed/removable restorations and<br>implants (i.e., transferring impression posts and bridge<br>components)<br>_ Functional impressions<br>_ Fabricating crown and bridgework or inlays<br>_ Fabricating full or partial dentures<br>_ Reline impressions<br>_ Use in the simultaneous mixing technique as well as the<br>putty-wash and triple tray techniques<br>_ Transferring root posts when fabricating posts and cores<br>indirectly | Medium Body<br>HySil Mono is to be used as a medium-bodied tray or<br>syringeable impression material for:<br>_ Taking impressions over fixed/removable restorations<br>and implants (i.e., transferring impression posts and bridge<br>components)<br>_ Functional impressions<br>_ Fabricating crown and bridgework or inlays<br>_ Fabricating full or partial dentures<br>_ Reline impressions<br>_ Use in the simultaneous mixing technique as well as the<br>putty-wash and triple tray techniques<br>_ Transferring root posts when fabricating posts and cores<br>indirectly | |
| | Light Body<br>SPIRIAS, NICETY, REFLECT and PE Sil Light Body is to be used as<br>syringeable impression materials for:<br>Two-step putty-wash impression technique | Light Body<br>HySil Light is to be used as syringeable impression<br>materials for:<br>Two-step putty-wash impression technique | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for IMICRYL diş malzemeleri san. ve tic a.ş. The logo features the text "IMICRYL" in a bold, sans-serif font, with the trademark symbol to the right of the word. To the left of the text is a circular graphic with the letters "IMI" inside. Below the text is the phrase "diş malzemeleri san. ve tic a.ş." in a smaller font.
| | One-step putty-wash impression technique<br>Two-step impression technique using dual viscosities<br>Reline impressions<br>• Fabricating full or partial dentures | | One-step putty-wash impression technique<br>Two-step impression technique using dual viscosities<br>Reline impressions<br>• Fabricating full or partial dentures | | | | | | | |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|--------------------------------------------------------------------------------------------------------------|---------------------|---------------|---------------------|---------------|---------------------|
| | Extra Light Body<br>SPIRIAS, NICETY, REFLECT and PE Sil Extra Light Body is to be<br>used as syringeable impression materials for:<br>Two-step putty-wash impression technique<br>One-step putty-wash impression technique<br>Two-step impression technique using dual viscosities<br>Reline impressions<br>Fabricating full or partial dentures | | Extra Light Body<br>HySil Extra Light is to be used as syringeable impression<br>materials for:<br>Two-step putty-wash impression technique<br>One-step putty-wash impression technique<br>Two-step impression technique using dual viscosities<br>Reline impressions<br>Fabricating full or partial dentures | | | | | | | |
| | Bite Registration<br>SPIRIAS, NICETY, REFLECT and PE Sil Bite Registration is used<br>for impression as below.<br>Taking occlusal surfaces<br>Confirming occlusal surfaces<br>Recording after putting the articulator | | Bite<br>HySil Bite is used for impression as below.<br>Taking occlusal surfaces<br>Confirming occlusal surfaces<br>Recording after putting the articulator | | | | | | | |
| COMPOSITION | Putty Soft, Heavy Body, Medium/Monophase/Regular Body,<br>Light Body, Extra Light Body, Bite Registration<br>Vinylpolysiloxane | | Putty, Heavy, Mono, Light,<br>Extra-Light, Bite<br>Vinylpolysiloxane | | Composed with same affiliated<br>material, but ratios of each<br>component in use are slightly<br>different. | | | | | |
| PHYSICAL PROPERTIES | Putty Soft | Heavy Body | Medium /<br>Monophase / Regular<br>Body | Light Body | Extra Light | | | | | |
| | New<br>Device | Predicate<br>Device | New<br>Device | Predicate<br>Device | New<br>Device | Predicate<br>Device | New<br>Device | Predicate<br>Device | New<br>Device | Predicate<br>Device |
| Type | Type 0 | Type 0 | Type 1 | Type 1 | Type 2 | Type 2 | Type 3 | Type 3 | Type 3 | Type 3 |
| Consistency | 28,40 mm | 31,27 mm | 30,18 mm | 28,47 mm | 34,35 mm | 35,65 mm | 36,67 mm | 45,60 mm | 43,42 mm | 44,61 mm |
| Working Time | 90 sec | - | 90 sec | 2,022 min | 90 sec | 2,612 min | 90 sec | 2,850 min | 90 sec | 3,6 min |
| Mixing Time | 30 sec | 38 sec | Auto<br>(30 sec) | Pass | Auto<br>(30 sec) | Pass | Auto (30 sec) | Pass | Auto (30 sec) | Pass |
| Detail Reproduction | 50μm | Pass | 20μm | Pass | 20μm | Pass | 20μm | Pass | 20μm | Pass |
| Compatibility with<br>gypsum | 50μm | Pass | 20μm | Pass | 20μm | Pass | 20μm | Pass | 20μm | Pass |
| Linear Dimensional<br>change | 0,01% | 0,031 % | 0,01% | 0,02 % | 0,02% | 0,02 % | 0,02% | 0,03 % | 0,02% | 0,027 % |
| Elastic Recovery | 99,00% | 99,8 % | 99,60% | 98,80 % | 99,99% | 98,60 % | 99,99% | 98,70 % | 99,99% | 99,5 % |
| Strain-in-Compression | 1,6% | 1,6 % | 2,30% | 2,30 % | 3,68% | 3,00 % | 3,38 % | 2,30 % | 3,18% | 2,6 % |
| PHYSICAL PROPERTIES | | | | | | | | | | |
| | | | Bite Registration | | | | | | | |
| New Device | | | Predicate Device | | | | | | | |
| Type | | Type B | Type B | | | | | | | |
| Mixing Time | | 30 sec | Pass | | | | | | | |
| Working Time | | 90 sec | Pass | | | | | | | |
| Time in Mouth (minimum) | | 3 min | 1 min. 30 sec. | | | | | | | |
| Linear Dimensional change | 0,02% | | -0,16% | | | | | | | |
| Compression set | | 0.06mm | Pass | | | | | | | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for IMICRYL. The logo consists of the word "IMICRYL" in large, bold, blue letters, with a registered trademark symbol next to it. To the left of the word is a blue circle containing the letters "IMI" in white. Below the word "IMICRYL" is the text "DIŞ MALZEMELERI SAN. VE TIC A.Ş." in smaller, blue letters.
50 HD
55 HD
## Similarities
Hardness
• IMICRYL Impression Materials; it is exactly similar to the in terms of composition, indications for use, function and physical properties.
• We believe that the prior use of these components in legally marketed devices and the performance data and results support the safety and effectiveness of IMICRYL Impression Materials for the intended use.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image is a logo for IMICRYL. The logo consists of a blue circle with a white tooth-like shape inside, with the letters "IMI" in blue inside the tooth. To the right of the circle is the word "IMICRYL" in blue, with a registered trademark symbol in the upper right corner. Below the logo is the text "DIŞ MALZEMELERI SAN. VE TIC A.Ş." in blue.
# NON-CLINICAL PERFORMANCE TESTING:
#### Biocompatibility Testing:
Cytotoxicity, irritation and sensitization tests of the device were performed. The device is biocompatible.
## Physical Testing:
In-vitro bench tests were performed on the IMICRYL Impression Materials according to the requirements in ISO 4823:2021 Dentistry-Elastomeric impression and bite registration materials.
Bench tests included in support of the substantial equivalence of IMICRYL Impression Materials are:
- Consistency
- Working Time
- Mixing Time .
- Detail Reproduction
- Compatibility with Gypsum
- . Linear Dimensional Change
- Elastic Recovery
- Strain-in-Compression
- Time in Mouth
- . Compression Set
- Hardness ●
# CONCLUSION
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 872.3660, and based on the information provided in this pre-market notification, IMICRYL DIS MALZEMELERI SANAYI VE TICARET A.S. concludes that the IMICRYL Impression Materials is safe, effective and substantially equivalent to the predicate devices as described herein. It does not introduce new indications for use, has similar technological characteristics and does not introduce new potential hazards or risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.