Carestation 750/750c
K213867 · Datex-Ohmeda, Inc. · BSZ · Mar 28, 2023 · Anesthesiology
Device Facts
| Record ID | K213867 |
| Device Name | Carestation 750/750c |
| Applicant | Datex-Ohmeda, Inc. |
| Product Code | BSZ · Anesthesiology |
| Decision Date | Mar 28, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 868.5160 |
| Device Class | Class 2 |
| Attributes | Therapeutic, Pediatric |
Intended Use
The Carestation750/750c anesthesia systems are intended to provide monitored anesthesia care, general inhalation anesthesia and/ or ventilatory support to a wide range of patients (neonatal, pediatric, and adult). The anesthesia systems are suitable for use in a patient environment, such as hospitals, surgical centers, or clinics. The systems are intended to be operated by a clinician qualified in the administration of general anesthesia.
Device Story
Anesthesia system providing inhalation anesthesia and ventilatory support; inputs include medical gases (O2, Air, N2O) and anesthetic agents (Isoflurane, Sevoflurane, Desflurane). Features electronic gas mixing and microprocessor-controlled ventilation (VCV, PCV, SIMV, PSVPro, PCV-VG, CPAP+PSV). Operated by clinicians in hospitals/clinics via 15-inch touchscreen interface. System monitors airway pressure, inspired O2, and exhaled volume; includes automated lung recruitment maneuvers. Output displayed as numerical digits and waveforms; assists clinicians in managing patient ventilation and gas delivery. Benefits include precise gas control, ergonomic mobility, and integrated patient monitoring to support safe anesthesia administration.
Clinical Evidence
No clinical data. Substantial equivalence established through non-clinical bench testing, including design verification, software validation, biocompatibility (ISO 18562), electrical safety (IEC 60601-1), and electromagnetic compatibility (IEC 60601-1-2).
Technological Characteristics
Anesthesia workstation with electronic gas mixer (O2, Air, N2O). Features 15-inch P-Cap touchscreen, microprocessor-controlled 7900 ventilator, and Selectatec vaporizer manifold. Connectivity includes integrated respiratory gas modules (CARESCAPE series). Materials evaluated per ISO 18562. Software is major level of concern. Complies with IEC 60601-1, IEC 60601-1-8, ISO 80601-2-13, and IEC 62304.
Indications for Use
Indicated for monitored anesthesia care, general inhalation anesthesia, and ventilatory support for neonatal, pediatric, and adult patients in hospitals, surgical centers, or clinics.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
Predicate Devices
- Carestation 620/650/650c (K151570)
Reference Devices
- Avance CS2 (K131945)
- Datex-Ohmeda Aestiva (K000706)
- Aespire View (K122445)
- Aisys CS2 (K170872)
- GE Monitor Series B650 (K102239)
Related Devices
- K123125 — GE DATEX-OHMEDA AVANCE CS2 · Datex-Ohmeda, Inc. · Feb 6, 2013
- K092864 — AESPIRE 7900, AESPIRE VIEW · Datex-Ohmeda Inc., A Division of Instrumentarium C · Jan 12, 2010
- K132530 — GE DATEX-OHMEDA AISYS CS2 (AISYS CS^2) · Datex-Ohmeda, Inc. · Dec 20, 2013
- K110213 — GE DATEX-OHMEDA AISYS · Datex-Ohmeda, Inc. · Jul 20, 2011
- K170872 — Aisys CS2 · Datex-Ohmeda, Inc. · Aug 15, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the words "U.S. Food & Drug Administration" in a smaller font size.
March 28, 2023
Datex-Ohmeda, Inc. Kimberly Mangum Regulatory Affairs Director 3030 Ohmeda Drive, PO Box 7550 Madison, Wisconsin 53707-7550
Re: K213867
Trade/Device Name: Carestation 750/750c Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine For Anesthesia Or Analgesia Regulatory Class: Class II Product Code: BSZ Dated: February 17, 2023 Received: February 22, 2023
Dear Kimberly Mangum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee -S
James J. Lee, Ph.D. Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K213867
Device Name Carestation 750/750c
### Indications for Use (Describe)
The Carestation750/750c anesthesia systems are intended to provide monitored anesthesia care, general inhalation anesthesia and/ or ventilatory support to a wide range of patients (neonatal, pediatric, and adult). The anesthesia systems are suitable for use in a patient environment, such as hospitals, surgical centers, or clinics. The systems are intended to be operated by a clinician qualified in the administration of general anesthesia.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of the GE monogram on the left and the words "GE HealthCare" on the right. The monogram is a stylized version of the letters "GE" inside of a circle. The text is in a simple, sans-serif font.
## Premarket Notification 510(k) Summary
As required by section 807.92 Carestation 750/750c GENERAL COMPANY INFORMATION as required by 807.92(a)(1) K213867
#### I. Submitter
### Company Name/Address/Phone/Fax:
Mailing Address: Datex-Ohmeda, Inc. 3030 Ohmeda Drive PO Box 7550 Madison, WI 53707-7550 USA
Physical Address: Datex-Ohmeda, Inc. 3030 Ohmeda Drive Madison, WI 53718 USA Tel: 608-334-4281
### Name of Contact:
Kimberly Mangum Regulatory Affairs Director - Anesthesia and Respiratory Care (ARC) Patient Care Solutions - GE HealthCare Kimberly.Mangum@ge.com Phone: 267-400-5180 www.gehealthcare.com
### Date Prepared: March 24, 2023
#### II. DEVICE
Name of Device: Carestation 750/750c Common or Usual Name: Gas Machine, Anesthesia (21 CFR 868.5160) Regulatory Class: II Product Code: BSZ
#### III. PREDICATE DEVICE
Primary Predicate: Carestation 620/650/650c, K151570 Reference Device: Avance CS2, K131945
#### DEVICE DESCRIPTION IV.
The GE Carestation 750/750c anesthesia machines (Carestation 750 series) are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonate, pediatric, and adult). The
{4}------------------------------------------------
anesthesia systems are suitable for use in a patient environment such as hospitals, surgical centers, or clinics. They represent one of the systems in a long line of products based on the Datex-Ohmeda Aestiva (K000706), Aespire View (K122445), Aisys CS2 (K170872), Avance CS2 (K131945), Carestation 620/650/650c (Carestation 600 series) (K151570) Anesthesia Systems. The Carestation 750 series anesthesia systems are intended to be operated by a clinician qualified in the administration of general anesthesia.
The Carestation™ 750/750c anesthesia systems combine advanced anesthesia delivery, patient monitoring. and care information management. The contemporary, compact design allows for easy mobility and addresses many ergonomic considerations including an effective cable management solution, aesthetic covers, and an expandable work surface area. Optional integrated features include auxiliary common gas outlet, auxiliary 02 outlet, auxiliary 02+Air outlet, suction control, and respiratory gas monitoring. The system provides integration of ventilation and gas delivery on a 15-inch color graphical touchscreen interface. This system also features electronic gas mixing of oxygen and a balance gas of either N2O or Air. The Carestation 750 series represents one of the systems in a long line of products based on the Datex-Ohmeda Aestiva (K000706), AespireView (K122445), Aisys CS2 (K170872), Avance CS2 (K131945), and Carestation 600 Series (K151570) Anesthesia Systems.
This anesthesia system is designed for mixing and delivering inhalation anesthetics (Isoflurane, Sevoflurane, or Desflurane), Air, O2, and N2O. This anesthesia system has Recruitment maneuvers, a feature to perform automated lung recruitment maneuver in a single step or in multi steps.
This anesthesia system uses electronic flow valve ventilation technology offering Volume Control Ventilation with tidal volume compensation and electronic PEEP. This technology also features Pressure Control Ventilation, optional Pressure Support Ventilation with an Apnea Backup (PSVPro™) that is used for spontaneously breathing patients, Synchronized Intermittent
Mandatory Ventilation (SIMV) modes, Pressure Control Ventilation-Volume Guarantee (PCVG), Continuous Positive Airway Pressure + Pressure Support Ventilation (CPAP + PSV), and VCV Cardiac Bypass. In Volume Control Ventilation, a patient can be ventilated using a minimal tidal volume of 20 ml. In Pressure Control Ventilation, volumes as low as 5 ml can be measured. These advanced features allow for the ventilation of a broad patient range. The device includes the following basic components:
The Carestation 750 series anesthesia systems supply set flows of medical gases to the breathing system using an electronic gas mixer (O2 with Air or O2 with N2O). Gas flows are adjusted by the user on the touchscreen, the flows are displayed on the system graphical user interface assembly as numerical digits and as electronic representations of flow meters. The system provides an option for auxiliary mixed Oxygen + Air flow delivery where 02 with Air are blended and delivered to an auxiliary port used to support spontaneously breathing patients using a nasal cannula. An optional auxiliary O2 supply includes a separate O2 flow tube and needle valve flow control that delivers O2 flow to an auxiliary port used to support spontaneously breathing patients using a nasal cannula. The gas flow from the optional auxiliary O2 subsystem does not flow through the electronic gas mixer.
The Carestation 750 series models include up to 3 breathing gases with O2 and Air as standard, and N20 as an optional breathing gas. The systems include two vaporizer positions compatible with, Isoflurane, Sevotlurane, and Desflurane vaporizers. The Carestation 750 is available with up to three back-up gas cylinder connections. The Carestation 750 series systems are also available in pendant (Carestation 750c) models.
The system uses touchscreen technology, hard keys, and a Comwheel to access system functions, menus, and settings on a 15'' color graphical user interface assembly (aka display). The graphical user interface
{5}------------------------------------------------
assembly is mounted on an arm on the left side of the machine. It can be rotated via the arm toward, or away from, the system to adjust the horizontal position. The arm is available allowing the display to be tilted up or down to adjust the vertical viewing angle or be tilted left or right to adjust the horizontal position of the display. An optional arm can be raised/lowered and rotated 360 degrees. A split screen field can be set to show gas trends, Spirometry loops, a Paw gauge, airway compliance, and optional ecoFlow information. If none is selected, the waveforms expand to fill the split screen area.
The Carestation 750 series systems accept Tec 7, and Tec 820/850 series vaporizers on a 2position Selectatec manifold. Safety features and devices within the systems are designed to decrease the risk of hypoxic mixtures, multiple anesthetic agent mixtures, complete power failure, or sudden gas supply The Carestation 750 series systems are available with optional integrated respiratory gas failures. monitoring which can be physically integrated into the system, receive electronic power from the Carestation 750/750c, and communicate measured values to the Carestation 750/750c for display on the system graphical user interface assembly. When supplied as an option, integrated respiratory gas monitoring is provided via the GE CARESCAPE series (EsCAiO or E-sCAiOV) respiratory airway modules (GE Healthcare Finland Oy, CE 0537) which is identical to the module used on Avance CS2.
The Anesthesia Ventilator used in the Carestation 750 series is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. This version of the GE 7900 ventilator is equipped with a built-in system for monitoring inspired oxygen (using an optional 02 cell or optional integrated gas module), patient airway pressure, and exhaled volume. Flow sensors in the breathing circuit are used to monitor and control patient ventilation.
This allows for the compensation of gas and tubing compression losses, fresh gas contribution, and small gas leakage from the breathing absorber, bellows, and pneumatic system connections. User settings and microprocessor calculations control breathing patterns. The user interface keeps ventilation settings in memory. The user may change settings with a simple ventilation parameter setting sequence. A bellows contains breathing gasses to be delivered to the patient and provides a barrier keeping patient gas separate from the ventilatory drive gas. Positive End Expiratory Pressure (PEEP) is regulated electronically. Positive pressure is maintained in the breathing system so that occurs is outward from the patent breathing circuit.
This ventilator comes with a standard ventilation mode as well as optional ventilation modes.
### Standard ventilation modes:
- VCV (Time Cycled, Volume Controlled Ventilation) .
- . PCV (Time Cycled, Pressure Control Ventilation)
### Optional ventilation modes:
- VCV-SIMV (Synchronized Intermittent Mandatory Ventilation Volume Control)
- . PCV-SIMV (Synchronized Intermittent Mandatory Ventilation Pressure Control)
- PSVPro (Pressure supported ventilation with apnea backup)
- . PCV-VG (Pressure Controlled ventilation - Volume Guaranteed)
- . PCV-VG-SIMV (Synchronized Intermittent Mandatory Ventilation, Pressure Controlled ventilation - Volume Guaranteed)
- . CPAP+PSV (Continuous Positive Airway Pressure/Pressure Support)
The system can include an internal, factory installed, suction regulator and control visible from the front of the machine. It can mount different monitors using an arm or shelf mounts. The mounting is achieved through a combination of GE Healthcare adapters and other third-party mounts, including one that allows
{6}------------------------------------------------
for the physical integration of the GE Monitor Series B650 (K102239). The Carestation 750 system also includes an optional cable management solution, which can help user to manage the various cables attached to the system.
#### V. INDICATIONS FOR USE
The Carestation750/750c anesthesia systems are intended to provide monitored anesthesia care, general inhalation anesthesia and/ or ventilatory support to a wide range of patients (neonatal, pediatric, and adult). The anesthesia systems are suitable for use in a patient environment, such as hospitals, surgical centers, or clinics. The systems are intended to be operated by a clinician qualified in the administration of general anesthesia.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Carestation 750 series is primarily based on the predicate device Carestation 600 series feature set (K151570) and contains identical or comparable hardware and software components. A comparison of the CS750 and predicate device technological characteristics is provided as Table 1 below.
{7}------------------------------------------------
| Specification | Proposed Device Carestation<br>750/750c | GE Datex-Ohmeda Predicate<br>Device<br>Carestation 620/650/650c | Reference Device Avance<br>CS2 (K131945) |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pneumatic System | Supplies set flows of medical gases to the breathing<br>system using electronic controlled gas mixing of 2<br>gasses (O2 with Air or O2 with N2O). | Supplies set flows of medical gases to<br>the breathing system using mechanical<br>controlled gas mixing of 2 gasses (O2<br>with Air or O2 with N2O) or 3 gases<br>(O2, Air, and N2O). | Supplies set flows of medical<br>gases to the breathing system<br>using electronic controlled<br>gas mixing of 2 gasses (O2<br>with Air or O2 with N2O). |
| Gas Supply | The gas mixing subsystem is a two gas mixer<br>providing a mixture of O2 and Air or O2 and N2O<br>(100% O2 or 100% Air (21% O2) can also be<br>selected). | The gas mixing subsystem is a three gas<br>mixer providing a mixture of O2 and<br>Air or O2 and N2O or O2, Air and N2O<br>(100% O2 or 100% Air (21% O2) can<br>also be selected). | The gas mixing subsystem is<br>a two gas mixer providing a<br>mixture of O2 and Air or O2<br>and<br>N2O (100% O2 or 100% Air<br>(21% O2) can also be<br>selected). |
| O2 Flow | A pressurized gas source (pipeline, or regulated<br>cylinder), supplies O2 directly to the gas mixing<br>subsystem. | A secondary pressure regulator in the<br>gas mixing subsystem regulates<br>constant O2 gas pressure and supplies it<br>to the O2 flow control valve. | A pressurized gas source<br>(pipeline, or regulated<br>cylinder), supplies O2 directly<br>to the gas mixing subsystem. |
| Air and N2O flow | Air gas flows from a pressurized gas source<br>(pipeline or regulated cylinder) to the Air/N2O<br>selector valve on the gas mixing subassembly. | Air gas flows from a pressurized gas<br>source (pipeline or regulated cylinder)<br>to the Air selector valve or a secondary<br>pressure regulator (optional) on the gas<br>mixing subassembly. | Air gas flows from a<br>pressurized gas source<br>(pipeline or regulated<br>cylinder) to the Air/N2O<br>selector valve on the gas<br>mixing subassembly. |
| Gas Mixer | Digital display and control of flows using an<br>electronic mass air flow sensor to provide<br>electronic mixing and proportioning of gases (O2<br>and Air or N2O). | Digital display of flows with traditional<br>needle valve flow control and flow head<br>mixing and proportioning of gases (Air,<br>O2, N2O). | Digital display and control of<br>flows using an electronic<br>mass air flow sensor to<br>provide electronic mixing and<br>proportioning of gases (O2<br>and Air or N2O). |
| Table 1: Carestation 750/750c Technological Characteristics Comparison | |
|------------------------------------------------------------------------|--|
|------------------------------------------------------------------------|--|
{8}------------------------------------------------
| Specification | Proposed Device Carestation<br>750/750c | GE Datex-Ohmeda Predicate<br>Device<br>Carestation 620/650/650c | Reference Device Avance<br>CS2 (K131945) |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aux O2 option | An auxiliary O2 gas supply to provide O2 gas to<br>the patient through internal connections and<br>control with an auxiliary O2 flowmeter. | An auxiliary O2 gas supply to provide<br>O2 gas to the patient through internal<br>connections and control with an<br>auxiliary O2 flowmeter. | Not a comparison point |
| Aux O2+Air gas<br>delivery option | The Aux O2+Air gas delivery option provides an<br>auxiliary O2+Air gas mixture to the patient<br>through internal connections and control with the<br>O2 and Air fresh gas controls. | The Aux O2+Air gas delivery option<br>provides an auxiliary O2+Air gas<br>mixture to the patient through internal<br>connections and control with the O2 and<br>Air fresh gas controls. | Not a comparison point |
| ACGO | Use the optional Auxiliary Common Gas Outlet<br>(ACGO) switch to direct the fresh gas flow through<br>the ACGO port on the front of the system. | Use the optional Auxiliary Common<br>Gas Outlet (ACGO) switch to direct the<br>fresh gas flow through the ACGO port<br>on the front of the system. | Not a comparison point |
| O2 Flush | The O2 flush button supplies a high flow of O2 to<br>the breathing system. | The O2 flush button supplies a high<br>flow of O2 to the breathing system. | Not a comparison point |
| ALT 02 | The Alternate O2 control is not an auxiliary<br>source of O2. The Alternate O2 is intended to be<br>used when in certain failures, such as the<br>electronic gas mixer is not available or the display<br>screen has failed. | None | ALT O2 as the auxiliary<br>source of O2 for electronic<br>gas mixer |
| Vaporizers<br>Compatibility | Carestation 750 series provide a two position<br>vaporizer manifold to support up to 2 active<br>positions for Tec series GE vaporizers. | Carestation 620/650/650c provides a<br>two position vaporizer manifold to<br>support up to 2 active positions for Tec<br>series GE vaporizers. | Not a comparison point |
| Specification | Proposed Device Carestation<br>750/750c | GE Datex-Ohmeda Predicate<br>Device<br>Carestation 620/650/650c | Reference Device Avance<br>CS² (K131945) |
| Respiratory<br>Gas Monitors<br>Compatibility | Airway Gas Option is:<br>CARESCAPE series: E-sCAiO and E-<br>sCAiOV.<br>The CAiOV feature set of the<br>CARESCAPE series E-sCAiOVX and EsCAiOVE<br>Airway Modules | Airway Gas Option is :<br>CARESCAPE series: E-sCAiO and<br>E-sCAiOV.<br>The CAiOV feature set of the<br>CARESCAPE series E-sCAiOVX<br>and E-sCAiOVE Airway Modules<br>N-CAiO (without Spirometry<br>connector) | Airway Gas Option is:<br>CARESCAPE series:<br>EsCAiO and E-sCAiOV.<br>The CAiOV feature set of the<br>CARESCAPE series<br>EsCAiOVX and E-sCAiOVE<br>Airway Modules |
| Optional<br>Integrated<br>Suction<br>Regulation | The vacuum suction regulator uses an external<br>vacuum supply. The venturi suction regulator uses<br>the system air supply source. | The vacuum suction regulator uses an<br>external vacuum supply. The venturi<br>suction regulator uses the system air<br>supply source. | Not a comparison point |
| Optional Gas<br>Scavenging | Yes | Yes | Not a comparison point |
| Breathing System | The integrated circle breathing system can be used<br>for both manual (bag) and mechanical (automatic)<br>patient ventilation. The breathing system includes<br>bellows assembly, APL valve, bag-to-vent switch,<br>CO2 bypass assembly. | The integrated circle breathing system<br>can be used for both manual (bag) and<br>mechanical (automatic) patient<br>ventilation. The breathing system<br>includes bellows assembly, APL valve,<br>bag-to-vent switch, CO2 bypass<br>assembly. | Not a comparison point |
| Flow Sensors | Variable orifice flow sensor (autoclavable) with<br>pneumatic and electrical connections located in the<br>inspiratory outlet and expiratory inlet inside the<br>breathing system. The outer chamber of the sensor<br>is heated. | Variable orifice flow sensor<br>(autoclavable) with pneumatic and<br>electrical connections located in the<br>inspiratory outlet and expiratory inlet<br>inside the breathing system. The outer<br>chamber of the sensor is heated. | Not a comparison point |
| CO2 Absorber<br>Canister | CO2 absorber canister absorbs CO2 in the gas path<br>from bellows. | CO2 absorber canister absorbs CO2 in<br>the gas path from bellows. | Not a comparison point |
| Specification | Proposed Device Carestation<br>750/750c | GE Datex-Ohmeda Predicate<br>Device<br>Carestation 620/650/650c | Reference Device Avance<br>CS² (K131945) |
| Ventilator Engine | The ventilator engine is located in the middle pan<br>of the system, under the user work surface. A<br>precision flow servo system control gas flow to<br>the patient. | The ventilator engine is located in the<br>middle pan of the system, under the<br>user work surface. A precision flow<br>servo system control gas flow to the<br>patient. | Not a comparison point |
| System<br>Mountings | Carestation 750 is a trolley mounted configuration<br>while the Carestation 750c can be configured and<br>installed with a pendant mounted version. | Carestation 650 and Carestation 620 are<br>a trolley mounted configuration while<br>the Carestation 650c can be configured<br>and installed with either a pendant or<br>wall mounted version. | Not a comparison point |
| Cable<br>Management | Clip and cubby are to fix cables, rear cover could<br>shield hoses and cables. | Clips are to fix cables. | Not a comparison point |
| Electrical<br>Subsystem | The electrical system consists of two main<br>computing units: the Graphical user interface<br>assembly (previously known as the display unit)<br>and the Anesthesia Computer Board (ACB).<br>The other key electrical boards are Power<br>Management Board, Sensor Interface Board, and<br>Mixer Interface Board. | The electrical system consists of two<br>main computing units: the Graphical<br>user interface assembly (previously<br>known as the display unit) and the<br>Anesthesia Computer Board (ACB).<br>The other key electrical boards are<br>Power Management Board, Sensor<br>Interface Board, and Frame Interface<br>Board. | Mixer Interface Board to<br>support electronic mixer |
| Touch Screen | Projective Capacitive (P-Cap) touch screen detects<br>touch by measuring the capacitance at each<br>addressable electrode.<br><br>P-Cap touch screen software driver. | Surface acoustic wave (SAW) detects<br>touch by measuring SAW on the route<br>SAW P-Cap touch screen software<br>driver. | Not a comparison point |
| Specification | Proposed Device Carestation<br>750/750c | GE Datex-Ohmeda Predicate<br>Device<br>Carestation 620/650/650c | Reference Device Avance<br>CS2 (K131945) |
| Software<br>Subsystem | The software provides user interface, power<br>management, and controls the delivery of<br>therapy to patients. As the segregation strategy,<br>the Carestation 750 series software is comprised<br>of below top-level Software Items/Units.<br>Display Computer/Display Unit Power Supply Controller/Power<br>Subsystem Front Panel Controller Anesthesia Control Board/Anesthesia Computer Mixer Interface Board Subsystem Sensor Interface Board Subsystem Mixer | The software provides user<br>interface, power management, and<br>controls the delivery of therapy to<br>patients. As the segregation<br>strategy, the Carestation 600 series<br>software is comprised of below top-<br>level Software Items/Units. Display Computer/Display Unit Power Supply Controller/Power<br>Subsystem Front Panel Controller Anesthesia Control Board/Anesthesia Computer Frame Interface Board Subsystem Sensor Interface Board Subsystem | Not a comparison point |
| ecoFLOW | ecoFLOW option within the split screen that<br>shows a visual indicator reference for the total gas<br>flows, agent usage, cost and approximate<br>minimum O2 flow needed to maintain the<br>clinician's desired inspired O2 concentration.<br>Carestation 750 series flows are limited to only O2,<br>Air or N2O. | ecoFLOW option within the split screen<br>that shows a visual indicator reference<br>for the total gas flows, agent usage, cost<br>and approximate minimum O2 flow<br>needed to maintain the clinician's<br>desired inspired O2 concentration.<br>Carestation 620/650/650c allows all 3<br>gases to be present due to traditional<br>needle valve flow control. | ecoFLOW option within the<br>split screen that shows a<br>visual indicator reference for<br>the total gas flows, agent<br>usage, cost and approximate<br>minimum O2 flow needed to<br>maintain the clinician's<br>desired inspired O2<br>concentration.<br>Avance CS2 flows are limited<br>to only O2, Air or N2O. |
| User Interface | Color display with Com Wheel or touch activation.<br>Quick keys provide quick activation access via the<br>touch screen user interface. Seven quick keys<br>available for ventilation and three quick keys<br>available for gas. | Color display with Com Wheel or touch<br>activation. Quick keys provide quick<br>activation access via the touch screen<br>user interface. Seven quick keys<br>available for ventilation. | Not a comparison point |
| Specification | Proposed Device Carestation<br>750/750c | GE Datex-Ohmeda Predicate<br>Device<br>Carestation 620/650/650c | Reference Device Avance<br>CS2 (K131945) |
| Operating System | The graphical user interface operating system on<br>Carestation 750 series is WEC 2013 (<br>Windows). | The graphical user interface operating<br>system on Carestation 600 series is CE6<br>(Windows). | Not a comparison point |
| Display Arm | User Interface is mounted on an arm on the left<br>side of the machine. It can be rotated left or right<br>via the arm toward or away from the system to<br>adjust the horizontal position. Both standard arm<br>and premium arm can be tilted up or down to<br>adjust the vertical angle or be tilted left or right to<br>adjust the horizontal angle of the User Interface.<br>Premium arm can also be raised/lowered and<br>rotated 360 degrees. | User Interface is mounted on an arm on<br>the left side of the machine. It can be<br>rotated left or right via the arm toward<br>or away from the system to adjust the<br>horizontal position. For the Carestation<br>650 and 650c it can also be tilted up or<br>down to adjust the vertical angle or be<br>tilted left or right to adjust the<br>horizontal angle of the User Interface.<br>The Carestation 620 uses a basic arm<br>that allows the User Interface to move<br>left and right but does not tilt. | Not a comparison point |
| HDX GUI<br>Style | Added Modern Interface Style (HDX), user can<br>enable in Super User System Config menu. | Classic GUI Interface Style on<br>Carestation 600. | Not a comparison point |
| Super user and<br>service mode<br>passwords | User could change super user password and service<br>password. New Password Entry item will be<br>restricted to a 6-22 digit password.<br>User could set new password to protect the device<br>with only authorized personnel entering the Super<br>User mode and Service mode. | The different 5 digit passwords of Super<br>user and Service mode are fixed, user<br>could not change the passwords. | Not a comparison point |
| Specification | Proposed Device Carestation 750/750c | GE Datex-Ohmeda Predicate Device Carestation 620/650/650c | Reference Device Avance CS² (K131945) |
| Recruitment<br>Maneuver | Use the Single Step recruitment maneuver to<br>deliver a pressure breath for a set time. | Use the Vital Capacity procedure to<br>deliver a pressure breath for a set time. | Not a comparison point |
| | Use the Multi Step recruitment maneuver to deliver<br>pressure breaths through a series of ventilation<br>steps. | Use the Cycling procedure to deliver<br>pressure breaths through a series of<br>ventilation steps. | Not a comparison point |
| Direct access to<br>important functions | A button bar is at the right side of the user<br>graphical interface to allow direct user access to 10<br>important functions:<br>Audio Pause, Alarm Setup, Auto Limits,<br>System Setup, Next page, Trends, Recruitment<br>Maneuver, Pause Gas Flow, Timer function , End<br>case. | A button bar is at the right side of the<br>user graphical interface to allow direct<br>user access to 10 important functions:<br>Audio Pause, Alarm Setup, Auto<br>Limits, System Setup, Next page,<br>Trends, Spirometry, Procedures, Timer<br>function, End case. | Not a comparison point |
| Recourse & Recovery | When Display Computer (DC) software<br>error occurs a processor reset will initiate the DC<br>reboot. However, the Anesthesia Computer (AC)<br>will keep working by using the current parameters,<br>when the system reboot successfully it will go to<br>Therapy mode directly by using setting values<br>recovered by Anesthesia<br>Computer software. DISPLAY<br>COMPUTER RESTARTED alarm will be<br>displayed on screen. | None | Not a comparison point |
| Checkout<br>failure<br>troubleshooting<br>message panel | If checkout tests failed, it will display a<br>troubleshooting message panel with problem<br>statement and step by step messages to instruct user<br>to remove the failure. | The preoperative checkout and<br>calibration are digitally shown and<br>guided according to the menu. | Not a comparison point |
| DatexOhmeda COM<br>changes | DatexOhmeda COM exists on<br>Carestation 700 (updates made) | DatexOhmeda COM exists on<br>Carestation 600. | Not a comparison point |
| Specification | Proposed Device Carestation<br>750/750c | GE Datex-Ohmeda Predicate<br>Device<br>Carestation 620/650/650c | Reference Device Avance<br>CS2 (K131945) |
| New Mixer and<br>Alternate O2<br>checkout test | Electronic Mixer and Alternate O2 checkout | None | Electronic Mixer and<br>Alternate O2 checkout |
| Gas Setup and<br>Preset menu,<br>Alt O2, Air Only and<br>related Alarms | Electronic mixer, alternate O2, and Air only mode | None | Electronic mixer, alternate<br>O2, and Air only mode |
| Remove latch alarm | The latched alarms could be removed because<br>alarm log history could be used instead. | Latched alarms exist as a feature in<br>Carestation 600 series. | Not a comparison point |
| Principles of<br>Operation:<br>Control and<br>Phase<br>Variables | Anesthesia machine settings are separated into five<br>categories:<br>Gas Flow (Electronic gas mixer) and Breathing Circuit Controls Main Ventilation Parameters Breath Timing Patient Ventilation Synchrony Safety | Anesthesia machine settings are<br>separated into five categories:<br>Gas Flow (Mechanical gas mixer) and Breathing Circuit Controls Main Ventilation Parameters Breath Timing Patient Ventilation Synchrony Safety | Gas Flow (Electronic gas<br>mixer) |
| Ventilation<br>Modes | VCV, PCV, VCV-SIMV,<br>PCV-SIMV, PSVPro,<br>PCV-VG, PCV-VG-SIMV,<br>CPAP-PSV | VCV, PCV, VCV-SIMV,<br>PCV-SIMV, PSVPro,<br>PCV-VG, PCV-VG-SIMV, CPAP-<br>PSV | Not a comparison point |
| Safety<br>Standards | Compliance with applicable FDA recognized<br>consensus standards…