K213803 · Prosidyan, Inc. · MQV · Aug 23, 2022 · Orthopedic
Device Facts
Record ID
K213803
Device Name
FIBERGRAFT Aeridyan Matrix Bone Graft Substitute
Applicant
Prosidyan, Inc.
Product Code
MQV · Orthopedic
Decision Date
Aug 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FIBERGRAFT® Aeridyan™ Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® Aeridyan™ Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® Aeridyan™ Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine. FIBERGRAFT® Aeridyan™ Matrix must be hydrated with saline or blood for pelvis and extremity applications.
Device Story
FIBERGRAFT® Aeridyan™ Matrix is a resorbable, osteoconductive bone void filler; composed of 45S5 bioactive glass granules, boron bioactive glass microspheres, and bovine type I collagen. Device supplied as a matrix; hydrated with saline, blood, or bone marrow aspirate (BMA) by clinician; molded into shape; packed into bony voids or gaps in the skeletal system. In posterolateral spine, used with autograft and BMA; in pelvis/extremities, hydrated with saline or blood. Acts as a scaffold; resorbs over time; replaced by host bone during healing. Benefits patient by filling osseous defects; promoting bone regeneration. Used in clinical/surgical settings by orthopedic surgeons.
Clinical Evidence
No human clinical data. Evidence includes bench testing (physical/chemical properties, biocompatibility per ISO 10993, endotoxin testing, shelf-life/packaging) and two animal studies. Rabbit posterolateral spine study (n=71) compared device to predicate/controls over 26 weeks using radiographic, histomorphometric, and biomechanical data. Rabbit femoral condyle defect study (n=44) compared device to predicates/references over 26 weeks using radiographic, histological, and histomorphometric data. Results demonstrated substantial equivalence in performance.
Technological Characteristics
Composition: 45S5 bioactive glass, boron bioactive glass, bovine type I collagen. Form: Non-setting putty/matrix. Bioactive, osteoconductive, resorbable. Sterile. Packaging: Thermo-form tray. No software or electronic components.
Indications for Use
Indicated for patients with surgically created osseous defects or osseous defects from traumatic injury to the bone. Used for bony voids or gaps not intrinsic to skeletal stability in the posterolateral spine, extremities, and pelvis.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
August 23, 2022
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are simple and professional, and they are likely used to identify the organizations on official documents and websites.
Prosidyan, Inc % Janice M. Hogan Regulatory Counsel Hogan Lovells US LLP 1735 Market Street, Suite 2300 Philadelphia, Pennsylvania 19103
Re: K213803
Trade/Device Name: FIBERGRAFT® Aeridyan™ Matrix Bone Graft Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: July 19, 2022 Received: July 19, 2022
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement on last page
510(k) Number (if known) K213803
Device Name
FIBERGRAFT® Aeridyan™ Matrix Bone Graft Substitute
Indications for Use (Describe)
FIBERGRAFT® Aeridyan™ Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® Aeridyan™ Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® Aeridyan™ Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine. FIBERGRAFT® Aeridyan™ Matrix must be hydrated with saline or blood for pelvis and extremity applications.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act(PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conductor sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
## 510(k) SUMMARY
# Prosidyan, Inc.'s FIBERGRAFT® Aeridyan™ Matrix - Bone Graft Substitute
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Prosidyan, Inc. 41 Spring Street #107 New Providence, NJ 07974 Phone: (908) 517-3666 Facsimile: (908) 396-1151 Contact Person: Gina M. Nagvajara, PhD
Date Prepared: July 19, 2022
#### Name of Device and Name
FIBERGRAFT® Aeridyan™ Matrix Bone Graft Substitute
## Common or Usual Name
Filler, Bone Void, Calcium Compound Classification Name/CFR Regulation/Product Code Resorbable Calcium Salt Bone Void Filler, 21 CFR 888.3045, product code MQV
#### Predicate Device
- · Prosidyan Inc, FIBERGRAFT BG Matrix (K180080)
## Reference Devices
- · NovaBone Products LLC, NovaBone MacroFORM (K140946)
- Prosidyan, Inc., FIBERGRAFT® AERIDYAN Matrix (K182670)
## Intended Use / Indications for Use
FIBERGRAFT® Aeridyan™ Matrix - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT® Aeridyan™ Matrix is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT® Aeridyan™ Matrix must be used with autogenous bone marrow aspirate and autograft in posterolateral spine. FIBERGRAFT® Aeridyan™ Matrix must be hydrated with saline or blood for pelvis and extremity applications.
## Device Description
FIBERGRAFT® Aeridyan™ Matrix product is composed of 4585 bioactive glass (M-45 granules), boron bioactive glass (MS-B microspheres/Borospheres™) and bovine type I collagen. After hydration with saline, blood or bone marrow aspirate (BMA), the Aeridyan™ Matrix can be applied directly to the defect site or molded into the desired shape and gently packed into the defect site as a non-setting putty. In posterolateral spine fusion applications, the product is intended to be hydrated with bone marrow aspirate (BMA) and mixed with autograft in a recommended 1:1 ratio. In pelvis and extremity applications, the product must be hydrated with saline or blood.
{4}------------------------------------------------
The FIBERGRAFT® AERIDY AN Matrix product was previously cleared in K182670. There have been no technological changes to the device since this last clearance - the formulation, raw materials and critical raw material suppliers remain the same. The intent of this submission is to expand the indication in the pelvis and extremities based on completion of an additional functional animal study.
#### Technological Characteristics
The FIBERGRAFT Aeridyan Matrix is of the same composition as the previously cleared FIBERGRAFT Aeridyan Matrix in K182670; and also similar in composition to the previously cleared FIBERGRAFT BG Matrix (K180080). All products are resorbable, osteoconductive, bioactive bone void filler devices that are composed of bioactive glass and bovine collagen. The comparison among the devices is shown in the table below.
{5}------------------------------------------------
| | Prosidyan's FIBERGRAFT®<br>Aeridyan ™ Matrix<br>(Subject Device) | Prosidyan's FIBERGRAFT® BG<br>Matrix (K180080)<br>(Predicate Device) | NovaBone MacroFORM (K140946)<br>(Reference Device) | Prosidyan's FIBERGRAFT®<br>AERIDYAN Matrix (K182670)<br>(Reference Device) | | Prosidyan's FIBERGRAFT®<br>Aeridyan ™ Matrix<br>(Subject Device) | Prosidyan's FIBERGRAFT® BG<br>Matrix (K180080)<br>(Predicate Device) | NovaBone MacroFORM (K140946)<br>(Reference Device) | Prosidyan's FIBERGRAFT®<br>AERIDYAN Matrix (K182670)<br>(Reference Device) |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|------------------------------------------------------------------|----------------------------------------------------------------------|----------------------------------------------------|----------------------------------------------------------------------------|
| Intended Use | Bone void filler for orthopedic<br>applications | Bone void filler for orthopedic<br>applications | Bone void filler for orthopedic<br>applications | Bone void filler for orthopedic<br>applications | Primary Component | 45S5 Bioactive Glass, Boron<br>bioactive glass | 45S5 Bioactive Glass | 45S5 Bioactive Glass | 45S5 Bioactive Glass, Boron<br>bioactive glass |
| Indications for Use | FIBERGRAFT® Aeridyan ™ Matrix<br>- Bone Graft Substitute is indicated<br>only for bony voids or gaps that are<br>not intrinsic to the stability of the<br>bony structure. FIBERGRAFT®<br>Aeridyan™ Matrix is indicated to<br>be gently packed into bony voids<br>or gaps of the skeletal system (i.e.,<br>the posterolateral spine,<br>extremities and pelvis). These<br>defects may be surgically created<br>osseous defects or osseous<br>defects created from traumatic<br>injury to the bone. The product<br>provides a bone void filler that<br>resorbs and is replaced with bone<br>during the healing process.<br>FIBERGRAFT® Aeridyan ™ Matrix<br>must be used with autogenous<br>bone marrow aspirate and<br>autograft in posterolateral spine.<br>FIBERGRAFT® Aeridyan ™ Matrix<br>must be hydrated with saline or<br>blood for pelvis and extremity<br>applications. | FIBERGRAFT® BG Matrix - Bone<br>Graft Substitute is indicated only<br>for bony voids or gaps that are not<br>intrinsic to the stability of the bony<br>structure. FIBERGRAFT® BG<br>Matrix is indicated to be gently<br>packed into bony voids or gaps of<br>the skeletal system (i.e.,<br>posterolateral spine, extremities<br>and pelvis). These defects may be<br>surgically created osseous defects<br>or osseous defects created from<br>traumatic injury to the bone. The<br>product provides a bone void filler<br>that resorbs and is replaced with<br>bone during the healing process.<br>FIBERGRAFT® BG Matrix must<br>be used with autogenous bone<br>marrow aspirate and autograft in<br>posterolateral spine. | NovaBone MacroFORM<br>graft devices are indicated only for<br>bony voids or gaps that are not<br>intrinsic to the stability of the bony<br>structure. NovaBone MacroFORM<br>is indicated to be gently packed<br>into bony voids or gaps of the<br>skeletal system (i.e. the<br>extremities and pelvis). These<br>defects may be surgically created<br>osseous defects or osseous<br>defects created from traumatic<br>injury to the bone. NovaBone<br>MacroFORM must be hydrated<br>with autogenous bone marrow<br>prior to implantation. The product<br>provides a bone void filler that<br>resorbs and is replaced with bone<br>during the healing process. | AERIDYAN<br>FIBERGRAFT®<br>Matrix - Bone Graft Substitute is<br>indicated only for bony voids or<br>gaps that are not intrinsic to the<br>stability of the bony structure.<br>FIBERGRAFT®<br>AERIDYAN<br>Matrix is indicated to be gently<br>packed into bony voids or gaps of<br>the skeletal system (i.e.,<br>posterolateral spine). These<br>defects may be surgically created<br>osseous defects or osseous<br>defects created from traumatic<br>injury to the bone. The product<br>provides a bone void filler that<br>resorbs and is replaced with bone<br>during the healing process.<br>FIBERGRAFT® AERIDYAN<br>Matrix must be used with<br>autogenous bone marrow aspirate<br>and autograft in posterolateral<br>spine. | Collagen<br>composition | Type I Collagen from bovine | Type I Collagen from bovine | Type I Collagen from bovine | Type I Collagen from bovine |
| Target Population | Individuals with surgically created<br>osseous defects or osseous<br>defects due to traumatic injury to<br>bone. | Individuals with surgically created<br>osseous defects or osseous<br>defects due to traumatic injury to<br>bone. | Individuals with surgically created<br>osseous defects or osseous<br>defects due to traumatic injury to<br>bone. | Individuals with surgically created<br>osseous defects or osseous<br>defects due to traumatic injury to<br>bone. | Osteoconductive | Yes | Yes | Yes | Yes |
| Anatomical<br>Placement | Posterolateral spine, Extremities<br>and Pelvis | Posterolateral spine, Extremities<br>and Pelvis | Extremities and Pelvis | Posterolateral spine | Bioactive | Yes | Yes | Yes | Yes |
| Resorbs | Yes | Yes | Yes | Yes | | | | | |
| Sterility | Sterile | Sterile | Sterile | Sterile | | | | | |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible | Biocompatible | | | | | |
| Packaging | Thermo-form tray | Thermo-form tray | Thermo-form tray and sealed cups | Thermo-form tray | | | | | |
{6}------------------------------------------------
# K213803
{7}------------------------------------------------
## Performance Data
The performance of the FIBERGRAFT® Aeridyan Matrix has been established by undertaking physical and chemical property evaluation studies, functional performance animal studies and biocompatibility tests. The physical and chemical property studies confirmed the in vitro functionality and bioactivity of the Aeridyan Matrix. The in vitro bioactivity test results have not been correlated to clinical performance. The biocompatibility of the FIBERGRAFT® Aeridyan Matrix has been demonstrated by ISO 10993 testing and the long history of clinical use of the bioactive glass material for the same intended use. In addition, the Aeridyan Matrix is composed of the same bioactive glass material and the same type and duration of patient as the predicates. Packaging evaluations, shelf life testing and real time aging testing were performed with passing results. Bacterial endotoxin testing was performed using the limulus amebocyte lysate (LAL) method and showed that the device meets the endotoxin limits of established guidelines.
The FIBERGRAFT® Aeridyan Matrix product was evaluated in a rabbit posterolateral spine fusion study to support device performance for its indications for use in the FIBERGRAFT® Aeridyan Matrix product was compared to the FIBERGRAFT® BG Matrix predicate device as well as controls. The animal study evaluated device performance in critical sized cancellous bone in the posterolateral spine of 71 skeletally mature rabbits. The performance was evaluated using radiographic, histomorphometric, and biomechanical data. Testing of the FIBERGRAFT® Aeridyan Matrix in the rabbit model is representative of the indications for use and range of anatomical sites proposed for the subject device. The results of the study through 26 weeks of follow-up demonstrated that the FIBERGRAFT® Aeridyan Matrix device performs substantially equivalently to the predicate device and positive control, and any minor technological differences between the device groups do not raise new types of safety or effectiveness concerns.
The FIBERGRAFT® Aeridyan Matrix product has also been evaluated in a rabbit femoral condyle defect study in skeletally mature rabbits to further support device performance for use in extremities and pelvis applications. The FIBERGRAFT® Aeridyan Matrix product was compared to the BG Matrix and NovaBone MacroFORM predicate and reference devices. The animal study evaluated device performance in critical sized defects in the distal femur of 44 skeletally mature rabbits. The performance was evaluated using radiographic, histological, and histomorphometric data. Testing of the FIBERGRAFT® Aeridyan Matrix in the rabbit model is representative of the indications for use and range of anatomical sites proposed for the subject device. The results of the study through 26 weeks of follow up demonstrated that the FIBERGRAFT® Aeridyan Matrix device performs substantially equivalently to the predicate and reference devices, and any minor technological differences between the device groups do not adversely impact performance.
Therefore, performance testing demonstrated that the FIBERGRAFT® Aeridyan Matrix device functions as intended and meets the requirements of class II bone void fillers as compared to the predicate device and reference devices.
#### Substantial Equivalence
As demonstrated in performance testing, the FIBERGRAFT® Aeridyan Matrix has the same intended use and similar indications, technological characteristics, and principles of operation as the predicate and reference devices. The minor technological differences between Aeridyan Matrix and its predicate device do not raise any new issues of safety or effectiveness. The data demonstrated that Aeridyan Matrix is substantially equivalent to the predicate device.
## Conclusion
FIBERGRAFT® Aeridyan Matrix is an osteoconductive, resorbable bone graft substitute composed of bioactive glass, mixed with type I collagen. The FIBERGRAFT® Aeridyan Matrix is substantially equivalent to its predicate and reference devices for its intended use as a synthetic bone void filler. Performance
{8}------------------------------------------------
testing, including in vivo data, demonstrated that the device functions as intended, supporting substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.