K213778 · B.Braun Medical, Inc. · FPA · Mar 3, 2022 · General Hospital
Device Facts
Record ID
K213778
Device Name
IV Administration Set
Applicant
B.Braun Medical, Inc.
Product Code
FPA · General Hospital
Decision Date
Mar 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The IV Administration Sets are intravenous administration sets intended for delivery of fluids from a container into a patient's vascular system. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.
Device Story
Gravity-fed, single-use, disposable IV administration set; delivers fluids from container to patient vascular system. Components include insertion spike, drip chamber, roller clamp, tubing, luer connections, and in-line filter. Used in clinical settings by healthcare providers. Device facilitates fluid therapy; modifications from predicate include fixed male luer lock, roller clamp with spike docking, and addition of in-line filter. Performance verified via bench testing against ISO standards.
Clinical Evidence
Bench testing only. No clinical studies performed. Testing included ISO 8536-4 (physical requirements), ISO 8536-8, ISO 80369-7/20 (connectors), microbial ingress, particulate testing (USP <788>), stress cracking, roller clamp function, joint testing, simulated use, and lipid compatibility.
Technological Characteristics
Gravity-fed infusion set. Materials: PVC (drip chamber, tubing), ABS (spike, clamp), MABS/PP (luer), MABS/ABS/PPSU/PVP/PTFE (filter). Standards: ISO 8536-4, ISO 8536-8, ISO 80369-7, ISO 80369-20, ISO 10993, ISO 11135. Sterilization: Ethylene Oxide (SAL 10^-6).
Indications for Use
Indicated for delivery of fluids from a container into a patient's vascular system. Suitable for any patient population, provided vascular anatomy is adequate and the solution/therapy duration is appropriate.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K173361 — IV Administration Set · B.Braun Medical, Inc. · Jun 26, 2018
K191640 — TK Intravascular Administration Set · Anhui Tiankang Medical Technology Co., Ltd. · Aug 26, 2020
K202437 — Infusion Sets for Single Use · Jiangsu Suyun Medical Materials Co., Ltd. · Mar 25, 2022
K163160 — Sterile Single-use Infusion Set · Jiangxi Hongda Medical Equipment Group , Ltd. · Apr 20, 2017
K162601 — Infusion Set for Single Use · Shinva Ande Healthcare Apparatus Co., Ltd. · May 31, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 3, 2022
B.Braun Medical Inc. Tracy Larish Sr. Regulatory Affairs Specialist 901 Marcon Blvd. Allentown, Pennsylvania 18109
## Re: K213778
Trade/Device Name: IV Administration Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: November 24, 2021 Received: December 3, 2021
### Dear Tracy Larish:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gang Peng for Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213778
Device Name IV Administration Set
Indications for Use (Describe)
The IV Administration Sets are intravenous administration sets intended for delivery of fluids from a container into a patient's vascular system. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="font-weight: bold;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="font-weight: normal;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K213778 510(k) SUMMARY
#### SUBMITTER INFORMATION:
| Name: | B. Braun Medical Inc. |
|-------------------|-------------------------------------------------|
| Address: | 901 Marcon Boulevard |
| | Allentown, PA 18109-9341 |
| Telephone Number: | 610-266-0500, ext. 2966 |
| Contact Person: | Tracy Larish, Sr. Regulatory Affairs Specialist |
| Telephone Number: | (610) 596-2941 |
| Fax Number: | (610) 849-9286 |
| Email: | tracy.larish@bbraunusa.com |
| Date Prepared: | February 3, 2021 |
#### DEVICE NAME:
| Device Trade Name: | IV Administration Set |
|----------------------|----------------------------------------------------------------------------------|
| Common Name: | Intravascular Administration Set |
| Classification Name: | Intravascular administration set, 21 CFR §880.5440<br>Class II, Product code FPA |
#### PREDICATE DEVICES:
- · K173361 IV Administration Set, B. Braun Medical Inc.
### DEVICE DESCRIPTION
IV Administration Sets are gravity, single use, disposable, intravenous administration sets used to deliver fluids from a container into a patient's vascular system. These sets may be comprised of various components which are broadly used throughout industry including insertion spike, drip chamber, clamp, luer access device, stopcock, manifold, tubing, luer connections (connector, adaptor), filter, and hand pump. IV Administration sets are configured to ensure the intended use of the device is met.
### INDICATIONS FOR USE:
The IV Administration Sets are intravenous administration sets intended for delivery of fluids from a container into a patient's vascular system. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.
#### TECHNOLOGICAL CHARACTERISTICS:
The IV Administration Set is substantially equivalent to the B. Braun Medical Inc. IV Administration Set (K173361). The predicate device used for comparison to the IV Administration Set was utilized to demonstrate the substantial equivalence in Indications for Use, intended use, and design between the proposed device and the predicate devices.
The IV Administration Set has the same intended use, the same principle of operation, the same fundamental scientific technology and Indications for Use as the predicate device. It comprised of similar component types and meet the same relevant performance specifications as the predicate devices.
{4}------------------------------------------------
The fundamental technology of the proposed set is unchanged when compared to the predicate device. The differences between the proposed IV Administration Set and predicate device are listed below. These differences do not raise new issues of safety and effectiveness.
- The Spin lock Luer connector with primestop cap has been modified to be a fixed male luer lock and the material of construction have been changed.
- The roller clamp design has been modified to include a docking location for the insertion spike and the materials of construction have been changed.
- An in-line filter was added to the set. ●
A table summarizing the comparison between the IV Administration Sets and the predicate devices is provided below:
| | Proposed IV Administration Set<br>(K213778) | Predicate Device (K173361) IV<br>Administration Set with HCO tubing | Differences |
|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The IV Administration Sets are<br>intravenous administration sets<br>intended for delivery of fluids from a<br>container into a patient's vascular<br>system.<br>These devices may be used for any<br>patient population with consideration<br>given to adequacy of vascular anatomy<br>and appropriateness for the solution<br>being infused and duration of therapy | The IV Administration Sets are<br>intravenous administration sets intended<br>for delivery of fluids from a container<br>into a patient's vascular system.<br>These devices may be used for any<br>patient population with consideration<br>given to adequacy of vascular anatomy<br>and appropriateness for the solution<br>being infused and duration of therapy | Same |
| Intended Use | Delivery of fluids from a container<br>into a patient's vascular system. | Delivery of fluids from a container into<br>a patient's vascular system. | Same |
| Mode of Fluid<br>Delivery | Gravity | Gravity | Same |
| Administration<br>Set Components/<br>materials | Drip chamber PVC<br>Insertion spike-ABS<br>Roller Clamp-ABS<br>Tubing- PVC w/ DEHT and HCO<br>Luer connection- MABS/PP<br>In line filters-MABS/ABS<br>PPSU/PVP/PTFE | Drip chamber-PVC<br>Insertion spike-ABS<br>Roller clamp-ABS<br>Tubing -PVC w/ DEHT and HCO<br>Luer connections-MABS/PC<br>Injection site-Acryllic/Polyisoprene<br>Needleless Y-site - PC<br>Back check valve- MABS | Design and material<br>change to the roller clamp<br>and luer connector.<br>Addition of in-line filter.<br>Biocompatibility and<br>performance testing were<br>performed to demonstrate<br>substantial equivalence<br>(see below). |
| Summary of<br>nonclinical tests<br>for determination<br>of substantial<br>equivalence | ISO 8536-4 Physical Requirements<br>● Particulate<br>● Leakage<br>● Flow Rate<br>● Tensile Strength<br>● Closure-piercing device<br>● Air-inlet device<br>● Tubing<br>● Drip chamber<br>● Flow Regulator<br>● Male Conical Fitting<br>● Protective Caps<br>ISO 8536-8<br>ISO 80369-7:2016<br>ISO 80369-20:2015<br>Internal Testing<br>● Microbial Ingress<br>● Particulate Testing per USP <788><br>● Stress Cracking<br>● Roller Clamp function<br>● Joint testing<br>● Simulated use Testing<br>● Lipid Compatibility | ISO 8536-4 Physical Requirements<br>● Particulate<br>● Leakage<br>● Flow Rate<br>● Tensile Strength<br>● Closure-piercing device<br>● Air-inlet device<br>● Tubing<br>● Drip chamber<br>● Flow Regulator<br>● Male Conical Fitting<br>● Protective Caps | Additional tests performed<br>to confirm substantial<br>equivalence |
| Patient Contact<br>category/<br>Duration | Externally Communicating, Blood<br>Path Indirect prolonged exposure | Externally Communicating, Blood Path<br>Indirect prolonged exposure | Same |
| | Proposed IV Administration Set<br>(K213778) | Predicate Device (K173361) IV<br>Administration Set with HCO tubing | Differences |
| Biocompatibility | Conforms to ISO 10993 | Conforms to ISO 10993 | Same |
| Pyrogenicity | Pyrogenicity <0.5 EU/mL | Pyrogenicity <0.5 EU/mL | Same |
| Sterilization | Ethylene Oxide SAL 10-6 | Ethylene Oxide SAL 10-6 | Same |
{5}------------------------------------------------
# NONCLINICAL TESTING
Bench testing performed on IV Administration Sets to demonstrates that the device performs as intended. No clinical testing was performed as these devices does not require clinical studies to demonstrate substantial equivalence with the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- · ISO 8536-4: "Infusion equipment for medical use Part 4: Infusion sets for single use, gravity feed"
- · ISO 8536-8: "Infusion equipment for medical use Part 8: Infusion sets for single use with pressure infusion apparatus"
- · ISO 10993-1: "Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process"
- · ISO 10993-7 2nd Edition 2008-10-15 Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals [Including: Technical Corrigendum 1 (2009)]
- · ISO 11135, "Sterilization of health care products Ethylene oxide Requirements for development, validation and routine control of a sterilization process for medical devices"
- · ISO 80369-7: "Small-bore connectors for liquids and gases in healthcare applications
- · ISO 80369-20: "Small-bore connectors for liquids and gases in healthcare applications-Part 20:Common test methods
## CONCLUSION:
Results of the functional and performance testing conducted on the proposed devices demonstrate that the IV Administration Set is equivalent to the predicate device. The differences, between subject device and predicate device, do not raise any new issues of safety and effectiveness. The subject IV Administration Set has demonstrated to be substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.