Dental: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the maxillofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients. Neuro: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the cranial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients.
Device Story
Resorbable bone fixation system; includes plates, meshes, and screws; composed of poly lactide and poly glycolide copolymer. Used for fixation of craniofacial, midfacial, and cranial bones following trauma or reconstruction. Operated by surgeons in clinical settings. Implants provide structural stabilization during bone healing; material resorbs over time. System not for mandible or full load-bearing applications. Output is physical stabilization of bone segments. Device is provided sterile; packaging updated to double-pouch configuration with inner pouch as primary sterile barrier.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including sterilization validation (ISO 11135:2014), EO residual testing (ISO 10993-7:2008), endotoxin testing (ISO 10993-1/ANSI/AAMI ST72:2019), and summative usability validation of packaging (n=19 users).
Technological Characteristics
Materials: Copolymer of poly lactide and poly glycolide. Fixation: Screws. Sterilization: Ethylene Oxide (EO) per ISO 11135:2014. Packaging: Double-pouch system with inner pouch as sterile barrier. No changes to implant design or material formulation.
Indications for Use
Indicated for fixation of maxillofacial, midfacial, and cranial skeletal bones affected by trauma or reconstruction in adult and pediatric patients. Not intended for use in the mandible or full load-bearing procedures.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
2/4/2022
Stryker Leibinger Micro Implants Gregory Gohl Senior Regulatory Affairs Specialist 750 Trade Centre Way -Suite 200 Portage, Michigan 49002
Re: K213777
Trade/Device Name: Stryker Resorbable Fixation System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY, GXR, HBW Dated: January 21, 2022 Received: January 26, 2022
Dear Gregory Gohl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K213777
Device Name Stryker Delta Resorbable Fixation System
### Indications for Use (Describe)
Dental: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the maxillofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients.
Neuro: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the cranial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
- SUBMITTER [§807.92(a)(1)] I.
| 510(k) Owner: | Stryker Leibinger GmbH& Co. KG<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Germany |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Submitter/ Contact<br>Person: | Gregory Gohl<br>Sr. Regulatory Affairs Specialist<br>Stryker Craniomaxillofacial (CMF)<br>1941 Stryker Way<br>Portage, MI 49002 |
| Phone: 269-370-1476<br>Fax: 877-648-7114 | |
| Date prepared: | February 4, 2022 |
- DEVICE [§807.92(a)(2)] K213777 II.
| Trade Name: | Stryker Resorbable Fixation System |
|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Abbreviated Name: | Delta System / Delta Resorbable Fixation System |
| Common or Usual Name: | Resorbable Bone Plating System |
| Device: | Stryker Resorbable Fixation System |
| Classification Name &<br>Regulation Description: | Plate, Fixation, Bone; per 21 CFR §888.3030 |
| Regulation Medical<br>Specialty & Review Panel: | Dental (OHT1/DHT1B – Office of Ophthalmic,<br>Anesthesia, Respiratory, ENT, & Dental); Neurology<br>(OHT5/DHT5A – Office of Neurological and Physical<br>Medicine Devices) |
| Primary Product Code: | JEY |
| Subsequent Prod. Codes: | GXR, HBW |
| Regulatory Device Class: | Class II |
| *Note the company Stryker or legacy name Stryker Leibinger precedes the<br>product/trade name and predicate device in some documentation. | |
{4}------------------------------------------------
### III. PREDICATE DEVICE [§807.92(a)(3)]
- A. Submission Branch of Predicate Device: Division of Surgical, Orthopedic, and Restorative Devices; Office of Device Evaluation Center for Devices and Radiological Health
- B. Predicate Device: Stryker Resorbable Fixation System K113109
# IV. DEVICE DESCRIPTION [§807.92(a)(4)]
- A. Submission Branch of Subject Device: Dental (OHT1/DHT1B-Office of Ophthalmic, Anesthesia, Respiratory, ENT, & Dental); Neurology (OHT5/DHT5A - Office of Neurological and Physical Medicine Devices)
- B. Subject Device: Stryker Resorbable Fixation System (also referred to as Stryker Delta Resorbable Fixation System or Delta System
The Stryker Resorbable Fixation System is a cranio-maxillofacial plating system intended for use in the fixation of bones of the craniofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients but is not intended for use in the mandible and/or full load bearing procedures. It consists of resorbable bone fixation plates, meshes, and screws made of a copolymer of poly lactide and poly glycolide.
- V. INDICATIONS FOR USE [§807.92(a)(5)]
| | Predicate Device - K113109 | Subject Device |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Delta Resorbable Fixation System is<br>intended for use in the fixation of bones<br>of the craniofacial and midfacial<br>skeleton affected by trauma or for<br>reconstruction. The system can be used<br>in both adult and pediatric patients but is<br>not intended for use in the mandible<br>and/or full load bearing procedures. | Dental: The Delta Resorbable Fixation<br>System is intended for use in the fixation<br>of bones of the maxillofacial and<br>midfacial skeleton affected by trauma or<br>for reconstruction. The system can be<br>used in both adult and pediatric patients.<br>Neuro: The Delta Resorbable Fixation<br>System is intended for use in the fixation<br>of bones of the cranial skeleton affected<br>by trauma or for reconstruction. The<br>system can be used in both adult and<br>pediatric patients. |
### TABLE 5-1: COMPARISON OF INDICATIONS FOR USE
{5}------------------------------------------------
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE [§807.92(a)(6)]
The subject device is compared to its predicate device for substantial equivalence of technological characteristics based on the following criteria:
- A. Principles of Operation
- B. Technological Characteristics
# C. Principles of Operation / Operating Principle
The basic operational principle of the Stryker Resorbable Fixation System remains the same as for the predicate: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the craniofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients but is not intended for use in the mandible and/or full load bearing procedures. This submission does not propose a change in the operating principle as there is no change to the implant design. interface with accessories or implant materials.
# D. Technological Characteristics
The technological characteristics remain the same as the predicate:
- -Same Intended Use: The Delta Resorbable Fixation System is intended for use in the fixation of bones of the craniofacial and midfacial skeleton affected by trauma or for reconstruction. The system can be used in both adult and pediatric patients but is not intended for use in the mandible and/or full load bearing procedures.
- -Same Mode of Fixation: fixated to the native bone with Stryker Delta Resorbable screws.
- Same Materials of Construction: Implants are made of a copolymer of poly lactide and poly glycolide. No change to the formulation of the polymer.
- -Same Design: The change of the sterilization supplier and location does not change the implant design by any means.
# VII. PERFORMANCE DATA [§807.92(b)(7)]
The modifications described in this submission are the change to a new Ethylene Oxide (EO) sterilization supplier and location and a modification to the packaging. The new EO sterilization cycle is equivalent to the predicate device sterilization cycle operating at lower EO gas concentrations. There is no change to the sterilization method regarding the final end-product. The new supplier location and cycle were proven to be as substantially equivalent for sterilization as the predicate location.
Packaging components, design, and materials remain the same. Labeling requirements also remain unchanged. The modification of the packaging is an update of the sterile barrier layout of the double pouch packaging system (i.e., only the inner pouch is the sterile barrier instead of the outer and inner pouch as for the predicate devices). This change is
{6}------------------------------------------------
communicated to the customer through an inclusion of a warning sticker in regard. Useability validation confirms the warning sticker label mitigates the risk associated with the change in the sterile barrier packaging.
A risk analysis was performed, and sterilization testing was performed in support of the substantial equivalence determination.
There is no change in the design, material, manufacturing process, or duration/location of contact. The tests necessary to evaluate the change regarding the new sterilization location and supplier are listed below. The following tests were performed to adopt the Stryker Resorbable Fixation System plates and meshes into the existing EO Cycle.
| Characteristic | Test | Result | Standards |
|-------------------------------------|-----------------------------------|--------|-------------------------------------|
| Sterilization<br>Validation | Product Adoption and<br>Sterility | Passed | ISO 11135:2014<br>+AMD1:2018 |
| Sterilization /<br>Biocompatibility | EO Residual | Passed | ISO 10993-7:2008<br>+AMD1:2019 |
| Sterilization /<br>Biocompatibility | Endotoxin Testing | Passed | ISO 10993-1<br>+ANSI/AAMI ST72:2019 |
TABLE 5-2: STERILIZATION AND BIOCOMPATIBILITY TESTING
The aseptic transfer (i.e., the transfer of the devices from the non-sterile field into the sterile field during surgery) is modified by the change in the sterile barrier declaration. The aseptic transfer has been identified as the only critical task to be subjected to a summative usability validation of packaging. The human factors testing/study was designed to evaluate in particular possible adverse effects caused by negative transfer. The study was conducted with a sample size of 19 representative users including experienced as well as new users. The full packaging configuration was subjected to test. All test participants were able to perform the aseptic transfer correctly. No use errors or use difficulties were observed. The packaging change was evaluated to be non-critical.
TABLE 5-3: SUMMATIVE USABILITY VALIDATION OF PACKAGING
| Characteristic | Test | Result |
|----------------------|---------------------------------|--------|
| Summative Validation | Aseptic transfer<br>performance | Passed |
# Performance Bench Testing
Performance bench testing was not required as a basis for substantial equivalence.
{7}------------------------------------------------
### Animal Testing
Animal testing was not required as a basis for substantial equivalence.
### Clinical Testing [§807.92(b)(2)]
Clinical testing was not required as a basis for substantial equivalence.
### VIII. CONCLUSION [§807.92(b)(3)]
In summary, the Stryker Resorbable Fixation System is substantially equivalent to its predicate device. The design of implants has not changed, nor has the manufacturing of them changed. There is no change to the sterilization method, and the sterilization cycles of the subject and predicate device are evaluated to be equivalent throughout this submission. The fundamental scientific technology has not changed from the predicate device since the design of the implants or subcomponents has not changed. The intended use, principle of operation, mechanism of action, technological characteristics, and materials of construction also have not changed. In reference to packaging and labeling, specifically, the packaging components, design, and materials remain the same. The moisture barrier change in packaging of the subject devices led to change in labeling. Human factors data supports that the modified packaging, including the change in labeling, is substantially equivalent to the predicate. Overall, the modifications do not raise new questions of safety or effectiveness. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein. it is concluded that the information included in this submission supports substantial equivalence to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.