← Product Code [LNH](/productcode/LNH) · K213693

# MAGNETOM Vida with syngo MR XA50A (K213693)

_Siemens Medical Solutions USA, Inc. · LNH · Feb 25, 2022 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K213693

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Feb 25, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML

## Indications for Use

The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.

## Device Story

MAGNETOM Vida with syngo MR XA50A is an MRI system; inputs include raw k-space data and MR signals; transforms data via deep learning-based reconstruction algorithms (Deep Resolve Swift Brain, Deep Resolve Boost) and pulse sequence modifications (BLADE diffusion, HASTE diffusion) to produce diagnostic images/spectra. Used in clinical settings by trained physicians/technicians. Output viewed by clinicians to assist in diagnosis of head, body, or extremities. Benefits include reduced acquisition time, improved SNR, reduced susceptibility artifacts, and enhanced image sharpness. Supports interventional procedures with MR-compatible hardware.

## Clinical Evidence

No clinical trials conducted. Evidence consists of bench testing, software verification/validation, and sample clinical images comparing new/modified features to the predicate. Clinical publications were referenced to support the utility of deep learning-accelerated reconstruction and specific pulse sequences.

## Technological Characteristics

Magnetic resonance diagnostic device; software-based image reconstruction (deep learning-based); pulse sequences (EPI, TSE, TGSE, SE, DIXON); connectivity via DICOM; conforms to IEC 60601-1, IEC 60601-2-33, ISO 14971, IEC 62304, and NEMA MS standards.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- MAGNETOM Vida with syngo MR XA31A ([K203443](/device/K203443.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

February 25, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left, and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Siemens Medical Solutions USA, Inc. % Karthik Pillai, Ph.D. Senior Regulatory Affairs Professional 40 Liberty Boulevard Mail Code 65-1A MALVERN PA 19355

Re: K213693

Trade/Device Name: MAGNETOM Vida with syngo MR XA50A Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LNI and MOS Dated: February 1, 2022 Received: February 2, 2022

Dear Karthik Pillai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

Submission Number (if known)

K213693

Device Name

MAGNETOM Vida with syngo MR XA50A

Indications for Use (Describe)

The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K213693 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR ട്ട 807.92.

- 1. General Information

| Establishment: | Siemens Medical Solutions USA, Inc. |
|----------------|-------------------------------------|
|                | 40 Liberty Boulevard                |
|                | Mail Code 65-1A                     |
|                | Malvern, PA 19355, USA              |
|                | Registration Number: 2240869        |

- Date Prepared: November 22, 2021
- Siemens Healthcare GmbH Manufacturer: Henkestr. 127 91052 Erlangen Germany Registration Number: 3002808157

Siemens Shenzhen Magnetic Resonance LTD. Siemens MRI Center Hi-Tech Industrial park (middle) Gaoxin C. Ave., 2nd Shenzhen 518057 P.R. CHINA Registration Number: 3004754211

## 2. Contact Information

Karthik Pillai, PhD Senior Regulatory Affairs Professional Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Mail Code 65-1A Malvern, PA 19355, USA +1(317)371-8593 Phone: Fax: (610) 448-1787 E-mail: karthik.pillai@siemens-healthineers.com

## 3. Device Name and Classification

| Device/ Trade name:   | MAGNETOM Vida with syngo MR XA50A           |
|-----------------------|---------------------------------------------|
| Classification Name:  | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology                                   |
| CFR Code:             | 21 CFR § 892.1000                           |
| Classification:       | II                                          |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.

| Product Code: | Primary: LNH        |  |
|---------------|---------------------|--|
|               | Secondary: LNI. MOS |  |

#### 4. Legally Marketed Predicate Device

| Trade name:           | MAGNETOM Vida                               |
|-----------------------|---------------------------------------------|
| 510(k) Number:        | K203443                                     |
| Classification Name:  | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology                                   |
| CFR Code:             | 21 CFR § 892.1000                           |
| Classification:       | II                                          |
| Product Code:         | Primary: LNH<br>Secondary: LNI, MOS         |

#### 5. Intended Use

The indications for use for the subject devices are the same as the predicate device:

The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.

#### 6. Device Description

MAGNETOM Vida with software syngo MR XA50A includes new and modified software compared to the predicate device, MAGNETOM Vida with software syngo MR XA31A. A high-level summary of the new and modified hardware and software is provided below:

#### Software

New Features and Applications

- Deep Resolve Swift Brain is a protocol for fast routine brain imaging primarily based on echo planar imaging (EPI) pulse sequences. Its main enablers are multi-shot (ms) EPI pulse sequence types and a deep learning-based image reconstruction.
- Deep Resolve Boost is a novel deep learning-based image reconstruction algorithm for 2D TSE data, which reconstructs images from k-space raw-data.
- BLADE diffusion is a multi-shot imaging method based on TSE or TGSE (when EPI factor > 1) readout and a BLADE trajectory with diffusion preparation to enable diffusion weighted imaging with reduced sensitivity to B0 inhomogeneity and reduced T2 decay caused image blurring.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.

- HASTE diffusion (HASTE DIFF) is a single-shot imaging method based on TSE readout with diffusion preparation to enable diffusion weighted imaging with reduced sensitivity to B0 inhomogeneity.
## Modified Features and Applications

- Fast GRE RefScan: A speed-optimized reference scan for GRAPPA and SMS kernel calibration for echo planar imaging pulse sequence types.
- Static Field Correction is a reconstruction option reducing susceptibilityinduced distortions and intensity variations.
- Deep Resolve Gain is a reconstruction option which improves the SNR of the scanned images. The functionality has been extended to pulse sequence types SE and TSE DIXON.
- Deep Resolve Sharp is an interpolation algorithm which increases the perceived sharpness of the interpolated images. The functionality has been extended to pulse sequence types SE and TSE DIXON.
- The myExam Angio Advanced Assist provides an assisted and quided workflow for peripheral angiography examination using care bolus. The main advantage of this new workflow is a simplified and improved planning procedure of multi-station peripherical angiography measurements.

## Other Modifications and / or Minor Changes

- TSE MoCo is an image-based motion correction in the average-dimension for the TSE pulse sequence type.
- MR Breast Biopsy is improved with an automatic fiducial detection.

## 7. Substantial Equivalence

MAGNETOM Vida with software syngo MR XA50A is substantially equivalent to the following predicate device:

| Predicate Device                     | FDA Clearance Number<br>and Date   | Product<br>Code | Manufacturer               |
|--------------------------------------|------------------------------------|-----------------|----------------------------|
| MAGNETOM Vida with<br>syngo MR XA31A | K203443, cleared March<br>31, 2021 | LNH<br>LNI, MOS | Siemens Healthcare<br>GmbH |

## 8. Comparison of technological characteristics with the predicate device

The subject device, MAGNETOM Vida with software syngo MR XA50A, is substantially equivalent to the predicate device with regard to the operational environment, programming language, operating system and performance.

The subject devices conform to the standard for medical device software (IEC 62304) and other relevant IEC and NEMA standards.

There are some differences in technological characteristics between the subject device and predicate device, including modified software. These differences have been tested and the conclusions from the non-clinical data suggests that the features bear an equivalent safety and performance profile to that of the predicate device.

## 9. Nonclinical Tests

The following performance testing was conducted on the subject devices.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.

| Performance Test                                                                                                                               | Tested Hardware or Software                                                                                                                                    | Source/Rationale for test                                                                            |
|------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Sample clinical images                                                                                                                         | new and modified software features                                                                                                                             | Guidance for Submission of<br>Premarket Notifications for                                            |
| Image quality assessments by<br>sample clinical images. In<br>some cases a comparison of<br>the image quality / quantitative<br>data was made. | - new / modified pulse sequence<br>types and algorithms.<br>- comparison images between<br>the new / modified features<br>and the predicate device<br>features | Magnetic Resonance<br>Diagnostic Devices                                                             |
| Software verification and<br>validation                                                                                                        | mainly new and modified<br>software features                                                                                                                   | Guidance for the Content of<br>Premarket Submissions for<br>Software Contained in Medical<br>Devices |

The results from each set of tests demonstrate that the devices perform as intended and are thus substantially equivalent to the predicate device to which it has been compared.

## 10.Clinical Tests / Publications

No additional clinical tests were conducted to support substantial equivalence for the subject devices; however, as stated above, sample clinical images were provided. Clinical publications were referenced to provide information on the use of the following features and functions.

| Feature / Function                                                                                                                                                                                                                                                                     | Clinical Publication                                                                                                                                                                                                                                                                      |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Deep Resolve Swift<br>Brain                                                                                                                                                                                                                                                            | [1] Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE:<br>Sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952-962.                                                                                                                                                    |
|                                                                                                                                                                                                                                                                                        | [2] Hyun CM, Kim HP, Lee SM, Lee S, Seo JK. Deep learning for<br>undersampled MRI reconstruction. Phys Med Biol. 2018;63(13):135007.<br>doi:10.1088/1361-6560/aac71a                                                                                                                      |
|                                                                                                                                                                                                                                                                                        | [3] Wang S, Su Z, Ying L, et al. Accelerating magnetic resonance imaging<br>via deep learning. In: 2016 IEEE 13th International Symposium on<br>Biomedical Imaging (ISBI). Prague, Czech Republic: IEEE; 2016:514-517.<br>doi:10.1109/ISBI.2016.7493320                                   |
|                                                                                                                                                                                                                                                                                        | [4] Yu S, Park B, Jeong J. Deep iterative down-up CNN for image<br>denoising. In: Proc. IEEE Conf. Comput. Vis. Pattern Recognit.; 2019:9.                                                                                                                                                |
|                                                                                                                                                                                                                                                                                        | [5] Hammernik K, Schlemper J, Qin C, Duan J, Summers RM, Rueckert D.<br>Σ-net: Systematic evaluation of iterative deep neural networks for fast<br>parallel MR image reconstruction. ArXiv191209278 Cs Eess. December<br>2019. http://arxiv.org/abs/1912.09278. Accessed January 9, 2020. |
|                                                                                                                                                                                                                                                                                        | [6] Hammernik K, Schlemper J, Qin C, Duan J, Summers RM, Rueckert D.<br>Systematic evaluation of iterative deep neural networks for fast parallel<br>MRI reconstruction with sensitivity-weighted coil combination. Magn.<br>Reson. Med. 2021;86(4):1859-1872. doi:10.1002/mrm.28827      |
| [7] Wang Z, Bovik AC, Sheikh HR, Simoncelli EP. Image quality<br>assessment: From error visibility to structural similarity. IEEE Trans Image<br>Process. 2004;13(4):600-612. doi:10.1109/TIP.2003.819861                                                                              |                                                                                                                                                                                                                                                                                           |
| [8] Schlemper J, Caballero J, Hajnal JV, Price AN, Rueckert D. A deep<br>cascade of convolutional neural networks for dynamic MR image<br>reconstruction," IEEE Trans. Med. Imag. 2018;37:491-503                                                                                      |                                                                                                                                                                                                                                                                                           |
| [9] Zbontar J, Knoll F, Sriram A, et al. fastMRI: An open dataset and<br>benchmarks for accelerated MRI. arXiv:181108839 [physics, stat].<br>December 2019. http://arxiv.org/abs/1811.08839. Accessed March 5,<br>2020.                                                                |                                                                                                                                                                                                                                                                                           |
| [10] Demir et al., Optimization of Magnetization Transfer Contrast for EPI<br>FLAIR Brain Imaging, Proceedings of the ISMRM 2021, abstract 4179                                                                                                                                        |                                                                                                                                                                                                                                                                                           |
| [11] Clifford et al., Clinical evaluation of an Al-accelerated two-minute<br>multi-shot EPI protocol for comprehensive high-quality brain imaging,<br>Proceedings of the ISMRM 2021, abstract 300                                                                                      |                                                                                                                                                                                                                                                                                           |
| [12] Filho et al., Clinical Evaluation of An Al-Accelerated Two-Minute<br>Multi-Shot EPI Protocol For Comprehensive High-Quality Brain MRI In An<br>Emergency And Inpatient Setting, Proceedings of the RSNA 2021,<br>accepted, abstract 16890                                         |                                                                                                                                                                                                                                                                                           |
| [13] Pistocchi et al., Al-enhanced multi-shot multi-contrast EPI protocol:<br>Preliminary clinical experience, Proceedings of the ECR 2022, submitted                                                                                                                                  |                                                                                                                                                                                                                                                                                           |
| [14] Gassenmaier S et al., Deep learning-accelerated T2-weighted<br>imaging of the prostate: Reduction of acquisition time and improvement<br>of image quality, European Journal of Radiology, 137 (2021)                                                                              |                                                                                                                                                                                                                                                                                           |
| [15] Herrmann J et al., Diagnostic Confidence and Feasibility of a Deep<br>Learning Accelerated HASTE Sequence of the Abdomen in a Single<br>Breath-Hold, Investigative Radiology, Volume 56, Number 5, May 2021                                                                       |                                                                                                                                                                                                                                                                                           |
| [16] Shanbhogue K et al. Accelerated single-shot T2-weighted fat-<br>suppressed (FS) MRI of the liver with deep learning-based image<br>reconstruction: qualitative and quantitative comparison of image quality<br>with conventional T2-weighted FS sequence. Eur Radiol. 2021 May 7. |                                                                                                                                                                                                                                                                                           |
| Deep Resolve Boost                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                           |
| [17] Herrmann J et al., Development and Evaluation of Deep Learning-<br>Accelerated Single-Breath-Hold Abdominal HASTE at 3 T Using Variable<br>Refocusing Flip Angles. Invest Radiol. 2021 Apr 22.                                                                                    |                                                                                                                                                                                                                                                                                           |
| [18] Gassenmaier S et al., Accelerated T2-Weighted TSE Imaging of the<br>Prostate Using Deep Learning Image Reconstruction: A Prospective<br>Comparison with Standard T2-Weighted TSE Imaging, Cancers, 13,<br>3593 (2021)                                                             |                                                                                                                                                                                                                                                                                           |

{7}------------------------------------------------

{8}------------------------------------------------

|                 | [19] Herrmann J et al., Feasibility and implementation of a Deep Learning<br>MR reconstruction for TSE sequences in musculoskeletal imaging,<br>Diagnostics, 11, 1484 (2021)                                                                                                                                |  |  |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
|                 | Publications under review:                                                                                                                                                                                                                                                                                  |  |  |
|                 | [20] Judith Herrmann et al., Feasibility and diagnostic confidence of deep<br>learning reconstructed TSE imaging of the knee at 1.5 and 3 T. A<br>prospective study.                                                                                                                                        |  |  |
|                 | [21] Alsop, D. C. (1997). Phase insensitive preparation of single<br>shot RARE: application to diffusion imaging in humans. Magnetic<br>resonance in medicine, 527-533.                                                                                                                                     |  |  |
| BLADE_Diffusion | [22] Fu, Q., Kong, X.-c., Liu, D.-x., Guo, Y.-h., Zhou, K., Lei, Z.-q., &<br>Zheng, C.-s. (2021). Clinical utility of turbo gradient and spin echo<br>BLADE-DWI (TGSE-BLADE-DWI) for orbital tumors compared<br>with readout-segmented echo-planar DWI. Proc. Intl. Soc. Mag.<br>Reson. Med., (p. 3933).    |  |  |
|                 | [23] Fu, Q., Kong, X.-c., Liu, D.-x., Guo, Y.-h., Zhou, K., Lei, Z.-q., &<br>Zheng, C.-s. (2021). The efficacy 2D turbo gradient- and spin-echo<br>diffusion-weighted imaging for cerebellopontine angle tumors.<br>Proc. Intl. Soc. Mag. Reson. Med., (p. 3934).                                           |  |  |
|                 | [24] Hu, H. H., McAllister, A. S., Jin, N., Lubeley, L. J., Selvaraj, B., Smith,<br>M., ... Zhou, K. (2019). Comparison of 2D BLADE turbo gradient-and<br>spin-Echo and 2D spin-Echo Echo-planar diffusion-weighted brain MRI at<br>3 T: preliminary experience in children. Academic radiology, 1597-1604. |  |  |
|                 | [25] Pipe, J. G., Farthing, V. G., & Forbes, K. P. (2002). Multishot<br>Diffusion-Weighted FSE UsingPROPELLER MRI. Magnetic Resonance<br>in Medicine, 42-52.                                                                                                                                                |  |  |
|                 | [26] Sheng, Y., Hong, R., Sha, Y., Zhang, Z., Zhou, K., & Fu, C. (2020).<br>Performance of TGSE BLADE DWI compared with RESOLVE DWI in the<br>diagnosis of cholesteatoma. BMC medical imaging, 1-9.                                                                                                         |  |  |
|                 | [27] Srinivasan, G., Rangwala, N., & Zhou, X. J. (2018). Steer-PROP: a<br>GRASE-PROPELLER sequence with interecho steering gradient pulses.<br>Magnetic resonance in medicine, 2533-2541.                                                                                                                   |  |  |
|                 | [28] Yuan, T., Sha, Y., Zhang, Z., Liu, X., Ye, X., Sheng, Y., ... Fu, C.<br>(2020). TGSE diffusion-weighted pulse sequence in the evaluation of<br>optic neuritis: Acomprehensive comparison of image quality with<br>RESOLVE DWI. Proc. Intl. Soc. Mag. Reson. Med., (p. 4137).                           |  |  |
|                 | [29] Zhou, K., Liu, W., & Cheng, S. (2018). Non-CPMG PROPELLER<br>diffusion imaging: comparison of phase insensitive preparation with split<br>acquisition. Proc. Intl. Soc. Mag. Reson. Med., (p. 5320).                                                                                                   |  |  |

# 11.Safety and Effectiveness

The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.

Risk Management is ensured via a risk analysis in compliance with ISO 14971, to identify and provide mitigation of potential hazards early in the design cycle

{9}------------------------------------------------

and continuously throughout the development of the product. Siemens Healthcare GmbH adheres to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards. Furthermore, the device is intended for healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.

MAGNETOM Vida with software syngo MR XA50A conforms to the following FDA recognized and international IEC, ISO and NEMA standards:

| Recognition<br>Number | Product<br>Area            | Title of Standard                                                                                                                                                                                        | Reference<br>Number and date              | Standards<br>Development<br>Organization |
|-----------------------|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|------------------------------------------|
| 19-4                  | General II<br>(ES/<br>EMC) | C1:2009/(R)2012 and<br>A2:2010/(R)2012 (Consolidated<br>Text) Medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance (IEC<br>60601-1:2005, MOD) | ES60601-<br>1:2005/(R)2012<br>and A1:2012 | ANSI AAMI                                |
| 12-295                | Radiology                  | Medical electrical equipment -<br>Part 2-33: Particular<br>requirements for the basic<br>safety and essential<br>performance of magnetic<br>resonance equipment for<br>medical diagnosis                 | 60601-2-33 Ed. 3.2 IEC<br>b:2015          | IEC                                      |
| 5-40                  | General I<br>(QS/<br>RM)   | Medical devices - Application of<br>risk management to medical<br>devices                                                                                                                                | 14971 Second<br>edition 2007-03-01        | ISO                                      |
| 5-114                 | General I<br>(QS/<br>RM)   | Medical devices - Part 1:<br>Application of usability<br>engineering to medical devices                                                                                                                  | 62366-1:2015                              | ANSI AAMI<br>IEC                         |
| 13-79                 | Software/<br>Informatics   | Medical device software -<br>Software life cycle processes<br>[Including Amendment 1 (2016)]                                                                                                             | 62304:2006/A1:201<br>6                    | ANSI AAMI<br>IEC                         |
| 12-232                | Radiology                  | Acoustic Noise Measurement<br>Procedure for Diagnosing<br>Magnetic Resonance Imaging<br>Devices                                                                                                          | MS 4-2010                                 | NEMA                                     |
| 12-288                | Radiology                  | Standards Publication<br>Characterization of Phased<br>Array Coils for Diagnostic<br>Magnetic Resonance Images                                                                                           | MS 9-2008 (R2014)                         | NEMA                                     |
| 12-300                | Radiology                  | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set 03/16/2012<br>Radiology                                                                                                                 | PS 3.1 - 3.20<br>(2016)                   | NEMA                                     |
| 12-331                | Radiology                  | Characterization of<br>Radiofrequency (RF) Coil<br>Heating in Magnetic Resonance<br>Imaging Systems                                                                                                      | Standards<br>Publication MS 14-<br>2019   | NEMA                                     |

# 12.Conclusion as to Substantial Equivalence

MAGNETOM Vida with software syngo MR XA50A has the same intended use and same basic technological characteristics than the predicate device system,

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots.

MAGNETOM Vida with syngo MR XA31A, with respect to the magnetic resonance features and functionalities. While there are some differences in technical features compared to the predicate device, the differences have been tested and the conclusions from all verification and validation data suggest that the features bear an equivalent safety and performance profile to that of the predicate device.

Siemens believes that MAGNETOM Vida with software syngo MR XA50A is substantially equivalent to the currently marketed device MAGNETOM Vida with software syngo MR XA31A (K203443, cleared on March 31, 2021).

---

**Source:** [https://fda.innolitics.com/device/K213693](https://fda.innolitics.com/device/K213693)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
