32Ch AIR Open Coil Suite: 1.5T AIR Open Head Neck Posterior 9ch / 3.0T AIR Open Head Neck Posterior 9ch, 1.5T AIR Open Neck Chest Anterior 7ch / 3.0T AIR Open Neck Chest Anterior 7ch, 1.5T AIR Open Head Anterior 16ch / 3.0T AIR Open Head Anterior 16ch
K213687 · Neocoil, LLC · MOS · Mar 11, 2022 · Radiology
Device Facts
Record ID
K213687
Device Name
32Ch AIR Open Coil Suite: 1.5T AIR Open Head Neck Posterior 9ch / 3.0T AIR Open Head Neck Posterior 9ch, 1.5T AIR Open Neck Chest Anterior 7ch / 3.0T AIR Open Neck Chest Anterior 7ch, 1.5T AIR Open Head Anterior 16ch / 3.0T AIR Open Head Anterior 16ch
Applicant
Neocoil, LLC
Product Code
MOS · Radiology
Decision Date
Mar 11, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The 32Ch AIR Open Coil Suite are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended to be used by trained medical professionals, for adult patients, in combination with and controlled by 1.5T or 3.0T GE Healthcare MRI system software. The 32Ch AIR Open Coil Suite is intended for use with specified couchtop and patient immobilization devices depending on the region of interest. When used as intended, the 32Ch AIR Open Coil Suite is used to produce diagnostic images of the head, neck, and brachial plexus structures. The images can be interpreted by medical professionals or facilitate Radiation Therapy (RT) planning.
Device Story
32Ch AIR Open Coil Suite consists of receive-only phased array RF coils (1.5T and 3.0T variants) for head, neck, and brachial plexus imaging; used with GE Healthcare MRI scanners. Coils receive magnetic resonance signals from hydrogen nuclei; signals amplified and transferred to scanner for tomographic image reconstruction. Operated by trained medical professionals in clinical settings. Used with couchtop/immobilization devices for patient positioning. Output images used for diagnostic interpretation or RT planning. Benefits include improved SNR and uniform anatomical coverage.
Clinical Evidence
Clinical performance evaluation conducted to assess image quality across 1.5T and 3.0T field strengths. No adverse events reported. Bench testing included biocompatibility, electrical safety (IEC 60601-1, 60601-2-33, 60601-1-2), surface temperature, and NEMA MS 6-2008 compliance for SNR and image uniformity. Results demonstrate performance consistent with predicate.
Technological Characteristics
Receive-only phased array RF coils; 1.5T and 3.0T field strengths. Materials assessed for biocompatibility. Energy source: MRI scanner (no internal power). Connectivity: System cable to GE Healthcare MRI scanner. Standards: IEC 60601-1, 60601-2-33, 60601-1-2, NEMA MS 6-2008.
Indications for Use
Indicated for adult patients requiring diagnostic MRI imaging of head, neck, and brachial plexus structures, or for Radiation Therapy (RT) planning, using 1.5T or 3.0T GE Healthcare MRI systems.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
March 11, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
NeoCoil, LLC % Katie Gonzalez Quality Systems and Regulatory Manager N27 W23910A Paul Road PEWAUKEE WI 53072
## Re: K213687
Trade/Device Name: 32Ch AIR Open Coil Suite: 1.5T AIR Open Head Neck Posterior 9ch / 3.0T AIR Open Head Neck Posterior 9ch, 1.5T AIR Open Neck Chest Anterior 7ch / 3.0T AIR Open Neck Chest Anterior 7ch, 1.5T AIR Open Head Anterior 16ch / 3.0T AIR Open Head Anterior 16ch Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: February 23, 2022 Received: February 24, 2022
Dear Katie Gonzalez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213687
#### Device Name
32Ch AIR Open Coil Suite: 1.5T AIR Open Head Neck Posterior 9ch / 3.0T AIR Open Head Neck Posterior 9ch, 1.5T AIR Oven Neck Chest Anterior 7ch / 3.0T AIR Open Neck Chest Anterior 7ch, 1.5T AIR Open Head Anterior 16ch / 3.0T AIR Open Head Anterior 16ch
#### Indications for Use (Describe)
The 32Ch AIR Open Coil Suite are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended to be used by trained medical professionals. for adult patients, in combination with and controlled by 1.5T or 3.0T GE Healthcare MRI system software.
The 32Ch AIR Open Coil Suite is intended for use with specified couchtop and patient immobilization devices depending on the region of interest.
When used as intended, the 32Ch AIR Open Coil Suite is used to produce diagnostic images of the head, neck, and brachial plexus structures. The interpreted by medical professionals or facilitate Radiation Therapy RT planning.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for NeoCoil. To the left of the company name is a black circular design with a white crescent shape inside. The company name is written in a sans-serif font, with "Neo" in blue and "Coil" in black.
## Traditional 510(k) Summary
Applicant
NeoCoil, LLC N27 W23910A Paul Rd Pewaukee, WI 53072 USA
Contact
Katie Gonzalez Quality Systems and Regulatory Manager 262-522-6124 (office) 262-347-1251 (fax) Katie.Gonzalez@neocoil.com
Preparation Date
November 18, 2021
Name of Device
- . Trade/Proprietary name(s):
32Ch AIR Open Coil Suite
1.5T 32Ch AIR Open Coil Suite
1.5T AIR Open Head Neck Posterior 9ch 1.5T AIR Open Neck Chest Anterior 7ch 1.5T AIR Open Head Anterior 16ch
3.0T 32Ch AIR Open Coil Suite
3.0T AIR Open Head Neck Posterior 9ch 3.0T AIR Open Neck Chest Anterior 7ch 3.0T AIR Open Head Anterior 16ch
- Common name: Magnetic Resonance Specialty Coil
- . Classification name: 21 CFR 892.1000, Magnetic resonance diagnostic device, Product Code MOS
Predicate Device
3.0T GEM RT Open Array, K143389 cleared on 03/06/2015
Device Description
The 32Ch AIR Open Coil Suite is comprised of MRI receive-only phased array RF coils designed for clinically acceptable signal-to-noise ratio (SNR) and uniform coverage of the Head, Neck and Brachial Plexus anatomy for use with GE Healthcare Magnetic Resonance Imaging (MRI)
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape with an arrow pointing to the right, followed by the word "NeoCoil" in blue and black. The word "Neo" is in blue, while the word "Coil" is in black.
scanners. The 32Ch AIR Open Coil Suite consists of the following arrays available in 1.5T and 3.0T field strengths:
- AIR Open Head Neck Posterior 9ch
- AIR Open Neck Chest Anterior 7ch
- AIR Open Head Anterior 16ch .
The coils receive magnetic resonance signals generated in hydrogen nuclei (protons) in the Head, Neck and Brachial Plexus anatomy while blocking the high-frequency magnetic field applied by the MRI scanner at specified timings. The received signals are amplified before being transferred to the MRI scanner through the coil's system cable. The amplified signals are processed by the MRI scanner into tomographic images of the Head, Neck and Brachial Plexus anatomy by the MRI scanner. Images are tvpically generated as axial, sagittal, coronal oblique slices. There are no accessories associated with the 32Ch AIR Open Coil Suite. Depending on the region of interest, the coils may be used in combination with the CIVCO RT Universal Couchtop™ MR Overlay for GE Motus, Kizuna, and GEM as well as the associated patient immobilization devices (e.g., thermoplastic mask, straps provided with the MRI scanner)
The 1.5T 32Ch AIR Open Coil Suite and the 3.0T 32Ch AIR Open Coil Suite are tuned to receive RF frequency corresponding to the proton precession in a 1.5 tesla and 3.0 tesla magnetic field (respectively), which is governed by the Larmor equation.
The 32Ch AIR Open Coil Suite is intended for use in a manner that is identical to the predicate device described in this submission.
Proposed labeling is documented in the Instructions for Use manual for the 32Ch AIR Open Coil Suite (NC139IFU-en).
Intended Use, including indications
The 32Ch AIR Open Coil Suite are Magnetic Resonance Imaging (MRI) RF Receive-Only Coils intended to be used by trained medical professionals, for adult patients, in combination with and controlled by 1.5T or 3.0T GE Healthcare MRI system software.
The 32Ch AIR Open Coil Suite is intended for use with specified couchtop and patient immobilization devices depending on the region of interest.
When used as intended, the 32Ch AIR Open Coil Suite is used to produce diagnostic images of the head, neck, and brachial plexus structures. The images can be interpreted by medical professionals or facilitate Radiation Therapy (RT) planning.
### Technological Characteristics
32Ch AIR Open Coil Suite coils are similar in design, material, chemical composition and energy source to the legally marketed device, the 3.0T GEM RT Open Array, K143389 cleared on 03/06/2015.
At a high level, the 1.5T 32Ch AIR Open Coil Suite and the 3.0T 32Ch AIR Open Coil Suite included as part of this submission, and the predicate device are based on the following same technological elements:
- Prescription use;
- Coil designs are receive-only phased array coils; ●
- . Decoupling methodology:
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black coil-like symbol on the left, followed by the word "NeoCoil" in blue and black. The "Neo" part of the word is in blue, while the "Coil" part is in black.
- . Patient contacting materials and chemical composition are known materials that have been assessed for compliance with recognized biocompatibility standards;
- Energy source for the coils is the MRI scanner; ●
- No energy is supplied by the coils; ●
- Coils designs are targeted for the head, neck and brachial plexus anatomy; ●
- Mechanical designs are contoured for the head, neck and brachial plexus anatomy: ●
- Facilitate Radiation Therapy (RT) planning;
- Manufactured for use with the same MRI scanner manufacturer.
The following technological differences exist between the subject and predicate device:
- Optimized coil dimensions and geometry to support use on state-of-the-art MRI scanners ● and improve usability.
- Field strength for the 1.5T 32Ch AIR Open Coil Suite .
- . Channel count
The Indications for Use for the 1.5T 32Ch AIR Open Coil Suite and 32Ch AIR Open Coil Suite are similar to the predicate device, the 3.0T GEM RT Open Array, K143389 cleared on 03/06/2015.
Clinical and non-clinical testing demonstrates that the safety and/or effectiveness of the 32Ch AIR Open Coil Suites compared to the predicate device is not adversely affected as a result of the differences.
### Testing
A combination of clinical and non-clinical performance data is included, referenced, or relied on to demonstrate that the 32Ch AIR Open Coil Suite is safe and effective and performs in a manner that demonstrates substantial equivalence to the predicate device, the 3.0T GEM RT Open Array, K143389 cleared on 03/06/2015.
### Performance testing - Bench:
A Test Report Summary for non-clinical Bench Testing performed, including testing to FDArecognized consensus standards identified as relevant in FDA quidance document Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, issued November 18, 2016, is outlined below:
| Test<br>Performed | Objective(s) of the<br>Test | Test Method<br>Description | Pre-defined<br>pass/fail<br>criteria | Results<br>Summary | Discussions /<br>Conclusions |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility<br>Assessment | Assess potential<br>biological risks | Evaluation of<br>data; historical<br>use, biologic<br>testing, where<br>warranted | Acceptable<br>level of risk | Pass | No identified significant<br>risks. |
| Electrical<br>Safety | Basic electrical<br>safety/essential<br>performance, 60601-1 | Test Lab | Pre-defined<br>performance<br>standards | Pass | Applicable<br>requirements for basic<br>electrical safety and<br>essential performance<br>met. |
| Electrical<br>Safety | Particular electrical<br>requirements; MR<br>equipment, 60601-2-<br>33 | Test Lab | Pre-defined<br>performance<br>standards | Pass | Applicable<br>requirements of the<br>particular standard<br>were met. |
| Test<br>Performed | Objective(s) of the<br>Test | Test Method<br>Description | Pre-defined<br>pass/fail<br>criteria | Results<br>Summary | Discussions /<br>Conclusions |
| Electrical<br>Safety | Collateral electrical<br>safety/essential<br>performance, 60601-<br>1-2 | Test Lab /<br>Bench Testing | Pre-defined<br>performance<br>standards | Pass | Applicable<br>requirements of the<br>collateral standard were<br>met. |
| Usability<br>Assessment | Devices meet<br>customer, end user<br>and patient needs | Actual,<br>simulated or<br>retrospective<br>evaluation of<br>the device<br>and/or data | Pre-defined<br>requirements | Pass | The devices met the<br>needs of the customer,<br>end user and patient. |
| Entrapment,<br>Trapping Zone<br>and Cable<br>Looping<br>(assessment w/<br>scanner) | Assess the device for<br>pinch points,<br>entrapment, cable<br>looping - interfacing<br>with MRI scanner | Evaluation of<br>coil-to-scanner<br>entrapment,<br>trapping and<br>cable looping<br>not covered by<br>test lab<br>assessments. | Requirements<br>based on pre-<br>defined<br>requirements<br>in 60601-1<br>and customer<br>requirements | Pass | Requirements were<br>met. |
| Surface<br>Temperature | Surface temperatures<br>do not exceed limits | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | Surface temperatures<br>were within IEC limits. |
| Unplugged<br>Surface<br>Temperature | Devices remain safe in<br>first fault condition | MRI scanner<br>test | Acceptable<br>level of risk | Pass | Surface temperatures<br>were within IEC limits<br>when the coil is left<br>unplugged in the MRI<br>scanner. |
| Blocking<br>Network<br>Analysis | Ensures devices are<br>designed with<br>adequate active and<br>passive transmit<br>decoupling | Theoretical<br>calculations | Adequate<br>transmit<br>decoupling | Pass | Blocking network<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| Maximum B1<br>Peak | Demonstrate the<br>devices can withstand<br>the maximum B1 peak<br>without obvious signs<br>of arcing, burning,<br>voltage breakdown | MRI scanner<br>test and visual<br>inspection | Pre-defined<br>performance<br>standards | Pass | Coils were able to<br>withstand maximum B1<br>peak without obvious<br>signs of arcing, burning<br>or voltage breakdown. |
| B1 Field<br>Distortion | Measure amount of<br>distortion produced<br>due to presence of an<br>RF coil in the scanner | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | B1 field inhomogeneity<br>meets performance<br>requirements and<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| B0 Filed<br>Distortion | Measure amount of<br>distortion produced<br>due to presence of an<br>RF coil in the scanner | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | B0 field inhomogeneity<br>meets performance<br>requirements and<br>demonstrates adequate<br>active and passive<br>transmit decoupling. |
| Test<br>Performed | Objective(s) of the<br>Test | Test Method<br>Description | Pre-defined<br>pass/fail<br>criteria | Results<br>Summary | Discussions<br>Conclusions |
| NEMA MS 6-<br>2008 | Evaluate single-<br>channel non-volume<br>special purpose<br>radiofrequency (RF)<br>coils for use with<br>magnetic resonance<br>(MR) imaging (MRI)<br>systems | MRI scanner<br>test | Pre-defined<br>performance<br>standards | Pass | SNR and Image<br>Uniformity are<br>consistent with the<br>requirements for<br>indications for use. |
{7}------------------------------------------------
N27 W23910A Paul Rd Pewaukee, WI 53072 USA Direct: (262) 347-1250 Fax: (262) 347-1251
Image /page/6/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black circular shape on the left and the word "NeoCoil" in blue on the right. The circular shape appears to be a stylized representation of a coil or spiral.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for NeoCoil. The logo consists of a black circular shape with an arrow pointing to the right, followed by the word "NeoCoil" in blue. The wordmark is in a sans-serif font and is slightly larger than the circular shape.
### Performance testing - Clinical:
Clinical data exhibits a mix of technical factors and anatomy as recommended in FDA guidance; Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices issued November 18, 2016.
No adverse events were reported during clinical performance testing.
Image /page/7/Picture/6 description: The image shows a comparison of MRI scans taken with 1.5T and 3.0T 32Ch AIR Open Coil Suites. The top row displays coronal T2-weighted images without fat saturation, while the bottom row shows sagittal T2-weighted images, also without fat saturation. The left column shows images from the 1.5T suite, and the right column shows images from the 3.0T suite, allowing for a direct visual comparison of image quality and detail between the two different MRI systems.
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for NeoCoil. The logo consists of a black coil-like symbol on the left and the word "NeoCoil" in blue on the right. The coil symbol has an arrow pointing to the right, suggesting movement or flow.
Image /page/8/Figure/2 description: The image shows a comparison of MRI scans taken with 1.5T and 3.0T 32Ch AIR Open Coil Suites. The top row displays axial T2-weighted images of the brain without fat saturation, while the bottom row shows axial T2-weighted images of the neck without fat saturation. The left column shows images taken with the 1.5T coil, and the right column shows images taken with the 3.0T coil. The images show the differences in image quality and resolution between the two coil suites.
### Conclusion
This submission demonstrates by means of nonclinical and clinical testing that the 32Ch AIR Coil Suite are safe and effective and perform as well as or better than the predicate device, the 3.0T GEM RT Open Array, K143389 cleared on 03/06/2015 for the indications claimed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.