K213631 · Medtronic, Inc. · LIT · Dec 15, 2021 · Cardiovascular
Device Facts
Record ID
K213631
Device Name
Chocolate PTA Balloon Catheter
Applicant
Medtronic, Inc.
Product Code
LIT · Cardiovascular
Decision Date
Dec 15, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Chocolate PTA Balloon Catheter is intended for balloon dilatation of lesions in the peripheral vasculature, including the iliac, femoral, ilio-femoral, popliteal, and infra-popliteal arteries.
Device Story
Over-the-wire balloon dilatation catheter; braided shaft; atraumatic tapered tip; semi-compliant balloon; nitinol constraining structure (CS) for fast deflation and uniform re-wrap. Used in peripheral vasculature (iliac, femoral, popliteal, infra-popliteal arteries) for lesion dilatation. Operated by physicians. Balloon expands to known diameters at specific pressures; CS removed from vessel upon deflation. Radiopaque markers assist positioning. Single-use; sterile; non-pyrogenic. Benefits patient by restoring vessel patency via controlled dilatation.
Clinical Evidence
Bench testing only. No clinical data. Biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, hemocompatibility) per ISO 10993-1:2018 passed. Packaging validation, aging, and sterilization testing (SAL 10^-6) performed to confirm safety of IFU/labeling changes.
Technological Characteristics
Over-the-wire catheter; braided shaft; semi-compliant balloon; nitinol constraining structure. Materials: Grilamid L25, HDPE (Petrothene), PE-Orevac 18300M. Radiopaque markers. Sterilization: ISO 11135-1. Compliance: ISO 10993. Dimensions: 120-150 cm length; 0.014" and 0.018" systems.
Indications for Use
Indicated for balloon dilatation of lesions in the peripheral vasculature, including iliac, femoral, ilio-femoral, popliteal, and infra-popliteal arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
December 15, 2021
Medtronic, Inc. Laura Hanson Regulatory Affairs Specialist 2300 Berkshire Lane North, Suite 5 Plymouth, Minnesota 55441
Re: K213631
Trade/Device Name: Chocolate PTA Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: LIT Dated: November 15, 2021 Received: November 17, 2021
Dear Laura Hanson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213631
Device Name Chocolate PTA Balloon Catheter
Indications for Use (Describe)
The Chocolate PTA Balloon Catheter is intended for balloon dilatation of lesions in the peripheral vasculature, including the iliac, femoral, ilio-femoral, popliteal, and infra-popliteal arteries.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Medtronic
### Special 510(k) Summary Chocolate PTA Balloon Catheter
| 510(k) Summary | This summary is being submitted in accordance with the<br>requirements of 21 CFR § 807.92. | | |
|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Applicant/ Submitter | Medtronic, Inc.<br>2300 Berkshire Lane North, Suite 5<br>Plymouth, MN 55441<br>Tel: 763-591-3024 | | |
| Contact Person | Laura Hanson<br>Regulatory Affairs Specialist<br>Tel: 763-591-3024<br>Email: laura.r.hanson@medtronic.com | | |
| Secondary Contact Person | Brandon Casa de Calvo<br>Regulatory Affairs Senior Manager<br>Tel: 763-591-3024<br>Email: brandon.m.casa.de.calvo@medtronic.com | | |
| Date Prepared | November 9, 2021 | | |
| Device Trade Name | Chocolate PTA Balloon Catheter | | |
| Submission 510(k) Number | K213631 | | |
| Device Common Name | Angioplasty Catheter | | |
| Classification Name | Percutaneous catheter, Cardiovascular | | |
| Regulation Number | 21 CFR 870.1250 | | |
| Classification | Class II | | |
| Classification Panel | Cardiovascular | | |
| Product Code | LIT | | |
| Primary Predicate Device | Chocolate PTA Balloon Catheter | | |
| Primary Predicate 510(k)<br>Number | K130414 | | |
| Primary Predicate<br>Regulation Number | 21 CFR 870.1250 | | |
| Indications for Use | The Chocolate PTA balloon catheter is intended for balloon<br>dilatation of lesions in the peripheral vasculature, including the<br>iliac, femoral, ilio-femoral, popliteal, and infra-popliteal arteries. | | |
| Device Description | The Chocolate PTA Balloon Catheter is an "over-the-wire" balloon<br>dilatation catheter with a braided shaft and an atraumatic tapered tip.<br>The product family consists of 0.014" and 0.018" systems that are<br>compatible with 0.014" and 0.018" guidewires, respectively. Overall<br>catheter lengths range from 120-150 cm. | | |
| | The distal end of the catheter has a semi-compliant balloon that<br>expands to known diameters at specific pressures. The balloon is<br>constrained by a nitinol constraining structure (CS) which provides<br>fast deflation and uniform re-wrap. Upon deflation, the CS is removed<br>from the vessel along with the balloon catheter. The balloon is<br>available in multiple sizes and contains two radiopaque markers to<br>assist with positioning. The proximal end of the device is a common<br>balloon catheter design of a braided shaft connected to a plastic hub<br>and strain relief. The hub has two ports; the balloon port is used to<br>inflate the balloon and the guidewire port connects to the guidewire<br>lumen. | | |
| | The Chocolate PTA balloon catheter is intended for single-use only<br>and is provided sterile and non-pyrogenic. | | |
| Comparison of<br>Technological<br>Characteristics | The subject Chocolate PTA Balloon Catheter components will have<br>minor changes to the materials used which are equivalent to the<br>materials used in the predicate device. These material changes are<br>being implemented alongside supplier changes to ensure a more robust<br>supply chain.<br>Additionally, in order to support EU MDR requirements, Chocolate's<br>IFU is being changed in size and stitching to accommodate additional<br>global language requirements with a change to eliminate renal arteries<br>from the indications for use.<br>The subject Chocolate device shares the following technological<br>characteristics with the predicate device. These characteristics are<br>considered identical between the subject and predicate. | | |
| | ● Intended Use | | |
| | ● Principle of Operation | | |
| | ● Balloon technical specifications including available<br>diameters, lengths, materials, radiopaque markers, rated burst<br>pressure, nominal pressure and constraining structure material | | |
| | ● Catheter technical specifications including available lengths,<br>guidewire compatibility, introducer compatibility, and all<br>materials except for the inner member component materials<br>listed in Table 7.1. | | |
| | ● Packaging materials | | |
| | ● Testing and Compliance to ISO 10993 and ISO 11135-1 | | |
| Table 7.1: Chocolate PTA Balloon Catheter Predicate to Subject Equivalencies | | | |
| | | | |
{4}------------------------------------------------
{5}------------------------------------------------
#### ● Sterilization Method
Table 7.1 below outlines the characteristics between the predicate and subject Chocolate PTA Balloon Catheters that are not identical but are considered similar and equivalent.
| | Predicate Device | Subject Device |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Chocolate PTA Balloon Catheter<br>(Primary Predicate) | Chocolate PTA Balloon Catheter |
| Indications for<br>Use | The Chocolate®* PTA Balloon<br>Catheter is intended for balloon<br>dilatation of lesions in the peripheral<br>vasculature, including the iliac,<br>femoral, ilio-femoral, popliteal,<br>infra-popliteal, and renal arteries. | The Chocolate™* PTA balloon<br>catheter is intended for balloon<br>dilatation of lesions in the peripheral<br>vasculature, including the iliac,<br>femoral, ilio-femoral, popliteal, and<br>infra-popliteal arteries. |
| Inner Member<br>component - outer<br>layer resin | Resin Change<br>Inner Member materials:<br>Outer layer NY-L25 Grilamid (Grilamid<br>L25 natural) Colorant 833125 Transblue FDA Inner layer PE-LR734 HDPE LR7340-01<br>Petrothene High Density Intermediate layer (tie-layer)<br>PE-Orevac 18300 | Resin Change<br>Inner Member materials:<br>Outer layer NY-L25 Grilamid (Grilamid L25<br>natural) Inner layer PE-LR734 HDPE LR7340-02<br>Petrothene High Density Intermediate layer (tie-layer)<br>PE-Orevac 18300M |
| IFU<br>(specifications) | Folded & stapled<br>8.5" x 11.0" | Saddle stitch booklet<br>10" x 10" |
| Compliance Card | 5.0" x 6.0" | 4.0" x 5.0"<br>Artwork Update |
| Pouch Label<br>(dimensions) | 6.25" x 8.5" | 6.38" x 8.00" |
| Carton Label<br>(dimensions) | 6.25" x 10.40" | 6.61" x 8.75" |
Performance data Due to the physical differences between the subject device IFU and predicate device IFU, the subject Chocolate PTA Balloon Catheter underwent packaging validation, packaging aging, and sterilization testing with a 72-page IFU and eIFU leaflet to resemble worst-case packaging configuration. This 72-page test IFU and eIFU leaflet exceeds the size of the IFU and eIFU leaflet in the subject device. All units passed their respective testing requirements and no additional risks resulted from the changes to the IFU. Medtronic is also performing 3year real-time aging tests to confirm the accelerated aging results. See
{6}------------------------------------------------
Section 16, Table 16.2 for a summary of completed testing for the subject Chocolate PTA Balloon Catheter IFU changes.
The proposed changes in this submission do not require a change in the sterilization process or equipment (sterilization assurance level: 10-6). Due to successfully completing the testing described in section 15.3.8, it was determined that no sterilization process changes are required due to the proposed changes.
To support the safety and performance of the subject Chocolate PTA Balloon Catheter device, new biocompatibility testing was completed in report D00557778. Biocompatibility Evaluation Report for Chocolate PTA Balloon Catheter for Supplier and Material Changes. The subject Chocolate PTA Balloon Catheter was subjected to the following biocompatibility testing:
- Cytotoxicity ●
- Sensitization ●
- Irritation/intracutaneous reactivity
- Acute systemic toxicity
- Material mediated pyrogenicity
- o Hemocompatibility testing
Based on the biocompatibility data gathered, it was concluded that the subject Chocolate PTA balloon catheter, as guided by ISO 10993-1:2018, passed the required tests and does not present a patient health hazard. As a result of this testing and the associated risk analysis, it was determined that there are no new or increased biocompatibility concerns.
The technological characteristics and performance criteria of the Chocolate PTA Balloon Catheter are equivalent to the predicate device, and the subject device performs in a manner equivalent to the predicate device currently on the market.
- Based on the intended use, technological characteristics, device design, Conclusion operating principle, shelf-life testing, packaging validation and sterilization testing. Medtronic concludes the subject Chocolate PTA Balloon Catheter to be substantially equivalent to the predicate Chocolate PTA Balloon Catheter, K130414.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.