K213629 · Nu Eyne Co., Ltd. · QGL · Feb 12, 2023 · Neurology
Device Facts
Record ID
K213629
Device Name
SMILE
Applicant
Nu Eyne Co., Ltd.
Product Code
QGL · Neurology
Decision Date
Feb 12, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5898
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The SMILE external Trigeminal Nerve Stimulation (eTNS) System is indicated for treatment of pediatric Attention Deficit Hyperactivity Disorder (ADHD) as a monotherapy in patients ages 7 through 12 years old who are not currently taking prescription ADHD medications. The device is to be used for patient treatment by prescription only and is intended to be used in the home under the supervision of a caregiver during periods of sleep.
Device Story
SMILE is a prescription-only, home-use transcutaneous electrical nerve stimulator (eTNS) for pediatric ADHD. Device consists of a main stimulation unit, charging cable, and cutaneous electrodes applied to the forehead. During sleep (7-9 hours), device delivers non-invasive electrical stimulation to the trigeminal nerve. Operation is supervised by a caregiver. Stimulation parameters include 30s ON/OFF cycles with 1s ramp-up/down. Device is battery-powered and provides a direct connection to brain structures implicated in ADHD. Output is intended to modulate neural activity to benefit patients with ADHD symptoms. Healthcare providers prescribe the device; caregivers manage nightly treatment sessions.
Clinical Evidence
Bench testing only. No clinical data presented. Evidence includes electrical safety (IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-10), software validation (IEC 62304), and usability testing (IEC 60601-1-6, IEC 62366-1).
Technological Characteristics
Internally powered (rechargeable Li-ion battery) ME equipment, Type BF applied part. Enclosure: ABS (V-0); PCB: FR4 (V-0). Dimensions: 60x44x17.6mm; Weight: 20.71g. Max output current: 10mA; Phase duration: 250us. Connectivity: USB charging. Sterilization: Not applicable. Software: Nueyne_SMILE v1.00, Moderate level of concern.
Indications for Use
Indicated for pediatric ADHD treatment as monotherapy in patients ages 7-12 years not currently taking prescription ADHD medications.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for attention deficit hyperactivity disorder (ADHD) is a prescription device that stimulates transcutaneously or percutaneously through electrodes placed on the forehead.
Special Controls
In combination with the general controls of the FD&C Act, the transcutaneous electrical nerve stimulator for Attention Deficit Hyperactivity Disorder (ADHD) is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Performance testing must demonstrate the electromagnetic compatibility and electrical, mechanical, and thermal safety of the device.
(3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be performed:
(i) Electrical performance testing must validate electrical output and duration of stimulation;
(ii) Battery performance testing must be performed; and
(iii) Adhesive integrity testing of the electrodes must be conducted.
(4) The technical parameters of the device including waveform, maximum output current and voltage, pulse duration, frequency, net charge per pulse, maximum current density, maximum average current, and maximum average power density must be fully characterized.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Shelf life testing of the electrodes must be performed to demonstrate continued package integrity and component functionality over the labeled shelf life.
(7) Labeling must include the following:
(i) A contraindication for patients with an implanted metallic or electronic device in the head, a cardiac pacemaker, or an implanted or wearable defibrillator;
(ii) A warning that the device is only for use on clean, intact skin;
(iii) Information on how the device operates and the typical sensations experienced during treatment;
(iv) A detailed summary of the device technical parameters;
(v) A shelf life for the electrodes;
(vi) Instructions for use, including placement of the device on the patient; and
(vii) Cleaning instructions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
February 12, 2023
Nu Eyne Co., Ltd. Sung jin Jung Manager #608, 28, Digital-ro 30-gil, Guro-gu Seoul, 08389 Korea. South
Re: K213629
Trade/Device Name: Smile Regulation Number: 21 CFR 882.5898 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Attention Deficit Hyperactivity Disorder Regulatory Class: Class II Product Code: OGL Dated: January 10, 2023 Received: January 13, 2023
Dear Dong Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Pamela D. Scott -S
Pamela D. Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213629
Device Name SMILE (Model: NUEYNE P022)
Indications for Use (Describe)
The SMILE external Trigeminal Nerve Stimulation (eTNS) System is indicated for treatment of pediatric Attention Deficit Hyperactivity Disorder (ADHD) as a monotherapy in patients ages 7 through 12 years old who are not currently taking prescription ADHD medications.
The device is to be used for patient treatment by prescription only and is intended to be used in the home under the supervision of a caregiver during periods of sleep.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b> Prescription Use (Part 21 CFR 801 Subpart D)</b></span> | <span style="text-decoration: underline;"><b> Over-The-Counter Use (21 CFR 801 Subpart C)</b></span> |
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for NuEyene, a medical R&D company. The logo features a stylized, interconnected blue design above the company name. The name "Nu" is in a darker blue, while "Eyene" is in a lighter blue, matching the color of the design above. Below the name, the text "A MEDICAL R&D COMPANY" is written in smaller, all-caps letters.
# 510(k) SUMMARY
This summary of 510(k) -safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: January 10, 2023
#### 1. INFORMATION
1.1 Submitter Information
- l Submitter Name: Nu Eyne Co., Ltd.
- l Address
- : #608, 28, Digital-ro 30-gil, Guro-gu, Seoul, 08389, Republic of Korea
- Telephone Number: +82-2-6953-8120
- -Fax: +82-303-3447-0017
- I Email: sungjin.jung(@nueyne.com
1.2 Contact Person
- I Name: Sung jin Jung (Manager / Nu Eyne Co.,Ltd.)
- I Address: #608, 28, Digital-ro 30-gil, Guro-gu, Seoul, 08389, Republic of Korea
- Telephone Number: +82-2-6953-8120 I . Fax: +82-303-3447-0017
- E-mail: sungjin.jung@nueyne.com
#### 2. DEVICE INFORMATION
- 2.1 Trade Name / Proprietary Name: SMILE (Models: NUEYNE P022)
2.2 Common Name: Transcutaneous electrical nerve stimulator for Attention Deficit Hyperactive Disorder
2.3 Classification Name: Transcutaneous Electrical Nerve Stimulator For Attention
Deficit Hyperactivity Disorder
- 2.4 Product Code: QGL
- 2.5 Classification Regulation: 21 CFR 882.5898
- 2.6 Device Class: Class II (Special Controls)
- 2.7 Classification Panel: Neurology
| Predicate Device | |
|--------------------------|---------------------|
| Manufacturer | NeuroSigma, Inc. |
| Device Name (Trade Name) | Monarch eTNS System |
| De Novo Number | DEN180041 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Nu Eyne, a medical R&D company. The logo features a stylized graphic of interconnected blue shapes resembling molecules or neural networks, with two dark blue circles within the design. Below the graphic, the company name "Nu Eyne" is written in a modern, sans-serif font, with "Nu" in a darker blue and "Eyne" in a lighter blue. Underneath the company name, the tagline "A MEDICAL R&D COMPANY" is written in smaller, uppercase letters.
## 4. SUBJECT DEVICE DESCRIPTION
Transcutaneous electrical nerve stimulator for Attention Deficit Hyperactivity Disorder. A transcutaneous electrical nerve stimulator for Attention Deficit Hyperactivity Disorder (ADHD) is a prescription device that stimulates transcutaneously or percutaneously through electrodes placed on the forehead.
#### 4.1 Device Identification
| Component | Description |
|--------------------|---------------------|
| Instruction Manual | Instruction for Use |
| SMILE Device | Main Body |
| Cable | Charging Cable |
## 4.2 Device Characteristics
| Trade Name | SMILE (Models: NUEYNE P022) |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name | Transcutaneous electrical nerve stimulator for Attention Deficit<br>Hyperactivity Disorder |
| Common Name | Transcutaneous Electrical Nerve Stimulator For ADHD |
| Classification Regulation | 21 CFR 882.5898 |
| Regulation Description | Transcutaneous electrical nerve stimulator for Attention Deficit<br>Hyperactive Disorder |
| Device Class | Class II (Special Control) |
| Panel | Neurology |
| Product Code | QGL |
| Definition | A transcutaneous electrical nerve stimulator for Attention<br>Deficit Hyperactivity Disorder (ADHD) is a prescription device<br>that stimulates transcutaneously or percutaneously through<br>electrodes placed on the forehead. |
| Physical State | Electrical stimulation unit with leads and cutaneous electrodes |
| Technical Method | Applies an electrical current through electrodes on patient's skin |
| Target Area | Trigeminal nerve |
| Software | Software Name: Nueyne_SMILE<br>Software Version: 1.00<br>Level-of Concern: Class B (Moderate Level) |
The SMILE is not related to biologics, drugs, coatings, additives, single-use, and sterile.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features the company name in two lines, with "Nu" in a dark blue, and "Eyne" in a lighter blue. Above the name is a stylized graphic of interconnected circles and lines, also in the lighter blue color, suggesting a network or molecular structure.
#### 4.3 Environment of Use
The device is to be used for patient treatment by prescription only and is intended to be used in the home under the supervision of a caregiver during periods of sleep.
#### 4.4 Description of the Device
(1) Explanation of how the device works/principle of operation
The SMILE eTNS System treatment protocol is administered each night while the patient is sleeping, for 7-9 hours. The device is designed to provide non-invasive electrical stimulation of the trigeminal nerve. The trigeminal nerve is the largest cranial nerve and has three major sensory divisions of the face, all of which are bilateral.
The trigeminal nerve provides a direct connection to multiple brain structures implicated in ADHD and other neurologic and neuropsychiatric disorders.
Image /page/5/Figure/7 description: This image shows a diagram of a device contained in a plastic enclosure. The diagram includes a rechargeable 3.7V battery connected to a main board. A power switch and control switch are connected to the main board, and an electrode pad is also connected to the main board. The diagram also includes labels A, B, and C, which may indicate different connections or pathways within the device.
#### (2) Mechanism of action
(3) Any necessary feature to determine SE or device performance
The SMILE external Trigeminal Nerve Stimulation (eTNS) System is indicated for treatment of pediatric Attention Deficit Hyperactivity Disorder (ADHD) as a monotherapy in patients ages 7 through 12 years old who are not currently taking prescription ADHD medications.
#### 4.5 Materials of Use
The device is manufactured using the materials listed below. The device does not contain Phthalates and not incorporate medicinal substances, tissues, or blood products.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features the company name in a stylized font, with "Nu" in a darker blue and "Eyne" in a lighter blue. Above the name is a graphic element consisting of interconnected circles, also in blue, resembling a network or molecular structure. Below the name, the tagline "A MEDICAL R&D COMPANY" is written in smaller, uppercase letters.
| Product | Part | Technical data | Manufacturer |
|-----------------|-------------|-----------------|------------------------------------------|
| SMILE<br>Device | Battery | TW342431-240mAh | SHENZEN<br>TAIWOO<br>BATTERY CO.,<br>LTD |
| | PCB | FR4, V-0 | JDM CO LTD |
| | LED plastic | LUPOY PC, V-2 | LG Chem |
| | Enclosure | ABS, V-0 | LG Chem |
We design the enclosure material of SMILE with Acrylonitrile butadiene styrene (ABS).
## 4.6 Key Performance Specifications/Characteristics of the Device
| | | (1) SMILE device (TPD-PS02) |
|--|--|-----------------------------|
|--|--|-----------------------------|
| Classification | Protection against electric shock: Internally powered ME Equipment<br>Applied part: Type BF |
|------------------------|---------------------------------------------------------------------------------------------|
| Dimensions (WxHxD) | 60.00mm x 44.00mm x 17.60mm |
| Weight | 20.71g |
| Power source | Rechargeable battery |
| Maximum output current | 10mA |
| Lifetime | 1.5 years |
| Operating condition | Temperature: 10°C ~ 40°C<br>Relative humidity: 5% ~ 85 %<br>Pressure: 700hPa ~ 1060hPa |
| Storage condition | Temperature: -10°C ~ 45°C<br>Relative humidity: 5% ~ 85 %<br>Pressure: 500hPa ~ 1060hPa |
## (2) Cable (CB-02)
| Dimensions (Length) | 1000mm |
|---------------------|--------|
| Current Rating | 2A |
| Voltage Rating | 250V |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Nu Eyne, a medical R&D company. The logo features a stylized graphic of interconnected blue circles above the company name. The name "Nu" is in a dark blue, modern font, while "Eyne" is in a lighter blue, futuristic font. Below the name, the tagline "A MEDICAL R&D COMPANY" is written in smaller, uppercase letters.
## 5. INTENDED USE
SMILE is a prescription device that stimulates nerves transcutaneously through electrodes placed on the forehead and is intended for pediatric ADHD.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for NuEyene, a medical R&D company. The logo features a stylized blue symbol resembling interconnected molecules or cells. Below the symbol is the company name, "NuEyene," in a simple, sans-serif font, with the tagline "A MEDICAL R&D COMPANY" underneath in smaller letters.
## 6. SUBSTANTIAL EQUIVALENCE
| Items | Subject Device | Predicate Device | Comparison<br>Result |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacturer | Nu Eyne Co., Ltd. | NeuroSigma, Inc. | Different |
| Device | Transcutaneous electrical nerve<br>stimulator for Attention Deficit<br>Hyperactive Disorder | Transcutaneous electrical nerve<br>stimulator for Attention Deficit<br>Hyperactive Disorder | Same |
| Trade/Device Name | SMILE | Monarch eTNS System | Different |
| 510(k)/Denovo Number | K213629 | DEN180041 | Different |
| Regulation Number | 21 CFR 882.5898 | 21 CFR 882.5898 | Same |
| Regulation Description | Transcutaneous electrical nerve<br>stimulator for Attention Deficit<br>Hyperactive Disorder | Transcutaneous electrical nerve<br>stimulator for Attention Deficit<br>Hyperactive Disorder | Same |
| Regulatory Class | Class II | Class II | Same |
| Product Code | OGL | QGL | Same |
| Intended Use | SMILE is the device that stimulates<br>nerves and is intended for ADHD | Monarch eTNS System is<br>Transcutaneous electrical nerve<br>stimulator for Attention Deficit<br>Hyperactivity Disorder. | Same |
| Definition | A transcutaneous electrical nerve<br>stimulator for Attention Deficit<br>Hyperactivity Disorder (ADHD) is a<br>prescription device that stimulates<br>transcutaneously or percutaneously | A transcutaneous electrical nerve<br>stimulator for Attention Deficit<br>Hyperactivity Disorder (ADHD) is a<br>prescription device that stimulates<br>transcutaneously or percutaneously | Same |
| | through electrodes placed on the<br>forehead. | through electrodes placed on the<br>forehead. | |
| Review Panel | Neurology | Neurology | Same |
| Physical State | Electrical stimulation unit with leads<br>and cutaneous electrodes. | Electrical stimulation unit with leads<br>and cutaneous electrodes. | Same |
| Technical Method | Applies an electrical current through<br>electrodes on patient's skin. | Applies an electrical current through<br>electrodes on patient's skin. | Same |
| Target Area | Trigeminal nerve | Trigeminal nerve | Same |
| Indications for Use | The SMILE external Trigeminal Nerve<br>Stimulation (eTNS) System is indicated<br>for treatment of pediatric Attention<br>Deficit Hyperactivity Disorder (ADHD)<br>as a monotherapy in patients ages 7<br>through 12 years old who are not<br>currently taking prescription ADHD<br>medications.<br>The device is to be used for patient<br>treatment by prescription only and is<br>intended to be used in the home under<br>the supervision of a caregiver during<br>periods of sleep. | The Monarch external Trigeminal<br>Nerve Stimulation (eTNS) System is<br>indicated for treatment of<br>pediatric Attention Deficit<br>Hyperactivity Disorder (ADHD) as a<br>monotherapy in patients ages 7<br>through 12 years old who are not<br>currently taking prescription ADHD<br>medications.<br>The device is to be used for patient<br>treatment by prescription only and is<br>intended to be used in<br>the home under the supervision of a<br>caregiver during periods of sleep. | Same |
| | | | |
| Picture | | | Different |
| Power Source | Rechargeable battery | Rechargeable battery | Same |
| Computerized | Yes | Yes | Same |
| S/W provided | MODERATE level of concern | MODERATE level of concern | Same |
| Max output current | 10mA | 10mA | Same |
| Patient Override Control Method | On/Off button | On/Off button | Same |
| Max Leakage Current | None (battery operated) | None (battery operated) | Same |
| Indicator display: Unit functioning | Yes | Yes | Same |
| Low battery indicator | Yes | Yes | Same |
| Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-1-11<br>IEC 60601-2-10<br>IEC 62366-1<br>IEC 62304<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-1-11<br>IEC 60601-2-10<br>IEC 62366-1<br>IEC 62304<br>ISO 10993-1 | Same |
| | | ISO 10993-5<br>ISO 10993-10 | |
| Picture | Image: head with device on forehead | Image: person putting device on another person's head | Different |
| Power Source | Rechargeable battery | Rechargeable battery | Same |
| Computerized | Yes | Yes | Same |
| S/W provided | MODERATE level of concern | MODERATE level of concern | Same |
| Max output current | 10mA | 10mA | Same |
| Patient Override Control Method | On/Off button | On/Off button | Same |
| Max Leakage Current | None (battery operated) | None (battery operated) | Same |
| Indicator display: Unit functioning | Yes | Yes | Same |
| Low battery indicator | Yes | Yes | Same |
| Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-1-11<br>IEC 60601-2-10<br>IEC 62366-1<br>IEC 62304 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-1-11<br>IEC 60601-2-10<br>IEC 62366-1<br>IEC 62304 | Same |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Nu Eyne, a medical R&D company. The logo features a stylized, interconnected network of blue shapes resembling a neural network or molecular structure. Below the symbol, the company name "Nu Eyne" is written in a modern, sans-serif font, with the tagline "A MEDICAL R&D COMPANY" appearing in smaller letters underneath.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features a stylized blue symbol resembling interconnected nodes or a network. Below the symbol, the company name "Nu Eyne" is written in a modern, sans-serif font, with "Nu" in a darker shade of blue. Underneath the name, the tagline "A MEDICAL R&D COMPANY" is written in smaller, uppercase letters.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image contains a logo for a company called Nu Eyne, which is described as "A MEDICAL R&D COMPANY". The logo features a stylized, interconnected design in blue, resembling a network or a chain of molecules. The company name "Nu Eyne" is written in a modern, sans-serif font, with "Nu" in a slightly larger size and a darker shade of blue compared to "Eyne".
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features a stylized, interconnected blue design above the company name. The company name, "Nu Eyne," is written in a modern font, with "Nu" in a darker blue and "Eyne" in a lighter blue. Below the name, the tagline "A MEDICAL R&D COMPANY" is written in smaller, uppercase letters.
| | | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | | |
|-----------------------------|-----------------------------------------------|--------------------------------------------|---------------------------------------------|---------------------------------------------|-----------|
| | Timer Setting | Yes | Yes | Same | |
| Device | Weight | 20.71g | 145 g (without battery) | Different | |
| | | Dimensions | 60.00mm x 44.00mm x 17.60mm | 69mm x 115mm x 27mm | Different |
| | | Expected Service Life | 1.5 years | 5 years | Different |
| | | Electrical Protection | Type BF | Type BF | Same |
| Battery | Battery Type | Lithium ion Battery | Lithium ion Battery | Same | |
| | | Expected Service Life | 300 cycles of complete charge-<br>discharge | 300 charges per battery<br>(10 months each) | Same |
| | | Maximum input voltage<br>(USB connector) | 5.25 Vdc | 5.36 Vdc | Different |
| Material | | | | | |
| | Device housing materials | | Plastic ABS | Plastic ABS | Same |
| Stimulation Characteristics | | | | | |
| | Maximum charge per phase | | 2.5 uC (Max 10 mA) | 2.5 uC | Same |
| | Net Charge per pulse | | 0 | 0 | Same |
| | Peak and peak-to-peak current<br>Peak voltage | | ± 10 mA | ± 10 mA | Same |
| | Phase duration | | 250 us | 250 us | Same |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image features the logo for "Nu Eyne, A MEDICAL R&D COMPANY". The logo consists of a stylized, interconnected blue design above the company name. The design appears to be a network of connected nodes, suggesting a focus on research and development. The company name is written in a simple, modern font, with "Nu Eyne" in a larger font size than "A MEDICAL R&D COMPANY".
| Maximum average power density | 20 uW/cm^2 | 7.5 mW/cm^2 | Different Larger<br>surface areamakes<br>power density smaller |
|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Maximum average current density | 66.6 uA/cm^2 | 1.4 mA.cm^2 | Different Larger<br>surface area makes<br>current density<br>smaller |
| Skin contact surface area of the stimulating<br>electrode | 9 cm^2 | 7.1 cm^2 | Different |
| And include the stimulation modulation<br>specifications (Ramp up, Ramp down, on, and<br>off and times for ramp up, ramp down, on, and<br>off. | 30 Sec ON, 1 Sec Ramp Down / 30 sec OFF,<br>1 Sec Ramp Up<br>Steady 8 hours | 30 Sec ON: 1 Sec Ramp Down: 30 Sec<br>OFF: 1 Sec Ramp Up<br>Steady 7-9 hours | Samet |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image features the logo for "Nu Eyne, A MEDICAL R&D COMPANY". The logo consists of a stylized graphic above the company name. The graphic is a blue, interconnected network-like design with two prominent circular nodes. The text "Nu Eyne" is in a bold, dark blue font, and below it, in a smaller font, is the phrase "A MEDICAL R&D COMPANY".
### 7. NON-CLINICAL DATA
#### 7.1 Electrical Safety and EMC Test
The electrical safety tests were performed to protect patients from undue risks arise from any hazards associated with final device. The tests were performed in accordance with the following standards.
| No. | Test Items | Standards |
|-----|-------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| 1 | General requirement for basic safety and<br>essential performance | - IEC 60601-1:2005/A1: 2012<br>(AAMI/ANSI ES 60601-1: 2005/A1: 2012) |
| 2 | General requirement for safety -<br>Electromagnetic disturbances | - IEC 60601-1-2:2014 |
| 3 | General requirement for safety - Medical<br>electrical equipment used in the home<br>healthcare environment | - IEC 60601-1-11:2015 and<br>- FDA Guidance ("Design Considerations for<br>Devices Intended for Home Use") |
| 4 | Particular requirement for safety - Nerve<br>and muscle stimulators | - IEC 60601-2-10:2012/Amd1:2016 |
#### 7.2 Performance Test
The following tests were performed to assess effectiveness of performance of the device. The tests were performed in accordance with following standards.
| No. | Test Items | Standards |
|-----|---------------------------------------------------------------------|---------------------------------|
| 1 | Particular requirement for safety – Nerve<br>and muscle stimulators | - IEC 60601-2-10:2012/Amd1:2016 |
| 2 | Technical Test | - IEC 60601-2-10:2012/Amd1:2016 |
#### 7.3 Usability V&V
The following tests were performed to assess effectiveness of usability of the test was performed in accordance with following standards.
| No. | Test Items | Standards and FDA Guidance documents |
|-----|--------------------------------------------|-------------------------------------------------------------------------------------------|
| 1 | General requirement for safety – Usability | - IEC 60601-1-6:2013 |
| | | - IEC 62366-1:2015 and |
| | | - FDA Guidance ("Applying Human Factors and<br>Usability Engineering to Medical Devices") |
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features a stylized graphic above the company name. The graphic consists of interconnected blue shapes resembling molecules or a network. The company name, "Nu Eyne," is written in a modern font, with "Nu" in a darker blue and "Eyne" in a lighter blue. Below the name, the tagline "A MEDICAL R&D COMPANY" is written in smaller, uppercase letters.
## 7.4 Software
The following tests were performed to assess effectiveness of software of the device. The test was performed in accordance with following standards.
| No. | Test Items | Standards and FDA Guidance Documents |
|-----|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | General requirement for safety -<br>Programmable electrical medical systems<br>(PEMS) | - IEC 62304:2006/AMD1:2015<br>- FDA Guidance (“Guidance for the Content of<br>Premarket Submissions for Software Contained in<br>Medical Devices”) |
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Nu Eyne, a medical R&D company. The logo features a stylized, interconnected blue design resembling a network or a chain of molecules. Below the graphic is the company name, "Nu Eyne," in a simple, sans-serif font, with the tagline "A MEDICAL R&D COMPANY" underneath in smaller letters.
## 8. CONCLUSION
Under the comparing substantial equivalence between the subject device and the predicate device, there are the same points such as general information, some technical and material information. Although there are some differences, the safety and performance test reports are supported to the safety and effectiveness of the subject device.
In this regard, we conclude that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.