K213616 · Becton, Dickinson and Company · FQH · Mar 23, 2022 · General Hospital
Device Facts
Record ID
K213616
Device Name
BD Surgiphor Antimicrobial Irrigation System
Applicant
Becton, Dickinson and Company
Product Code
FQH · General Hospital
Decision Date
Mar 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5475
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BD Surgiphor™ Antimicrobial Irrigation System is intended to mechanically loosen and remove debris, and foreign materials, including microorganisms, from wounds.
Device Story
BD Surgiphor Antimicrobial Irrigation System is a 2-step wound irrigation and debridement system. It consists of two 450 mL bottles: one containing 0.5% povidone iodine solution (Surgiphor Solution) to loosen debris, and one containing sterile saline (SurgiRinse Solution) to rinse the wound. The device is used by clinicians in a clinical setting. The mechanism of action is purely mechanical: the user squeezes the bottle to dispense fluid, and the pressure of the fluid moving across the wound surface physically removes debris, foreign materials, and microorganisms. The povidone iodine acts as a preservative to prevent microbial growth within the solution bottle. The system is provided terminally sterile and is intended for single use. By facilitating the removal of wound contaminants, the device aids in wound cleansing, potentially supporting the healing process.
Clinical Evidence
Bench testing only. Evidence includes USP <51> antimicrobial effectiveness, USP <85>/<161> endotoxin/pyrogen testing, ISO 10993-1 biocompatibility, fluid pressure testing, and nonclinical wound healing studies. Sterilization validation performed per ANSI/AAMI/ISO 11137 and TIR13004. Packaging and shelf-life validated per ISO 11607, ASTM F1980, ASTM F2096, and ASTM D4169. Stability testing conducted per ICH Q1A(R2).
Technological Characteristics
Two-part system: 0.5% povidone iodine solution and 0.9% sterile saline. Applicator: Polycarbonate cap with spike on polypropylene bottle. Sterilization: Gamma irradiation (SAL 10^-6) per ANSI/AAMI/ISO 11137/TIR13004. Biocompatibility: ISO 10993-1 compliant. Preservative: 0.5% Povidone Iodine. Storage: Room temperature, avoid freezing/excessive heat.
Indications for Use
Indicated for mechanical loosening and removal of debris, foreign materials, and microorganisms from wounds.
Regulatory Classification
Identification
A jet lavage is a device used to clean a wound by a pulsatile jet of sterile fluid. The device consists of the pulsing head, tubing to connect to a container of sterile fluid, and a means of propelling the fluid through the tubing, such as an electric roller pump.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.
Atteris Antimicrobial Skin and Wound Cleanser (K160192)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
March 23, 2022
Becton, Dickinson and Company Shellev Wilcox Staff Regulatory Affairs Specialist 75 N Fairway Dr Vernon Hills, Illinois 60061
Re: K213616
Trade/Device Name: BD Surgiphor Antimicrobial Irrigation System Regulation Number: 21 CFR 880.5475 Regulation Name: Jet Lavage Regulatory Class: Class II Product Code: FOH, FRO Dated: February 15, 2022 Received: February 16, 2022
Dear Shelley Wilcox:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julie Morabito Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K213616
Device Name
BD Surqiphor™ Antimicrobial Irrigation System (910100)
Indications for Use (Describe)
BD Surgiphor™ Antimicrobial Irrigation System is intended to mechanically loosen and remove debris, and foreign materials, including microorganisms, from wounds.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| Preparation Date: | 22 MAR 2021 |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number: | K213616 |
| Applicant: | BD<br>75 N Fairway Dr<br>Vernon Hills, IL 60061 |
| Contact Person: | Shelley Wilcox<br>Staff Specialist, Regulatory Affairs<br>Tel: (303) 324-3816 |
| Device Trade Name: | BD Surgiphor™ Antimicrobial Irrigation System |
| Classification Name: | Jet Lavage |
| Device Classification: | Class II (21 CFR 880.5475)<br>Unclassified (Pre-amendment) |
| Product Code: | FQH; FRO |
| Primary Predicate Device: | Irrisept® Antimicrobial Wound Lavage Product<br>System Product Code: FQH (Jet Lavage); Class II<br>(21 CFR 880.5475)<br>Product Code FRO (Dressing, Wound,<br>Drug); Unclassified (pre-amendment)<br>Applicant: Irrimax Corporation<br>K210536 |
| Secondary Predicate Device: | Surgiphor™ Wound Irrigation System<br>Product Code: FQH (Jet Lavage); Class II<br>(21 CFR 880.5475)<br>Applicant: Orthophor LLC<br>K202071 |
| Reference Device: | Atteris Antimicrobial Skin and Wound Cleanser<br>Product Code: FRO (Dressing, Wound,<br>Drug); Unclassified (pre-amendment)<br>Applicant: Rochal Industries, LLC<br>K160192 |
{4}------------------------------------------------
#### Device Description:
The BD Surgiphor™ Antimicrobial Irrigation System is a 2-step system of aqueous solutions for irrigation and debridement of wounds. The 2-step process includes one bottle of Surgiphor™ Solution (0.5% Povidone Iodine) which is used first to loosen wound debris, and one bottle of SurgiRinse™ Solution (saline solution, USP 99.95%) which is used second to rinse the loosened debris from the wound. The mechanical action of fluid moving across the wound provides the mechanism of action and aids in the removal of debris, and foreign materials, including microorganisms, from wounds. The BD Surgiphor™ Antimicrobial Irrigation System is provided as a two-part terminally sterilized system with 450 mL of each solution. The povidone in the Surgiphor™ Solution serves as a preservative to ensure that no unwanted microbial growth occurs in the solution after the bottle is open.
### Indications for Use:
BD Surgiphor™ Antimicrobial Irrigation System is intended to mechanically loosen and remove debris, and foreign materials, including microorganisms, from wounds.
## Comparison of Technological Characteristics:
The subject BD Surgiphor™ Antimicrobial Irrigation System is substantially equivalent to its secondary predicate, predecessor Surgiphor Wound Irrigation System (K202071), primary predicate Irrisept® Antimicrobial Wound Lavage (K210536) and reference device, Atteris Antimicrobial Skin and Wound Cleanser (K160192).
The Surgiphor solution comprises 0.5% povidone iodine in sterile saline while the Irrisept solution comprises 0.05% chlorhexidine gluconate and the Atteris solution, polyaminopropyl biguanide (PHMB) in purified water. The composition of all three solutions acts as a preservative to inhibit microbial growth in the respective solutions.
The mechanism of action is the same between the proposed and the predicates, specifically, the mechanical action of fluid across the wound removes wound debris, including microorganisms. The mechanism of action is defined by the fluid pressure of the solution, dispensed upon a wound. which is demonstrated by testing provided in this 510(k).
The BD Surgiphor™ Antimicrobial Irrigation System is unchanged from the legally marketed predicate Surgiphor™ Wound Irrigation System (K202071) in its intended use, performance, and technological characteristics except as noted in subsequent comparison of technological characteristics table. The removal of wound debris and foreign materials, including microorganisms, is consistent with the primary predicate Irrisept® Antimicrobial Wound Lavage (K210536) and reference predicate device Atteris Antimicrobial Skin & Wound Cleanser (K160192). All three predicate devices have intended use and apply a solution to the wound for removal of wound debris and foreign materials, including microorganisms.
This 510(k) is to specify the removal of microorganisms through mechanical cleansing action within the indications for use statement of the BD Surgiphor™ Antimicrobial Irrigation System consistent with predicates Irrisept Antimicrobial Wound Lavage and the Atteris Antimicrobial Skin & Wound Cleanser.
{5}------------------------------------------------
| Comparison of Technological Characteristics | | | | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison Feature | Proposed Device | Primary Predicate Device | Secondary Predicate Device | Reference Device |
| | BD Surgiphor™ Antimicrobial<br>Irrigation System | Irrisept® Antimicrobial<br>Wound Lavage | Surgiphor™ Wound Irrigation<br>System | Atteris Antimicrobial Skin<br>& Wound Cleanser |
| 510(K) Number | TBD | K210536 | K202071 | K160192 |
| Product Code | FQH, Jet Lavage<br>FRO, Dressing, Wound, Drug | FQH, Jet Lavage<br>FRO, Dressing, Wound, Drug | FQH, Jet Lavage | FRO, Dressing, Wound, Drug |
| Product Classification | Class II (21 CFR 880.5475)<br>Unclassified (Pre-Amendment) | Class II (21 CFR 880.5475)<br>Unclassified (Pre-Amendment) | Class II (21 CFR 880.5475) | Unclassified (Pre-Amendment) |
| Intended Use | Intended for wound cleansing<br>and removal of wound debris | Intended for wound cleansing<br>and removal of wound debris | Intended for wound cleansing<br>and removal of wound debris | Intended for wound cleansing<br>and removal of wound debris |
| Indications For Use | BD Surgiphor™ Antimicrobial<br>Irrigation System is intended to<br>mechanically loosen and remove<br>debris, and foreign materials,<br>including microorganisms, from<br>wounds. | Irrisept® Antimicrobial Wound<br>Lavage is intended for<br>mechanical cleansing and<br>removal of debris, dirt and<br>foreign materials, including<br>microorganisms from wounds. | The Surgiphor™ Wound<br>Irrigation System is a wound<br>cleansing delivery system<br>intended to loosen and remove<br>wound debris. | (Rx Use) Atteris Antimicrobial<br>Skin & Wound Cleanser is<br>intended for mechanical<br>cleansing and removal of debris,<br>dirt and foreign materials,<br>including microorganisms from<br>wounds such as stage I-IV<br>pressure ulcers, diabetic foot<br>ulcers, post-surgical wounds, first<br>and second-degree burns, grafted<br>and donor sites.<br><br>(OTC Use) Atteris Antimicrobial<br>Skin & Wound Cleanser is<br>intended for physical cleaning<br>and removal of dirt and debris,<br>from skin scrapes, cuts,<br>lacerations, minor irritations, exit<br>sites and unbroken skin. |
| Type Of Use | Prescription use only | Prescription use only | Prescription use only | Both prescription use and over-<br>the-counter use |
| Comparison of Technological Characteristics | | | | |
| Comparison Feature | Proposed Device | Primary Predicate Device | Secondary Predicate Device | Reference Device |
| | BD Surgiphor™ Antimicrobial<br>Irrigation System | Irrisept® Antimicrobial<br>Wound Lavage | Surgiphor™ Wound Irrigation<br>System | Atteris Antimicrobial Skin<br>& Wound Cleanser |
| Mechanism Of Action | The mechanical action of fluid<br>across the wound removes wound<br>debris, including<br>microorganisms.<br><br>The change in the indications for<br>use to include removal of<br>microorganisms as a type<br>of wound debris does not change<br>the intended use of the device<br>and does not raise new<br>safety and effectiveness<br>concerns. Substantial equivalence<br>has been confirmed through<br>performance testing of the fluid<br>pressure. | The mechanical action of fluid<br>across the wound removes wound<br>debris. | The mechanical action of fluid<br>across the wound removes wound<br>debris. | The mechanical action of fluid<br>across the wound removes wound<br>debris. |
| Solution | 1 bottle of Surgiphor Solution<br>0.5% povidone iodine plus<br>vitamin E TPGS in 0.9% saline,<br>pH 4.6 – 7.0<br><br>1 bottle of sterile saline, USP,<br>pH 4.0 – 7.0 | 0.05% chlorhexidine gluconate in<br>99.95% sterile water for<br>irrigation, USP | 1 bottle of Surgiphor Solution<br>0.5% povidone iodine plus<br>vitamin E TPGS in 0.9% saline,<br>USP, pH 4.6 – 6.4<br><br>1 bottle of sterile saline, USP,<br>pH 4.0 – 6.5 | Purified water, poloxamer 407,<br>sodium chloride,<br>ethylhexylglycerin,<br>hypromellose, octane-1,2-diol,<br>polyaminopropyl biguanide<br>[PHMB] |
| Solution<br>Antimicrobial<br>Preservative | 0.5% Povidone Iodine | 0.05% Chlorhexidine Gluconate | 0.5% Povidone Iodine | Polyaminopropyl Biguanide<br>[PHMB] |
| How Supplied | 1 – 450 mL bottle of Surgiphor<br>Solution 0.5% povidone iodine<br>plus vitamin E TPGS in 0.9%<br>saline, USP; 1 – 450 mL bottle of<br>sterile saline, USP; packed within<br>a PETG tray heat-sealed with a<br>Tyvek® cover and sterilized by<br>gamma irradiation to achieve a<br>SAL of 10-6. IFU are included<br>with the system. | Provided for single use. A 450<br>mL bottle of Irrisept is double<br>wrapped in CSR and sealed<br>within an outer Tyvek® pouch.<br>The bottle exterior, CSR wraps,<br>and applicator are sterilized by<br>EO gas. The bottle contains<br>aseptically processed Irrisept<br>solution. | 1 – 450 mL bottle of Surgiphor<br>Solution 0.5% povidone iodine<br>plus vitamin E TPGS in 0.9%<br>saline, USP; 1 – 450 mL bottle of<br>sterile saline, USP; packed within<br>a PETG tray heat-sealed with a<br>Tyvek® cover and sterilized by<br>gamma irradiation to achieve a<br>SAL of 10-6. IFU are included<br>with the system. | Provided Non-Sterile In 8 Fluid<br>Ounce Bottle with Sprayer. |
| Comparison of Technological Characteristics | | | | |
| Comparison Feature | Proposed Device | Primary Predicate Device | Secondary Predicate Device | Reference Device |
| | BD Surgiphor™ Antimicrobial<br>Irrigation System | Irrisept® Antimicrobial<br>Wound Lavage | Surgiphor™ Wound Irrigation<br>System | Atteris Antimicrobial Skin<br>& Wound Cleanser |
| Applicator | Polycarbonate cap with a spike<br>threads onto a polypropylene<br>bottle. The user squeezes the<br>bottle to dispense the solution<br>onto the wound. | Multiport applicator that threads<br>onto the Irrisept bottle. | Polycarbonate cap with a spike<br>threads onto a polypropylene<br>bottle. The user squeezes the<br>bottle to dispense the solution<br>onto the wound. | Sprayer Attachment |
| Sterilization | The BD Surgiphor™<br>Antimicrobial Irrigation System<br>is provided terminally sterile to a<br>SAL of 10-6 by gamma<br>irradiation. The system is<br>validated in accordance with<br>ANSI/AAMI/ISO TIR13004:<br>Sterilization of health care<br>products- Radiation -<br>Substantiation of a selected<br>sterilization dose: Method<br>VDmaxSD | Bottle exterior, CSR wraps, and<br>outer packaging conforms to ISO<br>11135-7 for EO sterilization and<br>ISO 10993-7 for EO residuals<br>whereas the solution is<br>aseptically processed. | The Surgiphor Sterile Wound<br>Irrigation System is provided<br>terminally sterile to a SAL of<br>10-6 by gamma irradiation. The<br>system is validated in accordance<br>with ANSI/AAMI/ISO TIR13004:<br>Sterilization of health care<br>products- Radiation -<br>Substantiation of a selected<br>sterilization dose: Method<br>VDmaxSD | Provided non-sterile |
| Biocompatibility | Biocompatible per ISO 10993-1<br>Biological evaluation of medical<br>devices - Part 1: Evaluation and<br>testing within a risk management<br>process, Annex A (informative)<br>Biological evaluation tests as a<br>surface device (breached or<br>compromised) with limited<br>contact (≤ 24 hours).<br>Added Teflon coating to the mold<br>to manufacture the cap. The<br>components of the mold that are<br>Teflon coated do not come into<br>contact with the fluid-contacting<br>portions of the cap. | Biocompatible per ISO 10993<br>testing for a surface device with<br>breached or compromised surface<br>contact and a limited duration (≤<br>24 hours) | Biocompatible per ISO 10993-1<br>Biological evaluation of medical<br>devices - Part 1: Evaluation and<br>testing within a risk management<br>process, Annex A (informative)<br>Biological evaluation tests as a<br>surface device (breached or<br>compromised) with limited<br>contact (≤ 24 hours).<br>No Teflon Coating on the mold to<br>manufacture the cap. | Biocompatible per ISO 10993<br>testing |
| Preservative<br>Effectiveness over<br>Shelf-Life | Demonstrated per USP <51><br>testing. | Demonstrated per USP <51><br>testing. | Demonstrated per USP <51><br>testing. | Demonstrated per USP <51><br>testing. |
| Storage Temperature | Store at room temperature. Avoid<br>freezing and excessive heat<br>above 40°C (104°F) | 10°C to 30°C | 10°C to 30°C | Unknown |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
## Performance Testing
There are no technological differences between the two versions of the Surgiphor System. The change to the BD labeled indication for use to include removal of "microorganisms" as a type of wound debris does not change the intended use of the device and does not raise new safety and effectiveness concerns. Substantial equivalence has been confirmed through performance testing as summarized below.
The following tests were submitted under K202071 and remain valid for the subject device:
- Preservative Antimicrobial Effectiveness ●
- o USP <51> Antimicrobial Effectiveness Testing
- . Endotoxins and Pyrogens
- o USP <85> Bacterial Endotoxins Test
- 0 USP <151> Pyrogen Test (USP Rabbit Test)
- o USP <161> Medical Devices- Bacterial Endotoxin and Pyrogen Tests
- Biocompatibility ●
- ISO 10993-1 Biological Evaluation of Medical Devices o
- Fluid Pressure Testing ●
- Nonclinical Wound Healing Study ●
The following tests were conducted to support the transfer of the Surgiphor System to BD:
- Sterilization ●
- o ANSI/AAMI/ISO TIR13004:2013, Sterilization of Health Care Products -Radiation - Substantiation of a Selected Sterilization Dose: Method VDmax®
- ANSI/AAMI/ISO 11137:2006/ (R) 2015 & A1:2013 & A2:2019. Sterilization O of Health Care Products - Radiation - Part1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ANSI/AAMI/ISO 11137-1:2018/: Sterilization of Health Care Products - O Microbiological methods - Part 1: Determination of the population of microorganisms on product
- ANSI/AAMI/ISO 11137-2:2019/: Sterilization of medical devices - O Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- ANSI/AAMI/ISO 11137-2:2013/ (R) 2019, Sterilization of Health Care O Products - Radiation - Part 2: Establishing the sterilization dose.
- (AAMI) ST72:2002/R2010, Bacterial Endotoxins-Test Methodologies, O Routine Monitoring, and Alternatives to Batch Testing
- . Packaging and Shelf-Life
- ISO 11607 Packaging for Terminally Sterilized Medical Devices O
- ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier O Systems for Medical Devices
- ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in O Packaging by Internal Pressurization (Bubble Test)
- ASTM D4169-16 Standard Practice for Performance Testing of Shipping O Containers and Systems
{9}------------------------------------------------
- Stability Testing ●
- ICH Q1A(R2) Stability Testing of New Drug Substances and Products O
- USP <51> Antimicrobial Effectiveness Testing O
- Free Iodine Determination o
- % Available Iodine Determination o
- o Osmolality Determination
- USP <791> pH Determination o
- Biocompatibility ●
- ISO 10993-1:2018 Biological evaluation of medical devices Part 1: o Evaluation and testing within a risk management process
## Substantial Equivalence Conclusion
The performance testing conducted demonstrates that the subject BD Surgiphor™ Antimicrobial Irrigation System is substantially equivalent to the primary predicate, Irrisept device and the secondary predicate, Surgiphor Wound Irrigation System in intended use and technological characteristics and this premarket notification supports the addition of "microorganisms" as a type of wound debris removed by the device in the labeled indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.