← Product Code [DWF](/productcode/DWF) · K213540

# Quantum SuperPAC Tubing Set, Quantum SuperPAC Cardioplegia Set (K213540)

_Qura S.R.L · DWF · May 20, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K213540

## Device Facts

- **Applicant:** Qura S.R.L
- **Product Code:** [DWF](/productcode/DWF.md)
- **Decision Date:** May 20, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

Quantum SuperPAC tubing set is a set of tubing intended for use during cardiopulmonary bypass for a duration up to 6 hours. Quantum SuperPAC Cardioplegia Set is a tubing set used for the infusion of cardioplegia solutions and physiological fluids during cardiac surgery procedures on the heart and great vessels for up to six hours.

## Device Story

Quantum SuperPAC Tubing and Cardioplegia sets are single-use, sterile, non-pyrogenic tubing circuits used during cardiopulmonary bypass and cardiac surgery. The tubing connects various extracorporeal components (oxygenators, pumps, reservoirs, filters) to facilitate blood circulation or the infusion of cardioplegia solutions and physiological fluids. The devices are operated by perfusionists in a clinical setting. The sets are constructed primarily of DOP-free polyvinyl chloride (PVC) and polycarbonate, with blood-contacting surfaces treated with a phosphorylcholine-based coating to improve biocompatibility. The device functions as a passive conduit for fluids; it does not perform active processing or analysis. By providing a reliable, biocompatible pathway for extracorporeal support, the device enables the maintenance of circulation and myocardial protection during complex cardiac procedures.

## Clinical Evidence

No clinical data included. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), mechanical integrity, pressure drop, spallation, tubing life, connection strength (ISO 15676), shelf-life, and sterilization validation (ISO 11135).

## Technological Characteristics

Materials: DOP-free PVC, polycarbonate, phosphorylcholine-based coating. Form factor: Tubing sets with various configurations/dimensions. Energy source: None (passive). Connectivity: None. Sterilization: Ethylene oxide (EtO) per ISO 11135:2014. Biocompatibility: ISO 10993-1 compliant. Performance standards: ISO 15676.

## Regulatory Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

## Predicate Devices

- Tubing Pack ([K171979](/device/K171979.md))

## Reference Devices

- COBE SMARXT TUBING AND CONNECTORS ([K981613](/device/K981613.md))
- MYOtherm XP Cardioplegia Delivery System With Cortiva BioActive Surface ([K162774](/device/K162774.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

May 20, 2022

Qura S.r.1. Raffaella Tommasini QA&RA Director Via di Mezzo, 23 Mirandola, Modena 41037 Italy

Re: K213540

Trade/Device Name: Quantum SuperPAC Tubing Set, Quantum SuperPAC Cardioplegia Set Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, Or Tubing Regulatory Class: Class II Product Code: DWF Dated: April 20, 2022 Received: April 22, 2022

Dear Raffaella Tommasini:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213540

Device Name Quantum SuperPAC Tubing Set

Indications for Use (Describe)

Quantum SuperPAC Tubing Set is a set of tubing intended for use during cardiopulmonary bypass for a duration up to 6 hours.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213540

Device Name Quantum SuperPAC Cardioplegia Set

Indications for Use (Describe)

Quantum SuperPAC Cardioplegia Set is a tubing set used for the infusion of cardioplegia solutions and physiological fluids during cardiac surgery procedures on the heart and great vessels for up to six hours.

Type of Use (Select one or both, as applicable)

X | Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Qura advanced medical solutions. The logo features a stylized human figure with outstretched arms, enclosed within a curved line. To the right of the figure, the word "QURA" is written in a bold, sans-serif font. Below "QURA", the words "advanced medical solutions" are written in a smaller font.

# 510(K) SUMMARY

| I. SUBMITTER           |                                                                                                                                                                                                                                                                    |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name:        | Qura S.r.l.                                                                                                                                                                                                                                                        |
| Submitter Address:     | Via di Mezzo, 23 41037 Mirandola (MO) Italy                                                                                                                                                                                                                        |
| Contact Person:        | Raffaella Tommasini, QA&RA Director - Qura s.r.l.                                                                                                                                                                                                                  |
| Phone:                 | +39 0535 1803050                                                                                                                                                                                                                                                   |
| e-mail:                | raffaella.tommasini@quramed.com                                                                                                                                                                                                                                    |
| Fax:                   | +39 0535 1803051                                                                                                                                                                                                                                                   |
| Date Summary Prepared: | April 20th, 2022                                                                                                                                                                                                                                                   |
| II. DEVICES            |                                                                                                                                                                                                                                                                    |
| Proprietary Name:      | Quantum SuperPAC Tubing set<br>Quantum SuperPAC Cardioplegia set                                                                                                                                                                                                   |
| Common Name:           | Tubing set<br>Cardioplegia set                                                                                                                                                                                                                                     |
| Classification Name:   | Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass                                                                                                                                                                                                     |
| Regulatory Class:      | II                                                                                                                                                                                                                                                                 |
| Product Code:          | DWF                                                                                                                                                                                                                                                                |
| Panel:                 | Cardiovascular<br>Medical Specialty 74 - 21 CFR Part 870<br>Office of Product Evaluation and Quality / Office of Health Technology 2<br>(OHT2 Cardiovascular Devices) / Division of Health Technology 2B<br>(Circulatory Support, Structural and Vascular Devices) |

#### III. PREDICATE AND REFERENCE DEVICES

# Predicate device:

| Trade Name:                    | Tubing Pack                                                    |
|--------------------------------|----------------------------------------------------------------|
| Registered Establishment Name: | Medtronic, Inc.                                                |
| Regulation Number:             | 870.4210                                                       |
| Regulation Name:               | Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass |
| Regulatory Class:              | II                                                             |
| Product Code:                  | DWF                                                            |
| 510(k) Number                  | K171979                                                        |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed within a blue arc. To the right of the figure is the word "QURA" in bold, blue letters, with the words "advanced medical solutions" in smaller font underneath.

### Reference devices:

| For Quantum SuperPAC Tubing set: |                                      |
|----------------------------------|--------------------------------------|
| Trade Name:                      | COBE SMARXT TUBING AND CONNECTORS    |
| Registered Establishment Name:   | COBE CARDIOVASCULAR, INC             |
| Regulation Number:               | 870.4390                             |
| Regulation Name:                 | Tubing, Pump, Cardiopulmonary Bypass |
| Regulatory Class:                | II                                   |
| Product Code:                    | DWE                                  |
| 510(k) Number                    | K981613                              |

### For Quantum SuperPAC Cardioplegia set:

| Trade Name:                    | MYOtherm XP Cardioplegia Delivery System With Cortiva BioActive Surface,<br>MYOtherm XP Cardioplegia Delivery System With Cortiva BioActive Surface<br>(Bridge System) |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registered Establishment Name: | Medtronic, Inc                                                                                                                                                         |
| Regulation Number:             | 870.4240                                                                                                                                                               |
| Regulation Name:               | Cardiopulmonary Bypass Heat Exchanger                                                                                                                                  |
| Regulatory Class:              | II                                                                                                                                                                     |
| Product Code:                  | DTR                                                                                                                                                                    |
| 510(k) Number                  | K162774                                                                                                                                                                |

#### IV. DEVICE DESCRIPTION

Quantum SuperPAC Tubing set devices are designed to connect different devices that are not provided in the Quantum SuperPAC tubing set such as oxygenators, pumps, reservoirs, filters, and other cardiopulmonary bypass components into circuits used in surgical procedures requiring extracorporeal support, for a maximum duration of 6 hours.

Quantum SuperPAC Cardioplegia set devices are designed to connect different devices that are not provided in the Quantum SuperPAC tubing set for the infusion of cardioplegia solutions and physiological fluids during cardiac surgery procedures on the heart and great vessels for up to six hours.

Quantum SuperPAC devices (all variants) are non-toxic, non-pyrogenic, and sterilized by ethylene oxide. Devices are intended for single use only and are not to be resterilized by the user.

All the device surfaces in contact with blood are treated with a phosphorylcholine-based coating.

Quantum SuperPAC devices (all variants) are mainly constituted of polyvinyl chloride (PVC) DOP free tubing and additional components composing the set; different variants are available, varying for tubing dimension and set configuration in order to address customer and surgical procedure specifications.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with outstretched arms, enclosed within a curved line, next to the word "QURA" in a bold, sans-serif font. Below "QURA" are the words "advanced medical solutions" in a smaller, lighter font.

#### V. INTENDED USE / INDICATIONS FOR USE

- l Quantum SuperPAC Tubing set:
Quantum SuperPAC tubing set is a set of tubing intended for use during cardiopulmonary bypass for a duration up to 6 hours.

- Quantum SuperPAC Cardioplegia set:
Quantum SuperPAC Cardioplegia Set is a tubing set used for the infusion of cardioplegia solutions and physiological fluids during cardiac surgery procedures on the heart and great vessels for up to six hours.

#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

An extensive and complete comparison between Quantum SuperPAC Tubing set and Quantum SuperPAC Cardioplegia set devices and the predicate/reference devices has been conducted. Devices has the same intended use and operating principle as the predicate and reference devices.

| Device                                        | Proposed Devices                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device –<br>Medtronic, Inc.<br>Tubing Pack                                                                                                                                                                                                                                                    | Reference Device<br>– Cobe<br>Cardiovascular Inc<br>COBE SMARXT<br>Tubing and<br>Connectors                                                                                                                                                                      | Reference Device<br>– Medtronic, Inc.<br>MYOtherm XP<br>Cardioplegia<br>Delivery System                                                                                                                                                                                                                                                                                                                                                                                                         |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                               | – Quantum<br>SuperPAC Tubing<br>Set                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                               | Quantum                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                               | SuperPAC<br>Cardioplegia set                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Name                                          | Quantum<br>SuperPAC Tubing<br>Set<br>Quantum<br>SuperPAC<br>Cardioplegia<br>Tubing Set                                                                                                                                                                                                                                                                                                                             | Tubing,<br>Connectors and<br>Accessories                                                                                                                                                                                                                                                                | COBE SMARXT<br>Tubing and<br>Connectors                                                                                                                                                                                                                          | MYOtherm XP<br>Cardioplegia<br>Delivery System<br>With Cortiva<br>BioActive Surface,<br>MYOtherm XP<br>Cardioplegia<br>Delivery System<br>With Cortiva<br>BioActive Surface<br>(Bridge System)                                                                                                                                                                                                                                                                                                  |
| 510(k) Number                                 | N.A.                                                                                                                                                                                                                                                                                                                                                                                                               | K171979                                                                                                                                                                                                                                                                                                 | K981613                                                                                                                                                                                                                                                          | K162774                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device description                            | Quantum<br>SuperPAC Tubing<br>Set is a tubing and<br>components set<br>connecting<br>different medical<br>devices used<br>during the<br>extracorporeal<br>procedure up to<br>six hours.<br>Quantum<br>SuperPAC<br>Cardioplegia<br>Tubing Set is a                                                                                                                                                                  | Medtronic Tubing<br>Packs consist of<br>coated and<br>uncoated tubing,<br>connectors and<br>various medical<br>devices that are<br>pre-assembled<br>into user-<br>specified<br>extracorporeal<br>cardiopulmonary<br>bypass perfusion<br>circuits. Tubing<br>Packs are<br>intended for use<br>within the | The COBE SMART<br>Tubing and<br>Connectors are<br>used in connecting<br>oxygenators,<br>reservoirs, filters<br>and other<br>cardiopulmonary<br>bypass<br>components into<br>circuits used in<br>surgical<br>procedures<br>requiring<br>extracorporeal<br>support | The MYOtherm XP<br>cardioplegia<br>delivery system<br>with Cortiva<br>bioactive surface is<br>comprised of<br>polycarbonate<br>housing and comes<br>preconnected to<br>tubing and<br>connector<br>components<br>comprised of<br>polycarbonate,<br>polyvinyl chloride<br>and plastisol<br>materials and                                                                                                                                                                                          |
| Device                                        | Proposed Devices<br>- Quantum<br>SuperPAC Tubing<br>Set<br>Quantum<br>SuperPAC<br>Cardioplegia set<br><br>tubing and<br>components set<br>used during<br>extracorporeal<br>circulation to<br>infuse<br>cardioplegia/physi<br>ologic fluids<br>solution up to six<br>hours.                                                                                                                                         | Predicate Device –<br>Medtronic, Inc.<br>Tubing Pack<br><br>extracorporeal<br>blood pathway<br>during<br>cardiopulmonary<br>bypass surgical<br>procedures.                                                                                                                                              | Reference Device<br>– Cobe<br>Cardiovascular Inc<br>COBE SMARXT<br>Tubing and<br>Connectors                                                                                                                                                                      | Reference Device<br>– Medtronic, Inc.<br>MYOtherm XP<br>Cardioplegia<br>Delivery System<br><br>provided in Y-type,<br>straight, reducer<br>and tubing<br>configurations.                                                                                                                                                                                                                                                                                                                        |
| Regulation #                                  | 870.4210                                                                                                                                                                                                                                                                                                                                                                                                           | 870.4210                                                                                                                                                                                                                                                                                                | 870.4390                                                                                                                                                                                                                                                         | 870.4240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Regulation Name                               | Cardiopulmonary<br>bypass vascular<br>catheter, cannula,<br>or tubing                                                                                                                                                                                                                                                                                                                                              | Cardiopulmonary<br>bypass vascular<br>catheter, cannula,<br>or tubing                                                                                                                                                                                                                                   | Cardiopulmonary<br>bypass pump<br>tubing                                                                                                                                                                                                                         | Cardiopulmonary<br>bypass heat<br>exchanger.                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code                                  | DWF                                                                                                                                                                                                                                                                                                                                                                                                                | DWF                                                                                                                                                                                                                                                                                                     | DWE                                                                                                                                                                                                                                                              | DTR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification                                | II                                                                                                                                                                                                                                                                                                                                                                                                                 | II                                                                                                                                                                                                                                                                                                      | II                                                                                                                                                                                                                                                               | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Indication for Use                            | Quantum<br>SuperPAC tubing<br>set is a set of<br>tubing intended<br>for use during<br>extracorporeal<br>circulation for a<br>duration up to 6<br>hours.<br>Quantum<br>SuperPAC<br>Cardioplegia Set is<br>a tubing set used<br>for the infusion of<br>cardioplegia<br>solutions and<br>physiological fluids<br>during cardiac<br>surgery<br>procedures on the<br>heart and great<br>vessels for up to<br>six hours. | The Medtronic<br>Tubing Pack is<br>indicated for use in<br>the extracorporeal<br>circuit during<br>cardiopulmonary<br>bypass (CPB)<br>surgical<br>procedures.                                                                                                                                           | The COBE SMART<br>Tubing and<br>Connectors are<br>intended to be<br>used in surgical<br>procedures<br>requiring<br>extracorporeal<br>support for periods<br>of up to six hours                                                                                   | The MYOtherm XP<br>with Cortiva<br>bioactive surface is<br>a device intended<br>for the mixing,<br>warming/cooling<br>and delivery of<br>oxygenated<br>blood/cardioplegia<br>solution in a<br>predetermined<br>ratio during<br>routine<br>cardiopulmonary<br>bypass (CPB)<br>procedures up to 6<br>hours in duration.<br>Blood/cardioplegia<br>solution is<br>delivered to the<br>patient through<br>the cardioplegia<br>delivery system<br>and appropriate<br>cannula by the<br>operation of a |
| Device                                        | Proposed Devices<br>- Quantum<br>SuperPAC Tubing<br>Set<br>Quantum<br>SuperPAC<br>Cardioplegia set                                                                                                                                                                                                                                                                                                                 | Predicate Device –<br>Medtronic, Inc.<br>Tubing Pack                                                                                                                                                                                                                                                    | Reference Device<br>- Cobe<br>Cardiovascular Inc<br>COBE SMARXT<br>Tubing and<br>Connectors                                                                                                                                                                      | Reference Device<br>- Medtronic, Inc.<br>MYOtherm XP<br>Cardioplegia<br>Delivery System                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                               |                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                  | single occlusive<br>roller pump.                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Target population                             | Patient requiring<br>extracorporeal<br>circulation<br>procedures for<br>period up to 6<br>hours                                                                                                                                                                                                                                                                                                                    | Patient requiring<br>CPB procedures for<br>period up to 6<br>hours                                                                                                                                                                                                                                      | Patient requiring<br>surgical<br>procedures<br>requiring<br>extracorporeal<br>support for periods<br>of up to six hours                                                                                                                                          | Patient requiring<br>cardiopulmonary<br>bypass (CPB)<br>procedures up to 6<br>hours in duration.                                                                                                                                                                                                                                                                                                                                                                                                |
| Target User                                   | Perfusionist                                                                                                                                                                                                                                                                                                                                                                                                       | Perfusionist                                                                                                                                                                                                                                                                                            | Perfusionist                                                                                                                                                                                                                                                     | Perfusionist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Biocompatibility<br>Requirements              | In compliance with<br>ISO 10993 series                                                                                                                                                                                                                                                                                                                                                                             | In compliance with<br>ISO 10993 series                                                                                                                                                                                                                                                                  | In compliance with<br>ISO 10993 series                                                                                                                                                                                                                           | In compliance with<br>ISO 10993 series                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Main<br>Contacting<br>Materials               | DOP-free polyvinyl<br>chloride (PVC)<br>Polycarbonate (PC)<br>Phosphorylcholine<br>coating                                                                                                                                                                                                                                                                                                                         | DOP-free polyvinyl<br>chloride (PVC)<br>Polycarbonate (PC)<br>Cortiva, Trillium<br>and Balance<br>coating                                                                                                                                                                                               | DOP-free polyvinyl<br>chloride (PVC)<br>Polycarbonate (PC)<br>SMART coating                                                                                                                                                                                      | Polyvinyl chloride<br>(PVC)<br>Polycarbonate (PC)<br>Plastisol materials                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Blood side<br>Connection Type                 | 3/8x3/32"<br>1/2x3/32"<br>1/4x1/16"<br>3/16x1/16"                                                                                                                                                                                                                                                                                                                                                                  | 0.085x.062 mm<br>0.107x.040mm<br>1/2X1/8"<br>1/2x3/32"<br>1/4X1/16"<br>3/16x1/16"<br>5/16x3/32"                                                                                                                                                                                                         | From 0.075x0.028"<br>to 5/8x1/8"                                                                                                                                                                                                                                 | 3/8" - 3/16"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Single-use                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Sterile Condition                             | Sterile                                                                                                                                                                                                                                                                                                                                                                                                            | Sterile                                                                                                                                                                                                                                                                                                 | Sterile                                                                                                                                                                                                                                                          | Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Description of the<br>sterilization<br>method | EtO sterilization<br>process                                                                                                                                                                                                                                                                                                                                                                                       | EtO sterilization<br>process                                                                                                                                                                                                                                                                            | EtO sterilization<br>process                                                                                                                                                                                                                                     | EtO sterilization<br>process                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed within a curved, blue line. To the right of the figure is the word "QURA" in bold, blue letters, with the words "advanced medical solutions" in smaller font underneath.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed within a curved line. To the right of the figure, the word "QURA" is written in a bold, sans-serif font. Below "QURA", the words "advanced medical solutions" are written in a smaller, sans-serif font.

Table 5-1 – Comparative Data

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Qura Advanced Medical Solutions. The logo features a stylized human figure with arms outstretched, enclosed within a curved line. To the right of the figure is the word "QURA" in a bold, sans-serif font, with the words "advanced medical solutions" in a smaller font underneath.

#### VII. PERFORMANCE DATA

## NON-CLINICAL TESTING

The following non-clinical testing was performed to support the substantial equivalence Quantum SuperPAC Tubing set and Quantum SuperPAC Cardioplegia set devices to the legally marketed predicate devices. This testing included biocompatibility evaluation, mechanical and performance verification, labeling and Instructions for Use (IFU), and verification and validation tests.

All testing passed by meeting the established requirements set for the devices. The following data were provided:

- Performance tests, mainly related to: -
	- Operating Parameters, O
	- Mechanical Integrity, O
	- o Device pressure drop,
	- O Spallation and Tubing Life,
	- Connection strength, o

and mainly performed according to ISO 15676.

- -Evaluation of product shelf life, including product's sterility according to EP/UPS requirements;
- -Validation of the EtO Sterilization process, according to ISO 11135:2014 [Recognition Nr.: 14-529],
- -Packaging Validation tests, according to ISO 11607-1:2019 [Previous Edition Recognition Nr. 14-454], ASTM F1886/F1886M-16 [Recognition Nr. 14-501], EN 868-5 and ASTM F1929-15 [Recognition Nr. 14-484],
- -Biocompatibility of the finished product, according to ISO 10993-1:2018 [Recognition Nr. 2-258] and FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process".

### Animal Study

No animal studies have been performed except for mandatory biocompatibility tests according to International Standard ISO 10993-1:2018 [Recognition Nr. 2-258] and FDA Guidance "Use of International ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

### CLINICAL TESTING

No clinical data have been included in the current Traditional 510(k) submission to support substantial equivalence to legally marketed predicate and reference devices.

#### VIII. CONCLUSIONS

Based on the indications for use, technological characteristics, results of non-clinical testing, and comparison to predicate and reference devices, Quantum SuperPAC Tubing set and Quantum SuperPAC Cardioplegia set devices have been shown to be substantially equivalent to legally marketed predicate and reference devices.

---

**Source:** [https://fda.innolitics.com/device/K213540](https://fda.innolitics.com/device/K213540)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
