DEKA ACE Pump System

K213536 · Deka Research and Development · QFG · Jul 25, 2023 · Clinical Chemistry

Device Facts

Record IDK213536
Device NameDEKA ACE Pump System
ApplicantDeka Research and Development
Product CodeQFG · Clinical Chemistry
Decision DateJul 25, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5730
Device ClassClass 2
AttributesTherapeutic

Intended Use

The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages 13 and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription. The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator-entered blood glucose, insulin to carbohydrate ratio, target glucose values, and current insulin on board.

Device Story

DEKA ACE Pump System is an Alternate Controller Enabled (ACE) infusion pump for subcutaneous insulin delivery; manages diabetes mellitus in patients 13+. System components: durable pump (rechargeable Li-ion battery), single-use cassette (microfluidic valves, reservoir, Acoustic Volume Sensing), and wireless remote controller (touch display). Inputs: user-entered data (carbs, blood glucose, insulin parameters) and commands from compatible automated insulin dosing software/iCGM. Operation: pump executes basal/bolus delivery based on programmed rates or external controller commands; AVS technology monitors fluid volume. Used in home environment by patient. Output: insulin delivery; data logs (delivery, alarms, connectivity). Benefits: precise, automated insulin management; interoperability with iCGM/automated dosing systems. Healthcare providers use logs for clinical decision-making. Safety features: occlusion detection, failsafe communication protocols, and alarm systems.

Clinical Evidence

No clinical data was obtained in support of this premarket submission. Substantial equivalence is supported by bench testing, including basal/bolus accuracy, occlusion testing, fault insertion, reliability, drug compatibility, and environmental testing per IEC 60601 series standards.

Technological Characteristics

Microprocessor-controlled micro-dosing pump with Acoustic Volume Sensing (AVS) feedback. Powered by rechargeable lithium-ion battery. Ingress protection IP28. Materials compliant with ISO-10993. Sterilization via gamma radiation (ISO 11137-1). Connectivity: wireless communication with iCGM and iAGC. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, IEC 60601-2-24, IEC 62304, ISO 14971.

Indications for Use

Indicated for subcutaneous insulin delivery for diabetes mellitus management in patients ages 13 and older. Requires prescription. Bolus calculator indicated for dosage determination based on carbohydrates, blood glucose, insulin sensitivity, insulin-to-carbohydrate ratio, target glucose, and insulin on board.

Regulatory Classification

Identification

An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.

Special Controls

Alternate controller enabled infusion pumps must comply with the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following: (i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions. (A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes ( *e.g.,* head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices. (ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration ( *e.g.,* occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability ( *e.g.,* temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low. (v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance. (vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment. (vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated. (viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing. (2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following: (i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented. (ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design. (3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following: (i) Secure authentication (pairing) to external devices. (ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices. (iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers ( *e.g.,* battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified. (v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol. (4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption ( *e.g.,* reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include: (i) A record of all drug delivery (ii) Commands issued to the pump and pump confirmations (iii) Device malfunctions (iv) Alarms and alerts and associated acknowledgements (v) Connectivity events ( *e.g.,* establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use. (7) Device labeling must include the following: (i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate. (ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters. (iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is <25 percent, 25 percent to <75 percent, 75 percent to <95 percent, 95 percent to <105 percent, 105 percent to <125 percent, 125 percent to <175 percent, 175 to 250 percent, and >250 percent of the commanded amount. (iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point. (v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps. (vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device. (vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K213536 B Applicant DEKA Research and Development C Proprietary and Established Names DEKA ACE Pump System D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QFG | Class II | 21 CFR 880.5730 - Alternate Controller Enabled Pump | CH - Clinical Chemistry | E Purpose for Submission: New device ## II Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. B Indication(s) for Use: The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages 13 and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator-entered blood glucose, insulin sensitivity, insulin to carbohydrate ratio, target glucose values, and current insulin on board. ## C Special Conditions for Use Statement(s): This device is for prescription use only. The DEKA ACE Pump System is not intended for anyone unable or unwilling to: - Test blood glucose (BG) levels as recommended by their healthcare team. - Maintain sufficient diabetes self-care skills. - See their healthcare team regularly. - Demonstrate adequate carbohydrate-counting skills (preferred, not required). The user must not: - Have difficulty seeing. - Have difficulty hearing. - Have deficient cognitive capabilities. - Have physical impairments which would make operating the pump or controller difficult. The device must be removed before Magnetic Resonance Imaging (MRI). Exposure to MRI may cause the pump and controller to move or lead to electric shocks and may result in severe injury. The device is designed to use rapid-acting U-100 insulin. The DEKA ACE Pump System is compatible with the following U-100 insulins: Humalog. ## III Device Description The DEKA ACE Pump is an Alternate Controller Enabled (ACE) infusion pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages 13 and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription. The system is able to be integrated with a Dexcom G6 interoperable Continuous Glycemic Controller (iCGM). The DEKA ACE Pump System consists of the following components: - Pump: A durable pump that incorporates fluid delivery algorithms and interfaces to an DEKA ACE Pump cassette, Remote Interface, iCGM, and iAGC. The pump is powered by a rechargeable lithium-ion battery. - Cassette: A single use pumping cassette that combines microfluidic valves, a pump chamber, insulin reservoir, and Acoustic Volume Sensing (AVS) measurement chamber. The cassette interfaces to an DEKA ACE Pump and off-the-shelf infusion set. - Remote Interface (Controller): A wireless controller that serves as the user interface to the DEKA ACE Pump system. This includes a large color touch display for ease of use. K213536 - Page 2 of 11 {2} The Controller is used in conjunction with the pump, Dexcom G6 CGM, and blood glucose monitor devices to program and monitor insulin delivery and blood glucose levels in your body. ## IV Substantial Equivalence Information: A Predicate Device Name(s): t:slim X2 insulin pump with interoperable technology B Predicate 510(k) Number(s): DEN180058 C Comparison with Predicate(s): | Device & Predicate Device(s): | K213536 | DEN180058 | | --- | --- | --- | | Device Trade Name | DEKA ACE Pump System | t:slim X2 insulin pump with interoperable technology | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | Intended for the subcutaneous delivery of insulin at variable rates for the management of diabetes mellitus in people requiring insulin. Intended to be interoperable with connected devices including CGMs and automated insulin dosing algorithms. | Same | | Operating Environment | Home Use | Same | | Insulin Delivery Modes | Both basal and bolus | Same | | General Device Characteristic Differences | | | K213536 - Page 3 of 11 {3} | Specific Drug/Biologic Use | U-100 Insulin. System has been tested with Humalog | U-100 Insulin. System has been tested with NovoLog and Humalog. | | --- | --- | --- | | Intended Population | Persons with Diabetes Mellitus ages 13 and up | Persons with Diabetes Mellitus ages 6 and up | | Insulin Basal Rate Delivery Range | 0 – 30 U/hour | 0 – 15 U/hour | | Insulin Bolus Delivery Range | Programmable from 0.05 - 25.00 Units in 0.01 Unit increments. | 0.01 U at volumes greater than 0.05 U, Max bolus volume 25 U | V Standards/Guidance Documents Referenced: - FDA Guidance “Infusion Pumps Total Product Life Cycle” dated December 2, 2014 - FDA Guidance “Applying Human Factors and Usability Engineering to Medical Devices” dated February 3, 2016 - FDA Guidance “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices” dated May 11, 2005 - FDA Guidance “Content of Premarket Submissions for Management of Cybersecurity in Medical Devices” dated October 2, 2014 - FDA Guidance “General Principles of Software Validation” dated January 11, 2002 - ISO 10993-1:2009 Biological Evaluation of Medical Devices: Evaluation and Testing - ISO 10993-3:2014 Biological Evaluation of Medical Devices – Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive Toxicity - ISO 10993-4:2002(A-2006) Biological Evaluation of Medical Devices – Part 4: Selection of Tests for Interactions with Blood - ISO 10993-5:2009 Biological Evaluation of Medical Devices – Part 5: Tests for in Vitro Cytotoxicity - ISO 10993-10:2010 Biological Evaluation of Medical Devices – Part 10: Tests for Irritation and Skin Sensitization - ISO 10993-11:2006 Biological Evaluation of Medical Devices – Part 11: Tests for Systemic Toxicity - ISO 14971 Second Edition 2007-03-01 Medical Devices – Application of Risk Management to Medical Devices - IEC 60601-1:2005+AMD1:2012 Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance - IEC 60601-1-2 Ed. 4.0 b: 2014 Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbances- Requirements and Tests K213536 - Page 4 of 11 {4} - IEC 60601-1-6 Ed. 3.1 b. 2013 Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability - IEC 60601-1-8 Ed. 2.1 b: 2012 Medical Electrical Equipment – Part 1-8: General Requirements for Basic Safety and Essential Performance – Collateral Standard: General Requirements, Tests and Guidance for Alarm Systems in Medical Electrical Equipment and Medical Electrical Systems - IEC 60601-1-11 Issued: 2015/01/20 Ed.2 Medical Electrical Equipment – Part 1-11: General Requirements for Basic Safety and Essential Performance – Collateral Standard- Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment - IEC 62304:2006 Medical Device Software – Software Life Cycle Processes - IEC 62366-1 Ed. 1.0 Medical Devices – Part 1: Application Of Usability Engineering To Medical Devices - ISO 11607-1:2006 Packaging for Terminally Sterilized Medical Devices – Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems - ISO 11137-1: 2006 Sterilization of health care productions – Radiation – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. - ISO 11137-2: 2013 Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose ## VI Performance Characteristics: ### A. Analytical Performance #### 1. Basal delivery accuracy To assess basal delivery accuracy, 32 pumps were tested by delivering at minimum, intermediate, and maximum basal rates (0.1, 1.0, and 30 U/hr). Sixteen of the pumps were unaged and 16 had been aged to simulate three years of regular use. For both aged and unaged pumps, eight pumps were tested with an unaged cassette, and eight with a cassette which underwent one year of accelerated aging. Delivery accuracy was assessed by pumping insulin into a container on a scale and measuring the weight of the liquid at 1h, 6h, and 12h intervals for minimum and intermediate rates (0.1 and 1.0 U/hr) and 1h and 6h intervals for maximum rate (30 U/hr). The following tables report the typical basal performance (median) observed, along with the lowest and highest results observed for minimum, intermediate, and maximum basal rate settings for all pumps tested. K213536 - Page 5 of 11 {5} Table 1: Amount of fluid delivered after 1, 6, and 12 hours with 0.1 U/hr (minimum) basal rate setting | 0.1 U/hr Basal Duration | 1 hour | 6 hours | 12 hours | | --- | --- | --- | --- | | Total expected delivery volume | 0.1 U | 0.6 U | 1.2 U | | Median amount delivered [min, max] | 0.12 U [0.09, 0.17] | 0.62 U [0.57, 0.66] | 1.22 U [1.16, 1.31] | Table 2: Amount of fluid delivered after 1, 6, and 12 hours with 1 U/hr (intermediate) basal rate setting | 1 U/hr Basal Duration | 1 hour | 6 hours | 12 hours | | --- | --- | --- | --- | | Total expected delivery volume | 1.0 U | 6.0 U | 12.0 U | | Median amount delivered [min, max] | 1.02 U [0.98, 1.09] | 6.05 U [5.84, 6.22] | 12.07 U [11.73, 12.33] | Table 3: Amount of fluid delivered after 1, 6, and 12 hours with 30 U/hr (high) basal rate setting | 30 U/hr Basal Duration | 1 hour | 6 hours | | --- | --- | --- | | Total expected delivery volume | 30 U | 180 U | | Median amount delivered [min, max] | 30.16 U [29.80, 30.61] | 181.05 U [178.94, 184.46] | 2. Bolus delivery accuracy To assess bolus delivery accuracy, 32 pumps were tested by delivering at minimum, intermediate, and maximum bolus volumes (0.05, 5.0, and 25 U). Each pump delivered 25 minimum and intermediate volume boluses and 7 maximum volume boluses interspersed during periods of active basal delivery. For each pump and cassette combination, a head height, orientation, infusion set, and insulin type was chosen. An approximately even distribution of parameters were used across the 32 combinations. Sixteen of the pumps were unaged and sixteen had been aged to simulate three years of regular use. For both aged and unaged pumps, eight pumps were tested with an unaged cassette, and eight with a cassette which underwent one year of accelerated aging. Tables 4-6 below show the number (and %) of boluses within the specified range of each target bolus volume. K213536 - Page 6 of 11 {6} Table 4: Amount of fluid delivered after a 0.05 U bolus request (n=800 boluses) | Units delivered after a 0.05 U bolus request (% of commanded units) | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | <0.0125 | 0.0125 - 0.0375 | 0.0375- 0.045 | 0.045- 0.0475 | 0.0475- 0.0525 | 0.0525- 0.0550 | 0.055- 0.0625 | 0.0625- 0.0875 | 0.0875- 0.125 | >0.125 | | | (<25%) | (25-75%) | (75-90%) | (90-95%) | (95-105%) | (105-110%) | (110-125%) | (125-175%) | (175-250%) | (>250%) | | Number and percent of boluses | 0/800 (0.0%) | 53/800 (6.6%) | 202/800 (25.3%) | 107/800 (13.4%) | 278/800 (34.8%) | 80/800 (10.0%) | 69/800 (8.6%) | 11/800 (1.4%) | 0/800 (0%) | 0/800 (0%) | Table 5: Amount of fluid delivered after a 5.0 U bolus request (n=800 boluses) | Units delivered after a 5.0 U bolus request (% of commanded units) | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | <1.25 | 1.25-3.75 | 3.75-4.50 | 4.50-4.75 | 4.75-5.25 | 5.25-5.50 | 5.50-6.25 | 6.25-8.75 | 8.75-2.50 | >12.50 | | | (<25%) | (25-75%) | (75-90%) | (90-95%) | (95-105%) | (105-110%) | (110-125%) | (125-175%) | (175-250%) | (>250%) | | Number and percent of boluses | 0/800 (0.0%) | 0/800 (0.0%) | 0/800 (0.0%) | 0/800 (0.0%) | 800/800 (100%) | 0/800 (0.0%) | 0/800 (0.0%) | 0/800 (0.0%) | 0/800 (0.0%) | 0/800 (0.0%) | Table 6: Amount of fluid delivered after a 25 U bolus request (n=224 boluses) | Units delivered after a 25.0 U bolus request (% of commanded units) | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | <6.25 | 6.25-18.75 | 18.75-22.5 | 22.5-23.75 | 23.75-26.25 | 26.25-27.5 | 27.5-31.25 | 31.25-43.75 | 43.75-62.5 | >62.5 | | | (<25%) | (25-75%) | (75-90%) | (90-95%) | (95-105%) | (105-110%) | (110-125%) | (125-175%) | (175-250%) | (>250%) | | Number and percent of boluses | 0/224 (0.0%) | 0/224 (0.0%) | 0/224 (0.0%) | 0/224 (0.0%) | 222/224 (99.1%) | 2/224 (0.9%) | 0/224 (0.0%) | 0/224 (0.0%) | 0/224 (0.0%) | 0/224 (0.0%) | K213536 - Page 7 of 11 {7} K213536 - Page 8 of 11 # 3. Occlusion detection To assess occlusion detection performance and unintended bolus upon occlusion release performance, 32 pumps were tested. Sixteen of the pumps were unaged and 16 had been aged. For both aged and unaged pumps, eight pumps were tested with an unaged cassette, and eight with a cassette which underwent one year of accelerated aging. The following table depicts the typical (average) time to occlusion detection for three different situations when using U-100 insulin. Table 7: Timing of occlusion detection alarms | | Typical time to occlusion detection | Maximum time to occlusion detection | | --- | --- | --- | | 2.0 U Bolus | 1 minute 52 seconds | 2 minutes 17 seconds | | 1.0 U/hr Basal | 19 minutes 33 seconds | 27 minutes 22 seconds | | 0.1 U/hr Basal | 1 hour 55 minutes | 3 hours 9 minutes | After pumps alarmed, the occlusions were cleared and the total amount of fluid delivered was measured. Typical volumes were 0.25 U, and maximum volumes were 0.4 U. # B. Other Supportive Instrument Performance Characteristics Data ## 1. Hazard Analysis A comprehensive hazard analysis was reviewed, in which design inputs and outputs, risks, and risk mitigations for hardware and software associated with proper functioning of the insulin pump were reviewed. The sponsor performed a hazard analysis to account for the unique intended use, design elements, and risks of their ACE pump. This analysis identified hazards which could reasonably be anticipated to impact the proper use of the device, traced all identified risks to adequate design controls, and demonstrated that design features were appropriately implemented and validated. ## 2. Human Factors Human factors validation tests were conducted with the DEKA ACE Pump System. All study participants received training that was consistent with the training that patients would receive with the commercial product. Usability evaluations assessed comprehension and usability of the device for critical device tasks. Results of the study demonstrated that the DEKA ACE Pump System is validated for its intended use. {8} K213536 - Page 9 of 11 3. Biocompatibility Biocompatibility testing was performed per ISO 10993-1:2018, FDA Guidance Document: Use of International Standard ISO 10993-1 “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process,” and FDA special controls for alternate controller enabled infusion pumps for the disposable cassette, pump and the disposable cassette packaging and dust cover. All endpoints were tested adequately, and results were acceptable. 4. Sterility The sterility information was reviewed and found to be acceptable. 5. Insulin Compatibility and Stability The DEKA ACE Pump system is compatible with U-100 insulin Humalog (insulin lispro) for up to 3 days. 6. Additional Bench Testing In addition to the performance testing described above, mechanical testing, simulated use testing, and other device verification testing was conducted to demonstrate that the system meets its intended use and is safe, reliable, and all safety and reliability critical requirements have been adequately verified. Summaries for reliability, safety, and verification testing follow: | Testing to Support System Reliability | | --- | | Durability of pumping mechanism | | Acoustic Volume Sensor | | Pump Button | | Pump Alarm | | Pump and cassette mating | | Pump Battery | | Charger | | Power Button | | Remote Interface | | Testing to Support System Safety | | --- | | Environmental Safety Testing to 60601-1-11 | | Safety and Essential Performance Testing to 60601-1 | | Occlusion Detection Testing | | Design specific tests: Volume Cross Checks, Inter and Intra Delivery Leaks | | Alarms Testing | | Data Handling Testing | | Pump Activation and Deactivation Testing | {9} | Pump/Controller Connectivity Testing | | --- | | User Guide Testing | | Insulin Delivery Verification Testing | | Incidental Delivery | | Worst Case Accuracy | | Fault Insertion Testing | | Testing to Support System Verification | | --- | | Environmental Safety Testing to 60601-1-11 | | Safety and Essential Performance Testing to 60601-1 | | Occlusion Detection Testing | | Alarms Testing | | Data Handling Testing | | Pump Activation and Deactivation Testing | | Pump/Controller Connectivity Testing | | User Guide Testing | | Insulin Delivery Verification Testing | | Environmental conditions Testing | # 7. Electromagnetic Compatibility and Wireless Coexistence Electromagnetic compatibility, electromagnetic immunity and wireless coexistence testing was performed for the pump. All tests demonstrated that the device would perform as expected in the home healthcare environment. # 8. Basic Safety and Essential Performance (Electrical Safety) The sponsor provided verification evidence for compliance with the IEC 60601-1 and applicable collateral standards. Verification results support the finding of substantial equivalent for this device. # 9. Data Logging The sponsor provided a summary of pump and controller logging capability which enable the device to record critical events including insulin delivery, pump commands and confirmations, connectivity states, malfunctions, and alarms. These were reviewed and found to be adequate. # 10. Interoperability A plan and approach for interoperability were provided according to the FDA Guidance "Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices - Guidance for Industry and Food and Drug Administration Staff" and determined to be adequate to support and clearly specify expectations, requirements, and interface specifications to potential interoperable devices. In addition, their plan covered their approach to working with connected device companies regarding contractual K213536 - Page 10 of 11 {10} approaches, interfaces for data communication and exchange, and post-market reporting procedures and responsibilities (e.g., who is responsible for investigating and reporting complaints, malfunctions, and adverse events). The sponsor additionally provided validated software protocols intended to ensure secure, accurate, and reliable communication with digital interfacing devices, as well as failsafe design features to mitigate the risks associated with interruption of communication with digitally connected devices. These protocols were reviewed and found to be adequate. ## 11. Software and Cybersecurity Detailed Information on software and cybersecurity of the device was reviewed and found acceptable. ## VII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. ## VIII Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K213536 - Page 11 of 11
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...