Sequential Circulators SC-1004-DL, SC-1008-DL, SC-2004-DL, SC-2008-DL, SC-4004-DL, SC-4008-DL

K213533 · Bio Compression Systems, Inc. · IRP · Feb 16, 2022 · Physical Medicine

Device Facts

Record IDK213533
Device NameSequential Circulators SC-1004-DL, SC-1008-DL, SC-2004-DL, SC-2008-DL, SC-4004-DL, SC-4008-DL
ApplicantBio Compression Systems, Inc.
Product CodeIRP · Physical Medicine
Decision DateFeb 16, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5650
Device ClassClass 2
AttributesTherapeutic

Intended Use

Intended to temporarily relieve minor muscle aches and to temporarily increase circulation to the treated area in people who are in good health.

Device Story

Sequential Circulators are powered inflatable tube massagers consisting of a pneumatic pump and segmented garment. The pump cyclically inflates garment chambers from distal to proximal, simulating kneading and stroking to increase circulation. Intended for home use by healthy individuals. The pump utilizes a motor, air compressor, and disc valve assembly to control air flow. Models feature either softkeys with LED display or touch screen LCD. The device provides adjustable pressure, cycle time, and treatment time. Healthcare providers or patients use the device to manage minor muscle aches. Benefits include temporary relief of muscle aches and improved circulation. The device is operated by the user in a home setting.

Clinical Evidence

No clinical or animal studies were submitted or relied upon. Substantial equivalence is supported by bench testing, including electrical safety (ANSI/AAMI ES60601-1, IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and functional verification (cycle time, treatment time, pressure setting accuracy, and mode operation).

Technological Characteristics

Powered inflatable tube massager; Class II; Type BF applied part; IP21 ingress protection. Materials: 200 Denier Nylon Oxford coated with 3 mil Polyurethane. Energy source: 100-240V mains power. Connectivity: Standalone. Software: IEC 62304 Class A. Standards: ANSI/AAMI ES60601-1, ANSI/AAMI HA60601-1-11, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11.

Indications for Use

Indicated for healthy individuals to temporarily relieve minor muscle aches and pains and increase circulation to treated areas.

Regulatory Classification

Identification

A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text. February 16, 2022 Bio Compression Systems, Inc. Marc Somelofski Director of RA/OA 120 West Commercial Ave Moonachie, New Jersey 07074 Re: K213533 Trade/Device Name: Sequential Circulators SC-1004-DL, SC-1008-DL, SC-2004-DL, SC-2008-DL, SC-4004-DL. SC-4008-DL Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: January 17, 2022 Received: January 19, 2022 Dear Marc Somelofski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K213533 Device Name Sequential Circulator models SC-1004-DL, SC-1008-DL, SC-2004-DL, SC-2008-DL, SC-4004-DL, SC-4008-DL Indications for Use (Describe) Intended to temporarily relieve minor muscle aches and to temporarily increase circulation to the treated area in people who are in good health. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K213533 Date Prepared: January 24, 2022 ### I. SUBMITTER Bio Compression Systems, Inc. 120 West Commercial Avenue Moonachie, NJ 07074, USA Phone: +1-201-939-0716 Contact Person: Marc L. Somelofski ### II. DEVICE | Device Name | Sequential Circulators SC-1004-DL, SC-1008-DL, SC-2004-DL, SC-2008-DL, SC-4004-DL, SC-4008-DL | |------------------------|-----------------------------------------------------------------------------------------------| | Trade/Proprietary Name | Sequential Circulators SC-1004-DL, SC-1008-DL, SC-2004-DL, SC-2008-DL, SC-4004-DL, SC-4008-DL | | Common/Usual Name | Sequential Circulators SC-1004-DL, SC-1008-DL, SC-2004-DL, SC-2008-DL, SC-4004-DL, SC-4008-DL | | Classification Name | Powered inflatable tube massager (21 CFR 890.5650) | | Product Code | IRP | | Class | Class II | ### III. PREDICATE DEVICE 2004-OC Massage System, 2008-OC Massage System (K122112). Reference devices K030437, K122154, K133483, K160608, K183169, and K210967 are mentioned. {4}------------------------------------------------ ### IV. DEVICE DESCRIPTION Bio Compression Systems' Sequential Circulators are powered inflatable tube massagers which consist of a segmented pneumatic sleeve ("garment") connected to a pneumatic compression pump ("pump"). The pump cyclically inflates the garment's segments ("chambers") in sequence from the distal end toward the trunk of the body. The sequential inflation of the garment simulates kneading and stroking of tissues with the hands, increasing circulation on the limb worn. The core components of the pump are the motor, air compressor, disc valves, and micro switch. The air compressor generates air flow into a stationary disc valve. The motor moves a rotating disc valve. The geometry of the disc valve directs air flow and cyclically triggers the micro switch. The pump is controlled by softkeys and an LED display (models SC-1004-DL, SC-1008-DL, SC-2004-DL, SC-2008-DL) or by a touch screen LCD (models SC-4004-DL, SC-4008-DL). The device uses the Predicate Device's garments. | | Subject Device<br>(K213533) | Predicate Device<br>(K122112) | Comparison | | | |-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Indications for<br>Use | Intended to temporarily<br>relieve minor muscle<br>aches and/or pains, and to<br>temporarily increase<br>circulation to the treated<br>area in people who are in<br>good health. | The 2004-OC and 2008-OC<br>Massage Systems are<br>powered inflatable tube<br>massagers intended to<br>temporarily relieve minor<br>muscle aches and/or pains,<br>and to temporarily increase<br>circulation to the treated areas.<br>The devices are intended for<br>home use by people who are<br>in good health. | Subject Device<br>indications are more<br>concise. Other<br>devices in the<br>category with similar<br>technological<br>characteristics have<br>been cleared for<br>these indications<br>(Reference Devices<br>K160608, K210967). | | | | | Prescription/<br>over-the-<br>counter use | Over-The-Counter | Over-The-Counter | Identical | | | | | | Subject Device<br>(K213533) | Predicate Device<br>(K122112) | Comparison | | | | Operating Principle | Same as Predicate | An air compressor<br>generates air flow in a<br>stationery disc valve.<br>A motor turns the top<br>disc valve which<br>directs air and trigger's<br>a micro switch. | Identical | | | Power Supply<br>Rating | 100-240V, 50/60 Hz | 120 VAC, 60 Hz | Equivalent - the Subject<br>Device and Predicate Device<br>both connect to mains power.<br>This is the same as other<br>similar cleared devices<br>(Reference Device K160608). | | | | Rated Input | 12VDC, 3A | 120 VAC, 0.5 A | Equivalent - the Subject<br>Device has a lower input<br>voltage. This is the same as<br>other similar cleared devices<br>(Reference Device K160608). | | | | Electrical<br>Classification | Class II | Class II | Identical | | | | Applied Part | Type BF | Type BF | Identical | | | | Ingress Protection | IP21 | IP21 | Identical | | | | Cycle Time | · 1000 series: 60-120<br>seconds in 30-<br>second increments<br>• 2000/4000 series:<br>60-120 seconds in<br>15-second<br>increments | • 2004-OC:<br>90 seconds<br>• 2008-OC:<br>45 seconds | Equivalent | | | | Treatment Time | Continuous or timed<br>as follows:<br>· 1000 series: 30-120<br>minutes adjustable<br>in 30-minute<br>increments<br>• 2000/4000 series:<br>10-120 minutes in 5-<br>minute increments | Continuous | Equivalent | | | | Subject Device<br>(K213533) | Predicate Device<br>(K122112) | Comparison | | | | Pressure Range | • 20-100 mmHg,<br>adjustable in 5<br>mmHg increments<br>• 10-120 mmHg,<br>adjustable in 1<br>mmHg increments<br>• 0-120 mmHg,<br>adjustable in 1<br>mmHg increments | 30-80 mmHg,<br>adjustable in 1 mmHg<br>increments | The Subject Device has a<br>broader range of pressure<br>selections. The Subject<br>Device's ranges are similar to<br>other cleared devices in this<br>category with the same<br>technological characteristics<br>and indications (Reference<br>Devices K160608, K183169,<br>K210967). | | | | Pressure Gradient<br>(decrease/proximal<br>chamber) | -1 mmHg (8-chamber<br>models) or -2 mmHg<br>(4-chamber models) | -1 mmHg | Equivalent - the Subject<br>Device's 8-chamber models<br>are identical, and the 4-<br>chamber models have the<br>same total gradient as the<br>Predicate Device's model<br>2008-OC | | | | Weight | • SC-1004-DL, SC-<br>2004-DL: 3.3 lbs.<br>(1.5 kg)<br>• SC-1008-DL, SC-<br>2008-DL: 3.65 lbs.<br>(1.66 kg)<br>• SC-4004-DL: 3.5 lbs.<br>(1.59 kg)<br>• SC-4008-DL: 3.85<br>lbs. (1.75 kg) | 8 lbs. (3.63 kg) | Similar - The Subject<br>Device's models are lighter | | | | Dimensions | • 1000 series: 4.5" x<br>11.75" x 7.75" (114<br>mm x 298 mm x 197<br>mm)<br>• 2000/4000 series:<br>4.5" x 12" x 7.34"<br>(114 mm x 304 mm<br>x 186 mm) | 5.5" x 12" x 8"<br>(140 mm x 305 mm x<br>203 mm) | Similar | | | | Inflatable Sleeves<br>("Garments") | Uses Predicate's<br>garments | 200 Denier Nylon<br>Oxford, coated with 3<br>mil of Polyurethane | Identical or identical material<br>covering additional anatomic<br>areas (shoulder, back,<br>hand/wrist, ankle, elbow,<br>knee). This is the same<br>anatomic areas as other<br>cleared devices in this<br>category with similar<br>technological characteristics<br>and indications (K030437,<br>K122154, K133483) | | | | | Subject Device<br>(K213533) | Predicate Device<br>(K122112) | Comparison | | | | Features | • Adjustable Cycle<br>Time<br>• Adjustable Pressure<br>• Usage Meter<br>• Focus Therapy<br>Mode (2000/4000<br>series)<br>• Pause<br>• Pre-Therapy Mode<br>(SC-4008-DL)<br>• Timed Treatment<br>• Individual Chamber<br>Adjustment (4000<br>series) | • Adjustable Pressure | The Subject Device has<br>additional features. These<br>features are similar or the<br>same as other cleared<br>devices in this category with<br>similar technological<br>characteristics and<br>indications (Reference<br>Devices K160608, K183169,<br>K210967). | | | | Software Safety<br>Class (IEC 62304) | A | N/A | The Subject Device has a<br>printed circuit board and<br>simple software. This is the<br>same as other cleared<br>devices in this category with<br>similar technological<br>characteristics and the same<br>indications K160608,<br>K183169, K210967). | | | ## V. INDICATIONS FOR USE {5}------------------------------------------------ ## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE {6}------------------------------------------------ {7}------------------------------------------------ The Subject Device has the same technological characteristics with respect to design, materials used, and construction. - Subject Device and Predicate Device have the same operating principle and use . the same disc valve assembly for operation - . Components used in Subject and Predicate Devices are identical, the same material, or have the same specifications - The Predicate Device uses the Subject Device's compression garments. ● - Subject Device and Predicate Device have the same or similar performance specifications The differences between the Subject Device and Predicate Device exist in similar cleared devices and do not raise any different questions of safety and/or effectiveness. - The difference in rated input and power supply rating is equivalent and the same as similar cleared devices (Reference Device K160608). - The Subject Device's broader range of pressure selections is similar to other ● cleared devices in this category with the same technological characteristics and indications (Reference Devices K160608, K183169, K210967). - The Subject Device's additional features are the same as other cleared devices in this category with similar technological characteristics and the same indications (Reference Devices K160608, K183169, K210967). - The Subject Device has a printed circuit board and simple software. This is the ● same as other cleared devices in this category with similar technological {8}------------------------------------------------ characteristics and the same indications K160608, K183169, K210967). Additionally, the Subject Device and Predicate device have the same operating principle. - The Subject Device has additional garment models covering additional anatomic ● areas. These are the same areas as other cleared devices in this category with similar technological characteristics and the same indications (K030437, K122154, K133483). ### VII. PERFORMANCE DATA The following testing and performance data were provided in support of the substantial equivalence determination. ### Electrical safety and electromagnetic compatibility (EMC) - . ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012 - ANSI/AAMI HA60601-1-11:2015-08 ● - IEC 60601-1:2005/AMD1:2012 ● - IEC 60601-1-2:2014 ● - . IEC 60601-1-6:2010/AMD1:2013 - IEC 60601-1-11:2015 ### Predicate Device routine acceptance tests conducted on Subject Device - Observation of operation ● - Pressure testing ● - HiPot (dielectric withstand test) testing ● ### Functional verification and validation testing - Cycle time verification and validation . - Treatment time verification and validation . - Pressure setting endpoint testing . - . Operation to confirm all modes, settings, and mode/setting changes function as intended {9}------------------------------------------------ ### Software Verification and Validation Testing Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". The software for this device is considered as a "minor" level of concern and there are no cybersecurity risks. #### Clinical/Animal Studies Clinical/animal studies were not submitted, referenced, or relied on in this premarket notification submission for a determination of substantial equivalence. ### VIII. CONCLUSION The data included in this submission demonstrates that the Subject Device is substantially equivalent to the legally marketed Predicate Device and performs comparably to the Predicate Device that is currently marketed for the same intended use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...