Wrist Blood Pressure Monitor, models AOJ-35A, AOJ-35B, AOJ-35C, AOJ-35D, AOJ-35E
K213503 · Shenzhen AOJ Medical Technology Co., Ltd. · DXN · Feb 18, 2022 · Cardiovascular
Device Facts
| Record ID | K213503 |
| Device Name | Wrist Blood Pressure Monitor, models AOJ-35A, AOJ-35B, AOJ-35C, AOJ-35D, AOJ-35E |
| Applicant | Shenzhen AOJ Medical Technology Co., Ltd. |
| Product Code | DXN · Cardiovascular |
| Decision Date | Feb 18, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1130 |
| Device Class | Class 2 |
Intended Use
The Wrist Blood Pressure Monitor is intended to measure and diastolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique by an inflatable cuff wrapped around the wrist at medical facilities or at home.
Device Story
Battery-powered, automatic wrist-worn blood pressure monitor; uses oscillometric technique via inflatable cuff to measure systolic/diastolic pressure and pulse rate. Operated by patients or clinicians in home or medical settings. Device performs automated inflation, deflation, and measurement; displays results (mmHg) on LCD screen. Includes low-voltage indication and auto-shutdown. Data used for monitoring blood pressure trends; assists healthcare providers in clinical decision-making regarding hypertension management. Benefits include convenient, non-invasive, self-administered blood pressure monitoring.
Clinical Evidence
Clinical validation study conducted per ISO 81060-2:2013 with 100 adult subjects (49 female, 51 male, age >12). Study met requirements for wrist size and blood pressure distribution. Results showed mean error <5.0 mmHg and standard deviation <8.0 mmHg, confirming clinical accuracy.
Technological Characteristics
Battery-powered (3Vdc, 2xAAA) wrist-worn device. Oscillometric measurement principle. LCD display. Biocompatible materials (cytotoxicity, sensitization, irritation tested per ISO 10993). Complies with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, and ISO 80601-2-30. Dimensions: 90x66x28.5mm or 79.6x70x26.8mm depending on model.
Indications for Use
Indicated for adult patients to measure systolic and diastolic blood pressure and pulse rate via non-invasive oscillometric technique using a wrist-worn inflatable cuff. Suitable for use in medical facilities or home environments.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
- BW-602 Automatic Wrist Electronic Blood Pressure Monitor (K193628)
Reference Devices
- AOJ-30B Electronic Blood Pressure Monitor (K191180)
Related Devices
- K161845 — Electronic Blood Pressure Monitor · Shenzhen Pango Electronic Co., Ltd. · Jul 31, 2017
- K172972 — Wrist Type Automatic Blood Pressure Monitor · Shenzhen Jiacom Technology CO ., Ltd. · Mar 2, 2018
- K191180 — Electronic Blood Pressure Monitor · Shenzhen AOJ Medical Technology Co., Ltd. · Aug 23, 2019
- K141683 — ELECTRONIC BLOOD PRESSURE MONITOR · Shenzhen Urion Technology Co., Ltd. · Apr 3, 2015
- K231310 — Electronic Blood Pressure Monitor · Shenzhen Jumper Medical Equipment Co., Ltd. · Oct 27, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.
February 18, 2022
Shenzhen AOJ Medical Technology Co., Ltd. Queena Chen Regulatory Director Room 301&4F, Blk A, Building A, Jingfa IM Park, Xiaweiyuan, Gushu Community, Xixiang, Baoan Shenzhen, Guangdong 518126 China
Re: K213503
Trade/Device Name: Wrist Blood Pressure Monitor, models AOJ-35B, AOJ-35C, AOJ-35D, AOJ-35E Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: December 20, 2021 Received: December 22, 2021
Dear Queena Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213503
Device Name
Wrist Blood Pressure Monitor, models AOJ-35A, AOJ-35B, AOJ-35C, AOJ-35D and AOJ-35E
Indications for Use (Describe)
The Wrist Blood Pressure Monitor is intended to measure and diastolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique by an inflatable cuff wrapped around the wrist at medical facilities or at home.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Shenzhen AOJ Medical Technology Co., Ltd.<br>Room 301&4F, Block A, Building A,<br>Jingfa Intelligent Manufacturing Park, Xiaweiyuan, Gushu Community, Xixiang<br>Street, Bao'an District, 518126, Shenzhen, China<br>TEL: 86 755-27786026 |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Queena Chen |
| Prepare date: | December 18, 2021 |
| 2. Device name and<br>classification: | Device Name: Wrist Blood Pressure Monitor<br>Models: AOJ-35A, AOJ-35B, AOJ-35C, AOJ-35D and AOJ-35E<br>Regulation No.: 21 CFR 870.1130<br>Review Panel: Cardiovascular<br>Classification Name: Cardiovascular Diagnostic Devices<br>Product code: DXN<br>Regulatory Class: Class II |
| 3. Reason for<br>Submission: | New Application. No prior submission for this device before. |
| 3. Class III device<br>statement | Not applicable, the subject device is a Class II device. |
| 4. Predicate Devices: | Primary predicate: Dongguan E-TEST Technology Co., Ltd., BW-602 Automatic<br>Wrist Electronic Blood Pressure Monitor cleared under K193628.<br><br>Reference predicate: Shenzhen AOJ Medical Technology Co., Ltd., AOJ-30B<br>Electronic Blood Pressure Monitor cleared under K191180. |
| 5. Device Description: | The predicates has not been subject to any recall before.<br>AOJ-35 series wrist blood pressure monitor is designed as a battery driven<br>automatic on-invasive blood pressure monitor. It can automatically complete the<br>inflation, deflation and measurement, which can measure systolic and diastolic<br>blood pressure as well as the pulse rate of adult person at wrist within its claimed<br>range and accuracy via the oscillometric technique. The result will be displayed in<br>the international unit mmHg.<br><br>The device also has low voltage indication, which will be triggered when the battery<br>is low.<br><br>All the models included in this submission follow the same software, same<br>measurement principle and same specifications. The main differences are color of<br>the face shell and keys, which will not affect the safety and effectiveness of the<br>device. |
| 6. Indications for<br>Use: | The Wrist Blood Pressure Monitor is intended to measure the systolic pressure and<br>diastolic pressure, as well as the pulse rate of adult person via non-invasive<br>oscillometric technique by an inflatable cuff wrapped around the wrist at medical<br>facilities or at home. |
{4}------------------------------------------------
### 7. Predicate Device Comparison
Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate devices.
Please refer to following table to find differences between the subject device and predicate device. All the differences do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.
| ITEM | Proposed Device | Predicate Device | Comparison<br>Result |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| | AOJ-35A series | BW-602/K193628 | |
| Manufacturer | Shenzhen AOJ Medical<br>Technology Co., Ltd. | DONGGUAN E-TEST<br>TECHNOLOGY CO., LTD | --- |
| Intended<br>Use/Indications for<br>Use | The Wrist Blood Pressure<br>Monitor is intended to measure<br>the systolic pressure and<br>diastolic pressure, as well as the<br>pulse rate of adult person via<br>non-invasive oscillometric<br>technique by an inflatable cuff<br>wrapped around the wrist at<br>medical facilities or at home. | Automatic Wrist Blood Pressure<br>Monitor is for use by medical<br>professionals or at home and is a<br>non invasive blood pressure<br>measurement system intended to<br>measure the diastolic and<br>systolic blood pressures and<br>pulse rate of an adult individual<br>by using a non-invasive<br>technique in which an inflatable<br>cuff is wrapped around the<br>wrist. The cuff circumference is<br>limited to 13.5 cm~19.5 cm. | Different |
| Operational Specifications | | | |
| Principle | Oscillometric | Oscillometric | Same |
| Measurement Item | SYS, DYS, Pulse Rate | SYS, DYS, Pulse Rate | Same |
| Patient population | Adult | Adult | Same |
| Measurement site | Wrist | Wrist | Same |
| Blood pressure<br>Measurement range | 0 - 295 mmHg | 0 - 294 mmHg | Different |
| Blood pressure<br>accuracy | $\pm$ 3 mmHg | $\pm$ 3 mmHg | Same |
| Pulse rate<br>measurement range | 40-199 bpm | 40-199 bpm | Same |
| Pulse rate accuracy | $\pm$ 5% of reading | $\pm$ 5% of reading | Same |
| Cuff size | 13.5~19.5cm | 13.5~19.5cm | Same |
| Display | Blood Pressure (Systolic and<br>Diastolic), Pulse rate, Time,<br>Date, BP Indicating Bar, Low<br>Battery Icon, Heart Icon,<br>Memory Record Number | Blood Pressure (Systolic and<br>Diastolic), Pulse rate, Date,<br>Time, BP Indicating Bar, Low<br>Battery Icon, Heart Icon,<br>Memory Record Number | Same |
| Screen | LCD | LCD | Same |
| Auto shutdown | YES | YES | Same |
| Operating<br>environment | Temperature: 5°C~40°C<br>Humidity: 15%-90% RH,<br>Atmospheric pressure:<br>70 kPa - 106 kPa | Temperature: 5°C~40°C<br>Humidity: 15%-90% RH,<br>Atmospheric pressure:<br>86 kPa - 106 kPa | Different |
| Storage environment | Ambient Temperature:<br>20°C to 55°C | Ambient Temperature:<br>20°C to 65°C | Same |
| | Relative Humidity: 10-93% RH,<br>Atmospheric pressure:<br>70 kPa - 106 kPa | Relative Humidity: 15-95% RH,<br>Atmospheric pressure:<br>86 kPa - 106 kPa | |
| Battery type | 3Vdc (2 *AAA batteries) | 3Vdc (2 *AAA batteries) | Same |
| Weight | Approx. 126 g without battery | Approx. 150 g without battery | Different |
| Dimensions | AOJ-35A/AOJ-35B/AOJ-35D:<br>90 mm x 66 mm x 28.5 mm<br>AOJ-35C/AOJ-35E:<br>79.6 mm x 70 mm x 26.8 mm | 72.69 mm x 64 mm x 28 mm | |
| Patient Contacting | Surface-contacting,<br>Less than 24 h | Surface-contacting,<br>Less than 24 h | Same |
| Biocompatibility<br>evaluation | Cytotoxicity, skin sensitization<br>and irritation | Cytotoxicity, skin sensitization<br>and irritation | Same |
| Electrical safety | IEC 60601-1<br>IEC 60601-1-11<br>ISO 80601-2-30 | IEC 60601-1<br>IEC 60601-1-11<br>ISO 80601-2-30 | Same |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Same |
Table 1 Comparison between predicate BW-602 and the subject device
{5}------------------------------------------------
Table 2 Comparison between predicate AOJ-30B and the subject device
| ITEM | Proposed Device<br>AOJ-35A series | Predicate Device<br>AOJ-30B/K191180 | Comparison<br>Result |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacturer | Shenzhen AOJ Medical Technology Co., Ltd. | Shenzhen AOJ Medical Technology Co., Ltd. | --- |
| Intended<br>Use/Indications for<br>Use | The Wrist Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique by an inflatable cuff wrapped around the wrist at medical facilities or at home. | The Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home. | Same |
| Operational Specifications | | | |
| Principle | Oscillometric | Oscillometric | Same |
| Measurement Item | SYS, DYS, Pulse Rate | SYS, DYS, Pulse Rate | Same |
| Patient population | Adult | Adult | Same |
| Measurement site | Wrist | Arm | Same |
| Blood pressure<br>measurement range | 0 - 295 mmHg | 0 - 295 mmHg | Same |
| Accuracy | ± 3 mmHg | ± 3 mmHg | Same |
| Heart rate<br>measurement range | 40-199 bpm | 40-199 bpm | Same |
| Accuracy | ± 5% of reading | ± 5% of reading | Same |
| Cuff size | 13.5~19.5cm | 22~36cm | Different |
| Display | Blood Pressure (Systolic and Diastolic), Pulse rate, Date, Time, WHO BP Classification Indicating Bar, Low Battery | Blood Pressure (Systolic and Diastolic), Pulse rate, Date, Time, WHO BP Classification Indicating Bar, Low Battery | Same |
{6}------------------------------------------------
| | Icon, Heart Icon, Memory Record Number | Icon, Heart Icon, Memory Record Number | |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------|
| Auto shutdown | YES | YES | Same |
| Operating environment | Temperature: 5°C~ 40°C<br>Humidity: 15%-90% RH,<br>Atmospheric pressure:<br>70 kPa - 106 kPa | Temperature: 5°C~ 40°C<br>Humidity: 15%-90% RH,<br>Atmospheric pressure:<br>70 kPa - 106 kPa | Same |
| Storage environment | Ambient Temperature:<br>-20°C to 55°C<br>Relative Humidity: 10-93% RH,<br>Atmospheric pressure:<br>70 kPa - 106 kPa | Ambient Temperature:<br>-20°C to 55°C<br>Relative Humidity: 10-93% RH,<br>Atmospheric pressure:<br>70 kPa - 106 kPa | |
| Battery type | 3Vdc (2 *AAA batteries) | 6Vdc (4 *AA batteries) | Different |
| Weight | Approx. 126 g without battery | Approx. 483.8 g without battery | Different |
| Dimensions | AOJ-35A/AOJ-35B/AOJ-35D:<br>90 mm x 66 mm x 28.5 mm<br>AOJ-35C/AOJ-35E:<br>79.6 mm x 70 mm x 26.8 mm | 138 mm x 120 mm x 59 mm | |
| Patient Contacting | Surface-contacting,<br>Less than 24 h | Surface-contacting,<br>Less than 24 h | Same |
| Biocompatibility evaluation | Cytotoxicity, skin sensitization and irritation | Cytotoxicity, skin sensitization and irritation | Same |
| Electrical safety | IEC 60601-1<br>IEC 60601-1-11<br>ISO 80601-2-30 | IEC 60601-1<br>IEC 60601-1-11<br>ISO 80601-2-30 | Same |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Same |
As seen in the comparison tables, the subject and predicate devices have almost the same design features and performance specifications. The differences between the subject and predicate devices will not raise different questions of safety or effectiveness. Moreover, as demonstrated in the bench testing, the different technological characteristics do not affect the safety and effectiveness of the subject device.
### 8. Performance Testing:
Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.
#### Non-Clinical Data:
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility testing
The biocompatibility evaluation for the device were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the battery of testing included the following tests:
- □ Cytotoxicity
- □ Skin Sensitization
- □ Skin Irritation
#### Electrical safety and electromagnetic compatibility (EMC)
{7}------------------------------------------------
Electrical safety and EMC testing were conducted on the device complies with the IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety, IEC 60601-1-11 MEDICAL ELECTRICAL EQUIPMENT -Part 1-11: General requirements for basic safety and essential performance -Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, and the IEC 60601-1-2 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Reguirements and tests standard for EMC.
#### Bench Testing
Bench testing was conducted on the device, consisting of all the accessories in the system. The system complies with the ISO 80601-2-30 Medical electrical equipment — Part 2-30: Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers for performance effectiveness.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injurv or death to the patient or operator.
#### Clinical data:
The clinical testing has been conducted per ISO 81060-2: 2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type on the monitor. There are 100 subjects involved in the stufy, of which 49 femal and 51 male. All the subjects are aged greater than 12 years old. And the wrist size distribution meets the requirements described in clause 5.1.4 of the standard as shown in Table 3. Also the blood pressure distribution matches the requirements defined in clause 5.1.5 of the standard shown in the following Table 4. And themean value is less than 5.0 mmHg, and standard deviation is also less than 8.0 mmHg after final analyzed.
| Size distribution | Pecentage | Compared to the requirements in standards |
|---------------------------|-----------|-------------------------------------------|
| Upper half of the cuff | 54% | At least 40%, complied |
| Lower half of the cuff | 46% | At least 40%, complied |
| Upper quarter of the cuff | 29% | At least 20%, complied |
| Lower quarter of the cuff | 24% | At least 20%, complied |
| Upper octal of the cuff | 11% | At least 10%, complied |
| Lower octal of the cuff | 10% | At least 10%, complied |
#### Table 3 Size distribution of the subjects
Table 4 Blood pressure distribution of the subjects
| | Range | Percentage | Compared to the requirements in standards |
|-----------------------------|-----------|------------|-------------------------------------------|
| Systolic blood<br>pressure | ≤100 mmHg | 12% | At least 5%, complied |
| | ≥160 mmHg | 13% | At least 5%, complied |
| | ≥140 mmHg | 33% | At least 20%, complied |
| Diastolic blood<br>pressure | ≥100 mmHg | 16% | At least 5%, complied |
| | ≤60 mmHg | 10% | At least 5%, complied |
| | ≥85mmHg | 38% | At least 20%, complied |
#### Summarv
Based on the non-clinical performance data as documented in the device development, the subject devices were found to have a safety and effectiveness profile that is similar to the predicate device.
{8}------------------------------------------------
## 9. Conclusion:
Verification and validation testing was conducted on the subject device and all testing passed pre-specified criteria. This premarket notification submission demonstrates that the AOJ-35 series Wrist Blood Pressure Monitor is substantially equivalent to the predicate devices.