← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K213470

# PERLA® TL Posterior Osteosynthesis System (K213470)

_Spineart · NKB · Jan 21, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K213470

## Device Facts

- **Applicant:** Spineart
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Jan 21, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The PERLA® TL Posterior Osteosynthesis System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). When used for posterior noncervical pedicle screw fixation in pediatric patients, the PERLA® TL Posterior Osteosynthesis System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The PERLA® TL Posterior Osteosynthesis System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

## Device Story

PERLA® TL Posterior Osteosynthesis System is a spinal fixation system comprising screws, rods, set screws, hooks, rod connectors, and cross-connectors. It provides rigid immobilization of spinal segments to facilitate fusion. The system is used by surgeons in clinical settings to treat spinal instabilities or deformities. The device is implanted via a posterior approach. The system is manufactured from medical-grade titanium alloy and cobalt chromium. This submission adds cannulated fenestrated sagittal screws to the existing product line to accommodate varying patient anatomies and surgical techniques. The device is intended for use with autograft or allograft. It is not indicated for use with bone cement or in patients with poor bone quality (e.g., osteoporosis, osteopenia); use is restricted to saline or radiopaque dye.

## Clinical Evidence

Bench testing only. Static compression bending, static torsion, and dynamic compression bending tests were performed in accordance with ASTM F1717-18. No clinical data was provided.

## Technological Characteristics

System components include screws, rods, set screws, hooks, and connectors. Materials: medical-grade titanium alloy (ASTM F136) and cobalt chromium (ASTM F1537). Delivery: non-sterile or sterile (gamma sterilization). Pyrogenicity: USP standard, endotoxin limit 20 EU/device. Mechanical testing per ASTM F1717-18.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- PERLA® TL Posterior Osteosynthesis System ([K203506](/device/K203506.md))
- PERLA® TL Posterior Osteosynthesis System ([K203222](/device/K203222.md))
- PERLA® TL Posterior Osteosynthesis System ([K193396](/device/K193396.md))
- ROMEO® 2 Posterior Osteosynthesis System ([K172101](/device/K172101.md))
- CD Horizon Spinal System ([K130646](/device/K130646.md))
- XIA 3 Spinal System ([K013823](/device/K013823.md))
- XIA 3 Spinal System ([K091291](/device/K091291.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 21, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Spineart Franck Pennesi Chief Technical Officer 3 Chemin du Pré Fleuri Plan Les Ouates, Geneva 1228 Switzerland

Re: K213470

Trade/Device Name: PERLA® TL Posterior Osteosynthesis System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP Dated: October 25, 2021 Received: October 28, 2021

Dear Franck Pennesi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

### Indications for Use

510(k) Number (if known)

### K213470

Device Name

PERLA® TL Posterior Osteosynthesis System

### Indications for Use (Describe)

The PERLA® TL Posterior Osteosynthesis System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). When used for posterior noncervical pedicle screw fixation in pediatric patients, the PERLA® TL Posterior Osteosynthesis System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The PERLA® TL Posterior Osteosynthesis System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

# Traditional 510k Perla® TL Posterior Osteosynthesis System

Image /page/3/Picture/2 description: The image shows the logo for Spineart. The logo consists of the word "spineart" in a modern, sans-serif font. The letters "s", "p", "i", "n", and the first "e" are in black, while the "art" portion of the word is in a purple color that matches the butterfly above the text. Above the text is a stylized butterfly in purple, positioned to the right of the word "spineart".

## 510(k) SUMMARY

| 510k                                                                                                                                                                                                                                                                                                                                                                                                                                | TRADITIONAL                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Basis for submission                                                                                                                                                                                                                                                                                                                                                                                                                | Product Line Extension of the PERLA® TL Posterior Osteosynthesis System                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | previously cleared under K193396; K203222; K203506                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Submitted by                                                                                                                                                                                                                                                                                                                                                                                                                        | SPINEART                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | 3 Chemin du Pré Fleuri                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | 1228 PLAN LES OUATES                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | GENEVA SWITZERLAND                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Contacts                                                                                                                                                                                                                                                                                                                                                                                                                            | Franck PENNESI Chief Technical Officer                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | Phone: +41 22 570 1200 Fax: +41 22 594 8306                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | Mail: fpennesi@spineart.com                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | Regulatory contact:                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | Idée Consulting - Dr Isabelle DRUBAIX (PhD) idee-consulting@bbox.fr                                                                                                                                                                                                                                                                                                                                                                                                 |
| Date Prepared                                                                                                                                                                                                                                                                                                                                                                                                                       | January 19, 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Common Name                                                                                                                                                                                                                                                                                                                                                                                                                         | Pedicle screw spinal system                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Trade Name                                                                                                                                                                                                                                                                                                                                                                                                                          | PERLA® TL Posterior Osteosynthesis System                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification Name                                                                                                                                                                                                                                                                                                                                                                                                                 | Thoracolumbosacral pedicle screw system                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Class                                                                                                                                                                                                                                                                                                                                                                                                                               | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code                                                                                                                                                                                                                                                                                                                                                                                                                        | NKB, KWP                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| CFR section                                                                                                                                                                                                                                                                                                                                                                                                                         | 888.3070                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device panel                                                                                                                                                                                                                                                                                                                                                                                                                        | ORTHOPEDIC                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Legally marketed<br>predicate devices                                                                                                                                                                                                                                                                                                                                                                                               | Primary predicate: PERLA® TL Posterior Osteosynthesis System by Spineart<br>(K203506)                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | Additional predicates: PERLA® TL Posterior Osteosynthesis System by Spineart<br>(K203222, K193396); ROMEO® 2 Posterior Osteosynthesis System by Spineart<br>(K172101); CD Horizon Spinal System by Medtronic Sofamor Danek USA, Inc<br>(K130646); XIA 3 Spinal System by Stryker Spine (K013823, K091291)                                                                                                                                                           |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | Indications for use                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative<br>disc disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal<br>tumor; and failed previous fusion (pseudarthrosis). When used for posterior non-<br>cervical pedicle screw fixation in pediatric patients, the PERLA® TL posterior<br>osteosynthesis system is indicated as an adjunct to fusion to treat adolescent |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Description of the medical device                                                                                                                                                                                                                                                                                                                                                                                                   | idiopathic scoliosis. The PERLA® TL posterior osteosynthesis system is intended to<br>be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited<br>to a posterior approach.                                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | The PERLA® TL posterior osteosynthesis system consists of a range of screws, rods,<br>set screws, hooks, rod connectors and cross-connectors. These connecting<br>components can be rigidly locked to the rod in a variety of configurations to be<br>adapted for the individual case. The Perla®TL system is manufactured from<br>medical grade titanium alloy and medical grade cobalt chromium conforming<br>respectively to standards ASTM F136 and ASTM F1537. |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | The PERLA® TL posterior osteosynthesis implants are delivered either non-sterile<br>or sterile (gamma sterilization) and supplied with dedicated surgical instruments.                                                                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | Bacterial endotoxin testing as specified in USP standard is used for pyrogenicity<br>testing to achieve the Endotoxin limit of 20 EU / device.                                                                                                                                                                                                                                                                                                                      |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | The subject product line extension of the PERLA® TL Posterior Osteosynthesis<br>System manufactured by Spineart (K193396; K203222; K203506) consists of<br>addition of cannulated fenestrated sagittal screws in various sizes and designs to<br>accommodate different patient anatomies or surgeon's techniques.                                                                                                                                                   |
| Technological<br>characteristics<br>compared to the<br>predicate devices                                                                                                                                                                                                                                                                                                                                                            | As was established in this submission, the PERLA® TL added components are<br>substantially equivalent and have the same technological characteristics to<br>predicate devices in areas including indications for use, function, material<br>composition, design, range of sizes and mechanical performance.                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     | The safety and effectiveness of cannulated fenestrated screws have not been<br>established when used in conjunction with bone cement or for use in patients with<br>poor bone quality (e.g., osteoporosis, osteopenia). This device is intended only to<br>be used with saline or radiopaque dye.                                                                                                                                                                   |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Discussion of Testing                                                                                                                                                                                                                                                                                                                                                                                                               | Static compression bending, static torsion and dynamic compression bending<br>testing were conducted in conformance to ASTM F1717-18. Results demonstrated<br>substantial equivalence between Cannulated Fenestrated Sagittal Screw added<br>component to the PerlaⓇTL system and the identified predicate devices.                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Conclusion                                                                                                                                                                                                                                                                                                                                                                                                                          | Based on the design features, technological characteristics, feature comparisons,<br>indications for use, and non-clinical performance testing, the added components<br>to the PERLA® TL posterior osteosynthesis system has demonstrated substantial<br>equivalence to the identified predicate devices.                                                                                                                                                           |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K213470](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K213470)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
