Fort Defiance Industries FRONT-LINE Field Sterilizer FL120, Fort Defiance Industries FRONT-LINE Field Sterilizer FL135

K213457 · Fort Defiance Industries, LLC · FLE · Mar 2, 2022 · General Hospital

Device Facts

Record IDK213457
Device NameFort Defiance Industries FRONT-LINE Field Sterilizer FL120, Fort Defiance Industries FRONT-LINE Field Sterilizer FL135
ApplicantFort Defiance Industries, LLC
Product CodeFLE · General Hospital
Decision DateMar 2, 2022
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.6880
Device ClassClass 2

Indications for Use

The Fort Defiance Industries FRONT-LINE Field Sterilizers are autoclaves designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped and unwrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities.

Device Story

FRONT-LINE Field Sterilizers (FL120, FL135) are tabletop autoclaves for healthcare facilities. Devices use electric resistive heaters to generate steam within a chamber boiler for sterilization. Operation involves Steam Flush Pressure Pulse (SFPP) cycles to purge air, followed by steam exposure and a drying phase. Users interact via a touchscreen controller to select from 10 pre-programmed cycles. Safety features include door interlocks, overtemperature/overpressure protection, and a pressure safety relief valve. Exhaust air passes through a 0.3-micron HEPA filter and heat exchanger for moisture reclamation. The device ensures sterility of medical/dental instruments, supporting clinical infection control workflows.

Clinical Evidence

Bench testing only. Validation performed per ANSI/AAMI ST55:2016. Testing included sterilization effectiveness (overkill method), biological performance (textiles, wrapped instruments, dental handpieces), moisture retention, and software verification. All tests passed.

Technological Characteristics

Steam sterilizer with 316L stainless steel chamber and 6061-T6 aluminum door. Uses electric resistive heaters for steam generation. Features touchscreen PLC controller, 0.3-micron HEPA filter, and heat exchanger. Complies with ASME Section VIII Div 1, IEC 61010-1, IEC 61010-2-040, and IEC 61326-1. Dimensions: FL120 (12"x23"), FL135 (13.5"x24").

Indications for Use

Indicated for sterilization of heat- and moisture-stable medical, dental, and surgical materials, including wrapped/unwrapped solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities.

Regulatory Classification

Identification

A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION". March 2, 2022 Fort Defiance Industries LLC Chris Coleman Quality Management Representative 2411 Maremont Parkway Loudon, Tennessee 37774 ### Re: K213457 Trade/Device Name: Fort Defiance Industries FRONT-LINE Field Sterilizer FL120. Fort Defiance Industries FRONT-LINE Field Sterilizer FL135 Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: October 26, 2021 Received: October 27, 2021 Dear Chris Coleman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K213457 ### Device Name Fort Defiance Industries FRONT-LINE Field Sterilizer FL120, Fort Defiance Industries FRONT-LINE Field Sterilizer FL135 Indications for Use (Describe) The Fort Defiance Industries FRONT-LINE Field Sterilizers are autoclaves designed for sterilizing heat- and moisturestable medical, dental, and surgical materials, including wrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities. | MODEL | CYCLE | MAXIMUM<br>LOAD | EXPOSURE<br>TEMP (F/C)* | EXPOSURE<br>TIME (MIN) | NUMBER OF<br>DYNAMIC<br>AIR REMOVAL<br>PULSES | DRY TIME<br>(MINUTES) | |-------|----------------------------------|-----------------------|-------------------------|------------------------|-----------------------------------------------|-----------------------| | FL120 | IMMEDIATE<br>USE (IUSS) | 20 lbs. | 270 / 132 | 4 | 3 | 0 | | | | | 275 / 135 | 3 | 3 | 0 | | FL120 | TEXTILES | 3 TEXTILE<br>PACKS ** | 250 / 121 | 30 | 3 | 60 | | | | | 270 / 132 | 4 | 3 | 60 | | | | | 275 / 135 | 3 | 3 | 60 | | FL120 | WRAPPED INSTRUMENTS<br>/ POUCHES | 20 lbs. | 250 / 121 | 30 | 3 | 60 | | | | | 270 / 132 | 4 | 3 | 60 | | | | | 275 / 135 | 3 | 3 | 60 | | FL120 | HANDPIECES | 12 HANDPIECES | 270 / 132 | 4 | 5 | 60 | | | | | 275 / 135 | 3 | 5 | 60 | | FL135 | IMMEDIATE<br>USE (IUSS) | 25 lbs | 270 /132 | 4 | 3 | 0 | | | | | 275 / 135 | 3 | 3 | 0 | | FL135 | TEXTILES | 3 TEXTILE<br>PACKS ** | 250 / 121 | 30 | 3 | 60 | | | | | 270 / 132 | 4 | 3 | 60 | | | | | 275 / 135 | 3 | 3 | 60 | | FL135 | WRAPPED INSTRUMENTS<br>/ POUCHES | 25 lbs | 250 / 121 | 30 | 3 | 60 | | | | | 270 / 132 | 4 | 3 | 60 | | | | | 275 / 135 | 3 | 3 | 60 | | FL135 | HANDPIECES | 12 HANDPIECES | 270 / 132 | 4 | 5 | 60 | | | | | 275 / 135 | 3 | 5 | 60 | NOTES: * THE ANSI/AAMI ST55:2016 MINIMUM TEMPERATURE IS LISTED FOR ALL CYCLES. ** AAMI STANDARD TEXTILE PACK (9 TOWELS - ST 55 SECTION 5.7.1.1). {3}------------------------------------------------ Prescription Use (Part 21 CFR 801 Subpart D) - X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all capitalized and in a sans-serif font. The logo is simple and modern. The color is bright and eye-catching. # 510(k) Summary for Fort Defiance Industries FRONT-LINE Field Sterilizer FL120 & FL135 K213457 Submitter / 510(k) Owner: Fort Defiance Industries LLC 2411 Maremont Parkway Loudon, TN 37774 Phone: (865) 408-0100 Fax No: (866) 732-8086 Contact: Chris Coleman Quality Management Representative Fort Defiance Industries LLC Phone: (865) 657-6108 (866) 732-8086 Fax: Submission Date: March 2nd, 2022 {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for FDI in orange. The letters are large and bold. There are white lines that cut through the letters, giving the logo a modern look. #### 1. Device Name | Trade Name: | Fort Defiance Industries FRONT-LINE Field Sterilizer FL120, Fort<br>Defiance Industries FRONT-LINE Field Sterilizer FL135 | |----------------------|---------------------------------------------------------------------------------------------------------------------------| | Model Name(s): | FL120 / FL135 | | Common/usual Name: | Electronic autoclave, steam sterilizer | | Classification Name: | Steam Sterilizer<br>Class II Device - 21 CFR 880.6880 Product Code FLE | #### 2. Predicate Device K111736, Tuttnauer EZ11 Plus Electronic Tabletop Autoclave, product code FLE, cleared December 8, 2011 #### 3. Description of Device The Fort Defiance Industries FRONT-LINE Field Sterilizers are autoclaves designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped and unwrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities. The devices use electrical resistive heaters to produce steam as the sterilizing agent. The devices adhere to sterilization processes and requirements that are universally accepted sterilization standards for Tabletop Steam Sterilizers. A touchscreen controller is used for monitoring and control of the device. The devices are equipped with multiple safety features. The door includes a safety locking mechanism and door switch. The door switch does not permit cycle operation unless the door is sealed closed and does not permit the door to be opened until the internal chamber pressure is at atmospheric pressure. Additional safety features include electronic overtemperature and overpressure protection, an independent high temperature cutoff switch, and a pressure safety relief valve for the pressure vessel. Only United States Food and Drug Administration cleared sterilization accessories should be used with these autoclaves. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all capitalized and in a sans-serif font. The logo is simple and modern. The color is bright and eye-catching. #### 4. Design and Materials The FRONT-LINE Field Sterilizers, models FL120 and FL135, are steam sterilizers composed of a stainless-steel chamber, an aluminum chamber door, electric resistive heaters, a plumbing system with a clean water tank, control system, and an aluminum frame providing a housing for all components. The devices are electronically controlled, and each device includes 10 pre-programmed sterilization cycles. All materials in the sterilization system and plumbing system are deemed acceptable for steam sterilizers. #### న. Technology The FRONT-LINE series of autoclaves are steam sterilizers composed of a pressure vessel that utilizes electric resistive heaters to produce the sterilizing agent, steam, from a chamber boiler. The autoclaves contain an electronic control system that automatically initiates programmed Steam Flush Pressure Pulse (SFPP) phases to purge atmospheric air from the chamber followed by a timed steam exposure and drying phase. Exhaust air purged during the dry phase passes through a heat exchanger to condense water vapor and reclaim as much condensate as possible for reuse. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for FDI. The logo is orange and features the letters "FDI" in a bold, sans-serif font. The letters are connected to each other, and there are white lines running through the letters, creating a sense of movement. The logo is simple and modern, and the orange color is eye-catching. #### 6. Intended Use The Fort Defiance Industries FRONT-LINE Field Sterilizers are autoclaves designed for sterilizing heat- and moisture-stable medical, dental, and surgical materials, including wrapped and unwrapped, solid, porous, and hollow items (e.g., dental handpieces, suction pipes) in healthcare facilities. | Model | Cycle | Maximum<br>Load | Exposure<br>Temperature¹<br>(°F/°C) | Exposure<br>Time<br>(minutes) | Dry Time<br>(minutes) | |-------|-------------------------------------|------------------|-------------------------------------|-------------------------------|-----------------------| | FL120 | Immediate Use<br>(IUSS) | 20 lbs. | 270/132<br>275/135 | 4<br>3 | 0 | | | Textiles | 3 Textile Packs² | 250/121<br>270/132<br>275/135 | 30<br>4<br>3 | 60 | | | Wrapped<br>Instruments /<br>Pouches | 20 lbs. | 250/121<br>270/132<br>275/135 | 30<br>4<br>3 | 60 | | FL120 | Handpieces | 12 Handpieces | 270/132 | 4 | 60 | | | | | 275/135 | 3 | | | FL135 | Immediate Use<br>(IUSS) | 25 lbs. | 270/132<br>275/135 | 4<br>3 | 0 | | | Textiles | 3 Textile Packs² | 250/121<br>270/132<br>275/135 | 30<br>4<br>3 | 60 | | | Wrapped<br>Instruments /<br>Pouches | 25 lbs. | 250/121<br>270/132<br>275/135 | 30<br>4<br>3 | 60 | | FL135 | Handpieces | 12 Handpieces | 270/132 | 4 | 60 | | | | | 275/135 | 3 | | 1 The ANSI/AAMI ST55:2016 minimum temperature is listed for all cycles. 2 AAMI Standard Textile Pack (9 towels - ST 55 section 5.7.1.1). {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for FDI. The logo is orange and features the letters "FDI" in a bold, sans-serif font. The "F" and "D" are connected by a horizontal line, and there is a diagonal line cutting through the top right corner of the "I". The logo is simple and modern. #### 7. Comparison of Technology The previously cleared device and the FRONT-LINE Field Sterilizers FL120 and FL135 are traditional steam sterilizers of similar chamber size using AAMI approved Steam Flush Pressure Pulse (SFPP) dynamic air removal cycles. The previously cleared device and the FRONT-LINE series sterilizers use electric heaters to generate steam and touchscreen controllers for cycle control. Based on the information provided below, there are negligible differences noted between the FRONT-LINE series and the predicate with regards to chamber size, cycle parameters, and configuration of steam generation. The differences do not affect safety or effectiveness of the new devices compared to the predicate. Table 1 provides a summary of performance and safety characteristics of the proposed new devices and the predicate. | Performance<br>and/or Safety<br>Characteristic | New Device<br>Fort Defiance Industries<br>FRONT-LINE Field Sterilizers<br>FL120 & FL135<br>K213457 | Predicate Device<br>Electronic Tabletop Autoclave<br>Tuttnauer EZ11 Plus<br>K111736 | |----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | The Fort Defiance Industries<br>FRONT-LINE Field Sterilizers<br>are autoclaves designed for<br>sterilizing heat- and moisture-<br>stable medical, dental, and<br>surgical materials, including<br>wrapped and unwrapped, solid,<br>porous, and hollow items (e.g.,<br>dental handpieces, suction pipes)<br>in healthcare facilities. | The EZ Plus series of autoclaves are<br>autoclaves designed for sterilizing<br>medical and surgical goods,<br>including both wrapped and<br>unwrapped, solid, hollow, and<br>porous products and goods defined<br>as hollow (e.g. dental handpieces,<br>suction pipes) in ophthalmic, dental,<br>and medical clinics; in first aid<br>rooms; and in small laboratories. | | Operating<br>Principle | Steam Flush Pressure Pulse<br>(SFPP).<br>Steam is the sterilization agent.<br>(traditional method). | Steam Flush Pressure Pulse<br>(SFPP).<br>Steam is the sterilization agent<br>(traditional method). | | Performance<br>and Safety<br>Standards<br>Followed | ANSI/AAMI ST55<br>ASME Sec. VIII Div 1<br>IEC 61010-1<br>IEC 61326-1<br>IEC 61010-2-040 | ANSI/AAMI ST55<br>ASME Sec. VIII Div 1<br>UL 61010-1<br>EN 61326-1<br>UL 61010-2-040 | | Performance<br>and/or Safety<br>Characteristic | New Device<br>Fort Defiance Industries<br>FRONT-LINE Field Sterilizers<br>FL120 & FL135<br>K213457 | Predicate Device<br>Electronic Tabletop Autoclave<br>Tuttnauer EZ11 Plus<br>K111736 | | | Factory Pre-Programmed<br>Steam Flush Pressure Pulse<br>(SFPP) Conditioning Phase | Factory Pre-Programmed<br>Steam Flush Pressure Pulse<br>(SFPP) Conditioning Phase | | Sterilization<br>Cycles | 30 min exposure @ 250°F<br>60 min dry time - Default | 30 min exposure @ 250°F<br>1 min dry time - Default | | | 4 min exposure @ 270°F<br>60 min dry time - Default | 3 or 4 min exposure @ 270°F<br>Dry time dependent on load | | | 3 min exposure @ 275°F<br>60 min dry time - Default | | | Chamber Size<br>& Volume | FL120<br>12" dia. X 23" lg.<br>1.6 cf. (45.3 L)<br>FL135<br>13.5" dia. X 24" lg.<br>2.00 cf. (56.35 L) | 11" dia. X 19.8" lg.<br>1.0 cf. (28.5 L) | | Chamber<br>Design and<br>Construction | Single Wall Chamber<br>ASME Section VIII, Div. I<br>certified<br>Stainless steel (316L) chamber<br>and aluminum door (6061-T6) | Single Wall Chamber<br>ASME Section VIII, Div. I<br>certified<br>Stainless steel (316L) chamber<br>and stainless-steel door | | Shelves / Trays | Hard anodized aluminum shelf | Stainless Steel wire trays | | Steam Source | Internal chamber boiler via<br>partitioned section | Internal chamber boiler along<br>length of chamber | | Steam<br>Generation<br>Mechanism | Electric resistive heaters<br>submerged in chamber boiler | External chamber wall electric band<br>heaters | | Dying Process | External air pulled through 0.3-<br>micron HEPA filter then heated<br>for drying | External air pulled through 0.2-<br>micron HEPA filter then heated<br>for drying | | Control<br>Technology | Electronic controller touch screen<br>(PLC) | Electronic controller touch screen<br>(PLC) | | Performance<br>and/or Safety<br>Characteristic | New Device<br>Fort Defiance Industries<br>FRONT-LINE Field Sterilizers<br>FL120 & FL135 | Predicate Device<br>Electronic Tabletop Autoclave<br>Tuttnauer EZ11 Plus<br>K111736 | | Process<br>Monitors | Chamber pressure transmitter.<br>Dual element chamber<br>temperature sensor. | Chamber pressure transmitter.<br>Dual element chamber<br>temperature sensor. | | Door Failsafe | Automatic Door Interlock | Automatic Door Interlock | | Primary<br>Overtemp and<br>Overpressure<br>Control | PLC Controller -<br>Utilizing RTDs and Pressure<br>Transmitter | PLC Controller -<br>Utilizing RTDs and Pressure<br>Transmitter | | Secondary<br>Overtemp<br>Control | Thermocouple Temperature<br>Cut-off Switch (TCO) | Capillary Temperature<br>Cut-off Switch (TCO) | | Secondary<br>Overpressure<br>Control | ASME Pressure Safety Valve | ASME Pressure Safety Valve | | Electrical<br>Input | 120VAC +/- 10%,<br>1Ø, 50/60 Hz, 15 A<br>230-240VAC +/- 10%,<br>1Ø, 50/60 Hz, 15 A | 120VAC +/- 5%, 60 Hz | ### Table 1: Comparison of Technology {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all capitalized and in a sans-serif font. There is a white line that cuts through the top of the letters. {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for FDI, which is an acronym. The letters are orange and have a unique design. The letters are connected by a white line that runs through the top of the letters. #### 8. Summary of Non-Clinical Testing Fort Defiance Industries conducted validation studies in accordance with AAMI/ANSI ST55:2016 (FDA Recognition Number 14-518). Design outputs and testing demonstrate that the FRONT-LINE Models FL120 and FL135 autoclaves meet all aspects of the standard. Performance effectiveness of the sterilizer cycles and exposure time recommendations were demonstrated by the overkill method by passing 3 successive cycles at half the programmed exposure time to guarantee a sterility assurance level (SAL) of at least 10 to probability of survival. Fort Defiance Industries validates its sterilization cycles using recommended practices, standards and guidelines developed by independent organizations such as the Association for the Advancement of Medical Instrumentation (AAMI). The Fort Defiance Industries FRONT-LINE Field Sterilizers FL120 and FL135 have been validated to meet the requirements of ANSI/AAMI ST55:2016, Tabletop Steam Sterilizers (FDA Recognition Number 14-518). Table 2 summarizes FRONT-LINE Field Sterilizer FL120 and FL135 effectiveness testing and results performed by Fort Defiance Industries. {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are connected to each other, and there is a white line running through the top of the letters. The logo is simple and modern. ### Table 2: Performance Testing Summary | Test Procedure | AAMI ST55:2016<br>Requirements | AAMI ST55:2016 Tests | Measured<br>Performance | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------| | Sterilizer Temperature<br>Control<br>&<br>Pressure Measurement<br>TP-5070<br>Note:<br>TR-5072 reports<br>Sterilizer Resolutions of<br>Temperature<br>Measurement<br>TR-5075 reports<br>Sterilizer Pressure<br>Measurement<br>TR-5077 reports Cold<br>Point Relationship | 4.4.2.1 Temperature monitoring<br>4.4.2.2 Relationship between<br>RTD and chamber cold point<br>4.4.2.4 Temperature graduations<br>of 1°C or less<br>4.4.3 -0/+6°F temp control<br>4.4.5.1 Pressure gauge of +/- 3%<br>of full scale, resolution of 1<br>PSIG or less | 5.4.2.1 - Verified by<br>inspection<br>5.4.2.2 - Verified by<br>inspection and test 5.4.3<br>5.4.2.4 - Verified by<br>inspection<br>5.4.3 - Temperature<br>instrumented chamber tests<br>to confirm steady state<br>temperature control range<br>and documentation of cold<br>point<br>5.4.5.1 - Verified by<br>inspection and by testing<br>against certified standards | PASS | | Temperature Accuracy<br>TP-5240 | 4.4.2.3 temperature accuracy of<br>+/- 1℃ over designated range | 5.4.2.3 - Verified by testing<br>against calibrated standards | PASS | | Air Removal<br>TP-5080 | 4.6.1 Air removal for dynamic<br>air removal sterilizers | A.4.6.1 - Alternate test for<br>Steam Flush Pressure Pulse<br>Sterilizers | PASS | | Software Verification<br>and Validation<br>TP-5100: Sterilizer Log<br>Verification<br>TP-5110: Sterilizer Fault<br>Conditions<br>TP-5120: General<br>Functions<br>TP-5130: Full Cycle<br>Verification | 4.4.7 Sterilizer fault conditions<br>4.4.8 Cycle documentation | 5.4.7 - Verified by inspection<br>5.4.8 - Verified by inspection | PASS | | Biological Performance<br>- Textiles<br>TP-5141 | 4.5 Biological performance of<br>sterilizers | 5.5.2 Biological performance<br>with a textile pack | PASS | | Biological Performance<br>- Wrapped Instrument<br>TP-5160 | 4.5 Biological performance of<br>sterilizers | 5.5.4 Biological performance<br>with wrapped instruments | PASS | | Biological Performance<br>- Dental Handpieces<br>TP-5170 | 4.5 Biological performance of<br>sterilizers | 5.5.5 Biological performance<br>with dental handpieces | PASS | | Test Procedure | AAMI ST55:2016<br>Requirements | AAMI ST55:2016 Tests | Measured<br>Performance | | Moisture Retention –<br>Textiles<br>TP-5190 | 4.7 Moisture Retention | 5.7.1 Textile test packs | PASS | | Moisture Retention –<br>Wrapped Instrument<br>TP-5200 | 4.7 Moisture Retention | 5.7.2 Wrapped Instr. Test<br>trays | PASS | | Moisture Retention –<br>Pouches Drying Test<br>TP-5210 | 4.7 Moisture Retention | 5.7.3 Paper-plastic peel<br>pouches | PASS | | Endotoxin and Water<br>Quality<br>TP-5220 | A.4.2.6 Water supply reservoir | A.4.2.6 Water supply<br>reservoir | PASS | | FDI Test Procedure or<br>3rd Party | Standard | Tests | Measured<br>Performance | | Labeling<br>TP-5010<br>(FDI) | AAMI ST55:2016 - 4.1.1<br>Device Markings<br>AAMI ST55:2016 - 4.1.2<br>Information Manual<br>AAMI ST55:2016 - 4.1.3<br>Service Manual | AAMI ST55:2016 - 5.1 Visual<br>inspection only | PASS | | Sterilizer Safety<br>TP-5050<br>(FDI) | AAMI ST55:2016 - 4.3.1<br>Interlock<br>AAMI ST55:2016 - 4.3.2<br>Thermal hazards<br>AAMI ST55:2016 - 4.3.3<br>Aborting cycles | AAMI ST55:2016 - 5.3.1<br>Interlock functions under<br>conditions described<br>AAMI ST55:2016 - 5.3.2<br>Thermal hazard testing per UL-<br>61010<br>AAMI ST55:2016 - 5.3.3<br>Verified by inspection | PASS | | General Inspection<br>TP-5060<br>(FDI) | 4.2.1 & 4.2.2 Pressure vessel<br>requirements<br>4.2.4 Corrosion resistance<br>4.2.5 Air filters<br>4.2.6 Water supply reservoir<br>4.4.4 Sterilizer exposure timer<br>4.4.6 Cycle completion | AAMI ST55:2016 - 5.2.1<br>Verified by compliance to<br>ASME BPVC Section VIII<br>Division 1 Code, requirements<br>in UL 61010-1 and IEC 61010-<br>2-040<br>AAMI ST55:2016 - 5.2.4<br>Material corrosion resistance<br>review<br>AAMI ST55:2016 - 5.2.5 Air<br>filter inspection necessary<br>filtration efficiency and filter<br>readily accessible<br>AAMI ST55:2016 - 5.2.6<br>Verified by inspection<br>AAMI ST55:2016 - 5.4.4<br>Verified by inspection and<br>timer testing to NIST standard<br>AAMI ST55:2016 - 5.6.6<br>Verified by inspection | PASS | | 3rd Party Testing<br>Electrical Safety | IEC 61010-1 Edition 3.1<br>2017-01 | 3rd Party test protocol | PASS | | 3rd Party Testing<br>Electrical/Sterilizer<br>Specific Safety | IEC 61010-2-040:2020 | 3rd Party test protocol | PASS | | 3rd Party Testing<br>EMC/EMI | IEC 61326-1:2020 | 3rd Party test protocol | PASS | {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are connected to each other, and the letter D is in the center. There is a white line that cuts through the top of the letters F and D. The logo is simple and modern. #### 9. Safety The Fort Defiance FRONT-LINE Field Steam Sterilizers have been designed, constructed, and tested to meet the safety and performance requirements of various national safety codes and standards. The device complies with the following standards: - ANSI/AAMI ST55:2016, Tabletop Steam Sterilizers (FDA Recognition Number 14-518). - IEC 61010-1 Edition 3.1 2017-01, "Safety Requirements for Electrical Equipment ● for Measurement, Control and Laboratory Use - Part 1 General Requirements". (FDA Recognition Number 19-34) - IEC 61326-1:2020, "Electrical Equipment for Measurement, Control, and . Laboratory Use – EMC Requirements – Part 1: General Requirements" (General Use) - . IEC 61010-2-040:2020, "Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 2-040: Particular Requirements for Sterilizers and Washer-disinfectors used to Treat Medical Materials" (General Use) - ASME Boiler and Pressure Vessel Code, Section VIII, Division I, 2019 Edition, . Rules for Construction of Pressure Vessels (General Use) {13}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the logo for FDI in orange. The letters are large and bold. A white line cuts through the top right corner of the D and the I. Table 3 summarizes FRONT-LINE Field Sterilizer FL120 and FL135 safety testing results performed by Fort Defiance Industries or a Third Party where indicated. ### Table 3: Safety Testing Summary {14}------------------------------------------------ Image /page/14/Picture/0 description: The image shows the logo for FDI in orange. The logo is made up of the letters F, D, and I. The letters are all capitalized and in a sans-serif font. The logo is simple and modern. The color is bright and eye-catching. #### 10. Conclusion The conclusion drawn from the nonclinical testing demonstrates that the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device, K111736, Class II (21 CFR 880.6880, Product code FLE).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...