Systematic literature review; ACCESS-PTS trial data; CAVA study data
A systematic literature review, including data from the ACCESS-PTS trial and CAVA studies, was used to support the expansion of the indications for use for the EkoSonic and EKOS+ Endovascular devices to include the treatment of deep vein thrombosis (DVT).
Systematic literature review; Deep vein thrombosis; Clinical evidence
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
ACCESS-PTS trial and CAVA studies; Systematic literature review of clinical studies
Patients with deep vein thrombosis (DVT)
Not applicable for this study
Safety and effectiveness for DVT treatment
Indications for Use
The EkoSonic™ Endovascular System is indicated for the: · Ultrasound facilitated, controlled and selective intravascular infusion of physician-specified fluids, including thrombolytics, for the treatment of pulmonary embolism and/or deep vein thrombosis · Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the pulmonary arteries and/or peripheral vasculature. The EKOS+ Endovascular System is indicated for the: · Ultrasound facilitated, controlled and selective intravascular infusion of physician-specified fluids, including thrombolytics, for the treatment of pulmonary embolism and/or deep vein thrombosis · Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the pulmonary arteries and/or peripheral vasculature.
Device Story
System consists of infusion catheter (IC) and ultrasonic core (USC); USC fits within IC central lumen. Device delivers physician-specified fluids and ultrasound energy to treatment zone. Used in clinical settings for percutaneous/endovascular procedures. Ultrasound facilitates delivery of thrombolytic agents into vascular blood clots. System includes temperature monitoring, feedback, and control to limit IC surface temperature to 43°C. Output power limited by software to ~50W. Healthcare providers use device to treat pulmonary embolism and deep vein thrombosis; ultrasound energy enhances thrombolytic efficacy, potentially improving patient outcomes by accelerating clot dissolution.
Clinical Evidence
No new bench or clinical testing performed. Substantial equivalence supported by systematic literature review, including data from the ACCESS-PTS trial and CAVA studies, to validate the inclusion of deep vein thrombosis (DVT) in the indications for use.
Technological Characteristics
Sterile, single-use catheter system. IC with multiple side-holes; USC with ultrasound transducers (6-30 for EkoSonic, 8-20 for EKOS+). Frequency: 2.05-2.5 MHz (EkoSonic), 1.58-2.00 MHz (EKOS+). Energy: R/F electrical converted to ultrasound. Compatible with 0.035" guide wire. Temperature feedback control limits surface to 43°C. Software-limited power output (~50W). Radiopaque markers on IC and USC elements.
Indications for Use
Indicated for patients requiring ultrasound-facilitated, controlled, selective intravascular infusion of physician-specified fluids (including thrombolytics) for treatment of pulmonary embolism and/or deep vein thrombosis, or infusion into pulmonary arteries and peripheral vasculature.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 14, 2021
Boston Scientific Corporation Ambreen Athar Regulatory Affairs, Principal Specialist Three Scimed Place Maple Grove, Minnesota 55311
Re: K213422
Trade/Device Name: EkoSonic Endovascular Device, EKOS+ Endovascular Device Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEY, KRA Dated: October 18, 2021 Received: October 20, 2021
Dear Ambreen Athar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213422
Device Name EkoSonic Endovascular Device EKOS+ Endovascular Device
Indications for Use (Describe)
The EkoSonicTM Endovascular System is indicated for the:
· Ultrasound facilitated, controlled and selective intravascular infusion of physician-specified fluids, including thrombolytics, for the treatment of pulmonary embolism and/or deep vein thrombosis
· Controlled and selective infusion of physician-specified fluids, into the pulmonary arteries and or peripheral vasculature.
The EKOS+ Endovascular System is indicated for the:
· Ultrasound facilitated, controlled and selective intravascular infusion of physician-specified fluids, including thrombolytics, for the treatment of pulmonary embolism and/or deep vein thrombosis
· Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the pulmonary arteries and/ or peripheral vasculature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Premarket Notification 510(k) Summary for K213422
### l. Submitter
Boston Scientific Corporation Three Scimed Place Maple Grove, MN 55311
Phone: 763.995.8171
Contact Person: Ambreen Athar Alternate Contact Person: Brent Hathcock
Date Prepared: December 13, 2021
#### II. Devices
| Proprietary Name: | EkoSonic Endovascular Device |
|--------------------------------|-----------------------------------------------------|
| Common or Usual Name: | Continuous Flush Catheter |
| Primary Classification Name: | Mechanical Thrombolysis Catheter (21 CFR §870.5150) |
| Primary Product Code: | QEY |
| FDA Panel/Device Class: | Cardiovascular; Class II |
| Secondary Classification Name: | Catheter, Continuous Flush (21 CFR §870.1210) |
| Secondary Product Code: | KRA |
| FDA Panel/Device Class: | Cardiovascular; Class II |
| Proprietary Name: | EKOS+ Endovascular Device |
| Common or Usual Name: | Continuous Flush Catheter |
| Primary Classification Name: | Mechanical Thrombolysis Catheter (21 CFR §870.5150) |
| Primary Product Code: | QEY |
| FDA Panel/Device Class: | Cardiovascular; Class II |
| Secondary Classification Name: | Catheter, Continuous Flush (21 CFR §870.1210) |
| Secondary Product Code: | KRA |
| FDA Panel/Device Class: | Cardiovascular; Class II |
#### lll. Predicate Devices
The proposed modification is to the Indications For Use for both the EkoSonic Endovascular Device (K191119) and EKOS PE Endovascular Device (K200648). The modification to the indications is to include Deep Vein Thrombosis (DVT). A systematic literature review is presented to support the appropriateness of the change. There is no change to the design or manufacture of the device as a result of this change, therefore, the predicate devices are the current versions of the legally marketed devices. The predicate devices will be EkoSonic Endovascular Device (K191119) and the EKOS PE Endovascular Device (K200648).
{4}------------------------------------------------
No reference devices were used in this notification.
### IV. Device Description
### EkoSonic Endovascular Device -
The sterile, single-use EkoSonic Endovascular Device consists of an Infusion Catheter (IC) and an Ultrasonic Core (USC). The USC fits within the central lumen of the IC. Together, they are used to deliver physician-specified fluids and ultrasound energy to the treatment zone.
## EKOS+ Endovascular Device -
The sterile, single-use EKQS+ Endovascular Device consists of an Infusion Catheter (IC),and an Ultrasonic Core (USC). The USC fits within the central lumen of the IC. Together, they are used to deliver physician-specified fluids and ultrasound energy to the treatment zone.
### V. Proposed Intended Use/Indications for use
The EkoSonic™ Endovascular System is indicated for the:
- Ultrasound facilitated, controlled and selective intravascular infusion of physician-. specified fluids, including thrombolytics, for the treatment of pulmonary embolism and/or deep vein thrombosis
- . Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the pulmonary arteries and/or peripheral vasculature.
The EKOS+ Endovascular System is indicated for the:
- Ultrasound facilitated, controlled and selective intravascular infusion of physician-● specified fluids, including thrombolytics, for the treatment of pulmonary embolism and/or deep vein thrombosis
- Controlled and selective infusion of physician-specified fluids, including . thrombolytics, into the pulmonary arteries and/or peripheral vasculature.
| | PREDICATE DEVICES | | SUBJECT DEVICES | |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | EkoSonic Endovascular<br>Device<br>(Predicate Device) | EKOS PE Endovascular<br>Device with CU4.0<br>(Predicate Device) | EkoSonic Endovascular<br>Device<br>(Subject Device) | EKOS+ Endovascular<br>Device<br>(Subject Device) |
| 510(k) Number | K191119 | K200648 | To be assigned | To be assigned |
| Product Code | QEY, KRA | QEY, KRA | QEY, KRA | QEY, KRA |
| Product Name | EkoSonic Endovascular<br>Device | EKOS PE Endovascular<br>Device | EkoSonic Endovascular<br>Device | EKOS+ Endovascular<br>Device |
| | PREDICATE DEVICES | | SUBJECT DEVICES | |
| | EkoSonic Endovascular | EKOS PE Endovascular | EkoSonic Endovascular | EKOS+ Endovascular |
| | Device | Device with CU4.0 | Device | Device |
| | (Predicate Device) | (Predicate Device) | (Subject Device) | (Subject Device) |
| Indications for<br>Use | The EkoSonic<br>Endovascular Device with<br>CU4.0 is indicated for the:<br>· Ultrasound facilitated,<br>controlled and selective<br>infusion of physician-<br>specified fluids, including<br>thrombolytics, into the<br>vasculature for the<br>treatment of pulmonary<br>embolism.<br>· Infusion of solutions into<br>the pulmonary arteries.<br>· Controlled and selective<br>infusion of physician-<br>specified fluids, including<br>thrombolytics, into the<br>peripheral vasculature. | The EKOS PE<br>Endovascular System [with<br>CU4.0] is indicated for the:<br>· Ultrasound facilitated,<br>controlled and selective<br>infusion of physician-<br>specified fluids,<br>including thrombolytics,<br>into the vasculature for<br>the treatment of<br>pulmonary embolism.<br>• Infusion of solutions into<br>the pulmonary arteries.<br>· Controlled and selective<br>infusion of physician-<br>specified fluids,<br>including thrombolytics,<br>into the peripheral<br>vasculature. | The EkoSonic™<br>Endovascular System is<br>indicated for the:<br>Ultrasound facilitated,<br>controlled and<br>selective intravascular<br>infusion of physician-<br>specified fluids,<br>includina<br>thrombolytics, for the<br>treatment of<br>pulmonary embolism<br>and/or deep vein<br>thrombosis<br>Controlled and<br>selective infusion of<br>physician-specified<br>fluids, including<br>thrombolytics, into the<br>pulmonary arteries<br>and/or peripheral<br>vasculature. | The EKOS+ Endovascular<br>System is indicated for the:<br>Ultrasound facilitated,<br>controlled and<br>selective intravascular<br>infusion of physician-<br>specified fluids,<br>including<br>thrombolytics, for the<br>treatment of<br>pulmonary embolism<br>and/or deep vein<br>thrombosis<br>Controlled and<br>selective infusion of<br>physician-specified<br>fluids, including<br>thrombolytics, into the<br>pulmonary arteries<br>and/or peripheral<br>vasculature. |
| Principle of<br>Operation | The EkoSonic<br>Endovascular System<br>employs ultrasound to<br>facilitate the delivery of<br>thrombolytic agents into<br>vascular blood clots. | The EKOS PE<br>Endovascular System<br>[EKOS PE Endovascular<br>Device with CU4.0]<br>employs ultrasound to<br>facilitate the delivery of<br>thrombolytic agents into<br>vascular blood clots. | The EkoSonic<br>Endovascular System<br>employs ultrasound to<br>facilitate the delivery of<br>thrombolytic agents into<br>vascular blood clots. | The EKOS+ Endovascular<br>System employs ultrasound<br>to facilitate the delivery of<br>thrombolytic agents into<br>vascular blood clots. |
| Infusion Hole<br>Pattern | Multiple side-holes | Multiple side-holes | Multiple side-holes | Multiple side-holes |
| Catheter<br>Working<br>Length | 106 cm or 135 cm | 106 cm or 135 cm | 106 cm or 135 cm | 106 cm or 135 cm |
| Treatment<br>Zone Length | 6 cm - 50 cm | 8 cm - 20 cm | 6 cm - 50 cm | 8 cm - 20 cm |
| Compatible<br>Guide Wire | 0.035" | 0.035" | 0.035" | 0.035" |
| Outer Diameter | 5.4 Fr | 7.7 Fr | 5.4 Fr | 7.7 Fr |
| Placement<br>Mode | Percutaneous/endovascular | Percutaneous/endovascular | Percutaneous/endovascular | Percutaneous/endovascular |
| Packaged<br>Sterile | Yes - EkoSonic<br>Endovascular Device | Yes - EKOS PE<br>Endovascular Device | Yes - EkoSonic<br>Endovascular Device | Yes - EKOS+<br>Endovascular Device |
| | PREDICATE DEVICES | | SUBJECT DEVICES | |
| | EkoSonic Endovascular<br>Device<br>(Predicate Device) | EKOS PE Endovascular<br>Device with CU4.0<br>(Predicate Device) | EkoSonic Endovascular<br>Device<br>(Subject Device) | EKOS+ Endovascular<br>Device<br>(Subject Device) |
| Single-Use<br>Disposable | Yes - EkoSonic<br>Endovascular Device | Yes - EKOS PE<br>Endovascular Device | Yes - EkoSonic<br>Endovascular Device | Yes - EKOS+<br>Endovascular Device |
| Materials<br>Biocompatible | Yes - EkoSonic<br>Endovascular Device | Yes - EKOS PE<br>Endovascular Device | Yes - EkoSonic<br>Endovascular Device | Yes - EKOS+<br>Endovascular Device |
| Radiopaque<br>Markers | Yes on IC<br>USC ultrasound elements<br>are radiopaque | Yes on IC<br>USC ultrasound elements<br>are radiopaque | Yes on IC<br>USC ultrasound elements<br>are radiopaque | Yes on IC<br>USC ultrasound elements<br>are radiopaque |
| Mechanism of<br>Action | Ultrasound | Ultrasound | Ultrasound | Ultrasound |
| Energy Source | R/F electrical from CU<br>converted to ultrasound | R/F electrical from CU<br>converted to ultrasound | R/F electrical from CU<br>converted to ultrasound | R/F electrical from CU<br>converted to ultrasound |
| Ultrasound<br>Transducer(s)<br>in Catheter | 6 to 30 | 8 to 20 | 6 to 30 | 8 to 20 |
| Acoustic<br>Characteristics | Frequency = 2.05 - 2.5<br>MHz | Frequency = 1.58-2.00<br>MHz | Frequency = 2.05 - 2.5<br>MHz | Frequency = 1.58-2.00<br>MHz |
| Maximum<br>Output Power<br>Limit | Power is available for<br>~100W Pulses. The power<br>output is limited by software<br>to ~50W. | Power is available for<br>~100W Pulses. The power<br>output is limited by software<br>to ~50W. | Power is available for<br>~100W Pulses. The power<br>output is limited by software<br>to ~50W. | Power is available for<br>~100W Pulses. The power<br>output is limited by software<br>to ~50W. |
| Maximum<br>EkoSonic<br>Device<br>Temperature | Temperature monitoring,<br>feedback and control<br>system limits the surface<br>temperature of the IC to<br>43°C during operation. | Temperature monitoring,<br>feedback and control<br>system limits the surface<br>temperature of the IC to<br>43°C during operation. | Temperature monitoring,<br>feedback and control<br>system limits the surface<br>temperature of the IC to<br>43°C during operation. | Temperature monitoring,<br>feedback and control<br>system limits the surface<br>temperature of the IC to<br>43°C during operation. |
### VI. Comparison of Technological Characteristics with the Predicate Devices
{5}------------------------------------------------
{6}------------------------------------------------
There are no device modifications (individually or cumulatively) to the subject devices that impact the technological characteristics of the devices since the last 510(k) submission including:
- Product specification requirements
- Raw material composition or constituents .
- . Product chemical composition or the physical and chemical properties
- Device packaging (with the exception of BSC branding on the labels) ●
The only change is to the indications for use to include the DVT as a treatment and a change to the trade name of EKOS PE to EKOS+.
### VII. Performance Data
Device testing was not required to support the change to the indications for use changes. There are no changes to the device technological characteristics. The devices are, therefore, substantially equivalent to the predicate device.
#### VIII. Clinical Data
{7}------------------------------------------------
A systematic literature review, including data from the ACCESS-PTS trial and CAVA studies, were used to support the inclusion of DVT to the indications for use with both the EkoSonic Endovascular and EKOS+ Endovascular devices.
### IX. Conclusions
The EkoSonic Endovascular Device and the EKOS+ Endovascular Device are substantially equivalent to the predicate devices. The expansion to the indications for use and the change to the trade name do not impact the technological characteristics for the system.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.