← Product Code [MAI](/productcode/MAI) · K213415

# OSSIOfiber Suture Anchor (K213415)

_OSSIO , Ltd. · MAI · Mar 28, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K213415

## Device Facts

- **Applicant:** OSSIO , Ltd.
- **Product Code:** [MAI](/productcode/MAI.md)
- **Decision Date:** Mar 28, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The OSSIOfiber® Suture Anchors are indicated for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair and Bunionectomy. Knee: Anterior Cruciate Ligament Repair (4.75-5.5 Anchors Only), Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary or adjunct fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 Anchors only). Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction. Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair (Tennis Elbow).

## Device Story

Device consists of eyelet and anchor body preloaded on inserter; used for soft tissue-to-bone fixation in orthopedic procedures. Anchor body/eyelet composed of PLDLA polymer reinforced with mineral fibers; material degrades via hydrolysis into alpha-hydroxy acids and minerals found in natural bone. Load transfers to surrounding anatomy during healing; degradation occurs over ~18 months, eliminating need for hardware removal. Used by surgeons in clinical settings. Provides secure fixation for various tendon/ligament repairs; benefits patient by avoiding secondary removal surgery.

## Clinical Evidence

Bench testing only. Static and cyclic pull-out testing performed to verify fixation strength compared to predicate. Torsional strength, driving torque, and insertion testing conducted at time zero. In-vitro degradation profile characterized. Biocompatibility established via reference to previously cleared devices and ISO 10993 standards.

## Technological Characteristics

Anchor/eyelet material: PLDLA reinforced with mineral fibers. Sterile, single-use, non-pyrogenic. Supplied with inserter and optional sutures/needles. Sterilization: Ethylene Oxide (EtO). Mechanical performance verified via static/cyclic pull-out and torque testing.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex SwiveLock Suture Anchor ([K193503](/device/K193503.md))

## Reference Devices

- OSSIO® Pin Product Family ([K181180](/device/K181180.md))
- OSSIOfiber® Hammertoe Fixation System ([K190652](/device/K190652.md), [K201803](/device/K201803.md))
- OSSIOfiber® Compression Screw ([K193660](/device/K193660.md))
- OSSIOfiber® Cannulated Trimmable Fixation Nail ([K203465](/device/K203465.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 28, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

OSSIO Ltd. % David McGurl Director, Regulatory Affairs MCRA, LLC 803 7th Street NW Washington, District of Columbia 20001

# Re: K213415

Trade/Device Name: OSSIOfiber® Suture Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI Dated: March 1, 2022 Received: March 1, 2022

# Dear David McGurl:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K213415

Device Name

# OSSIOfiber® Suture Anchor

Indications for Use (Describe)

The OSSIOfiber® Suture Anchors are indicated for fixation of suture (soft tissue) to bone in the shoulder, footlankle, knee, hand/wrist, and elbow in the following procedures:

- Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
- I Foot/Ankle: Lateral Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair and Bunionectomy.
- I Knee: Anterior Cruciate Ligament Repair (4.75-5.5 Anchors Only), Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary or adjunct fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 Anchors only).
- . Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.
- . Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair (Tennis Elbow).

| Type of Use (Select one or both, as applicable)                                                                |                                                                                                               |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span>                     ☒ Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                     □ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number

{3}------------------------------------------------

### 510(k) SUMMARY

# OSSIOfiber® Suture Anchor

#### Submitter

# Ossio Ltd.

8 HaTochen Street, Caesarea, Israel, 3079861 Phone: +972-4-9986600 Facsimile: +972-4-9986601 Contact Person: Taly Lindner Date Prepared: March 01, 2022

Name of Device: OSSIOfiber® Suture Anchor

Common or Usual Name: Fastener, Fixation, Biodegradable, Soft Tissue

Classification Name: Single/multiple component metallic bone fixation appliances and accessories

Requlatory Class: Class II, 21 C.F.R. § 888.3030

#### Product Code: MAI

#### Predicate Devices

Arthrex SwiveLock Suture Anchor (K193503) - Primary Predicate

# Reference Devices

OSSIO® Pin Product Family (K181180) OSSIOfiber® Hammertoe Fixation System (K190652, K201803) OSSIOfiber® Compression Screw (K193660) OSSIOfiber® Cannulated Trimmable Fixation Nail (K203465)

# Device Description

The OSSIOfiber® Suture Anchor consists of an eyelet and anchor body preloaded on an inserter. The anchor body and eyelet are made from poly (L-lactide-co-D,L-lactide) (PLDLA) reinforced with continuous mineral fibers. OSSIOfiber® implants have been shown to be biocompatible. The polymer content degrades by hydrolysis into alpha-hydroxy acids that are metabolized by the fibers are made from minerals that are found in natural bone. As the OSSIOfiber® implants degrade, the load transfers to the surrounding anatomy throughout the healing period of the bone. Substantial degradation takes place within approximately 18 months as shown in pre-clinical studies, thus eliminating the requirement for future hardware removal surgery. Sutures, needles and suture snare may also be provided with the device depending on configuration.

The OSSIOfiber® Suture Anchors are sterile, single-use, and non-pyrogenic.

{4}------------------------------------------------

#### Indications for Use

The OSSIOfiber® Suture Anchors are indicated for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, and elbow in the following procedures:

- .. Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.
- Knee: Anterior Cruciate Ligament Repair (4.75-5.5 Anchors Only). Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis and Quadriceps Tendon Repair. Secondary or adjunct fixation of ACL/PCL reconstruction or repair (4.75 - 5.5 Anchors only).
- Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.
- Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair (Tennis Elbow).

### Summary of Technological Characteristics

The OSSIOfiber® Suture Anchor has the same intended use, indications for use, and principles of operation, and similar material composition and design characteristics as the predicate device Arthrex SwiveLock Suture Anchor (K193503).

The material and manufacturing methods of the OSSIOfiber® anchor and eyelet are the same as that of the cleared reference devices (K181180, K190652, K201803, K193660, K203465). Sutures with or without needles are assembled with the anchor and eyelet on an inserter and sterilized with EtC.

Both the device and the predicate are supplied sterilized by EtO. Although there are slight design (e.g. length, diameters) and material differences compared to the predicate, mechanical testing demonstrated at least equivalent performance both initially and after in vitro degradation. Any differences between OSSIOfiber® Suture Anchor and its predicate device do not raise different questions of safety and effectiveness.

# Non-Clinical Data

Static pull-out and cyclic pull-out testing were performed to verify the strength and fixation properties of the OSSIOfiber® Suture Anchor, and to compare them to those of the predicate device (K193503). Testing was done initially and following in-vitro degradation. The in-vitro degradation profile was characterized. Torsional strength, driving torque, and insertion testing at time zero were also conducted for the OSSIOfiber® Suture Anchor.

Biocompatibility for the device was primarily established by use of material identical to the reference devices or use of well-established material. Biocompatibility for the implants (Anchor and Eyelet) was established primarily based on the referenced ISO 10993 data from the previously cleared reference

{5}------------------------------------------------

devices (K181180, K190652, K201803). Biocompatibility for the sutures and needles was established within their 510(k) clearances.

Biocompatibility testing for the inserter included cytotoxicity, endotoxin pyrogenicity testing, material mediated pyrogenicity testing, irritation, acute systemic toxicity and sensitization testing.

A rationale was provided to support the MR safe labeling of the device.

# Conclusions

The OSSIOfiber® Suture Anchor is as safe and effective as its predicate device, Arthrex SwiveLock Suture Anchor (K193503). The OSSIOfiber® Suture Anchors have the same intended use, indications for use, and principles of operation, and similar material composition and design characteristics as the predicate device Arthrex SwiveLock Suture Anchor (K193503). The minor differences do not alter the intended surgical use of the device and do not affect its safety and effectiveness when used as labeled. Non-clinical testing data demonstrate that the OSSIOfiber® Suture Anchor is at least as safe and effective as the predicate device. Thus, the OSSIOfiber® Suture Anchor is substantially equivalent.

---

**Source:** [https://fda.innolitics.com/device/K213415](https://fda.innolitics.com/device/K213415)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
