← Product Code QSY · K213413

# Hemostatic Dressing (K213413)

_Jiangsu Newvalue Medical Products Co., Ltd. · QSY · Sep 1, 2023 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K213413

## Device Facts

- **Applicant:** Jiangsu Newvalue Medical Products Co., Ltd.
- **Product Code:** QSY
- **Decision Date:** Sep 1, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

Hemostatic Dressing is intended for the external, temporary control of severely bleeding wounds. Hemostatic Dressing is intended for temporary external use to stop bleeding of minor wounds, minor cuts and minor abrasions.

## Device Story

Hemostatic Dressing is a sterile, non-woven fabric dressing composed of hydrophilic fibers (chitosan and chitosan derivative). Applied directly to a wound with pressure, the dressing absorbs blood and exudates, transforming into a gel to seal the wound and promote coagulation. Used in clinical or home settings for temporary bleeding control. Healthcare providers or patients apply the dressing manually; the gel-like transformation provides a physical seal and aids hemostasis. Benefits include rapid bleeding control and exudate management.

## Clinical Evidence

Bench testing only. Testing included absorbency, gelling characteristics, pH, loss on drying, residue on ignition, heavy metals, bacterial endotoxins, packaging integrity, and sterility. Biocompatibility testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, subacute toxicity, and hemolysis. In-vivo hemostatic performance, sterilization validation, and real-time shelf-life stability testing were also conducted.

## Technological Characteristics

Sterile, single-layer, needle-punched non-woven fabric dressing. Material: mixture of chitosan and chitosan derivatives. Mechanism: hydrophilic fiber absorption and gel formation. Sterilization: Gamma radiation (SAL 10^-6). Packaging: Foil pouch. Single-use.

## Regulatory Identification

To temporarily control bleeding and cover external wounds.

## Predicate Devices

- HemCon ChitoGauze™ ([K090026](/device/K090026.md))

## Reference Devices

- SURECELL® Gelling Fiber Wound Dressing ([K173005](/device/K173005.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".

September 1, 2023

Jiangsu NewValue Medical Products Co., Ltd. Lu Wang Vice Manager Building G35.the east of KouTai Road and the north of XinYang Road.CMC. TaiZhou. Jiangsu 225300 China

Re: K213413

Trade/Device Name: Hemostatic Dressing Regulatory Class: Unclassified Product Code: FRO Dated: March 22, 2023 Received: April 7, 2023

### Dear Lu Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# David Krause -S

David Krause, Ph.D. Deputy Director OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for New Value. The logo consists of the letters 'NV' in a stylized font, with the 'N' in a darker shade and the 'V' in a lighter shade. To the right of the letters is the word 'NewValue' in a cursive font.

Section 4

# Indications for Use Statement

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

Device Name Hemostatic Dressing

Indications for Use (Describe)

Indications for Use (Rx): Hemostatic Dressing is intended for the external, temporary control of severely bleeding wounds.

Indications for Use (OTC):

Hemostatic Dressing is intended for temporary external use to stop bleeding of minor wounds, minor cuts and minor

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# continue on a separate page if needed.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review in the sellow of influmation is estimated to breins per response, including the linding the linding the and review the collection of information. Send comments regarding the data needed and complete
of this information collection . including for succined this burden estimate or of this information of including suggestions for reducing this burden, this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for New Value. The logo consists of the letters 'NV' in a stylized font, with the 'N' in dark blue and the 'V' in orange. To the right of the letters is the word 'NewValue' in dark blue.

# 510(k) Summary

#### SUBMITTER: I.

Jiangsu NewValue Medical Products Co., Ltd. Building G35, No.1 Avenue, China Medical City, Taizhou, Jiangsu, 225300, China Contact Person: Lu Wang Title: Vice Manager Tel: 86-015951155167 Email: wanglu@new-value.cn

Summary prepared: 08/29/2023

II. DEVICE

Name of Device: Hemostatic Dressing (K213413) Common Name: Dressing, Wound, Drug Regulatory Class: II Product Code: FRO

III. PREDICATE DEVICE

Primary predicate device: HemCon ChitoGauze™ (K090026)

IV. REFERENCE DEVICE

Reference Device: SURECELL® Gelling Fiber Wound Dressing (K173005)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for New Value. The logo consists of the letters "NV" in blue and orange, followed by the words "New Value" in blue. The "N" in "NV" is blue, and the "V" is orange. The words "New Value" are in a sans-serif font.

## V. DEVICE DESCRIPTION

Hemostatic Dressing is a sterile non-woven fabric dressing composed of hydrophilic fiber (a mixture of chitosan and chitosan derivative). When applied directly over a wound with pressure, this soft and highly absorbent dressing quickly absorbs blood and turns into a gel to seal the wound. It promotes control of wound bleeding and exudates absorption and promotes coagulation.

## VI. INDICATIONS FOR USE

Hemostatic Dressing is a hemostatic dressing for the external, temporary control of severely bleeding wounds.

Hemostatic dressing Over-the-Counter is indicated for temporary external use to stop bleeding of minor wounds, minor cuts and minor abrasions.

## VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Hemostatic dressing is compared with the predicate device (HemCon ChitoGauze™ (K090026)). The results are shown below in the Technological Characteristics Comparison Table:

| Item                                                                                                                                                                             | Subject Device<br>Hemostatic Dressing                                                                                                                                                                        | Predicate Device<br>ChitoGauze™<br>(HemCon Medical<br>Technologies, Inc.)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Reference Device<br>SURECELL® Gelling<br>Fiber Wound Dressing                                                                                                                                                                                                                                                                                                         |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number                                                                                                                                                                    | K213413                                                                                                                                                                                                      | K090026                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K173005                                                                                                                                                                                                                                                                                                                                                               |
| Classification                                                                                                                                                                   | Unclassified                                                                                                                                                                                                 | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Unclassified                                                                                                                                                                                                                                                                                                                                                          |
| Product Code                                                                                                                                                                     | FRO                                                                                                                                                                                                          | FRO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | FRO                                                                                                                                                                                                                                                                                                                                                                   |
| Common name                                                                                                                                                                      | Dressing, Wound, Drug                                                                                                                                                                                        | Dressing, Wound, Drug                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Dressing, Wound,<br>Drug                                                                                                                                                                                                                                                                                                                                              |
| Intended use                                                                                                                                                                     | External hemostatic                                                                                                                                                                                          | External hemostatic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Wound dressing                                                                                                                                                                                                                                                                                                                                                        |
| Indications for use                                                                                                                                                              | Indications for Use (Rx<br>Only): Hemostatic<br>Dressing is a hemostatic<br>dressing for the external,<br>temporary control of<br>severely bleeding wounds.                                                  | Indications for Use (Rx<br>Only): ChitoGauze is a<br>hemostatic dressing for<br>the external, temporary<br>control of severely<br>bleeding wounds.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Indications for Use (Rx<br>Only): Under professional<br>medical care,<br>SURECELL™<br>Gelling Fiber Wound Dressing<br>may<br>be used for the                                                                                                                                                                                                                          |
| <b>Indications for Use (OTC):</b><br>Hemostatic<br>Dressing is intended for<br>temporary external use to<br>stop bleeding of minor<br>wounds, minor cuts and<br>minor abrasions. | <b>Indications for Use (OTC):</b> ChitoGauze is<br>intended for temporary<br>external use to stop<br>bleeding of minor<br>wounds, minor cuts and<br>minor abrasions.                                         | management of leg ulcers<br>(Stage I-IV), pressure ulcers,<br>diabetic ulcers, surgical<br>wounds (e.g. post-operative,<br>donor sites, dermatological),<br>Burns (1st and 2nd degree),<br>surgical or traumatic wounds<br>which have been left to heal<br>by secondary<br>intention. SURECELLTM<br>Gelling Fiber Wound<br>Dressing<br>maintains a moist wound<br>environment.<br>SURECELLTM Gelling Fiber<br>Wound Dressing may also be<br>used for the local<br>management of wounds such a<br>wounds that have been<br>surgically or mechanically<br>debrided, donor sites,<br>and traumatic wounds.<br><b>Indications for Use<br/>(OTC):</b> SURECELLTM<br>Gelling Fiber Wound<br>Dressing<br>Over-the-Counter is<br>indicated for<br>minor Burns, minor abrasions<br>and minor Lacerations, minor<br>superficial cuts.<br>SURECELLTM Gelling Fiber<br>Wound Dressing is<br>intended for the maintenance of<br>a moist wound environment. |                                                                                                                                                                                                                                                                                                                                                                       |
| Duration of Use                                                                                                                                                                  | Temporary                                                                                                                                                                                                    | Temporary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Prolonged exposure                                                                                                                                                                                                                                                                                                                                                    |
| Physical<br>Composition                                                                                                                                                          | Single layer needle<br>punched chitosan<br>nonwoven fabric dressing                                                                                                                                          | polyester/rayon blend<br>non-woven medical gauze<br>that is coated with<br>chitosan                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Single layer needle<br>punched chitosan<br>nonwoven fabric<br>dressing                                                                                                                                                                                                                                                                                                |
| Hemostatic<br>Material                                                                                                                                                           | A mixture of chitosan and<br>chitosan derivatives                                                                                                                                                            | Chitosan                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | A mixture of<br>chitosan and<br>chitosan derivatives                                                                                                                                                                                                                                                                                                                  |
| Mechanism of<br>Action                                                                                                                                                           | Hemostatic Dressing will<br>turn into a gel-like<br>condition to absorb the<br>blood and seal the wound.<br>It promotes control of<br>wound bleeding and<br>exudates absorption and<br>promotes coagulation. | ChitoGauze™ will turn<br>into a gel-like condition to<br>absorb the blood. It<br>promotes control of<br>wound bleeding and<br>exudates absorption and<br>promotes coagulation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | SURECELL™<br>Gelling Fiber<br>Wound Dressing<br>vertically absorbs<br>wound exudate and<br>creates a<br>comfortable clear<br>gel, helps reduce<br>the risk of<br>periwound<br>maceration,<br>maintains a moist<br>environment for<br>optimal wound<br>healing, and<br>removes dead and<br>damaged tissue<br>from wound<br>without damaging<br>newly formed<br>tissue. |
| Chitosan Form                                                                                                                                                                    | Needle punched chitosan<br>nonwoven fabric dressing                                                                                                                                                          | Chitosan coated on gauze                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Needle punched<br>chitosan nonwoven<br>fabric dressing                                                                                                                                                                                                                                                                                                                |
| Single Use                                                                                                                                                                       | Single Use                                                                                                                                                                                                   | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Single Use                                                                                                                                                                                                                                                                                                                                                            |
| Biocompatibility                                                                                                                                                                 | Biocompatible                                                                                                                                                                                                | Biocompatible                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Biocompatible                                                                                                                                                                                                                                                                                                                                                         |
| Sterilization                                                                                                                                                                    | Gamma radiation                                                                                                                                                                                              | Gamma radiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Gamma radiation                                                                                                                                                                                                                                                                                                                                                       |
| Sterility                                                                                                                                                                        | SAL 10-6                                                                                                                                                                                                     | SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | SAL 10-6                                                                                                                                                                                                                                                                                                                                                              |
| Primary<br>Packaging                                                                                                                                                             | Foil pouch                                                                                                                                                                                                   | Foil pouch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Paper-plastic pouch                                                                                                                                                                                                                                                                                                                                                   |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for New Value. The logo consists of the letters 'NV' in a stylized font, with the 'N' in blue and the 'V' in orange. To the right of the letters is the word 'NewValue' in blue, with the letters connected in a cursive style.

# K213413

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for NewValue. The logo consists of the letters "NV" in a stylized font, with the first "N" in blue and the "V" in orange. To the right of the letters, the word "NewValue" is written in blue, in a font that is similar to the "NV" letters.

K213413

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for New Value. The logo consists of the letters 'NV' in a stylized font, with the 'N' in blue and the 'V' in orange. To the right of the letters is the text 'New Value' in a blue, sans-serif font. The text is slightly smaller than the letters 'NV'.

#### NONCLINICAL DATA VIII.

To verify that the Hemostatic dressing is as safe and effective as the predicate device, representative samples of Hemostatic dressing were underwent a series of tests including bench testing (absorbency, gelling characteristics, pH, loss on drying, residue on ignition, heavy metals, bacterial endotoxins, packaging sealing, and sterility), biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, subacute toxicity, hemolysis), in-vivo hemostatic, sterilization validation testing, and real-time shelf life stability testing.

## IX. CONCLUSION

JIANGSU NEWVALUE Medical PRODUCTS CO.,LTD. considers Hemostatic dressing to be equivalent to the predicate devices listed above. This conclusion is based upon the devices' similarities in intended use, design, mechanisms of action, technology and materials. The slight differences between Hemostatic dressing and the predicate devices do not raise any questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K213413](https://fda.innolitics.com/device/K213413)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
