SleepCheckRx is indicated to record a patient's respiratory pattern during sleep for the purpose of pre-screening patients for obstructive sleep apnea (OSA) syndrome. The device is designed for use in home-screening of adults (≥ 22 years of age) with suspected possible sleep breathing disorders. Results can be used to assist the healthcare professional in determining the need for further diagnosis and evaluation. The system is not intended as a substitute for full polysomnography when additional parameters such as sleep stages, limb movements, or EEG activity are required.
Device Story
SleepCheckRx is a smartphone-based software application for home pre-screening of obstructive sleep apnea (OSA). The device uses the smartphone's native microphone to record breathing and snore sounds for at least 6 hours while the patient sleeps. Patients input demographic data (age, sex, height, weight, neck circumference) and complete a STOP-Bang questionnaire. The recorded audio and patient data are processed by locked machine learning algorithms to assess OSA risk. The system provides a binary output: 'Minimal to Mild Risk' (AHI < 15) or 'Moderate to Severe Risk' (AHI ≥ 15). Healthcare professionals (HCPs) manage access via a secure cloud portal, providing codes to patients. Results are shared with both the patient and the HCP to guide decisions regarding further diagnostic evaluation. The device benefits patients by facilitating early identification of OSA risk in a home setting, potentially streamlining the path to clinical diagnosis.
Clinical Evidence
Prospective, multi-center study (n=228 subjects, 220 analyzable). Compared SleepCheckRx algorithm output against simultaneous Type II (in-home) PSG diagnosis (AASM 2017 guidelines). Primary endpoint: agreement between device and PSG. Results at AHI ≥ 15 threshold: sensitivity 89.3%, specificity 77.6%. No adverse events reported.
Technological Characteristics
Software-only application running on consumer smartphones. Uses native smartphone microphone for acoustic sensing. Connectivity via secure cloud portal for HCP management. Locked machine learning algorithm. No direct patient contact. No specific materials or energy sources beyond smartphone hardware.
Indications for Use
Indicated for adults (≥ 22 years) with suspected sleep breathing disorders for pre-screening of obstructive sleep apnea (OSA) syndrome in the home environment. Not intended as a substitute for full polysomnography.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 5, 2022
ResApp Health Neroli Anderson VP Clinical, Quality And Regulatory Level 12, 100 Creek Street Brisbane, Queensland 4000 Australia
Re: K213360
Trade/Device Name: SleepCheckRx Regulation Number: 21 CFR 868.2375 Regulation Name: Ventilatory Effort Recorder Regulatory Class: Class II Product Code: MNR Dated: June 9, 2022 Received: June 10, 2022
Dear Neroli Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213360
Device Name SleepCheckRx
Indications for Use (Describe)
Sleep CheckRx is indicated to record a patient's respiratory pattern during sleep for the purpose of pre-screening patients for obstructive sleep apnea (OSA) syndrome.
The device is designed for use in home-screening of adults (≥ 22 years of age) with suspected possible sleep breathing disorders. Results can be used to assist the healthcare professional in determining the need for further diagnosis and evaluation.
The system is not intended as a substitute for full polysomnography when additional parameters such as sleep stages, limb movements, or EEG activity are required.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K213360 510(k) Summary (As required by section 807.92(c))
# SLEEPCHECKRx
| I. GENERAL INFORMATION | |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | ResApp Health<br>Level 12, 100 Creek Street<br>Brisbane QLD 4000<br>Australia |
| Contact Person: | Neroli Anderson<br>VP Clinical, Quality and Regulatory<br>Level 12, 100 Creek Street<br>Brisbane QLD 4000<br>Australia<br>Phone: +61 416 220 239<br>Email: neroli@resapphealth.com.au |
| Date Prepared: | July 02, 2022 |
| II. DEVICE DETAILS | |
| Trade Name: | SleepCheckRx |
| Common or Usual Name: | Breathing Frequency Monitor |
| Classification Name: | Device: Ventilatory Effort Recorder |
| Regulation: | 21 CFR 868.2375 |
| Regulatory Class: | II |
| Product Code: | MNR |
| Review Panel: | Anesthesiology |
| III. PREDICATE DEVICE | |
| Predicate Device(s): | DROWZLE (K173974) |
{4}------------------------------------------------
## IV. DEVICE DESCRIPTION
SleepCheckRx is a software application used by healthcare providers to pre-screen adults at home for the risk of Obstructive Sleep Apnea (OSA). The application is accessed via a compatible smartphone and uses the device's inbuilt microphone to record breathing and snore sounds while the user is asleep, over a minimum duration of 6 hours.
This audio recording, in conjunction with basic information about the user, is analyzed by locked machine learnt software algorithms to provide a risk assessment of the presence of OSA. Prior to use, the patient installs SleepCheckRx on a compatible smartphone and activates the application by entering a code issued to their Healthcare Professional (HCP) through a secure cloud server portal.
The patient enters basic information including date of birth, biological sex, height, weight, and neck circumference. The patient is also required to answer a STOP-Bang questionnaire. The phone is placed on a nightstand as directed, and when the user is ready to go to sleep, they initiate the software that begins a recording.
SleepCheckRx then captures breathing and snore sounds continuously throughout the night until the user stops the recording once they have awoken in the morning (a minimum of 6 hours of recording is required).
In the event of a successful recording, SleepCheckRx provides the user with a binary outcome of either 'Minimal to Mild Risk of OSA' (AHI<15) or 'Moderate to Severe Risk of OSA' (AHI ≥15).
SleepCheckRx requires a healthcare professional (HCP) to access a secure cloud server portal to provide the patient with an access code to use the app downloaded from the Apple App Store.
The results of the patient's recording and the STOP-Bang questionnaire are shared with the patient and the HCP, for the HCP to then determine if further investigation and evaluation is required.
## V. INDICATIONS FOR USE
SleepCheckRx is indicated to record a patient's respiratory pattern during sleep for the purpose of pre-screening patients for obstructive sleep apnea (OSA) syndrome.
The device is designed for use in home-screening of adults (≥ 22 years of age) with suspected possible sleep breathing disorders. Results can be used to assist the healthcare professional in determining the need for further diagnosis and evaluation.
The system is not intended as a substitute for full polysomnography when additional parameters such as sleep stages, limb movements, or EEG activity are required.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
A summary comparison of technological characteristics between the SleepCheckRx and the predicate device is provided below.
{5}------------------------------------------------
# ResApp Health. ResApp Health.
K213360 SleepCheckRx 510(k) Premarket Notification (Traditional)
| Device Name | DROWZLE | SleepCheckRx | Comparison |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k number | K173974 | K213360 | N/A |
| Regulation<br>Number | 21 CFR§ 868.2375 | 21 CFR§ 868.2375 | Same regulation<br>number. |
| Regulation<br>Name | Breathing Frequency<br>Monitor | Breathing Frequency<br>Monitor | Same regulation name |
| Product Code | MNR | MNR | Same product code. |
| Intended Use | Home-use for screening<br>patients with possible<br>sleep disorders. | Home-use for screening<br>patients with possible<br>sleep disorders. | Equivalent. Both<br>devices intended to<br>home screen patients<br>with possible sleep<br>disorders. |
| Indications for<br>Use | DROWZLE is indicated<br>to record a patient's<br>respiratory pattern during<br>sleep for the purpose of<br>prescreening patients for<br>obstructive sleep apnea<br>(OSA) syndrome.<br>The device is designed<br>for use in home-<br>screening of adults with<br>suspected possible sleep<br>breathing disorders.<br>Results are used to<br>assist the healthcare<br>professional in<br>determining the need for<br>further diagnosis and<br>evaluation.<br>The system is not<br>intended as a substitute<br>for full polysomnography<br>when additional<br>parameters such as<br>sleep stages, limb<br>movements, or EEG<br>activity are required. | SleepCheckRx is<br>indicated to record a<br>patient's respiratory<br>pattern during sleep for<br>the purpose of pre-<br>screening patients for<br>obstructive sleep apnea<br>(OSA) syndrome.<br>The device is designed<br>for use in home-<br>screening of adults (≥ 22<br>years of age) with<br>suspected possible sleep<br>breathing disorders.<br>Results can be used to<br>assist the healthcare<br>professional in<br>determining the need for<br>further diagnosis and<br>evaluation.<br>The system is not<br>intended as a substitute<br>for full polysomnography<br>when additional<br>parameters such as<br>sleep stages, limb<br>movements, or EEG<br>activity are required. | Equivalent indications<br>for use. Both devices<br>intended to record a<br>patient's respiratory<br>pattern during sleep<br>for the purpose of pre-<br>screening patients for<br>obstructive sleep<br>apnea (OSA)<br>syndrome. |
| Target<br>Population | Adults | Adults | Equivalent. Both<br>devices intended for<br>use on adults. |
| Intended Use<br>Environment | Home Environment | Home Environment | Equivalent. Both<br>devices intended for<br>the home environment |
| Device Name | DROWZLE | SleepCheckRx | Comparison |
| Type of Use | Prescription Use (Part 21<br>CFR 801 Subpart D) | Prescription Use (Part 21<br>CFR 801 Subpart D) | Both devices are<br>prescription only use. |
| Method of<br>Measurement | Acoustic analysis of<br>breathing and snoring<br>sounds. | Acoustic analysis of<br>breathing and snoring<br>sounds. | Equivalent. Both<br>devices analyze<br>breathing and snoring<br>sounds. |
| Mode of Action | Analyses breathing<br>sounds to identify<br>respiratory events<br>indicative of OSA or other<br>disorders. | Analyses breathing<br>sounds to identify<br>respiratory events<br>indicative of OSA or other<br>disorders. | Equivalent. Both<br>devices identify<br>respiratory events<br>indicative of OSA or<br>other disorders. |
| Sensor<br>Placement Site | Smartphone placed within<br>24 inches of pillow. | Smartphone placed next<br>to the bedside (on a<br>nightstand). | Equivalent. Both<br>devices utilize smart<br>phones placed in close<br>proximity while<br>sleeping. |
| Sensor<br>Elements | Microphone(s) native to<br>smartphone. | Microphone(s) native to<br>smartphone. | Equivalent. Both<br>devices utilize<br>microphone(s) native<br>to the smartphone. |
| Patient Contact | Software only. No direct<br>patient contact. | Software only. No direct<br>patient contact. | Equivalent. Software<br>only. Neither device<br>patient contacting. |
| Portability | Yes | Yes | Equivalent. Both<br>devices downloaded<br>onto a smartphone. |
| Measured<br>Variable | Oral and nasal breath<br>sounds. | Oral and nasal breath<br>sounds. | Equivalent. Both<br>devices detect oral and<br>nasal breath sounds. |
| Breathing<br>Events | Breath sound gaps >10<br>seconds. | Breathing pauses of<br>duration ≥ 6s but < 60<br>seconds. | Equivalent. Both<br>devices detect<br>breathing events<br>(however the<br>algorithms have been<br>trained to detect events<br>of different durations). |
| Sensor<br>Attachment | N/A | N/A | Equivalent. Neither<br>device has a sensor<br>attachment. |
| Display Type | Smartphone display | Smartphone display | Equivalent. Both<br>devices utilize the<br>smartphone display. |
| Signal Loss<br>Indicator | N/A | N/A | Equivalent. Neither<br>device has a signal<br>loss indicator. |
| Reported<br>Metrics | Risk of OSA based on the<br>following AHI estimates: | Risk of OSA based on the<br>following AHI estimates: | Equivalent. Both<br>devices report the risk |
| Device Name | DROWZLE | SleepCheckRx | Comparison |
| | Minimal to Mild Risk of OSA (AHI<15) Moderate to Severe Risk of OSA (AHI ≥15). Risk classification based on standard questionnaires:<br>a. STOP-Bang<br>b. Epworth Sleepiness Scale<br>c. Calculated Resonea Index | Minimal to Mild Risk of OSA (AHI<15) Moderate to Severe Risk of OSA (AHI ≥15). STOP-Bang questionnaire is also reported. | of OSA based on a<br>binary output, i.e.,<br>either minimal to mild risk of OSA (AHI<15),<br>or moderate to severe risk of OSA (AHI ≥15).<br>Both devices report the STOP-Bang score.<br>SleepCheckRx does not report the Epworth<br>scale, nor does it display a proprietary<br>index score. |
| Reporting of<br>Results to HCP | Results are available for<br>review by the clinician via<br>email. | Results are available for<br>review by the clinician via<br>the portal. | Equivalent. Both<br>devices may be<br>reviewed via a clinician<br>portal. |
| Sleep Night Use | Can be used multiple<br>nights. | Can be used multiple<br>nights. | Equivalent. Both<br>devices may be used<br>over multiple nights. |
| Maximum Run-<br>time | No maximum run-time. | No maximum run-time. | Equivalent. Neither<br>device specifies a<br>maximum run-time. |
| Minimum Time<br>Required for<br>result | 2 hours | 6 hours | SleepCheckRx is<br>required to be used for<br>a longer period of time<br>(6 hours) to ensure<br>more accurate results. |
| Controller | Smartphone<br>microprocessor. | Smartphone<br>microprocessor. | Equivalent. Both<br>devices utilize a<br>smartphone. |
| Physical<br>Characteristics | Software runs on user's<br>smartphone. | Software runs on user's<br>smartphone. | Equivalent. Both<br>devices utilize a<br>smartphone. |
| Power | Smartphone plugged into<br>wall outlet with built-in<br>battery back-up. | Smartphone plugged into<br>wall outlet with built-in<br>battery back-up. | Equivalent. Both<br>devices utilize a<br>smartphone. |
{6}------------------------------------------------
# ResApp Health. ResApp Health.
K213360 SleepCheckRx 510(k) Premarket Notification (Traditional)
{7}------------------------------------------------
# ResApp Health. ResApp Health.
K213360 SleepCheckRx 510(k) Premarket Notification (Traditional)
{8}------------------------------------------------
### VII. NON-CLINICAL PERFORMANCE DATA
SleepCheckRx performance data included various verification and validation activities to demonstrate compliance with the applicable FDA guidance documents, including:
- · Software verification and validation testing was performed in accordance with the applicable requirements outlined in the FDA quidance for the Content of Premarket Submissions for Software Contained in Medical Devices (2005).
- · Cybersecurity verification and validation testing in accordance with the applicable requirements outlined in the FDA guidance Content of Premarket Submissions for Management of Cybersecurity in Medical Devices - Guidance for Industry and Food and Drug Administration Staff (2014).
- · Device Compatibility testing was performed to ensure that compatible smartphones support equivalent performance.
- An evaluation of performance of SleepCheckRx under conditions part of intended use environment (noise, distances, direction, etc.) to verify device effectiveness.
- · Human Factors Engineering (HFE) / Usability evaluation was performed in accordance with the applicable requirements of FDA Guidance. Applying Human Factors and Usability Engineering to Medical Devices. Guidance for Industry and Food and Drug Administration Staff. February 3, 2016.
### VIII. CLINICAL DATA
A prospective, multi-center, single-staged study on the SleepCheckRx algorithm was conducted on 228 adult analyzable subjects.
The clinical study used a binomial endpoint comparing the presence and severity of OSA, by using sleep sounds captured and analyzed by the SleepCheckRx algorithm and comparing them to a simultaneous PSG diagnosis (gold standard). PSG diagnosis was established by independent scorers, in accordance with the Type II (in-home) American Academy of Sleep Medicine (AASM) 2017 Guidelines.
Sleep sound recordings were captured by the SleepCheckRx algorithm throughout the night (minimum of 6 hours) according to the overnight sleep study protocol. Sleep sounds were recorded by a compatible smartphone (loaded with the SleepCheckRx sound recording algorithm) placed on the nightstand, simultaneously with PSG. Each sleep study was scored by a qualified independent sleep scorer.
The mean age of all subjects was 50 years, with a higher percentage of males in the study (63%). The mean BMI was 29.8 and the mean neck circumference of 39.4.
Results of the study (n=220) demonstrated an agreement between SleepCheckRx and clinical diagnosis (using 2 independent PSG scorers) of 109 from 122 subjects (sensitivity 89.3%) with moderate to severe risk of OSA (AHI ≥15), and an agreement of 76 from 98 subjects (specificity 77.6%) with minimal to mild risk of OSA (AHI<15).
There were no adverse events or adverse device effects reported during the study.
### IX. Substantial Equivalence discussion
SleepCheckRx has the same intended use as the predicate DROWZLE. Both software platforms are intended to record a patient's respiratory pattern during sleep for the purpose of prescreening patients for obstructive sleep apnea (OSA) syndrome.
{9}------------------------------------------------
Both devices are designed for use in home-screening adults with suspected possible sleep breathing disorders. The results of both devices are used to assist the healthcare professional in determining the need for further diagnosis and evaluation.
Both devices report the risk of OSA based on a binary output, i.e., either minimal to mild risk of OSA (AHI<15), or moderate to severe risk of OSA (AHI ≥15).
SleepCheckRx is supported by clinical data demonstrating a sensitivity and specificity at the AHI ≥ 15 threshold of 89.3% and 77.6% respectively (n=220) when compared to clinical diagnosis. These results are comparable to the DROWZLE device which reported a sensitivity of 93.7% and specificity of 63% (AHI>15).
#### CONCLUSIONS X.
Based on the information provided in this submission, the SleepCheckRx is substantially equivalent to the predicate DROWZLE device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.