K213332 · Paradigm Medical Corporation · ONG · Jun 3, 2022 · General, Plastic Surgery
Device Facts
Record ID
K213332
Device Name
Multifrax Laser System
Applicant
Paradigm Medical Corporation
Product Code
ONG · General, Plastic Surgery
Decision Date
Jun 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1550 nm: The Multifrax 1550 nm laser is indicated for use in dermatological procedures requiring coagulation of soft tissue, as well as for skin resurfacing procedures. It is also indicated for treatment of dyschromia and cutaneous lesions, such as, but not limited to lentigos (age spots), solar lentigos (sun spots), actinic keratosis, and melasma, and for treatment of periorbital wrinkles, acne scars and surgical scars. 1927 nm: The Multifrax 1927 nm laser is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephiledes (freckles).
Device Story
Multifrax Laser System is a portable, non-ablative, fractional diode laser; emits 1550 nm and 1927 nm wavelengths. Device delivers light pulses in rows via handpiece to create micro-thermal zones (MTZ) in skin. Features include adjustable treatment width, adjustable pulse spacing, and integrated movement/skin contact tracking sensors. Operated by healthcare professionals in clinical settings; powered by rechargeable batteries with belt clip. System controlled via color touchscreen. Output allows for precise coagulation and resurfacing; benefits patients through treatment of pigmented lesions, wrinkles, and scars.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), laser safety (IEC 60601-2-22, IEC 60825-1), software validation (IEC 62304), and biocompatibility (ISO 10993-5, -10, -23).
Indicated for dermatological procedures requiring soft tissue coagulation, skin resurfacing, and treatment of dyschromia, cutaneous lesions (lentigos, solar lentigos, actinic keratosis, melasma, ephiledes), periorbital wrinkles, acne scars, and surgical scars. For use by trained healthcare professionals.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 3, 2022
Paradigm Medical Corporation % Maureen O'connell President O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, Massachusetts 02180
Re: K213332
Trade/Device Name: Multifrax Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONG Dated: May 6, 2022 Received: May 9, 2022
Dear Maureen O'connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213332
Device Name Multifrax Laser System
#### Indications for Use (Describe)
1550 nm: The Multifrax 1550 nm laser is indicated for use in dermatological procedures requiring coagulation of soft tissue, as well as for skin resurfacing procedures. It is also indicated for treatment of dyschromia and cutaneous lesions, such as, but not limited to lentigos (age spots), solar lentigos (sun spots), and melasma, and for treatment of periorbital wrinkles, acne scars and surgical scars.
1927 nm: The Multifrax 1927 mm laser is indicated for use in dermatological procedures requiring the coagulation of soft tissue, treatment of actinic keratosis, and treatment of pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephiledes (freckles).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
### Paradigm Medical Corporation Multifrax Laser System K213332
### 510(k) Owner
Paradigm Medical 7371 Los Brazos San Diego, CA 92127
### Submission Correspondent
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham. MA 02180 Phone: 978-207-1245
Date Prepared: June 3, 2022
### Trade Name of Device
Multifrax Laser System
### Common or Usual Name
Powered laser surgical instrument with microbeam/fractional output
#### Classification Name
Laser surgical instrument for use in general and plastic surgery and dermatology; 21 C.F.R. \$878.4810 Class II Product Code: ONG
#### Predicate Device
Solta Medical, Inc. Fraxel Dual 1550/1927 Laser System cleared in K130193
#### Device Description
The Multifrax Laser System provides non-ablative fractional treatment using the 1550 nm and 1927 nm wavelengths. The mode of action is delivering a series of light pulses in a row as the handpiece tip is moved across the skin which leaves a pattern of microdots (Micro Thermal Zone (MTZ)) of irradiated skin.
Multifrax Laser System is a dual-wavelength diode laser system comprising two laser sources emitting at 1550 and 1927 nm and generating fine pulses of energy.
- Multifrax is a non-ablative and non-invasive fractional laser device. ●
{4}------------------------------------------------
- Multifrax is a light portable laser combining a small laser unit and 2 long lasting ● rechargeable batteries equipped with a belt clip.
- Multifrax offers optimized technical characteristics: adjustable treatment width, adjustable ● distance between laser pulses, distal tip with integrated high-performance movement and skin contact tracking sensors, and color touch screen display.
### Indications for Use
1550 nm: The Multifrax 1550 nm laser is indicated for use in dermatological procedures requiring coagulation of soft tissue, as well as for skin resurfacing procedures. It is also indicated for treatment of dyschromia and cutaneous lesions, such as, but not limited to lentigos (age spots), solar lentigos (sun spots), actinic keratosis, and melasma, and for treatment of periorbital wrinkles, acne scars and surgical scars.
1927 nm: The Multifrax 1927 nm laser is indicated for use in dermatological procedures requiring the coagulation of soft tissue. treatment of actinic keratosis, and treatment of pigmented lesions such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and ephiledes (freckles).
### Substantial Equivalence
Paradigm Medical believes that the Multifrax described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to a legally marketed predicate device that is a Class II medical device which is the Fraxel Dual 1550/1927 nm Laser System cleared in K130193. The table below compares the properties of the two devices.
| Characteristic | Multifrax | Fraxel Dual 1550/1927 nm<br>Laser System |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Paradigm Medical | Solta Medical, Inc. |
| 510(k)<br>Number | - | K130193 |
| Product Code | ONG | GEX |
| Regulation | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Indications<br>for Use | 1550 nm: The Multifrax 1550 nm laser<br>is indicated for use in dermatological<br>procedures requiring coagulation of<br>soft tissue, as well as for skin<br>resurfacing procedures. It is also<br>indicated for treatment of dyschromia<br>and cutaneous lesions, such as, but not<br>limited to lentigos (age spots), solar<br>lentigos (sun spots), actinic keratosis,<br>and melasma, and for treatment of<br>periorbital wrinkles, acne scars and<br>surgical scars.<br><br>1927 nm: The Multifrax 1927 nm laser<br>is indicated for use in dermatological<br>procedures requiring the coagulation of<br>soft tissue, treatment of actinic<br>keratosis, and treatment of pigmented<br>lesions such as, but not limited to | 1550: The Fraxel 1550 nm laser is<br>indicated for use in dermatological<br>procedures requiring coagulation of soft<br>tissue, as well as for skin resurfacing<br>procedures. It is also indicated for<br>treatment of dyschromia and cutaneous<br>lesions, such as, but not limited to<br>lentigos (age spots), solar lentigos (sun<br>spots), actinic keratosis, and melasma,<br>and for treatment of periorbital wrinkles,<br>acne scars and surgical scars.<br><br>1927 nm: The Fraxel 1927 nm laser is<br>indicated for use in dermatological<br>procedures requiring the coagulation of<br>soft tissue, treatment of actinic keratosis,<br>and treatment of pigmented lesions such<br>as, but not limited to lentigos (age spots), |
{5}------------------------------------------------
| | lentigos (age spots), solar lentigos (sun spots) and ephiledes (freckles). | solar lentigos (sun spots) and ephiledes (freckles). |
|-------------------------|----------------------------------------------------------------------------|------------------------------------------------------|
| Clearance<br>Type | Prescription | Prescription |
| User | Healthcare Professional | Healthcare Professional |
| Operating<br>Principles | Non-ablative treatment | Non-ablative treatment |
| Laser Source | Laser diode | Erbium fiber<br>Thulium fiber |
| Delivery<br>System | Handpiece | Fiber delivery<br>to Handpiece |
| Laser<br>Wavelength | 1550 nm<br>1927 nm | 1550 nm<br>1927 nm |
| Pulse Energy | 4 to 70 mJ (1550 nm)<br>1 to 20 mJ (1927 nm) | 4 to 70 mJ (1550 nm)<br>5 to 20 mJ (1927 nm) |
| Tissue<br>Contact | Disposable contact tip | Disposable contact tips |
| Power source | 2 rechargeable batteries | Power supply |
The intended use of the Multifrax as well as the indications for use are identical. Both are prescription devices for use by trained healthcare professionals for non-ablative dermatological laser treatment. The devices have the same technological characteristics. Both lasers are 1550 nm and 1927 nm diode lasers. Both are delivered to the patient via a handpiece with a contact tip. The upper pulse energy of both devices is the same for both wavelengths. The scaming shape is a single line for both devices. Neither device has an aiming beam. Both devices are software controlled and operated via a touchscreen.
# Performance Data
Non-clinical testing was performed to verify that the proposed device met all design specifications and is substantially equivalent to the predicate device. Testing was performed to the following standards and the Multifrax Laser System was found to meet the requirements:
- Electrical safety per ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, ● C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- . Electromagnetic compatibility per IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests
- Usability per IEC 60601-1-6 2010, AMD 1: 2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Usability
- Laser equipment safety per IEC 60601-2-22 Medical electrical equipment Part 2: . Particular requirements for the basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.
- Laser safety per IEC 60825-1:2014 Safety of laser products-Part 1: Equipment ● classification and requirements
{6}------------------------------------------------
- Software validation and verification per IEC 62304:2006 Medical device software-. Software life cycle processes
- Biocompatibility per ISO 10993-23:2021 Biological evaluation of medical devices-Part . 23: Tests for irritation; ISO 10993-10:2021 Biological evaluation of medical devices-Part 10 :Tests for irritation and skin sensitization; and ISO 10993-5 :2009 Third edition Biological evaluation of medical devices-Part 5 : Tests for in vitro cytotoxicity
## Conclusion
Based on the substantial equivalence discussion and the performance testing, the Multifrax Laser System is substantially equivalent to the Fraxel Dual 1550/1927 nm Laser System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.