K213282 · Energy Resources International Co., Ltd. Hsinchu Branch · EHD · Apr 4, 2022 · Dental
Device Facts
Record ID
K213282
Device Name
ERI Handheld Dental X-ray System (Model AG100)
Applicant
Energy Resources International Co., Ltd. Hsinchu Branch
Product Code
EHD · Dental
Decision Date
Apr 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Attributes
Pediatric
Indications for Use
ERI Handheld Dental X-Ray System (Model AG100) is intended for the purpose of using it for both adult and pediatric patients by qualified operators (e.g. dentist and/ or radiographer, etc.) regulated by competent authority of each country for producing diagnostic dental X-ray images.
Device Story
Handheld dental X-ray system; produces diagnostic images for dental procedures (caries, root canal, implants). Operates via fixed tube current (0.6 mA) and voltage (60 kVp). Used by dentists or radiographers in clinical settings. Compatible with CR, DR, or traditional dental film. Operator triggers exposure via handset button. Device includes scatter shield for radiation protection. Benefits include portability for intraoral imaging. System is battery-powered (14.4 V DC Li-polymer).
Clinical Evidence
Bench testing only. No clinical data conducted. Verification and validation included electrical performance evaluations, usability testing, and adherence to international standards (IEC 60601 series, ISO 10993, 21 CFR 1020).
Technological Characteristics
Extraoral X-ray source; 60 kVp/0.6 mA fixed output. 14.4 V DC Li-polymer battery. 20 cm source-to-skin distance; 6 cm cone diameter. 153mm Pb-filled acrylic scatter shield. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2 (EMC), IEC 60601-1-3 (Radiation), IEC 60601-2-65 (Dental X-ray). Biocompatibility per ISO 10993. Software classified as Moderate level of concern.
Indications for Use
Indicated for adult and pediatric patients for producing diagnostic dental X-ray images by qualified operators (dentists/radiographers). Used for preoperative, intraoperative, and postoperative evaluation of dental caries, root canal treatment, and dental implant positioning.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{1}------------------------------------------------
April 4, 2022
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
Energy Resources International Co., Ltd. Hsinchu Branch % Ms. Judy Chiang QA&RA Specialist 2F-2, No. 6-1, Sec. 2, Zhengyi Rd Zhubei City. Hsinchu County 30261 TAIWAN
### Re: K213282
Trade/Device Name: ERI Handheld Dental X-ray System (Model AG100) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: EHD, MUH Dated: September 27, 2021 Received: October 1, 2021
#### Dear Ms. Chiang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K213282
#### Device Name
ERI Handheld Dental X-Ray System (Model AG100)
Indications for Use (Describe)
mandheld Dental X-Ray System (Model AG100) is intended for the purpose of using it for both adult and pediatric patients by qualified operators (e.g. dentist and/ or radiographer, etc.) regulated by competent authority of each country for producing diagnostic dental X-ray images.
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|--|--|
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and comblete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | |
| Food and Drug Administration<br>Office of Chief Information Officer | Department of Health and Human Services<br>Paperwork Reduction Act (PRA) Staff | | |
| PRAStaff(@fda.hhs.gov | | | |
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{4}------------------------------------------------
# ENERGY RESOURCES INTERNATIONAL CO., LTD. 510(k) Submission AG100
# Section 5
510(k) Summary
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a document with the logo of Energy Resources International Co., Ltd. Hsinchu Branch. The document includes a table with the following information: Product: AG100, 510(k) Summary, page: 5-2, and Version: 2.0. The document appears to be a summary related to the product AG100.
The assigned 510(K) number: K213282 Date Prepared: 09/27/2021
#### Submitter Information I.
| Submitter: |
|--------------------------------------------------------------------------------------------------|
| ENERGY RESOURCES INTERNATIONAL CO., LTD. Hsinchu Branch |
| Address: 2F-2, No.6-1, Sec. 2, Zhengyi Rd., Zhubei City, Hsinchu County 30261,<br>Taiwan(R.O.C.) |
| Phone Number: +886-3-6585156 |
| Fax Number: +886-3-6585802 |
| Contact Person: Judy Chiang |
| Email: judy.chiang@eriray.com |
| Phone Number: +886-3-6585156 ext.112 |
| Fax Number: +886-3-6585802 |
#### Subject Device II.
Trade Name: ERI Handheld Dental X-Ray System (Model AG100)
Regulation Name and Classification:
| No. | Product<br>Code | Regulation Name | Regulatory<br>Class | Regulation<br>Number | Panel |
|-----|-----------------|-------------------------------|---------------------|----------------------|-----------|
| 1 | EHD | Extraoral Source X-Ray System | II | 872.1800 | Radiology |
| 2 | MUH | Extraoral Source X-Ray System | II | 872.1800 | Radiology |
#### III. Predicate Device
| 510(K) Number | K173319 |
|-------------------|----------------------------------------|
| Manufacturer | Aribex |
| Device Name | KaVo NOMAD Pro 2 Handheld X-Ray System |
| Regulation Name | Extraoral Source X-Ray System |
| Regulation Number | 872.1800 |
| Product Code | EHD |
| Regulatory Class | II |
### IV. Device Description
It can be used with computed radiography (CR) detection film, digital radiography (DR) detection film or traditional film (dental film) to take intraoral X-rays and provide the information needed by the dentist for diagnosis.
As a preoperative, intraoperative, and postoperative evaluation for identifying dental caries or auxiliary root canal treatment, as well as confirming the position of the implant during dental implant surgery. It can take pictures of the front and back teeth, as well as the bite of
{6}------------------------------------------------
| 能 資 國 際 股 份 有 限 公 司 新 竹 分 公 司<br>Energy Resources International Co., Ltd. Hsinchu Branch | | | | |
|------------------------------------------------------------------------------------------|-------|----------------|---------|-----|
| Product | AG100 | 510(k) Summary | page | 5-3 |
| | | | Version | 2.0 |
the back teeth. Applicable objects include adults and children.
Due to the medical software included in radiation-emitting device. ERI handheld system is of Moderate level of concern.
The AG100 device, as well as, the predicate device (Trade Name: KaVo NOMAD Pro 2 Handheld X-Ray System manufactured by Aribex (K173319)) utilize a fixed tube current and voltage (KVp) to achieve their intended use.
#### v. Indications for Use
ERI Handheld Dental X-Ray System (Model AG100) is intended for the purpose of using it for both adult and pediatric patients by qualified operators (e.g. dentist and/ or radiographer, etc.) regulated by competent authority of each country for producing diagnostic dental X-ray images.
#### VI. Comparison of Technology Characteristics with the Predicate Device
Energy Resources International claims that the AG100 device is substantially equivalent to the KaVo NOMAD Pro 2 Handheld X-Ray System cleared by the FDA in K173319. Energy Resources International claims substantial equivalence because the AG100 has an equivalent intended use, operating principles, and operational specifications as compared to the predicate device.
The specific details regarding the similarities and differences between the AG100 and KaVo NOMAD Pro 2 Handheld X-Ray System have been identified and explained in the Comparison Table, see the following table 3. A summary of these similarities and differences is included below. These differences do not present any new issues related to safety and effectiveness.
| Device name | Kavo NOMAD Pro 2 Handheld<br>X-ray system<br>(Predicate Device: K173319) | ERI Handheld Dental X-Ray<br>System (Model AG100) |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification,<br>regulation<br>number, product<br>code | Extraoral Source X-ray System,<br>21 CFR 872.1800,<br>EHD | Extraoral Source X-ray System,<br>21 CFR 872.1800,<br>EHD/MUH |
| Intended Use | The KaVo NOMAD Pro 2<br>Handheld X-ray System is<br>indicated for use only by a trained<br>and qualified dentist or dental<br>technician for both adult and<br>pediatric subjects as an extraoral<br>diagnostic dental X-ray source to<br>produce X-ray images using<br>intraoral image receptors. | ERI Handheld Dental X-Ray<br>System (Model AG100) is intended<br>for the purpose of using it for both<br>adult and pediatric patients by<br>qualified operators (e.g. dentist and/<br>or radiographer, etc.) regulated by<br>competent authority of each country<br>for producing diagnostic dental X-<br>ray images. |
{7}------------------------------------------------
| Product | AG100 | 510(k) Summary | page 5-4<br>Version 2.0 |
|----------------------------------|------------------------------------------------------|----------------|----------------------------------------------------------------------------------------------------------------------------------|
| Source to skin distance | 20 cm | | 20 cm |
| Cone diameter | 6 cm | | 6 cm |
| Backscatter radiation protection | 6.75" dia. Pb-filled acrylic plastic scatter shield | | 153mm(6.02") dia. Pb-filled acrylic plastic scatter shield |
| Exposure switch | Trigger located on handset | | Trigger located on front top |
| | Electrical | | |
| Energy Source | Rechargeable 21.6 V DC Li-ion battery core pack | | Rechargeable 14.4 V DC Li-polymer battery core pack |
| Capacity | 1.7 A-hr | | 2.4 A-hr (34.56Wh) |
| Casing | Hard shell case | | Hard shell case |
| Recharge capability | 70% remaining capacity after 300 cycles | | After 300 cycles, more than 70% |
| Exposure Time | 0.02 – 1.0 seconds in 0.01 increments | | 0.1 – 2.0 seconds in 0.01 increments |
| mA | 2.5 mA fixed | | 0.6 mA fixed |
| kVp | 60 kVp fixed | | 60 kVp fixed |
| Duty Cycle | 1:60 | | maximum 60 sec after previous shot |
| Electrical Safety Standards | AAMI ES60601-1:2005/(R)2012 And A1:2012 | | ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) |
| EMI Standards | IEC60601-1-2 Ed. 4 | | IEC60601-1-2:2014 Ed. 4 |
| | X-ray Performance | | |
| Performance Standard | 21 CFR 1020.30, 1020.31; IEC60601-1-3; IEC60601-2-65 | | 21 CFR 1020.30, 1020.31; IEC 60601-1-3 Edition 2.1; IEC 60601-2-65 Edition 1.1 CONSOLIDATED VERSION; IEC 61223-3-4 First edition |
{8}------------------------------------------------
| Image: Energy Resources International Co., Ltd. Hsinchu Branch logo | | | | |
|---------------------------------------------------------------------|-------|----------------|---------|-----|
| Product | AG100 | 510(k) Summary | page | 5-5 |
| | | | Version | 2.0 |
Hence, the subject device is similar to the predicate device in terms of the indications for use and technological application.
# VII. Non-clinical Test Data
Testing was performed in accordance with the following international standards:
- IEC 62304 Edition 1.1 [2015-06] Medical device software - Software life cycle processes
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2 Edition 4.0 [2014-02] Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
- IEC 60601-1-3 Edition 2.1 [2013-04] Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- IEC 60601-2-65 Edition 1.1 [2017-05] Medical electrical equipment - Part 2-65: Particular requirements for the basic safety and essential performance of dental intra-oral-X-ray equipment
- IEC 61223-3-4 First edition [2000-03] Evaluation and routine testing in medical imaging departments - Part 3-4: Acceptance tests - Imaging performance of dental X-ray
- ISO14971 Third Edition [2019-12] Medical devices - Applications of risk management to medical devices
- ISO10993-2 Second edition [2006-07-15] Biological Evaluation of medical devices - Part 2: Animal welfare requirements
- ISO10993-5 Third edition [2009-06-01] Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO10993-10 Third Edition [2010-08-01] Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- ISO10993-12 Fourth edition [2012-07-01] Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
- 21 CFR 1020.30: Diagnostic x-ray system and their major components
- 21 CFR 1020.31: Radiographic Equipment
### VIII. Clinical Performance Data
The function and performance of the AG100 have been evaluated through non-clinical design verification and validation tests. Testing included electrical performance evaluations and usability test. The results of the AG100 performance evaluations demonstrate that the device design is well suited for its intended use. Clinical testing has not been conducted on this product, and successful bench testing results should be sufficient in demonstrating substantial equivalence for the AG100 handheld system.
{9}------------------------------------------------
| 117 | 能 資 國 際 股 份有限公司 新 竹 分 公 司<br>Energy Resources International Co., Ltd. Hsinchu Branch | | |
|-----------------|--------------------------------------------------------------------------------------|---------|-----|
| Product AG100 | | page | 5-6 |
| | 510(k) Summary | Version | 2.0 |
#### IX. Conclusions
The Energy Resources International AG100 has the same intended use and same basic technology as the predicate device, thus is able to achieve the same effectiveness and safety as the predicate device. The AG100 contains in some combination similar features and design as the predicates. Other differences include device design such as exposure time, size and user interface. The subject device is substantially equivalent to the predicate device with its intended use, mechanical and electrical performance as described.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.