Dynamis Pro Family

K213267 · Fotona D.O.O. · GEX · Sep 8, 2022 · General, Plastic Surgery

Device Facts

Record IDK213267
Device NameDynamis Pro Family
ApplicantFotona D.O.O.
Product CodeGEX · General, Plastic Surgery
Decision DateSep 8, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Dynamis Er:YAG laser is intended for surgical incision/excision, cutting ablation, vaporization and coagulation of soft and hard tissue. All soft tissue is included, such as skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs, and glands. Dermatology and Plastic Surgery Indications: Epidermal nevi, actinic cheilitis, verrucae, skin tags, keratoses and soft tissue resurfacing; Soft tissue resurfacing with S22 and S22-T scanner; ENT Surgery Indications: ENT lesions, cysts, polyps, hyperkeratosis, oral leukoplakia; Oral/Maxillofacial Indications: Oral and glossal lesions; Uvulopalatoplasty by laser resurfacing Ophtalmology Indications: Soft tissue surrounding the eye; Intra-oral soft tissue incision, excision, ablation, coagulation; General Surgery Indications: Surgical incision/excision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation is necessary; Gynecology Indications: Herpes simplex, endometrial adhesion, CIN (Cervical intraepithelial neoplasia), cysts, condiloma; Genitourinary Indications: lesions of the external genitalia, urethra and anus, penis, scrotum and urethra, vulvar lesions, polyps and familial polyps of the colon; Podiatry Indications: Warts, plantar verrucae, large mosaic verrucae, matrixectomy; The Fotona F-22 Handpiece is intended for: In fractionated mode: Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece; In non-fractionated mode: General Surgery Indications: Surgical incision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation is necessary; The Fotona FS-01 and X-Restart Handpieces are intended for: Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece; The Dynamis Nd:YAG laser is intended for incision, vaporization coagulation and hemostasis of vascular lesions and soft tissue in various dermatological and surgical areas, and for permanent reduction of unwanted hair in Fitzpatrick skin types I - VI. Surgical incison, excision, vaporization, ablation and coagulation of soft tissue. All soft tissue is included, striated and smooth tissue, muscle, cartilage, mucous membrane, lymph vessels and nodes, organs and glands, fibroma removal, frenectomy and frenotomy; Treatment of Aphthous Ulcers; Excision and Vaporization of Herpex Simplex I and II; Laser assisted uvulopalatoplasty (LAUP); Laser assisted lipolysis; Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number if hair regrowing when measured at 6, 9 and 12 months after the completion of a treament regime; Treatment of wrinkles; Treatment of wrinkles with S11 scanner; Treatment of mild to moderate inflammatory acne vulgaris; Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins; Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including: Matrixectomy; Radical nail excision; Periungual and subungual warts; Plantar warts; Neuromas; Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.); Endo Venous Laser Therapy of superficial incompetent tributary veins associated with varicose veins and varicosities. Nd:YAG laser (1064 nm wavelength) therapy: Temporary relief of muscle and joint pain and stiffness, arthritis pain or muscle spasm, temporary increase in local blood circulation and/or promoting relaxation of muscle.

Device Story

Dynamis Pro Family is a dual-wavelength laser system (Er:YAG 2940 nm and Nd:YAG 1064 nm) housed in a single console. System uses flashlamp-pumped solid-state crystals; red diode aiming beam (650 nm) assists targeting. Er:YAG energy delivered via articulated arm to manual or scanner handpieces; Nd:YAG energy delivered via optical fiber to manual/scanner handpieces or bare fiber. Operated by clinicians in clinical settings. Device transforms electrical power into pulsed laser energy for tissue interaction (incision, ablation, coagulation, hemostasis). Output allows precise tissue management, hair reduction, and pain relief. Clinician monitors procedure via touchscreen interface. Benefits include versatile surgical/dermatological applications, reduced recovery time for specific procedures, and non-invasive therapeutic options.

Clinical Evidence

No clinical data provided. Substantial equivalence established via bench testing, verification, and validation against safety standards including IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and ISO 14971.

Technological Characteristics

Solid-state laser system; Er:YAG (2940 nm) and Nd:YAG (1064 nm) wavelengths. Pulsed output mode. Beam delivery via articulated arm (Er:YAG) or optical fiber (Nd:YAG). Red diode aiming beam (650 nm). LCD touchscreen interface. Powered by facility electrical source. Complies with IEC 60601-1, 60601-2-22, 60825-1, and ISO 14971/10993-1.

Indications for Use

Indicated for surgical incision, excision, ablation, vaporization, coagulation, and hemostasis of soft/hard tissues across dermatology, plastic surgery, ENT, oral/maxillofacial, ophthalmology, general surgery, gynecology, genitourinary, and podiatry. Includes permanent hair reduction for Fitzpatrick skin types I-VI, treatment of acne, wrinkles, vascular/pigmented lesions, onychomycosis, and temporary pain relief. Contraindications not specified.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. September 8, 2022 Fotona d.o.o. Marko Berdajs Quality Assurance and Regulatory Affairs Stegne 7 Ljubljana, 1000 Slovenia Re: K213267 Trade/Device Name: Dynamis Pro Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 27, 2022 Received: July 27, 2022 Dear Marko Berdajs: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K213267 Device Name Dynamis Pro Family #### Indications for Use (Describe) Er: YAG laser (2940 nm wavelength): The Dynamis Er: Y AG laser is intended for surgical incision/excision, vaporization and coagulation of soft and hard tissue. All soft tissue is included, such as skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs, and glands. - Dermatology and Plastic Surgery Indications: Epidermal nevi, actinic cheilitis, verrucae, skin tags, keratoses and soft tissue resurfacing; - Soft tissue resurfacing with S22 and S22-T scanner; - ENT Surgery Indications: ENT lesions, cysts, polyps, hyperkeratosis, oral leukoplakia; - Oral/Maxillofacial Indications: Oral and glossal lesions; - Uvulopalatoplasty by laser resurfacing - Ophtalmology Indications: Soft tissue surrounding the eye ; - Intra-oral soft tissue incision, excision, ablation, coagulation; - General Surgery Indications: Surgical incision/excision, vaporization of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation is necessary; - Gynecology Indications: Herpes simplex, endometrial adhesion, CIN (Cervical intraepithelial neoplasia), cysts, condiloma: - Genitourinary Indications:lesions of the external genitalia, urethra and anus, penis, scrotum and urethra, vulvar lesions, polyps and familial polyps of the colon; - Podiatry Indications: Warts, plantar verrucae, large mosaic verrucae, matrixectomy; - The Fotona F-22 Handpiece is intended for: - In fractionated mode: - Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece; - In non-fractionated mode: - General Surgery Indications: Surgical incision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation is necessary; - The Fotona FS-01 and X-Restart Handpieces are intended for: - Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece; #### Nd:YAG laser (1064 nm wavelength): The Dynamis Nd: Y AG laser is intended for incision, ablation vaporization and hemostasis of vascular lesions and soft tissue in various dermatological areas, and for permanent reduction of unwanted hair in Fitzpatrick skin types I - VI. - Surgical incison, excision, vaporization, ablation of soft tissue. All soft tissue is included, striated and smooth tissue, muscle, cartilage, meniscus, mucous membrane, lymph vessels and nodes, organs and glands, fibroma removal, frenectomy and frenotomy; - Treatment of Aphthous Ulcers: - Excision and Vaporization of Herpex Simplex I and II; - Laser assisted uvulopaletoplasty (LAUP); {3}------------------------------------------------ - Laser assisted lipolysis; - Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned thair reduction is defined as the long-term. stable reduction in the number if hair regrowing when measured at 6.9 and 12 months after the completion of a treament regime: - Treatment of wrinkles: - Treatment of wrinkles with S11 scanner; - Treatment of mild to moderate inflammatory acne vulgaris; - Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins; - Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including: - Matrixectomy - Radical nail excision - Periungual and subungual warts - Plantar warts - Neuromas - Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.) - Endo Venous Laser Therapy of superficial incompetent tributary veins associated with varicose veins and varicosities. Nd:YAG laser (1064 nm wavelength) therapy: - Temporary relief of muscle and joint pain and stiffhess, arthritis pain or muscle spasm, temporary increase in local blood circulation and/or promoting relaxation of muscle. Type of Use (Select one or both, as applicable) |X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510k Summary # K213267 ### SUBMITTER'S INFORMATION ### Submitter: Fotona d.o.o. Stegne 7 1000 Ljubljana, Slovenia Phone: +386 1 5009 100 Fax: + 386 1 5009 200 ### Contact Person: Marko Berdajs, Quality Assurance and Regulatory Affairs Manager Phone: + 386 1 5009 100 E-mail: marko.berdajs@fotona.com ### Date: September 07, 2022 ### DEVICE INFORMATION | Device Trade Name: | <b>Dynamis Pro Family</b> | |----------------------|-------------------------------------------------------------------------------------------------| | Common name: | Medical Laser System | | Classification name: | GEX-Powered Laser Surgical Instrument, General and Plastic Surgery<br>21 CFR 878.4810, Class II | | Product Code: | GEX | ### PREDICATE DEVICE LightWalker Laser System Family (K202985) Dynamis Pro Family (K182088) ### DEVICE DESCRIPTION SUMMARY The Dynamis Pro Family is based on Er:YAG (2940 nm) and Nd:YAG (1064 nm) laser technology. The laser unit and controls are contained in a single console. Electrical power is supplied to the console by the facility's power source. The unit combines two flashlamppumped laser sources in one housing, with optical cavities containing the Er: Y AG and Nd: Y AG crystals. A red diode aiming beam (650 nm) is combined with both therapeutic laser beams. The combined therapeutic and aiming beams are guided through an articulated arm to an optical manual or scanner handpiece (in the case of the Er:YAG laser), or through an optical fiber delivery system to an optical manual or scanner handpiece (in the case of the Nd:YAG laser). {5}------------------------------------------------ Optionally, the Nd:YAG therapeutic and aiming laser beams can be guided through a fiber having a connector on the proximal end and a bare fiber on the distal end. The Dynamis Pro Family is designed to operate in single wavelength (Nd: YAG or Er.Y AG) configurations (models) and dual wavelength (Nd: Y AG and Er: Y AG) configurations (models). ### INTENDED USE/INDICATIONS FOR USE The Dynamis Pro Family and its accessories will have the same intended use as predicate device and will be marketed for the following indications for use: ### Er: YAG laser (2940 nm wavelength): The Dynamis Er:YAG laser is intended for surgical incision/excision, cutting ablation, vaporization and coagulation of soft and hard tissue. All soft tissue is included, such as skin, cutaneous tissue, subcutaneous tissue, striated and smooth tissue, muscle, cartilage meniscus, mucous membrane, lymph vessels and nodes, organs, and glands. - Dermatology and Plastic Surgery Indications: Epidermal nevi, actinic cheilitis, verrucae, skin tags, keratoses and soft tissue resurfacing; - Soft tissue resurfacing with S22 and S22-T scanner; - ENT Surgery Indications: ENT lesions, cysts, polyps, hyperkeratosis, oral leukoplakia; - Oral/Maxillofacial Indications: Oral and glossal lesions; - Uvulopalatoplasty by laser resurfacing - Ophtalmology Indications: Soft tissue surrounding the eye; - Intra-oral soft tissue incision, excision, ablation, coagulation - General Surgery Indications: Surgical incision/excision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation is necessary: - Gynecology Indications: Herpes simplex, endometrial adhesion, CIN (Cervical intraepithelial neoplasia), cysts, condiloma; - Genitourinary Indications: lesions of the external genitalia, urethra and anus, penis, scrotum and urethra, vulvar lesions, polyps and familial polyps of the colon; - Podiatry Indications: Warts, plantar verrucae, large mosaic verrucae, matrixectomy; - The Fotona F-22 Handpiece is intended for: - In fractionated mode: Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece; · In non-fractionated mode: - General Surgery Indications: - Surgical incision/excision. vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation is necessary; {6}------------------------------------------------ - The Fotona FS-01 and X-Restart Handpieces are intended for: - Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece; ### Nd:YAG laser (1064 nm wavelength): The Dynamis Nd:YAG laser is intended for incision, vaporization coagulation and hemostasis of vascular lesions and soft tissue in various dermatological and surgical areas, and for permanent reduction of unwanted hair in Fitzpatrick skin types I - VI. - Surgical incison, excision, vaporization, ablation and coagulation of soft tissue. All soft tissue is included, striated and smooth tissue, muscle, cartilage, mucous membrane, lymph vessels and nodes, organs and glands, fibroma removal, frenectomy and frenotomy; - Treatment of Aphthous Ulcers; - Excision and Vaporization of Herpex Simplex I and II; - Laser assisted uvulopalatoplasty (LAUP); - Laser assisted lipolysis; - Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number if hair regrowing when measured at 6, 9 and 12 months after the completion of a treament regime; - Treatment of wrinkles; - Treatment of wrinkles with S11 scanner; - Treatment of mild to moderate inflammatory acne vulgaris; - Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins; - Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including: - Matrixectomy - Radical nail excision - Periungual and subungual warts - Plantar warts - Neuromas - Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.) - Endo Venous Laser Therapy of superficial incompetent tributary veins associated with varicose veins and varicosities. ### Nd:YAG laser (1064 nm wavelength) therapy: - Temporary relief of muscle and joint pain and stiffness, arthritis pain or muscle spasm, temporary increase in local blood circulation and/or promoting relaxation of muscle. {7}------------------------------------------------ ### INDICATIONS FOR USE COMPARISON The indications for use from the predicate devices that are now included in the Dynamis Pro Family do not raise new types of questions regarding safety and effectiveness. New indications with the Er: YAG (2940 nm) and Nd: YAG (1064 nm) laser output have been introduced, based on previously cleared indications for the substantially equivalent predicate devices. # TECHNOLOGICAL COMPARISON A comparison of the technical specifications for the intended use of the Dynamis Pro Family with the previously cleared device is provided in Table 1. Table 1: The comparison of technical capabilities and characteristics between Dynamis Pro Family (K182088), LightWalker Laser System Family (K202985) and Dynamis Pro Family (this submission) for the Er:YAG and Nd:YAG laser wavelengths | | LightWalker Laser System<br>Family | | Dynamis Pro Family | | Dynamis Pro Family | | |--------------------|------------------------------------------------------------------|--------------------|-----------------------|-----------------------|-----------------------|-----------------------| | | (K202985) | | (K182088) | | (this submission) | | | | Er:YAG | Nd:YAG | Er:YAG | Nd:YAG | Er:YAG | Nd:YAG | | Wavelength | 2940 nm | 1064 nm | 2940 nm | 1064 nm | 2940 nm | 1064 nm | | Laser media | Solid state<br>Er:YAG | Solid state Nd:YAG | Solid state<br>Er:YAG | Solid state<br>Nd:YAG | Solid state<br>Er:YAG | Solid state<br>Nd:YAG | | Aiming beam | Laser diode 635 nm/650 nm (red) or<br>520-532 nm (green); < 1 mW | | 650 nm (red) ; < 1 mW | | 650 nm (red) ; < 1 mW | | | Output mode | Pulsed | Pulsed | Pulsed | Pulsed | Pulsed | Pulsed | | Pulse energy | Up to 1.5 J | Up to 20 J | Up to 3 J | Up to 50 J | Up to 3 J | Up to 50 J | | Pulsewidth | 0.025 - 1 ms | 0.1 - 25 ms | 0.1 - 1.5 ms | 0.1 - 50 ms | 0.1 - 1.5 ms | 0.1 - 50 ms | | Repetition<br>rate | Up to 50 Hz | Up to 100 Hz | Up to 50 Hz | Up to 100 Hz | Up to 50 Hz | Up to 100 Hz | | Power | Up to 20 W | Up to 30 W | Up to 20 W | Up to 80 W | Up to 20 W | Up to 80 W | | Beam<br>Delivery | Articulated<br>Arm | Fiber | Articulated Arm | Fiber | Articulated Arm | Fiber | | User<br>Interface | LCD Touchscreen | | LCD Touchscreen | | LCD Touchscreen | | {8}------------------------------------------------ ### NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY AND CONCLUSIONS ### Non-Clinical Summary: The Dynamis Pro Family has been evaluated via verification and validation tests and inspections for conformance to the applicable regulations and safety standards. The Dynamis Pro Family is designed, tested and will be manufactured in accordance with both, mandatory and voluntary standards: #### IEC 60601-1:2005 + A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance. #### IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests. #### IEC 60601-2-22:2007 + A1:2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment. ### IEC 60601-1-6:2010 + A1:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability. #### IEC 60601-1-9:2007 + A1:2013 Medical electrical equipment - Part 1-9: General requirements for basic safety and essential performance - Collateral standard: Requirements for environmentally conscious design. ### IEC 60825-1:2014 Safety of laser products. Part 1: Equipment classification and requirements. #### IEC 62366:2007 + A1:2014 Medical devices - Application of usability engineering to medical devices. ### IEC 62304:2006 + A1:2015 Medical device Software - software life-cycle process. #### ISO standards: #### ISO 14971:2007 Medical devices - Application of risk management to medical devices ### ISO 17664:2004 * Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical devices. #### ISO 17664:2017 (state-of-the-art edition)* Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices #### ISO 10993-1:2009 * Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. #### ISO 10993-1:2018 (state-of-the-art edition)* Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process * With fulfilling the requirements of the latest standard edition, all requirements of previous standard edition are still fulfilled. {9}------------------------------------------------ ### Clinical Summary: N/A ### Conclusions: The Dynamis Pro Family indications for use and technological characteristics do not raise new types of questions regarding safety and efficacy when compared to the predicates. Based on its technical characteristics, design, functional features, performance test data, and its indications for use as listed above, the Dynamis Pro Family is considered to be substantially equivalent to the predicate devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%