CARESCAPE ONE and CARESCAPE DOCK F0, Model Numbers MBZ323 and MFA101, CARESCAPE ECG, Model Number MKE101,CARESCAPE TEMP, Model Number MKT101, CARESCAPE PRES, Model Number MKP101, CARESCAPE SPO2, Model Numbers MKS101 and MKS201
K213234 · Ge Medical Systems Information Technologies, Inc. · MHX · Apr 12, 2022 · Cardiovascular
Device Facts
Record ID
K213234
Device Name
CARESCAPE ONE and CARESCAPE DOCK F0, Model Numbers MBZ323 and MFA101, CARESCAPE ECG, Model Number MKE101,CARESCAPE TEMP, Model Number MKT101, CARESCAPE PRES, Model Number MKP101, CARESCAPE SPO2, Model Numbers MKS101 and MKS201
Applicant
Ge Medical Systems Information Technologies, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Apr 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
CARESCAPE ONE is both a multi-parameter physiological patient monitor and an accessory to a multi-parameter patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility. CARESCAPE ONE is indicated for the monitoring of hemodynamic and respiratory physiological parameters. When the CARESCAPE ONE is operated as a standalone multi-parameter physiological patient monitor, it provides the following physiological parameters: - ECG (heart rate, ST segment, and arrhythmia detection) - Pulse oximetry (pulse rate, functional oxygen saturation [SpO2]) - Non-invasive blood pressure (systolic, diastolic, and mean arterial pressures) - Invasive pressure (pulse rate and systolic, diastolic, and mean pressures) - Temperature - Respiratory carbon dioxide (EtCO2, FiCO2, and respiration rate) - Impedance respiration When the CARESCAPE ONE is connected as an accessory to a compatible host monitor, it provides the following physiological parameters to the host monitor: - ECG (heart rate, ST segment, and arrhythmia detection) - Pulse oximetry (pulse rate, function [SpO2], and total hemoglobin concentration [SpHB]] - Non-invasive blood pressure (systolic, diastolic, and mean arterial pressures) - Invasive pressure (pulse rate and systolic, diastolic, and mean pressures) - Regional oxygen saturation (rSO2) - Temperature - Respiratory carbon dioxide (EtCO2, FiCO2, and respiration rate) - Impedance respiration When the CARESCAPE ONE is connected as an accessory to a compatible host monitor, visual and audible alarms, user controls, and user interface are provided on the compatible host monitor and not on CARESCAPE ONE. CARESCAPE ONE is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. Regional oxygen saturation (rSO2) is an adjunct parameter for nonitoring of cerebral/somatic regional oximetry of blood in the brain or other tissue beneath the sensor. It is intended to be used on patients greater than 40 kg (88 lbs) at risk for reduced-flow or no-flow ischemic states. CARESCAPE ONE is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in the proper use of the equipment in a professional healthcare facility. Contraindications for using CARESCAPE ONE: The CARESCAPE ONE is not intended for use within a controlled MR environment.
Device Story
CARESCAPE ONE is a multi-parameter physiological monitor for bedside use and intra-hospital transport. It functions as a standalone monitor or as an acquisition accessory to a compatible host monitor. Input data is acquired via plug-and-play Active Cable Modules (ACM) for ECG, SpO2, invasive pressure, temperature, CO2, and rSO2; non-invasive blood pressure (NIBP) is integrated into the main unit. The device processes these signals to provide waveforms, vital sign values, and alarms. When used as an accessory, the device transmits data to a host monitor, which manages the user interface and alarms. Operated by trained healthcare personnel, the device provides real-time physiological data to support clinical decision-making. Benefits include flexible, modular configuration for patient-specific monitoring needs and continuous data acquisition during transport.
Clinical Evidence
No clinical data required; substantial equivalence established through bench testing, including software verification, hardware performance, and compliance with recognized consensus standards (IEC 60601 series).
Technological Characteristics
Modular patient monitor with 7-inch TFT LCD; internal battery and AC power. Uses Active Cable Modules (ACM) for parameter acquisition. Connectivity via custom medical-grade USB ports and Ethernet. Software runs on Linux. Complies with IEC 60601-1, IEC 60601-1-2, and ISO 80601 series standards.
Indications for Use
Indicated for hemodynamic and respiratory monitoring of adult, pediatric, and neonatal patients in professional healthcare facilities and during intra-hospital transport. rSO2 monitoring indicated for patients >40 kg at risk for ischemic states. Contraindicated for use in MR environments.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 26, 2022
GE Medical Systems Information Technologies, Inc. Joel Kent Sr. Manager Regulatory Affairs 9900 Innovation Drive Wauwatosa, Wisconsin 53226
### Re: K213234
Trade/Device Name: CARESCAPE ONE and CARESCAPE DOCK F0, Model Numbers MBZ323 and MFA101, CARESCAPE ECG, Model Number MKE101,CARESCAPE TEMP, Model Number MKT101, CARESCAPE PRES, Model Number MKP101, CARESCAPE SPO2, Model Numbers MKS101 and MKS201 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: MHX, BZQ, CCK, DPS, DPZ, DQA, DQK, DRT, DSI, DSJ, DSK, DXN, FLL, MLD, MUD, QEM
Dear Joel Kent:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 12, 2022. Specifically, FDA is updating this SE Letter to add a missing product code as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Shih, OHT2: Office of Cardiovascular Devices, Jennifer.Shih(@fda.hhs.gov.
Sincerely,
# Jennifer W. Shih -S
Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 12, 2022
GE Medical Systems Information Technologies, Inc. Joel Kent Sr. Manager Regulatory Affairs 9900 Innovation Drive Wauwatosa, Wisconsin 53226
Re: K213234
Trade/Device Name: CARESCAPE ONE Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, BZQ, CCK, DPS, DPZ, DQA, DRT, DSI, DSK, DXN, FLL, MLD, MUD, OEM Dated: March 16, 2022 Received: March 17, 2022
Dear Joel Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer W.Shih -S
Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213234
Device Name CARESCAPE ONE
#### Indications for Use (Describe)
CARESCAPE ONE is both a multi-parameter physiological patient monitor and an accessory to a multi-parameter patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility. CARESCAPE ONE is indicated for the monitoring of hemodynamic and respiratory physiological parameters. When the CARESCAPE ONE is operated as a standalone multi-parameter physiological patient monitor, it provides the following physiological parameters:
- · ECG (heart rate, ST segment, and arrhythmia detection)
- · Pulse oximetry (pulse rate, functional oxygen saturation [SpO2])
- · Non-invasive blood pressure (systolic, diastolic, and mean arterial pressures)
- · Invasive pressure (pulse rate and systolic, diastolic, and mean pressures)
- · Temperature
- · Respiratory carbon dioxide (EtCO2, FiCO2, and respiration rate)
- · Impedance respiration
When the CARESCAPE ONE is connected as an accessory to a compatible host monitor, it provides the following physiological parameters to the host monitor:
- · ECG (heart rate, ST segment, and arrhythmia detection)
- · Pulse oximetry (pulse rate, function [SpO2], and total hemoglobin concentration [SpHB]]
- · Non-invasive blood pressure (systolic, diastolic, and mean arterial pressures)
- · Invasive pressure (pulse rate and systolic, diastolic, and mean pressures)
- · Regional oxygen saturation (rSO2)
- · Temperature
- · Respiratory carbon dioxide (EtCO2, FiCO2, and respiration rate)
- · Impedance respiration
When the CARESCAPE ONE is connected as an accessory to a compatible host monitor, visual and audible alarms, user controls, and user interface are provided on the compatible host monitor and not on CARESCAPE ONE.
CARESCAPE ONE is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. Regional oxygen saturation (rSO2) is an adjunct parameter for nonitoring of cerebral/somatic regional oximetry of blood in the brain or other tissue beneath the sensor. It is intended to be used on patients greater than 40 kg (88 lbs) at risk for reduced-flow or no-flow ischemic states.
CARESCAPE ONE is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in the proper use of the equipment in a professional healthcare facility.
Contraindications for using CARESCAPE ONE:
The CARESCAPE ONE is not intended for use within a controlled MR environment.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a circle. The circle is surrounded by decorative swirls or flourishes. The logo is black and white.
GE Medical Systems Information Technologies, Inc. 9900 Innovation Drive Wauwatosa, WI 53226, USA
#### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
#### Owner/Contact/Date (807.92(a)(1)):
| | Date: April 7, 2022 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------|
| Owner/Submitter: | GE Medical Systems Information Technologies, Inc.<br>9900 Innovation Drive<br>Wauwatosa, WI 53226, USA |
| Primary Contact Person: | Joel Kent<br>Director, Regulatory Affairs Strategy<br>GE Healthcare<br>Phone: 617-851-0943<br>E-mail: joel.kent@ge.com |
| Secondary Contact Person: | Monica Morrison<br>Regulatory Affairs Executive<br>GE Healthcare<br>Phone: 608-515-3077 |
Device names (807.92(a)(2)):
Trade Name: CARESCAPE ONE Common/Usual Name: Multiparameter patient monitor (monitor, physiological, patient (with arrhythmia detection or alarms)
E-mail: monica.morrison(@ge.com
{6}------------------------------------------------
Classification Names: 21 CFR 870.1025 monitor, physiological, patient (with
arrhythmia detection or alarms)
21 CFR 868.2375 monitor, breathing frequency
21 CFR 868.1400 Carbon dioxide gas analyzer
21 CFR 870.2340 Electrocardiograph
21 CFR 870.2710 oximeter, ear
21 CFR 870.2700 oximeter
21 CFR 870.1425 Programmable diagnostic computer
21 CFR 870.2300 monitor, cardiac (incl. cardiotachometer & rate alarm)
21 CFR 870.1025 detector and alarm, arrhythmia
21 CFR 870.1100 alarm, blood-pressure
21 CFR 870.1110 computers, blood-pressure
21 CFR 870.1130 system, measurement, blood-pressure, noninvasive
21 CFR 880.2910 Clinical electronic thermometer
21 CFR 870.1025 monitor, St Segment with alarm
21 CFR 870.2700 Oximeter, Tissue Saturation
21 CFR 870.2700 Cerebral Oximeter
Product Code: MHX
| Subsequent Product Codes: | BZQ |
|---------------------------|-----|
| | CCK |
| | DPS |
| | DPZ |
| | DQA |
| | DQK |
| | DRT |
| | DSI |
| | DSJ |
| | DSK |
| | DXN |
| | FLL |
| | MLD |
| | MUD |
| | QEM |
{7}------------------------------------------------
The primary predicate for this submission is K200494, Predicate Device(s) (807.92(a)(3)): CARESCAPE ONE
> Additional reference devices: K191149, CARESCAPE B650 K191323, CARESCAPE B850 K191249, CARESCAPE B450 K182868, INVOS PM7100 Patient Monitor, INVOS Adult rSO2 Sensor K110028, MASIMO RADICAL Y PULSE CO-OXIMETER
Device Description (807.92(a)(4)):
CARESCAPE ONE belongs to the CARESCAPE patient monitor family. The concept of the CARESCAPE ONE is to provide a flexible bedside monitor that can also be used during intra-hospital transport. The flexibility of the CARESCAPE ONE allows the user to configure the monitor's vital sign acquisition for only the parameters they require. This is achieved using plug and play Active Cable Modules (ACM) that connect via medical grade USB ports on the CARESCAPE ONE monitor. Note that the USB ports are not compatible with commercial USB items on the market due to a custom connector design. Each ACM is dedicated to measuring a particular vital sign, currently there are ECG/Respiration, Invasive Blood Pressure, Temperature, SpO2, CO2, and rSO2 (rSO2 monitoring is only supported when CARESCAPE ONE is operating as an acquisition device. rSO2 values are not displayed on CARESCAPE ONE). The only exception is the Non-Invasive Blood Pressure (NIBP) measurement which does not require a separate ACM since the capability to measure NIBP is built-in to the CARESCAPE ONE monitor itself.
CARESCAPE ONE provides the users the acquired display values, waveforms, alarms and status messages in compact footprint monitor that runs on an internal battery as well as AC power when connected to the docking station.
When connected to a compatible host monitor, CARESCAPE ONE operates as an acquisition device. In this mode, CARESCAPE ONE screen and user interface is effectively disabled and it transmits data received form the Active Cable Modules to the host monitor, which is responsible for managing clinical configuration settings, and displaying values, waveforms, alarms, and status messages.
{8}------------------------------------------------
#### Intended Use: (807.92(a)(5)):
#### Indications
CARESCAPE ONE is both a multi-parameter physiological patient monitor and an accessory to a multi-parameter patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility.
CARESCAPE ONE is indicated for the monitoring of hemodynamic and respiratory physiological parameters.
When the CARESCAPE ONE is operated as a standalone multiparameter physiological patient monitor, it provides the following physiological parameters:
·ECG (heart rate, ST segment, and arrhythmia detection)
·Pulse oximetry (pulse rate, functional oxygen saturation [SpO2])
·Non-invasive blood pressure (systolic, diastolic, and mean arterial pressures)
·Invasive pressure (pulse rate and systolic, diastolic, and mean pressures)
·Temperature
•Respiratory carbon dioxide (EtCO2, FiCO2, and respiration rate)
·Impedance respiration
When the CARESCAPE ONE is connected as an accessory to a compatible host monitor, it provides the following physiological parameters to the host monitor:
·ECG (heart rate, ST segment, and arrhythmia detection)
•Pulse oximetry (pulse rate, functional oxygen saturation [SpO2], and total hemoglobin concentration [SpHb])
•Non-invasive blood pressure (systolic, diastolic, and mean arterial pressures)
•Invasive pressure (pulse rate and systolic, diastolic, and mean pressures)
·Regional oxygen saturation (rSO2)
·Temperature
•Respiratory carbon dioxide (EtCO2, FiCO2, and respiration rate)
·Impedance respiration
When the CARESCAPE ONE is connected as an accessory to a compatible host monitor, visual and audible alarms, user controls, and user interface are provided on the compatible host monitor and not on CARESCAPE ONE.
CARESCAPE ONE is intended for use on adult, pediatric, and
{9}------------------------------------------------
neonatal patients and on one patient at a time. Regional oxygen saturation (rSO2) is an adjunct parameter for noninvasive monitoring of cerebral/somatic regional oximetry of blood in the brain or other tissue beneath the sensor. It is intended to be used on patients greater than 40 kg (88 lbs) at risk for reduced-flow or no-flow ischemic states. CARESCAPE ONE is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in the proper use of the equipment in a professional healthcare facility. Contraindications for using CARESCAPE ONE: The CARESCAPE ONE is not intended for use within a controlled MR environment. Technology (807.92(a)(6)): The CARESCAPE ONE uses the same fundamental technology and functionality as the predicate CARESCAPE ONE (K200494). The main differences with respect to the predicate discussed throughout the submission are summarized below: - Compatibility with CARESCAPE Bx50 host monitors running the latest version of software - Compatibility with new OEM CARESCAPE Parameter Modules - Extended support for Masimo parameters - Updates to the Indications for Use (We have revised the Indications for Use statement from the predicate to reflect the changes that have been made for clarity and to reflect the inclusion of the acquisition of OEM parameter rSO2 with the CARESCAPE rSO2 – INVOS™ Parameter module and extended functionality with CARESCAPE SpO2 - Masimo measurement device - Continuous improvements in Cybersecurity - Workflow enhancements in the clinical software - Workflow enhancements in the monitor service and installation features - Updates to the list of accessories - Updates to the user documentation The CARESCAPE ONE remains substantially equivalent to the predicates and the device itself (software and hardware) remains nearly identical to the version cleared under K200494. There are no significant changes to the CARESCAPE ONE monitored parameters, and all measurement algorithms are identical to those that were included in K200494.
{10}------------------------------------------------
A summary of the main changes compared to the predicate are listed below in the comparison table.
# Product Comparison versus Predicate Main features
| Specification | PREDICATE CARESCAPE<br>ONE (K190008) | CARESCAPE ONE | Discussion of Differences |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The CARESCAPE ONE is<br>both a multi-parameter<br>physiological patient monitor<br>and an accessory to a multi-<br>parameter patient monitor<br>intended for use in multiple<br>areas and intra-hospital<br>transport within a professional<br>healthcare facility. | CARESCAPE ONE is both a<br>multi-parameter physiological<br>patient monitor and an<br>accessory to a multi-parameter<br>patient monitor intended for<br>use in multiple areas and<br>intra-hospital transport within<br>a professional healthcare<br>facility. | Equivalent<br>Added functionality to pass total<br>hemoglobin concentration [SpHb]<br>from CARESCAPE SpO2 – Masimo<br>(previously cleared K110028) and<br>regional oxygen saturation (rSO2)<br>from CARESCAPE rSO2 – INVOS |
| | The CARESCAPE ONE is<br>indicated for the monitoring of<br>hemodynamic (including<br>ECG, ST segment, arrhythmia<br>detection, invasive pressure,<br>non-invasive blood pressure, | CARESCAPE ONE is<br>indicated for the monitoring<br>of hemodynamic and<br>respiratory physiological<br>parameters. | (previously cleared K182868) through<br>to a host monitor. These parameters<br>are not displayed on the CARESCAPE<br>ONE. |
| | SpO2, pulse rate, and<br>temperature), and respiratory<br>(impedance respiration and<br>CO2 airway gas) physiological<br>parameters. | When the CARESCAPE<br>ONE is operated as a<br>standalone multi-parameter<br>physiological patient<br>monitor, it provides the | There were some minor text edits<br>made to improve clarity.<br>This change is substantially equivalent |
| | The CARESCAPE ONE can<br>be used as a standalone<br>monitor. In this mode of | following physiological<br>parameters: | to the predicate device. |
| | operation, the CARESCAPE<br>ONE provides ECG, ST<br>segment, arrhythmia detection, | • ECG (heart rate, ST<br>segment, and arrhythmia<br>detection) | |
| | invasive pressure, non-<br>invasive blood pressure, SpO2,<br>pulse rate, temperature,<br>impedance respiration, and<br>CO2 airway gas parameter | • Pulse oximetry (pulse rate,<br>functional oxygen saturation<br>[SpO2]) | |
| | acquisition and monitoring.<br>The CARESCAPE ONE can | •Non-invasive blood<br>pressure (systolic, diastolic,<br>and mean arterial | |
| | be connected as an accessory<br>to a compatible CARESCAPE | pressures) | |
| | monitor. In this mode of<br>operation, the CARESCAPE<br>ONE provides ECG, ST | • Invasive pressure (pulse<br>rate and systolic, diastolic,<br>and mean pressures) | |
| | segment, arrhythmia detection,<br>invasive pressure, non- | •Temperature | |
| | invasive blood pressure, SpO2,<br>pulse rate, temperature,<br>impedance respiration, and<br>CO2 airway gas parameter<br>acquisition. | •Respiratory carbon dioxide<br>(EtCO2, FiCO2, and<br>respiration rate)<br>•Impedance respiration | |
| | Visual and audible alarms,<br>user controls, and user<br>interface on the CARESCAPE<br>ONE are not active in this<br>mode. | When the CARESCAPE<br>ONE is connected as an<br>accessory to a compatible<br>host monitor, it provides the | |
| Specification | PREDICATE CARESCAPE<br>ONE (K190008) | CARESCAPE ONE | Discussion of Differences |
| | The CARESCAPE ONE is<br>indicated for use on adult,<br>pediatric, and neonatal patients<br>and on one patient at a time.<br><br>The CARESCAPE ONE is<br>indicated for use under the<br>direct supervision of a licensed<br>healthcare practitioner, or by<br>personnel trained in the proper<br>use of the equipment in a<br>professional healthcare<br>facility.<br><br>Contraindications for using | following physiological<br>parameters to the host<br>monitor:<br>•ECG (heart rate, ST<br>segment, and arrhythmia<br>detection)<br><br>•Pulse oximetry (pulse rate,<br>functional oxygen saturation<br>[SpO2], and total<br>hemoglobin concentration<br>[SpHb])<br><br>•Non-invasive blood | |
| | CARESCAPE ONE:<br>The CARESCAPE ONE is not<br>intended for use within a<br>controlled MR environment. | pressure (systolic, diastolic,<br>and mean arterial<br>pressures) | |
| | | •Invasive pressure (pulse<br>rate and systolic, diastolic,<br>and mean pressures) | |
| | | •Regional oxygen saturation<br>(rSO2) | |
| | | •Temperature | |
| | | •Respiratory carbon dioxide<br>(EtCO2, FiCO2, and<br>respiration rate) | |
| | | •Impedance respiration | |
| | | When the CARESCAPE<br>ONE is connected as an<br>accessory to a compatible<br>host monitor, visual and<br>audible alarms, user<br>controls, and user interface<br>are provided on the<br>compatible host monitor<br>and not on CARESCAPE<br>ONE. | |
| | | CARESCAPE ONE is<br>indicated for use on adult,<br>pediatric, and neonatal<br>patients and on one patient at<br>a time. | |
| | | Regional oxygen saturation<br>(rSO2) is an adjunct<br>parameter for noninvasive<br>monitoring of<br>cerebral/somatic regional<br>oximetry of blood in the<br>brain or other tissue<br>beneath the sensor. It is | |
| Specification | PREDICATE CARESCAPE<br>ONE (K190008) | CARESCAPE ONE | Discussion of Differences |
| | | intended to be used on<br>patients greater than 40 kg<br>(88 lbs) at risk for reduced-<br>flow or no-flow ischemic<br>states. | |
| | | CARESCAPE ONE is<br>intended for use under the<br>direct supervision of a<br>licensed healthcare<br>practitioner, or by personnel<br>trained in the proper use of the<br>equipment in a professional<br>healthcare facility. | |
| | | Contraindications for using<br>CARESCAPE ONE:…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.